Open Access 17.04.2020 | Original Article

Simulation Training in Neuroangiography—Validation and Effectiveness

verfasst von: Kornelia Kreiser, Lea Ströber, Kim G. Gehling, Frederick Schneider, Stefan Kohlbecher, Christian M. Schulz, Claus Zimmer, Jan S. Kirschke

Erschienen in: Clinical Neuroradiology | Ausgabe 2/2021

Abstract

Purpose

Simulators are increasingly used in the training of endovascular procedures; however, for the use of the Mentice vascular interventional system trainer (VIST) simulator in neuroradiology, the validity of the method has not yet been proven. The study was carried out to test the construct validity of such a simulator by demonstrating differences between beginner and expert neurointerventionalists and to evaluate whether a training effect can be demonstrated in repeated cases for different levels of experience.

Methods

In this study 4 experts and 6 beginners performed 10 diagnostic angiographies on the VIST simulator (Mentice AB, Gothenburg, Sweden). Of the cases four were non-recurring, whereas three were repeated once and ten subjects performed all tasks. Additionally, another expert performed only five non-recurring cases. The simulator recorded total time, fluoroscopy time, amount of contrast medium and number of material changes. Furthermore, gaze direction and heart rate were recorded, and subjects completed a questionnaire on workload.

Results

Beginners and experts showed significant differences in total duration time, fluoroscopy time and amount of contrast agent (all p < 0.05). Gaze direction, dwell time and heart rate were similar between both groups. Only beginners improved during training with respect to total duration time, fluoroscopy time and amount of contrast agent. If a case was previously known to them, the total duration and fluoroscopy time were significantly shortened (p < 0.001).

Conclusion

This study demonstrated both the construct validity of a diagnostic neuroangiography simulator as well as a significant training effect for beginners. Therefore, in particular beginner neurointerventionalists should use such simulation tools more extensively in their initial training.
Hinweise
All authors contributed to the study conception and design. Material preparation, data collection and analysis were performed by Lea Ströber, Kim Gehling, Frederick Schneider, Stefan Kohlbecher, Christian Schulz, Jan Kirschke und Kornelia Kreiser. The first draft of the manuscript was written by Kornelia Kreiser and all authors commented on previous versions of the manuscript. All authors read and approved the final manuscript.
Abkürzungen
bpm
Beats per minute
F
French
IQR
Interquartile range
NASA TLX
National Aeronautics and Space Administration Task Load Index
RTLX
RAW TLX (unweighted task load index)
rtPA
Recombinant tissue plasminogen activator
SD
Standard deviation
STL
Stereolithography (file format)
VER
Vertebralis catheter

Background/Introduction

The number of endovascularly treated patients suffering from stroke or aneurysms has steadily increased over the last few years [1]. In 2015, five large randomized stroke trials established the superiority of mechanical thrombectomy in combination with intravenous recombinant tissue plasminogen activator (i.v. rt-PA) over i.v. rt-PA alone [26]. Since then, the number of treated patients has risen continuously. Moreover, the time window has been recently broadened [7, 8], which will further increase the number of patients eligible for interventional treatment. Although mechanical thrombectomy is performed in more and more patients at centers with great expertise, demand is increasing for an on-call service around the clock at multiple smaller centers. Hence, a sufficient number of interventional neuroradiologists must be trained at a high level. Using high-fidelity simulators offers the possibility of structured training without endangering patients [9, 10]. Although the first prototypes of simulators specifically for neuroradiological applications were developed many years ago [11, 12], they are not yet widely used in neuroradiology. After proof of face and content validity, based on expert opinion [13] the next crucial step to establish such a new approach includes the demonstration of its construct validity, i.e. beginners and advanced users can be distinguished via the parameters measured by the simulator. Only when this prerequisite is met should such a simulator be used for training interventionalists. Concerning interventional procedures, construct validity was previously shown for renal [14, 15] and cardiac procedures [16, 17] as well as for carotid stenting [18, 19]. After the validity or neuroangiographies of another make, the ANGIO Mentor Express (Simbionix, Cleveland, OH, USA) has already been proven [20], the present study aimed at establishing the construct validity of the Mentice VIST simulator; however, construct validity is not directly related to the effectiveness of a simulator. To be used as an effective training tool in practice, improvements in psychomotor skills should be measurable [21, 22]. Therefore, the second aim of the study was to quantify improvements of different quality measures of a diagnostic neuroangiography, resulting from repeated simulator training. Of interest was, in addition to an already confirmed training effect for beginners and intermediates [10, 20, 23, 24], whether this impact could also be demonstrated for experts.

Material and Methods

Hardware

All neuroangiographic simulations were performed with the vascular intervention system trainer VIST C from Mentice (Mentice AB, Gothenburg, Sweden), integrated in the VIST LAB. This is a stationary unit whose surface is designed in the form of a patient silhouette (Fig. 1). The VIST LAB consists of a control screen with a touch screen function for selecting scenarios and materials, plus three additional monitors. On two of them, the examiner can follow the live image on two planes and on the third monitor can scroll through the recorded series. A laptop and the simulator itself are positioned under the surface. The only visible detail of the simulator is an insertion sheath at the height of the manikin’s groin, which serves as the access point to the simulator for all materials. Pushing, pulling, and rotating movements of the introduced wires and catheters are detected by three sensors, which can also generate resistance in the sense of force feedback. These movements are transferred into a virtual patient anatomy, consisting of arteries and bones, and displayed on the screens in real time. A foot switch with two pedals, a panel with several options for controlling the virtual C-arms and other settings (shutter, zoom, etc.), and a tube with a syringe for the injection of air, simulating the injection of contrast medium, are also connected to the simulator.
Parameters stored by the device are total duration of the procedure, total fluoroscopy time, number of series, total series time, amount of contrast agent, and number of material changes.

Endovascular Devices

Real angiographic materials, such as a hydrophilic wire (0.035″ Glidewire, Terumo, Somerset, NJ, USA) and a 5F vertebralis diagnostic catheter (VER, Cordis, Santa Clara, CA, USA) were used. All terminal curvatures had been cut off as required by the manufacturer of the simulator.

Software

The source data of magnetic resonance (MR) angiographies of real patients with different anatomical types of the aortic arch were segmented semi-automatically by using IntelliSpacePortal (Philips, Best, The Netherlands). The resulting 3D model was imported to the simulator as a Stereolithography (STL) file. The integrated Case-it software module (Mentice AB, Gothenburg, Sweden) connected the model of the aortic arch and the supra-aortic branches to a template of the descending aorta and the iliac arteries down to the superficial femoral artery.

