Validity of Goniometric Elbow Measurements: Comparative Study with a Radiographic Method

verfasst von: Julien Chapleau, MD, Fanny Canet, MScA, Yvan Petit, PhD, G-Yves Laflamme, MD, FRCSC, Dominique M. Rouleau, MD, MSc, FRCSC

Erschienen in: Clinical Orthopaedics and Related Research® | Ausgabe 11/2011

Einloggen, um Zugang zu erhalten

Abstract

Background

A universal goniometer is commonly used to measure the elbow’s ROM and carrying angle; however, some authors question its poor intertester reliability.

Questions/purposes

We (1) assessed the validity of goniometric measurements as compared with radiographic measurements in the evaluation of ROM of the elbow and (2) determined the reliability of both.

Methods

The ROM and carrying angle of 51 healthy subjects (102 elbows) were measured using two methods: with a universal goniometer by one observer three times and on radiographs by two independent examiners. Paired t-test and Pearson’s correlation were used to compare and detect the relationship between mean ROM. The maximal error was calculated according to the Bland and Altman method.

Results

The intraclass correlation coefficients (ICC) ranged from 0.945 to 0.973 for the goniometric measurements and from 0.980 to 0.991 for the radiographic measurements. The two methods correlated when measuring the total ROM in flexion and extension. The maximal errors of the goniometric measurement were 10.3° for extension, 7.0° for flexion, and 6.5° for carrying angle 95% of the time. We observed differences for maximum flexion, maximal extension, and carrying angle between the methods.

Conclusion

Both measurement methods differ but they correlate. When measured with a goniometer, the elbow ROM shows a maximal error of approximately 10°.

