Can Microcomputed Tomography Measure Retrieved Polyethylene Wear? Comparing Fixed-bearing and Rotating-platform Knees

verfasst von: Charles A. Engh Jr, MD, Rebecca L. Zimmerman, MS, Robert H. Hopper Jr, PhD, Gerard A. Engh, MD

Erschienen in: Clinical Orthopaedics and Related Research® | Ausgabe 1/2013

Einloggen, um Zugang zu erhalten

Abstract

Background

Wear of total knee polyethylene has been quantified gravimetrically with thickness measurements and evaluation of surface wear modes. However, these techniques do not localize volumetric wear.

Questions/Purposes

We used micro-CT scans of retrieved total knee liners and unworn, new liners to determine the volume and location of wear.

Methods

We retrieved 12 fixed and 12 rotating-platform bearings after a mean 52 months of use. Inserts were weighed and thickness was measured. Micro-CT scans of retrieved and matched new liners were superimposed to compare the location and magnitude of wear.

Results

The average total wear was 254 ± 248 mm3. The average wear rate was 58 ± 41 mm3/year. Wear was 69% of penetration, demonstrating the contribution of deformation to knee wear. Rotating-platform wear rate was 43 ± 25 mm3/year and the fixed-bearing rate was 74 ± 49 mm3/year. Five percent of the rotating-platform wear rate came from the backside compared with 14% of the fixed-bearing wear rate.