Accessories

An eye-tracking camera (EyeSeeCam Sci, EyeSeeTec GmbH, München, Germany) was used to record the viewing direction of the test persons. The pulse rate was recorded with a heart rate belt (Zephyr™ BioModule™ Sensor, BioHarnessTM 3, Medtronic, Dublin, Ireland). Free MATLAB®-based (MathWorks®, Natick, MA, USA) ARTiiFACT software [25] was used to evaluate the mean heart rate. The whole set-up is presented in Fig. 1.

Test Persons

Participants were recruited from the radiology and the neuroradiology departments and divided into two groups. The expert group consisted of five physicians with advanced angiography experience from more than 100 cerebral angiographies each. One of them only completed the first five cases and then left the study. These data could only be included in the analysis of validity, but not in the evaluations of the training effect. The beginner group consisted of one medical student without any experience in angiography, three neuroradiological residents with little neuroangiographic experience (<30) and two radiological residents with exclusively peripheral angiographic experience.
The local ethics committee approved the study (172/14), and each participant provided informed consent before participating.

Study Design

Each participant had to complete ten cases. Every case consisted of a complete cerebral angiography in one patient, i.e., the selective probing and visualization of the internal carotid artery, external carotid artery, and vertebral artery with their respective dependent branches on both sides. The ten cases were split in two study parts, where the first part included five unknown cases and the second part included two new cases and three repeated cases from the first part. In detail, the first case was repeated in case six, the fourth case in case nine and the fifth case in case ten. All cases were classified into five levels of difficulty by a neuroradiologist not participating in the study. The classification was based on the following criteria: level 1 was easy to perform with a VER catheter without using a roadmap due to standard anatomy; level 2 was also feasible with a VER catheter, but a roadmap was required; level 3 included an arterial norm variant, i.e., the left vertebral artery originated directly from the aortic arch. In level 4, a sidewinder catheter was required to solve the task. The highest level 5 required advanced catheter maneuvers and changes, plus precise knowledge of the vessel anatomy, variants, and possible pathologies. In each part of the study there was one case of each difficulty level. The participants were informed neither about the level nor about whether the presented case was a known or unknown one.
Before the first session, a technical introduction to the simulator was given, together with a shortened exercise session on a case that was not part of the study. The participant was equipped with a heart rate belt and a head-mounted camera. The individual simulation sessions were limited to a maximum of 2h to avoid any possible fatigue effects. The operation of the simulator beyond the catheter guidance was the responsibility of a student. Each angiography was started with a hydrophilic guidewire (35° curvature) and VER as standard, but the material could be changed at any time on request. At the end of each case, the investigators filled out a questionnaire to record the subjectively perceived workload (NASA-TLX, National Aeronautics and Space Administration Task Load Index, German translation, [26]) as rating with respect to mental demands, physical demands, temporal demands, satisfaction with performance, effort, and frustration. Each rating scale ranged from 0 to 20.

Statistics

Statistical analysis was performed by descriptive and exploratory statistics. Ordinal data (number of series, number of material changes, results of NASA-TLX) and metric data (total duration, total fluoroscopy time, total series time, amount of contrast agent, mean heart rate and percentage of gaze direction lower than 30°) were compared by using a median test or a Mann-Whitney U‑test, respectively. Ordinal data are presented in median (interquartile range, IQR), metric data are presented as mean (± standard deviation, SD) if not stated differently. A p-value < 0.05 was considered as statistically significant. All analyses were performed using IBM SPSS Statistics, Version 23.0 (IBM, Armonk, NY, USA).

Results

Proof of Validation

Comparative results of the experts and beginners of all 10 cases are listed in Table 1. Total duration time, total fluoroscopy time and amount of contrast agent differed significantly. With respect to the subjective task requirements, significant differences were observed in the assessment of satisfaction and effort. Beginners were more dissatisfied and had to make greater efforts than the experts. Nevertheless, heart rate and viewing direction did not differ between experts and beginners in the cases.
Table 1
Comparison of experts and beginners in a summary of all 10 simulations (one expert only performed simulations 1–5)
 
Simulations 1–10
 
Experts (n = 5)
Beginners (n = 6)
p-value
Simulation data:
Mean
±SD
Mean
±SD
Total duration (min)
19.65
7.73
31.96
15.07
p0.001
Fluoroscopy time (min)
11.91
4.59
18.78
7.72
p0.001
Total time of series (min)
1.90
0.80
1.75
0.79
p = 0.269
Contrast agent (ml)
52.04
21.31
67.20
30.82
p=0.013
 
Median
IQR
Median
IQR
Number of series
16
10
16.5
9
p = 0.976
Number of material changes
1
1
2
2
p = 0.056
NASA-TLX (range 1–20):
Median
IQR
Median
IQR
Mental demand
9.5
8
10.5
7
p = 0.795
Physical demand
4.5
4
5
4
p = 0.780
Temporal demand
7.5
7
6
6
p = 0.324
Satisfaction with performance
17
4
15
6
p=0.008
Effort
11
7
15
5
p0.001
Frustration
6
4
5
5
p = 0.095
Physiological data:
Mean
±SD
Mean
±SD
Heart rate (beats per minute/bpm)
79.00
8.92
81.47
13.86
p = 0.669
Viewing direction (%)a
29.38
14.40
30.08
14.60
p = 0.502
Median and IQR for number of series, number of material changes and results of NASA-TLX, p-values of mediantest <0.05 are in bold type. Mean for total duration, fluoroscopy time, total time of series and amount of contrast agent, heart rate and gaze direction, p-values of Mann Whitney U‑test <0.05 are in bold type
aPercentage of deviation of more than 30° from the horizontal view downwards.

