Validation of cardiac output estimation using the fourth-generation FloTrac/EV1000™ system in patients undergoing robotic-assisted off-pump coronary artery bypass surgery

verfasst von: Nien-Hsun Wu, Tsung-Han Hsieh, Chun-Yu Chang, Ping-Chen Shih, Ming‐Chang Kao, Han-Yu Lin

Erschienen in: Heart and Vessels | Ausgabe 3/2023

Einloggen, um Zugang zu erhalten

Abstract

The pulmonary artery catheter (PAC)—despite its invasiveness—remains the gold standard for cardiac output (CO) monitoring. The FloTrac system, a less invasive hemodynamic monitor has been developed, which estimates CO using arterial pressure waveform analysis without external calibration. Recently, an upgraded version of FloTrac system with improved algorithm to follow changes in vascular resistance was introduced into the market. The aim of this study was to assess the reliability of the CO estimated from the fourth-generation FloTrac/EV1000 system (COFT) compared to that measured with PAC using the thermodilution method (COPAC) during robotic-assisted off-pump coronary artery bypass (OPCAB) surgery. COFT and COPAC were obtained simultaneously at 4 predefined time points during robotic-assisted OPCAB: 5 min after the induction of general anesthesia (T1), after starting one-lung ventilation (T2), after capnothorax (T3), and after mini-thoracotomy was performed (T4). The agreement of data was investigated by Bland–Altman analysis. Thirty-four patients were initially enrolled. After exclusion, 32 patients and a total of 128 paired CO measurements were obtained. The overall bias was 1.46 L/min, the 95% limits of agreements were − 3.40 to 6.33 L/min, and the percentage error was 72.98%. Regression analysis of the systemic vascular resistance index (SVRI) and the bias between COPAC and COFT showed that the bias was moderately correlated with the SVRI (r2 = 0.43; p < 0.0001). Despite a software upgrade, the reliability of the fourth-generation FloTrac/EV1000™ system during robotic-assisted OPCAB to estimate CO was not acceptable, especially in patients with low SVRI.
Literatur
1.
Zurück zum Zitat Whellan DJ, McCarey MM, Taylor BS, Rosengart TK, Wallace AS, Shroyer AL, Gammie JS, Peterson ED (2016) Trends in robotic-assisted coronary artery bypass grafts: a study of the society of thoracic surgeons adult cardiac surgery database, 2006 to 2012. Ann Thorac Surg 102:140–146CrossRef Whellan DJ, McCarey MM, Taylor BS, Rosengart TK, Wallace AS, Shroyer AL, Gammie JS, Peterson ED (2016) Trends in robotic-assisted coronary artery bypass grafts: a study of the society of thoracic surgeons adult cardiac surgery database, 2006 to 2012. Ann Thorac Surg 102:140–146CrossRef
2.
Zurück zum Zitat Yanagawa F, Perez M, Bell T, Grim R, Martin J, Ahuja V (2015) Critical outcomes in nonrobotic vs robotic-assisted cardiac surgery. JAMA Surg 150:771–777CrossRef Yanagawa F, Perez M, Bell T, Grim R, Martin J, Ahuja V (2015) Critical outcomes in nonrobotic vs robotic-assisted cardiac surgery. JAMA Surg 150:771–777CrossRef
3.
Zurück zum Zitat Do QB, Goyer C, Chavanon O, Couture P, Denautt A, Cartier R (2002) Hemodynamic changes during off-pump CABG surgery. Eur J Cardiothorac Surg 21:385–390CrossRef Do QB, Goyer C, Chavanon O, Couture P, Denautt A, Cartier R (2002) Hemodynamic changes during off-pump CABG surgery. Eur J Cardiothorac Surg 21:385–390CrossRef
4.
Zurück zum Zitat Funk DJ, Moretti EW, Gan TJ (2009) Minimally invasive cardiac output monitoring in the perioperative setting. Anesth Analg 108:887–897CrossRef Funk DJ, Moretti EW, Gan TJ (2009) Minimally invasive cardiac output monitoring in the perioperative setting. Anesth Analg 108:887–897CrossRef
5.
Zurück zum Zitat Ganz W, Donoso R, Marcus HS, Forrester JS, Swan HJ (1971) A new technique for measurement of cardiac output by thermodilution in man. Am J Cardiol 27:392–396CrossRef Ganz W, Donoso R, Marcus HS, Forrester JS, Swan HJ (1971) A new technique for measurement of cardiac output by thermodilution in man. Am J Cardiol 27:392–396CrossRef
6.
Zurück zum Zitat American Society of Anesthesiologists Task Force on Pulmonary Artery Catheterization (2003) Practice guidelines for pulmonary artery catheterization: an updated report by the American society of anesthesiologists task force on pulmonary artery catheterization. Anesthesiology 99:988–1014CrossRef American Society of Anesthesiologists Task Force on Pulmonary Artery Catheterization (2003) Practice guidelines for pulmonary artery catheterization: an updated report by the American society of anesthesiologists task force on pulmonary artery catheterization. Anesthesiology 99:988–1014CrossRef
