Open Access 07.06.2017 | Original Research Article

Wavelet decomposition analysis in the two-flash multifocal ERG in early glaucoma: a comparison to ganglion cell analysis and visual field

verfasst von: Livia M Brandao, Matthias Monhart, Andreas Schötzau, Anna A Ledolter, Anja M Palmowski-Wolfe

Erschienen in: Documenta Ophthalmologica | Ausgabe 1/2017

Abstract

Purpose

To further improve analysis of the two-flash multifocal electroretinogram (2F-mfERG) in glaucoma in regard to structure–function analysis, using discrete wavelet transform (DWT) analysis.

Methods

Sixty subjects [35 controls and 25 primary open-angle glaucoma (POAG)] underwent 2F-mfERG. Responses were analyzed with the DWT. The DWT level that could best separate POAG from controls was compared to the root-mean-square (RMS) calculations previously used in the analysis of the 2F-mfERG. In a subgroup analysis, structure–function correlation was assessed between DWT, optical coherence tomography and automated perimetry (mf103 customized pattern) for the central 15°.

Results

Frequency level 4 of the wavelet variance analysis (144 Hz, WVA-144) was most sensitive (p < 0.003). It correlated positively with RMS but had a better AUC. Positive relations were found between visual field, WVA-144 and GCIPL thickness. The highest predictive factor for glaucoma diagnostic was seen in the GCIPL, but this improved further by adding the mean sensitivity and WVA-144.

Conclusions

mfERG using WVA analysis improves glaucoma diagnosis, especially when combined with GCIPL and MS.

Introduction

Early diagnosis in glaucoma is still a challenge. While optical coherence tomography (OCT) technology detects ganglion cell loss before a defect is visualized in the standard automated perimetry (SAP) in some patients [1], in others the SAP defect appears before any ganglion cell layer thinning is detected. Visual electrophysiology offers additional noninvasive tests and provides more objective measures to monitor retinal cell activity.
The multifocal electroretinogram (mfERG, two-flash pattern) technology which allows topographic examination of retinal function is an objective method shown to be sensitive in glaucoma [28]. Also, it complements visual field information and may be more sensitive in some patients, especially in cases where defects are seen in OCT but not yet in the visual field (poster ARVO 2014: IOVS. 2014; 55(13):975).
The entire process of retinal visual processing involves the phototransduction cascade with different groups of cells and circuits from the photoreceptors to the ganglion cells. Thus, electrical signals produced by different biological structures contribute to the retinal response of the mfERG that is recorded from the cornea [812]. In the standard mfERG, amplitude and implicit time are often analyzed [13]. In the two-flash mfERG, the response is smaller and there are more adaptive influences interfering with traditional analysis [7]. Due to the complexity of the two-flash mfERG waveform, disease-specific local variations may be difficult to identify. The responses from vertical and horizontal pathways are not uniform in time. Therefore, individual wave components, important in, e.g., glaucoma, may be hidden inside the whole summed sequence of activity recorded over a period of time, under a certain stimulus.
Applying mathematical methods, which work as a “filter” in order to identify “hidden characteristics” inside the electrical signal, have been proposed to optimize analysis of electroretinogram (ERG) responses. Bach and Meigen [14] analyzed the application of discrete fourier transform (DFT) in steady-state evoked potentials (i.e., pattern ERG, VEP). Although its application demonstrated good results, DFT has not gained wide acceptance in the analysis of the mfERG [15, 16].
Wavelet analysis, which takes both frequency and time into consideration, is not a new topic in biomedical signals. Its use has been increasingly studied and applied either to improve the interpretation of the biosignal itself (ECG [17], EMG [18]) or to exclude interference between signals [19]. In electrophysiology of vision, this method has been introduced a decade ago [20, 21]. It has been used to analyze the photopic negative response in the full-field ERG of healthy subjects [22]. Barraco et al. [23] compared different analytical approaches (including application of the continuous wavelet transform, CWT) to analyze the a-wave of the full-field ERG, demonstrating consistently altered photoreceptor behavior in various diseases. Nair and Joseph [24] could differentiate patients with congenital stationary night blindness, rod–cone dystrophy and central retinal occlusion from healthy controls using wavelet analysis of the full-field ERG. Discrete wavelet transform (DWT) analysis of the ERG waveform has been reported to be superior to the traditional time-domain analysis [21, 25]. Also, Gauvin et al. [25] investigated the luminance dependence of the full-field ERG response applying various DWT descriptors and introducing the use of the Hölder exponent. Lately, he applied DWT to assess the contribution of the oscillatory potentials to the full-field ERG [26]. With specific focus in glaucoma Miguel-Jiménez et al. [27] applied different types of wavelet analysis to the global flash mfERG response in glaucoma and found that CWT analysis, discrete wavelet transform (DWT) [28] and discrete wavelet packet transform [29] can separate advanced glaucoma from control and give additional information to the defect seen on Humphrey visual fields from these patients.
The aim of this study was to apply the DWT analysis to the two-flash multifocal electroretinogram (2F-mfERG) responses in primary open-angle glaucoma (POAG), in order to identify, among different mother wavelets, the most suitable one for the 2F-mfERG signal shape and to test its performance to differentiate glaucoma from control. In addition, we evaluated how the DWT analysis compares to the root-mean-square (RMS) calculations previously used in our studies for the analysis of the 2F-mfERG. For the most sensitive descriptor of the DWT, we then investigated the association with the ganglion cell–inner plexiform layer (GCIPL) thickness measured on OCT, as well as the visual field sensitivity using a customized visual field pattern based on the stimulus grid of the mfERG (mf103-pattern).

