Skip to main content
Erschienen in: Journal of NeuroEngineering and Rehabilitation 1/2005

Open Access 01.12.2005 | Research

Influence of a portable audio-biofeedback device on structural properties of postural sway

verfasst von: Marco Dozza, Lorenzo Chiari, Becky Chan, Laura Rocchi, Fay B Horak, Angelo Cappello

Erschienen in: Journal of NeuroEngineering and Rehabilitation | Ausgabe 1/2005

Abstract

Background

Good balance depends on accurate and adequate information from the senses. One way to substitute missing sensory information for balance is with biofeedback technology. We previously reported that audio-biofeedback (ABF) has beneficial effects in subjects with profound vestibular loss, since it significantly reduces body sway in quiet standing tasks.

Methods

In this paper, we present the effects of a portable prototype of an ABF system on healthy subjects' upright stance postural stability, in conditions of limited and unreliable sensory information. Stabilogram diffusion analysis, combined with traditional center of pressure analysis and surface electromyography, were applied to the analysis of quiet standing tasks on a Temper foam surface with eyes closed.

Results

These analyses provided new evidence that ABF may be used to treat postural instability. In fact, the results of the stabilogram diffusion analysis suggest that ABF increased the amount of feedback control exerted by the brain for maintaining balance. The resulting increase in postural stability was not at the expense of leg muscular activity, which remained almost unchanged.

Conclusion

Examination of the SDA and the EMG activity supported the hypothesis that ABF does not induce an increased stiffness (and hence more co-activation) in leg muscles, but rather helps the brain to actively change to a more feedback-based control activity over standing posture.
Hinweise

Electronic supplementary material

The online version of this article (doi:10.​1186/​1743-0003-2-13) contains supplementary material, which is available to authorized users.

Competing interests

The authors of this paper declare that they applied for a patent (patent application PCT/IB2004/001679) concerning the audio-biofeedback device used for the study described in this paper. The patent is property of the University of Bologna and of the Oregon Health & Science University.
Abkürzungen
ABF
= audio biofeedback
Acc-RMS
= root mean square of the acceleration
AP
= anterior-posterior
BF
= biofeedback
COP
= center of pressure
COP-RMS
= root mean square of the COP
EMG
= electromyography
F95%
= frequency comprising the 95% of the power
GA
= gastrocnemius
ML
= medial-lateral
SDA
= stabilogram diffusion analysis
SO
= soleus
TI
= tibialis

Background

Maintaining balance is a complex task accomplished by the brain through the fusion and interpretation of sensory information. When sensory information from vestibular, somatosensory, and visual systems [13] are not accurate and/or adequate, balance will be compromised. Although, in many cases, the loss of peripheral sensory information is not curable or reversible, the brain can compensate for the loss of sensory information by relying more on the other sensory channels [4, 5].
The purpose of biofeedback (BF) systems for postural control is to provide additional sensory information about body equilibrium to the brain [6]. In the last few years, different sensors, encoding algorithms, and information restitution devices have been combined to develop promising BF systems for postural control [79]. The major design goals were focused on portability, usability, economy, and effectiveness in balance improvements [8, 1012].
The development of these BF systems has been facilitated by the availability of lightweight, miniaturized, and economical sensors such as accelerometers, inclinometers, and gyroscopes [13]. The use of these sensors makes BF devices inexpensive, unsusceptible to shadowing effect, and not limited in the measurement field, in contrast to dynamometric platforms and motion analysis systems, which are commonly used in laboratory settings [14, 15]. In addition, due to their size and weight, these sensors can measure body segment movement without hindering natural motor execution.
More detail is needed in understanding how biofeedback information interacts with the brain or, from a neuroscience perspective, how the brain uses artificial BF information and combines it with natural sensory information. We believe that understanding this interaction is fundamental for further developing effective BF systems.
An interesting analysis in the understanding of how the brain may use BF information for postural control was proposed by Collins and De Luca [16]. These authors developed a statistical-biomechanics method for analyzing force platform data recorded during quiet standing, called stabilogram diffusion analysis (SDA). SDA was applied to center of pressure (COP) data and it disclosed that COP tends to drift away from a relative equilibrium point over short-term observation intervals (less than 1-second long), whereas COP tends to return to a relative equilibrium point over long-term observation intervals. These results took Collins and De Luca to suggest that the motion of the COP is not purely random, and that SDA may be able to give insight on the amount of open-loop and closed-loop postural control applied by the central nervous system for maintaining balance [17]. SDA was used in several contexts, e.g. to evaluate the effect of spaceflight [18], visual input [19, 20], and age-related changes [21, 22] on postural stability. Chiari el al [20] developed and validated a new nonlinear model for extracting parameters from SDA diagrams, reducing from 6 to 2 the number of the parameters used to characterize the structural properties of COP. Rocchi et al. [23] found that these new parameters may be useful adjuncts to evaluate postural control strategies in patients with Parkinson's disease and may allow the comparison of different deep brain stimulation electrode sites based on their effect on structural properties of the COP.
In this paper, we investigate the effect on postural stability of a portable, accelerometry-based, audio biofeedback (ABF) system recently developed by the authors [9]. Standing with eyes closed on Temper™ foam will be used to evaluate the effects of artificial auditory cues to enhance the limited (from the eyes) and unreliable (from the feet) natural sensory information. Measurements include COP recorded by a force platform under the feet, trunk acceleration measured by the ABF sensors, and EMG signals from the leg muscles. SDA according to [20], traditional COP analysis [24], and muscle activation analysis according to [25] were performed in order to evaluate the effect of ABF on healthy young subject's upright posture.
These analyses were aimed to answer two questions: (1) do structural properties of postural sway change with ABF? And, if so, (2) in which way will this help in understanding the mechanisms underlying ABF efficacy and in improving the design of a rehabilitation strategy for balance disorders?
In this paper, we present evidence that supports the hypothesis that ABF does not induce a purely biomechanical increase in stiffness (and hence more co-activation) in the leg muscles, but rather ABF helps the brain actively adapt its control activity over standing posture.

