Abstract

Purpose

The purpose of this study was to compare in vivo kinematics, range of motion, and functional outcomes in patients that received either a high-flexion cruciate retaining or a high-flexion cruciate substituting knee.

Methods

Forty-eight high-flexion cruciate retaining (CR) and 47 high-flexion cruciate substituting (PS) knees were included in this study and followed for an average of 27 months (24–33). Weight-bearing and non-weight-bearing maximal flexions and functional scores were compared between two groups. For kinematics evaluations, amount of posterior femoral roll-back and internal tibial rotation from 0° to maximal flexion using lateral radiographs under weight-bearing conditions were also compared.

Results

Average weight-bearing maximal flexion was 126.3° in the PS group, which was significantly higher than the 115.0° in the CR group. Average functional scores showed no significant difference between the two groups. In terms of kinematics, the average amount of posterior femoral roll-back during full flexion was 9.6 mm in the PS group and 6.1 mm in the CR group, which was a significant difference. However, internal tibial rotation during full flexion was not significantly different in the two groups.

Conclusion

PS high-flexion TKA provided greater weight-bearing maximal flexion and posterior femoral roll-back than CR high-flexion TKA, although no difference in clinical outcomes was observed between the two prosthesis designs.

Level of evidence

I.
Literatur
1.
Zurück zum Zitat Anouchi YS, McShane M, Kelly F Jr et al (1996) Range of motion in total knee replacement. Clin Orthop Relat Res 331:87–92PubMedCrossRef Anouchi YS, McShane M, Kelly F Jr et al (1996) Range of motion in total knee replacement. Clin Orthop Relat Res 331:87–92PubMedCrossRef
2.
Zurück zum Zitat Banks SA, Markovich GD, Hodge WA (1997) In vivo kinematics of cruciate-retaining and -substituting knee arthroplasties. J Arthroplasty 12:297–304PubMedCrossRef Banks SA, Markovich GD, Hodge WA (1997) In vivo kinematics of cruciate-retaining and -substituting knee arthroplasties. J Arthroplasty 12:297–304PubMedCrossRef
3.
Zurück zum Zitat Becker MW, Insall JN, Faris PM (1991) Bilateral total knee arthroplasty. One cruciate retaining and one cruciate substituting. Clin Orthop Relat Res 271:122–124PubMed Becker MW, Insall JN, Faris PM (1991) Bilateral total knee arthroplasty. One cruciate retaining and one cruciate substituting. Clin Orthop Relat Res 271:122–124PubMed
4.
Zurück zum Zitat Bellemans J, Banks S, Victor J et al (2002) Fluoroscopic analysis of the kinematics of deep flexion in total knee arthroplasty. Influence of posterior condylar offset. J Bone Joint Surg Br 84:50–53PubMedCrossRef Bellemans J, Banks S, Victor J et al (2002) Fluoroscopic analysis of the kinematics of deep flexion in total knee arthroplasty. Influence of posterior condylar offset. J Bone Joint Surg Br 84:50–53PubMedCrossRef
5.
Zurück zum Zitat Cates HE, Komistek RD, Mahfouz MR et al (2008) In vivo comparison of knee kinematics for subjects having either a posterior stabilized or cruciate retaining high-flexion total knee arthroplasty. J Arthroplasty 23:1057–1067PubMedCrossRef Cates HE, Komistek RD, Mahfouz MR et al (2008) In vivo comparison of knee kinematics for subjects having either a posterior stabilized or cruciate retaining high-flexion total knee arthroplasty. J Arthroplasty 23:1057–1067PubMedCrossRef
6.
Zurück zum Zitat Dennis DA, Komistek RD, Colwell CE Jr et al (1998) In vivo anteroposterior femorotibial translation: a multicenter analysis. Clin Orthop Relat Res 356:47–57PubMedCrossRef Dennis DA, Komistek RD, Colwell CE Jr et al (1998) In vivo anteroposterior femorotibial translation: a multicenter analysis. Clin Orthop Relat Res 356:47–57PubMedCrossRef
7.
Zurück zum Zitat Dennis DA, Komistek RD, Mahfouz MR et al (2003) Multicenter determination of in vivo kinematics after total knee arthroplasty. Clin Orthop Relat Res 416:37–57PubMedCrossRef Dennis DA, Komistek RD, Mahfouz MR et al (2003) Multicenter determination of in vivo kinematics after total knee arthroplasty. Clin Orthop Relat Res 416:37–57PubMedCrossRef
8.
Zurück zum Zitat Dennis DA, Komistek RD, Mahfouz MR et al (2004) A multicenter analysis of axial femorotibial rotation after total knee arthroplasty. Clin Orthop Relat Res 428:180–189PubMedCrossRef Dennis DA, Komistek RD, Mahfouz MR et al (2004) A multicenter analysis of axial femorotibial rotation after total knee arthroplasty. Clin Orthop Relat Res 428:180–189PubMedCrossRef
9.
Zurück zum Zitat Dennis DA, Komistek RD, Stiehl JB et al (1998) Range of motion following total knee arthroplasty: the effect of implant design and weight-bearing conditions. J Arthroplasty 13:748–752PubMedCrossRef Dennis DA, Komistek RD, Stiehl JB et al (1998) Range of motion following total knee arthroplasty: the effect of implant design and weight-bearing conditions. J Arthroplasty 13:748–752PubMedCrossRef
10.
Zurück zum Zitat Hartford JM, Banit D, Hall K et al (2001) Radiographic analysis of low contact stress meniscal bearing total knee replacements. J Bone Joint Surg Am 83:229–234PubMed Hartford JM, Banit D, Hall K et al (2001) Radiographic analysis of low contact stress meniscal bearing total knee replacements. J Bone Joint Surg Am 83:229–234PubMed
11.
Zurück zum Zitat Kim YH, Choi Y, Kwon OR et al (2009) Functional outcome and range of motion of high-flexion posterior cruciate-retaining and high-flexion posterior cruciate-substituting total knee prostheses. A prospective, randomized study. J Bone Joint Surg Am 91:753–760PubMedCrossRef Kim YH, Choi Y, Kwon OR et al (2009) Functional outcome and range of motion of high-flexion posterior cruciate-retaining and high-flexion posterior cruciate-substituting total knee prostheses. A prospective, randomized study. J Bone Joint Surg Am 91:753–760PubMedCrossRef