Materials and methods

The study protocol was approved by the Ethics Committee of the University of Basel. All procedures followed the tenets of the Declaration of Helsinki. Informed consent was signed before participation.
Thirty-five healthy individuals and 25 POAG patients were included in this study. Inclusion criteria for all individuals included: visual acuity of 0.8 or better and refractive error between ±6 diopters of hyperopia or myopia. All patients were recruited at their regular glaucoma specialist consultation. Patients diagnosed with glaucoma presented glaucomatous optic neuropathy at fundus examination, significant localized thinning of the neuroretinal rim on the RNFL thickness map and, for POAG patients, a reproducible visual field defect (minimum three tests, Octopus 101, G2 protocol). Normal parameters in Octopus visual fields are: mean defect (MD) under 2.2 dB and a square root loss of variance (sLV) under 2.5 dB. Pre-perimetric glaucoma (PPG) patients presented with glaucomatous optic nerve characteristics associated with significant thinning of the RNFL on OCT, but no visual field defect.
Individuals diagnosed with systemic diseases which may affect the eye (e.g., diabetes), with regular use of medications that can influence retinal cell electrical activity (e.g., antidepressant, chloroquine, anticonvulsants) or who had previous ocular surgeries (e.g., cataract extraction, glaucoma surgery), were excluded.
All participants underwent visual acuity testing, slit lamp, fundus examination and tonometry (Goldmann). In addition a 2F-mfERG was performed to assess DWT and to decide on the best mfERG DWT descriptor to separate glaucoma patients from normal subjects.
The 2F-mfERG protocol used in this study has been described previously [3, 6, 30]. Briefly, it was recorded with VERIS Science 6.06, FMSIII (Electro-Diagnostic Imaging, USA). Pupils were dilated with a solution (Tropicamide 0.5%, Phenylephrine 1%, Spital Pharmazie USB, Switzerland), and a Burian Allen bipolar contact lens was used after application of a gel interface (Methocel 2%, OmniVision AG, Switzerland). The 103 hexagons stimulated the central 50° of the retina according to an m-sequence of length 213 − 1 (L max 100 cd/m2, L min < 1 cd/m2). Each m-sequence step (M) was followed by two global flashes of 200 cd/m2 (F) at an interval of 26 ms. This interval was created by interposing dark frames at <1 cd/m2 (O), thus creating the stimulation sequence MOFOFO as shown in Fig. 1. We used a bandpass filter of 1–300 Hz. Total recording time was 10 min and 55 s, divided into 16 segments. Recordings with poor signal or contaminated by ocular movements were discarded and re-recorded. Artifact rejection filtering incorporated in the VERIS software was applied twice, as suggested by the manufacturer. Spatial averaging was not applied. The addition of a global flash to the standard mfERG m-sequence stimulus increases the contribution from the inner retina to the electrophysiological response [31] which is thought to enhance glaucoma detection [32, 33]. For the waveform analysis, cross-correlation between the m-sequence and the raw data recorded allowed individual focal waveforms to be derived. These were exported from VERIS using the command “export analyzed data.” For the DWT, the first-order 2F-mfERG response was analyzed from 15 to 105 ms for each focal response within the central 15° of the retina (19 responses).
In order to compare our results from the DWT to our results from previous studies, we also analyzed the RMS of the 19 central responses (15°) filtered at 1–200 Hz as this has shown best differentiation between POAG and control [2, 3, 30]. For this comparison, we analyzed the response to the m-sequence step (MOFOFO), the direct component, DC, at 15–45 ms and two induced components IC1 at 45–75 ms and IC2 at 75–105 ms (Fig. 1).

Discrete wavelet analysis

In general, DWT represents discrete time series such as biosignals as real-valued functions of time and temporal frequency. An initial wavelet template, the “mother wavelet,” is changed in scaling (temporal frequency) and location (time). Changing width and location of the template creates a wavelet family that is correlated with the signal. The values of these correlations are used as coefficients to characterize the signal in frequency and time. Coefficients can be allocated to decomposition levels of descending frequency levels (high to low). The number of frequency levels depends on the length of the time series. Further technical information can be found in references [25, 26, 34].
Figure 1 shows an example of DWT analysis applied to a recorded signal from a control and a patient in this study. The original signal from one hexagon of the mfERG (waveform in the top box) can be decomposed into many frequency levels, depending on the length of the time series. The first level (1211 Hz) corresponds to high frequencies (noise), while the highest level (11 Hz) corresponds to the lowest frequencies. For each frequency level, the vertical lines represent individual wavelet coefficients. For each frequency level, the variance between these coefficients is computed and subjected to further analysis as the wavelet variance analysis (WVA).
Review of the literature showed that several mother wavelets have been applied in DWT of the ERG response, such as the “Daubechies” wavelets [20, 21, 25], the “Haar” wavelet [24, 35] and the “Mexican hat” [23, 36, 37]. For glaucoma, Miguel-Jiménez et al. have successfully applied DWT to the global flash mfERG response in advanced glaucoma. They analyzed a number of mother templates (not specified) and on visual comparison found the mother template Bior 3.1 to have the best performance [29]. In a later paper, they applied continuous wavelet transform using the Morlet waveform [27] with good results. In our study, we first compared the performance of different possible mother wavelets, such as the “Haar” wavelet and Daubechies S6, S8 and S10. Wavelet form Bior 3.1 was not contained in our software package and thus not tried. Performance was quantified as the statistical difference based on p values from mixed effects models. “Haar” wavelet, “Daubechies S6” and “Daubechies S10” showed larger p values comparing glaucoma against the control group. Larger p values are conventionally less significant when considering acceptance of a null hypothesis. Thus, decomposition was done using the “Daubechies S8” wavelet, which is default in the applied software package.
In the present study, seven frequency decomposition levels (1211–11 Hz) were evaluated.
In order to discriminate between study groups [controls and POAG (high-tension glaucoma (HTG), normal-tension glaucoma (NTG) and PPG)], various descriptors (e.g., describing factors: variance, energy, median, min, max IQR) were derived based on the coefficients. Here the best descriptor was the WVA which is in agreement with Gauvin et al. [25], who demonstrated the advantages of WVA application when using DWT (Daubechies wavelet) in ERG.
Figure 2 summarizes our decomposition results. For each group, the box plots show the distribution of the WVA considering each location (19 focal mfERG waveforms) for each subject within each frequency level analyzed. Variance at frequency level 4 (144 Hz) was the most sensitive distinguishing parameter (p = 0.015, red box). Thus, we focused our subsequent analysis on WVA at frequency level 4 (WVA-144). In order to reduce the effect of potential outliers based on edge effects, WVA was explored on a log scale.

Structure function analysis

In a subgroup of 31 subjects (16 POAG, 15 controls), 2F-mfERG, OCT and customized visual fields could be obtained on the same day. In these subjects, we investigated the association between the most sensitive parameter of the DWT (DWT descriptor) and the GCIPL thickness as well as the visual field sensitivity. From each of these diagnostic exams, we analyzed the central 15° degrees of the retina (central 19 points, Fig. 3), as this is the area covered by the ganglion cell analysis in the OCT (4.0 mm vertical and 4.8 mm horizontal diameter). This area is also known to comprise the highest density of ganglion cells in the retina [38] and is the area previously reported as most sensitive in the 2F-mfERG in glaucoma [3].

GCIPL analysis

GCIPL analysis was done using Cirrus SD-OCT [Carl Zeiss, USA, macular cube protocol (512 × 128)]. GCIPL thickness was calculated directly by the Cirrus software (version 6.5.0.722).

Customized automated perimetry pattern

The customized automated perimetry pattern, mf103-pattern, (poster ARVO 2015: Invest. Ophthalmol Vis Sci 2015; 56(7):629) is implemented in the Octopus Perimeter (Octopus 900, Haag-Streit, Switzerland) using the same test parameters as in SAP (size III; 100-ms duration; background 10 cd/m2; 0 dB equaled to 4000 asb). Its main difference to SAP is that each stimulus presented to the patient is positioned to correspond to the center of each hexagon in the mfERG grid (103 stimulus locations covering 50°). Therefore, the sensitivity of each point from the mf103-pattern relates directly to the same hexagon from the 2F-mfERG response. Sensitivity values from each location were calculated individually in decibels transformed into linear units and compared either individually or averaged for the entire 15°.

