Skip to main content
Erschienen in: Journal of Cardiothoracic Surgery 1/2011

Open Access 01.12.2011 | Case report

Delayed intracardial shunting and hypoxemia after massive pulmonary embolism in a patient with a biventricular assist device

verfasst von: Thomas Weig, Michael E Dolch, Lorenz Frey, Dirk Bruegger, Peter Boekstegers, Ralf Sodian, Michael Irlbeck

Erschienen in: Journal of Cardiothoracic Surgery | Ausgabe 1/2011

download
DOWNLOAD
print
DRUCKEN
insite
SUCHEN

Abstract

We describe the interdisciplinary management of a 34-year-old woman with dilated cardiomyopathy three months postpartum on a cardiac biventricular assist device (BVAD) as bridge to heart transplantation with delayed onset of intracardial shunting and subsequent hypoxemia due to massive pulmonary embolism. After emergency surgical embolectomy pulmonary function was highly compromised (PaO2/FiO2 54) requiring bifemoral veno-venous extracorporeal membrane oxygenation. Transesophageal echocardiography detected atrial level hypoxemic right-to-left shunting through a patent foramen ovale (PFO). Percutaneous closure of the PFO was achieved with a PFO occluder device. After placing the PFO occluder device oxygenation increased significantly (Δ paO2 119 Torr). The patient received heart transplantation 20 weeks after BVAD implantation and was discharged from ICU 3 weeks after transplantation.
An increase in pulmonary vascular resistance in patients on BVAD can reopen a PFO resulting in atrial right-to-left shunting and subsequent hypoxemia. The case demonstrates the usefulness of transesophageal echocardiography examinations in the detection of this unexpected event. Percutaneous placement of a PFO occluder device is an appropriate strategy to stop intracardiac shunting through PFO in fixed elevation of pulmonary vascular resistance.
Hinweise

Electronic supplementary material

The online version of this article (doi:10.​1186/​1749-8090-6-133) contains supplementary material, which is available to authorized users.

Competing interests

The authors declare that they have no competing interests.

Authors' contributions

TW reviewed the case, conducted a review of the literature and drafted the manuscript. TW and MI performed the echocardiographic studies and participated in the design of the case report. RS and PB performed the operation and intervention described. MD, LF and DB confirmed the patient's diagnosis and revised the manuscript, contributing important intellectual content. All authors read and approved the final manuscript.

Background

In a literature review, few cases of atrial level right-to-left shunt in patients with left ventricular assist devices are described. All these cases were detected either intraoperatively [13] or within the first postoperative days [47]. We describe a case of delayed onset of atrial level right-to-left shunt after massive pulmonary embolism on biventricular assist device (BVAD) support.

Case Presentation

A 34 year old female patient was admitted to our hospital with dilated cardiomyopathy three months after birth of her third child. She had a known history of familial dilated cardiomyopathy. Recompensation was not achieved despite maximum medical therapy and insertion of an intra-aortic balloon pump. BVAD [Excor, Berlin Heart, Berlin, Germany] was implanted using a bi-atrial cannulation technique as bridge to heart transplantation. Perioperative transesophageal echocardiography did not show a patent foramen ovale (PFO). Postoperative recovery was immediate and the patient was discharged from the ICU on the third postoperative day.
Four weeks after device implantation the patient developed fulminant pulmonary embolism despite therapeutic anticoagulation. Emergency surgical embolectomy for massive pulmonary embolism was performed since thrombolysis was not an option after recent implantation of an artificial heart (Figure 1). Pulmonary function was highly compromised after embolectomy and veno-venous extracorporeal membrane oxygenation (ECMO) [Bio-Console, Medtronic, Minneapolis, USA] was established using a bifemoral venous access. Weaning from veno-venous ECMO was achieved over the following week but after removal oxygenation failure reoccurred. FiO2 of 1.0 was necessary to achieve sufficient oxygen saturation (paO2/FiO2 54). Modification of ventilator setting with adjustments of PEEP and peak inspiratory pressure did not lastingly improve oxygenation. Transesophageal echocardiography detected atrial level intracardial shunting (Figure 2). There was no improvement after application of inhaled pulmonary vasodilatators. CT-scan after surgical embolectomy showed residual emboli in the pulmonary vascular system. Invasive procedures such as re-embolectomy, topical thrombolysis or catheter fragmentation were considered as too harmful or not effective. Since right heart function was secured even with high pulmonary vascular resistance, percutaneous placement of a PFO occluder device [Amplatzer PFO Occluder®, AGA Medical, Plymouth, USA] was performed (Figure 2, Additional file 1). Oxygenation increased significantly after placement without change of respirator settings (Δ paO2 119 Torr). Weaning from mechanical ventilation was successful after 15 weeks.
Additional file 1: Transesophageal echocardiogram. Transesophageal echocardiogram before and after patent foramen ovale closure with a PFO occluder device [Amplatzer PFO Occluder®, AGA Medical, Plymouth, USA]. (MOV 2 MB)
After 5 weeks of therapeutic anticoagulation the residual emboli diminished and pulmonary vascular resistance was measured at 184 dyne•s/cm5 with activated assist device and 160 dyne•s/cm5 with deactivated assist device.
Heart transplantation was performed 20 weeks after implantation of the BVAD and 16 weeks after pulmonary embolism and placement of the PFO occluder device. Discharge from ICU was 3 weeks after transplantation. Informed consent for publication was obtained from the patient.