Conclusions

Micro-CT can determine the volume and location of wear of retrieved tibial liners. Because the magnitude of the manufacturing tolerances was approximately half the magnitude of the total wear on average, accounting for the potential influence of tolerances is important to accurately measure volumetric wear if the unworn (preimplantation) geometry of the insert is unknown. Without accounting for tolerances, this technique may not be applicable for retrievals with a short followup that have low wear. However, application of micro-CT could be of value in determining the exact location of wear in knee simulator studies in which the same insert is measured repeatedly and manufacturing tolerances are not a concern.
Literatur
1.
Zurück zum Zitat Atwood SA, Currier JH, Mayor MB, Collier JP, Van Citters DW, Kennedy FE. Clinical wear measurement on low contact stress rotating platform knee bearings. J Arthroplasty. 2008;23:431–440.PubMedCrossRef Atwood SA, Currier JH, Mayor MB, Collier JP, Van Citters DW, Kennedy FE. Clinical wear measurement on low contact stress rotating platform knee bearings. J Arthroplasty. 2008;23:431–440.PubMedCrossRef
2.
Zurück zum Zitat Benjamin J, Szivek J, Dersam G, Persselin S, Johnson R. Linear and volumetric wear of tibial inserts in posterior cruciate-retaining knee arthroplasties. Clin Orthop Relat Res. 2001;392:131–138.PubMedCrossRef Benjamin J, Szivek J, Dersam G, Persselin S, Johnson R. Linear and volumetric wear of tibial inserts in posterior cruciate-retaining knee arthroplasties. Clin Orthop Relat Res. 2001;392:131–138.PubMedCrossRef
3.
Zurück zum Zitat Bland JM, Altman DG. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet. 1986;1:307–310.PubMedCrossRef Bland JM, Altman DG. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet. 1986;1:307–310.PubMedCrossRef
4.
Zurück zum Zitat Blunn GW, Joshi AB, Minns RJ, Lidgren L, Lilley P, Ryd L, Engelbrecht E, Walker PS. Wear in retrieved condylar knee arthroplasties. A comparison of wear in different designs of 280 retrieved condylar knee prostheses. J Arthroplasty. 1997;12:281–290.PubMedCrossRef Blunn GW, Joshi AB, Minns RJ, Lidgren L, Lilley P, Ryd L, Engelbrecht E, Walker PS. Wear in retrieved condylar knee arthroplasties. A comparison of wear in different designs of 280 retrieved condylar knee prostheses. J Arthroplasty. 1997;12:281–290.PubMedCrossRef
5.
Zurück zum Zitat Bowden AE, Kurtz SM, Edidin AA. Validation of a micro-CT technique for measuring volumetric wear in retrieved acetabular liners. J Biomed Mater Res Part B Appl Biomater. 2005;75:205–209.PubMedCrossRef Bowden AE, Kurtz SM, Edidin AA. Validation of a micro-CT technique for measuring volumetric wear in retrieved acetabular liners. J Biomed Mater Res Part B Appl Biomater. 2005;75:205–209.PubMedCrossRef
6.
Zurück zum Zitat Collier MB, Engh CA, Hatten KM, Ginn SD, Sheils TM, Engh GA. Radiographic assessment of the thickness lost from polyethylene tibial inserts that had been sterilized differently. J Bone Joint Surg Am. 2008;90:1543–1552.PubMedCrossRef Collier MB, Engh CA, Hatten KM, Ginn SD, Sheils TM, Engh GA. Radiographic assessment of the thickness lost from polyethylene tibial inserts that had been sterilized differently. J Bone Joint Surg Am. 2008;90:1543–1552.PubMedCrossRef
7.
Zurück zum Zitat Collier MB, Engh CA, Mcauley JP, Engh GA. Factors associated with the loss of thickness of polyethylene tibial bearings after knee arthroplasty. J Bone Joint Surg Am. 2007;89:1306–1314.PubMedCrossRef Collier MB, Engh CA, Mcauley JP, Engh GA. Factors associated with the loss of thickness of polyethylene tibial bearings after knee arthroplasty. J Bone Joint Surg Am. 2007;89:1306–1314.PubMedCrossRef
8.
Zurück zum Zitat Collier MB, Jewett BA, Engh CA. Clinical assessment of tibial polyethylene thickness: comparison of radiographic measurements with as-implanted and as-retrieved thicknesses. J Arthroplasty. 2003;18:860–866.PubMedCrossRef Collier MB, Jewett BA, Engh CA. Clinical assessment of tibial polyethylene thickness: comparison of radiographic measurements with as-implanted and as-retrieved thicknesses. J Arthroplasty. 2003;18:860–866.PubMedCrossRef
9.
Zurück zum Zitat Conditt MA, Thompson MT, Usrey MM, Ismaily SK, Noble PC. Backside wear of polyethylene tibial inserts: mechanism and magnitude of material loss. J Bone Joint Surg Am. 2005;87:326–331.PubMedCrossRef Conditt MA, Thompson MT, Usrey MM, Ismaily SK, Noble PC. Backside wear of polyethylene tibial inserts: mechanism and magnitude of material loss. J Bone Joint Surg Am. 2005;87:326–331.PubMedCrossRef
10.
Zurück zum Zitat Engh GA, Engh GA, Zimmerman RL, Zimmerman RL, Parks NL, Parks NL, Engh CA, Engh CA. Analysis of wear in retrieved mobile and fixed bearing knee inserts. J Arthroplasty. 2009;24:28–32.PubMedCrossRef Engh GA, Engh GA, Zimmerman RL, Zimmerman RL, Parks NL, Parks NL, Engh CA, Engh CA. Analysis of wear in retrieved mobile and fixed bearing knee inserts. J Arthroplasty. 2009;24:28–32.PubMedCrossRef
11.
Zurück zum Zitat Garcia RM, Kraay MJ, Messerschmitt PJ, Goldberg VM, Rimnac CM. Analysis of retrieved ultra-high-molecular-weight polyethylene tibial components from rotating-platform total knee arthroplasty. J Arthroplasty. 2009;24:131–138.PubMedCrossRef Garcia RM, Kraay MJ, Messerschmitt PJ, Goldberg VM, Rimnac CM. Analysis of retrieved ultra-high-molecular-weight polyethylene tibial components from rotating-platform total knee arthroplasty. J Arthroplasty. 2009;24:131–138.PubMedCrossRef