Statistical analysis

As WVAs have a skewed distribution, they were log-transformed in order to achieve an approximately normal distribution.
Comparisons between study groups were done using linear mixed effects models using age and gender as adjusting covariates. The mixed effects model generates averages over the 19 central focal WVA-144 values (144 Hz) for each subject. These averages are then compared between the study groups. Results are presented as differences of mean WVAs between study groups with corresponding 95% confidence intervals and p values. A p value <0.05 is considered significant. All evaluations were done using the statistical software R version 3.3.0 [39]. Wavelet analysis was performed using the package “wmtsa” contained in software R. Details are described by Walden [34].
ROC curves with corresponding AUCs were calculated to compare the diagnostic performance between different parameters and parameter combinations. Sensitivities and specificities were not calculated as this should only be done in large validated studies, which is not the case here [40]. In order to prevent overfitting due to small sample sizes, AUCs were calculated using penalized logistic regression with internal tenfold cross-validation (repeated 10 times) using the package “glmnet.” Penalized regression shrinks the regression coefficients, which prevents overoptimistic results based on small sample sizes. This calculation and comparisons between AUCs were done within the package “caret,” which is a convenient tool to examine predictive performances of statistical models. Within this package, the AUCs were internally 3 × 10-fold cross-validated and all pairwise differences were computed and tested to assess whether the difference is equal to zero [41, 42].
Calculations were examined with age-matched disease versus control groups to avoid the expected strong age influence on disease diagnosis which could mask the influence of study parameters. Matching was done based on propensity scores implemented in the package “nonrandom.”

Results

Demographics from the analysis group are listed in Table 1. Average overall MD was 3.6 dB for POAG and 0.1 dB for control (p < 0.01). All patients presented with a controlled intraocular pressure, under topical medication if needed. In all statistical calculations, data were adjusted for age and gender.
Table 1
Demographic results from DWT analysis group
Group
Controls
(n = 35)
PPG
(n = 7)
NTG
(n = 12)
HTG
(n = 6)
p value
(anova)
Age (years)
Mean ± SD
50.8
±12.1
63.8
±14.7
60.1
±13.7
62
±10.2
0.017*
Gender (M/F)
11/24
6/1
9/3
5/1
 
BCVA (decimal)
Mean ± SD
1.0
±0.1
0.99
±0.13
0.96
±0.06
0.91
±0.13
0.035*
Refraction (mean 95% CI)
 Diopters
−0.6
−1.3/0.1
−0.4
−1.3/0.4
−0.02
−1.5/1.4
0.5
−1.4/2.6
0.558
 Cylinder
−0.9
−1.1/−0.7
−0.6
−1.2/0.04
−0.6
−0.8/−0.4
−1.2
−2.1/−0.3
0.188
IOP (mmHg)
Mean ± SD
14.2
±2.9
13.4
±1.5
12.8
±1.7
12
±3.7
0.066
CD
Mean ± SD
0.3
±0.07
0.7
±0.1
0.7
±0.1
0.7
±0.1
0.000*
SD standard deviation; PPG pre-perimetric glaucoma; NTG normal-tension glaucoma; HTG high-tension glaucoma; BVCA best corrected visual acuity; 95% CI 95% confidence interval; IOP intraocular pressure; CD cup-to-disk ratio; * p value <0.05 was considered statistically significant

Discrete wavelet analysis

DWT analysis was applied to the 2F-mfERG, and the “Daubechies S8” wavelet was identified as the mother wavelet with the best performance for this dataset (see methods). Starting with the “Daubechies S8” wavelet, we looked for the best descriptor to differentiate glaucoma from controls by analyzing the variance of the wavelet coefficients at different frequency levels. Significant differences between control and POAG were seen at a frequency of 144 Hz (p = 0.015), 66.6 Hz (p = 0.025) and 600 Hz (p = 0.046) (Fig. 2). In our further analysis, we used the most sensitive descriptor we found, WVA-144, which corresponds to the decomposition level of 144 Hz.
In addition, we also analyzed WVA-144 in the individual epochs, DC, IC1 and IC2 but did not find a higher significance than seen in the DWT of the overall waveform between 15 and 105 ms.
We compared the predictive diagnostic performance of WVA-144 to RMS using the AUC of the ROC curve in an age-matched subgroup of 20 controls and 20 POAG patients. In these patients, RMS showed a significant positive correlation with WVA-144 (DC: p < 0.001, IC1: p < 0.001, IC2: p < 0.001). In the ROC analysis, WVA-144 and DC had the highest AUC values of 0.692 and 0.650, respectively. Both RMS-based AUC and WVA-144-based AUC could differentiate POAG from control. However, the DeLong test [43] demonstrated a statistical significant difference between the RMS-based AUC at each of the epochs (DC, IC1 and IC2) and the WVA-144-based AUC (p < 0.001). This suggests that the WVA-144 was more sensitive than the RMS measures, as it has the highest AUC.

Structure–function analysis

In 15 controls and 16 POAG patients, 2F-mfERG, mf103-pattern and macular OCTs were obtained on the same day in order to investigate structure–function relationships.
Table 2 shows the characteristics of this population: Total MD was 4.1 dB in the glaucoma group and 0.5 dB in controls (p < 0.01). GCIPL was thinner in POAG [68.0 µm (±8.7)] than in controls (80.9 µm (±4.8), p < 0.01). WVA-144 was significantly lower in POAG than in controls (p = 0.038). Both GCIPL thickness and visual field linear sensitivity showed a significant positive relationship with WVA-144 and also between GCIPL and the mf103-pattern (Fig. 4).
Table 2
Demographic results from the structure–function analysis group
Group
Controls (n = 15)
PPG (n = 2)
NTG (n = 9)
HTG (n = 5)
p value
Age (years)
Mean ± SD
49.3
± 7.3
53.0
± 25.5
61.1
± 16.5
61.0
± 6.3
0.102
Gender (M/F)
4/11
1/1
6/3
4/1
 
MD (dB)
Mean (95% CI)
0.12
(−0.8/1.0)
1.7
(3.0/5.9)
3.4
(2.2/5.4)
5.5
(3.8/17.3)
0.007*
MD 15°
Mean (95% CI)
−0.07
(−1.1/0.5)
1.8
(0.6/2.2)
1.5
(0.6/2.2)
3.6
(1.0/3.7)
0.028*
GCIPL (µm)
Mean ± SD
80.9
±4.8
82.8
±5.1
68.4
±6.4
61.2
±5.2
0.000*
WVA-144
323
±112
164
± 97.1
184
± 141
145
± 60.4
0.038*
Legend: SD standard deviation; PPG pre-perimetric glaucoma; NTG normal-tension glaucoma; HTG high-tension glaucoma; 95% CI 95% confidence interval; MD overall mean defect in decibels; MD 15° average mean defect from the central 15°; GCIPL ganglion cell–inner plexiform layer in µm; WVA-144 wavelet variance analysis frequency level 144 Hz; * p value <0.05 was considered statistically significant
For the individual types of glaucoma, Table 3 demonstrates the consistent significant group differences seen in WVA-144 and in GCIPL thickness, but not in the central mf103-pattern sensitivity. Within the central 15° (19 points), 56.2% of POAG patients had no defect, while 25% (n = 4) had 3 or more points with p < 0.05, the remainder showing only 1 or 2 individual points of probability <0.05.
Table 3
Mixed effects model results from group comparison between WVA-144, GCIPL and MS
 