Clinical Relevance

The goniometer is a reasonable and simple clinical tool, but for research protocols, we suggest using the radiographic method because of the higher level of precision required.
Literatur
1.
Zurück zum Zitat Armstrong AD, MacDermid JC, Chinchalkar S, Stevens RS, King GJ. Reliability of range-of-motion measurement in the elbow and forearm. J Shoulder Elbow Surg. 1998;7:573–580.PubMedCrossRef Armstrong AD, MacDermid JC, Chinchalkar S, Stevens RS, King GJ. Reliability of range-of-motion measurement in the elbow and forearm. J Shoulder Elbow Surg. 1998;7:573–580.PubMedCrossRef
2.
Zurück zum Zitat Awan R, Smith J, Boon AJ. Measuring shoulder internal rotation range of motion: a comparison of 3 techniques. Arch Phys Med Rehabil. 2002;83:1229–1234.PubMedCrossRef Awan R, Smith J, Boon AJ. Measuring shoulder internal rotation range of motion: a comparison of 3 techniques. Arch Phys Med Rehabil. 2002;83:1229–1234.PubMedCrossRef
3.
Zurück zum Zitat Bland JM, Altman DG. A note on the use of the intraclass correlation coefficient in the evaluation of agreement between two methods of measurement. Comput Biol Med. 1990;20:337–340.PubMedCrossRef Bland JM, Altman DG. A note on the use of the intraclass correlation coefficient in the evaluation of agreement between two methods of measurement. Comput Biol Med. 1990;20:337–340.PubMedCrossRef
4.
Zurück zum Zitat Bland JM, Altman DG. Applying the right statistics: analyses of measurement studies. Ultrasound Obstet Gynecol. 2003;22:85–93.PubMedCrossRef Bland JM, Altman DG. Applying the right statistics: analyses of measurement studies. Ultrasound Obstet Gynecol. 2003;22:85–93.PubMedCrossRef
5.
Zurück zum Zitat Boone DC, Azen SP. Normal range of motion of joints in male subjects. J Bone Joint Surg Am. 1979;61:756–759.PubMed Boone DC, Azen SP. Normal range of motion of joints in male subjects. J Bone Joint Surg Am. 1979;61:756–759.PubMed
6.
Zurück zum Zitat Docherty MA, Schwab RA, Ma OJ. Can elbow extension be used as a test of clinically significant injury? South Med J. 2002;95:539–541.PubMed Docherty MA, Schwab RA, Ma OJ. Can elbow extension be used as a test of clinically significant injury? South Med J. 2002;95:539–541.PubMed
7.
Zurück zum Zitat Fish DR, Wingate L. Sources of goniometric error at the elbow. Phys Ther. 1985;65:1666–1670.PubMed Fish DR, Wingate L. Sources of goniometric error at the elbow. Phys Ther. 1985;65:1666–1670.PubMed
8.
Zurück zum Zitat Glasgow C, Wilton J, Tooth L. Optimal daily total end range time for contracture: resolution in hand splinting. J Hand Ther. 2003;16:207–218.PubMedCrossRef Glasgow C, Wilton J, Tooth L. Optimal daily total end range time for contracture: resolution in hand splinting. J Hand Ther. 2003;16:207–218.PubMedCrossRef
9.
Zurück zum Zitat Golden DW, Wojcicki JM, Jhee JT, Gilpin SL, Sawyer JR, Heyman MB. Body mass index and elbow range of motion in a healthy pediatric population: a possible mechanism of overweight in children. J Pediatr Gastroenterol Nutr. 2008;46:196–201.PubMedCrossRef Golden DW, Wojcicki JM, Jhee JT, Gilpin SL, Sawyer JR, Heyman MB. Body mass index and elbow range of motion in a healthy pediatric population: a possible mechanism of overweight in children. J Pediatr Gastroenterol Nutr. 2008;46:196–201.PubMedCrossRef
10.
Zurück zum Zitat Goodwin J, Clark C, Deakes J, Burdon D, Lawrence C. Clinical methods of goniometry: a comparative study. Disabil Rehabil. 1992;14:10–15.PubMedCrossRef Goodwin J, Clark C, Deakes J, Burdon D, Lawrence C. Clinical methods of goniometry: a comparative study. Disabil Rehabil. 1992;14:10–15.PubMedCrossRef
11.
Zurück zum Zitat Greene BL, Wolf SL. Upper extremity joint movement: comparison of two measurement devices. Arch Phys Med Rehabil. 1989;70:288–290.PubMed Greene BL, Wolf SL. Upper extremity joint movement: comparison of two measurement devices. Arch Phys Med Rehabil. 1989;70:288–290.PubMed
12.
Zurück zum Zitat Gunal I, Kose N, Erdogan O, Gokturk E, Seber S. Normal range of motion of the joints of the upper extremity in male subjects, with special reference to side. J Bone Joint Surg Am. 1996;78:1401–1404.PubMed Gunal I, Kose N, Erdogan O, Gokturk E, Seber S. Normal range of motion of the joints of the upper extremity in male subjects, with special reference to side. J Bone Joint Surg Am. 1996;78:1401–1404.PubMed
13.
Zurück zum Zitat Hayes K, Walton JR, Szomor ZR, Murrell GA. Reliability of five methods for assessing shoulder range of motion. Aust J Physiother. 2001;47:289–294.PubMed Hayes K, Walton JR, Szomor ZR, Murrell GA. Reliability of five methods for assessing shoulder range of motion. Aust J Physiother. 2001;47:289–294.PubMed
14.
Zurück zum Zitat Jaeger GH, Marcellin-Little DJ, Depuy V, Lascelles BD. Validity of goniometric joint measurements in cats. Am J Vet Res. 2007;68:822–826.PubMedCrossRef Jaeger GH, Marcellin-Little DJ, Depuy V, Lascelles BD. Validity of goniometric joint measurements in cats. Am J Vet Res. 2007;68:822–826.PubMedCrossRef
15.
Zurück zum Zitat Lee J, Koh D, Ong CN. Statistical evaluation of agreement between two methods for measuring a quantitative variable. Comput Biol Med. 1989;19:61–70.PubMedCrossRef Lee J, Koh D, Ong CN. Statistical evaluation of agreement between two methods for measuring a quantitative variable. Comput Biol Med. 1989;19:61–70.PubMedCrossRef
16.
Zurück zum Zitat Lennon RI, Riyat MS, Hilliam R, Anathkrishnan G, Alderson G. Can a normal range of elbow movement predict a normal elbow x ray? Emerg Med J. 2007;24:86–88.PubMedCrossRef Lennon RI, Riyat MS, Hilliam R, Anathkrishnan G, Alderson G. Can a normal range of elbow movement predict a normal elbow x ray? Emerg Med J. 2007;24:86–88.PubMedCrossRef
17.
Zurück zum Zitat London JT. Kinematics of the elbow. J Bone Joint Surg Am. 1981;63:529–535.PubMed London JT. Kinematics of the elbow. J Bone Joint Surg Am. 1981;63:529–535.PubMed
18.
Zurück zum Zitat Macedo LG, Magee DJ. Differences in range of motion between dominant and nondominant sides of upper and lower extremities. J Manipulative Physiol Ther. 2008;31:577–582.PubMedCrossRef Macedo LG, Magee DJ. Differences in range of motion between dominant and nondominant sides of upper and lower extremities. J Manipulative Physiol Ther. 2008;31:577–582.PubMedCrossRef
19.
