Open Access 01.12.2019 | Research article

Ultrasound measurements versus invasive intracranial pressure measurement method in patients with brain injury: a retrospective study

verfasst von: Jing Zhou, Jing Li, Tiantian Ye, Yanrong Zeng

Erschienen in: BMC Medical Imaging | Ausgabe 1/2019

Abstract

Background

The invasive method for intracranial pressure measurement is ‘gold standard’ but not always feasible because the intraventricular catheter/ intraparenchymal micro transducer used in the measurement of intracranial pressure measurement may cause complications. Imaging modalities with clinical examination protocol have a lack of specificity and accuracy. The objective of the study was to compare the accuracy of diagnostic parameters of ultrasound measurements in patients with brain injury underwent invasive intracranial pressure measurement method.

Methods

Data of invasive intracranial pressure measurement method and ultrasound measurements of 185 patients with brain injury who required admission diagnosis were included in the analysis. Pearson correlation was tested for diagnostic parameters. Logistical regression analysis was performed for diagnostic parameters of death patients to evaluate independent parameter of mortality.

Results

Straight sinus flow velocities, middle cerebral artery flow velocities, and optic nerve sheath diameter were correlated with intracranial pressure (p < 0.0001 for all). Arterial blood pressure (p = 0.127) and middle cerebral artery pulsatility index (p = 0.06) were not correlated with intracranial pressure. A total of 47 patients died during the study period. Intracranial pressure (p = 0.015) and optic nerve sheath diameter (p = 0.035) were found to be independent predictor of mortality.

Conclusions

Ultrasound measurement especially optic nerve sheath diameter can be successfully used instead of invasive intracranial pressure measurement method in patients with brain injury.

Level of evidence

III.
Hinweise

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Abkürzungen
FVd
Diastolic flow velocity
FVm
Mean flow velocity
FVs
Systolic flow velocity
STROBE
Strengthening the reporting of observational studies in epidemiology

Background

Intracranial hypertension is a deadly complication and one of the contributing factor in secondary brain injury because its intensity and duration are associated with a fatal outcome [1].
Imaging modalities with clinical examination protocol have a lack of specificity [2] and insufficient accuracy [3]. Therefore, the invasive intracranial pressure measurement method is recommended [4] but a multicenter, controlled trial on patients with brain injury comparing care focused on imaging and clinical examination with care focused on maintaining intracranial pressure at 20 mmHg or less has found no significant differences [2].
The use of an intraventricular catheter/ intraparenchymal micro transducer may cause complications like infection or hemorrhage [5]. Therefore, the invasive intracranial pressure measurement method is only used in severe traumatic brain injury when systemic complications are absent [6]. When invasive intracranial pressure measurement method is contraindicated, imaging modalities with clinical examination would be preferred [4]. A prospective, observational study on patients with traumatic brain injuries recommended ocular sonography as a screening test for raised intracranial pressure measurements [7]. A retrospective study suggested transcranial Doppler for measurement of intracranial pressure [8]. Two-depth transcranial Doppler has better diagnostic reliability for intracranial pressure measurements [9, 10].
The primary aim of the retrospective study was to compare the accuracy of diagnostic parameters of ultrasound measurements in patients with brain injury underwent invasive intracranial pressure measurement method. The secondary endpoint of the study was to evaluate independent diagnostic parameters for mortality of patients in a Chinese setting.

Methods

Inclusion criteria

Patients age 18 years or older with severe traumatic brain injury, intraparenchymal hemorrhage, aneurysmal subarachnoid hemorrhage, or stroke (decision of neurosurgeon) were included in the analysis. Among these patients, who required sedation, intracranial pressure monitoring or mechanical ventilation with an admission diagnosis were included in the analysis.

Exclusion criteria

Patients were excluded from the analysis who had optic nerve injuries, skull base fracture (with cerebrospinal fluid leaks), an ocular pathology, inaccessible ultrasound windows, and no radiological or clinical susceptibility of cerebral venous vasospasm or thrombosis. Patients whose intracranial pressure mean values changed > 2 mmHg during ultrasound measurements and who had a history of glaucoma were not included in the analysis.

