What are the limitations of methods to measure carbon monoxide in biological samples?

verfasst von: Stefania Oliverio, Vincent Varlet

Erschienen in: Forensic Toxicology | Ausgabe 1/2020

Einloggen, um Zugang zu erhalten

Abstract

Purpose

Carbon monoxide (CO) is one of the most important toxic gases in the atmosphere. Its high affinity for hemoglobin has made carboxyhemoglobin (COHb) the most appropriate biomarker for CO poisoning. COHb is measured using spectrophotometric (ultraviolet-spectrophotometry, CO-oximetry) or gas chromatographic (GC) methods combined with flame ionization or mass spectrometry (MS) detectors. However, inconsistencies in many cases have been reported between measured values and reported symptoms, raising doubts as to the suitability of COHb as a biomarker and the accuracy and reliability of its measurement methods. Therefore, we aimed to review the accuracy of current methods used to measure CO and to determine their sources of error and their effects on the interpretation process.

Methods

A detailed search of PubMed was performed in November 2018 using relevant keywords. After exclusion criteria were applied, 46 articles out of 191 initial hits were carefully reviewed.

Results

While optical methods are highly influenced by changes in blood quality due to degradation of samples during storage, GC methods are less affected. However, measurement of COHb does not quantify free CO, which is mainly responsible for toxicity mechanisms other than hypoxia, such as inhibition of hemoproteins, thus underestimating the true CO burden. Therefore, measurement of COHb is not sufficiently accurate for diagnosis of CO poisoning.