Zurück zum Zitat Macedo LG, Magee DJ. Effects of age on passive range of motion of selected peripheral joints in healthy adult females. Physiother Theory Pract. 2009;25:145–164.PubMedCrossRef Macedo LG, Magee DJ. Effects of age on passive range of motion of selected peripheral joints in healthy adult females. Physiother Theory Pract. 2009;25:145–164.PubMedCrossRef
20.
Zurück zum Zitat Mayerson NH, Milano RA. Goniometric measurement reliability in physical medicine. Arch Phys Med Rehabil. 1984;65:92–94.PubMed Mayerson NH, Milano RA. Goniometric measurement reliability in physical medicine. Arch Phys Med Rehabil. 1984;65:92–94.PubMed
21.
Zurück zum Zitat Morrey BF, Sanchez-Sotelo J. The Elbow and its Disorders. Ed 4. Philadelphia, PA: Saunders Elsevier; 2008. Morrey BF, Sanchez-Sotelo J. The Elbow and its Disorders. Ed 4. Philadelphia, PA: Saunders Elsevier; 2008.
22.
Zurück zum Zitat Paraskevas G, Papadopoulos A, Papaziogas B, Spanidou S, Argiriadou H, Gigis J. Study of the carrying angle of the human elbow joint in full extension: a morphometric analysis. Surg Radiol Anat. 2004;26:19–23.PubMedCrossRef Paraskevas G, Papadopoulos A, Papaziogas B, Spanidou S, Argiriadou H, Gigis J. Study of the carrying angle of the human elbow joint in full extension: a morphometric analysis. Surg Radiol Anat. 2004;26:19–23.PubMedCrossRef
23.
Zurück zum Zitat Park W, Ramachandran J, Weisman P, Jung ES. Obesity effect on male active joint range of motion. Ergonomics. 2010;53:102–108.PubMedCrossRef Park W, Ramachandran J, Weisman P, Jung ES. Obesity effect on male active joint range of motion. Ergonomics. 2010;53:102–108.PubMedCrossRef
24.
Zurück zum Zitat Petherick M, Rheault W, Kimble S, Lechner C, Senear V. Concurrent validity and intertester reliability of universal and fluid-based goniometers for active elbow range of motion. Phys Ther. 1988;68:966–969.PubMed Petherick M, Rheault W, Kimble S, Lechner C, Senear V. Concurrent validity and intertester reliability of universal and fluid-based goniometers for active elbow range of motion. Phys Ther. 1988;68:966–969.PubMed
25.
Zurück zum Zitat Pique-Vidal C, Maled-Garcia I, Arabi-Moreno J, Vila J. Radiographic angles in hallux valgus: differences between measurements made manually and with a computerized program. Foot Ankle Int. 2006;27:175–180.PubMed Pique-Vidal C, Maled-Garcia I, Arabi-Moreno J, Vila J. Radiographic angles in hallux valgus: differences between measurements made manually and with a computerized program. Foot Ankle Int. 2006;27:175–180.PubMed
26.
Zurück zum Zitat Riddle DL, Rothstein JM, Lamb RL. Goniometric reliability in a clinical setting: shoulder measurements. Phys Ther. 1987;67:668–673.PubMed Riddle DL, Rothstein JM, Lamb RL. Goniometric reliability in a clinical setting: shoulder measurements. Phys Ther. 1987;67:668–673.PubMed
27.
Zurück zum Zitat Rothstein JM, Miller PJ, Roettger RF. Goniometric reliability in a clinical setting: elbow and knee measurements. Phys Ther. 1983;63:1611–1615.PubMed Rothstein JM, Miller PJ, Roettger RF. Goniometric reliability in a clinical setting: elbow and knee measurements. Phys Ther. 1983;63:1611–1615.PubMed
28.
Zurück zum Zitat Shaaban H, Pereira C, Williams R, Lees VC. The effect of elbow position on the range of supination and pronation of the forearm. J Hand Surg Eur Vol. 2008;33:3–8.PubMedCrossRef Shaaban H, Pereira C, Williams R, Lees VC. The effect of elbow position on the range of supination and pronation of the forearm. J Hand Surg Eur Vol. 2008;33:3–8.PubMedCrossRef
29.
Zurück zum Zitat Shiba R, Sorbie C, Siu DW, Bryant JT, Cooke TD, Wevers HW. Geometry of the humeroulnar joint. J Orthop Res. 1988;6:897–906.PubMedCrossRef Shiba R, Sorbie C, Siu DW, Bryant JT, Cooke TD, Wevers HW. Geometry of the humeroulnar joint. J Orthop Res. 1988;6:897–906.PubMedCrossRef
30.
Zurück zum Zitat Steindler A. Kinesiology of the Human Body Under Normal and Pathological Conditions. Ed 5. Springfield, IL: Charles C. Thomas; 1977. Steindler A. Kinesiology of the Human Body Under Normal and Pathological Conditions. Ed 5. Springfield, IL: Charles C. Thomas; 1977.
31.
Zurück zum Zitat Szulc P, Lewandowski J. Verification of selected anatomical landmarks used as reference points for universal goniometer positioning during elbow joint mobility range measurements. Folia Morphol (Warsz). 2003;62:353–355. Szulc P, Lewandowski J. Verification of selected anatomical landmarks used as reference points for universal goniometer positioning during elbow joint mobility range measurements. Folia Morphol (Warsz). 2003;62:353–355.
32.
Zurück zum Zitat Terwee CB, De Winter AF, Scholten RJ, Jans MP, Deville W, van Schaardenburg D, Bouter LM. Interobserver reproducibility of the visual estimation of range of motion of the shoulder. Arch Phys Med Rehabil. 2005;86:1356–1361.PubMedCrossRef Terwee CB, De Winter AF, Scholten RJ, Jans MP, Deville W, van Schaardenburg D, Bouter LM. Interobserver reproducibility of the visual estimation of range of motion of the shoulder. Arch Phys Med Rehabil. 2005;86:1356–1361.PubMedCrossRef
33.
Zurück zum Zitat van de Pol RJ, van Trijffel E, Lucas C. Inter-rater reliability for measurement of passive physiological range of motion of upper extremity joints is better if instruments are used: a systematic review. J Physiother. 2010;56:7–17.PubMedCrossRef van de Pol RJ, van Trijffel E, Lucas C. Inter-rater reliability for measurement of passive physiological range of motion of upper extremity joints is better if instruments are used: a systematic review. J Physiother. 2010;56:7–17.PubMedCrossRef
34.
Zurück zum Zitat Zampagni ML, Casino D, Zaffagnini S, Visani AA, Marcacci M. Estimating the elbow carrying angle with an electrogoniometer: acquisition of data and reliability of measurements. Orthopedics. 2008;31:370.PubMed Zampagni ML, Casino D, Zaffagnini S, Visani AA, Marcacci M. Estimating the elbow carrying angle with an electrogoniometer: acquisition of data and reliability of measurements. Orthopedics. 2008;31:370.PubMed
Metadaten
Titel
Validity of Goniometric Elbow Measurements: Comparative Study with a Radiographic Method
verfasst von
Julien Chapleau, MD
Fanny Canet, MScA
Yvan Petit, PhD
G-Yves Laflamme, MD, FRCSC
Dominique M. Rouleau, MD, MSc, FRCSC
Publikationsdatum
01.11.2011
Verlag
Springer-Verlag
Erschienen in
Clinical Orthopaedics and Related Research® / Ausgabe 11/2011
Print ISSN: 0009-921X
Elektronische ISSN: 1528-1132
DOI
https://doi.org/10.1007/s11999-011-1986-8

