Refractive predictability using two optical biometers and refraction types for intraocular lens power calculation in cataract surgery

verfasst von: Huanhuan Cheng, Jianbing Li, Bing Cheng, Mingxing Wu

Erschienen in: International Ophthalmology | Ausgabe 7/2020

Einloggen, um Zugang zu erhalten

Abstract

Purpose

To evaluate the accuracy of intraocular lens (IOL) power calculation in relation to optical biometry devices and refraction types.

Methods

Patients undergoing cataract phacoemulsification and insertion of the MX60 IOL were enrolled. Optical biometric measurements were performed with both IOLMaster 700 and Lenstar 900. Biometry measurements were compared between devices. A subsample of 133 eyes (81.1%) had examination for both autorefraction and subjective refraction postoperatively. The differences between the postoperative refraction and the refraction predicted by eight formulas (Kane, Hill-RBF 2.0, Barrett Universal II, Olsen, Haigis, SRK/T, Holladay 1 and Hoffer Q) were calculated.

Results

Overall, this study comprised 164 eyes of 164 patients. High agreement between the two biometers for axial length, average keratometry readings, anterior chamber depth, lens thickness and central corneal thickness was found (interclass correlation confidents: 0.999, 0.988, 0.965, 0.865 and 0.972, respectively, all P < 0.001). The absolute prediction error calculated with IOLMaster 700 measurements was significantly lower than that calculated with Lenstar 900 measurements for Olsen (P = 0.003), Haigis (P < 0.001) and Hoffer Q (P = 0.028). OPD-Scan III gave slightly more negative readings than subjective refraction (mean difference − 0.107 ± 0.553, P = 0.003 for spherical equivalent). However, no significant difference in absolute prediction error was found between the two refraction types per each formula.