Discussion

The problem with PFO and left ventricular assist device leading to atrial level right-to-left shunt with consecutive hypoxemia is well described [17]. PFO has an incidence up to 27% in normal healthy adults as well as in adult cardiac surgical patients [8, 9]. If left ventricular assist device (LVAD) is activated, left atrial unloading leads to a decrease in left atrial pressure [10]. Right atrial pressure exceeds left atrial pressure and with PFO atrial level right-to-left shunt occurs. Depending on the shunt fraction hypoxemia may occur [11].
Therefore, intraoperative transesophageal echocardiography with colour Doppler imaging and contrast with agitated saline is highly recommended before cardiopulmonary bypass and after LVAD activation [12, 13]. Alternatively, manual occlusion of the pulmonary artery shortly before activation of the LVAD by the surgeon and transesophageal echocardiography studies as described are performed [14]. If PFO is detected before weaning from cardiopulmonary bypass, immediate operative closure is recommended. If shunting is detected after weaning from cardiopulmonary bypass, delayed interventional closure after stabilization is preferred if oxygenation failure is tolerable, since failure of the right heart in LVAD implantation or bleeding complications due to coagulopathy after reapplied bypass can deteriorate outcome [2]. PFO closure improved oxygenation in all known cases as it did in our patient.
There is only one other case of delayed onset of atrial level right-to-left shunt in patients on ventricular assist device [15]. In this case report, atrial level right-to-left shunt with hypoxemia occurred after replacement of the valves of a LVAD [LVAS, Novacor, Salt Lake City, USA] which had been implanted one year before. The management consisted of reduction of right atrial pressure by conservative means.
Persisting elevation of right atrial pressure due to persisting change of the pulmonary vascular resistance in a patient with a BVAD has not been described. An etiologic reason for persisting elevation of pulmonary vascular resistance can be massive pulmonary embolism as described in our case. Our report is the first description of a patient surviving massive pulmonary embolism while on BVAD, followed by successful orthotopic heart transplantation. To the best of our knowledge there is only one other published case of pulmonary embolism in a patient with a BVAD. This patient died shortly after the event [16].
Emergency surgical embolectomy is recommended in hemodynamic unstable patients with massive pulmonary embolism in a facility with cardiac surgical capabilities [17]. Catheter embolectomy should be performed in absence of cardiothoracic surgical backup [17]. In our case, thrombolysis was contraindicated. Therefore emergency surgical embolectomy was the treatment of choice. The reported median reduction of pulmonary vascular resistance achieved by surgical embolectomy is from 893 ± 443.5 dyne•s/cm5 to 285 ± 214 dyne•s/cm5[18], a result that was achieved in our patient.
With regard to the planned heart transplantation, chronic thromboembolic pulmonary hypertension would have been an exclusion criterion.

Conclusion

Diagnostic transesophageal echocardiography must be performed with relevant change in the hemodynamic situation and recurring hypoxemia in patients with VAD since increase in pulmonary vascular resistance can reopen PFO resulting in atrial level right-to-left shunting and consecutive hypoxemia.
Written informed consent was obtained from the patient for publication of this Case report and any accompanying images. A copy of the written consent is available for review by the Editor-in-Chief of this journal.
This article is published under license to BioMed Central Ltd. This is an Open Access article distributed under the terms of the Creative Commons Attribution License (http://​creativecommons.​org/​licenses/​by/​2.​0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.

