Temperature changes during cortical bone drilling with a newly designed step drill and an internally cooled drill

verfasst von: Goran Augustin, Slavko Davila, Toma Udilljak, Tomislav Staroveski, Danko Brezak, Slaven Babic

Erschienen in: International Orthopaedics | Ausgabe 7/2012

Einloggen, um Zugang zu erhalten

Abstract

Purpose

Bone drilling causes an increase in bone temperature, and a temperature above 47°C is critical because it causes thermal bone necrosis. Thermal osteonecrosis is common with the drill diameter of ≥4.5 mm without cooling. The aim of this study was to determine the increase of bone temperature during drilling using newly contructed two-step and internally cooled drills.

Methods

An experiment was set up according to a central composite design. An internally cooled drill (3.4 mm and 4.5 mm) and a two-step drill (2.5/3.4 and 3.4/4.5 mm) were used in combination with feed rates of (0.02, 0.04, 0.10, 0.16 and 0.18 mm/rev) and cutting speeds (1.18, 10.68, 33.61, 56.55 and 66.05 m/min) with and without cooling with water of 24°C. Bone temperatures were measured with thermocouples. Drilling was performed on pig diaphyses with a three-axis mini milling machine.

Results

Bone temperatures in all combinations of parameters with internal cooling were below the critical 47°C (p = 0.05). The highest temperatures were detected using a 4.5-mm drill (40.5°C). A statistically significant effect other than cooling was found with the drill diameter and feed. A drill diameter of 3.4 mm with internal cooling developed a maximum temperature of 38.5°C and without cooling 46.3°C. For the same conditions a drill with diameter of 4.5 mm reached temperatures of 40.5°C and 55.7°C, respectively. The effect of feed rate is inversely proportional to the increase in bone temperature. With the feed rate 0.16 mm/rev, temperature was below critical even using the 4.5-mm drill (46.4°C, p = 0.05). Using the 3.4-mm drill all temperatures were below critical (46.2°C, p = 0.05). The two-step drill compared to a standard drill with the same diameter did not show statistical differences in maximum bone temperatures for all combinations of parameters (p = 0.05).

