Use of stochastic resonance methods for improving laparoscopic surgery performance

verfasst von: Robert Hoskins, Jinling Wang, Caroline G. L. Cao

Erschienen in: Surgical Endoscopy | Ausgabe 10/2016

Einloggen, um Zugang zu erhalten

Abstract

Background

Vibrotactile feedback (VIB) has been utilized in previous research as sensory augmentation to improve performance during minimally invasive surgical tasks. Stochastic resonance (SR), introduced into the human control system as white noise at a subthreshold level, has shown promise to improve the sensitivity of tactile receptors resulting in performance enhancement for sensorimotor tasks. The purpose of this study was to determine whether SR could improve performance (accuracy, speed) in a simulated laparoscopic palpation task.

Methods

Sixteen subjects performed a palpation task using a laparoscopic tool to detect the presence of tumors (compacted felt) embedded in simulated tissue samples (silicone gel) inside a laparoscopic trainer box. Subjects were randomly assigned to one of the four different conditions: (1) SR, (2) VIB, (3) VIB + SR, and (4) Control. The VIB and SR signals were administered via two separate haptic actuators attached to the subjects’ dominant upper arms and forearms, respectively. All subjects were presented with 36 tissue samples with no sensory augmentation (Control) to establish baseline, followed by another 36 samples under one of the randomly assigned vibration conditions (SR, VIB, VIB + SR, or Control).

Results

Results show a significantly larger improvement in tumor detection accuracy in the SR group compared to the VIB and Control groups. There was no difference in the time to task completion, indicating that there was no speed–accuracy trade-off.