Competing interests

The authors declare that they have no competing interests.

Authors' contributions

TW reviewed the case, conducted a review of the literature and drafted the manuscript. TW and MI performed the echocardiographic studies and participated in the design of the case report. RS and PB performed the operation and intervention described. MD, LF and DB confirmed the patient's diagnosis and revised the manuscript, contributing important intellectual content. All authors read and approved the final manuscript.
Anhänge

Authors’ original submitted files for images

Below are the links to the authors’ original submitted files for images.
Literatur
1.
Zurück zum Zitat Baker JE, Stratmann G, Hoopes C, Donateillo R, Tseng E, Russell IA: Profound hypoxemia resulting from shunting across an inadvertent atrial septal tear after left ventricular assist device placement. Anesth Analg. 2004, 98 (4): 937-940.CrossRefPubMed Baker JE, Stratmann G, Hoopes C, Donateillo R, Tseng E, Russell IA: Profound hypoxemia resulting from shunting across an inadvertent atrial septal tear after left ventricular assist device placement. Anesth Analg. 2004, 98 (4): 937-940.CrossRefPubMed
2.
Zurück zum Zitat Kavarana MN, Rahman FA, Recto MR, Dowling RD: Transcatheter closure of patent foramen ovale after left ventricular assist device implantation: intraoperative decision making. J Heart Lung Transplant. 2005, 24 (9): 1445-CrossRefPubMed Kavarana MN, Rahman FA, Recto MR, Dowling RD: Transcatheter closure of patent foramen ovale after left ventricular assist device implantation: intraoperative decision making. J Heart Lung Transplant. 2005, 24 (9): 1445-CrossRefPubMed
3.
Zurück zum Zitat Nguyen DQ, Das GS, Grubbs BC, Bolman RM, Park SJ: Transcatheter closure of patent foramen ovale for hypoxemia during left ventricular assist device support. J Heart Lung Transplant. 1999, 18 (10): 1021-1023. 10.1016/S1053-2498(99)00064-9.CrossRefPubMed Nguyen DQ, Das GS, Grubbs BC, Bolman RM, Park SJ: Transcatheter closure of patent foramen ovale for hypoxemia during left ventricular assist device support. J Heart Lung Transplant. 1999, 18 (10): 1021-1023. 10.1016/S1053-2498(99)00064-9.CrossRefPubMed
4.
Zurück zum Zitat Bonvini RF, Verin V, Lerch R, Gerard I, Sierra J, Spratt JC: Percutaneous closure of patent foramen ovale in a patient presenting arterial hypoxaemia and supported with bi-ventricular assist device. Intensive Care Med. 2005, 31 (4): 602-603. 10.1007/s00134-004-2466-3.CrossRefPubMed Bonvini RF, Verin V, Lerch R, Gerard I, Sierra J, Spratt JC: Percutaneous closure of patent foramen ovale in a patient presenting arterial hypoxaemia and supported with bi-ventricular assist device. Intensive Care Med. 2005, 31 (4): 602-603. 10.1007/s00134-004-2466-3.CrossRefPubMed
5.
Zurück zum Zitat Loeffelbein F, Schlensak C, Beyersdorf F, Dittrich S: Successful interventional closure of a patent foramen ovale in a pediatric patient supported with a biventricular assist device. Interact Cardiovasc Thorac Surg. 2007, 6 (6): 778-779. 10.1510/icvts.2007.155622.CrossRefPubMed Loeffelbein F, Schlensak C, Beyersdorf F, Dittrich S: Successful interventional closure of a patent foramen ovale in a pediatric patient supported with a biventricular assist device. Interact Cardiovasc Thorac Surg. 2007, 6 (6): 778-779. 10.1510/icvts.2007.155622.CrossRefPubMed
6.
