Validation of post-stenting fractional flow reserve with intravascular ultrasound parameters for optimal stent deployment

Zeitschrift:
The International Journal of Cardiovascular Imaging
Autoren:
Sung Gyun Ahn, Soonchang Hong, Jung-Woo Son, Jun-Won Lee, Young Jin Youn, Min-Soo Ahn, Jang-Young Kim, Byung-Su Yoo, Seung-Hwan Lee, Junghan Yoon
Wichtige Hinweise
Sung Gyun Ahn and Soonchang Hong have contributed equally to the work.

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Abstract

We aimed to evaluate the relationship between post-stent fractional flow reserve (FFR) and intravascular ultrasound (IVUS) parameters for stent optimization. IVUS and FFR measurements were performed in 101 coronary lesions after successful percutaneous coronary intervention (PCI). Quantitative IVUS parameters for stent optimization such as minimal intra-stent cross sectional area (CSA), proximal, and distal reference lumen CSA, and plaque burden (PB) at both edges of the stent related to the post-stent FFR. Mean post-stent FFR was 0.91 ± 0.06. Post-stent FFR ≤ 0.90 was observed in 46 (45.5%) lesions. The 55 lesions with post-stent FFR > 0.9 had larger proximal CSA (9.2 ± 2.9 mm2 vs. 7.4 ± 2.3 mm2, p = 0.001), distal reference lumen CSA (7.4 ± 2.7 mm2 vs. 6.3 ± 2.1 mm2, p = 0.017), minimum intra-stent lumen CSA (7.4 ± 2.5 mm2 vs. 6.1 ± 1.7 mm2, p = 0.004), and lower proximal reference PB (45 ± 8.2% vs. 50 ± 11%, p = 0.017) than did those with post-stent FFR ≤ 0.90. Post-stent FFR had an inverse correlation with age, left anterior descending artery (LAD) against non-LAD, and proximal reference PB and a positive correlation with pre-stent FFR, proximal and distal reference lumen CSA, and minimum intra-stent lumen CSA. Meanwhile, no correlation was found between post-stent FFR and stent length. Pre-stent FFR (β = 0.544), proximal reference lumen CSA (β = 0.530) and age (β = − 0.251) were found to be independently associated with the post-stent FFR. Post-procedural FFR might partly reflect the appropriateness of PCI results in terms of the adequate expansion of the stent and full lesion coverage on IVUS.

Bitte loggen Sie sich ein, um Zugang zu diesem Inhalt zu erhalten

★ PREMIUM-INHALT
e.Med Interdisziplinär

Mit e.Med Interdisziplinär erhalten Sie Zugang zu allen CME-Fortbildungen und Fachzeitschriften auf SpringerMedizin.de. Zusätzlich können Sie eine Zeitschrift Ihrer Wahl in gedruckter Form beziehen – ohne Aufpreis.

Weitere Produktempfehlungen anzeigen
Literatur
Über diesen Artikel
  1. Sie können e.Med Innere Medizin 14 Tage kostenlos testen (keine Print-Zeitschrift enthalten). Der Test läuft automatisch und formlos aus. Es kann nur einmal getestet werden.


 

Neu im Fachgebiet Kardiologie

Mail Icon II Newsletter

Bestellen Sie unseren kostenlosen Newsletter Update Kardiologie und bleiben Sie gut informiert – ganz bequem per eMail.

Bildnachweise