Proof of Training Effectiveness

In a comparison of the first five cases with the last five cases participants improved significantly in terms of total duration, fluoroscopy time and perceived effort (total duration 31.94 min (SD 16.41 min) to 21.15 (SD 7.18 min), p < 0.001; fluoroscopy time 18.15 min (SD 8.55 min) to 13.42 min (SD 4.87 min), p = 0.003; effort [median] 15 (IQR 7) to 13 (IQR 7), p = 0.031); however, when results are separated for experts and beginners (Table 2), then this effect remained significant only in beginners. Fig. 2 visualizes the decrease of fluoroscopy time during the sequence of the training, although the degree of difficulty increased from 1 to 5 and 6 to 10 respectively. In addition, the beginners became more satisfied with their performance during the course of training. Similar differences were found when comparing unknown cases with known cases, where total duration time and fluoroscopy time differed significantly and participants were more satisfied with their performance (total duration 35.91 min (SD 16.03 min) to 22.73 min (SD 6.37 min), p < 0.001; fluoroscopy time 20.66 min (SD 7.90 min) to 14.39 min (SD 5.19 min), p = 0.003; satisfaction with performance [median] 15.5 (IQR 5) to 16 (IQR 3), p = 0.022).
Table 2
Comparison of experts and beginners of simulations 1–5 and 6–10 separately
 
Experts
Beginners
 
Simulation (n = 5)
1–5
Simulation 6–10
(n = 4)
p-value
Simulation 1–5 (n = 6)
Simulation 6–10 (n = 6)
p-value
Simulation data:
Mean
±SD
Mean
±SD
Mean
±SD
Mean
±SD
Total duration
(min)
21.88
8.79
16.99
5.29
p=0.045
40.00
16.73
23.92
6.99
p0.001
Fluoroscopy time
(min
12.49
5.27
11.23
3.61
p = 0.465
22.68
7.98
14.87
5.10
p0.001
Total time of series
(min)
1.99
0.84
1.79
0.76
p = 0.588
1.84
0.79
1.67
0.80
p = 0.492
Contrast agent
(ml)
54.05
22.41
49.62
20.20
p = 0.423
75.82
34.14
58.58
24.75
p=0.045
 
Median
IQR
Median
IQR
Median
IQR
Median
IQR
Number of series
17
9
16
12
p = 0.978
17
6
16
11
p = 0.439
Number of material changes
1
1
1
2
p = 0.839
2
2
1.5
1
p = 0.267
NASA-TLX
(range 1–20):
Median
IQR
Median
IQR
Median
IQR
Median
IQR
Mental demand
10
9
10
8
p = 0.762
11.5
9
10
6
p = 0.439
Physical demand
4
4
5
6
p = 0.762
4
2
5.5
5
p = 0.434
Temporal demand
8
8
7
7
p = 0.762
6.5
7
6
5
p = 1.000
Satisfaction with performance
17
2
16
4
p = 0.226
13.5
6
16
2
p=0.009
Effort
11.5
7
11
6
p = 0.978
15.5
3
13.5
5
p = 0.796
Frustration
5.5
5
7
3
p = 0.392
5
7
4
5
p = 0.595
Physiological data:
Mean
±SD
Mean
±SD
Mean
±SD
Mean
±SD
Heart rate (bpm)
78.06
7.74
80.14
10.25
p = 0.220
84.07
14.88
78.86
12.47
p = 0.322
Viewing direction (%)a
27.47
12.00
31.68
16.69
p = 0.377
31.55
18.31
28.60
9.70
p = 0.982
Median and IQR for number of series, number of material changes and results of NASA-TLX, p-values of mediantest <0.05 are in bold type. Mean for total duration, fluoroscopy time, total time of series and amount of contrast agent, heart rate and gaze direction, p-values of Mann Whitney-U-Test <0.05 are in bold type
aPercentage of deviation of more than 30° from the horizontal view downwards.
Again, these differences remained significant only for beginners (Table 3). Neither heart rate, nor viewing direction differed in experts or beginners between known and unknown cases.
Table 3
Comparison of experts and beginners regarding unknown and known cases
 
Experts (n = 4)
Beginners (n = 6)
 
Unknown
cases
Known cases
p-value
Unknown
cases
Known cases
p-value
Simulation data:
Mean
±SD
Mean
±SD
Mean
±SD
Mean
±SD
Total duration
(min
20.90
8.42
16.34
4.19
p = 0.131
35.91
16.03
22.73
6.37
p0.001
Fluoroscopy time
(min)
12.36
5.03
10.71
2.98
p = 0.397
20.66
7.90
14.39
5.19
p=0.003
Total time of
series (min)
1.93
0.83
1.83
074
p = 0.765
1.81
0.78
1.64
0.82
p = 0.529
Contrast agent
(ml)
52.84
22.04
49.89
19.98
p = 0.745
71.95
30.90
56.12
28.43
p = 0.058
 
Median
IQR
Median
IQR
Median
IQR
Median
IQR
Number of series
17
12
16
7
p = 0.878
17
7
16
11
p = 0.398
Number of material changes
1
1
1
2
p = 0.673
2
2
1
1
p = 0.183
NASA-TLX
(range 1–20):
Median
IQR
Median
IQR
Median
IQR
Median
IQR
Mental demand
10
9
8
8
p = 0.735
11.5
9
8
6
p = 0.398
Physical demand
4
4
5
6
p = 0.735
5
4
5
4
p = 0.865
Temporal demand
8
7
7
7
p = 0.735
6.5
7
6
5
p = 0.910
Satisfaction with performance
17
3
15.5
4
p = 0.338
14
6
16
2
p=0.022
Effort
12.5
5
9.5
6
p = 0.406
15.5
3
13
7
p = 0.611
Frustration
6
5
6.5
2
p = 0.975
5
8
4
3
p = 0.164
Physiological data:
Mean
±SD
Mean
±SD
Mean
±SD
Mean
±SD
Heart rate (bpm)
78.77
8.41
79.61
10.54
p = 0.524
82.4
14.30
79.09
12.85
p = 0.594
Viewing direction (%)a
27.75
13.36
33.73
16.38
p = 0.234
30.79
16.43
28.41
9.23
p = 0.949
Median and IQR for number of series, number of material changes and results of NASA-TLX, p-values of median test <0.05 are in bold type. Mean for total duration, fluoroscopy time, total time of series and amount of contrast agent, heart rate and gaze direction, p-values of Mann Whitney U‑test <0.05 are in bold type
aPercentage of deviation of more than 30° from the horizontal view downwards