Zurück zum Zitat Loyalka P, Idelchik GM, Kar B: Percutaneous left ventricular assist device complicated by a patent foramen ovale: importance of identification and management. Catheter Cardiovasc Interv. 2007, 70 (3): 383-386. 10.1002/ccd.21185.CrossRefPubMed Loyalka P, Idelchik GM, Kar B: Percutaneous left ventricular assist device complicated by a patent foramen ovale: importance of identification and management. Catheter Cardiovasc Interv. 2007, 70 (3): 383-386. 10.1002/ccd.21185.CrossRefPubMed
7.
Zurück zum Zitat Srinivas CV, Collins N, Borger MA, Horlick E, Murphy PM: Hypoxemia complicating LVAD insertion: novel application of the Amplatzer PFO occlusion device. J Card Surg. 2007, 22 (2): 156-158. 10.1111/j.1540-8191.2007.00370.x.CrossRefPubMed Srinivas CV, Collins N, Borger MA, Horlick E, Murphy PM: Hypoxemia complicating LVAD insertion: novel application of the Amplatzer PFO occlusion device. J Card Surg. 2007, 22 (2): 156-158. 10.1111/j.1540-8191.2007.00370.x.CrossRefPubMed
8.
Zurück zum Zitat Augoustides JG, Weiss SJ, Weiner J, Mancini J, Savino JS, Cheung AT: Diagnosis of patent foramen ovale with multiplane transesophageal echocardiography in adult cardiac surgical patients. J Cardiothorac Vasc Anesth. 2004, 18 (6): 725-730. 10.1053/j.jvca.2004.08.009.CrossRefPubMed Augoustides JG, Weiss SJ, Weiner J, Mancini J, Savino JS, Cheung AT: Diagnosis of patent foramen ovale with multiplane transesophageal echocardiography in adult cardiac surgical patients. J Cardiothorac Vasc Anesth. 2004, 18 (6): 725-730. 10.1053/j.jvca.2004.08.009.CrossRefPubMed
9.
Zurück zum Zitat Hagen PT, Scholz DG, Edwards WD: Incidence and size of patent foramen ovale during the first 10 decades of life: an autopsy study of 965 normal hearts. Mayo Clin Proc. 1984, 59 (1): 17-20.CrossRefPubMed Hagen PT, Scholz DG, Edwards WD: Incidence and size of patent foramen ovale during the first 10 decades of life: an autopsy study of 965 normal hearts. Mayo Clin Proc. 1984, 59 (1): 17-20.CrossRefPubMed
10.
Zurück zum Zitat Nakatani S, Thomas JD, Savage RM, Vargo RL, Smedira NG, McCarthy PM: Prediction of right ventricular dysfunction after left ventricular assist device implantation. Circulation. 1996, 94 (9 Suppl): II216-221.PubMed Nakatani S, Thomas JD, Savage RM, Vargo RL, Smedira NG, McCarthy PM: Prediction of right ventricular dysfunction after left ventricular assist device implantation. Circulation. 1996, 94 (9 Suppl): II216-221.PubMed
11.
Zurück zum Zitat Godart F, Rey C, Prat A, Vincentelli A, Chmait A, Francart C, Porte H: Atrial right-to-left shunting causing severe hypoxaemia despite normal right-sided pressures - Report of 11 consecutive cases corrected by percutaneous closure. Eur Heart J. 2000, 21 (6): 483-489. 10.1053/euhj.1999.1944.CrossRefPubMed Godart F, Rey C, Prat A, Vincentelli A, Chmait A, Francart C, Porte H: Atrial right-to-left shunting causing severe hypoxaemia despite normal right-sided pressures - Report of 11 consecutive cases corrected by percutaneous closure. Eur Heart J. 2000, 21 (6): 483-489. 10.1053/euhj.1999.1944.CrossRefPubMed
12.
Zurück zum Zitat Chumnanvej S, Wood MJ, MacGillivray TE, Melo MF: Perioperative echocardiographic examination for ventricular assist device implantation. Anesth Analg. 2007, 105 (3): 583-601. 10.1213/01.ane.0000278088.22952.82.CrossRefPubMed Chumnanvej S, Wood MJ, MacGillivray TE, Melo MF: Perioperative echocardiographic examination for ventricular assist device implantation. Anesth Analg. 2007, 105 (3): 583-601. 10.1213/01.ane.0000278088.22952.82.CrossRefPubMed
13.
