Introduction
Clinical indications for the performance of VCUG
Operator issues: the radiologist’s VCUG technique to reduce radiation exposure
Modification or optimization of the fluoroscope
Current research in pulsed fluoroscopy
Variable | Continuous fluoroscopy | Grid-controlled variable-rate pulsed fluoroscopy |
F test |
P value |
---|---|---|---|---|
Total radiation exposure (mGy) | ||||
Group 1 | 0.60±0.05 | 0.13±0.05 | 62.87 | 0.004* |
Group 2 | 1.71±0.10 | 0.26±0.04 | 342.97 | <0.001* |
Group 3 | 3.92±0.20 | 0.52±0.07 | 357.85 | <0.001* |
Fluoroscopic radiation exposure (mGy) | ||||
Group 1 | 0.32±0.04 | 0.07±0.03 | 22.99 | 0.17* |
Group 2 | 1.05±0.13 | 0.16±0.04 | 105.91 | 0.02* |
Group 3 | 2.31±0.47 | 0.26±0.03 | 72.03 | <0.01* |
Recorded image radiation exposure (mGy) | ||||
Group 1 | 0.28±0.02 | 0.07±0.03 | 121.02 | 0.002* |
Group 2 | 0.77±0.06 | 0.10±0.01 | 113.35 | 0.002* |
Group 3 | 1.62±0.18 | 0.26±0.03 | 72.03 | 0.003* |
Fluoroscopy time (min) | ||||
Group 1 | 0.75±0.05 | 1.02±0.15 | 2.50 | 0.21 |
Group 2 | 1.18±0.14 | 1.34±0.21 | 2.40 | 0.22 |
Group 3 | 1.35±0.32 | 1.45±0.23 | 0.07 | 0.82 |
Radiologist and girth group | Continuous fluoroscopy | Grid-controlled variable-rate pulsed fluoroscopy |
P value |
---|---|---|---|
Radiologist A | |||
Group 1 | 1.5 (1–) | 1 (1–) | 0.04* |
Group 2 | 1 (1–) | 1 (1–) | 0.18 |
Group 3 | 2 (1–) | 1 (1–) | 0.32 |
Radiologist B | |||
Group 1 | 2 (1–) | 1.5 (1–) | 0.25 |
Group 2 | 1 (1–) | 1 (1–) | 0.56 |
Group 3 | 1 (1–) | 1 (1–) | 0.32 |
Radiologist and girth group | Continuous fluoroscopy | Grid-controlled variable-rate pulsed fluoroscopy |
P valuea
|
---|---|---|---|
Radiologist A | |||
Group 1 | 2 (1–) | 2 (1–) | >0.99 |
Group 2 | 2.5 (2–) | 2 (1–) | 0.56 |
Group 3 | 2 (1–) | 2 (2–) | 0.65 |
Radiologist B | |||
Group 1 | 2 (1–) | 2 (1–) | 0.26 |
Group 2 | 2 (1–) | 2 (1–) | 0.98 |
Group 3 | 1 (1–) | 1 (1–) | 0.32 |