Conclusions

The results have implications for the design of instruments and methods for increasing detection accuracy such as in palpation tasks. This technology could help surgeons better identify tumors located in healthy surrounding tissue.
Literatur
1.
Zurück zum Zitat Cuschieri A, Buess G (1992) Nature and scope of endoscopic surgery. Operative manual of endoscopic surgery. Springer, Berlin Heidelberg, pp 9–13CrossRef Cuschieri A, Buess G (1992) Nature and scope of endoscopic surgery. Operative manual of endoscopic surgery. Springer, Berlin Heidelberg, pp 9–13CrossRef
2.
Zurück zum Zitat Way LW, Stewart L, Gantert W, Liu K, Lee CM, Whang K, Hunter JG (2003) Causes and prevention of laparoscopic bile duct injuries: analysis of 252 cases from a human factors and cognitive psychology perspective. Ann Surg 237(4):460–469PubMedPubMedCentral Way LW, Stewart L, Gantert W, Liu K, Lee CM, Whang K, Hunter JG (2003) Causes and prevention of laparoscopic bile duct injuries: analysis of 252 cases from a human factors and cognitive psychology perspective. Ann Surg 237(4):460–469PubMedPubMedCentral
3.
Zurück zum Zitat Picod G, Jambon AC, Vinatier D, Dubois P (2005) What can the operator actually feel when performing a laparoscopy? Surg Endosc Other Interv Tech 19(1):95–100CrossRef Picod G, Jambon AC, Vinatier D, Dubois P (2005) What can the operator actually feel when performing a laparoscopy? Surg Endosc Other Interv Tech 19(1):95–100CrossRef
4.
Zurück zum Zitat Strasberg SM (2002) Avoidance of biliary injury during laparoscopic cholecystectomy. J Hepato-Biliary-Pancreat Surg 9(5):543–547CrossRef Strasberg SM (2002) Avoidance of biliary injury during laparoscopic cholecystectomy. J Hepato-Biliary-Pancreat Surg 9(5):543–547CrossRef
5.
Zurück zum Zitat Kitagawa M, Dokko D, Okamura AM, Yuh DD (2005) Effect of sensory substitution on suture-manipulation forces for robotic surgical systems. J Thorac Cardiovasc Surg 129(1):151–158CrossRefPubMed Kitagawa M, Dokko D, Okamura AM, Yuh DD (2005) Effect of sensory substitution on suture-manipulation forces for robotic surgical systems. J Thorac Cardiovasc Surg 129(1):151–158CrossRefPubMed
6.
Zurück zum Zitat Schoonmaker RE, Cao CGL (2006) Vibrotactile force feedback system for minimally invasive surgical procedures. In: IEEE international conference on systems, man and cybernetics, 2006. SMC’06, vol 3, pp 2464–2469 Schoonmaker RE, Cao CGL (2006) Vibrotactile force feedback system for minimally invasive surgical procedures. In: IEEE international conference on systems, man and cybernetics, 2006. SMC’06, vol 3, pp 2464–2469
7.
Zurück zum Zitat Zhou M, Cao CGL (2009) Vibrotactile feedback improves laparoscopic palpation skills. In: Proceedings of the human factors and ergonomics society annual meeting, vol 53, pp 738–742 Zhou M, Cao CGL (2009) Vibrotactile feedback improves laparoscopic palpation skills. In: Proceedings of the human factors and ergonomics society annual meeting, vol 53, pp 738–742
8.
Zurück zum Zitat Bholat OS, Haluck RS, Murray WB, Gorman PJ, Krummel TM (1999) Tactile feedback is present during minimally invasive surgery. J Am Coll Surg 189(4):349–355CrossRefPubMed Bholat OS, Haluck RS, Murray WB, Gorman PJ, Krummel TM (1999) Tactile feedback is present during minimally invasive surgery. J Am Coll Surg 189(4):349–355CrossRefPubMed
9.
Zurück zum Zitat Bark K, McMahan W, Remington A, Gewirtz J, Wedmid A, Lee DI, Kuchenbecker KJ (2013) In vivo validation of a system for haptic feedback of tool vibrations in robotic surgery. Surg Endosc Other Interv Tech 27(2):656–664CrossRef Bark K, McMahan W, Remington A, Gewirtz J, Wedmid A, Lee DI, Kuchenbecker KJ (2013) In vivo validation of a system for haptic feedback of tool vibrations in robotic surgery. Surg Endosc Other Interv Tech 27(2):656–664CrossRef
10.
Zurück zum Zitat McMahan W, Romano JM, Abdul Rahuman AM, Kuchenbecker KJ (2010) High frequency acceleration feedback significantly increases the realism of haptically rendered textured surfaces. In: Haptics symposium, 2010 IEEE, pp 141–148 McMahan W, Romano JM, Abdul Rahuman AM, Kuchenbecker KJ (2010) High frequency acceleration feedback significantly increases the realism of haptically rendered textured surfaces. In: Haptics symposium, 2010 IEEE, pp 141–148
11.
Zurück zum Zitat McDonnell MD, Abbott D (2009) What is stochastic resonance? definitions, misconceptions, debates, and its relevance to biology. PLoS Comput Biol 5(5):1–9CrossRef McDonnell MD, Abbott D (2009) What is stochastic resonance? definitions, misconceptions, debates, and its relevance to biology. PLoS Comput Biol 5(5):1–9CrossRef
12.
Zurück zum Zitat Kurita Y, Shinohara M, Ueda J (2011) Wearable sensorimotor enhancer for a fingertip based on stochastic resonance. In: 2011 IEEE international conference on robotics and automation (ICRA), pp 3790–3795 Kurita Y, Shinohara M, Ueda J (2011) Wearable sensorimotor enhancer for a fingertip based on stochastic resonance. In: 2011 IEEE international conference on robotics and automation (ICRA), pp 3790–3795