Zurück zum Zitat Liao KK, Miller L, Toher C, Ormaza S, Herrington CS, Bittner HB, Park SJ: Timing of transesophageal echocardiography in diagnosing patent foramen ovale in patients supported with left ventricular assist device. Ann Thorac Surg. 2003, 75 (5): 1624-1626. 10.1016/S0003-4975(02)04676-3.CrossRefPubMed Liao KK, Miller L, Toher C, Ormaza S, Herrington CS, Bittner HB, Park SJ: Timing of transesophageal echocardiography in diagnosing patent foramen ovale in patients supported with left ventricular assist device. Ann Thorac Surg. 2003, 75 (5): 1624-1626. 10.1016/S0003-4975(02)04676-3.CrossRefPubMed
14.
Zurück zum Zitat Majd RE, Kavarana MN, Bouvette M, Dowling RD: Improved technique to diagnose a patent foramen ovale during left ventricular assist device insertion. Ann Thorac Surg. 2006, 82 (5): 1917-1918. 10.1016/j.athoracsur.2005.09.030.CrossRefPubMed Majd RE, Kavarana MN, Bouvette M, Dowling RD: Improved technique to diagnose a patent foramen ovale during left ventricular assist device insertion. Ann Thorac Surg. 2006, 82 (5): 1917-1918. 10.1016/j.athoracsur.2005.09.030.CrossRefPubMed
15.
Zurück zum Zitat Kilger E, Strom C, Frey L, Felbinger TW, Pichler B, Tichy M, Rank N, Wheeldon D, Kesel K, Schmitz C: Intermittent atrial level right-to-left shunt with temporary hypoxemia in a patient during support with a left ventricular assist device. Acta Anaesthesiol Scand. 2000, 44 (1): 125-127. 10.1034/j.1399-6576.2000.440122.x.CrossRefPubMed Kilger E, Strom C, Frey L, Felbinger TW, Pichler B, Tichy M, Rank N, Wheeldon D, Kesel K, Schmitz C: Intermittent atrial level right-to-left shunt with temporary hypoxemia in a patient during support with a left ventricular assist device. Acta Anaesthesiol Scand. 2000, 44 (1): 125-127. 10.1034/j.1399-6576.2000.440122.x.CrossRefPubMed
16.
Zurück zum Zitat Beitzke D, Wieselthaler G, Schima H, Loewe C: Pulmonary embolism in a patient with a biventricular assist device--imaging with multislice computed tomography. Eur J Cardiothorac Surg. 2011, 39 (3): 415.-10.1016/j.ejcts.2010.07.009.CrossRefPubMed Beitzke D, Wieselthaler G, Schima H, Loewe C: Pulmonary embolism in a patient with a biventricular assist device--imaging with multislice computed tomography. Eur J Cardiothorac Surg. 2011, 39 (3): 415.-10.1016/j.ejcts.2010.07.009.CrossRefPubMed
17.
Zurück zum Zitat Samoukovic G, Malas T, deVarennes B: The role of pulmonary embolectomy in the treatment of acute pulmonary embolism: a literature review from 1968 to 2008. Interact Cardiovasc Thorac Surg. 2010, 11 (3): 265-270. 10.1510/icvts.2009.228361.CrossRefPubMed Samoukovic G, Malas T, deVarennes B: The role of pulmonary embolectomy in the treatment of acute pulmonary embolism: a literature review from 1968 to 2008. Interact Cardiovasc Thorac Surg. 2010, 11 (3): 265-270. 10.1510/icvts.2009.228361.CrossRefPubMed
18.
Zurück zum Zitat Jamieson SW, Kapelanski DP, Sakakibara N, Manecke GR, Thistlethwaite PA, Kerr KM, Channick RN, Fedullo PF, Auger WR: Pulmonary endarterectomy: experience and lessons learned in 1,500 cases. Ann Thorac Surg. 2003, 76 (5): 1457-1462. 10.1016/S0003-4975(03)00828-2.CrossRefPubMed Jamieson SW, Kapelanski DP, Sakakibara N, Manecke GR, Thistlethwaite PA, Kerr KM, Channick RN, Fedullo PF, Auger WR: Pulmonary endarterectomy: experience and lessons learned in 1,500 cases. Ann Thorac Surg. 2003, 76 (5): 1457-1462. 10.1016/S0003-4975(03)00828-2.CrossRefPubMed
Metadaten
Titel
Delayed intracardial shunting and hypoxemia after massive pulmonary embolism in a patient with a biventricular assist device
verfasst von
Thomas Weig
Michael E Dolch
Lorenz Frey
Dirk Bruegger
Peter Boekstegers
Ralf Sodian
Michael Irlbeck
Publikationsdatum
01.12.2011
Verlag
BioMed Central
Erschienen in
Journal of Cardiothoracic Surgery / Ausgabe 1/2011
Elektronische ISSN: 1749-8090
DOI
https://doi.org/10.1186/1749-8090-6-133

