Skip to main content
Erschienen in: Trials 1/2016

Open Access 01.12.2016 | Research

Varied flushing frequency and volume to prevent peripheral intravenous catheter failure: a pilot, factorial randomised controlled trial in adult medical-surgical hospital patients

verfasst von: Samantha Keogh, Julie Flynn, Nicole Marsh, Gabor Mihala, Karen Davies, Claire Rickard

Erschienen in: Trials | Ausgabe 1/2016

Abstract

Background

Research has identified high failure rates of peripheral intravenous catheter (PIVC) and varied flushing practices.

Methods

This is a single-centre, pilot, non-masked, factorial randomised controlled trial. Participants were adults, with a PIVC of expected use ≥24 hours (n = 160), admitted to general medical or surgical wards of a tertiary referral hospital in Queensland (Australia). Patients were randomly allocated to one of four flush groups using manually prepared syringes and 0.9 % sodium chloride: 10 mL or 3 mL flush, every 24 or 6 hours. The primary endpoint was PIVC failure, a composite measure of occlusion, infiltration, accidental dislodgement and phlebitis.

Results

PIVC average dwell was 3.1 days. PIVC failure rates per 1000 hours were not significantly different for the volume intervention (4.84 [3 mL] versus 7.44 [10 mL], p = 0.06, log-rank). PIVC failure rates per 1000 hours were also not significantly different for the frequency intervention (5.06 [24 hour] versus 7.34 [6 hour], p = 0.05, log-rank). Cox proportional hazard regression found neither the flushing nor frequency intervention, or their interaction (p = 0.21) to be significantly associated with PIVC failure. However, female gender (hazard ratio [HR] 2.2 [1.3–3.6], p < 0.01), insertion in hand/posterior wrist (HR 1.7 [1.0–2.7], p < 0.05) and the rate per day of PIVC access (combined flushes and medication pushes) (HR 1.2 [1.1–1.4], p < 0.01) significantly predicted PIVC failure.

Conclusion

Neither increased flushing volume nor frequency significantly altered the risk of PIVC failure. Female gender, hand/posterior wrist placement and episodes of access (flushes and medication) may be more important. Larger, definitive trials are feasible and required.

Trial registration

Australian and New Zealand Clinical Trials Registry: ACTRN12615000025​538. Registered on 19 January 2015.

Background

Peripheral intravenous catheters (PIVCs) are the most commonly used invasive devices in hospitals; they are relied upon for therapy across nearly all medical and surgical specialties [13]. Yet failure prior to completion of therapy occurs in up to 69 % of patients [412]. This may be due to a range of complications, which can be mechanical, vascular or infectious. Mechanical complications include occlusion, infiltration and dislodgement. Vascular complications include venous thrombotic occlusion and phlebitis (irritation or inflammation of the vessel wall). Infectious complications may be bacterial or fungal, and local or systemic bloodstream infections. Complications lead to device failure and device replacement, which results in interrupted therapy, pain associated with resiting and increased health care costs for resources and staff time [13]. Bloodstream infections prolong hospitalisation and increase treatment costs and mortality [14, 15].
The interactions believed to cause mechanical and vascular complications are based on the following theoretical understandings. A fibrin coating forms within the PIVC lumen and catheter tip within 24 hours of placement. Fibrin can form the basis for thrombus development, which as well as being a nidus for infection, can occlude the PIVC lumen and even the vessel [16]. Similarly, inappropriate concentrations of injected/infused solutions, or incompatible mixtures, can cause fluids or medications to precipitate within the catheter lumen and may obstruct the catheter [17].
Current practice recommendations are to flush PIVCs before and after each medication administration, and at regular intervals when PIVCs are not in use [15, 18, 19]. The theoretical purpose of flushing is to maintain catheter patency by preventing internal luminal occlusion, reducing build-up of blood and other products on the PIVC internal surface and preventing interaction of incompatible fluids/medicines [18, 19].
Flushing is a historical practice, based more on derived scientific principles and tradition than on randomised controlled trials (RCTs), and current flushing practices vary widely [20, 21]. This has implications for costs and workload [13]. A large survey of flushing practices revealed a high level of policy awareness (72 %) but varied levels of adherence. Approximately half of respondents stated that there was no medical order or documentation for the flush. Twenty-five percent of respondents used a syringe smaller than the required 10 mL. Use of prefilled syringes was limited to 10 %. The frequency of flushing varied widely from 4th hourly to never, with the most common responses being pro re nata (23 %) or 6th hourly (23 %) [20].
There are no RCTs in adult patients comparing different flushing volumes and frequencies that are in common practice today. A single-site paediatric RCT found no significant difference in overall PIVC failure with varied flushing frequencies [22]. Large RCTs in varied patient populations are urgently needed to identify the ideal flushing methods including frequency and volume. This would inform practice to [13] prevent PIVC failure, thus preventing painful complications and reinsertions, and reduce organisational costs [13]. This pilot study aimed to test the effect of both volume and frequency on PIVC failure, in preparation for future large trials.

Methods

The study was a four-arm, 2 × 2 factorial design RCT comparing the effectiveness of different flushing frequencies (more versus less) and volumes (high versus low) in maintaining the patency of PIVCs. A factorial design enables more than one clinical question to be tested from a single RCT [23]. In addition to testing the feasibility of conducting a larger trial, our hypotheses were that both increased flushing volume and frequency would decrease PIVC failure.