Conclusions

A two-step drill does not have any advantages over a standard twist drill of the same diameter. An internally cooled drill causes a significantly smaller increase of bone temperature during drilling with water of 24°C. An internally cooled drill is currently the 'ideal' drill for traumatology/orthopaedics because it produces the smallest increase in bone drilling temperature. If internal cooling is used the regulation of other drilling parameters is of no importance.
Literatur
1.
Zurück zum Zitat Eriksson AR, Albrektsson T (1983) Temperature threshold levels for heat-induced bone tissue injury: a vital-microscopic study in the rabbit. J Prosthet Dent 50:101–107PubMedCrossRef Eriksson AR, Albrektsson T (1983) Temperature threshold levels for heat-induced bone tissue injury: a vital-microscopic study in the rabbit. J Prosthet Dent 50:101–107PubMedCrossRef
2.
Zurück zum Zitat Ercoli C, Funkenbusch PD, Lee HJ, Moss ME, Graser GN (2004) The influence of drill wear on cutting efficiency and heat production during osteotomy preparation for dental implants: a study of drill durability. Int J Oral Maxillofac Implants 19:335–349PubMed Ercoli C, Funkenbusch PD, Lee HJ, Moss ME, Graser GN (2004) The influence of drill wear on cutting efficiency and heat production during osteotomy preparation for dental implants: a study of drill durability. Int J Oral Maxillofac Implants 19:335–349PubMed
3.
Zurück zum Zitat Matthews LS, Hirsch C (1972) Temperature measured in human cortical bone when drilling. J Bone Joint Surg Am 54:297–308PubMed Matthews LS, Hirsch C (1972) Temperature measured in human cortical bone when drilling. J Bone Joint Surg Am 54:297–308PubMed
4.
Zurück zum Zitat Chacon GE, Bower DI, Larsen PE, McGlumphy EA, Beck FM (2006) Heat production by 3 implant drill systems after repeated drilling and sterilization. J Oral Maxillofac Surg 64:265–269PubMedCrossRef Chacon GE, Bower DI, Larsen PE, McGlumphy EA, Beck FM (2006) Heat production by 3 implant drill systems after repeated drilling and sterilization. J Oral Maxillofac Surg 64:265–269PubMedCrossRef
5.
Zurück zum Zitat Eriksson RA, Albrektsson T, Albrektsson B (1984) Heat caused by drilling cortical bone. Temperature measured in vivo in patients and animals. Acta Orthop Scand 55:629–631PubMedCrossRef Eriksson RA, Albrektsson T, Albrektsson B (1984) Heat caused by drilling cortical bone. Temperature measured in vivo in patients and animals. Acta Orthop Scand 55:629–631PubMedCrossRef
6.
Zurück zum Zitat Yacker MJ, Klein M (1996) The effect of irrigation on osteotomy depth and bur diameter. Int J Oral Maxillofac Implants 11:634–638PubMed Yacker MJ, Klein M (1996) The effect of irrigation on osteotomy depth and bur diameter. Int J Oral Maxillofac Implants 11:634–638PubMed
7.
Zurück zum Zitat Eriksson RA, Adell R (1986) Temperatures during drilling for the placement of implants using the osseointegration technique. J Oral Maxillofac Surg 44:4–7PubMedCrossRef Eriksson RA, Adell R (1986) Temperatures during drilling for the placement of implants using the osseointegration technique. J Oral Maxillofac Surg 44:4–7PubMedCrossRef
8.
Zurück zum Zitat Kondo S, Okada Y, Iseki H, Hori T, Takakura K, Kobayashi A, Nagata H (2000) Thermological study of drilling bone tissue with a high-speed drill. Neurosurgery 46:1162–1168PubMedCrossRef Kondo S, Okada Y, Iseki H, Hori T, Takakura K, Kobayashi A, Nagata H (2000) Thermological study of drilling bone tissue with a high-speed drill. Neurosurgery 46:1162–1168PubMedCrossRef
9.
Zurück zum Zitat Augustin G, Davila S, Mihoci K, Udiljak T, Vedrina DS, Antabak A (2008) Thermal osteonecrosis and bone drilling parameters revisited. Arch Orthop Trauma Surg 128:71–77PubMedCrossRef Augustin G, Davila S, Mihoci K, Udiljak T, Vedrina DS, Antabak A (2008) Thermal osteonecrosis and bone drilling parameters revisited. Arch Orthop Trauma Surg 128:71–77PubMedCrossRef
10.
Zurück zum Zitat Sener BC, Dergin G, Gursoy B, Kelesoglu E, Slih I (2009) Effects of irrigation temperature on heat control in vitro at different drilling depths. Clin Oral Implan Res 20:294–298CrossRef Sener BC, Dergin G, Gursoy B, Kelesoglu E, Slih I (2009) Effects of irrigation temperature on heat control in vitro at different drilling depths. Clin Oral Implan Res 20:294–298CrossRef
11.
Zurück zum Zitat Reingewirtz Y, Moncler S, Senger B (1997) Influence of different parameters on bone heating and drilling time in implantology. Clin Oral Implan Res 8:189–197CrossRef Reingewirtz Y, Moncler S, Senger B (1997) Influence of different parameters on bone heating and drilling time in implantology. Clin Oral Implan Res 8:189–197CrossRef
12.
Zurück zum Zitat Jacob CH, Berry JT, Pope MH (1976) A study of the bone machining process—drilling. J Biomech 9:343–349CrossRef Jacob CH, Berry JT, Pope MH (1976) A study of the bone machining process—drilling. J Biomech 9:343–349CrossRef
13.
Zurück zum Zitat Aerssens J, Boonen S, Lowet G, Dequeker J (1998) Interspecies differences in bone composition, density and quality: potential implications for in vivo bone. Endocrinology 139:663–670PubMedCrossRef Aerssens J, Boonen S, Lowet G, Dequeker J (1998) Interspecies differences in bone composition, density and quality: potential implications for in vivo bone. Endocrinology 139:663–670PubMedCrossRef
14.
Zurück zum Zitat Sedlin ED, Hirsch C (1966) Factors affecting the determination of the physical properties of femoral cortical bone. Acta Orthop Scand 37:29–48PubMedCrossRef Sedlin ED, Hirsch C (1966) Factors affecting the determination of the physical properties of femoral cortical bone. Acta Orthop Scand 37:29–48PubMedCrossRef
15.
Zurück zum Zitat Natali C, Ingle P, Dowell J (1996) Orthopaedic bone drills—can they be improved? Temperature changes near the drilling face. J Bone Joint Surg Br 78:357–362PubMed Natali C, Ingle P, Dowell J (1996) Orthopaedic bone drills—can they be improved? Temperature changes near the drilling face. J Bone Joint Surg Br 78:357–362PubMed
16.
Zurück zum Zitat Bachus KN, Rondina MT, Hutchinson DT (2000) The effects of drilling force on cortical temperatures and their duration: an in vitro study. Med Eng Phys 22:685–691PubMedCrossRef Bachus KN, Rondina MT, Hutchinson DT (2000) The effects of drilling force on cortical temperatures and their duration: an in vitro study. Med Eng Phys 22:685–691PubMedCrossRef
17.
Zurück zum Zitat Augustin G, Zigman T, Davila S et al (2011) Cortical bone drilling and thermal osteonecrosis. Clin Biomech, (Bristol, Avon). 2011 Nov 7. [Epub ahead of print] Augustin G, Zigman T, Davila S et al (2011) Cortical bone drilling and thermal osteonecrosis. Clin Biomech, (Bristol, Avon). 2011 Nov 7. [Epub ahead of print]
18.
Zurück zum Zitat Kirschner H, Meyer W (1975) Entwichlung einer Innenkuhlung fur chirurgische 1147 Bohrer. Dtsch Zahnaztl Zeitschrift 30:436–438 Kirschner H, Meyer W (1975) Entwichlung einer Innenkuhlung fur chirurgische 1147 Bohrer. Dtsch Zahnaztl Zeitschrift 30:436–438
19.
Zurück zum Zitat Benington IC, Biagioni PA, Briggs J, Sheridan S, Lamey PJ (2002) Thermal changes observed at implant sites during internal and external irrigation. Clin Oral Implan Res 13:293–297CrossRef Benington IC, Biagioni PA, Briggs J, Sheridan S, Lamey PJ (2002) Thermal changes observed at implant sites during internal and external irrigation. Clin Oral Implan Res 13:293–297CrossRef
20.
Zurück zum Zitat Bubeck KA, García-López J, Maranda LS (2009) In vitro comparison of cortical bone temperature generation between traditional sequential drilling and a newly designed step drill in the equine third metacarpal bone. Vet Comp Orthop Traumatol 22:442–447PubMed Bubeck KA, García-López J, Maranda LS (2009) In vitro comparison of cortical bone temperature generation between traditional sequential drilling and a newly designed step drill in the equine third metacarpal bone. Vet Comp Orthop Traumatol 22:442–447PubMed
Metadaten
Titel
Temperature changes during cortical bone drilling with a newly designed step drill and an internally cooled drill
verfasst von
Goran Augustin
Slavko Davila
Toma Udilljak
Tomislav Staroveski
Danko Brezak
Slaven Babic
Publikationsdatum
01.07.2012
Verlag
Springer-Verlag
Erschienen in
International Orthopaedics / Ausgabe 7/2012
Print ISSN: 0341-2695
Elektronische ISSN: 1432-5195
DOI
https://doi.org/10.1007/s00264-012-1491-z