Discussion

In this study it was possible to prove the construct validity of a neuroangiographic simulator by demonstrating significant performance differences between beginners and experts. We were furthermore able to show the effectiveness of such a simulator by directly measuring significant improvements in psychomotor skills of beginners in cases derived from real patient anatomy.
This is in line with previous studies on different scenarios, such as carotid artery stenting [13, 18] and several infra-aortic applications [9, 17] and addresses one major concern regarding neuroangiographic simulations: In 2008, Carroll and Messenger stated that “medical simulation has made the transition from an experimental technology to the clinical world”, and that “perhaps the most pressing issue […] regarding medical simulation is validation” [27].
To prove the construct validity, we had clearly separated groups of beginners with a maximum of 30 cerebral angiographies performed and of experts with at least 100 procedures. The simulator data for all 10 procedures evidently demonstrated this subdivision. As previously confirmed for cardiac angiography [28], beginners need more time to find and examine the target vessels. Thus, total duration and fluoroscopy time differed significantly. Beginners more often produced roadmaps; accordingly, the total amount of contrast agent was significantly higher for beginners. The fact that the simulator only shows the contrast of the arteries and contains no parenchyma or the venous phase may serve as a reason for the missing difference in the duration of the series. Thus, in this parameter no difference was present between experts and beginners. Overall, these results contradict the findings of Nguyen et al. who were only able to identify the amount of contrast agent as a distinguishing feature for the experience level [20]. A possible reason could be their small number of 2 compared tasks, while our study design consisted of 10 procedures.
Observations of gaze direction, blink frequency, pupil size and dwell time are recognized means of examining attention and cognitive stress [29]. Based on such data Richstone et al. were able to distinguish unequivocally between beginners and experts in surgical laparoscopy [30], but for endovascular cardiac interventions, Currie et al. found hardly any differences [31]. The assumption that experts would turn their gaze less often away from the X‑ray screen could therefore not be confirmed by them. The present study focused on the direction of each subject’s gaze; no differences between the two groups could be observed.
Heart rate was recorded, as it was assumed that it would differ between various levels of workload and thus between beginners and experts according to the effort made. In previous work on anesthesiologists, Martin et al. as well as Weinger et al. found significant differences in several heart rate parameters, including mean heart rate, during different phases of an anesthetic procedure [32, 42]. In a later work of the first group on pre-hospital emergency medicine, heart rate variability discriminated better between different levels of workload compared to the mean heart rate [33]. Currie et al. also measured numerous parameters such as heart rate variability, electrodermal activity and skin temperature in physicians undertaking cardiac endovascular procedures but could not find differences between experience levels [31]. Hence, mean heart rate seems to be an unreliable predictor, matching the results of our study, where it was not suitable to distinguish between beginners and experts.
The NASA-TLX is a frequently used tool to assess subjective workload, especially in anesthesia and the field of emergency care [34], but also in other areas, e.g. radiotherapy [35] or flight simulation [36]. In comparison with the previously mentioned complex technical tools, the NASA-TLX offers an easy and fast method for recording the workload. The modified version Raw-TLX (RTLX), without the weighting process of the subscales, is even easier to use [37]. In a study on surgical robotics, differences were found in NASA RTLX by examiners with different experience levels [38], just as in our participants. Apparently, there are no data on whether NASA RTLX changes differently through training of beginners compared to experts. The satisfaction of our participants only increased when the other measurable parameters also improved, which was only true for beginners. We therefore assume that the simulator is well suitable to give adequate feedback on one’s own performance.
An increase in performance with respect to the two essential parameters of total duration and fluoroscopy time has been demonstrated in our study for beginners and confirmed the results of Spiotta et al. and Zaika et al. [10, 24]. Among our participants, however, this effect could not be observed among experts. Thus, not every experience level benefits from simulator training that focuses exclusively on diagnostic angiography. This supports the assumption that in the best case skills are learned that an expert already possesses. For beginners, the training effect can be objectively read from the metrics of the simulator, as known for peripheral endovascular procedures [14, 39]. In addition, we have shown that repeated exercises of the same cases are helpful for beginners not only to increase their metric values, but also to increase their own satisfaction. Spiotta et al. also noted an increase in confidence in the acquisition of skills, both in terms of knowledge of the anatomy and the technique of vessel selection [23]. Not least for this reason, several centers have now begun to implement a structured milestone-based curriculum and propose to integrate simulation training into formal neuroendovascular training [23, 40]
When demonstrating a training effect, it is always difficult to distinguish the extent to which a real gain in specific psychomotor skills overlaps with increasing familiarity with the operation of the simulator [39]. An improvement merely by habituation would be recognizable in all subjects; however, since only the group of beginners improved their performance relevantly, this effect seems to be negligible here.
Notably, in the RTLX evaluation, the degree of frustration of the experts slightly increased during training, whereas it decreased among the beginners (both not significant). Comments from the participants showed that individual experts were disturbed by the differences between simulation and reality, whereas beginners were not. The participants of a simulator training should therefore be informed in advance with respect to the differences that can be expected in the behavior of the simulator in relation to reality. Otherwise, they approach the training with a variety of ideas and demands, and frustration and anger can easily arise [1].

Limitations

Construct validity should distinguish not only between beginners and experts, but ideally also between various levels of experience [16]. In our study, the small number of test persons prevented further division of study participants. Also, the number of cases, in particular of repeated cases was limited in this study. A higher number of cases might also show differences in expert neuroradiologists.

Conclusion

Construct validity of a high-tech simulator could be demonstrated for diagnostic neuroangiography and especially beginners showed a measurable training effect through repeated practice. Further studies should demonstrate the benefit of such simulation training for the patient.

Compliance with ethical guidelines

Conflict of interest

K. Kreiser conducts research in areas of interest similar to the business interests of Mentice AB (Gothenburg, Sweden). She gets minor travel grants, but no personal fees. She consults for phenox and has received speaker fees from Philips, Penumbra and Stryker. The Department of Neuroradiology has received research support and investigator fees for clinical studies from Biogen Idec, Quintiles, MSD Sharp & Dome, Boehringer Ingelheim, Inventive Health Clinical UK Ltd., Advance Cor, Brainsgate, Pfizer, Bayer-Schering, Novartis, Roche, Servier, Penumbra, WCT GmbH, Syngis, SSS International Clinical Research, PPD Germany GmbH, Worldwide Clinical Trials Ltd., Phenox, Covidien, Actelion, Medivation, Medtronic, Harrison Clinical Research, Concentric, Pharmtrace, Reverse Medical Corp., Premier Research Germany Ltd., Surpass Medical Ltd., GlaxoSmithKline, AXON Neuroscience, Bristol-Myers Squibb, Genentech, Acandis, EISAI, NeuroRx, Italfarmaco, Bioclinica, MIAC and IXICO. No patents issued and pending. C. Zimmer disclosed no relevant relationships regarding activities related to the present article. He has served on scientific advisory boards for Philips and Bayer Schering; serves as co-editor on the Advisory Board of Clinical Neuroradiology; has received speaker fees from Bayer-Schering and Philips. J.S. Kirschke received speaker fees from Philips. L. Ströber, K.G. Gehling, F. Schneider, S. Kohlbecher and C.M. Schulz declare that they have no competing interests.