Clinical management of patients

Patients were put on propofol (a continuous infusion) and the muscle relaxant (atracurium) was given when necessary. Mechanical ventilation was adjusted to maintain normocapnia (partial pressure of carbon dioxide: 38 ± 2 mmHg) and adequate oxygenation (peripheral oxygen saturation ≥ 91%). The adequate cerebral perfusion pressure had been maintained ≥61 mmHg by norepinephrine, vasopressors, and normal saline [11]. As per institutional guidelines, intracranial hypertension had been treated by optimizing arterial blood pressure and using hyperosmolar fluids and sedation infusion.

Invasive intracranial pressure measurement method

Intracranial pressure had been measured by intensive care physician (minimum 10-years of experience) of institutes using 1.3 mm Tunnelling intraparenchymal probe (3PN, Spiegelberg GmbH & Co. KG, Hamburg, Germany) or a catheter inserted into the ventricles of brain and was connected to an external drainage and pressure transducer and system (Camino®, Integra® Life Sciences Corporation, Plainsboro, NJ, USA). Admission Glasgow Coma Scale, demographical characteristics, comorbidities, severity and mechanism of brain injuries, and discharge Glasgow Outcome Scale were recorded for each patient [12].

Ultrasound measurements

Ultrasound measurements had been performed by ultrasound equipment (GE Healthcare, Beijing, China) by ultra-sonographers of the institutes (blinded regarding the results of the invasive intracranial pressure measurement method and minimum 10-years of experience). The mean values of five individual measurements were considered in the analysis. Venous transcranial Doppler measurements, optic nerve sheath diameter, and arterial transcranial doppler had been measured twice in a day from day 1–5 at 50–80 mm optic nerve blood pressure (as accessible ultrasound windows).

Venous transcranial Doppler measurements

Arterial blood pressure, cerebral perfusion pressure, the partial pressure of carbon dioxide, diastolic (FVd), mean (FVm), and systolic (FVs) middle cerebral artery flow velocities, FVd, FVm, and FVs straight sinus flow velocities were recorded by 2 MHz linear probes (GE Healthcare, Beijing, China) on the straight sinus through an occipital and transforaminal bone window [5].

Optic nerve sheath diameter measurements

Optic nerve sheath diameter was recorded by 7.5 MHz linear probes (GE Healthcare, Beijing, China) with the lowest possible acoustic power. The gel was applied on the surface of eyelids. The probe was oriented 300 on both closed eyes of the patients with the head elevated to 300 in the supine position at approximately 300 on the length of the plane and at the horizontal surface. The measurements of axial and sagittal planes of both eyes had been made such that the visible widest diameter behind the retina was 2.8 mm (Fig. 1) [13].

Arterial transcranial Doppler measurement

2 MHz linear probe (GE Healthcare, Beijing, China) was used to measure arterial transcranial Doppler measurement in the same supine position of the patients. It was performed bilaterally on middle cerebral artery using temporal window [14].
Events of mortality were collected from institutional records.

Statistical analysis

InStat GraphPad Software, San Diego, IL, USA was used for statistical analysis. Pearson correlation was tested for diagnostic parameters. Logistical regression analysis was performed for diagnostic parameters among the data of death patients. Inter-rater reliability was evaluated by intraclass correlation coefficient (> 0.8: optimal, 0.7–0.8: strong, 0.5–0.69: moderate, 0.3–0.49: fair, < 0.3: poor), calculated by a two-way random effects model in Stata 12.1 software (StataCorp, College Station, TX, USA) [15]. The results were considered significant at 95% of confidence level.

Results

Enrollment

Data of patients who required intracranial pressure monitoring or mechanical ventilation with an admission diagnosis from 12 February 2017 to 1 January 2019 of the Beijing Luhe Hospital, Affiliated to Capital Medical University, Beijing, China, and referring hospitals were reviewed. Among the available patients’ records, 15 patients had optic nerve injuries, 41 had skull base fracture, 13 had a known history of ocular pathology, 40 had inaccessible ultrasound windows, 11 had no radiological susceptibility of cerebral venous thrombosis, and 12 had no clinical susceptibility of cerebral thrombosis. Therefore, these data were excluded from the analysis. Also, 8 patients were on treatment that might affect cerebrospinal fluid pressure, so excluded from the analysis. Even, 15 patients had reported more than ±2 mmHg changes in mean values of intracranial pressure during ultrasound measurements. Therefore, these data were also excluded from the analysis. Data of invasive intracranial pressure measurement method and ultrasound measurements of 185 patients were included in the analysis. Flow chart of the study is presented in Fig. 2.