7.
Zurück zum Zitat Seifi A, Elliott RJ, Elsehety MA (2016) Usage of Swan-Ganz catheterization during the past 2 decades in United States. J Crit Care 35:213–214CrossRef Seifi A, Elliott RJ, Elsehety MA (2016) Usage of Swan-Ganz catheterization during the past 2 decades in United States. J Crit Care 35:213–214CrossRef
8.
Zurück zum Zitat Suehiro K, Tanaka K, Funao T, Matsuura T, Mori T, Nishikawa K (2013) Systemic vascular resistance has an impact on the reliability of the Vigileo-FloTrac system in measuring cardiac output and tracking cardiac output changes. Br J Anaesth 111:170–177CrossRef Suehiro K, Tanaka K, Funao T, Matsuura T, Mori T, Nishikawa K (2013) Systemic vascular resistance has an impact on the reliability of the Vigileo-FloTrac system in measuring cardiac output and tracking cardiac output changes. Br J Anaesth 111:170–177CrossRef
9.
Zurück zum Zitat Maeda T, Yoshitani K, Inatomi Y, Ohnishi Y (2014) Inaccuracy of the FloTrac/Vigileo system in patients with low cardiac index. J Cardiothorac Vasc Anesth 28:1521–1526CrossRef Maeda T, Yoshitani K, Inatomi Y, Ohnishi Y (2014) Inaccuracy of the FloTrac/Vigileo system in patients with low cardiac index. J Cardiothorac Vasc Anesth 28:1521–1526CrossRef
10.
Zurück zum Zitat Suehiro K, Tanaka K, Mikawa M, Uchihara Y, Matsuyama T, Matsuura T, Funao T, Yamada T, Mori T, Nishikawa K (2015) Improved performance of the fourth-generation flotrac/vigileo system for tracking cardiac output changes. J Cardiothorac Vasc Anesth 29:656–662CrossRef Suehiro K, Tanaka K, Mikawa M, Uchihara Y, Matsuyama T, Matsuura T, Funao T, Yamada T, Mori T, Nishikawa K (2015) Improved performance of the fourth-generation flotrac/vigileo system for tracking cardiac output changes. J Cardiothorac Vasc Anesth 29:656–662CrossRef
11.
Zurück zum Zitat Bland JM, Altman DG (1986) Statistical methods for assessing agreement between two methods of clinical measurement. Lancet 1:307–310CrossRef Bland JM, Altman DG (1986) Statistical methods for assessing agreement between two methods of clinical measurement. Lancet 1:307–310CrossRef
12.
Zurück zum Zitat Critchley LA, Critchley JA (1999) A meta-analysis of studies using bias and precision statistics to compare cardiac output measurement techniques. J Clin Monit Comput 15:85–91CrossRef Critchley LA, Critchley JA (1999) A meta-analysis of studies using bias and precision statistics to compare cardiac output measurement techniques. J Clin Monit Comput 15:85–91CrossRef
13.
Zurück zum Zitat Kusaka Y, Ohchi F, Minami T (2019) Evaluation of the fourth-generation FloTrac/Vigileo system in comparison with the intermittent bolus thermodilution method in patients undergoing cardiac surgery. J Cardiothorac Vasc Anesth 33:953–960CrossRef Kusaka Y, Ohchi F, Minami T (2019) Evaluation of the fourth-generation FloTrac/Vigileo system in comparison with the intermittent bolus thermodilution method in patients undergoing cardiac surgery. J Cardiothorac Vasc Anesth 33:953–960CrossRef
14.
Zurück zum Zitat Eisenried A, Klarwein R, Ihmsen H, Wehrfritz A, Tandler R, Heim C, Fechner J (2019) Accuracy and trending ability of the fourth-generation FloTrac/EV1000 system in patients with severe aortic valve stenosis before and after surgical valve replacement. J Cardiothorac Vasc Anesth 33:1230–1236CrossRef Eisenried A, Klarwein R, Ihmsen H, Wehrfritz A, Tandler R, Heim C, Fechner J (2019) Accuracy and trending ability of the fourth-generation FloTrac/EV1000 system in patients with severe aortic valve stenosis before and after surgical valve replacement. J Cardiothorac Vasc Anesth 33:1230–1236CrossRef
15.
Zurück zum Zitat Ylikauma LA, Ohtonen PP, Erkinaro TM, Vakkala MA, Liisanantti JH, Satta JU, Juvonen TS, Kaakinen TI (2022) Bioreactance and fourth-generation pulse contour methods in monitoring cardiac index during off-pump coronary artery bypass surgery. J Clin Monit Comput 36(3):879–888CrossRef Ylikauma LA, Ohtonen PP, Erkinaro TM, Vakkala MA, Liisanantti JH, Satta JU, Juvonen TS, Kaakinen TI (2022) Bioreactance and fourth-generation pulse contour methods in monitoring cardiac index during off-pump coronary artery bypass surgery. J Clin Monit Comput 36(3):879–888CrossRef
16.