Methods

Participants

Eight healthy subjects participated in this study (5 males and 3 females, aged 23.5 ± 3.0 yrs, range 21–28 yrs). All participants were free from any neurological, orthopaedic, hearing, or vestibular disorder. Informed consent form was obtained from each subject. The form was prepared in accordance with the Oregon Health and Science University Ethical Committee and respected the declaration of Helsinky, 1964.

Apparatus and procedure

Subjects performed 10, 60-second trials standing with eyes closed on Temper™, 4"-thick foam. COP displacement was recorded with an AMTI OR6-6 force plate. An ABF system [9] was used to provide subjects with additional balance information related to trunk acceleration. The ABF system used a sensor, based on 2-D accelerometers (Analog Device ADXL203) mounted on the subject's back (L5), to create an audio stereo sound representing the acceleration sensed along the anterior-posterior (AP) and the medial-lateral (ML) direction. A laptop, Toshiba Celeron 2.3 GHz, was dedicated to convert the accelerations into stereo sounds. Commercial headphones were used by the subjects to listen to the ABF sound. The ABF system is described in detail in [9] and illustrated in Figure 1. In short, the stereo sound provided by the ABF system consisted of two sine waves, one for the left ear channel and one for the right ear channel. Pitch, volume and left/right balance of the stereo sound were modulated to represent the 2-D acceleration information. Specifically, when the subject swayed forward, and consequently the acceleration increased in the anterior direction, the sound got louder in volume and higher in pitch. When the subject swayed backward, and consequently the acceleration increased in the posterior direction, the sound got louder in volume and lower in pitch. When the subject moved right and, consequently, the acceleration increased in the right direction, the sound got louder in the right ear channel and lower in the left one. When the subject moved left, and consequently the acceleration increased in the left direction, the sound got louder in the left ear channel and lower in the right one. The sound dynamics was optimized for each trial by taking as a reference the first 10-second recordings of each trial. The equations used for the pitch, volume, and left/right balance modulation can be found in [9]. Each subject was instructed to maintain balance during the trials by taking advantage of the ABF information, when available. Five trials with ABF and 5 trials without ABF were performed in randomized order by each subject. Before the experimental session, the subjects were instructed on how ABF codes trunk acceleration into sound, and performed free-movement trials until they felt confident in performing the full experiment.

Data recording

For each standing trial, ground reaction forces and torques were recorded from the force plate with a 100-Hz sampling frequency. COP displacement was computed offline from the force plate data after applying a 10-Hz cut-off, zero phase, low-pass Butterworth filter. Accelerations from the trunk along AP and ML direction were collected with a 100 Hz sampling frequency. EMG was recorded from right leg muscles, Tibialis (TI), Soleus (SO), and Gastrocnemius (GA) with two surface electrodes fixed about 6–8 cm apart along the length of each muscle belly; the ground electrode was fixed on a bony area of the right Hallux. The EMG signals were acquired with a 100-Hz sampling frequency, amplified 20000 times, band-pass filtered (71-2652 Hz), integrated with a 6th order Butterworth low-pass filter with a cut-off of 100 Hz (National Semiconductor MF6-100), and full-wave rectified.

Data analysis

From AP COP data, the root mean square distance (COP-RMS) and the frequency comprising the 95% of the power (F95%) were extracted according to Prieto et al. [24].
From the acceleration sensed at trunk level along AP direction we computed the root mean square value (Acc-RMS).
In addition, two stochastic parameters were included in the analyses. These parameters characterize a previously developed model that describes with continuity the transition among the different scaling regimes found in the COP time series [20] The model is described by the following equation:
V(Δt) = K Δt2H(Δt)
where V(Δt) is the variance of COP displacement, computed at time-lag Δt, and H is the scaling exponent, also called Hurst exponent. This is assumed to follow a sigmoid law in the time interval (Δt):
https://static-content.springer.com/image/art%3A10.1186%2F1743-0003-2-13/MediaObjects/12984_2005_Article_27_Equa_HTML.gif
In this way, the features extracted from COP data are the following (see [20] for more details):
K is an estimate of the diffusion coefficient of the random process obtained by sampling the COP time series at the sampling frequency 1/ΔTc.
ΔTc represents the time-lag at which the real process corresponds to a purely random behavior, and where it switches from a persistent (positively correlated, and hence interpreted in terms of feed-forward control) to an anti-persistent (negatively correlated, and hence interpreted in terms of feedback control) behavior [16].
Mean muscular activity was calculated from the full wave rectified EMG of each muscle. Muscle activity was expressed as percentage of the maximal recorded activity for each muscle in each subject. This procedure allowed a reliable comparison of muscle activity between-subjects. The EMG signals were further processed applying a zero phase, low pass-filter with a 2 Hz cut-off in order to obtain tension curves according to Olney and Winter [25]. These tension curves were cross-correlated to determine the amount of co-activation between the muscles recorded.

