Validation of a DIXON-based fat quantification technique for the measurement of visceral fat using a CT-based reference standard
Erschienen in:
Abdominal Radiology
|
Ausgabe 1/2019
Einloggen, um Zugang zu erhalten
Abstract
Purpose
The purpose of the study is to determine whether a novel semi-automated DIXON-based fat quantification algorithm can reliably quantify visceral fat using a CT-based reference standard.
Methods
This was an IRB-approved retrospective cohort study of 27 subjects who underwent abdominopelvic CT within 7 days of proton density fat fraction (PDFF) mapping on a 1.5T MRI. Cross-sectional visceral fat area per slice (cm2) was measured in blinded fashion in each modality at intervertebral disc levels from T12 to L4. CT estimates were obtained using a previously published semi-automated computational image processing system that sums pixels with attenuation − 205 to − 51 HU. MR estimates were obtained using two novel semi-automated DIXON-based fat quantification algorithms that measure visceral fat area by spatially regularizing non-uniform fat-only signal intensity or de-speckling PDFF 2D images and summing pixels with PDFF ≥ 50%. Pearson’s correlations and Bland–Altman analyses were performed.
Results
Visceral fat area per slice ranged from 9.2 to 429.8 cm2 for MR and from 1.6 to 405.5 cm2 for CT. There was a strong correlation between CT and MR methods in measured visceral fat area across all studied vertebral body levels (r = 0.97; n = 101 observations); the least (r = 0.93) correlation was at T12. Bland–Altman analysis revealed a bias of 31.7 cm2 (95% CI [− 27.1]–90.4 cm2), indicating modestly higher visceral fat assessed by MR.
Conclusion
MR- and CT-based visceral fat quantification are highly correlated and have good cross-modality reliability, indicating that visceral fat quantification by either method can yield a stable and reliable biomarker.