Ethical standards

The study has been approved by the local ethics committee (172/14). All procedures performed in studies involving human participants or on human tissue were in accordance with the ethical standards of the institutional and/or national research committee and with the 1975 Helsinki declaration and its later amendments or comparable ethical standards. Informed consent was obtained from all individual participants included in the study. The manuscript does not contain clinical studies.
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article’s Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article’s Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://​creativecommons.​org/​licenses/​by/​4.​0/​.

Unsere Produktempfehlungen

e.Med Interdisziplinär

Kombi-Abonnement

Jetzt e.Med zum Sonderpreis bestellen!

Für Ihren Erfolg in Klinik und Praxis - Die beste Hilfe in Ihrem Arbeitsalltag

Mit e.Med Interdisziplinär erhalten Sie Zugang zu allen CME-Fortbildungen und Fachzeitschriften auf SpringerMedizin.de.

Jetzt bestellen und 100 € sparen!

e.Med Neurologie & Psychiatrie

Kombi-Abonnement

Mit e.Med Neurologie & Psychiatrie erhalten Sie Zugang zu CME-Fortbildungen der Fachgebiete, den Premium-Inhalten der dazugehörigen Fachzeitschriften, inklusive einer gedruckten Zeitschrift Ihrer Wahl.

e.Med Neurologie

Kombi-Abonnement

Mit e.Med Neurologie erhalten Sie Zugang zu CME-Fortbildungen des Fachgebietes, den Premium-Inhalten der neurologischen Fachzeitschriften, inklusive einer gedruckten Neurologie-Zeitschrift Ihrer Wahl.