Study interventions

Patients were allocated to one of the four following study arms:
1.
High volume, high frequency (10 mL every 6 hours)
 
2.
High volume, low frequency (10 mL every 24 hours)
 
3.
Low volume, high frequency (3 mL every 6 hours)
 
4.
Low volume, low frequency (3 mL every 24 hours)
 
The volume and frequency parameters were derived from reported practice in a large cross-sectional survey, guidelines and literature [20]. Flushes were 0.9 % sodium chloride using syringes that were manually prepared by ward nursing staff (prefilled syringes were not used). Colour coded stickers indicating the flush prescription were placed in the regular order section of the patient’s medication chart. This order was in addition to the recommended practice of 5–10 mL pre- and post-medication flushes with 0.9 % sodium chloride. Confirmation of intravenous (IV) flush as per protocol was by the nursing signature against prescription in medication chart.

Insertion and care of PIVCs

Forty-five percent of all PIVCs were inserted by specialist nursing IV inserters. The remainder were inserted by clinical nursing or medical staff. The skin preparation used was chlorhexidine 2 % with isopropyl alcohol 70 % swabs (SOLU-IVMC/TM 3 M St Paul, MN, USA). BD Insyte™ Autoguard™ BC catheters were used with BD Connecta™ extension tubing and CareFusion SmartSite® needleless connectors (BD Medical, Franklin Lakes, NJ, USA). PIVCs were secured with standard simple polyurethane dressings. Nursing (not the research or IV team) and medical staff provided follow-up care. Catheter sites were labelled to identify study inclusion. PIVCs were replaced on clinical indication as per hospital policy, although occasionally some medical staff still requested routine replacement. The Research Nurses (RNs) visited patients daily to visually inspect the PIVC, remind staff of research protocol, gather data collection sheets (until two days after removal of the PIVC) and assess for outcome measures and adverse events. Other data were obtained by the RNs from patient charts, notes, computerised administration and pathology systems.

Outcome measures

This pilot study collected outcomes to establish feasibility of the protocol and processes. Feasibility data outcomes included the success of screening and recruitment strategies; ease of data collection processes and technology; and resources and research staff time.
The primary endpoint for both the volume and frequency hypotheses was PIVC failure, a composite of any unplanned PIVC removal prior to completion of therapy. This included: occlusion (PIVC will not infuse, or leakage occurs when fluid infused), infiltration (leakage of fluid into surrounding tissues), accidental dislodgement and phlebitis (score of 2 or more of pain/tenderness, redness, swelling, purulence and/or a palpable cord). A secondary outcome was infection (laboratory-confirmed local or bloodstream infection) [24, 25].

Setting and sample

The trial was undertaken in surgical and medical wards at a large tertiary metropolitan hospital in Brisbane, Australia, where 160 patients were recruited over four months. This number is adequate in pilot trials to represent the target population of larger RCTs for the purposes of assessing feasibility [26].
Potentially eligible patients were those aged 18 years or over, with a PIVC that had been inserted within 24 hours of recruitment, or was about to be inserted and expected to remain for >24 hours, and who gave written informed consent. Exclusion criteria included: non-English-speaking patients without an interpreter and patients receiving continuous IV therapy. Only one PIVC per patient was studied. The study received ethics committee approval from the Royal Brisbane and Women’s Hospital Human Research Ethics Committee (HREC/13/QRBW/420) and Griffith University Human Research Ethics Committee (NRS/12/14/HREC), and the trial was registered with the Australian and New Zealand Clinical Trials Registry: ACTRN12615000025538.

Recruitment, randomisation, allocation concealment and blinding

Staff education about study aims and processes was conducted prior to commencement of study on all wards. An RN screened patients daily, gained informed consent, performed randomisation and liaised with the ward nursing and medical teams. Eligible patients were invited to participate in the study and written informed consent gained. Randomisation (simple) was obtained via a centralised web-based service (https://​www151.​griffith.​edu.​au) and was generated in a 1:1:1:1 ratio for the four study groups. Allocation concealment was maintained until each patient’s study entry. A Project Manager undertook regular quality checks to ensure allocation integrity and data quality. Intravenous flush orders were not amenable to blinding of patients, clinical staff or RNs.

Data collection

The RNs collected and entered data in the clinical areas using preformatted case report forms and then entered the data into a purpose-built computer database (Research Electronic Data Capture, Vanderbilt University). All data were de-identified at this point and only identifiable within the database by specific study number. Patient characteristics collected by RNs at baseline included: age, sex, estimated weight category, diagnostic group, dominant limb, vein quality, skin integrity, co-morbidities, length of stay, immunodeficiency, current infection and intravenous therapy (including antimicrobial). PIVC characteristics collected were: device type, insertion site, PIVC gauge, side of insertion, clinical area/ward, inserter discipline, initial/subsequent PIVC and the addition of extension tubing and injection ports. The RNs visually inspected PIVCs daily and assessed for phlebitis. After PIVC removal, the following data were collected: reason for PIVC removal, dwell time, phlebitis with 48 hours of removal, hospital length of stay and hospital mortality. The number of PIVC accesses to administer medication/flushes was recorded from the patient’s medication charts. In the case of any suspected blood stream infections, clinical staff ordered blood and PIVC cultures from patients suspected as per usual practice and RNs accessed the results.

Data analysis

Data cleaning and analyses were performed in Stata (14.1, StataCorp, College Station, TX, USA). An intention-to-treat analysis framework was used. Mean values and standard deviations were reported for normally distributed data; median values and interquartile range were reported otherwise. As a pilot study, we tested our statistical comparison methods, but did not expect to find statistical differences. Incidence rates of PIVC failure per 1000 PIVC hours (95 % confidence intervals [CIs] were calculated, and Kaplan-Meier curves generated. The null hypotheses of no differences in PIVC survival with increased volume or increased frequency were tested with the log-rank test of equality. Cox proportional hazard models (univariable and multivariable) were fitted and hazard ratios (HRs) calculated, including an interaction term between volume and frequency. The interaction was confirmed with Altman and Bland’s method (results not presented) [27]. Covariates were selected for the multivariable model at p < 0.2 in the univariable model, and dropped if correlated with another covariate (at ρ > 0.5 and p < 0.05). The final multivariate model was built using manual stepwise backward removal of covariates at p < 0.05 [27]. The proportional hazards assumption was checked.