Zurück zum Zitat Cho YJ, Koo CH, Kim TK, Hong DM, Jeon Y (2016) Comparison of cardiac output measures by transpulmonary thermodilution, pulse contour analysis, and pulmonary artery thermodilution during off-pump coronary artery bypass surgery: a subgroup analysis of the cardiovascular anaesthesia registry at a single tertiary centre. J Clin Monit Comput 30:771–782CrossRef Cho YJ, Koo CH, Kim TK, Hong DM, Jeon Y (2016) Comparison of cardiac output measures by transpulmonary thermodilution, pulse contour analysis, and pulmonary artery thermodilution during off-pump coronary artery bypass surgery: a subgroup analysis of the cardiovascular anaesthesia registry at a single tertiary centre. J Clin Monit Comput 30:771–782CrossRef
17.
Zurück zum Zitat Maeda T, Hamaguchi E, Kubo N, Shimokawa A, Kanazawa H, Ohnishi Y (2019) The accuracy and trending ability of cardiac index measured by the fourth-generation FloTrac/Vigileo system™ and the fick method in cardiac surgery patients. J Clin Monit Comput 33:767–776CrossRef Maeda T, Hamaguchi E, Kubo N, Shimokawa A, Kanazawa H, Ohnishi Y (2019) The accuracy and trending ability of cardiac index measured by the fourth-generation FloTrac/Vigileo system™ and the fick method in cardiac surgery patients. J Clin Monit Comput 33:767–776CrossRef
18.
Zurück zum Zitat Mayer J, Boldt J, Poland R, Peterson A, Manecke GR Jr (2009) Continuous arterial pressure waveform-based cardiac output using the FloTrac/Vigileo: a review and meta-analysis. J Cardiothorac Vasc Anesth 23:401–406CrossRef Mayer J, Boldt J, Poland R, Peterson A, Manecke GR Jr (2009) Continuous arterial pressure waveform-based cardiac output using the FloTrac/Vigileo: a review and meta-analysis. J Cardiothorac Vasc Anesth 23:401–406CrossRef
19.
Zurück zum Zitat Lorsomradee S, Lorsomradee S, Cromheecke S, De Hert SG (2007) Uncalibrated arterial pulse contour analysis versus continuous thermodilution technique: effects of alterations in arterial waveform. J Cardiothorac Vasc Anesth 21:636–643CrossRef Lorsomradee S, Lorsomradee S, Cromheecke S, De Hert SG (2007) Uncalibrated arterial pulse contour analysis versus continuous thermodilution technique: effects of alterations in arterial waveform. J Cardiothorac Vasc Anesth 21:636–643CrossRef
20.
Zurück zum Zitat Scolletta S, Franchi F, Taccone FS, Donadello K, Biagioli B, Vincent JL (2011) An uncalibrated pulse contour method to measure cardiac output during aortic counterpulsation. Anesth Analg 113:1389–1395CrossRef Scolletta S, Franchi F, Taccone FS, Donadello K, Biagioli B, Vincent JL (2011) An uncalibrated pulse contour method to measure cardiac output during aortic counterpulsation. Anesth Analg 113:1389–1395CrossRef
21.
Zurück zum Zitat Matsuura K, Imamaki M, Ishida A, Shimura H, Fujita H, Niitsuma Y, Miyazaki M (2008) Low systemic vascular resistance state following off-pump coronary artery bypass grafting. Ann Thorac Cardiovasc Surg 14:15–21 Matsuura K, Imamaki M, Ishida A, Shimura H, Fujita H, Niitsuma Y, Miyazaki M (2008) Low systemic vascular resistance state following off-pump coronary artery bypass grafting. Ann Thorac Cardiovasc Surg 14:15–21
22.
Zurück zum Zitat Suehiro K, Tanaka K, Yamada T, Matsuura T, Mori T, Funao T, Nishikawa K (2015) The ability of the Vigileo-FloTrac system to measure cardiac output and track cardiac output changes during one-lung ventilation. J Clin Monit Comput 29:333–339CrossRef Suehiro K, Tanaka K, Yamada T, Matsuura T, Mori T, Funao T, Nishikawa K (2015) The ability of the Vigileo-FloTrac system to measure cardiac output and track cardiac output changes during one-lung ventilation. J Clin Monit Comput 29:333–339CrossRef
23.
Zurück zum Zitat Tsai YF, Su BC, Lin CC, Liu FC, Lee WC, Yu HP (2012) Cardiac output derived from arterial pressure waveform analysis: validation of the third-generation software in patients undergoing orthotopic liver transplantation. Transplant Proc 44:433–437CrossRef Tsai YF, Su BC, Lin CC, Liu FC, Lee WC, Yu HP (2012) Cardiac output derived from arterial pressure waveform analysis: validation of the third-generation software in patients undergoing orthotopic liver transplantation. Transplant Proc 44:433–437CrossRef
Metadaten
Titel
Validation of cardiac output estimation using the fourth-generation FloTrac/EV1000™ system in patients undergoing robotic-assisted off-pump coronary artery bypass surgery
verfasst von
Nien-Hsun Wu
Tsung-Han Hsieh
Chun-Yu Chang
Ping-Chen Shih
Ming‐Chang Kao
Han-Yu Lin
Publikationsdatum
01.10.2022
Verlag
Springer Japan
Erschienen in
Heart and Vessels / Ausgabe 3/2023
Print ISSN: 0910-8327
Elektronische ISSN: 1615-2573
DOI
https://doi.org/10.1007/s00380-022-02177-x