Difference of means
95% CI
p value
WVA-144
   
 PPG
−0.28
−0.52/−0.05
0.0180
 HTG
−0.31
−0.50/−0.12
0.0018
 NTG
−0.19
−0.37/0.00
0.0460
GCIPL
   
 PPG
−10.23
−18.9/−1.5
0.0220
 HTG
−11.05
−18.0/−4.0
0.0026
 NTG
−9.76
−16.5/−2.9
0.0058
MS
   
 PPG
−154.6
−586.4/276.9
0.4713
 HTG
−217.5
−520.3/85.3
0.4951
 NTG
−96.6
−381.2/188.0
0.1536
Results are presented as differences of means between study groups with corresponding 95% confidence intervals (95% CI) and p values
For the 95% CI, the 2.5th percentile and the 97.5th percentile are given
WVA-144 wavelet variance analysis frequency level 144 Hz (transformed into logarithm units); GCIPL ganglion cell–inner plexiform layer (µm); MS mean sensitivity (decibels transformed into linear units)
Analyzing the data on a more focal basis, that is correlating individual GCIPL sectors from the thickness map to the corresponding hexagons from the mf103-pattern and 2F-mfERG, we found a statistically significant relationship between most of GCIPL sectors and WVA-144, with exception of the most temporal inferior sector (Fig. 5).
As glaucoma incidence increases with age, we compared diagnostic performance in an age-matched subgroup based on the mean of tenfold cross-validated penalized logistic regression.
The mean AUCs of Age, GCIPL, MS and WVA-144, as well as for all examinations combined, were calculated. Age had a small mean AUC of 0.44, and MS and WVA-144 had moderate AUCs (0.65, 0.64). GCIPL had the highest predictive performance (0.880), outperformed only by the combination of all exams together GCIPL + MS + WVA-144 (AUC: 0.945), showing that MS and WVA-144 contribute to glaucoma diagnosis with additional information.

Discussion

In electrophysiology, electrical activity from retinal cells is recorded from the corneal surface. The resulting waveforms are traditionally analyzed in amplitude and latency (peak time) [44, 45]. In the global flash mfERG response, the root-mean-square (RMS) method is often applied [8, 46]. Decomposition of the mfERG signal into its time–frequency components, applying DWT analysis, offers an interesting objective and automated approach to further explore electrophysiological responses. In the full-field ERG, DWT has been shown to reveal additional contributions as shown by Gauvin et al. [26]. For example, Gauvin et al. [25, 47] were able to show a strong correlation between DWT descriptors and different parts of the full-field ERG such as the a-wave, the b-wave and also oscillatory potentials at 80 and at 160 Hz [26]. Thus, DWT offers a promising tool, in particular as the response to the global flash mfERG waveform does not show typical a- or b-waves due to the adaptive properties of the stimulus response.
In our study, we could differentiate patients with early, rather than advanced, POAG (including PPG) from controls using DWT analysis on the two-flash mfERG response. In the present study, we were able to directly compare electrophysiological findings not only to visual fields, but also to the OCT. When our DWT findings were compared to a waveform analysis that is more frequently applied by clinicians, the RMS [2, 3, 30], we found their performance to be similar, although DWT was statistically better in distinguishing POAG from control. WVA could be more powerful because RMS is based on overall variance, whereas WVA distinguishes between frequency levels. If there is more variance in a certain level, this could be detected in WVA, but blurred in RMS. Thus, the application of DWT to the 2F-mfERG is a valid additional diagnostic tool in glaucoma. It offers an objective approach to the electrical signal that takes into consideration not only time–frequency but also variance within the response. This is an advantage over time-domain analysis, where variance of amplitude can be analyzed at different frequencies following bandpass filtering.
In our population, DWT analysis of the two-flash mfERG using Daubechies S8 was able to differentiate glaucoma from control in OAG with an AUC of 0.64. While sensitivity and specificity appear lower than reported by Miguel-Jiménez et al., our analysis was restricted to the central retina in all subjects included. In addition, our patient group consisted of primarily early glaucoma with an overall average MD of 3.6 dB. Within the central 15° analyzed, mean MD was 2.7 dB in HTG, 1.3 dB in NTG and 1.0 dB in PPG (Table 2, [48]). In advanced glaucoma, Miguel-Jiménez et al. [28] were able to identify signal characteristics in a two-flash mfERG that could differentiate OAG from control (DWT, mother wavelet Bior 3.1). In addition to the 50 patients analyzed, they demonstrated a good topographic correlation between the two-flash mfERG DWT findings and the Humphrey fields in two patients with advanced OAG [28]. In a subsequent paper, using discrete wavelet packet decomposition analysis the same group focused on the induced component (60–90 ms) of their global flash mfERG. Analysis of the visual field sectors affected by glaucoma in 25 OAG patients found a sensitivity of 0.81 and a specificity of 0.73 [29]. Further, using CWT (Morlet waveform) in OAG and healthy subjects, sensitivity could be increased to 0.894 and specificity to 0.844, again only including sectors identified as glaucomatous and again looking at the induced component [27].
Some of the differences between our findings and those of Miguel-Jiménez et al. [2729] may be related not only to the different glaucoma stages analyzed, but also to the different selection of areas analyzed: only areas with field defect versus the central 15°, irrespective of the location of the field defect. Differences may also result from the use of different visual field patterns. In contrast to Miguel-Jiménez et al. [2729], we did not average hexagons to fit single visual field locations. Our customized visual field pattern (mf103-pattern) allowed direct spatial correlation between visual field sensitivity and the mfERG response.
In our early POAG patients, the WVA-144 was able to identify glaucomatous alterations in the central 15° even when the visual field could not. Similar findings have been observed in outer retinal disease, such as retinitis pigmentosa [49]. Our findings are in agreement with Takagi et al. [1], who suggested that a more sensitive functional test should be able to detect glaucoma earlier than the visual field as GCIPL thickness may be already compromised in glaucoma even when no defect is identified by standard automated perimetry.
Structure function investigation showed a significant positive relationship between the WVA-144 from the 2F-mfERG and the GCIPL, between the WVA-144 and the mf103-pattern, and also between GCIPL and the mf103-pattern. A detailed analysis demonstrated a significant relationship between WVA-144 and the GCIPL thickness map sectors except for the temporal inferior sector. It has been demonstrated that superior visual field defects (inferior retina) are more common in glaucoma than inferior field defects [5052]. In our POAG patients, there was no predominance of superior or inferior field defects. The lack of correlation in the temporal inferior quadrant is an interesting finding. Follow-up studies in a larger population are needed to assess whether that is an incidental finding.
AUC analysis with age-matched groups demonstrated that GCIPL alone had a high predictive performance, which improved further when MS and WVA-144 were added. Thus, we agree with a recent editorial from the American Academy of Ophthalmology, in which Medeiros and Tatham [53] suggest that for a correct early diagnosis it is better to combine different examinations rather than to keep searching for a single parameter.
Here, the most significant difference between POAG and controls was localized at 144 HZ, which coincides with the frequency range of the oscillatory potentials (OPs) [54]. OPs of full field are thought to reflect activity between bipolar, amacrine and ganglion cells and are generated within the IPL [54, 55]. In experimental glaucoma, swelling of ganglion cell dendrites in the IPL has been described as an earliest manifestation of glaucoma [56]. Thus, changes at the OP frequency level may be expected in glaucoma.
Indeed, OPs from the mfERG have previously been described as a sensitive measure of glaucomatous dysfunction, especially with use of a slow sequence mfERG [32, 57]. In the monkey and in humans, Rangaswamy et al. [57] found glaucoma-induced changes in fast OPs at about 143 Hz in all areas studied, even when field defects were only moderate (MD between −5 and −10 dB). These fast OPs showed a good positive correlation to the estimated ganglion cell density. Slow OPs at about 77 Hz were affected primarily in the center, when MD was larger than—10 dB [57]. In primates with experimental glaucoma, a two-flash mfERG (MOFOFO) showed changes primarily in a low-frequency component contributing to DC/IC1 but also in a high-frequency component at a little below 150 Hz [8].
Fortune et al. [58] could show that OPs in glaucoma are diminished in the first induced component of the one global flash mfERG, while in the two-flash mfERG Palmowski-Wolfe et al. [59] showed these to be primarily affected in IC2. Compared to standard mfERG, OPs from the global flash mfERG are more sensitive to detect glaucomatous dysfunction in human and experimental glaucoma [32].
In future studies, the performance of the DWT may be increased further by including various descriptors in the analysis in addition to only studying the most sensitive descriptor. We have not addressed the possibility that different descriptors are affected differently in HTG, NTG and PPG. Assessing ratios between different decomposition levels could reduce outside influences (such as noise) and thereby the coefficient of variation. This may further improve separation between patient groups and control. Another improvement might be to include the Hölder exponent, as proposed by Gauvin [25, 49] which characterizes the complexity of the waveform and was able to identify electrophysiological changes in the ERG prior to changes in the visual field in retinitis pigmentosa [49]. In addition, the use of scalograms [24, 25, 47] that summarize the particular contribution of individual coefficients over time may improve understanding of disease.
To date, electrophysiological measures are not incorporated in the clinical routine diagnosis of glaucoma, although they have been shown to be sensitive in this disease. This may change with the automatization of DWT analysis. DWT offers an exciting new approach with the potential of development of automatic analysis procedures and with visualization of results (e.g., scalograms) and thus facilitation of interpretation by clinicians.
A limitation in our study could be the imbalance in age between glaucoma and control groups. Therefore, all calculations were adjusted for age and AUC calculations were performed in perfectly age-matched groups. Another imbalance was seen in the gender distribution which we took into account in our statistical analysis. Nonetheless, previous studies on GCIPL could not identify gender differences in this retinal layer [60] (GCIPL). Cohn et al. observed that in the visual field, sensitivity of the left and right hemifield may differ in females but not males. When sensitivity values were averaged over the field, as done in our study, no gender differences were found [61].
In conclusion, we demonstrated that the application of wavelet analysis to the 2F-mfERG recordings improve glaucoma diagnosis when added to GCIPL and visual field analysis. It may improve follow-up as significant differences were seen between control and glaucoma but not in the visual fields when results were adjusted for age and gender. This confirms suggestions of early central retinal structure and function involvement even prior to central field defects.