Weitere Artikel der Ausgabe 7/2012

International Orthopaedics 7/2012 Zur Ausgabe

Arthropedia

Grundlagenwissen der Arthroskopie und Gelenkchirurgie. Erweitert durch Fallbeispiele, Videos und Abbildungen. 
» Jetzt entdecken

Mehr Frauen im OP – weniger postoperative Komplikationen

21.05.2024 Allgemeine Chirurgie Nachrichten

Ein Frauenanteil von mindestens einem Drittel im ärztlichen Op.-Team war in einer großen retrospektiven Studie aus Kanada mit einer signifikanten Reduktion der postoperativen Morbidität assoziiert.

„Übersichtlicher Wegweiser“: Lauterbachs umstrittener Klinik-Atlas ist online

17.05.2024 Klinik aktuell Nachrichten

Sie sei „ethisch geboten“, meint Gesundheitsminister Karl Lauterbach: mehr Transparenz über die Qualität von Klinikbehandlungen. Um sie abzubilden, lässt er gegen den Widerstand vieler Länder einen virtuellen Klinik-Atlas freischalten.

Klinikreform soll zehntausende Menschenleben retten

15.05.2024 Klinik aktuell Nachrichten

Gesundheitsminister Lauterbach hat die vom Bundeskabinett beschlossene Klinikreform verteidigt. Kritik an den Plänen kommt vom Marburger Bund. Und in den Ländern wird über den Gang zum Vermittlungsausschuss spekuliert.

TEP mit Roboterhilfe führt nicht zu größerer Zufriedenheit

15.05.2024 Knie-TEP Nachrichten

Der Einsatz von Operationsrobotern für den Einbau von Totalendoprothesen des Kniegelenks hat die Präzision der Eingriffe erhöht. Für die postoperative Zufriedenheit der Patienten scheint das aber unerheblich zu sein, wie eine Studie zeigt.

Update Orthopädie und Unfallchirurgie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.