Weitere Produktempfehlungen anzeigen
Literatur
1.
Zurück zum Zitat Berlis A, Morhard D, Weber W. On the basis of the DeGIR/DGNR register nationwide care for acute ischemic stroke patients in 2016 and 2017 using mechanical thrombectomy by radiologists and neuroradiologists. Rofo. 2019;191:613–7.PubMed Berlis A, Morhard D, Weber W. On the basis of the DeGIR/DGNR register nationwide care for acute ischemic stroke patients in 2016 and 2017 using mechanical thrombectomy by radiologists and neuroradiologists. Rofo. 2019;191:613–7.PubMed
2.
Zurück zum Zitat Saver JL, Goyal M, Bonafe A, Diener HC, Levy EI, Pereira VM, et al. Stent-retriever thrombectomy after intravenous t‑PA vs. t‑PA alone in stroke. N Engl J Med. 2015;372:2285–95.PubMed Saver JL, Goyal M, Bonafe A, Diener HC, Levy EI, Pereira VM, et al. Stent-retriever thrombectomy after intravenous t‑PA vs. t‑PA alone in stroke. N Engl J Med. 2015;372:2285–95.PubMed
3.
Zurück zum Zitat Jovin TG, Chamorro A, Cobo E, de Miquel MA, Molina CA, Rovira A, et al. Thrombectomy within 8 hours after symptom onset in ischemic stroke. N Engl J Med. 2015;372:2296–306.PubMed Jovin TG, Chamorro A, Cobo E, de Miquel MA, Molina CA, Rovira A, et al. Thrombectomy within 8 hours after symptom onset in ischemic stroke. N Engl J Med. 2015;372:2296–306.PubMed
4.
Zurück zum Zitat Goyal M, Demchuk AM, Menon BK, Eesa M, Rempel JL, Thornton J, et al. Randomized assessment of rapid endovascular treatment of ischemic stroke. N Engl J Med. 2015;372:1019–30.PubMed Goyal M, Demchuk AM, Menon BK, Eesa M, Rempel JL, Thornton J, et al. Randomized assessment of rapid endovascular treatment of ischemic stroke. N Engl J Med. 2015;372:1019–30.PubMed
5.
Zurück zum Zitat Berkhemer OA, Fransen PS, Beumer D, van den Berg LA, Lingsma HF, Yoo AJ, et al. A randomized trial of intraarterial treatment for acute ischemic stroke. N Engl J Med. 2015;372:11–20.PubMed Berkhemer OA, Fransen PS, Beumer D, van den Berg LA, Lingsma HF, Yoo AJ, et al. A randomized trial of intraarterial treatment for acute ischemic stroke. N Engl J Med. 2015;372:11–20.PubMed
6.
Zurück zum Zitat Campbell BC, Mitchell PJ, Kleinig TJ, Dewey HM, Churilov L, Yassi N, et al. Endovascular therapy for ischemic stroke with perfusion-imaging selection. N Engl J Med. 2015;372:1009–18.PubMed Campbell BC, Mitchell PJ, Kleinig TJ, Dewey HM, Churilov L, Yassi N, et al. Endovascular therapy for ischemic stroke with perfusion-imaging selection. N Engl J Med. 2015;372:1009–18.PubMed
7.
Zurück zum Zitat Nogueira RG, Jadhav AP, Haussen DC, Bonafe A, Budzik RF, Bhuva P, et al. Thrombectomy 6 to 24 hours after stroke with a mismatch between deficit and infarct. N Engl J Med. 2018;378:11–21.PubMed Nogueira RG, Jadhav AP, Haussen DC, Bonafe A, Budzik RF, Bhuva P, et al. Thrombectomy 6 to 24 hours after stroke with a mismatch between deficit and infarct. N Engl J Med. 2018;378:11–21.PubMed
8.
Zurück zum Zitat Albers GW, Marks MP, Kemp S, Christensen S, Tsai JP, Ortega-Gutierrez S, et al. Thrombectomy for stroke at 6 to 16 hours with selection by perfusion imaging. N Engl J Med. 2018;378:708–18.PubMedPubMedCentral Albers GW, Marks MP, Kemp S, Christensen S, Tsai JP, Ortega-Gutierrez S, et al. Thrombectomy for stroke at 6 to 16 hours with selection by perfusion imaging. N Engl J Med. 2018;378:708–18.PubMedPubMedCentral
9.
Zurück zum Zitat Schimmel DR, Sweis R, Cohen ER, Davidson C, Wayne DB. Targeting clinical outcomes: Endovascular simulation improves diagnostic coronary angiography skills. Catheter Cardiovasc Interv. 2016;87:383–8.PubMed Schimmel DR, Sweis R, Cohen ER, Davidson C, Wayne DB. Targeting clinical outcomes: Endovascular simulation improves diagnostic coronary angiography skills. Catheter Cardiovasc Interv. 2016;87:383–8.PubMed
10.
Zurück zum Zitat Spiotta AM, Rasmussen PA, Masaryk TJ, Benzel EC, Schlenk R. Simulated diagnostic cerebral angiography in neurosurgical training: a pilot program. J Neurointerv Surg. 2013;5:376–81.PubMed Spiotta AM, Rasmussen PA, Masaryk TJ, Benzel EC, Schlenk R. Simulated diagnostic cerebral angiography in neurosurgical training: a pilot program. J Neurointerv Surg. 2013;5:376–81.PubMed
11.
Zurück zum Zitat Chui CK, Li Z, Anderson JH, Murphy K, Venbrux A, Ma X, et al. Training and pretreatment planning of interventional neuroradiology procedures—initial clinical validation. Stud Health Technol Inform. 2002;85:96–102.PubMed Chui CK, Li Z, Anderson JH, Murphy K, Venbrux A, Ma X, et al. Training and pretreatment planning of interventional neuroradiology procedures—initial clinical validation. Stud Health Technol Inform. 2002;85:96–102.PubMed
12.
Zurück zum Zitat Wu X, Pegoraro V, Luboz V, Neumann PF, Bardsley R, Dawson S, et al. New approaches to computer-based interventional neuroradiology training. Stud Health Technol Inform. 2005;111:602–7.PubMed Wu X, Pegoraro V, Luboz V, Neumann PF, Bardsley R, Dawson S, et al. New approaches to computer-based interventional neuroradiology training. Stud Health Technol Inform. 2005;111:602–7.PubMed
13.
Zurück zum Zitat Nicholson WJ, Cates CU, Patel AD, Niazi K, Palmer S, Helmy T, et al. Face and content validation of virtual reality simulation for carotid angiography: results from the first 100 physicians attending the Emory NeuroAnatomy Carotid Training (ENACT) program. Simul Healthc. 2006;1:147–50.PubMed Nicholson WJ, Cates CU, Patel AD, Niazi K, Palmer S, Helmy T, et al. Face and content validation of virtual reality simulation for carotid angiography: results from the first 100 physicians attending the Emory NeuroAnatomy Carotid Training (ENACT) program. Simul Healthc. 2006;1:147–50.PubMed
14.
Zurück zum Zitat Coates PJ, Zealley IA, Chakraverty S. Endovascular simulator is of benefit in the acquisition of basic skills by novice operators. J Vasc Interv Radiol. 2010;21:130–4.PubMed Coates PJ, Zealley IA, Chakraverty S. Endovascular simulator is of benefit in the acquisition of basic skills by novice operators. J Vasc Interv Radiol. 2010;21:130–4.PubMed
15.
Zurück zum Zitat Aggarwal R, Black SA, Hance JR, Darzi A, Cheshire NJ. Virtual reality simulation training can improve inexperienced surgeons’ endovascular skills. Eur J Vasc Endovasc Surg. 2006;31:588–93.PubMed Aggarwal R, Black SA, Hance JR, Darzi A, Cheshire NJ. Virtual reality simulation training can improve inexperienced surgeons’ endovascular skills. Eur J Vasc Endovasc Surg. 2006;31:588–93.PubMed
16.
Zurück zum Zitat Jensen UJ, Jensen J, Olivecrona GK, Ahlberg G, Tornvall P. Technical skills assessment in a coronary angiography simulator for construct validation. Simul Healthc. 2013;8:324–8.PubMed Jensen UJ, Jensen J, Olivecrona GK, Ahlberg G, Tornvall P. Technical skills assessment in a coronary angiography simulator for construct validation. Simul Healthc. 2013;8:324–8.PubMed
17.
Zurück zum Zitat Lipner RS, Messenger JC, Kangilaski R, Baim DS, Holmes DR Jr, Williams DO, et al. A technical and cognitive skills evaluation of performance in interventional cardiology procedures using medical simulation. Simul Healthc. 2010;5:65–74.PubMed Lipner RS, Messenger JC, Kangilaski R, Baim DS, Holmes DR Jr, Williams DO, et al. A technical and cognitive skills evaluation of performance in interventional cardiology procedures using medical simulation. Simul Healthc. 2010;5:65–74.PubMed
18.
Zurück zum Zitat Berry M, Reznick R, Lystig T, Lonn L. The use of virtual reality for training in carotid artery stenting: a construct validation study. Acta Radiol. 2008;49:801–5.PubMed Berry M, Reznick R, Lystig T, Lonn L. The use of virtual reality for training in carotid artery stenting: a construct validation study. Acta Radiol. 2008;49:801–5.PubMed
19.
Zurück zum Zitat Van Herzeele I, Aggarwal R, Choong A, Brightwell R, Vermassen FE, Cheshire NJ. Virtual reality simulation objectively differentiates level of carotid stent experience in experienced interventionalists. J Vasc Surg. 2007;46:855–63.PubMed Van Herzeele I, Aggarwal R, Choong A, Brightwell R, Vermassen FE, Cheshire NJ. Virtual reality simulation objectively differentiates level of carotid stent experience in experienced interventionalists. J Vasc Surg. 2007;46:855–63.PubMed
20.
Zurück zum Zitat Nguyen N, Eagleson R, Boulton M, de Ribaupierre S. Realism, criterion validity, and training capability of simulated diagnostic cerebral angiography. Stud Health Technol Inform. 2014;196:297–303.PubMed Nguyen N, Eagleson R, Boulton M, de Ribaupierre S. Realism, criterion validity, and training capability of simulated diagnostic cerebral angiography. Stud Health Technol Inform. 2014;196:297–303.PubMed
21.
Zurück zum Zitat Ahmed K, Keeling AN, Fakhry M, Ashrafian H, Aggarwal R, Naughton PA, Darzi A, Cheshire N, Athanasiou T, Hamady M. Role of virtual reality simulation in teaching and assessing technical skills in endovascular intervention. J Vasc Interv Radiol. 2010;21:55–66.PubMed Ahmed K, Keeling AN, Fakhry M, Ashrafian H, Aggarwal R, Naughton PA, Darzi A, Cheshire N, Athanasiou T, Hamady M. Role of virtual reality simulation in teaching and assessing technical skills in endovascular intervention. J Vasc Interv Radiol. 2010;21:55–66.PubMed