13.
Zurück zum Zitat Mendez-Balbuena I, Manjarrez E, Schulte-Moenting J, Huethe F, Tapia JA, Hepp-Reymond M-C, Kristeva R (2012) Improved sensorimotor performance via stochastic resonance. J Neurosci 32(36):12612–12618CrossRefPubMed Mendez-Balbuena I, Manjarrez E, Schulte-Moenting J, Huethe F, Tapia JA, Hepp-Reymond M-C, Kristeva R (2012) Improved sensorimotor performance via stochastic resonance. J Neurosci 32(36):12612–12618CrossRefPubMed
14.
Zurück zum Zitat Collins JJ, Priplata AA, Gravelle DC, Niemi J, Harry J, Lipsitz LA (2003) IEEE noise-enhanced human sensorimotor function. Eng Med Biol Mag 22(2):76–83CrossRef Collins JJ, Priplata AA, Gravelle DC, Niemi J, Harry J, Lipsitz LA (2003) IEEE noise-enhanced human sensorimotor function. Eng Med Biol Mag 22(2):76–83CrossRef
15.
Zurück zum Zitat Mulavara AP, Fiedler MJ, Kofman IS, Wood SJ, Serrador JM, Peters B, Bloomberg JJ (2011) Improving balance function using vestibular stochastic resonance: optimizing stimulus characteristics. Exp Brain Res 210(2):303–312CrossRefPubMed Mulavara AP, Fiedler MJ, Kofman IS, Wood SJ, Serrador JM, Peters B, Bloomberg JJ (2011) Improving balance function using vestibular stochastic resonance: optimizing stimulus characteristics. Exp Brain Res 210(2):303–312CrossRefPubMed
16.
Zurück zum Zitat Collins A, Blackburn JT, Olcott C, Yu B, Weinhold P (2011) The impact of stochastic resonance electrical stimulation and knee sleeve on impulsive loading and muscle co-contraction during gait in knee osteoarthritis. Clin Biomech 26(8):853–858CrossRef Collins A, Blackburn JT, Olcott C, Yu B, Weinhold P (2011) The impact of stochastic resonance electrical stimulation and knee sleeve on impulsive loading and muscle co-contraction during gait in knee osteoarthritis. Clin Biomech 26(8):853–858CrossRef
17.
Zurück zum Zitat Kurita Y, Sueda Y, Tsuji T, Hattori M, Tokunaga M, Egi H, Ueda J (2012) Improvement of tactile sensitivity by stochastic resonance: application to vibrating forceps. In: ASME 2012 5th annual dynamic systems and control conference joint with the JSME 2012 11th motion and vibration conference. pp 465–469 Kurita Y, Sueda Y, Tsuji T, Hattori M, Tokunaga M, Egi H, Ueda J (2012) Improvement of tactile sensitivity by stochastic resonance: application to vibrating forceps. In: ASME 2012 5th annual dynamic systems and control conference joint with the JSME 2012 11th motion and vibration conference. pp 465–469
18.
Zurück zum Zitat Sueda Y, Hattori M, Sawada H, Egi H, Ohdan H, Ueda J, Kurita Y (2013) Improvement of tactile sensitivity by stochastic resonance effect-Applications to surgical grasping forceps. In: Engineering in medicine and biology society (EMBC), 2013 35th annual international conference of the IEEE, pp 4601–4604 Sueda Y, Hattori M, Sawada H, Egi H, Ohdan H, Ueda J, Kurita Y (2013) Improvement of tactile sensitivity by stochastic resonance effect-Applications to surgical grasping forceps. In: Engineering in medicine and biology society (EMBC), 2013 35th annual international conference of the IEEE, pp 4601–4604
19.
Zurück zum Zitat Ai L, Liu J, Liu J (2009) Using auditory noise to enhance the fine-motor of human’s hand due to cross-modal stochastic resonance. In: 2nd International Conference on biomedical engineering and informatics, 2009. BMEI’09, pp 1–4 Ai L, Liu J, Liu J (2009) Using auditory noise to enhance the fine-motor of human’s hand due to cross-modal stochastic resonance. In: 2nd International Conference on biomedical engineering and informatics, 2009. BMEI’09, pp 1–4
20.
Zurück zum Zitat Liu W, Lipsitz LA, Montero-Odasso M, Bean J, Kerrigan DC, Collins JJ (2002) Noise-enhanced vibrotactile sensitivity in older adults, patients with stroke, and patients with diabetic neuropathy. Arch Phys Med Rehabil 83(2):171–176CrossRefPubMed Liu W, Lipsitz LA, Montero-Odasso M, Bean J, Kerrigan DC, Collins JJ (2002) Noise-enhanced vibrotactile sensitivity in older adults, patients with stroke, and patients with diabetic neuropathy. Arch Phys Med Rehabil 83(2):171–176CrossRefPubMed
21.
Zurück zum Zitat Priplata AA, Patritti BL, Niemi JB, Hughes R, Gravelle DC, Lipsitz LA, Collins JJ (2006) Noise-enhanced balance control in patients with diabetes and patients with stroke. Ann Neurol 59(1):4–12CrossRefPubMed Priplata AA, Patritti BL, Niemi JB, Hughes R, Gravelle DC, Lipsitz LA, Collins JJ (2006) Noise-enhanced balance control in patients with diabetes and patients with stroke. Ann Neurol 59(1):4–12CrossRefPubMed
Metadaten
Titel
Use of stochastic resonance methods for improving laparoscopic surgery performance
verfasst von
Robert Hoskins
Jinling Wang
Caroline G. L. Cao
Publikationsdatum
31.12.2015
Verlag
Springer US
Erschienen in
Surgical Endoscopy / Ausgabe 10/2016
Print ISSN: 0930-2794
Elektronische ISSN: 1432-2218
DOI
https://doi.org/10.1007/s00464-015-4730-8