Characteristics of patients

The minimum admission Glasgow Coma Scale of the enrolled patients was 4. The other demographical and clinical characteristics of enrolled patients are presented in Table 1.
Table 1
Demographical and clinical characteristics of enrolled patients
Characteristics
Value
Data of patients included in the analysis
185
Gender
Male
86 (47)
Female
99 (53)
Age (years)
Minimum
38
Maximum
68
Mean ± SD
52.42 ± 4.53
Weight (kg)
58.42 ± 4.59
Height (cm)
161 ± 6
Pathology
 Traumatic brain injury
135 (73)
 Intraparenchymal hemorrhage
38 (21)
 Aneurysmal subarachnoid hemorrhage
12 (6)
Comorbidities
 Asthma
19 (10)
 Hypertension
18 (10)
 Depression
35 (19)
 Alcoholic
8 (4)
 Smokers
15 (8)
 History of myocardial infarction
3 (2)
 Admission Glasgow Coma Scale
Maximum
14
Minimum
4
Median
9
 Discharge Glasgow Outcome Scale
Maximum
5
Minimum
1
Median
3
 Chest infection
29 (16)
 Post-traumatic acute respiratory distress syndrome
6 (3)
 Sepsis
5 (3)
 Ventriculitis
2 (1)
Categorial variables are represented as number (percentage) and continuous variables are represented as mean ± SD or median

Diagnostic parameters

Straight sinus flow velocities, middle cerebral artery flow velocities, and optic nerve sheath diameter were correlated with intracranial pressure (p < 0.0001 for all). Arterial blood pressure (p = 0.127) and middle cerebral artery pulsatility index (p = 0.06) were not correlated with intracranial pressure (Table 2). Inter-rater reliability had strong agreement (intraclass correlation coefficient = 0.71) among ultra-sonographers.
Table 2
Diagnostic parameters
Characteristics
Value
r (Pearson correlation coefficient)
Significant range of ‘r’
p-value
Data of patients included in the analysis
185
Intracranial pressure (mmHg)
Minimum
5
Reference
Reference
Reference
Maximum
17
Mean ± SD
10.24 ± 3.51
aArterial blood pressure (mmHg)
Minimum
84
0.1125
0.2527–0.3233
0.127
Maximum
101
Mean ± SD
90.61 ± 5.29
Cerebral perfusion pressure (mmHg)
Minimum
71
0.9527
0.9243–0.957
< 0.0001
Maximum
88
Mean ± SD
78.13 ± 4.58
Partial pressure of carbon dioxide (mmHg)
Minimum
4.85
0.9027
0.872–0.9264
< 0.0001
Maximum
5.95
Mean ± SD
5.24 ± 0.27
Straight sinus systolic flow velocity (cm/s)
Minimum
23
0.8662
0.825–0.8982
< 0.0001
Maximum
40
Mean ± SD
31.11 ± 4.9
Straight sinus diastolic flow velocity (cm/s)
Minimum
12
0.5538
0.4451–0.6465
< 0.0001
Maximum
18
Mean ± SD
13.47 ± 1.31
Straight sinus means flow velocity (cm/s)
Minimum
15
0.9298
0.9072–0.947
< 0.0001
Maximum
25
Mean ± SD
20.16 ± 3.23
Middle cerebral artery systolic flow velocity (cm/s)
Minimum
95
0.8724
0.833–0.9031
< 0.0001
Maximum
115
Mean ± SD
103.27 ± 5.64
Middle cerebral artery diastolic flow velocity (cm/s)
Minimum
40
0.7681
0.7016–0.8214
< 0.0001
Maximum
55
Mean ± SD
47.38 ± 4.26
Middle cerebral artery means flow velocity (cm/s)
Minimum
60
0.9
0.8685–0.9243
< 0.0001
Maximum
76
Mean ± SD
69.1 ± 3.81
Optic nerve sheath diameter (mm)
Minimum
4
0.8717
0.832–0.9025
< 0.0001
Maximum
6
Mean ± SD
4.77 ± 0.43
aMiddle cerebral artery pulsatility index
Minimum
0.85
0.1423
0.2134–0.2809
0.06
Maximum
1.05
Mean ± SD
0.93 ± 0.08
Variables are represented as mean ± SD
Pearson correlation was performed with intracranial pressure
A p value less than 0.05 was considered significant
aInsignificant correlation with intracranial pressure