12.
Zurück zum Zitat Kim YH, Sohn KS, Kim JS (2005) Range of motion of standard and high-flexion posterior stabilized total knee prostheses. A prospective, randomized study. J Bone Joint Surg Am 87:1470–1475PubMedCrossRef Kim YH, Sohn KS, Kim JS (2005) Range of motion of standard and high-flexion posterior stabilized total knee prostheses. A prospective, randomized study. J Bone Joint Surg Am 87:1470–1475PubMedCrossRef
13.
Zurück zum Zitat Landis JR, Koch GG (1977) The measurement of observer agreement for categorical data. Biometrics 33:159–174PubMedCrossRef Landis JR, Koch GG (1977) The measurement of observer agreement for categorical data. Biometrics 33:159–174PubMedCrossRef
14.
Zurück zum Zitat Maloney WJ, Schurman DJ (1992) The effects of implant design on range of motion after total knee arthroplasty. Total condylar versus posterior stabilized total condylar designs. Clin Orthop Relat Res 278:147–152PubMed Maloney WJ, Schurman DJ (1992) The effects of implant design on range of motion after total knee arthroplasty. Total condylar versus posterior stabilized total condylar designs. Clin Orthop Relat Res 278:147–152PubMed
15.
Zurück zum Zitat Massin P, Gournay A (2006) Optimization of the posterior condylar offset, tibial slope, and condylar roll-back in total knee arthroplasty. J Arthroplasty 21:889–896PubMedCrossRef Massin P, Gournay A (2006) Optimization of the posterior condylar offset, tibial slope, and condylar roll-back in total knee arthroplasty. J Arthroplasty 21:889–896PubMedCrossRef
16.
Zurück zum Zitat Most E, Sultan PG, Park SE et al (2006) Tibiofemoral contact behavior is improved in high-flexion cruciate retaining TKA. Clin Orthop Relat Res 452:59–64PubMedCrossRef Most E, Sultan PG, Park SE et al (2006) Tibiofemoral contact behavior is improved in high-flexion cruciate retaining TKA. Clin Orthop Relat Res 452:59–64PubMedCrossRef
17.
Zurück zum Zitat Most E, Sultan PG, Park SE et al (2005) Kinematic analysis of conventional and high-flexion cruciate-retaining total knee arthroplasties: an in vitro investigation. J Arthroplasty 20:529–535PubMedCrossRef Most E, Sultan PG, Park SE et al (2005) Kinematic analysis of conventional and high-flexion cruciate-retaining total knee arthroplasties: an in vitro investigation. J Arthroplasty 20:529–535PubMedCrossRef
18.
Zurück zum Zitat Myles CM, Rowe PJ, Walker CR et al (2002) Knee joint functional range of movement prior to and following total knee arthroplasty measured using flexible electrogoniometry. Gait Posture 16:46–54PubMedCrossRef Myles CM, Rowe PJ, Walker CR et al (2002) Knee joint functional range of movement prior to and following total knee arthroplasty measured using flexible electrogoniometry. Gait Posture 16:46–54PubMedCrossRef
19.
Zurück zum Zitat Sharma A, Komistek RD, Scuderi GR et al (2007) High-flexion TKA designs: what are their in vivo contact mechanics? Clin Orthop Relat Res 464:117–126PubMed Sharma A, Komistek RD, Scuderi GR et al (2007) High-flexion TKA designs: what are their in vivo contact mechanics? Clin Orthop Relat Res 464:117–126PubMed
20.
Zurück zum Zitat Shi K, Hayashida K, Umeda N et al (2007) Kinematic comparison between mobile-bearing and fixed-bearing inserts in Nexgen legacy posterior stabilized flex total knee arthroplasty. J Arthroplasty 23:164–169PubMedCrossRef Shi K, Hayashida K, Umeda N et al (2007) Kinematic comparison between mobile-bearing and fixed-bearing inserts in Nexgen legacy posterior stabilized flex total knee arthroplasty. J Arthroplasty 23:164–169PubMedCrossRef
21.
Zurück zum Zitat Stiehl JB, Komistek RD, Dennis DA (1999) Detrimental kinematics of a flat on flat total condylar knee arthroplasty. Clin Orthop Relat Res 365:139–148PubMedCrossRef Stiehl JB, Komistek RD, Dennis DA (1999) Detrimental kinematics of a flat on flat total condylar knee arthroplasty. Clin Orthop Relat Res 365:139–148PubMedCrossRef
22.
Zurück zum Zitat Stiehl JB, Komistek RD, Dennis DA et al (1995) Fluoroscopic analysis after posterior cruciate retaining knee arthroplasty. J Bone Joint Surg Br 77:884–889PubMed Stiehl JB, Komistek RD, Dennis DA et al (1995) Fluoroscopic analysis after posterior cruciate retaining knee arthroplasty. J Bone Joint Surg Br 77:884–889PubMed
23.
Zurück zum Zitat Suggs JF, Hanson GR, Park SE et al (2008) Patient function after a posterior stabilizing total knee arthroplasty: cam-post engagement and knee kinematics. Knee Surg Sports Traumatol Arthrosc 16:290–296PubMedCrossRef Suggs JF, Hanson GR, Park SE et al (2008) Patient function after a posterior stabilizing total knee arthroplasty: cam-post engagement and knee kinematics. Knee Surg Sports Traumatol Arthrosc 16:290–296PubMedCrossRef
24.
Zurück zum Zitat Suggs JF, Kwon YM, Durbhakula SM et al (2009) In vivo flexion and kinematics of the knee after TKA: comparison of a conventional and a high flexion cruciate-retaining TKA design. Knee Surg Sports Traumatol Arthrosc 17:150–156PubMedCrossRef Suggs JF, Kwon YM, Durbhakula SM et al (2009) In vivo flexion and kinematics of the knee after TKA: comparison of a conventional and a high flexion cruciate-retaining TKA design. Knee Surg Sports Traumatol Arthrosc 17:150–156PubMedCrossRef
Metadaten
Titel
Comparisons of kinematics and range of motion in high-flexion total knee arthroplasty: cruciate retaining vs. substituting designs
verfasst von
Jong-Keun Seon
Ju-Kwon Park
Young-Joo Shin
Hyung-Yeon Seo
Keun-Bae Lee
Eun-Kyoo Song
Publikationsdatum
01.12.2011
Verlag
Springer-Verlag
Erschienen in
Knee Surgery, Sports Traumatology, Arthroscopy / Ausgabe 12/2011
Print ISSN: 0942-2056
Elektronische ISSN: 1433-7347
DOI
https://doi.org/10.1007/s00167-011-1434-z