Results

Feasibility outcomes

One hundred and seventy-three patients were screened for eligibility, with 160 recruited (Fig. 1). One hundred percent of recruited patients were randomised, giving an average of 40 patients per month. Staff on study wards were cooperative and supportive of the trial. RNs spent an average of 4 hours a day (Monday to Friday) recruiting and data collecting for the trial. Only follow-up data were collected on weekends for approximately 2 hours a day when RNs covered for a range of studies. All participants received the allocated intervention with no protocol violations, and no patients were lost to follow-up.

Patients and PIVCs

At baseline, groups were similar in demographic and clinical risk profiles (Table 1). PIVCs were used for an average 3.1 (median: 2.70) days across all study groups and total 11,911 PIVC hours were studied (Table 2). The overall PIVC failure rate was 46 % or 6.13/1000 device hours, with occlusion and phlebitis the most common reasons for PIVC failure. No confirmed PIVC-related infections occurred in any group. Two suspected PIVC infections occurred; one patient had gram-negative bacilli in both peripheral blood and urine, the second had a blood culture with probable skin contaminant.
Table 1
Patient and insertion characteristics at baseline (n = 160)
Characteristic
Total (n = 160)
Flush volume
Flush frequency
 
3 mL (n = 80)
10 mL (n = 80)
24 h (n = 79)
6 h (n = 81)
Group size
160 (100)
80 (50)
80 (50)
79 (49)
81 (51)
Age (median, IQR)
63 (24)
63 (22)
64 (30)
63 (26)
64 (23)
Gender (male)
93 (58)
53 (66)
40 (50)
47 (59)
46 (57)
Overweight/obesea
70 (44)
33 (41)
37 (46)
33 (42)
37 (46)
Comorbidities (≥2)
110 (69)
51 (64)
59 (74)
57 (72)
53 (65)
Skin integrity (poor)
24 (15)
12 (15)
12 (15)
12 (15)
12 (15)
Infection (any type)
43 (27)
22 (28)
21 (26)
18 (23)
25 (31)
Antibiotic treatment
115 (72)
57 (71)
58 (72)
60 (76)
55 (68)
Frequency of IV txb:
     
- 0
15 (11)
10 (12)
5 (6)
8 (10)
7 (9)
- 1–2 times daily
23 (17)
10 (12)
13 (16)
14 (18)
9 (11)
- 3–4 times daily
71 (54)
35 (44)
36 (45)
37 (47)
34 (42)
- ≥5 times daily
23 (17)
13 (16)
10 (12)
8 (10)
15 (19)
Inserted in dominant arm
80 (52)
37 (47)
36 (49)
36 (48)
37 (47)
Insertion point:
     
- post. wrist
39 (24)
22 (28)
17 (21)
18 (23)
21 (26)
- hand
38 (24)
18 (22)
20 (25)
18 (23)
20 (25)
- post. u/l forearm
30 (19)
15 (19)
15 (19)
17 (22)
13 (16)
- ant. u/l forearm
29 (18)
15 (19)
14 (18)
14 (18)
15 (19)
- cub. fossa/ant. u arm
24 (15)
10 (12)
14 (18)
12 (15)
12 (15)
Vein quality (poor)
85 (53)
41 (51)
44 (55)
39 (49)
46 (57)
Insertion in wardc
118 (74)
59 (74)
59 (74)
59 (75)
59 (73)
Inserted by:
     
- IV nurse inserter
72 (45)
36 (45)
36 (45)
39 (49)
33 (41)
- any doctor
68 (43)
35 (44)
33 (41)
35 (44)
33 (41)
- any nurse/unknown
20 (13)
9 (11)
11 (14)
5 (6)
15 (19)
Multiple insertion attempts
32 (20)
15 (19)
17 (22)
17 (22)
15 (19)
Device sized:
     
- 22 gauge
76 (48)
37 (46)
39 (49)
35 (44)
41 (51)
- 20 gauge
67 (42)
33 (41)
34 (42)
35 (44)
32 (40)
Extension tubing
82 (51)
41 (51)
41 (51)
43 (54)
39 (48)
3-way tap
118 (74)
59 (74)
59 (74)
59 (75)
59 (73)
n (%) shown unless otherwise indicated post. posterior, ant. anterior, cub. cubital, mL millilitres, h hours, tx treatment, uupper, l lower
a Estimated
b Frequencies may not add up to group size due to missing/omitted values
c Versus all other
d Other categories omitted
Table 2
Device outcomes at removal
Outcome
Total (n = 160)
Flush volume
Flush frequency
 
3 mL (n = 80)
10 mL (n = 80)
24 h (n = 79)
6 h (n = 81)
Device failure:
73 (46)
29 (36)
44 (55)
32 (41)
41 (51)
- occlusion or leaking
22 (14)
12 (15)
10 (12)
9 (11)
13 (16)
- infiltration
18 (11)
4 (5)
14 (18)
10 (13)
8 (10)
- phlebitis or too painful
22 (14)
7 (9)
15 (19)
8 (10)
14 (17)
- dislodgment
15 (9)
5 (6)
10 (12)
7 (9)
8 (10)
Device dwell time (days)a
2.70 (1.66–4.02)
2.23 (1.46–4.32)
2.73 (1.71–3.98)
2.83 (1.65–4.25)
2.26 (1.67–3.58)
Device-hours
11,911
5,998
5,914
6,327
5,584
Incidence rate (95 % CI)b
6.13 (4.87–7.71)
4.84 (3.36–6.96)
7.44 (5.54–10.0)
5.06 (3.58–7.15)
7.34 (5.41–9.97)
Log-rank test (p value)
-
0.063
0.054
n (%) shown unless otherwise noted
CI confidence interval
a Median and inter-quartile range
b Per 1000 device-hours