Prediction of mortality

From institutional records, it was found that 34 patients died during hospitalization and 13 patients died during the follow-up period. No significant association was observed between intracranial pressure and neuropsychological and functional outcomes during hospitalization and follow-up period for survived patients. Multivariate analysis was performed among diagnostic data of death patients. Intracranial pressure (p = 0.015) and optic nerve sheath diameter (p = 0.035) were found to be independent predictor of mortality (Table 3).
Table 3
Multivariate analysis for prediction of mortality
Characteristics
p-value
Data of patients included in the analysis
47
Intracranial pressurea
0.015
Arterial blood pressure
0.12
Cerebral perfusion pressure
0.072
Partial pressure of carbon dioxide
0.085
Straight sinus systolic flow velocity
0.062
Straight sinus diastolic flow velocity
0.067
Straight sinus means flow velocity
0.068
Middle cerebral artery systolic flow velocity
0.078
Middle cerebral artery diastolic flow velocity
0.086
Middle cerebral artery means flow velocity
0.082
Optic nerve sheath diametera
0.035
Middle cerebral artery pulsatility index
0.095
A p value less than 0.05 was considered significant
Data of survived patients were considered as the reference standard
aSignificant predictor of mortality

Discussion

Intracranial pressure measurement

In the study, all the enrolled patients required admission diagnosis and subjected to measure intracranial pressure. Also, elevated intracranial pressure was found a significant parameter of death (p = 0.015) and no association between intracranial pressure and neuropsychological outcome during follow-up period among survivors. The results of the study were consistent with prospective cohort studies [1, 5] and a multicenter, controlled trial [2]. The pathophysiological events and elevated intracranial pressure can have effects on neuropsychological outcome in patients with brain injury [16]. Intracranial pressure is an independent predictor of death in patients with brain injury.

Venous transcranial Doppler measurements

Straight sinus flow velocities and middle cerebral artery flow velocities had a correlation with intracranial pressure, those were measured invasively but failed to predict mortality. The results of the study were consistent with the clinical study [2, 17] and with prospective observational study [5]. Increasing intracranial pressure leads to venous hemodynamic changes but strongly in the low-pressure venous compartment only [5]. Transcranial insonation difficulties and anatomical variations in cerebral veins were responsible for the poor performance of straight sinus and middle cerebral artery flow velocities to predict intracranial pressure [5]. Venous transcranial Doppler measurements are poorly developed method of diagnosis.

Optic nerve sheath diameter measurements

Optic nerve sheath diameter was found an individual parameter of death (p = 0.035) as well as had correlation (p < 0.0001) with intracranial pressure, those were measured invasively. The results of the study were consistent with prospective observational studies [5, 7, 9, 18] and a multicenter, controlled trial [2]. The optic nerve is surrounded by cerebrospinal fluid [7]. If cerebrospinal fluid pressure increases, there will be enlargement of the optic nerve sheath [5]. Even, measurement of optic nerve sheath diameter is a simple, accurate, safe, and quick method of diagnosis. Clinicians would prefer the measurements of optic nerve sheath diameter in their intensive care units for patients with brain injury.

Arterial transcranial doppler measurement

Middle cerebral artery pulsatility index had no correlation (p = 0.06) with intracranial pressure and was not independent parameter of death (p = 0.095). The results of the study were consistent with the clinical studies [5, 17] but not consistent with case report [19]. The possible justification is that in physiological parameters may affect pulsatility index [17] while intracranial pressure is independent parameter [20]. Even, intracranial pressure is a dynamic dimension that changes rapidly with time [7] and it is unclear whether or how pulsatility index would change. Pulsatility index is not a reliable predictor of intracranial pressure.