Weitere Artikel der Ausgabe 3/2023

Heart and Vessels 3/2023 Zur Ausgabe

„Übersichtlicher Wegweiser“: Lauterbachs umstrittener Klinik-Atlas ist online

17.05.2024 Klinik aktuell Nachrichten

Sie sei „ethisch geboten“, meint Gesundheitsminister Karl Lauterbach: mehr Transparenz über die Qualität von Klinikbehandlungen. Um sie abzubilden, lässt er gegen den Widerstand vieler Länder einen virtuellen Klinik-Atlas freischalten.

„Jeder Fall von plötzlichem Tod muss obduziert werden!“

17.05.2024 Plötzlicher Herztod Nachrichten

Ein signifikanter Anteil der Fälle von plötzlichem Herztod ist genetisch bedingt. Um ihre Verwandten vor diesem Schicksal zu bewahren, sollten jüngere Personen, die plötzlich unerwartet versterben, ausnahmslos einer Autopsie unterzogen werden.

Hirnblutung unter DOAK und VKA ähnlich bedrohlich

17.05.2024 Direkte orale Antikoagulanzien Nachrichten

Kommt es zu einer nichttraumatischen Hirnblutung, spielt es keine große Rolle, ob die Betroffenen zuvor direkt wirksame orale Antikoagulanzien oder Marcumar bekommen haben: Die Prognose ist ähnlich schlecht.

Schlechtere Vorhofflimmern-Prognose bei kleinem linken Ventrikel

17.05.2024 Vorhofflimmern Nachrichten

Nicht nur ein vergrößerter, sondern auch ein kleiner linker Ventrikel ist bei Vorhofflimmern mit einer erhöhten Komplikationsrate assoziiert. Der Zusammenhang besteht nach Daten aus China unabhängig von anderen Risikofaktoren.

Update Kardiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.