Erschienen in: Herzschrittmachertherapie + Elektrophysiologie 4/2014

01.12.2014 | Schwerpunkt

Multielectrode Pulmonary Vein Ablation Catheter (PVAC®)

Current data on results and risks

verfasst von: Dr. Gerold Mönnig, Lars Eckardt

Erschienen in: Herzschrittmachertherapie + Elektrophysiologie | Ausgabe 4/2014

Einloggen, um Zugang zu erhalten

Abstract

Electrical isolation of pulmonary veins is the cornerstone of catheter ablation for patients with symptomatic atrial fibrillation. However, uncertainty surrounds the choice of energy source in pulmonary vein isolation (PVI). Various alternative techniques such as the Pulmonary Vein Ablation Catheter (PVAC®, Medtronic Inc., Minneapolis, MN, USA) have been developed to facilitate PVI. This over-the-wire multielectrode catheter is delivering duty-cycled bipolar and unipolar radiofrequency (RF) energy at relatively low power.
PVI with this “one-shot” PVACatheter can shorten the procedure duration and lower fluoroscopy time compared to irrigated RF. It enables mapping and ablation with the same array, but fails to show signals during RF energy delivery. The effectiveness of PVAC is comparable to other technologies in randomized studies. The overall complication rate of PVAC PVI is comparable to irrigated RF and possibly slightly higher for cryoballoon PVI. Special attention has to be paid to an effective anticoagulation throughout the ablation procedure, avoidance of embolic events and pulmonary venous stenosis.
The novel catheter design of the PVAC Gold® array may improve safety by reducing embolic events through avoidance of electrode 1-to-10 interaction and by better tissue contact due to the 20° forward tilt. Although clinical data with this new array are lacking so far, the PVAC system has been shown to be a promising tool for PVI. However, prospective studies especially with the novel array are required to determine its true role for catheter ablation of atrial fibrillation in the future.
Literatur
1.
Zurück zum Zitat Camm AJ, Lip GY, De Caterina R, Savelieva I, Atar D, Hohnloser SH, Hindricks G, Kirchhof P, Bax JJ, Baumgartner H, Ceconi C, Dean V, Deaton C, Fagard R, Funck-Brentano C, Hasdai D, Hoes A, Knuuti J, Kolh P, McDonagh T, Moulin C, Popescu BA, Reiner Z, Sechtem U, Sirnes PA, Tendera M, Torbicki A, Vahanian A, Windecker S, Vardas P, Al-Attar N, Alfieri O, Angelini A, Blomstrom-Lundqvist C, Colonna P, De Sutter J, Ernst S, Goette A, Gorenek B, Hatala R, Heidbuchel H, Heldal M, Kristensen SD, Le Heuzey JY, Mavrakis H, Mont L, Filardi PP, Ponikowski P, Prendergast B, Rutten FH, Schotten U, Van Gelder IC, Verheugt FW (2012) Focused update of the ESC guidelines for the management of atrial fibrillation: an update of the 2010 ESC guidelines for the management of atrial fibrillation. Developed with the special contribution of the European heart rhythm association. Eur Heart J 33:2719–2747CrossRef Camm AJ, Lip GY, De Caterina R, Savelieva I, Atar D, Hohnloser SH, Hindricks G, Kirchhof P, Bax JJ, Baumgartner H, Ceconi C, Dean V, Deaton C, Fagard R, Funck-Brentano C, Hasdai D, Hoes A, Knuuti J, Kolh P, McDonagh T, Moulin C, Popescu BA, Reiner Z, Sechtem U, Sirnes PA, Tendera M, Torbicki A, Vahanian A, Windecker S, Vardas P, Al-Attar N, Alfieri O, Angelini A, Blomstrom-Lundqvist C, Colonna P, De Sutter J, Ernst S, Goette A, Gorenek B, Hatala R, Heidbuchel H, Heldal M, Kristensen SD, Le Heuzey JY, Mavrakis H, Mont L, Filardi PP, Ponikowski P, Prendergast B, Rutten FH, Schotten U, Van Gelder IC, Verheugt FW (2012) Focused update of the ESC guidelines for the management of atrial fibrillation: an update of the 2010 ESC guidelines for the management of atrial fibrillation. Developed with the special contribution of the European heart rhythm association. Eur Heart J 33:2719–2747CrossRef