Statistical analysis

Paired T-tests were performed to determine the effect of ABF on the different parameters extracted from COP, acceleration and EMG data collected. The threshold for statistical significance was set to p = 0.05.

Results

Subjects' confidence and comfort

All participants reported ABF sound was comfortable and its way of representing the information was intuitive. In fact, none of the subjects needed more than two, free-movement trials before feeling ready to start the experiment.

Subjects' sway

ABF significantly influenced subjects' balance on the foam. The percentage change induced by ABF on all sway parameters, either measured at the trunk level with the accelerometer or at the feet level with the force platform, is shown in Figure 2. Figure 2 also reports significance levels of the parameter changes occurred while using the ABF. The general results shown in Figure 2 are specified in detail in the following.

Center of Pressure analysis

Center of pressure displacement in the AP direction was significantly influenced by ABF. T-test results revealed significant effects of ABF on COP-RMS (p = 0.015). This effect is shown by a consistent reduction of COP-RMS for 7 out of 8 subjects as shown in Table 1 (column 7). Average reduction of COP-RMS was 10.7%. Columns 1 and 4 of Table 1 also show the subject-by-subject values of COP-RMS without and with ABF, respectively. The last three subjects (#6, #7, #8) were females and showed smaller COP-RMS, as expected considering their smaller heights [26].
F95% increased with ABF for 7 out of 8 subjects (Table 1, column 8) but this result was not significant (p = 0.42). The values of F95% are also reported for each subject in both conditions (Table 1, columns 2 and 5). Average increase of F95% due to ABF was 6.2% as shown in Figure 2.
It is worth noting that subject #8 behaved as an outlier (Figure 3), compared to the other subjects since she was the only one who showed opposite changes in COP-RMS and F95% while using ABF. Performing the T-Tests, after eliminating this outlier, increased the significance of using ABF on COP-RMS and on F95% (p = 0.002 and p = 0.02, respectively). These results better match the results already published in [9]. The outlying behavior of subject #8 will be investigated further in the discussion.
Table 1
ABF effect on sway parameters Parameters. COP-RMS, F95%, and Acc-RMS are reported, subject-by-subject, for trials with and without ABF. Percentage differences between these two conditions are also reported. Standard deviations are indicated in parenthesis.
 
COP-RMS
(NO – ABF) [mm]
F95 %
(NO – ABF) [Hz]
Acc-RMS
(NO – ABF) [mm/s2]
COP-RMS
(ABF) [mm]
F95 %
(ABF) [Hz]
Acc-RMS
(ABF) [mm/s2]
COP-RMS
%difference
F95%
%difference
Acc-RMS
%difference
Subj. #1
10.79 (2.84)
0.99 (0.05)
137 (48)
9.57 (1.86)
1.18 (0.16)
118 (13)
-11.2
19.1
-14.1
Subj. #2
9.91 (2.77)
1.20 (0.29)
142 (27)
9.50 (2.26)
1.30 (0.20)
120 (23)
-4.1
8.7
-15.6
Subj. #3
9.21 (2.94)
1.16 (0.14)
121 (23)
8.61 (1.42)
1.37 (0.07)
113 (21)
-6.5
18.0
-7.0
Subj. #4
10.23 (1.50)
1.43 (0.08)
117 (30)
8.80 (1.74)
1.49 (0.12)
100 (12)
-13.9
4.1
-14.6
Subj. #5
8.50 (0.93)
1.49 (0.22)
143 (46)
6.90 (1.35)
1.53 (0.28)
115 (19)
-18.8
2.6
-19.3
Subj. #6
9.62 (1.55)
1.34 (0.30)
126 (43)
7.35 (0.88)
1.34 (0.09)
89 (20)
-23.6
0.0
-29.2
Subj. #7
6.37 (1.48)
1.60 (0.07)
64 (8.3)
5.19 (0.59)
1.94 (0.12)
51 (4.7)
-18.5
20.8
-20.1
Subj. #8
6.08 (1.19)
1.78 (0.25)
48 (6.3)
6.75 (1.41)
1.37 (0.16)
39 (3.8)
10.9
-23.1
-17.3
Average
8.84 (1.75)
1.37 (0.26)
112 (36)
7.83 (1.54)
1.44 (0.15)
93 (31)
-10.7 (10.92)
6.2 (14.4)
-17.2 (6.3)

Acceleration analysis

Acceleration sensed at trunk level (L5) in AP direction was significantly reduced by ABF. T-test results also revealed significant effects of ABF on Acc-RMS (p = 0.0009). Acc-RMS was reduced by ABF across all subjects, as shown in Table 1 (last column).
Average reduction of Acc-RMS was 17.2% (Figure 2). Columns 3 and 7 of Table 1 also show the subject-by-subject values of Acc-RMS without and with ABF, respectively. The last three subjects were females and showed smaller Acc-RMS, as expected considering their smaller heights [26].

Stabilogram diffusion analysis

SDA diagrams plotted from AP COP data, were also significantly influenced by ABF (Figure 4). As a consequence, the parameters K and ΔTc characterizing the SDA diagram were both significantly decreased by ABF (Figure 2). Average K reduction was 9.3% (p = 0.02), whereas average ΔTc reduction was 33.9% (p = 0.018). Table 2 reports the subject-by-subject values of K and ΔTc in both conditions tested. Subject #8 and subject #7 are the only ones who showed a slight increase in K.
Table 2
ABF effect on SDA parameters Parameters. K and ΔTc are reported, subject-by-subject, for trials with and without ABF. Percentage differences between these two conditions are also reported. Standard deviations are indicated in parenthesis.
 