Weitere Artikel der Ausgabe 12/2011

Knee Surgery, Sports Traumatology, Arthroscopy 12/2011 Zur Ausgabe

Arthropedia

Grundlagenwissen der Arthroskopie und Gelenkchirurgie. Erweitert durch Fallbeispiele, Videos und Abbildungen. 
» Jetzt entdecken

Knie-TEP: Kein Vorteil durch antibiotikahaltigen Knochenzement

29.05.2024 Periprothetische Infektionen Nachrichten

Zur Zementierung einer Knie-TEP wird in Deutschland zu über 98% Knochenzement verwendet, der mit einem Antibiotikum beladen ist. Ob er wirklich besser ist als Zement ohne Antibiotikum, kann laut Registerdaten bezweifelt werden.

Häusliche Gewalt in der orthopädischen Notaufnahme oft nicht erkannt

28.05.2024 Häusliche Gewalt Nachrichten

In der Notaufnahme wird die Chance, Opfer von häuslicher Gewalt zu identifizieren, von Orthopäden und Orthopädinnen offenbar zu wenig genutzt. Darauf deuten die Ergebnisse einer Fragebogenstudie an der Sahlgrenska-Universität in Schweden hin.

Fehlerkultur in der Medizin – Offenheit zählt!

28.05.2024 Fehlerkultur Podcast

Darüber reden und aus Fehlern lernen, sollte das Motto in der Medizin lauten. Und zwar nicht nur im Sinne der Patientensicherheit. Eine negative Fehlerkultur kann auch die Behandelnden ernsthaft krank machen, warnt Prof. Dr. Reinhard Strametz. Ein Plädoyer und ein Leitfaden für den offenen Umgang mit kritischen Ereignissen in Medizin und Pflege.

Mehr Frauen im OP – weniger postoperative Komplikationen

21.05.2024 Allgemeine Chirurgie Nachrichten

Ein Frauenanteil von mindestens einem Drittel im ärztlichen Op.-Team war in einer großen retrospektiven Studie aus Kanada mit einer signifikanten Reduktion der postoperativen Morbidität assoziiert.

Update Orthopädie und Unfallchirurgie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.