Acknowledgements

This study was presented in parts as an oral presentation at ISCEV 2016.

Funding

This study was supported by AMP: Swiss National Fund (SNF NMS 1823) and LHW Stiftung Lichtenstein. Both sponsors had no role in the design or conduct of this research.

Compliance with ethical standards

Conflict of interest

All authors certify that they have no affiliations with or involvement in any organization or entity with any financial interest (such as honoraria; educational grants; participation in speakers’ bureaus; membership, employment, consultancies, stock ownership or other equity interest; and expert testimony or patent-licensing arrangements) or non-financial interest (such as personal or professional relationships, affiliations, knowledge or beliefs) in the subject matter or materials discussed in this manuscript.

Human and animal rights

All procedures performed in this study involving human participants were approved and in accordance with the ethical standards of the Ethics Committee of the University of Basel, the national research committee and with the 1964 Helsinki declaration and its later amendments or comparable ethical standards. This article does not contain any studies with animals performed by any of the authors.

Ethical approval

All procedures performed in studies involving human participants were in accordance with the ethical standards of the institutional and/or national research committee and with the 1964 Helsinki Declaration and its later amendments or comparable ethical standards.
Informed consent was obtained from all individual participants included in the study.
Open AccessThis article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://​creativecommons.​org/​licenses/​by/​4.​0/​), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made.

Unsere Produktempfehlungen

e.Med Interdisziplinär

Kombi-Abonnement

Für Ihren Erfolg in Klinik und Praxis - Die beste Hilfe in Ihrem Arbeitsalltag

Mit e.Med Interdisziplinär erhalten Sie Zugang zu allen CME-Fortbildungen und Fachzeitschriften auf SpringerMedizin.de.