K (NO-ABF) [mm2]
Δtc (NO-ABF) [s]
K (ABF) [mm2]
Δtc (ABF) [s]
K % difference
Δtc % difference
Subj. #1
100 (57)
0.42 (0.21)
86 (15)
0.38 (0.17)
-14.6
-9.9
Subj. #2
70 (29)
0.51 (0.31)
66 (20)
0.41 (0.34)
-7.4
-20.5
Subj. #3
75 (41)
0.52 (0.29)
65 (20)
0.29 (0.12)
-13.3
-45.3
Subj. #4
80 (21)
0.81 (0.46)
70 (14)
0.39 (0.14)
-11.1
-52.0
Subj. #5
47 (13)
0.32 (0.08)
39 (10)
0.26 (0.16)
-18.1
-19.7
Subj. #6
64 (12)
0.27 (0.08)
61 (9)
0.20 (0.09)
-5.7
-26.1
Subj. #7
32 (7)
0.17 (0.06)
34 (9)
0.09 (0.01)
6.6
-47.4
Subj. #8
35 (14)
0.29 (0.09)
38 (13)
0.19 (0.06)
5.8
-34.3
Average
63 (23)
0.41 (0.20)
57 (18.5)
0.27 (0.11)
-9.3 (9.2)
-33.9 (15.3)

Muscle activity analysis

Muscle activity of TI, GA, and SO was not influenced by ABF. Overall, the mean activity, expressed as a percentage of the maximal activity recorded from each single muscle across all the trials of a subject, did not change significantly due to ABF (see Figure 5A). TI activity showed a trend toward increasing in trials with ABF (p = 0.17) but this change was particularly clear only for subjects #4 and #7.
Muscle co-activation of ankle agonists-antagonists did not change significantly due to the ABF (see Figure 5B). Co-activation between TI and GA was small both with (r2 = 0.11) and without (r2 = 0.08) ABF. Similarly small was the co-activation between TI and SO with (r2 = 0.14) and without (r2 = 0.09) ABF. As expected, co-activation between the agonists muscles, GA and SO, was instead large (r2 = 0.39 in trials with ABF and r2 = 0.46 in trials without ABF). Figure 5B reports the coefficient of determination r2, which indicates the amount of muscular co-activation, for all pairs of muscles analyzed in trials with and without ABF.

Discussion

Using the proposed ABF device, all healthy subjects included in this study could sway less when standing in a particularly challenging condition, with vision unavailable and somatosensation partly unreliable. All subjects, in fact, reduced their AP Acc-RMS (see Table 1). In this way, subjects were further from their stability limits and, consequently, more stable. Trunk stabilization resulted in smaller corrective torques at the ankles, and hence smaller COP displacements. All but one subjects (Subj. #8) showed a significant decrease in AP COP-RMS (Fig. 2). During ABF, postural corrections in leg muscles were smaller but more frequent in number, as suggested by the increase in F95% of the COP. Future studies involving more sophisticated techniques for the acquisition and analysis of the EMG signals will be needed to validate this hypothesis. This result suggests that ABF may partially substitute for the lack of visual and somatosensory information for postural control by taking the postural control system towards a new steady state associated with a different control strategy.
Examination of the SDA and the EMG activity supported the hypothesis that ABF does not induce an increased stiffness (and hence more co-activation) in leg muscles, but rather helps the brain to actively change to a more feedback-based control over standing posture. Representative SDA diagrams reported in Figure 4 suggest that ABF contributes to a general reduction of both the diffusion coefficient K and the transition time ΔTc. Downward shifts of the SDA diagrams, described by smaller diffusion coefficients, reflect a reduced stochastic activity of the COP, and hence a more tightly regulated control system [16]. Shorter transition times reflect an earlier switching between persistent and anti-persistent behaviors, and hence more prompt reactions to perturbations of the postural control system [27]. In summary, these results support the hypotheses that ABF: 1) increases postural stability in stance, and 2) results in a more prominent role for feedback control over feed-forward control. Hence, the solution proposed by the brain with ABF seems to involve more feedback control for a more stable sway.
Interestingly, our results differ from the results observed by Rougier in quiet stance experiments with visual BF [28]. With visual BF, SDA diagrams only changed some local properties (local slopes) over short or long observation intervals but did not shift significantly, consistent with little, if any, change in K. Furthermore, with visual BF, closed-loop control operated over longer observation-times, suggesting that feed-forward control expanded over feedback control. Such a different behavior between auditory and visual BF may be due to the peculiar, non-redundant role of different senses in multi-sensory integration for the control of posture [29]. Whereas vision provides information about the external environment, it allows predictions of forthcoming events in the scene (feed-forward control) [30]. In contrast, hearing, compared to vision, may be more important for postural reactions to disturbing stimuli (feedback control). This result can also be related to the different processing times required by the central nervous system for visual and auditory stimuli with auditory reaction times significantly faster than visual reaction times. Finally, another factor which may explain the different outcomes of the two BF-studies is the selection of two, different, input variables (COP for visual BF and Acceleration from the trunk for ABF). It is widely accepted that upper- and lower- body segments are controlled separately [31].
Both predictive (feed-forward) and reactive (feedback) control need to be used in order to have an adequate interaction with the environment for postural stability. For this reason, it's hard to determine the relative validity of audio and visual BF. Rather, it may be important, in a rehabilitation setting, to identify which one of the two components of postural control (feed-forward or feedback) needs more reinforcement or substitution in a particular patient, and consequently design an optimised BF treatment.
The outlying results observed for Subj. #8 need to be discussed individually. This woman in fact did not decrease COP-RMS and K, and did not increase F95%, although, similarly to the other subjects, she decreased Acc-RMS and ΔTc (these changes were consistent across the whole population). Hence, with ABF she actually swayed less and she showed the same increase of feedback control. Nonetheless, either due to her small body size or to a slightly different control scheme, she obtained these goals with a different strategy. Figure 6 reports her muscle activities and co-activations. It can be seen how she generally increases muscle activity with ABF (Figure 6A), in particular with a large increase in the activity of posterior muscles, GA and SO. It should be noted, however, that also the estimated co-activations (Figure 6B) look pretty dissimilar compared with the ones of the other subjects, shown in Figure 5B. Particularly low is the co-activation of agonists muscles GA-SO without ABF, which ABF partly contributes to enlarge. For all these reasons, her postural behavior in the proposed task should be looked as an outlying behavior and more analyses are needed, on a larger population, to assess the real influence of body size or usual control strategies on the responsiveness to ABF.
Many earlier biofeedback systems used audio alarms to notify the user of abnormal values of monitored parameters (e.g. [32]). The present ABF system is novel in the use of nonlinear coding functions and in the customization of these functions for each subject and task [9]. Although the current ABF system may interfere with use of hearing for communication, it may be quite useful during the rehabilitation and training process. Plans are underway to improve the current ABF system by making it wireless for increased portability and equipping it with a communication module for remote control, recording, and monitoring. Different sonification procedures will also be tested and compared in a near future. Specifically, 3-D generated sound with a HRTF (Head Related Transfer Function) or immersive sound may be even more effective signals for improving stance balance.