Primary endpoint

The incidence rate of PIVC failure in patients assigned to 3 mL flushing was 4.8 per 1000 hours compared with 7.4 per 1000 hours in those assigned to 10 mL flushing (p = 0.063, log-rank, Table 2 and Fig. 2). The incidence rate of PIVC failure in patients assigned to 24 hour flushing was 5.1 per 1000 hours compared with 7.3 per 1000 hours in those assigned to 6 hour flushing (p = 0.054, log-rank, Table 2 and Fig. 2). There was no significant interaction between the two study interventions (p = 0.21, Cox, Table 3); i.e. the results for comparison of the two volumes were not affected by the frequency, and the results for comparison between the two frequencies were also not affected by the volume. Univariable Cox regression found estimates of effect for both the volume and frequency interventions to be large but not statistically significant (Table 3). Failure was increased with 10 mL (versus 3 mL) volume (HR 1.56, 95 % CI 0.97–2.50, p = 0.06), and 6 hour (versus 24 hour) flushing (HR 1.58, 95 % CI 0.99–2.53, p = 0.06). On multivariable analysis, volume, frequency and their interaction remained non-significant, but female gender (HR 2.2, 95 % CI 1.3–3.6, p < 0.01), insertion in hand/posterior wrist (HR 1.7, 95 % CI 1.0–2.7, p < 0.05) and increased episodes rate per day (combined flushes and medication pushes) (HR 1.2, 95 % CI 1.1–1.4, p < 0.01) significantly predicted PIVC failure. Increasing the average daily episode rate by one (e.g. from three to four episodes per day) was associated with a 25 % increase in the relative risk of failure.
Table 3
Cox proportional hazards regression (n = 160)
 
Univariable
Multivariable
HR (95 % CI)
HR (95 % CI)
10 mL volume (referent: 3 mL)
1.56 (0.97–2.50)*
f
6 hr frequency (referent = 24 h)
1.58 (0.99–2.53)*
f
Interaction of volume and frequencya
0.54 (0.21–1.42)
f
Age (one year increase)
1.00 (0.99,1.02)
f
Gender (female)
1.90 (1.18,3.05)***
2.17 (1.33,3.55)***
Weight (over/obese)
1.29 (0.81,2.05)
f
Comorbidities (one category higher)b
1.03 (0.78,1.35)
f
Insertion on dominant side
1.67 (1.01,2.76)**
f
Infection at baseline (any)
1.56 (0.93,2.59)*
f
Antibiotics at baseline (yes)
0.99 (0.60,1.64)
f
IV treatment at baseline (yes)
0.95 (0.43,2.07)
f
Insertion at hand/posterior wristc
1.67 (1.04,2.70)**
1.66 (1.02,2.69)**
Insertion at forearma
0.58 (0.35,0.97)**
e
Insertion at cub. fossa/ant. upper armc
1.01 (0.53,1.93)
f
Inserted by IV servicec
0.95 (0.59,1.52)
f
Inserted by nursec
2.01 (1.11,3.63)**
f
Size (20 g versus 22 g)
1.14 (0.70,1.87)
f
Vein quality (poor)
1.26 (0.79,2.01)
f
Multiple insertion attempts (yes)
1.32 (0.78,2.23)
f
Extension tubing (yes)
1.25 (0.78,1.98)
f
3-way tap (yes)
1.33 (0.79,2.24)
f
IV treatment at device removal (yes)
2.61 (1.28,5.29)***
e
IV treatment rate at device removald
1.40 (1.07,1.82)**
e
IV antibiotics at device removal (yes)
2.07 (1.27,3.38)***
e
IV antibiotics rate at device removald
1.33 (1.09,1.62)***
e
Medication types administered:
  
- antibioticsc
1.12 (0.67,1.87)
f
- cephalosporinsc
0.80 (0.49,1.32)
f
- penicillin combinationc
0.75 (0.40,1.41)
f
Episode rate (medications and flushes)
1.22 (1.08,1.38)***
1.25 (1.10,1.41)***
HR hazard ratio, CI confidence interval, IV intravenous
*p value <0.2; **p value <0.05; ***p value <0.01
aModel for the interaction term results included the main effects
bOrdinal: 0 = none/one, 1 = two/three, 2 = four or more
cVersus all others
dPer day, ordinal: 0 = none, 1 = one/two, 2 = three/four, 3 = five or more
eExcluded due to correlation with other covariate
fCovariate not entered into multivariate model (at p ≥ 0.2), or removed during model building at p > 0.05. Only variables p < 0.2 on univariable analysis entered into the multivariable model