Literatur
2.
Zurück zum Zitat Kramer SA, Ledolter AA, Todorova MG, Schotzau A, Orgul S, Palmowski-Wolfe AM (2013) The 2-global flash mfERG in glaucoma: attempting to increase sensitivity by reducing the focal flash luminance and changing filter settings. Doc Ophthalmol Adv Ophthalmol 126(1):57–67. doi:10.1007/s10633-012-9360-z CrossRef Kramer SA, Ledolter AA, Todorova MG, Schotzau A, Orgul S, Palmowski-Wolfe AM (2013) The 2-global flash mfERG in glaucoma: attempting to increase sensitivity by reducing the focal flash luminance and changing filter settings. Doc Ophthalmol Adv Ophthalmol 126(1):57–67. doi:10.​1007/​s10633-012-9360-z CrossRef
3.
Zurück zum Zitat Ledolter AA, Monhart M, Schoetzau A, Todorova MG, Palmowski-Wolfe AM (2015) Structural and functional changes in glaucoma: comparing the two-flash multifocal electroretinogram to optical coherence tomography and visual fields. Doc Ophthalmol Adv Ophthalmol 130(3):197–209. doi:10.1007/s10633-015-9482-1 CrossRef Ledolter AA, Monhart M, Schoetzau A, Todorova MG, Palmowski-Wolfe AM (2015) Structural and functional changes in glaucoma: comparing the two-flash multifocal electroretinogram to optical coherence tomography and visual fields. Doc Ophthalmol Adv Ophthalmol 130(3):197–209. doi:10.​1007/​s10633-015-9482-1 CrossRef
4.
Zurück zum Zitat Palmowski AM, Allgayer R, Heinemann-Vernaleken B, Ruprecht KW (2002) Multifocal electroretinogram with a multiflash stimulation technique in open-angle glaucoma. Ophthalmic Res 34(2):83–89. doi:10.1159/000048333 CrossRefPubMed Palmowski AM, Allgayer R, Heinemann-Vernaleken B, Ruprecht KW (2002) Multifocal electroretinogram with a multiflash stimulation technique in open-angle glaucoma. Ophthalmic Res 34(2):83–89. doi:10.​1159/​000048333 CrossRefPubMed
5.
Zurück zum Zitat Palmowski AM, Ruprecht KW (2004) Follow up in open angle glaucoma. A comparison of static perimetry and the fast stimulation mfERG. Multifocal ERG follow up in open angle glaucoma. Doc Ophthalmol Adv Ophthalmol 108(1):55–60CrossRef Palmowski AM, Ruprecht KW (2004) Follow up in open angle glaucoma. A comparison of static perimetry and the fast stimulation mfERG. Multifocal ERG follow up in open angle glaucoma. Doc Ophthalmol Adv Ophthalmol 108(1):55–60CrossRef
6.
Zurück zum Zitat Palmowski-Wolfe AM, Todorova MG, Orguel S, Flammer J, Brigell M (2007) The ‘two global flash’ mfERG in high and normal tension primary open-angle glaucoma. Doc Ophthalmol Adv Ophthalmol 114(1):9–19. doi:10.1007/s10633-006-9033-x CrossRef Palmowski-Wolfe AM, Todorova MG, Orguel S, Flammer J, Brigell M (2007) The ‘two global flash’ mfERG in high and normal tension primary open-angle glaucoma. Doc Ophthalmol Adv Ophthalmol 114(1):9–19. doi:10.​1007/​s10633-006-9033-x CrossRef
7.
Zurück zum Zitat Shimada Y, Bearse MA Jr, Sutter EE (2005) Multifocal electroretinograms combined with periodic flashes: direct responses and induced components. Graefe’s Arch Clin Expe Ophthalmol Albrecht von Graefes Archiv fur klinische und experimentelle Ophthalmologie 243(2):132–141. doi:10.1007/s00417-004-1072-y CrossRef Shimada Y, Bearse MA Jr, Sutter EE (2005) Multifocal electroretinograms combined with periodic flashes: direct responses and induced components. Graefe’s Arch Clin Expe Ophthalmol Albrecht von Graefes Archiv fur klinische und experimentelle Ophthalmologie 243(2):132–141. doi:10.​1007/​s00417-004-1072-y CrossRef
9.
Zurück zum Zitat Hood DC, Greenstein V, Frishman L, Holopigian K, Viswanathan S, Seiple W, Ahmed J, Robson JG (1999) Identifying inner retinal contributions to the human multifocal ERG. Vis Res 39(13):2285–2291CrossRefPubMed Hood DC, Greenstein V, Frishman L, Holopigian K, Viswanathan S, Seiple W, Ahmed J, Robson JG (1999) Identifying inner retinal contributions to the human multifocal ERG. Vis Res 39(13):2285–2291CrossRefPubMed
10.
Zurück zum Zitat Hood DC, Frishman LJ, Viswanathan S, Robson JG, Ahmed J (1999) Evidence for a ganglion cell contribution to the primate electroretinogram (ERG): effects of TTX on the multifocal ERG in macaque. Vis Neurosci 16(3):411–416CrossRefPubMed Hood DC, Frishman LJ, Viswanathan S, Robson JG, Ahmed J (1999) Evidence for a ganglion cell contribution to the primate electroretinogram (ERG): effects of TTX on the multifocal ERG in macaque. Vis Neurosci 16(3):411–416CrossRefPubMed
11.
Zurück zum Zitat Hood DC, Frishman LJ, Saszik S, Viswanathan S (2002) Retinal origins of the primate multifocal ERG: implications for the human response. Invest Ophthalmol Vis Sci 43(5):1673–1685PubMed Hood DC, Frishman LJ, Saszik S, Viswanathan S (2002) Retinal origins of the primate multifocal ERG: implications for the human response. Invest Ophthalmol Vis Sci 43(5):1673–1685PubMed
12.
Zurück zum Zitat Hare WA, Ton H (2002) Effects of APB, PDA, and TTX on ERG responses recorded using both multifocal and conventional methods in monkey. Effects of APB, PDA, and TTX on monkey ERG responses. Doc Ophthalmol Adv Ophthalmol 105(2):189–222CrossRef Hare WA, Ton H (2002) Effects of APB, PDA, and TTX on ERG responses recorded using both multifocal and conventional methods in monkey. Effects of APB, PDA, and TTX on monkey ERG responses. Doc Ophthalmol Adv Ophthalmol 105(2):189–222CrossRef
13.
Zurück zum Zitat Hood DC, Bach M, Brigell M, Keating D, Kondo M, Lyons JS, Marmor MF, McCulloch DL, Palmowski-Wolfe AM, International Society For Clinical Electrophysiology of V (2012) ISCEV standard for clinical multifocal electroretinography (mfERG) (2011 edition). Doc Ophthalmol Adv Ophthalmol 124(1):1–13. doi:10.1007/s10633-011-9296-8 CrossRef Hood DC, Bach M, Brigell M, Keating D, Kondo M, Lyons JS, Marmor MF, McCulloch DL, Palmowski-Wolfe AM, International Society For Clinical Electrophysiology of V (2012) ISCEV standard for clinical multifocal electroretinography (mfERG) (2011 edition). Doc Ophthalmol Adv Ophthalmol 124(1):1–13. doi:10.​1007/​s10633-011-9296-8 CrossRef
14.
Zurück zum Zitat Bach M, Meigen T (1999) Do’s and don’ts in Fourier analysis of steady-state potentials. Doc Ophthalmol Adv Ophthalmol 99(1):69–82CrossRef Bach M, Meigen T (1999) Do’s and don’ts in Fourier analysis of steady-state potentials. Doc Ophthalmol Adv Ophthalmol 99(1):69–82CrossRef
15.