Conclusion

We have investigated the attributes of a portable instrument that feeds back trunk acceleration in order to help subjects to reduce their postural sway during stance. The instrument meets requirements for an adequate biofeedback system that may find interesting applications not only as a rehabilitation device in the clinic, but also in the home care setting, and when doing community mobility training outside the traditional clinical setting. In fact, it has appropriate bandwidth and sensitivity, smoothness and delay of the acoustic signal generator, as well as portability. Acoustic information related to trunk movement allowed subjects in the present experiment to increase postural stability when sensory information from both vision and the surface were compromised by eye closure and stance on foam. We provided evidence that the balance improvement was not a stiffening at the ankle, but rather the brain actively adapted its control strategy over standing posture with more feedback-based control.

Acknowledgements

We thank Jim Frank, Maria Grazia Benedetti, and Marco Knaflitz for the advice about electromyography analysis. We also thank Charles Russell and Andrew Owings for technical assistance.
Open Access This article is published under license to BioMed Central Ltd. This is an Open Access article is distributed under the terms of the Creative Commons Attribution License ( https://​creativecommons.​org/​licenses/​by/​2.​0 ), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.

Competing interests

The authors of this paper declare that they applied for a patent (patent application PCT/IB2004/001679) concerning the audio-biofeedback device used for the study described in this paper. The patent is property of the University of Bologna and of the Oregon Health & Science University.
Literatur
1.
Zurück zum Zitat Horak FB, Shupert CL, Mirka A: Components of postural dyscontrol in the elderly: a review. Neurobiol Aging. 1989, 10: 727-738. 10.1016/0197-4580(89)90010-9.CrossRefPubMed Horak FB, Shupert CL, Mirka A: Components of postural dyscontrol in the elderly: a review. Neurobiol Aging. 1989, 10: 727-738. 10.1016/0197-4580(89)90010-9.CrossRefPubMed
2.
Zurück zum Zitat Lord SR, Ward JA: Age-associated differences in sensori-motor function and balance in community dwelling women. Age Ageing. 1994, 23: 452-460.CrossRefPubMed Lord SR, Ward JA: Age-associated differences in sensori-motor function and balance in community dwelling women. Age Ageing. 1994, 23: 452-460.CrossRefPubMed
3.
Zurück zum Zitat Corriveau H, Prince F, Hebert R, Raiche M, Tessier D, Maheux P, Ardilouze JL: Evaluation of postural stability in elderly with diabetic neuropathy. Diabetes Care. 2000, 23: 1187-1191.CrossRefPubMed Corriveau H, Prince F, Hebert R, Raiche M, Tessier D, Maheux P, Ardilouze JL: Evaluation of postural stability in elderly with diabetic neuropathy. Diabetes Care. 2000, 23: 1187-1191.CrossRefPubMed
4.
Zurück zum Zitat Horak FB, Macpherson JM: Postural equilibrium and orientation. Published for the American Physiology Society by Oxford University Press. Edited by: Rowell RB, Shepherd JT. 1996, New York: Published for the American Physiology Society by Oxford University Press, 255-292. Horak FB, Macpherson JM: Postural equilibrium and orientation. Published for the American Physiology Society by Oxford University Press. Edited by: Rowell RB, Shepherd JT. 1996, New York: Published for the American Physiology Society by Oxford University Press, 255-292.
5.
Zurück zum Zitat Horak FB, Hlavacka F: Somatosensory loss increases vestibulospinal sensitivity. J Neurophysiol. 2001, 86: 575-585.PubMed Horak FB, Hlavacka F: Somatosensory loss increases vestibulospinal sensitivity. J Neurophysiol. 2001, 86: 575-585.PubMed
6.
Zurück zum Zitat Moore S, Woollacott M: The use of biofeedback to improve postural stability. Phys Ther Practice. 1993, 2: 1-19. Moore S, Woollacott M: The use of biofeedback to improve postural stability. Phys Ther Practice. 1993, 2: 1-19.
7.
Zurück zum Zitat Wall C, Weinberg MS, Schmidt PB, Krebs DE: Balance prosthesis based on micromechanical sensors using vibrotactile feedback of tilt. IEEE Trans Biomed Eng. 2001, 48: 1153-1161. 10.1109/10.951518.CrossRefPubMed Wall C, Weinberg MS, Schmidt PB, Krebs DE: Balance prosthesis based on micromechanical sensors using vibrotactile feedback of tilt. IEEE Trans Biomed Eng. 2001, 48: 1153-1161. 10.1109/10.951518.CrossRefPubMed
8.
Zurück zum Zitat Tyler M, Danilov Y, Bach YR: Closing an open-loop control system: vestibular substitution through the tongue. J Integr Neurosci. 2003, 2: 159-164. 10.1142/S0219635203000263.CrossRefPubMed Tyler M, Danilov Y, Bach YR: Closing an open-loop control system: vestibular substitution through the tongue. J Integr Neurosci. 2003, 2: 159-164. 10.1142/S0219635203000263.CrossRefPubMed
9.
Zurück zum Zitat Chiari L, Dozza M, Cappello A, Horak FB, Macellari V, Giansanti D: An accelerometry-based system for balance improvement using audio-biofeedback. IEEE Trans Biomed Eng. 2005, Chiari L, Dozza M, Cappello A, Horak FB, Macellari V, Giansanti D: An accelerometry-based system for balance improvement using audio-biofeedback. IEEE Trans Biomed Eng. 2005,