Conclusion

IOLMaster 700 and Lenstar 900 showed good agreement in biometric measurements with a trend toward better refractive outcome using IOLMaster 700. The accuracy of IOL calculation assessed with OPD autorefraction was equivalent to that assessed with subjective refraction.
Literatur
1.
Zurück zum Zitat Kanclerz P, Hoffer KJ, Rozema JJ, Przewłócka K, Savini G (2016) Biometry with a new swept-source optical coherence tomography biometer: repeatability and agreement with an optical low-coherence reflectometry device. J Cataract Refract Surg 42:577–581CrossRef Kanclerz P, Hoffer KJ, Rozema JJ, Przewłócka K, Savini G (2016) Biometry with a new swept-source optical coherence tomography biometer: repeatability and agreement with an optical low-coherence reflectometry device. J Cataract Refract Surg 42:577–581CrossRef
2.
Zurück zum Zitat Cruysberg LP, Doors M, Verbakel F, Berendschot TT, De Brabander J, Nuijts RM (2009) Evaluation of the Lenstar LS 900 non-contact biometer. Br J Ophthalmol 94:106–110CrossRef Cruysberg LP, Doors M, Verbakel F, Berendschot TT, De Brabander J, Nuijts RM (2009) Evaluation of the Lenstar LS 900 non-contact biometer. Br J Ophthalmol 94:106–110CrossRef
3.
Zurück zum Zitat Arriola-Villalobos P, Almendral-Gómez J, Garzón N, Ruiz-Medrano J, Fernández-Pérez C, Martínez-de-la-Casa JM, Díaz-Valle D (2017) Agreement and clinical comparison between a new swept-source optical coherence tomography-based optical biometer and an optical low-coherence reflectometry biometer. Eye 31:437–442CrossRef Arriola-Villalobos P, Almendral-Gómez J, Garzón N, Ruiz-Medrano J, Fernández-Pérez C, Martínez-de-la-Casa JM, Díaz-Valle D (2017) Agreement and clinical comparison between a new swept-source optical coherence tomography-based optical biometer and an optical low-coherence reflectometry biometer. Eye 31:437–442CrossRef
4.
Zurück zum Zitat Özyol P, Özyol E (2016) Agreement between swept-source optical biometry and Scheimpflug-based topography measurements of anterior segment parameters. Am J Ophthalmol 169:73–78CrossRef Özyol P, Özyol E (2016) Agreement between swept-source optical biometry and Scheimpflug-based topography measurements of anterior segment parameters. Am J Ophthalmol 169:73–78CrossRef
5.
Zurück zum Zitat Asena L, Akman A, Gungor SG, Dursun Altınörs D (2018) Comparison of keratometry obtained by a swept source OCT-based biometer with a standard optical biometer and Scheimpflug imaging. Curr Eye Res 43:882–888CrossRef Asena L, Akman A, Gungor SG, Dursun Altınörs D (2018) Comparison of keratometry obtained by a swept source OCT-based biometer with a standard optical biometer and Scheimpflug imaging. Curr Eye Res 43:882–888CrossRef
6.
Zurück zum Zitat El Chehab H, Agard E, Dot C (2018) Comparison of two biometers: a swept-source optical coherence tomography and an optical low-coherence reflectometry biometer. Eur J Ophthalmol 29:547–554CrossRef El Chehab H, Agard E, Dot C (2018) Comparison of two biometers: a swept-source optical coherence tomography and an optical low-coherence reflectometry biometer. Eur J Ophthalmol 29:547–554CrossRef
7.
Zurück zum Zitat Norrby S (2008) Sources of error in intraocular lens power calculation. J Cataract Refract Surg 34:368–376CrossRef Norrby S (2008) Sources of error in intraocular lens power calculation. J Cataract Refract Surg 34:368–376CrossRef
8.
Zurück zum Zitat Ostri C, Holfort SK, Fich MS, Riise P (2018) Automated refraction is stable 1 week after uncomplicated cataract surgery. Acta Ophthalmol 96:149–153CrossRef Ostri C, Holfort SK, Fich MS, Riise P (2018) Automated refraction is stable 1 week after uncomplicated cataract surgery. Acta Ophthalmol 96:149–153CrossRef
9.
Zurück zum Zitat Hoffer KJ, Aramberri J, Haigis W, Olsen T, Savini G, Shammas HJ, Bentow S (2015) Protocols for studies of intraocular lens formula accuracy. Am J Ophthalmol 160:403–405CrossRef Hoffer KJ, Aramberri J, Haigis W, Olsen T, Savini G, Shammas HJ, Bentow S (2015) Protocols for studies of intraocular lens formula accuracy. Am J Ophthalmol 160:403–405CrossRef
10.
Zurück zum Zitat Kane JX, Van Heerden A, Atik A, Petsoglou C (2016) Intraocular lens power formula accuracy: comparison of 7 formulas. J Cataract Refract Surg 42:1490–1500CrossRef Kane JX, Van Heerden A, Atik A, Petsoglou C (2016) Intraocular lens power formula accuracy: comparison of 7 formulas. J Cataract Refract Surg 42:1490–1500CrossRef
11.
Zurück zum Zitat Kane JX, Van Heerden A, Atik A, Petsoglou C (2017) Accuracy of 3 new methods for intraocular lens power selection. J Cataract Refract Surg 43:333–339CrossRef Kane JX, Van Heerden A, Atik A, Petsoglou C (2017) Accuracy of 3 new methods for intraocular lens power selection. J Cataract Refract Surg 43:333–339CrossRef
12.
Zurück zum Zitat Wallace HB, Misra SL, Li SS, McKelvie J (2018) Predicting pseudophakic refractive error: interplay of biometry prediction error, anterior chamber depth, and changes in corneal curvature. J Cataract Refract Surg 44:1123–1129CrossRef Wallace HB, Misra SL, Li SS, McKelvie J (2018) Predicting pseudophakic refractive error: interplay of biometry prediction error, anterior chamber depth, and changes in corneal curvature. J Cataract Refract Surg 44:1123–1129CrossRef
13.
Zurück zum Zitat McGinnigle S, Naroo SA, Eperjesi F (2014) Evaluation of the auto-refraction function of the Nidek OPD-Scan III. Clin Exp Optom 97:160–163CrossRef McGinnigle S, Naroo SA, Eperjesi F (2014) Evaluation of the auto-refraction function of the Nidek OPD-Scan III. Clin Exp Optom 97:160–163CrossRef
14.
Zurück zum Zitat Reitblat O, Levy A, Kleinmann G, Assia EI (2018) Accuracy of intraocular lens power calculation using three optical biometry measurement devices: the OA-2000, Lenstar-LS900 and IOLMaster-500. Eye 32:1244–1252CrossRef Reitblat O, Levy A, Kleinmann G, Assia EI (2018) Accuracy of intraocular lens power calculation using three optical biometry measurement devices: the OA-2000, Lenstar-LS900 and IOLMaster-500. Eye 32:1244–1252CrossRef
15.
Zurück zum Zitat Srivannaboon S, Chirapapaisan C (2019) Comparison of refractive outcomes using conventional keratometry or total keratometry for IOL power calculation in cataract surgery. Graefes Arch Clin Exp Ophthalmol 257:2677–2682CrossRef Srivannaboon S, Chirapapaisan C (2019) Comparison of refractive outcomes using conventional keratometry or total keratometry for IOL power calculation in cataract surgery. Graefes Arch Clin Exp Ophthalmol 257:2677–2682CrossRef
16.
Zurück zum Zitat Olsen T (2007) Calculation of intraocular lens power: a review. Acta Ophthalmol Scand 85:472–485CrossRef Olsen T (2007) Calculation of intraocular lens power: a review. Acta Ophthalmol Scand 85:472–485CrossRef
17.
Zurück zum Zitat Cooke DL, Cooke TL (2016) Comparison of 9 intraocular lens power calculation formulas. J Cataract Refract Surg 42:1157–1164CrossRef Cooke DL, Cooke TL (2016) Comparison of 9 intraocular lens power calculation formulas. J Cataract Refract Surg 42:1157–1164CrossRef
18.
Zurück zum Zitat Savini G, Hoffer KJ, Shammas HJ, Aramberri J, Huang J, Barboni P (2017) Accuracy of a new swept-source optical coherence tomography biometer for IOL power calculation and comparison to IOLMaster. J Refract Surg 33:690–695CrossRef Savini G, Hoffer KJ, Shammas HJ, Aramberri J, Huang J, Barboni P (2017) Accuracy of a new swept-source optical coherence tomography biometer for IOL power calculation and comparison to IOLMaster. J Refract Surg 33:690–695CrossRef
19.
Zurück zum Zitat Melles RB, Holladay JT, Chang WJ (2018) Accuracy of intraocular lens calculation formulas. Ophthalmology 125:169–178CrossRef Melles RB, Holladay JT, Chang WJ (2018) Accuracy of intraocular lens calculation formulas. Ophthalmology 125:169–178CrossRef
20.
Zurück zum Zitat Connell BJ, Kane JX (2019) Comparison of the Kane formula with existing formulas for intraocular lens power selection. BMJ Open Ophthalmol 4:e251CrossRef Connell BJ, Kane JX (2019) Comparison of the Kane formula with existing formulas for intraocular lens power selection. BMJ Open Ophthalmol 4:e251CrossRef
21.
Zurück zum Zitat Melles RB, Kane JX, Olsen T, Chang WJ (2019) Update on intraocular lens calculation formulas. Ophthalmology 126:1334–1335CrossRef Melles RB, Kane JX, Olsen T, Chang WJ (2019) Update on intraocular lens calculation formulas. Ophthalmology 126:1334–1335CrossRef
Metadaten
Titel
Refractive predictability using two optical biometers and refraction types for intraocular lens power calculation in cataract surgery
verfasst von
Huanhuan Cheng
Jianbing Li
Bing Cheng
Mingxing Wu
Publikationsdatum
15.04.2020
Verlag
Springer Netherlands
Erschienen in
International Ophthalmology / Ausgabe 7/2020
Print ISSN: 0165-5701
Elektronische ISSN: 1573-2630
DOI
https://doi.org/10.1007/s10792-020-01355-y

