18.06.2020 | Short Communication

Usefulness of a blood glucose and ketone monitoring device as a screening tool for lethal diabetic ketoacidosis

verfasst von: Pierre-Antoine Peyron, Maëlle Plawecki, Maisy Lossois, Manuela Lotierzo, Eric Baccino, Jean-Paul Cristol

Erschienen in: International Journal of Legal Medicine | Ausgabe 1/2021

Einloggen, um Zugang zu erhalten

Abstract

Diabetic ketoacidosis (DKA) is a frequent and life-threatening complication, whose diagnosis remains challenging in forensic practice. We aimed at assessing the performance of a commercially available blood glucose and ketone monitoring device (BGMD) in measuring glucose and ketone levels in post-mortem vitreous (VH) and blood samples, in order to determine if such a device can be used for screening lethal cases of DKA at autopsy. VH and blood samples were collected in cases of unexplained causes of death at autopsy. Glucose and β-hydroxybutyrate (BHB) were measured in VH and BHB in blood using the BGMD. The values were compared to those obtained with validated enzymatic methods. Values ≥ 10 mmol/L were considered to be elevated for glucose, and BHB values ≥ 2.5 mmol/L were considered to indicate ketoacidosis. There was a strong and significant correlation between VH glucose and blood BHB concentrations measured with the BGMD and the validated method (r = 0.78 and r = 0.80, p < 0.0001, respectively), whereas no correlation was found for VH BHB values (r = 0.19, p = 0.19). The sensitivity and specificity of the BGMD were both excellent (1.0) to detect elevated VH glucose levels with a threshold of 14.4 mmol/L, and to detect elevated blood BHB levels with a threshold of 2.85 mmol/L. In contrast, the specificity of the BGMD to detect high BHB levels in VH was poor (0.50) with an optimal threshold of 2.5 mmol/L. We showed that a commercially available BGMD is suitable for identifying cases of lethal DKA and other metabolic disorders at autopsy, through the investigation of vitreous glucose and blood BHB. We therefore recommend the systematic use of a BGMD for screening these conditions in cases of unexplained deaths.
Literatur
10.
Zurück zum Zitat (2019) Classification of diabetes mellitus. World Health Organization, Geneva (2019) Classification of diabetes mellitus. World Health Organization, Geneva
12.
Zurück zum Zitat Iten PX, Meier M (2000) Beta-hydroxybutyric acid--an indicator for an alcoholic ketoacidosis as cause of death in deceased alcohol abusers. J Forensic Sci 45:624–632CrossRef Iten PX, Meier M (2000) Beta-hydroxybutyric acid--an indicator for an alcoholic ketoacidosis as cause of death in deceased alcohol abusers. J Forensic Sci 45:624–632CrossRef
16.
Zurück zum Zitat Heimer J, Gascho D, Madea B, Steuer A, Martinez RM, Thali MJ, Zoelch N (2020) Comparison of the beta-hydroxybutyrate, glucose, and lactate concentrations derived from postmortem proton magnetic resonance spectroscopy and biochemical analysis for the diagnosis of fatal metabolic disorders. Int J Legal Med 134:603–612. https://doi.org/10.1007/s00414-019-02235-6CrossRefPubMed Heimer J, Gascho D, Madea B, Steuer A, Martinez RM, Thali MJ, Zoelch N (2020) Comparison of the beta-hydroxybutyrate, glucose, and lactate concentrations derived from postmortem proton magnetic resonance spectroscopy and biochemical analysis for the diagnosis of fatal metabolic disorders. Int J Legal Med 134:603–612. https://​doi.​org/​10.​1007/​s00414-019-02235-6CrossRefPubMed
17.
Zurück zum Zitat Lin LI (1989) A concordance correlation coefficient to evaluate reproducibility. Biometrics 45:255–268CrossRef Lin LI (1989) A concordance correlation coefficient to evaluate reproducibility. Biometrics 45:255–268CrossRef
18.
Zurück zum Zitat Sturner WQ, Gantner GE (1964) Postmortem vitreous glucose determinations. J Forensic Sci 9:485–491PubMed Sturner WQ, Gantner GE (1964) Postmortem vitreous glucose determinations. J Forensic Sci 9:485–491PubMed
19.
Zurück zum Zitat Gagajewski A, Murakami MM, Kloss J, Edstrom M, Hillyer M, Peterson GF, Amatuzio J, Apple FS (2004) Measurement of chemical analytes in vitreous humor: stability and precision studies. J Forensic Sci 49:371–374CrossRef Gagajewski A, Murakami MM, Kloss J, Edstrom M, Hillyer M, Peterson GF, Amatuzio J, Apple FS (2004) Measurement of chemical analytes in vitreous humor: stability and precision studies. J Forensic Sci 49:371–374CrossRef
Metadaten
Titel
Usefulness of a blood glucose and ketone monitoring device as a screening tool for lethal diabetic ketoacidosis
verfasst von
Pierre-Antoine Peyron
Maëlle Plawecki
Maisy Lossois
Manuela Lotierzo
Eric Baccino
Jean-Paul Cristol
Publikationsdatum
18.06.2020
Verlag
Springer Berlin Heidelberg
Erschienen in
International Journal of Legal Medicine / Ausgabe 1/2021
Print ISSN: 0937-9827
Elektronische ISSN: 1437-1596
DOI
https://doi.org/10.1007/s00414-020-02343-8

Weitere Artikel der Ausgabe 1/2021

International Journal of Legal Medicine 1/2021 Zur Ausgabe

Neu im Fachgebiet Rechtsmedizin