Ultrasound versus invasive intracranial pressure measurement

During the study, besides the invasive intracranial pressure measurement method, optic nerve sheath diameter and arterial and venous transcranial Doppler measurements were performed for all patients who required measurements of intracranial pressure. Invasive intracranial pressure measurement devices are ‘gold standard’ modalities [21] but due to the lack of neurosurgeons and/or contraindications, it is not always feasible [5, 19]. Arterial and venous transcranial ultrasound is fast and does not require specific dedicated hardware [8]. Ultrasonography device is available in the intensive care unit too [5]. Ultrasonography equipment is a hands-free tool for patients who need to measure intracranial pressure.

Limitations

Several limitations of the study have reported for examples, different types of brain injury patients were enrolled. Type of brain injury also has an effect on mortality. Effects of type of brain injury on mortality is not evaluated in multivariate analysis. There is a need for specialized training for image analysis of ultrasound. The threshold values for intracranial pressure and optic nerve sheath diameter are not well-defined for brain injuries [5]. Unlike an invasive method, ultra-sound has inter-and intra-observer variabilities.

Conclusion

Non-invasively measured straight sinus flow velocities, middle cerebral artery flow velocities, and optic nerve sheath diameter were correlated with invasively measured intracranial pressure. Optic nerve sheath diameter and intracranial pressure were independent parameters for mortality. Ultrasound measurement can be successfully used instead of invasive intracranial pressure measurement method in patients with brain injury. The large human trial is recommended to validate non-invasive techniques.

Acknowledgments

Authors are thankful to all medical and non-medical staff of the Beijing Luhe Hospital, Affiliated to Capital Medical University, Beijing, China.
The protocol of the study (BLH/CMU/CL/25/17 dated 2 February 2017) had been approved by the Beijing Luhe Hospital human ethics committee. An informed consent form had been signed by all participating patients or patient caretakers regarding imaging modalities, clinical examination, invasive intracranial pressure measurement, and anesthesia. The study had adhered to the law of China, the strengthening the reporting of observational studies in epidemiology (STROBE) statement, and Declaration of Helsinki (V2008).
An informed consent form had been signed by all participating patients or patient caretakers regarding the publication of the study in all formats of publication house including personal data of patients irrespective of time and language.

Competing interests

The authors declared that they have no competing interest or any the other competing interest regarding results and/or discussion reported in the research.
Open AccessThis article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://​creativecommons.​org/​licenses/​by/​4.​0/​), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The Creative Commons Public Domain Dedication waiver (http://​creativecommons.​org/​publicdomain/​zero/​1.​0/​) applies to the data made available in this article, unless otherwise stated.