Discussion

This pilot study was initiated in response to an unacceptably high PIVC failure rate in our organisation and resultant high morbidity and costs [10, 12]. This is one of the first RCTs to evaluate the impact of different flushing volumes and frequencies on PIVC outcome. Feasibility of a future large trial was indicated, with the interventions acceptable to ward staff and feasible to conduct in the clinical setting as demonstrated by the protocol adherence and successful recruitment rate. Initial group comparisons suggested that higher frequency and higher volume flushing were associated with increased PIVC failure. This is counter-intuitive to current practice, where it is believed more often and larger volume flushing equals less PIVC failure. However, univariable and multivariable regression showed that flushing volume and frequency were not significantly associated with failure. Retrospectively, the study power was 40 % and 14 % for the volume and frequency hypotheses respectively; larger, definitive trials would require 140 patients/group (volume), and 520 patients/group (frequency) (90 % power, alpha 0.05, https://​stat.​ubc.​ca/​~rollin/​stats/​ssize/​).
Similar to our results, a single-site trial (n = 397) of different flushing frequencies (3 mL twice daily versus 3 mL once daily) in Italy also found no significant difference in risk of PIVC failure (12.1 % versus 9.5 % p = 0.42) [22]. However, their sample was a select paediatric population (receiving no infusion therapy or intravenous antibiotics, and using prefilled flush syringes), quite different from our cohort. Additionally, their observed overall proportion of failure (8.7 %) was very low compared to other studies. This suggests that the use of prefilled and/or pressure limiting syringe technology needs to be explored to optimise flush and medication administration.
PIVC failure was significantly associated with the increasing episode rate of PIVC accesses per day. However, medication type (i.e. drug or antibiotic classification) was not a significant contributing factor to PIVC failure. So if it is not the flush volume, frequency or medication type that predicts failure, it may be the excessive injection pressure associated with manually prepared syringes that damages the vessel intima, by direct pressure and/or haemodilution activating the endothelium [17]. It could also be the manual handling of the PIVC during administration.
Our recent survey of flushing practices (n = 1178) revealed 23 % of nurses/midwives used a 5 mL or 2 mL syringe to deliver a flush, and almost all used manually prepared flush syringes [20]. This implies a lack of appreciation for the increase in pressure per square inch associated with the properties of smaller size syringes. The use of reduced pressure for flush delivery through a syringe with a larger diameter such as the standard 10 mL syringe is recommended to optimise flush outcomes and minimise damage to the vein [17, 28, 29]. These recommendations are largely derived from physics principles, and have not been tested in clinical trials. In recent times, commercially prepared prefilled flush syringes have become available, which negate the potential for operators to make an incorrect choice of a smaller size syringe, since they are produced in diameters consistent with a 10 mL syringe, but in 3, 5 and 10 mL volumes.
Female gender was significantly associated with PIVC failure (HR 2.2), consistent with a previous study where females were significantly more at risk for both phlebitis (HR 1.64, p < 0.001) and occlusion (HR 1.44, p < 0.001) [30]. While gender is a non-modifiable factor, staff should take this high risk factor into consideration for best-practice insertion, monitoring and maintenance regimens.
We also identified hand/wrist insertion as a significant risk for PIVC failure (HR 1.7). Previous research confirmed insertion site as a predictor of PIVC failure. Again, insertion in the hand compared to the forearm had an HR of 1.47 (p < 0.001) for occlusion and an HR of 2.45 (p < 0.001) for accidental removal in previous work [30]. Other research has linked insertion site with phlebitis [14, 31, 32]. PIVC insertion site is a variable over which clinicians can exercise judgement and choice, and they should therefore consider these risks in relation to their insertion practice. Guideline site recommendations have recently been updated to preference the forearm [18, 24, 33].
It is clear that the mechanisms of how IV flushing, medication and fluid administration impact on PIVC complications require further elucidation. The literature to date, including Schreiber and colleagues (2015) [22] and this pilot study, have not identified effective flushing regimens. There are few RCTs performed on PIVC flushing, and this study provides new, rigorous data that contribute to knowledge in this area. More recently, clinically indicated replacement has allowed catheters to be used for longer periods, so trials such as this are needed to investigate how improved maintenance can keep PIVCs functional over time [34].

Limitations

The small pilot sample and single site setting impact on the interpretation and generalisability of the results. It was not possible to mask the respective interventions, and so there was potential for outcome assessment bias. The use of dedicated research nurses, standardised data collection and blinded microbiologists minimised the risk. We did not control flushing volumes used pre- and post-medications, and this may have confounded the effect of the randomised flushing. However, analysing the number of total PIVC accesses per day appears to have controlled for this factor. The recommended syringe size for all flushes was a 10 mL syringe; however, we do not know if this was always used, and excessive pressure may have been a confounder if smaller sizes were used.

Conclusions and implications

Larger, definitive trials of flushing volume and flushing frequency of peripheral intravenous catheter management are feasible and required. Neither increased flushing volume nor frequency significantly altered the risk of PIVC failure in this pilot trial; however, the trial had inadequate power. (Female gender, hand/posterior wrist placement and frequency of access (flushes and medication) should be considered confounders in future trials.) The mechanisms of how IV flushing, medication and fluid administration impact on the cannula, vessel endothelium and blood components are poorly understood and require further explication. We need to explore how syringe technology and method of administration can make a difference to PIVC outcomes and therefore transform the delivery and patient experience of IV care.

Abbreviations

CI, confidence interval; HR, hazard ratio; IV, intravenous; PIVC, peripheral intravenous catheter; RCT, randomised controlled trial; RN, Research Nurse

Acknowledgements

The researchers are grateful to all the patients who participated in the study and the clinical staff who support and facilitated the study. The researchers acknowledge the funding that supported this study from an NHMRC Centre for Research Excellence in Nursing (NCREN) seed grant, and from BD, a global medical technology company that makes flushing products (through Becton Dickinson Pty Ltd in Australia) in the form of a grant-in-aid. BD did not design the study protocol, collect or analyse data, and did not prepare the manuscript.

Authors’ contributions

SK, JF, NM, GM and CR participated in the conception and design of the trial. SK and JF drafted the manuscript with revisions provided by NM, GM, KD and CR. SK, JF, NM and KD participated in management of the trial. SK, JF and NM participated in recruitment and treatment of patients in each centre. SK, JF, NM, GM and CR participated in data collection and analysis. All authors read and approved the final version to be published.