Conclusions

An alternative biomarker is needed, such as determining the total amount of CO in blood. Although further research is required, we recommend that toxicologists consider all sources of error that can alter COHb concentrations, and in more challenging cases, they should use GC–MS methods to confirm the results obtained by spectrophotometry.
Literatur
10.
11.
Zurück zum Zitat Penney DG (2007) Carbon monoxide poisoning, 1st edn. CRC Press, Boca Raton CrossRef Penney DG (2007) Carbon monoxide poisoning, 1st edn. CRC Press, Boca Raton CrossRef
14.
Zurück zum Zitat Vreman HJ, Stevenson DK, Oh W, Fanaroff AA, Wright LL, Lemons JA, Wright E, Shankaran S, Tyson JE, Korones SB (1994) Semiportable electrochemical instrument for determining carbon monoxide in breath. Clin Chem 40:1927–1933 (PMID: 7923774) PubMed Vreman HJ, Stevenson DK, Oh W, Fanaroff AA, Wright LL, Lemons JA, Wright E, Shankaran S, Tyson JE, Korones SB (1994) Semiportable electrochemical instrument for determining carbon monoxide in breath. Clin Chem 40:1927–1933 (PMID: 7923774) PubMed
16.
Zurück zum Zitat Ramieri A Jr, Jatlow P, Seligson D (1974) New method for rapid determination of carboxyhemoglobin by use of double-wavelength spectrophotometry. Clin Chem 20:278–281 (PMID: 4813007) PubMed Ramieri A Jr, Jatlow P, Seligson D (1974) New method for rapid determination of carboxyhemoglobin by use of double-wavelength spectrophotometry. Clin Chem 20:278–281 (PMID: 4813007) PubMed
24.
Zurück zum Zitat Mahoney JJ, Vreman HJ, Stevenson DK, Van Kessel AL (1993) Measurement of carboxyhemoglobin and total hemoglobin by five specialized spectrophotometers (CO-oximeters) in comparison with reference methods. Clin Chem 39:1693–1700 (PMID: 8353959) PubMed Mahoney JJ, Vreman HJ, Stevenson DK, Van Kessel AL (1993) Measurement of carboxyhemoglobin and total hemoglobin by five specialized spectrophotometers (CO-oximeters) in comparison with reference methods. Clin Chem 39:1693–1700 (PMID: 8353959) PubMed
35.
Zurück zum Zitat Vreman HJ, Stevenson DK (1994) Carboxyhemoglobin determined in neonatal blood with a CO-oximeter unaffected by fetal oxyhemoglobin. Clin Chem 40:1522–1527 (PMID: 7519133) PubMed Vreman HJ, Stevenson DK (1994) Carboxyhemoglobin determined in neonatal blood with a CO-oximeter unaffected by fetal oxyhemoglobin. Clin Chem 40:1522–1527 (PMID: 7519133) PubMed
46.
Zurück zum Zitat Vreman HJ, Kwong LK, Stevenson DK (1984) Carbon monoxide in blood: an improved microliter blood-sample collection system, with rapid analysis by gas chromatography. Clin Chem 30:1382–1386 (PMID: 6744592) PubMed Vreman HJ, Kwong LK, Stevenson DK (1984) Carbon monoxide in blood: an improved microliter blood-sample collection system, with rapid analysis by gas chromatography. Clin Chem 30:1382–1386 (PMID: 6744592) PubMed
48.
Zurück zum Zitat Rodkey FL, Collison HA (1970) An artifact in the analysis of oxygenated blood for its low carbon monoxide content. Clin Chem 16:896–899 (PMID: 5473549) PubMed Rodkey FL, Collison HA (1970) An artifact in the analysis of oxygenated blood for its low carbon monoxide content. Clin Chem 16:896–899 (PMID: 5473549) PubMed
51.
Zurück zum Zitat Collison HA, Rodkey FL, O’Neal JD (1968) Determination of carbon monoxide in blood by gas chromatography. Clin Chem 14:162–171 Collison HA, Rodkey FL, O’Neal JD (1968) Determination of carbon monoxide in blood by gas chromatography. Clin Chem 14:162–171
57.
Zurück zum Zitat Dubowski KM, Lu JL (1973) Measurement of carboxyhemoglobin and carbon monoxide in blood. Ann Clin Lab Sci 3:53–65 (PMID: 4691500) PubMed Dubowski KM, Lu JL (1973) Measurement of carboxyhemoglobin and carbon monoxide in blood. Ann Clin Lab Sci 3:53–65 (PMID: 4691500) PubMed
64.
Zurück zum Zitat Van Dam J, Daenens P (1994) Microanalysis of carbon monoxide in blood by head-space capillary gas chromatography. J Forensic Sci 39:473–478 (PMID: 8195758) PubMed Van Dam J, Daenens P (1994) Microanalysis of carbon monoxide in blood by head-space capillary gas chromatography. J Forensic Sci 39:473–478 (PMID: 8195758) PubMed
68.
Zurück zum Zitat Vaupel P, Zander R, Bruley DF (eds) (1994) Oxygen transport to tissue XV. Springer Science and Business Media, New York Vaupel P, Zander R, Bruley DF (eds) (1994) Oxygen transport to tissue XV. Springer Science and Business Media, New York
69.
Zurück zum Zitat Penney DG (ed) (2000) Carbon monoxide toxicity, 1st edn. CRC Press, Boca Raton Penney DG (ed) (2000) Carbon monoxide toxicity, 1st edn. CRC Press, Boca Raton
71.
Zurück zum Zitat Pojer R, Whitfield JB, Poulos V, Eckhard IF, Richmond R, Hensley WJ (1984) Carboxyhemoglobin, cotinine, and thiocyanate assay compared for distinguishing smokers from non-smokers. Clin Chem 30:1377–1380 (PMID: 6744590) PubMed Pojer R, Whitfield JB, Poulos V, Eckhard IF, Richmond R, Hensley WJ (1984) Carboxyhemoglobin, cotinine, and thiocyanate assay compared for distinguishing smokers from non-smokers. Clin Chem 30:1377–1380 (PMID: 6744590) PubMed
78.
Metadaten
Titel
What are the limitations of methods to measure carbon monoxide in biological samples?
verfasst von
Stefania Oliverio
Vincent Varlet
Publikationsdatum
01.07.2019
Verlag
Springer Singapore
Erschienen in
Forensic Toxicology / Ausgabe 1/2020
Print ISSN: 1860-8965
Elektronische ISSN: 1860-8973
DOI
https://doi.org/10.1007/s11419-019-00490-1

Weitere Artikel der Ausgabe 1/2020

Forensic Toxicology 1/2020 Zur Ausgabe

Neu im Fachgebiet Rechtsmedizin

Open Access 15.08.2022 | Mammakarzinom | CME

Prädiktive molekulare Diagnostik beim Mammakarzinom

Welche Anforderungen stellen sich heute und in Zukunft für die Pathologie?

10.08.2022 | Schwerpunkt: Akkreditierung und Zertifizierung (Qualitätssicherung)

Ringversuche – ein zentrales Mittel der externen Qualitätssicherung

03.08.2022 | Schwerpunkt: Akkreditierung und Zertifizierung (Qualitätssicherung)

Akkreditierung in der Pathologie und Neuropathologie

Wege und Pitfalls

25.07.2022 | Schwerpunkt: Akkreditierung und Zertifizierung (Qualitätssicherung)

Die Akkreditierung in der Pathologie und Neuropathologie in Deutschland

Rahmenbedingungen und Voraussetzungen