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Literatur
1.
Zurück zum Zitat Badri S, Chen J, Barber J, Temkin NR, Dikmen SS, Chesnut RM, Deem S, Yanez ND, Treggiari MM. Mortality and long-term functional outcome associated with intracranial pressure after traumatic brain injury. Intensive Care Med. 2012;38:1800–9.CrossRefPubMed Badri S, Chen J, Barber J, Temkin NR, Dikmen SS, Chesnut RM, Deem S, Yanez ND, Treggiari MM. Mortality and long-term functional outcome associated with intracranial pressure after traumatic brain injury. Intensive Care Med. 2012;38:1800–9.CrossRefPubMed
2.
Zurück zum Zitat Chesnut RM, Temkin N, Carney N, Dikmen S, Rondina C, Videtta W, Petroni G, Lujan S, Pridgeon J, Barber J, Machamer J, Chaddock K, Celix JM, Cherner M, Hendrix T. A trial of intracranial-pressure monitoring in traumatic brain injury. N Engl J Med. 2012;367:2471–81.CrossRefPubMedPubMedCentral Chesnut RM, Temkin N, Carney N, Dikmen S, Rondina C, Videtta W, Petroni G, Lujan S, Pridgeon J, Barber J, Machamer J, Chaddock K, Celix JM, Cherner M, Hendrix T. A trial of intracranial-pressure monitoring in traumatic brain injury. N Engl J Med. 2012;367:2471–81.CrossRefPubMedPubMedCentral
3.
Zurück zum Zitat Strumwasser A, Kwan RO, Yeung L, Miraflor E, Ereso A, Castro-Moure F, Patel A, Sadjadi J, Victorino GP. Sonographic optic nerve sheath diameter as an estimate of intracranial pressure in adult trauma. J Surg Res. 2011;170:265–71.CrossRefPubMed Strumwasser A, Kwan RO, Yeung L, Miraflor E, Ereso A, Castro-Moure F, Patel A, Sadjadi J, Victorino GP. Sonographic optic nerve sheath diameter as an estimate of intracranial pressure in adult trauma. J Surg Res. 2011;170:265–71.CrossRefPubMed
4.
Zurück zum Zitat Carney N, Totten AM, O'Reilly C, Ullman JS, Hawryluk GW, Bell MJ, Bratton SL, Chesnut R, Harris OA, Kissoon N, Rubiano AM, Shutter L, Tasker RC, Vavilala MS, Wilberger J, Wright DW, Ghajar J. Guidelines for the management of severe traumatic brain injury, fourth edition. Neurosurgery. 2017;80:6–15.PubMed Carney N, Totten AM, O'Reilly C, Ullman JS, Hawryluk GW, Bell MJ, Bratton SL, Chesnut R, Harris OA, Kissoon N, Rubiano AM, Shutter L, Tasker RC, Vavilala MS, Wilberger J, Wright DW, Ghajar J. Guidelines for the management of severe traumatic brain injury, fourth edition. Neurosurgery. 2017;80:6–15.PubMed
6.
Zurück zum Zitat Brain Trauma Foundation, American Association of Neurological Surgeons, Congress of Neurological Surgeons, Joint Section on Neurotrauma and Critical Care, AANS/CNS, Bratton SL, Chestnut RM, Ghajar J, McConnell Hammond FF, Harris OA, Hartl R, Manley GT, Nemecek A, Newell DW, Rosenthal G, Schouten J, Shutter L, Timmons SD, Ullman JS, Videtta W, Wilberger JE, Wright DW. Guidelines for the management of severe traumatic brain injury. IX. Cerebral perfusion thresholds. J Neurotrauma. 2007;24:S59–64.CrossRef Brain Trauma Foundation, American Association of Neurological Surgeons, Congress of Neurological Surgeons, Joint Section on Neurotrauma and Critical Care, AANS/CNS, Bratton SL, Chestnut RM, Ghajar J, McConnell Hammond FF, Harris OA, Hartl R, Manley GT, Nemecek A, Newell DW, Rosenthal G, Schouten J, Shutter L, Timmons SD, Ullman JS, Videtta W, Wilberger JE, Wright DW. Guidelines for the management of severe traumatic brain injury. IX. Cerebral perfusion thresholds. J Neurotrauma. 2007;24:S59–64.CrossRef
7.
Zurück zum Zitat Geeraerts T, Merceron S, Benhamou D, Vigué B, Duranteau J. Non-invasive assessment of intracranial pressure using ocular sonography in neurocritical care patients. Intensive Care Med. 2008;34:2062–7.CrossRefPubMed Geeraerts T, Merceron S, Benhamou D, Vigué B, Duranteau J. Non-invasive assessment of intracranial pressure using ocular sonography in neurocritical care patients. Intensive Care Med. 2008;34:2062–7.CrossRefPubMed