22.
Zurück zum Zitat Liebig T, Holtmannspötter M, Crossley R, Lindkvist J, Henn P, Lönn L, et al. Metric-based virtual reality simulation: a paradigm shift in training for mechanical thrombectomy in acute stroke. Stroke. 2018;49:e239–42.PubMed Liebig T, Holtmannspötter M, Crossley R, Lindkvist J, Henn P, Lönn L, et al. Metric-based virtual reality simulation: a paradigm shift in training for mechanical thrombectomy in acute stroke. Stroke. 2018;49:e239–42.PubMed
23.
Zurück zum Zitat Spiotta AM, Kellogg RT, Vargas J, Chaudry MI, Turk AS, Turner RD. Diagnostic angiography skill acquisition with a secondary curve catheter: phase 2 of a curriculum-based endovascular simulation program. J Neurointerv Surg. 2015;7:777–80.PubMed Spiotta AM, Kellogg RT, Vargas J, Chaudry MI, Turk AS, Turner RD. Diagnostic angiography skill acquisition with a secondary curve catheter: phase 2 of a curriculum-based endovascular simulation program. J Neurointerv Surg. 2015;7:777–80.PubMed
24.
Zurück zum Zitat Zaika O, Nguyen N, Boulton M, Eagleson R, de Ribaupierre S. Evaluation of user performance in simulation-based diagnostic cerebral angiography training. Stud Health Technol Inform. 2016;220:465–8.PubMed Zaika O, Nguyen N, Boulton M, Eagleson R, de Ribaupierre S. Evaluation of user performance in simulation-based diagnostic cerebral angiography training. Stud Health Technol Inform. 2016;220:465–8.PubMed
25.
Zurück zum Zitat Kaufmann T, Sütterlin S, Schulz SM, Vögele C. ARTiiFACT: a tool for heart rate artifact processing and heart rate variability analysis. Behav Res Methods. 2011;43:1161–70.PubMed Kaufmann T, Sütterlin S, Schulz SM, Vögele C. ARTiiFACT: a tool for heart rate artifact processing and heart rate variability analysis. Behav Res Methods. 2011;43:1161–70.PubMed
26.
Zurück zum Zitat Hart SG, Staveland LE. Development of NASA-TLX (Task Load Index): results of empirical and theoretical research. Adv Psychol. 1988;52:139–83. Hart SG, Staveland LE. Development of NASA-TLX (Task Load Index): results of empirical and theoretical research. Adv Psychol. 1988;52:139–83.
27.
Zurück zum Zitat Carroll JD, Messenger JC. Medical simulation: the new tool for training and skill assessment. Perspect Biol Med. 2008;51:47–60.PubMed Carroll JD, Messenger JC. Medical simulation: the new tool for training and skill assessment. Perspect Biol Med. 2008;51:47–60.PubMed
28.
Zurück zum Zitat Jensen UJ, Lagerquist B, Jensen J, Tornvall P. The use of fluoroscopy to construct learning curves for coronary angiography. Catheter Cardiovasc Interv. 2012;80:564–9.PubMed Jensen UJ, Lagerquist B, Jensen J, Tornvall P. The use of fluoroscopy to construct learning curves for coronary angiography. Catheter Cardiovasc Interv. 2012;80:564–9.PubMed
29.
Zurück zum Zitat Marshall SP. The Index of Cognitive Activity: measuring cognitive workload. Proceedings of the IEEE 7th Conference on Human Factors and Power Plants, Scottsdale, AZ, USA, 2002, pp. 7–7. Marshall SP. The Index of Cognitive Activity: measuring cognitive workload. Proceedings of the IEEE 7th Conference on Human Factors and Power Plants, Scottsdale, AZ, USA, 2002, pp. 7–7.
30.
Zurück zum Zitat Richstone L, Schwartz MJ, Seideman C, Cadeddu J, Marshall S, Kavoussi LR. Eye metrics as an objective assessment of surgical skill. Ann Surg. 2010;252:177–82.PubMed Richstone L, Schwartz MJ, Seideman C, Cadeddu J, Marshall S, Kavoussi LR. Eye metrics as an objective assessment of surgical skill. Ann Surg. 2010;252:177–82.PubMed
31.
Zurück zum Zitat Currie J, Bond RR, McCullagh P, Black P, Finlay DD, Gallagher S, et al. Wearable technology-based metrics for predicting operator performance during cardiac catheterisation. Int J Comput Assist Radiol Surg. 2019;14:645–57.PubMedPubMedCentral Currie J, Bond RR, McCullagh P, Black P, Finlay DD, Gallagher S, et al. Wearable technology-based metrics for predicting operator performance during cardiac catheterisation. Int J Comput Assist Radiol Surg. 2019;14:645–57.PubMedPubMedCentral
32.
Zurück zum Zitat Martin J, Schneider F, Kowalewskij A, Jordan D, Hapfelmeier A, Kochs EF, et al. Linear and non-linear heart rate metrics for the assessment of anaesthetists’ workload during general anaesthesia. Br J Anaesth. 2016;117:767–74.PubMed Martin J, Schneider F, Kowalewskij A, Jordan D, Hapfelmeier A, Kochs EF, et al. Linear and non-linear heart rate metrics for the assessment of anaesthetists’ workload during general anaesthesia. Br J Anaesth. 2016;117:767–74.PubMed
33.
Zurück zum Zitat Weinger MB, Reddy SB, Slagle JM. Multiple measures of anesthesia workload during teaching and nonteaching cases. Anesth Analg. 2004;98:1419–25.PubMed Weinger MB, Reddy SB, Slagle JM. Multiple measures of anesthesia workload during teaching and nonteaching cases. Anesth Analg. 2004;98:1419–25.PubMed
34.
Zurück zum Zitat Schneider F, Martin J, Hapfelmeier A, Jordan D, Schneider G, Schulz CM. The validity of linear and non-linear heart rate metrics as workload indicators of emergency physicians. PLoS One. 2017;12:e188635.PubMedPubMedCentral Schneider F, Martin J, Hapfelmeier A, Jordan D, Schneider G, Schulz CM. The validity of linear and non-linear heart rate metrics as workload indicators of emergency physicians. PLoS One. 2017;12:e188635.PubMedPubMedCentral
35.
Zurück zum Zitat Schneider F, Martin J, Schneider G, Schulz CM. The impact of the patient’s initial NACA score on subjective and physiological indicators of workload during pre-hospital emergency care. Plos One. 2018;13:e202215.PubMedPubMedCentral Schneider F, Martin J, Schneider G, Schulz CM. The impact of the patient’s initial NACA score on subjective and physiological indicators of workload during pre-hospital emergency care. Plos One. 2018;13:e202215.PubMedPubMedCentral
36.
Zurück zum Zitat Mazur LM, Mosaly PR, Jackson M, Chang SX, Burkhardt KD, Adams RD, et al. Quantitative assessment of workload and stressors in clinical radiation oncology. Int J Radiat Oncol Biol Phys. 2012;83:e571–6.PubMed Mazur LM, Mosaly PR, Jackson M, Chang SX, Burkhardt KD, Adams RD, et al. Quantitative assessment of workload and stressors in clinical radiation oncology. Int J Radiat Oncol Biol Phys. 2012;83:e571–6.PubMed
37.
Zurück zum Zitat Mansikka H, Virtanen K, Harris D. Comparison of NASATLX scale, modified cooper-harper scale and mean inter-beat interval as measures of pilot mental workload during simulated flight tasks. Ergonomics. 2019;62:246–54.PubMed Mansikka H, Virtanen K, Harris D. Comparison of NASATLX scale, modified cooper-harper scale and mean inter-beat interval as measures of pilot mental workload during simulated flight tasks. Ergonomics. 2019;62:246–54.PubMed
38.
Zurück zum Zitat Hart SG. Nasa-Task Load Index (NASA-TLX); 20 Years Later. Proceedings of the Human Factors and Ergonomics Society Annual Meeting, 2006;50:904–8. Hart SG. Nasa-Task Load Index (NASA-TLX); 20 Years Later. Proceedings of the Human Factors and Ergonomics Society Annual Meeting, 2006;50:904–8.
39.
Zurück zum Zitat Dulan G, Rege RV, Hogg DC, Gilberg-Fisher KM, Arain NA, Tesfay ST, Scott DJ. Proficiency-based training for robotic surgery: construct validity, workload, and expert levels for nin inanimate exercises. Surg Endosc. 2012;26:1516–21.PubMed Dulan G, Rege RV, Hogg DC, Gilberg-Fisher KM, Arain NA, Tesfay ST, Scott DJ. Proficiency-based training for robotic surgery: construct validity, workload, and expert levels for nin inanimate exercises. Surg Endosc. 2012;26:1516–21.PubMed
40.
Zurück zum Zitat Glaiberman CB, Jacobs B, Street M, Duncan JR, Scerbo MW, Pilgrim TK. Simulation in training: one-year experience using an efficiency index to assess interventional radiology fellow training status. J Vasc Interv Radiol. 2008;19:1366e71. Glaiberman CB, Jacobs B, Street M, Duncan JR, Scerbo MW, Pilgrim TK. Simulation in training: one-year experience using an efficiency index to assess interventional radiology fellow training status. J Vasc Interv Radiol. 2008;19:1366e71.
41.
Zurück zum Zitat Kreiser K, Gehling K, Zimmer C. Simulation in angiography—experiences from 5 years teaching, training, and research. Rofo. 2019;191:547–52.PubMed Kreiser K, Gehling K, Zimmer C. Simulation in angiography—experiences from 5 years teaching, training, and research. Rofo. 2019;191:547–52.PubMed
42.
Zurück zum Zitat Dieckmann P, Wehner T. Über Grundsätze zur Gestaltung von Simulatorsettings für Forschung und Lehre. Harburger Beiträge zur Psychologie und Soziologie der Arbeit, Vol. 31. 2002. Dieckmann P, Wehner T. Über Grundsätze zur Gestaltung von Simulatorsettings für Forschung und Lehre. Harburger Beiträge zur Psychologie und Soziologie der Arbeit, Vol. 31. 2002.
Metadaten
Titel
Simulation Training in Neuroangiography—Validation and Effectiveness
verfasst von
Kornelia Kreiser
Lea Ströber
Kim G. Gehling
Frederick Schneider
Stefan Kohlbecher
Christian M. Schulz
Claus Zimmer
Jan S. Kirschke
Publikationsdatum
17.04.2020
Verlag
Springer Berlin Heidelberg
Erschienen in
Clinical Neuroradiology / Ausgabe 2/2021
Print ISSN: 1869-1439
Elektronische ISSN: 1869-1447
DOI
https://doi.org/10.1007/s00062-020-00902-5