Competing interests

BD, a global medical technology company that makes flushing products (through Becton Dickinson Pty Ltd in Australia) partly funded this study in the form of a grant-in-aid. BD did not design the study protocol, collect or analyse data, and did not prepare the manuscript. The authors declare that they have no competing interests.
Open AccessThis article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://​creativecommons.​org/​licenses/​by/​4.​0/​), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The Creative Commons Public Domain Dedication waiver (http://​creativecommons.​org/​publicdomain/​zero/​1.​0/​) applies to the data made available in this article, unless otherwise stated.
Literatur
1.
Zurück zum Zitat Hadaway L. Short peripheral intravenous catheters and infections. J Infus Nurs. 2012;35(4):230–40.CrossRefPubMed Hadaway L. Short peripheral intravenous catheters and infections. J Infus Nurs. 2012;35(4):230–40.CrossRefPubMed
2.
Zurück zum Zitat Maki DG. Improving the safety of peripheral intravenous catheters. Br Med J. 2008;337:a630.CrossRef Maki DG. Improving the safety of peripheral intravenous catheters. Br Med J. 2008;337:a630.CrossRef
3.
Zurück zum Zitat Zingg W, Pittet D. Peripheral venous catheters: an under-evaluated problem. Internat J Antimicrobial Agents. 2009;34 Suppl 4:S38–42.CrossRef Zingg W, Pittet D. Peripheral venous catheters: an under-evaluated problem. Internat J Antimicrobial Agents. 2009;34 Suppl 4:S38–42.CrossRef
4.
Zurück zum Zitat Bausone-Gazda D, Lefaiver CA, Walters SA. A randomized controlled trial to compare the complications of 2 peripheral intravenous catheter-stabilization systems. J Infus Nurs. 2010;33(6):371–84.CrossRefPubMed Bausone-Gazda D, Lefaiver CA, Walters SA. A randomized controlled trial to compare the complications of 2 peripheral intravenous catheter-stabilization systems. J Infus Nurs. 2010;33(6):371–84.CrossRefPubMed
5.
Zurück zum Zitat Chico-Padron RM, Carrion-Garcia L, Delle-Vedove-Rosales L, Gonzalez-Vargas CS, Marrero-Perera M, Medina-Chico S, et al. Comparative safety and costs of transparent versus gauze wound dressings in intravenous catheterization. J Nurs Care Qual. 2011;26(4):371–6.PubMed Chico-Padron RM, Carrion-Garcia L, Delle-Vedove-Rosales L, Gonzalez-Vargas CS, Marrero-Perera M, Medina-Chico S, et al. Comparative safety and costs of transparent versus gauze wound dressings in intravenous catheterization. J Nurs Care Qual. 2011;26(4):371–6.PubMed
6.
Zurück zum Zitat Forni C, D'Alessandro F, Gambino O, Amodeo A, Pignotti E, Zanotti E, et al. Effectiveness of the transparent sterile dressing vs standard to fix the peripheral venous catheter (PVC) on the incidence of phlebitis. A randomized controlled trial. Assistenza infermieristica e ricerca: AIR. 2012;31(2):63–9.PubMed Forni C, D'Alessandro F, Gambino O, Amodeo A, Pignotti E, Zanotti E, et al. Effectiveness of the transparent sterile dressing vs standard to fix the peripheral venous catheter (PVC) on the incidence of phlebitis. A randomized controlled trial. Assistenza infermieristica e ricerca: AIR. 2012;31(2):63–9.PubMed
7.
Zurück zum Zitat Fujita T, Namiki N. Replacement of peripheral intravenous catheters. J Infus Nurs. 2008;17(18):2509–10. Fujita T, Namiki N. Replacement of peripheral intravenous catheters. J Infus Nurs. 2008;17(18):2509–10.
8.
Zurück zum Zitat Fujita T, Namiki T, Suzuki T, Yamamoto E. Normal saline flushing for maintenance of peripheral intravenous sites. J Infus Nurs. 2006;15(1):103–4. Fujita T, Namiki T, Suzuki T, Yamamoto E. Normal saline flushing for maintenance of peripheral intravenous sites. J Infus Nurs. 2006;15(1):103–4.
9.
Zurück zum Zitat Malyon L, Ullman AJ, Phillips N, Young J, Kleidon T, Murfield J, et al. Peripheral intravenous catheter duration and failure in paediatric acute care: a prospective cohort study. Emerg Med Aust. 2014;26(6):602–8.CrossRef Malyon L, Ullman AJ, Phillips N, Young J, Kleidon T, Murfield J, et al. Peripheral intravenous catheter duration and failure in paediatric acute care: a prospective cohort study. Emerg Med Aust. 2014;26(6):602–8.CrossRef
10.
Zurück zum Zitat Marsh N, Webster J, Flynn J, Mihala G, Hewer B, Fraser J, et al. Securement methods for peripheral venous catheters to prevent failure: a randomised controlled pilot trial. J Vasc Access. 2015;16(3):237–44.CrossRefPubMed Marsh N, Webster J, Flynn J, Mihala G, Hewer B, Fraser J, et al. Securement methods for peripheral venous catheters to prevent failure: a randomised controlled pilot trial. J Vasc Access. 2015;16(3):237–44.CrossRefPubMed
11.
Zurück zum Zitat Nassaji-Zavareh M, Ghorbani R. Peripheral intravenous catheter-related phlebitis and related risk factors. Singap Med J. 2007;48(8):733–6. Nassaji-Zavareh M, Ghorbani R. Peripheral intravenous catheter-related phlebitis and related risk factors. Singap Med J. 2007;48(8):733–6.