Weitere Artikel der Ausgabe 7/2020

International Ophthalmology 7/2020 Zur Ausgabe

Neu im Fachgebiet Augenheilkunde

Metastase in der periokulären Region

Metastasen Leitthema

Orbitale und periokuläre metastatische Tumoren galten früher als sehr selten. Aber mit der ständigen Aktualisierung von Medikamenten und Nachweismethoden für die Krebsbehandlung werden neue Chemotherapien und Strahlenbehandlungen eingesetzt. Die …

Staging und Systemtherapie bei okulären und periokulären Metastasen

Metastasen Leitthema

Metastasen bösartiger Erkrankungen sind die häufigsten Tumoren, die im Auge diagnostiziert werden. Sie treten bei ungefähr 5–10 % der Patienten mit soliden Tumoren im Verlauf der Erkrankung auf. Besonders häufig sind diese beim Mammakarzinom und …

CME: Wundheilung nach Trabekulektomie

Trabekulektomie CME-Artikel

Wird ein Glaukom chirurgisch behandelt, ist die anschließende Wundheilung von entscheidender Bedeutung. In diesem CME-Kurs lernen Sie, welche Pathomechanismen der Vernarbung zugrunde liegen, wie perioperativ therapiert und Operationsversagen frühzeitig erkannt werden kann.

„standard operating procedures“ (SOP) – Vorschlag zum therapeutischen Management bei periokulären sowie intraokulären Metastasen

Metastasen Leitthema

Peri- sowie intraokuläre Metastasen sind insgesamt gesehen selten und meist Zeichen einer fortgeschrittenen primären Tumorerkrankung. Die Therapie ist daher zumeist palliativ und selten kurativ. Zudem ist die Therapiefindung sehr individuell. Die …

Update Augenheilkunde

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.