Weitere Artikel der Ausgabe 10/2016

Surgical Endoscopy 10/2016 Zur Ausgabe

Update Chirurgie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.

S3-Leitlinie „Diagnostik und Therapie des Karpaltunnelsyndroms“

CME: 2 Punkte

Prof. Dr. med. Gregor Antoniadis Das Karpaltunnelsyndrom ist die häufigste Kompressionsneuropathie peripherer Nerven. Obwohl die Anamnese mit dem nächtlichen Einschlafen der Hand (Brachialgia parästhetica nocturna) sehr typisch ist, ist eine klinisch-neurologische Untersuchung und Elektroneurografie in manchen Fällen auch eine Neurosonografie erforderlich. Im Anfangsstadium sind konservative Maßnahmen (Handgelenksschiene, Ergotherapie) empfehlenswert. Bei nicht Ansprechen der konservativen Therapie oder Auftreten von neurologischen Ausfällen ist eine Dekompression des N. medianus am Karpaltunnel indiziert.

Prof. Dr. med. Gregor Antoniadis
Berufsverband der Deutschen Chirurgie e.V.

S2e-Leitlinie „Distale Radiusfraktur“

CME: 2 Punkte

Dr. med. Benjamin Meyknecht, PD Dr. med. Oliver Pieske Das Webinar S2e-Leitlinie „Distale Radiusfraktur“ beschäftigt sich mit Fragen und Antworten zu Diagnostik und Klassifikation sowie Möglichkeiten des Ausschlusses von Zusatzverletzungen. Die Referenten erläutern, welche Frakturen konservativ behandelt werden können und wie. Das Webinar beantwortet die Frage nach aktuellen operativen Therapiekonzepten: Welcher Zugang, welches Osteosynthesematerial? Auf was muss bei der Nachbehandlung der distalen Radiusfraktur geachtet werden?

PD Dr. med. Oliver Pieske
Dr. med. Benjamin Meyknecht
Berufsverband der Deutschen Chirurgie e.V.

S1-Leitlinie „Empfehlungen zur Therapie der akuten Appendizitis bei Erwachsenen“

CME: 2 Punkte

Dr. med. Mihailo Andric
Inhalte des Webinars zur S1-Leitlinie „Empfehlungen zur Therapie der akuten Appendizitis bei Erwachsenen“ sind die Darstellung des Projektes und des Erstellungswegs zur S1-Leitlinie, die Erläuterung der klinischen Relevanz der Klassifikation EAES 2015, die wissenschaftliche Begründung der wichtigsten Empfehlungen und die Darstellung stadiengerechter Therapieoptionen.

Dr. med. Mihailo Andric
Berufsverband der Deutschen Chirurgie e.V.