12.
Zurück zum Zitat Rickard CM, Webster J, Wallis MC, Marsh N, McGrail MR, French V, et al. Routine versus clinically indicated replacement of peripheral intravenous catheters: a randomised controlled equivalence trial. Lancet. 2012;380(9847):1066–74.CrossRefPubMed Rickard CM, Webster J, Wallis MC, Marsh N, McGrail MR, French V, et al. Routine versus clinically indicated replacement of peripheral intravenous catheters: a randomised controlled equivalence trial. Lancet. 2012;380(9847):1066–74.CrossRefPubMed
13.
Zurück zum Zitat Tuffaha H, Rickard CM, Wenbster J, Scuffham P, Marsh N, Gordon L. Cost effectiveness analysis of clinically indicated versus routine replacement of peripheral intravenous catheters. Appl Health Economics Health Policy. 2012;12(1):51–8.CrossRef Tuffaha H, Rickard CM, Wenbster J, Scuffham P, Marsh N, Gordon L. Cost effectiveness analysis of clinically indicated versus routine replacement of peripheral intravenous catheters. Appl Health Economics Health Policy. 2012;12(1):51–8.CrossRef
14.
Zurück zum Zitat Maki DG, Ringer M. Risk factors for infusion-related phlebitis with small peripheral venous catheters. A randomized controlled trial. Ann Intern Med. 1991;114(10):845–54.CrossRefPubMed Maki DG, Ringer M. Risk factors for infusion-related phlebitis with small peripheral venous catheters. A randomized controlled trial. Ann Intern Med. 1991;114(10):845–54.CrossRefPubMed
15.
Zurück zum Zitat Malach T, Jerassy Z, Rudensky B, Schlesinger Y, Broide E, Olsha O, et al. Prospective surveillance of phlebitis associated with peripheral intravenous catheters. Am J Infect Control. 2006;34(5):308–12.CrossRefPubMed Malach T, Jerassy Z, Rudensky B, Schlesinger Y, Broide E, Olsha O, et al. Prospective surveillance of phlebitis associated with peripheral intravenous catheters. Am J Infect Control. 2006;34(5):308–12.CrossRefPubMed
16.
Zurück zum Zitat Zhang L, Keogh S, Rickard CM. Reducing the risk of infection associated with vascular access devices through nanotechnology: a perspective. Int J Nanomedicine. 2013;8:4453–66.CrossRefPubMedPubMedCentral Zhang L, Keogh S, Rickard CM. Reducing the risk of infection associated with vascular access devices through nanotechnology: a perspective. Int J Nanomedicine. 2013;8:4453–66.CrossRefPubMedPubMedCentral
17.
Zurück zum Zitat Hadaway L. Technology of flushing vascular access devices. J Infus Nurs. 2006;29(3):129–45.CrossRef Hadaway L. Technology of flushing vascular access devices. J Infus Nurs. 2006;29(3):129–45.CrossRef
18.
Zurück zum Zitat Gorski L, Hadaway L, Hagle ME, McGoldrick M, Orr M, Doellman D. Infusion therapy standards of practice. J Infus Nurs. 2016;39(Supp1):S1–159. Gorski L, Hadaway L, Hagle ME, McGoldrick M, Orr M, Doellman D. Infusion therapy standards of practice. J Infus Nurs. 2016;39(Supp1):S1–159.
19.
Zurück zum Zitat Queensland Health. Insertion and management of peripheral intravenous catheters guideline. In: Centre for Healthcare Related Infection Surveillance and Prevention. Brisbane: Queensland Government; 2013. Queensland Health. Insertion and management of peripheral intravenous catheters guideline. In: Centre for Healthcare Related Infection Surveillance and Prevention. Brisbane: Queensland Government; 2013.
20.
Zurück zum Zitat Keogh S, Flynn J, Marsh N, Higgins N, Davies K, Rickard C. Nursing and midwifery practice for maintenance of vascular access device patency. A cross-sectional survey. Int J Nurs Stud. 2015;52(11):1678–85.CrossRefPubMed Keogh S, Flynn J, Marsh N, Higgins N, Davies K, Rickard C. Nursing and midwifery practice for maintenance of vascular access device patency. A cross-sectional survey. Int J Nurs Stud. 2015;52(11):1678–85.CrossRefPubMed
21.
Zurück zum Zitat New K, Marsh N, Hewer B, Webster J. Intravascular device utilization and complications: a point prevalence survey. Aust Health Rev. 2014;38(3):345–9.CrossRefPubMed New K, Marsh N, Hewer B, Webster J. Intravascular device utilization and complications: a point prevalence survey. Aust Health Rev. 2014;38(3):345–9.CrossRefPubMed
22.
Zurück zum Zitat Schreiber S, Zanchi C, Ronfani L, Delise A, Corbelli A, Bortoluzzi R, et al. Normal saline flushes performed once daily maintain peripheral intravenous catheter patency: a randomised controlled trial. Arch Dis Child. 2015;100(7):700–3.CrossRefPubMed Schreiber S, Zanchi C, Ronfani L, Delise A, Corbelli A, Bortoluzzi R, et al. Normal saline flushes performed once daily maintain peripheral intravenous catheter patency: a randomised controlled trial. Arch Dis Child. 2015;100(7):700–3.CrossRefPubMed
23.
Zurück zum Zitat Montgomery AA, Fahey T, Peters TJ. A factorial randomised controlled trial of decision analysis and an information video plus leaflet for newly diagnosed hypertensive patients. Br J Gen Pract. 2003;53(491):446–53.PubMedPubMedCentral Montgomery AA, Fahey T, Peters TJ. A factorial randomised controlled trial of decision analysis and an information video plus leaflet for newly diagnosed hypertensive patients. Br J Gen Pract. 2003;53(491):446–53.PubMedPubMedCentral