8.
Zurück zum Zitat Zweifel C, Czosnyka M, Carrera E, de Riva N, Pickard JD, Smielewski P. Reliability of the blood flow velocity pulsatility index for assessment of intracranial and cerebral perfusion pressures in head-injured patients. Neurosurgery. 2012;71:853–61.CrossRefPubMed Zweifel C, Czosnyka M, Carrera E, de Riva N, Pickard JD, Smielewski P. Reliability of the blood flow velocity pulsatility index for assessment of intracranial and cerebral perfusion pressures in head-injured patients. Neurosurgery. 2012;71:853–61.CrossRefPubMed
9.
Zurück zum Zitat Ragauskas A, Bartusis L, Piper I, Zakelis R, Matijosaitis V, Petrikonis K, Rastenyte D. Improved diagnostic value of a TCD-based non-invasive ICP measurement method compared with the sonographic ONSD method for detecting elevated intracranial pressure. Neurol Res. 2014;36:607–14.CrossRefPubMed Ragauskas A, Bartusis L, Piper I, Zakelis R, Matijosaitis V, Petrikonis K, Rastenyte D. Improved diagnostic value of a TCD-based non-invasive ICP measurement method compared with the sonographic ONSD method for detecting elevated intracranial pressure. Neurol Res. 2014;36:607–14.CrossRefPubMed
10.
Zurück zum Zitat Ragauskas A, Matijosaitis V, Zakelis R, Petrikonis K, Rastenyte D, Piper I, Daubaris G. Clinical assessment of noninvasive intracranial pressure absolute value measurement method. Neurology. 2012;78:1684–91.CrossRefPubMed Ragauskas A, Matijosaitis V, Zakelis R, Petrikonis K, Rastenyte D, Piper I, Daubaris G. Clinical assessment of noninvasive intracranial pressure absolute value measurement method. Neurology. 2012;78:1684–91.CrossRefPubMed
11.
Zurück zum Zitat Steiner T, Juvela S, Unterberg A, Jung C, Forsting M, Rinkel G, European Stroke Organization. European stroke organization guidelines for the management of intracranial aneurysms and subarachnoid haemorrhage. Cerebrovasc Dis. 2013;35:93–112.CrossRefPubMed Steiner T, Juvela S, Unterberg A, Jung C, Forsting M, Rinkel G, European Stroke Organization. European stroke organization guidelines for the management of intracranial aneurysms and subarachnoid haemorrhage. Cerebrovasc Dis. 2013;35:93–112.CrossRefPubMed
12.
Zurück zum Zitat Raj R, Siironen J, Skrifvars MB, Hernesniemi J, Kivisaari R. Predicting outcome in traumatic brain injury: development of a novel computerized tomography classification system (Helsinki computerized tomography score). Neurosurgery. 2014;75:632–46.CrossRefPubMed Raj R, Siironen J, Skrifvars MB, Hernesniemi J, Kivisaari R. Predicting outcome in traumatic brain injury: development of a novel computerized tomography classification system (Helsinki computerized tomography score). Neurosurgery. 2014;75:632–46.CrossRefPubMed
13.
Zurück zum Zitat Robba C, Cardim D, Donnelly J, Bertuccio A, Bacigaluppi S, Bragazzi N, Cabella B, Liu X, Matta B, Lattuada M, Czosnyka M. Effects of pneumoperitoneum and Trendelenburg position on intracranial pressure assessed using different non-invasive method. Br J Anaesth. 2016;117:783–91.CrossRefPubMed Robba C, Cardim D, Donnelly J, Bertuccio A, Bacigaluppi S, Bragazzi N, Cabella B, Liu X, Matta B, Lattuada M, Czosnyka M. Effects of pneumoperitoneum and Trendelenburg position on intracranial pressure assessed using different non-invasive method. Br J Anaesth. 2016;117:783–91.CrossRefPubMed
14.
Zurück zum Zitat Robba C, Bragazzi NL, Bertuccio A, Cardim D, Donnelly J, Sekhon M, Lavinio A, Duane D, Burnstein R, Matta B, Bacigaluppi S, Lattuada M, Czosnyka M. Effects of prone position and positive end-expiratory pressure on noninvasive estimators of ICP: a pilot study. J Neurosurg Anesthesiol. 2017;29:243–50.CrossRefPubMed Robba C, Bragazzi NL, Bertuccio A, Cardim D, Donnelly J, Sekhon M, Lavinio A, Duane D, Burnstein R, Matta B, Bacigaluppi S, Lattuada M, Czosnyka M. Effects of prone position and positive end-expiratory pressure on noninvasive estimators of ICP: a pilot study. J Neurosurg Anesthesiol. 2017;29:243–50.CrossRefPubMed