Weitere Artikel der Ausgabe 1/2011

Journal of Cardiothoracic Surgery 1/2011 Zur Ausgabe

Häusliche Gewalt in der orthopädischen Notaufnahme oft nicht erkannt

28.05.2024 Häusliche Gewalt Nachrichten

In der Notaufnahme wird die Chance, Opfer von häuslicher Gewalt zu identifizieren, von Orthopäden und Orthopädinnen offenbar zu wenig genutzt. Darauf deuten die Ergebnisse einer Fragebogenstudie an der Sahlgrenska-Universität in Schweden hin.

Fehlerkultur in der Medizin – Offenheit zählt!

28.05.2024 Fehlerkultur Podcast

Darüber reden und aus Fehlern lernen, sollte das Motto in der Medizin lauten. Und zwar nicht nur im Sinne der Patientensicherheit. Eine negative Fehlerkultur kann auch die Behandelnden ernsthaft krank machen, warnt Prof. Dr. Reinhard Strametz. Ein Plädoyer und ein Leitfaden für den offenen Umgang mit kritischen Ereignissen in Medizin und Pflege.

Mehr Frauen im OP – weniger postoperative Komplikationen

21.05.2024 Allgemeine Chirurgie Nachrichten

Ein Frauenanteil von mindestens einem Drittel im ärztlichen Op.-Team war in einer großen retrospektiven Studie aus Kanada mit einer signifikanten Reduktion der postoperativen Morbidität assoziiert.

TAVI versus Klappenchirurgie: Neue Vergleichsstudie sorgt für Erstaunen

21.05.2024 TAVI Nachrichten

Bei schwerer Aortenstenose und obstruktiver KHK empfehlen die Leitlinien derzeit eine chirurgische Kombi-Behandlung aus Klappenersatz plus Bypass-OP. Diese Empfehlung wird allerdings jetzt durch eine aktuelle Studie infrage gestellt – mit überraschender Deutlichkeit.

Update Chirurgie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.

S3-Leitlinie „Diagnostik und Therapie des Karpaltunnelsyndroms“

Karpaltunnelsyndrom BDC Leitlinien Webinare
CME: 2 Punkte

Das Karpaltunnelsyndrom ist die häufigste Kompressionsneuropathie peripherer Nerven. Obwohl die Anamnese mit dem nächtlichen Einschlafen der Hand (Brachialgia parästhetica nocturna) sehr typisch ist, ist eine klinisch-neurologische Untersuchung und Elektroneurografie in manchen Fällen auch eine Neurosonografie erforderlich. Im Anfangsstadium sind konservative Maßnahmen (Handgelenksschiene, Ergotherapie) empfehlenswert. Bei nicht Ansprechen der konservativen Therapie oder Auftreten von neurologischen Ausfällen ist eine Dekompression des N. medianus am Karpaltunnel indiziert.

Prof. Dr. med. Gregor Antoniadis
Berufsverband der Deutschen Chirurgie e.V.

S2e-Leitlinie „Distale Radiusfraktur“

Radiusfraktur BDC Leitlinien Webinare
CME: 2 Punkte

Das Webinar beschäftigt sich mit Fragen und Antworten zu Diagnostik und Klassifikation sowie Möglichkeiten des Ausschlusses von Zusatzverletzungen. Die Referenten erläutern, welche Frakturen konservativ behandelt werden können und wie. Das Webinar beantwortet die Frage nach aktuellen operativen Therapiekonzepten: Welcher Zugang, welches Osteosynthesematerial? Auf was muss bei der Nachbehandlung der distalen Radiusfraktur geachtet werden?

PD Dr. med. Oliver Pieske
Dr. med. Benjamin Meyknecht
Berufsverband der Deutschen Chirurgie e.V.

S1-Leitlinie „Empfehlungen zur Therapie der akuten Appendizitis bei Erwachsenen“

Appendizitis BDC Leitlinien Webinare
CME: 2 Punkte

Inhalte des Webinars zur S1-Leitlinie „Empfehlungen zur Therapie der akuten Appendizitis bei Erwachsenen“ sind die Darstellung des Projektes und des Erstellungswegs zur S1-Leitlinie, die Erläuterung der klinischen Relevanz der Klassifikation EAES 2015, die wissenschaftliche Begründung der wichtigsten Empfehlungen und die Darstellung stadiengerechter Therapieoptionen.

Dr. med. Mihailo Andric
Berufsverband der Deutschen Chirurgie e.V.