24.
Zurück zum Zitat O'Grady NP, Alexander M, Burns LA, Dellinger EP, Garland J, Heard SO, et al. Guidelines for the prevention of intravascular catheter-related infections. Am J Infect Control. 2011;39(4 Suppl 1):S1–34.CrossRefPubMed O'Grady NP, Alexander M, Burns LA, Dellinger EP, Garland J, Heard SO, et al. Guidelines for the prevention of intravascular catheter-related infections. Am J Infect Control. 2011;39(4 Suppl 1):S1–34.CrossRefPubMed
25.
Zurück zum Zitat Webster J, Clarke S, Paterson D, Hutton A, van Dyk S, Gale C, et al. Routine care of peripheral intravenous catheters versus clinically indicated replacement: randomised controlled trial. Br Med J. 2008;337(a339) Webster J, Clarke S, Paterson D, Hutton A, van Dyk S, Gale C, et al. Routine care of peripheral intravenous catheters versus clinically indicated replacement: randomised controlled trial. Br Med J. 2008;337(a339)
26.
Zurück zum Zitat Hertzog MA. Considerations in determining sample size for pilot studies. Res Nurs Health. 2008;31(2):180–91.CrossRefPubMed Hertzog MA. Considerations in determining sample size for pilot studies. Res Nurs Health. 2008;31(2):180–91.CrossRefPubMed
27.
Zurück zum Zitat Altman DG, Bland JM. Interaction revisited: the difference between two estimates. Br Med J. 2003;326(7382):219.CrossRef Altman DG, Bland JM. Interaction revisited: the difference between two estimates. Br Med J. 2003;326(7382):219.CrossRef
28.
Zurück zum Zitat Macklin D. What's physics got to do with it? J Vasc Access Devices. 1999;4(2):7–13.CrossRef Macklin D. What's physics got to do with it? J Vasc Access Devices. 1999;4(2):7–13.CrossRef
29.
Zurück zum Zitat Perucca R. Peripheral venous access devices. In: Alexander M, Corrigan A, Gorski L, Hankins J, Perucca R, editors. Infusion nursing: an evidence-based approach. 3rd ed. St Louis: Saunders/Elsevier; 2010. p. 456–79. Perucca R. Peripheral venous access devices. In: Alexander M, Corrigan A, Gorski L, Hankins J, Perucca R, editors. Infusion nursing: an evidence-based approach. 3rd ed. St Louis: Saunders/Elsevier; 2010. p. 456–79.
30.
Zurück zum Zitat Wallis MC, McGrail M, Webster J, Marsh N, Gowardman J, Playford EG, et al. Risk factors for peripheral intravenous catheter failure: a multivariate analysis of data from a randomized controlled trial. Infect Control Hosp Epidemiol. 2014;35(1):63–8.CrossRefPubMed Wallis MC, McGrail M, Webster J, Marsh N, Gowardman J, Playford EG, et al. Risk factors for peripheral intravenous catheter failure: a multivariate analysis of data from a randomized controlled trial. Infect Control Hosp Epidemiol. 2014;35(1):63–8.CrossRefPubMed
31.
Zurück zum Zitat Dillon MF, Curran J, Martos R, Walsh C, Walsh J, Al-Azawi D, et al. Factors that affect longevity of intravenous cannulas: a prospective study. QJM. 2008;101(9):731–5.CrossRefPubMed Dillon MF, Curran J, Martos R, Walsh C, Walsh J, Al-Azawi D, et al. Factors that affect longevity of intravenous cannulas: a prospective study. QJM. 2008;101(9):731–5.CrossRefPubMed
32.
Zurück zum Zitat Tagalakis V, Kahn SR, Libman M, Blostein M. The epidemiology of peripheral vein infusion thrombophlebitis: a critical review. Am J Med. 2002;113(2):146–51.CrossRefPubMed Tagalakis V, Kahn SR, Libman M, Blostein M. The epidemiology of peripheral vein infusion thrombophlebitis: a critical review. Am J Med. 2002;113(2):146–51.CrossRefPubMed
33.
Zurück zum Zitat Infusion Nurses Society. Infusion nursing standards of practice. J Infus Nurs. 2016;39(1S):S77–81. Infusion Nurses Society. Infusion nursing standards of practice. J Infus Nurs. 2016;39(1S):S77–81.
34.
Zurück zum Zitat Loveday HP, Wilson JA, Pratt RJ, Golsorkhi M, Tingle A, Bak A, et al. epic3: national evidence-based guidelines for preventing health care associated infections in NHS hospitals in England. J Hosp Infect. 2014;86 Suppl 1:S1–70.CrossRefPubMed Loveday HP, Wilson JA, Pratt RJ, Golsorkhi M, Tingle A, Bak A, et al. epic3: national evidence-based guidelines for preventing health care associated infections in NHS hospitals in England. J Hosp Infect. 2014;86 Suppl 1:S1–70.CrossRefPubMed
Metadaten
Titel
Varied flushing frequency and volume to prevent peripheral intravenous catheter failure: a pilot, factorial randomised controlled trial in adult medical-surgical hospital patients
verfasst von
Samantha Keogh
Julie Flynn
Nicole Marsh
Gabor Mihala
Karen Davies
Claire Rickard
Publikationsdatum
01.12.2016
Verlag
BioMed Central
Erschienen in
Trials / Ausgabe 1/2016
Elektronische ISSN: 1745-6215
DOI
https://doi.org/10.1186/s13063-016-1470-6

Weitere Artikel der Ausgabe 1/2016

Trials 1/2016 Zur Ausgabe