15.
Zurück zum Zitat Oberfoell S, Murphy D, French A, Trent S, Richards D. Inter-rater reliability of sonographic optic nerve sheath diameter measurements by emergency medicine physicians. J Ultrasound Med. 2017;36:1579–84.CrossRefPubMed Oberfoell S, Murphy D, French A, Trent S, Richards D. Inter-rater reliability of sonographic optic nerve sheath diameter measurements by emergency medicine physicians. J Ultrasound Med. 2017;36:1579–84.CrossRefPubMed
16.
Zurück zum Zitat Treggiari MM, Schutz N, Yanez ND, Romand JA. Role of intracranial pressure values and patterns in predicting outcome in traumatic brain injury: a systematic review. Neurocrit Care. 2007;6:104–12.CrossRefPubMed Treggiari MM, Schutz N, Yanez ND, Romand JA. Role of intracranial pressure values and patterns in predicting outcome in traumatic brain injury: a systematic review. Neurocrit Care. 2007;6:104–12.CrossRefPubMed
17.
Zurück zum Zitat Behrens A, Lenfeldt N, Ambarki K, Malm J, Eklund A, Koskinen LO. Transcranial Doppler pulsatility index: not an accurate method to assess intracranial pressure. Neurosurgery. 2010;66:1050–7.CrossRefPubMed Behrens A, Lenfeldt N, Ambarki K, Malm J, Eklund A, Koskinen LO. Transcranial Doppler pulsatility index: not an accurate method to assess intracranial pressure. Neurosurgery. 2010;66:1050–7.CrossRefPubMed
18.
Zurück zum Zitat Komut E, Kozacı N, Sonmez BM, Yılmaz F, Komut S, Yıldırım ZN, Beydilli I, Yel C. Bedside sonographic measurement of optic nerve sheath diameter as a predictor of intracranial pressure in ED. Am J Emerg Med. 2016;34:963–7.CrossRefPubMed Komut E, Kozacı N, Sonmez BM, Yılmaz F, Komut S, Yıldırım ZN, Beydilli I, Yel C. Bedside sonographic measurement of optic nerve sheath diameter as a predictor of intracranial pressure in ED. Am J Emerg Med. 2016;34:963–7.CrossRefPubMed
19.
Zurück zum Zitat Wakerley B, Yohana K, Luen Teoh H, Tan CW, Chan BP, Sharma VK. Non-invasive intracranial pressure monitoring with transcranial Doppler in a patient with progressive cerebral venous sinus thrombosis. J Neuroimaging. 2014;24:302–4.CrossRefPubMed Wakerley B, Yohana K, Luen Teoh H, Tan CW, Chan BP, Sharma VK. Non-invasive intracranial pressure monitoring with transcranial Doppler in a patient with progressive cerebral venous sinus thrombosis. J Neuroimaging. 2014;24:302–4.CrossRefPubMed
20.
Zurück zum Zitat Frattalone AR, Stevens RD. Intracranial pressure and its surrogates. Intensive Care Med. 2011;37:1051–3.CrossRefPubMed Frattalone AR, Stevens RD. Intracranial pressure and its surrogates. Intensive Care Med. 2011;37:1051–3.CrossRefPubMed
21.
Zurück zum Zitat Dubourg J, Javouhey E, Geeraerts T, Messerer M, Kassai B. Ultrasonography of optic nerve sheath diameter for detection of raised intracranial pressure: a systematic review and meta-analysis. Intensive Care Med. 2011;37:1059–68.CrossRefPubMed Dubourg J, Javouhey E, Geeraerts T, Messerer M, Kassai B. Ultrasonography of optic nerve sheath diameter for detection of raised intracranial pressure: a systematic review and meta-analysis. Intensive Care Med. 2011;37:1059–68.CrossRefPubMed
Metadaten
Titel
Ultrasound measurements versus invasive intracranial pressure measurement method in patients with brain injury: a retrospective study
verfasst von
Jing Zhou
Jing Li
Tiantian Ye
Yanrong Zeng
Publikationsdatum
01.12.2019
Verlag
BioMed Central
Erschienen in
BMC Medical Imaging / Ausgabe 1/2019
Elektronische ISSN: 1471-2342
DOI
https://doi.org/10.1186/s12880-019-0354-0

Weitere Artikel der Ausgabe 1/2019

BMC Medical Imaging 1/2019 Zur Ausgabe

Update Radiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.