10.
Zurück zum Zitat Bazzarelli M, Durdle N, Lou E, Raso J, Hill D: A wearable networked embedded system for the treatment of scoliosis. Stud Health Technol Inform. 2002, 91: 383-386.PubMed Bazzarelli M, Durdle N, Lou E, Raso J, Hill D: A wearable networked embedded system for the treatment of scoliosis. Stud Health Technol Inform. 2002, 91: 383-386.PubMed
11.
Zurück zum Zitat Kadkade PP, Benda BJ, Schmidt PB, Wall C: Vibrotactile display coding for a balance prosthesis. IEEE Trans Neural Syst Rehabil Eng. 2003, 11: 392-399. 10.1109/TNSRE.2003.819937.CrossRefPubMed Kadkade PP, Benda BJ, Schmidt PB, Wall C: Vibrotactile display coding for a balance prosthesis. IEEE Trans Neural Syst Rehabil Eng. 2003, 11: 392-399. 10.1109/TNSRE.2003.819937.CrossRefPubMed
12.
Zurück zum Zitat Dozza M, Chiari L, Horak FB: Audio- biofeedback improves balance in patients with bilateral vestibular loss. Arch Phys Med Rehabil. 2005, 86: 1401-1403. 10.1016/j.apmr.2004.12.036.CrossRefPubMed Dozza M, Chiari L, Horak FB: Audio- biofeedback improves balance in patients with bilateral vestibular loss. Arch Phys Med Rehabil. 2005, 86: 1401-1403. 10.1016/j.apmr.2004.12.036.CrossRefPubMed
13.
Zurück zum Zitat Luinge HJ, Veltink PH: Inclination measurement of human movement using a 3-D accelerometer with autocalibration. IEEE Trans Neural Syst Rehabil Eng. 2004, 12: 112-121. 10.1109/TNSRE.2003.822759.CrossRefPubMed Luinge HJ, Veltink PH: Inclination measurement of human movement using a 3-D accelerometer with autocalibration. IEEE Trans Neural Syst Rehabil Eng. 2004, 12: 112-121. 10.1109/TNSRE.2003.822759.CrossRefPubMed
14.
Zurück zum Zitat Mayagoitia RE, Lotters JC, Veltink PH, Hermens H: Standing balance evaluation using a triaxial accelerometer. Gait Posture. 2002, 16: 55-59. 10.1016/S0966-6362(01)00199-0.CrossRefPubMed Mayagoitia RE, Lotters JC, Veltink PH, Hermens H: Standing balance evaluation using a triaxial accelerometer. Gait Posture. 2002, 16: 55-59. 10.1016/S0966-6362(01)00199-0.CrossRefPubMed
15.
Zurück zum Zitat Mayagoitia RE, Nene AV, Veltink PH: Accelerometer and rate gyroscope measurement of kinematics: an inexpensive alternative to optical motion analysis systems. J Biomech. 2002, 35: 537-542. 10.1016/S0021-9290(01)00231-7.CrossRefPubMed Mayagoitia RE, Nene AV, Veltink PH: Accelerometer and rate gyroscope measurement of kinematics: an inexpensive alternative to optical motion analysis systems. J Biomech. 2002, 35: 537-542. 10.1016/S0021-9290(01)00231-7.CrossRefPubMed
16.
Zurück zum Zitat Collins JJ, De Luca CJ: Open-loop and closed-loop control of posture: a random-walk analysis of center-of-pressure trajectories. Exp Brain Res. 1993, 95: 308-318. 10.1007/BF00229788.CrossRefPubMed Collins JJ, De Luca CJ: Open-loop and closed-loop control of posture: a random-walk analysis of center-of-pressure trajectories. Exp Brain Res. 1993, 95: 308-318. 10.1007/BF00229788.CrossRefPubMed
17.
Zurück zum Zitat Collins JJ, De Luca CJ: Upright, correlated random walks: A statistical-biomechanics approach to the human postural control system. Chaos. 1995, 5: 57-63. 10.1063/1.166086.CrossRefPubMed Collins JJ, De Luca CJ: Upright, correlated random walks: A statistical-biomechanics approach to the human postural control system. Chaos. 1995, 5: 57-63. 10.1063/1.166086.CrossRefPubMed
18.
Zurück zum Zitat Collins JJ, De Luca CJ, Pavlik AE, Roy SH, Emley MS: The effects of spaceflight on open-loop and closed-loop postural control mechanisms: human neurovestibular studies on SLS-2. Exp Brain Res. 1995, 107: 145-150. 10.1007/BF00228026.CrossRefPubMed Collins JJ, De Luca CJ, Pavlik AE, Roy SH, Emley MS: The effects of spaceflight on open-loop and closed-loop postural control mechanisms: human neurovestibular studies on SLS-2. Exp Brain Res. 1995, 107: 145-150. 10.1007/BF00228026.CrossRefPubMed
19.
Zurück zum Zitat Collins JJ, De Luca CJ: The effects of visual input on open-loop and closed-loop postural control mechanisms. Exp Brain Res. 1995, 103: 151-163. 10.1007/BF00241972.CrossRefPubMed Collins JJ, De Luca CJ: The effects of visual input on open-loop and closed-loop postural control mechanisms. Exp Brain Res. 1995, 103: 151-163. 10.1007/BF00241972.CrossRefPubMed
20.
Zurück zum Zitat Chiari L, Bertani A, Cappello A: Classification of visual strategies in human postural control by stochastic parameters. Hum Mov Sci. 2000, 19: 817-842. 10.1016/S0167-9457(01)00024-0.CrossRef Chiari L, Bertani A, Cappello A: Classification of visual strategies in human postural control by stochastic parameters. Hum Mov Sci. 2000, 19: 817-842. 10.1016/S0167-9457(01)00024-0.CrossRef
21.
Zurück zum Zitat Collins JJ, De Luca CJ, Burrows A, Lipsitz LA: Age-related changes in open-loop and closed-loop postural control mechanisms. Exp Brain Res. 1995, 104: 480-492. 10.1007/BF00231982.CrossRefPubMed Collins JJ, De Luca CJ, Burrows A, Lipsitz LA: Age-related changes in open-loop and closed-loop postural control mechanisms. Exp Brain Res. 1995, 104: 480-492. 10.1007/BF00231982.CrossRefPubMed