12.
Zurück zum Zitat Kelly NH, Fu RH, Wright TM, Padgett DE. Wear damage in mobile-bearing TKA is as severe as that in fixed-bearing TKA. Clin Orthop Relat Res. 2011;469:123–130.PubMedCrossRef Kelly NH, Fu RH, Wright TM, Padgett DE. Wear damage in mobile-bearing TKA is as severe as that in fixed-bearing TKA. Clin Orthop Relat Res. 2011;469:123–130.PubMedCrossRef
13.
Zurück zum Zitat Kester MA, Herrera L, Wang A, Essner A. Knee bearing technology: where is technology taking us? J Arthroplasty. 2007;22(Suppl 3):16–20.PubMedCrossRef Kester MA, Herrera L, Wang A, Essner A. Knee bearing technology: where is technology taking us? J Arthroplasty. 2007;22(Suppl 3):16–20.PubMedCrossRef
14.
Zurück zum Zitat Kop AM, Swarts E. Quantification of polyethylene degradation in mobile bearing knees: a retrieval analysis of the Anterior-Posterior-Glide (APG) and Rotating Platform (RP) Low Contact Stress (LCS) knee. Acta Orthop. 2007;78:364–370.PubMedCrossRef Kop AM, Swarts E. Quantification of polyethylene degradation in mobile bearing knees: a retrieval analysis of the Anterior-Posterior-Glide (APG) and Rotating Platform (RP) Low Contact Stress (LCS) knee. Acta Orthop. 2007;78:364–370.PubMedCrossRef
15.
Zurück zum Zitat Lavernia CJ, Sierra RJ, Hungerford DS, Krackow K. Activity level and wear in total knee arthroplasty: a study of autopsy retrieved specimens. J Arthroplasty. 2001;16:446–453.PubMedCrossRef Lavernia CJ, Sierra RJ, Hungerford DS, Krackow K. Activity level and wear in total knee arthroplasty: a study of autopsy retrieved specimens. J Arthroplasty. 2001;16:446–453.PubMedCrossRef
16.
Zurück zum Zitat Lu Y-C, Huang C-H, Chang T-K, Ho F-Y, Cheng C-K, Huang C-H. Wear-pattern analysis in retrieved tibial inserts of mobile-bearing and fixed-bearing total knee prostheses. J Bone Joint Surg Br. 2010;92:500–507.PubMedCrossRef Lu Y-C, Huang C-H, Chang T-K, Ho F-Y, Cheng C-K, Huang C-H. Wear-pattern analysis in retrieved tibial inserts of mobile-bearing and fixed-bearing total knee prostheses. J Bone Joint Surg Br. 2010;92:500–507.PubMedCrossRef
17.
Zurück zum Zitat McEwen HMJ, Barnett PI, Bell CJ, Farrar R, Auger DD, Stone MH, Fisher J. The influence of design, materials and kinematics on the in vitro wear of total knee replacements. J Biomech. 2005;38:357–365.PubMedCrossRef McEwen HMJ, Barnett PI, Bell CJ, Farrar R, Auger DD, Stone MH, Fisher J. The influence of design, materials and kinematics on the in vitro wear of total knee replacements. J Biomech. 2005;38:357–365.PubMedCrossRef
18.
Zurück zum Zitat Muratoglu OK, Bragdon CR, Jasty M, O’Connor DO, Knoch Von RS, Harris WH. Knee-simulator testing of conventional and cross-linked polyethylene tibial inserts. J Arthroplasty. 2004;19:887–897.PubMedCrossRef Muratoglu OK, Bragdon CR, Jasty M, O’Connor DO, Knoch Von RS, Harris WH. Knee-simulator testing of conventional and cross-linked polyethylene tibial inserts. J Arthroplasty. 2004;19:887–897.PubMedCrossRef
19.
Zurück zum Zitat Muratoglu OK, Rubash HE, Bragdon CR, Burroughs BR, Huang A, Harris WH. Simulated normal gait wear testing of a highly cross-linked polyethylene tibial insert. J Arthroplasty. 2007;22:435–444.PubMedCrossRef Muratoglu OK, Rubash HE, Bragdon CR, Burroughs BR, Huang A, Harris WH. Simulated normal gait wear testing of a highly cross-linked polyethylene tibial insert. J Arthroplasty. 2007;22:435–444.PubMedCrossRef
20.
Zurück zum Zitat Sanzén L, Sahlström A, Gentz CF, Johnell IR. Radiographic wear assessment in a total knee prosthesis. 5- to 9-year follow-up study of 158 knees. J Arthroplasty. 1996;11:738–742.PubMedCrossRef Sanzén L, Sahlström A, Gentz CF, Johnell IR. Radiographic wear assessment in a total knee prosthesis. 5- to 9-year follow-up study of 158 knees. J Arthroplasty. 1996;11:738–742.PubMedCrossRef
21.
Zurück zum Zitat Teeter MG, Naudie DDR, McErlain DD, Brandt J-M, Yuan X, Macdonald SJ, Holdsworth DW. In vitro quantification of wear in tibial inserts using microcomputed tomography. Clin Orthop Relat Res. 2011;469:107–112.PubMedCrossRef Teeter MG, Naudie DDR, McErlain DD, Brandt J-M, Yuan X, Macdonald SJ, Holdsworth DW. In vitro quantification of wear in tibial inserts using microcomputed tomography. Clin Orthop Relat Res. 2011;469:107–112.PubMedCrossRef
22.
Zurück zum Zitat Thadani PJ, Vince KG, Ortaaslan SG, Blackburn DC, Cudiamat CV. Ten- to 12-year followup of the Insall-Burstein I total knee prosthesis. Clin Orthop Relat Res. 2000;380:17–29.PubMedCrossRef Thadani PJ, Vince KG, Ortaaslan SG, Blackburn DC, Cudiamat CV. Ten- to 12-year followup of the Insall-Burstein I total knee prosthesis. Clin Orthop Relat Res. 2000;380:17–29.PubMedCrossRef
Metadaten
Titel
Can Microcomputed Tomography Measure Retrieved Polyethylene Wear? Comparing Fixed-bearing and Rotating-platform Knees
verfasst von
Charles A. Engh Jr, MD
Rebecca L. Zimmerman, MS
Robert H. Hopper Jr, PhD
Gerard A. Engh, MD
Publikationsdatum
01.01.2013
Verlag
Springer-Verlag
Erschienen in
Clinical Orthopaedics and Related Research® / Ausgabe 1/2013
Print ISSN: 0009-921X
Elektronische ISSN: 1528-1132
DOI
https://doi.org/10.1007/s11999-012-2513-2