Zurück zum Zitat Mazinani BA, Repas T, Weinberger AW, Vobig MA, Walter P (2007) Amplitude calculation in multifocal ERG: comparison of repeatability in 30 Hz flicker and first order kernel stimulation. Graefe’s Arch Clin Exp Ophthalmol Albrecht von Graefes Archiv fur klinische und experimentelle Ophthalmologie 245(3):338–344. doi:10.1007/s00417-006-0423-2 CrossRef Mazinani BA, Repas T, Weinberger AW, Vobig MA, Walter P (2007) Amplitude calculation in multifocal ERG: comparison of repeatability in 30 Hz flicker and first order kernel stimulation. Graefe’s Arch Clin Exp Ophthalmol Albrecht von Graefes Archiv fur klinische und experimentelle Ophthalmologie 245(3):338–344. doi:10.​1007/​s00417-006-0423-2 CrossRef
16.
Zurück zum Zitat Todorova MG, Palmowski-Wolfe AM, Orguel S, Flammer J (2006) 30 Hz-flicker mfERG in primary open-angle glaucoma patients: 30 Hz-flicker-mfERG in POAG. Doc Ophthalmol Adv Ophthalmol 113(1):11–20. doi:10.1007/s10633-006-9008-y CrossRef Todorova MG, Palmowski-Wolfe AM, Orguel S, Flammer J (2006) 30 Hz-flicker mfERG in primary open-angle glaucoma patients: 30 Hz-flicker-mfERG in POAG. Doc Ophthalmol Adv Ophthalmol 113(1):11–20. doi:10.​1007/​s10633-006-9008-y CrossRef
18.
Zurück zum Zitat Kilby J, Hosseini HG (2006) Extracting effective features of SEMG using continuous wavelet transform. In: Conference proceedings: annual international conference of the IEEE engineering in medicine and biology society IEEE engineering in medicine and biology society conference 1:1704–1707. doi:10.1109/iembs.2006.260064 Kilby J, Hosseini HG (2006) Extracting effective features of SEMG using continuous wavelet transform. In: Conference proceedings: annual international conference of the IEEE engineering in medicine and biology society IEEE engineering in medicine and biology society conference 1:1704–1707. doi:10.​1109/​iembs.​2006.​260064
22.
Zurück zum Zitat Kundra H, Park JC, McAnany JJ (2016) Comparison of photopic negative response measurements in the time and time-frequency domains. Doc Opthalmol 133(2):91–98CrossRef Kundra H, Park JC, McAnany JJ (2016) Comparison of photopic negative response measurements in the time and time-frequency domains. Doc Opthalmol 133(2):91–98CrossRef
26.
Zurück zum Zitat Gauvin M, Dorfman AL, Trang N, Gauthier M, Little JM, Lina JM, Lachapelle P (2016) Assessing the contribution of the oscillatory potentials to the genesis of the photopic ERG with the discrete wavelet transform. Biomed Res 2016:2790194. doi:10.1155/2016/2790194 Gauvin M, Dorfman AL, Trang N, Gauthier M, Little JM, Lina JM, Lachapelle P (2016) Assessing the contribution of the oscillatory potentials to the genesis of the photopic ERG with the discrete wavelet transform. Biomed Res 2016:2790194. doi:10.​1155/​2016/​2790194
27.
Zurück zum Zitat Miguel-Jimenez JM, Blanco R, De-Santiago L, Fernandez A, Rodriguez-Ascariz JM, Barea R, Martin-Sanchez JL, Amo C, Sanchez-Morla E, Boquete L (2015) Continuous-wavelet-transform analysis of the multifocal ERG waveform in glaucoma diagnosis. Med Biol Eng Comput. doi:10.1007/s11517-015-1287-6 PubMed Miguel-Jimenez JM, Blanco R, De-Santiago L, Fernandez A, Rodriguez-Ascariz JM, Barea R, Martin-Sanchez JL, Amo C, Sanchez-Morla E, Boquete L (2015) Continuous-wavelet-transform analysis of the multifocal ERG waveform in glaucoma diagnosis. Med Biol Eng Comput. doi:10.​1007/​s11517-015-1287-6 PubMed
30.
Zurück zum Zitat Ledolter AA, Kramer SA, Todorova MG, Schotzau A, Palmowski-Wolfe AM (2013) The effect of filtering on the two-global-flash mfERG: identifying the optimal range of frequency for detecting glaucomatous retinal dysfunction. Doc Ophthalmol Adv Ophthalmol 126(2):117–123. doi:10.1007/s10633-012-9364-8 CrossRef Ledolter AA, Kramer SA, Todorova MG, Schotzau A, Palmowski-Wolfe AM (2013) The effect of filtering on the two-global-flash mfERG: identifying the optimal range of frequency for detecting glaucomatous retinal dysfunction. Doc Ophthalmol Adv Ophthalmol 126(2):117–123. doi:10.​1007/​s10633-012-9364-8 CrossRef
31.
Zurück zum Zitat Shimada Y, Li Y, Bearse MA Jr, Sutter EE, Fung W (2001) Assessment of early retinal changes in diabetes using a new multifocal ERG protocol. Br J Ophthalmol 85(4):414–419CrossRefPubMedPubMedCentral Shimada Y, Li Y, Bearse MA Jr, Sutter EE, Fung W (2001) Assessment of early retinal changes in diabetes using a new multifocal ERG protocol. Br J Ophthalmol 85(4):414–419CrossRefPubMedPubMedCentral
32.
Zurück zum Zitat Fortune B, Wang L, Bui BV, Cull G, Dong J, Cioffi GA (2003) Local ganglion cell contributions to the macaque electroretinogram revealed by experimental nerve fiber layer bundle defect. Invest Ophthalmol Vis Sci 44(10):4567–4579CrossRefPubMed Fortune B, Wang L, Bui BV, Cull G, Dong J, Cioffi GA (2003) Local ganglion cell contributions to the macaque electroretinogram revealed by experimental nerve fiber layer bundle defect. Invest Ophthalmol Vis Sci 44(10):4567–4579CrossRefPubMed
33.
34.
Zurück zum Zitat Walden AT (2001) Wavelet analysis of discrete time series. In: Casacuberta C, Miró-Roig RM, Verdera J, Xambó-Descamps S (eds) European congress of mathematics: Barcelona, July 10–14, 2000 Volume II. Birkhäuser Basel, Basel, pp 627–641. doi:10.1007/978-3-0348-8266-8_56 CrossRef Walden AT (2001) Wavelet analysis of discrete time series. In: Casacuberta C, Miró-Roig RM, Verdera J, Xambó-Descamps S (eds) European congress of mathematics: Barcelona, July 10–14, 2000 Volume II. Birkhäuser Basel, Basel, pp 627–641. doi:10.​1007/​978-3-0348-8266-8_​56 CrossRef
39.
Zurück zum Zitat R: A Language and Environment for Statistical Computing (2016) 3.0.2 edn. R foundation for statistical computing, Vienna, Austria R: A Language and Environment for Statistical Computing (2016) 3.0.2 edn. R foundation for statistical computing, Vienna, Austria
40.
Zurück zum Zitat Harrell F (2001) Regression modeling strategies with applications to linear models, logistic regression, and survival analysis. Springer series in statistics, 1st edn. Springer, New York. doi:10.1007/978-1-4757-3462-1 Harrell F (2001) Regression modeling strategies with applications to linear models, logistic regression, and survival analysis. Springer series in statistics, 1st edn. Springer, New York. doi:10.​1007/​978-1-4757-3462-1
42.
Zurück zum Zitat Eugster MJA, Hothorn T, Leisch F (2008) Exploratory and inferential analysis of benchmark experiments (trans: Statistics Do). Technical Report, vol 30. Ludwigs-Maximilians-Universitat Munchen Eugster MJA, Hothorn T, Leisch F (2008) Exploratory and inferential analysis of benchmark experiments (trans: Statistics Do). Technical Report, vol 30. Ludwigs-Maximilians-Universitat Munchen