22.
Zurück zum Zitat Laughton CA, Slavin M, Katdare K, Nolan L, Bean JF, Kerrigan DC, Philips E, Lipsitz LA, Collins JJ: Aging, muscle activity, and balance control: physiologic changes associated with balance impairment. Gait Posture. 2003, 18: 101-108. 10.1016/S0966-6362(02)00200-X.CrossRefPubMed Laughton CA, Slavin M, Katdare K, Nolan L, Bean JF, Kerrigan DC, Philips E, Lipsitz LA, Collins JJ: Aging, muscle activity, and balance control: physiologic changes associated with balance impairment. Gait Posture. 2003, 18: 101-108. 10.1016/S0966-6362(02)00200-X.CrossRefPubMed
23.
Zurück zum Zitat Rocchi L, Chiari L, Cappello A, Gross A, Horak FB: Comparison between subthalamic nucleus and globus pallidus internus stimulation for postural performance in Parkinson's disease. Gait Posture. 2004, 19: 172-183. 10.1016/S0966-6362(03)00059-6.CrossRefPubMed Rocchi L, Chiari L, Cappello A, Gross A, Horak FB: Comparison between subthalamic nucleus and globus pallidus internus stimulation for postural performance in Parkinson's disease. Gait Posture. 2004, 19: 172-183. 10.1016/S0966-6362(03)00059-6.CrossRefPubMed
24.
Zurück zum Zitat Prieto TE, Myklebust JB, Hoffmann RG, Lovett EG, Myklebust BM: Measures of postural steadiness: differences between healthy young and elderly adults. IEEE Trans Biomed Eng. 1996, 43: 956-966. 10.1109/10.532130.CrossRefPubMed Prieto TE, Myklebust JB, Hoffmann RG, Lovett EG, Myklebust BM: Measures of postural steadiness: differences between healthy young and elderly adults. IEEE Trans Biomed Eng. 1996, 43: 956-966. 10.1109/10.532130.CrossRefPubMed
25.
Zurück zum Zitat Olney SJ, Winter DA: Predictions of knee and ankle moments of force in walking from EMG and kinematic data. J Biomech. 1985, 18: 9-20. 10.1016/0021-9290(85)90041-7.CrossRefPubMed Olney SJ, Winter DA: Predictions of knee and ankle moments of force in walking from EMG and kinematic data. J Biomech. 1985, 18: 9-20. 10.1016/0021-9290(85)90041-7.CrossRefPubMed
26.
Zurück zum Zitat Chiari L, Rocchi L, Cappello A: Stabilometric parameters are affected by anthropometry and foot placement. Clin Biomech (Bristol, Avon). 2002, 17: 666-677. 10.1016/S0268-0033(02)00107-9.CrossRef Chiari L, Rocchi L, Cappello A: Stabilometric parameters are affected by anthropometry and foot placement. Clin Biomech (Bristol, Avon). 2002, 17: 666-677. 10.1016/S0268-0033(02)00107-9.CrossRef
27.
Zurück zum Zitat Chiari L, Cappello A, Lenzi D, Della CU: An improved technique for the extraction of stochastic parameters from stabilograms. Gait Posture. 2000, 12: 225-234. 10.1016/S0966-6362(00)00086-2.CrossRefPubMed Chiari L, Cappello A, Lenzi D, Della CU: An improved technique for the extraction of stochastic parameters from stabilograms. Gait Posture. 2000, 12: 225-234. 10.1016/S0966-6362(00)00086-2.CrossRefPubMed
28.
Zurück zum Zitat Rougier P: Influence of visual feedback on successive control mechanisms in upright quiet stance in humans assessed by fractional Brownian motion modelling. Neurosci Lett. 1999, 266: 157-160. 10.1016/S0304-3940(99)00272-4.CrossRefPubMed Rougier P: Influence of visual feedback on successive control mechanisms in upright quiet stance in humans assessed by fractional Brownian motion modelling. Neurosci Lett. 1999, 266: 157-160. 10.1016/S0304-3940(99)00272-4.CrossRefPubMed
29.
Zurück zum Zitat Maurer C, Mergner T, Peterka RJ: A multisensory posture control model of human upright stance. Prog Brain Res. 2003, 142: 189-201.CrossRefPubMed Maurer C, Mergner T, Peterka RJ: A multisensory posture control model of human upright stance. Prog Brain Res. 2003, 142: 189-201.CrossRefPubMed
30.
Zurück zum Zitat Cerminara NL, Edge AL, Marple-Horvat DE, Apps R: The lateral cerebellum and visuomotor control. Prog Brain Res. 2004, 148: 213-226.CrossRef Cerminara NL, Edge AL, Marple-Horvat DE, Apps R: The lateral cerebellum and visuomotor control. Prog Brain Res. 2004, 148: 213-226.CrossRef
31.
Zurück zum Zitat Buchanan JJ, Horak FB.: Voluntary control of postural equilibrium patterns. Behav Brain Res. 2003, 143 (2): 121-40. 10.1016/S0166-4328(03)00038-X.CrossRefPubMed Buchanan JJ, Horak FB.: Voluntary control of postural equilibrium patterns. Behav Brain Res. 2003, 143 (2): 121-40. 10.1016/S0166-4328(03)00038-X.CrossRefPubMed
32.
Zurück zum Zitat Wong AM, Lee MY, Kuo JK, Tang FT: The development and clinical evaluation of a standing biofeedback trainer. J Rehabil Res Dev. 1997, 34: 322-327.PubMed Wong AM, Lee MY, Kuo JK, Tang FT: The development and clinical evaluation of a standing biofeedback trainer. J Rehabil Res Dev. 1997, 34: 322-327.PubMed
Metadaten
Titel
Influence of a portable audio-biofeedback device on structural properties of postural sway
verfasst von
Marco Dozza
Lorenzo Chiari
Becky Chan
Laura Rocchi
Fay B Horak
Angelo Cappello
Publikationsdatum
01.12.2005
Verlag
BioMed Central
Erschienen in
Journal of NeuroEngineering and Rehabilitation / Ausgabe 1/2005
Elektronische ISSN: 1743-0003
DOI
https://doi.org/10.1186/1743-0003-2-13