Weitere Artikel der Ausgabe 1/2013

Clinical Orthopaedics and Related Research® 1/2013 Zur Ausgabe

Symposium: Papers Presented at the Annual Meetings of The Knee Society

Papers Presented at the Annual Meetings of The Knee Society: Editorial Comment

Symposium: Papers Presented at the Annual Meetings of the Knee Society

Postoperative Alignment and ROM Affect Patient Satisfaction After TKA

Arthropedia

Grundlagenwissen der Arthroskopie und Gelenkchirurgie. Erweitert durch Fallbeispiele, Videos und Abbildungen. 
» Jetzt entdecken

Häusliche Gewalt in der orthopädischen Notaufnahme oft nicht erkannt

28.05.2024 Traumatologische Notfälle Nachrichten

In der Notaufnahme wird die Chance, Opfer von häuslicher Gewalt zu identifizieren, von Orthopäden und Orthopädinnen offenbar zu wenig genutzt. Darauf deuten die Ergebnisse einer Fragebogenstudie an der Sahlgrenska-Universität in Schweden hin.

Fehlerkultur in der Medizin – Offenheit zählt!

Darüber reden und aus Fehlern lernen, sollte das Motto in der Medizin lauten. Und zwar nicht nur im Sinne der Patientensicherheit. Eine negative Fehlerkultur kann auch die Behandelnden ernsthaft krank machen, warnt Prof. Dr. Reinhard Strametz. Ein Plädoyer und ein Leitfaden für den offenen Umgang mit kritischen Ereignissen in Medizin und Pflege.

Mehr Frauen im OP – weniger postoperative Komplikationen

21.05.2024 Allgemeine Chirurgie Nachrichten

Ein Frauenanteil von mindestens einem Drittel im ärztlichen Op.-Team war in einer großen retrospektiven Studie aus Kanada mit einer signifikanten Reduktion der postoperativen Morbidität assoziiert.

„Übersichtlicher Wegweiser“: Lauterbachs umstrittener Klinik-Atlas ist online

17.05.2024 Klinik aktuell Nachrichten

Sie sei „ethisch geboten“, meint Gesundheitsminister Karl Lauterbach: mehr Transparenz über die Qualität von Klinikbehandlungen. Um sie abzubilden, lässt er gegen den Widerstand vieler Länder einen virtuellen Klinik-Atlas freischalten.

Update Orthopädie und Unfallchirurgie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.