Weitere Artikel der Ausgabe 2/2021

Clinical Neuroradiology 2/2021 Zur Ausgabe

Mammakarzinom: Brustdichte beeinflusst rezidivfreies Überleben

26.05.2024 Mammakarzinom Nachrichten

Frauen, die zum Zeitpunkt der Brustkrebsdiagnose eine hohe mammografische Brustdichte aufweisen, haben ein erhöhtes Risiko für ein baldiges Rezidiv, legen neue Daten nahe.

„Übersichtlicher Wegweiser“: Lauterbachs umstrittener Klinik-Atlas ist online

17.05.2024 Klinik aktuell Nachrichten

Sie sei „ethisch geboten“, meint Gesundheitsminister Karl Lauterbach: mehr Transparenz über die Qualität von Klinikbehandlungen. Um sie abzubilden, lässt er gegen den Widerstand vieler Länder einen virtuellen Klinik-Atlas freischalten.

Klinikreform soll zehntausende Menschenleben retten

15.05.2024 Klinik aktuell Nachrichten

Gesundheitsminister Lauterbach hat die vom Bundeskabinett beschlossene Klinikreform verteidigt. Kritik an den Plänen kommt vom Marburger Bund. Und in den Ländern wird über den Gang zum Vermittlungsausschuss spekuliert.

Darf man die Behandlung eines Neonazis ablehnen?

08.05.2024 Gesellschaft Nachrichten

In einer Leseranfrage in der Zeitschrift Journal of the American Academy of Dermatology möchte ein anonymer Dermatologe bzw. eine anonyme Dermatologin wissen, ob er oder sie einen Patienten behandeln muss, der eine rassistische Tätowierung trägt.

Update Radiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.