Weitere Artikel der Ausgabe 1/2005

Journal of NeuroEngineering and Rehabilitation 1/2005 Zur Ausgabe

Neu in den Fachgebieten Neurologie und Psychiatrie

Nicht Creutzfeldt Jakob, sondern Abführtee-Vergiftung

29.05.2024 Obstipation Nachrichten

Eine ältere Frau trinkt regelmäßig Sennesblättertee gegen ihre Verstopfung. Der scheint plötzlich gut zu wirken. Auf Durchfall und Erbrechen folgt allerdings eine Hyponatriämie. Nach deren Korrektur kommt es plötzlich zu progredienten Kognitions- und Verhaltensstörungen.

Hörschwäche erhöht Demenzrisiko unabhängig von Beta-Amyloid

29.05.2024 Hörstörungen Nachrichten

Hört jemand im Alter schlecht, nimmt das Hirn- und Hippocampusvolumen besonders schnell ab, was auch mit einem beschleunigten kognitiven Abbau einhergeht. Und diese Prozesse scheinen sich unabhängig von der Amyloidablagerung zu ereignen.

Fehlerkultur in der Medizin – Offenheit zählt!

28.05.2024 Fehlerkultur Podcast

Darüber reden und aus Fehlern lernen, sollte das Motto in der Medizin lauten. Und zwar nicht nur im Sinne der Patientensicherheit. Eine negative Fehlerkultur kann auch die Behandelnden ernsthaft krank machen, warnt Prof. Dr. Reinhard Strametz. Ein Plädoyer und ein Leitfaden für den offenen Umgang mit kritischen Ereignissen in Medizin und Pflege.

„Übersichtlicher Wegweiser“: Lauterbachs umstrittener Klinik-Atlas ist online

17.05.2024 Klinik aktuell Nachrichten

Sie sei „ethisch geboten“, meint Gesundheitsminister Karl Lauterbach: mehr Transparenz über die Qualität von Klinikbehandlungen. Um sie abzubilden, lässt er gegen den Widerstand vieler Länder einen virtuellen Klinik-Atlas freischalten.