43.
Zurück zum Zitat DeLong ER, DeLong DM, Clarke-Pearson DL (1988) Comparing the areas under two or more correlated receiver operating characteristic curves: a nonparametric approach. Biometrics 44(3):837–845CrossRefPubMed DeLong ER, DeLong DM, Clarke-Pearson DL (1988) Comparing the areas under two or more correlated receiver operating characteristic curves: a nonparametric approach. Biometrics 44(3):837–845CrossRefPubMed
44.
Zurück zum Zitat Hood DC, Bach M, Brigell M, Keating D, Kondo M, Lyons JS, Marmor MF, McCulloch DL, Palmowski-Wolfe AM (2012) ISCEV standard for clinical multifocal electroretinography. Doc Ophthalmol Adv Ophthalmol 124(1):1–13. doi:10.1007/s10633-011-9296-8 CrossRef Hood DC, Bach M, Brigell M, Keating D, Kondo M, Lyons JS, Marmor MF, McCulloch DL, Palmowski-Wolfe AM (2012) ISCEV standard for clinical multifocal electroretinography. Doc Ophthalmol Adv Ophthalmol 124(1):1–13. doi:10.​1007/​s10633-011-9296-8 CrossRef
45.
Zurück zum Zitat McCulloch DL, Marmor MF, Brigell MG, Hamilton R, Holder GE, Tzekov R, Bach M (2015) ISCEV standard for full-field clinical electroretinography (2015 update). Doc Ophthalmol Adv Ophthalmol 130(1):1–12. doi:10.1007/s10633-014-9473-7 CrossRef McCulloch DL, Marmor MF, Brigell MG, Hamilton R, Holder GE, Tzekov R, Bach M (2015) ISCEV standard for full-field clinical electroretinography (2015 update). Doc Ophthalmol Adv Ophthalmol 130(1):1–12. doi:10.​1007/​s10633-014-9473-7 CrossRef
47.
Zurück zum Zitat Gauvin M, Sustar M, Little JM, Brecelj J, Lina JM, Lachapelle P (2017) Quantifying the ON and OFF contributions to the flash ERG with the discrete wavelet transform. Biomed Res Int 6(1):3. doi:10.1167/tvst.6.1.3 Gauvin M, Sustar M, Little JM, Brecelj J, Lina JM, Lachapelle P (2017) Quantifying the ON and OFF contributions to the flash ERG with the discrete wavelet transform. Biomed Res Int 6(1):3. doi:10.​1167/​tvst.​6.​1.​3
49.
Zurück zum Zitat Gauvin M, Chakor H, Koenekoop RK, Little JM, Lina JM, Lachapelle P (2016) Witnessing the first sign of retinitis pigmentosa onset in the allegedly normal eye of a case of unilateral RP: a 30-year follow-up. Doc Ophthalmol Adv Ophthalmol 132(3):213–229. doi:10.1007/s10633-016-9537-y CrossRef Gauvin M, Chakor H, Koenekoop RK, Little JM, Lina JM, Lachapelle P (2016) Witnessing the first sign of retinitis pigmentosa onset in the allegedly normal eye of a case of unilateral RP: a 30-year follow-up. Doc Ophthalmol Adv Ophthalmol 132(3):213–229. doi:10.​1007/​s10633-016-9537-y CrossRef
50.
Zurück zum Zitat Schiefer U, Papageorgiou E, Sample PA, Pascual JP, Selig B, Krapp E, Paetzold J (2010) Spatial pattern of glaucomatous visual field loss obtained with regionally condensed stimulus arrangements. Invest Ophthalmol Vis Sci 51(11):5685–5689. doi:10.1167/iovs.09-5067 CrossRefPubMedPubMedCentral Schiefer U, Papageorgiou E, Sample PA, Pascual JP, Selig B, Krapp E, Paetzold J (2010) Spatial pattern of glaucomatous visual field loss obtained with regionally condensed stimulus arrangements. Invest Ophthalmol Vis Sci 51(11):5685–5689. doi:10.​1167/​iovs.​09-5067 CrossRefPubMedPubMedCentral
52.
Zurück zum Zitat Nicholas SP, Werner EB (1980) Location of early glaucomatous visual field defects. Can J Ophthalmol J Can D’ophtalmol 15(3):131–133 Nicholas SP, Werner EB (1980) Location of early glaucomatous visual field defects. Can J Ophthalmol J Can D’ophtalmol 15(3):131–133
54.
Zurück zum Zitat John R, Heckenlively GBA (eds) (2006) Principles and practice of clinical electrophysiology of vision, 2nd edn. The MIT Press, Cambridge John R, Heckenlively GBA (eds) (2006) Principles and practice of clinical electrophysiology of vision, 2nd edn. The MIT Press, Cambridge
55.
Zurück zum Zitat Wachtmeister L (1998) Oscillatory potentials in the retina: what do they reveal. Prog Retinal Eye Res 17(4):485–521CrossRef Wachtmeister L (1998) Oscillatory potentials in the retina: what do they reveal. Prog Retinal Eye Res 17(4):485–521CrossRef
56.
Zurück zum Zitat Weber AJ, Kaufman PL, Hubbard WC (1998) Morphology of single ganglion cells in the glaucomatous primate retina. Invest Ophthalmol Vis Sci 39(12):2304–2320PubMed Weber AJ, Kaufman PL, Hubbard WC (1998) Morphology of single ganglion cells in the glaucomatous primate retina. Invest Ophthalmol Vis Sci 39(12):2304–2320PubMed
58.
Zurück zum Zitat Fortune B, Bearse MA Jr, Cioffi GA, Johnson CA (2002) Selective loss of an oscillatory component from temporal retinal multifocal ERG responses in glaucoma. Invest Ophthalmol Vis Sci 43(8):2638–2647PubMed Fortune B, Bearse MA Jr, Cioffi GA, Johnson CA (2002) Selective loss of an oscillatory component from temporal retinal multifocal ERG responses in glaucoma. Invest Ophthalmol Vis Sci 43(8):2638–2647PubMed
59.
Zurück zum Zitat Palmowski-Wolfe AMOS, Todorova MG (2011) Multifocal oscillatory potentials in the ‘two global flash’ mfERG in high and normal tension primary open-angle glaucoma. J Clin Exp Ophthalmol 2:167. doi:10.4172/2155-9570.1000167 CrossRef Palmowski-Wolfe AMOS, Todorova MG (2011) Multifocal oscillatory potentials in the ‘two global flash’ mfERG in high and normal tension primary open-angle glaucoma. J Clin Exp Ophthalmol 2:167. doi:10.​4172/​2155-9570.​1000167 CrossRef
61.
Zurück zum Zitat Cohn H, DeAgostini M, Aron-Rosa D, Laloum L, Boller F (1994) Sex differences in the left and right hemifields of normal subjects with computerised static perimetry. Br J Ophthalmol 78(11):837–841CrossRefPubMedPubMedCentral Cohn H, DeAgostini M, Aron-Rosa D, Laloum L, Boller F (1994) Sex differences in the left and right hemifields of normal subjects with computerised static perimetry. Br J Ophthalmol 78(11):837–841CrossRefPubMedPubMedCentral
Metadaten
Titel
Wavelet decomposition analysis in the two-flash multifocal ERG in early glaucoma: a comparison to ganglion cell analysis and visual field
verfasst von
Livia M Brandao
Matthias Monhart
Andreas Schötzau
Anna A Ledolter
Anja M Palmowski-Wolfe
Publikationsdatum
07.06.2017
Verlag
Springer Berlin Heidelberg
Erschienen in
Documenta Ophthalmologica / Ausgabe 1/2017
Print ISSN: 0012-4486
Elektronische ISSN: 1573-2622
DOI
https://doi.org/10.1007/s10633-017-9593-y

Weitere Artikel der Ausgabe 1/2017

Documenta Ophthalmologica 1/2017 Zur Ausgabe

Neu im Fachgebiet Augenheilkunde

Update Augenheilkunde

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.