Weitere Artikel der Ausgabe 11/2011

Clinical Orthopaedics and Related Research® 11/2011 Zur Ausgabe

Arthropedia

Grundlagenwissen der Arthroskopie und Gelenkchirurgie. Erweitert durch Fallbeispiele, Videos und Abbildungen. 
» Jetzt entdecken

Mehr Frauen im OP – weniger postoperative Komplikationen

21.05.2024 Allgemeine Chirurgie Nachrichten

Ein Frauenanteil von mindestens einem Drittel im ärztlichen Op.-Team war in einer großen retrospektiven Studie aus Kanada mit einer signifikanten Reduktion der postoperativen Morbidität assoziiert.

„Übersichtlicher Wegweiser“: Lauterbachs umstrittener Klinik-Atlas ist online

17.05.2024 Klinik aktuell Nachrichten

Sie sei „ethisch geboten“, meint Gesundheitsminister Karl Lauterbach: mehr Transparenz über die Qualität von Klinikbehandlungen. Um sie abzubilden, lässt er gegen den Widerstand vieler Länder einen virtuellen Klinik-Atlas freischalten.

Klinikreform soll zehntausende Menschenleben retten

15.05.2024 Klinik aktuell Nachrichten

Gesundheitsminister Lauterbach hat die vom Bundeskabinett beschlossene Klinikreform verteidigt. Kritik an den Plänen kommt vom Marburger Bund. Und in den Ländern wird über den Gang zum Vermittlungsausschuss spekuliert.

TEP mit Roboterhilfe führt nicht zu größerer Zufriedenheit

15.05.2024 Knie-TEP Nachrichten

Der Einsatz von Operationsrobotern für den Einbau von Totalendoprothesen des Kniegelenks hat die Präzision der Eingriffe erhöht. Für die postoperative Zufriedenheit der Patienten scheint das aber unerheblich zu sein, wie eine Studie zeigt.

Update Orthopädie und Unfallchirurgie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.