2.
Zurück zum Zitat Gaita F, Leclercq JF, Schumacher B, Scaglione M, Toso E, Halimi F, Schade A, Froehner S, Ziegler V, Sergi D, Cesarani F, Blandino A (2011) Incidence of silent cerebral thromboembolic lesions after atrial fibrillation ablation may change according to technology used: comparison of irrigated radiofrequency, multipolar nonirrigated catheter and cryoballoon. J Cardiovasc Electrophysiol 22:961–968PubMedCrossRef Gaita F, Leclercq JF, Schumacher B, Scaglione M, Toso E, Halimi F, Schade A, Froehner S, Ziegler V, Sergi D, Cesarani F, Blandino A (2011) Incidence of silent cerebral thromboembolic lesions after atrial fibrillation ablation may change according to technology used: comparison of irrigated radiofrequency, multipolar nonirrigated catheter and cryoballoon. J Cardiovasc Electrophysiol 22:961–968PubMedCrossRef
3.
Zurück zum Zitat Herrera Siklody C, Deneke T, Hocini M, Lehrmann H, Shin DI, Miyazaki S, Henschke S, Fluegel P, Schiebeling-Romer J, Bansmann PM, Bourdias T, Dousset V, Haissaguerre M, Arentz T (2011) Incidence of asymptomatic intracranial embolic events after pulmonary vein isolation: comparison of different atrial fibrillation ablation technologies in a multicenter study. J Am Coll Cardiol 58:681–688PubMedCrossRef Herrera Siklody C, Deneke T, Hocini M, Lehrmann H, Shin DI, Miyazaki S, Henschke S, Fluegel P, Schiebeling-Romer J, Bansmann PM, Bourdias T, Dousset V, Haissaguerre M, Arentz T (2011) Incidence of asymptomatic intracranial embolic events after pulmonary vein isolation: comparison of different atrial fibrillation ablation technologies in a multicenter study. J Am Coll Cardiol 58:681–688PubMedCrossRef
4.
Zurück zum Zitat Haines DE, Stewart MT, Barka ND, Kirchhof N, Lentz LR, Reinking NM, Urban JF, Halimi F, Deneke T, Kanal E (2013) Microembolism and catheter ablation II: effects of cerebral microemboli injection in a canine model. Circ Arrhythm Electrophysiol 6:23–30PubMedCrossRef Haines DE, Stewart MT, Barka ND, Kirchhof N, Lentz LR, Reinking NM, Urban JF, Halimi F, Deneke T, Kanal E (2013) Microembolism and catheter ablation II: effects of cerebral microemboli injection in a canine model. Circ Arrhythm Electrophysiol 6:23–30PubMedCrossRef
5.
Zurück zum Zitat Haines DE, Stewart MT, Dahlberg S, Barka ND, Condie C, Fiedler GR, Kirchhof NA, Halimi F, Deneke T (2013) Microembolism and catheter ablation I: a comparison of irrigated radiofrequency and multielectrode-phased radiofrequency catheter ablation of pulmonary vein ostia. Circ Arrhythm Electrophysiol 6:16–22PubMedCrossRef Haines DE, Stewart MT, Dahlberg S, Barka ND, Condie C, Fiedler GR, Kirchhof NA, Halimi F, Deneke T (2013) Microembolism and catheter ablation I: a comparison of irrigated radiofrequency and multielectrode-phased radiofrequency catheter ablation of pulmonary vein ostia. Circ Arrhythm Electrophysiol 6:16–22PubMedCrossRef
6.
Zurück zum Zitat Verma A, Debruyne P, Nardi S, Deneke T, Degreef Y, Spitzer S, Balzer JO, Boersma L (2013) Evaluation and reduction of asymptomatic cerebral embolism in ablation of atrial fibrillation, but high prevalence of chronic silent infarction: results of the ERACE trial. Circ Arrhythm Electrophysiol 6:835–842 Verma A, Debruyne P, Nardi S, Deneke T, Degreef Y, Spitzer S, Balzer JO, Boersma L (2013) Evaluation and reduction of asymptomatic cerebral embolism in ablation of atrial fibrillation, but high prevalence of chronic silent infarction: results of the ERACE trial. Circ Arrhythm Electrophysiol 6:835–842
7.
Zurück zum Zitat Zellerhoff S, Ritter MA, Kochhauser S, Dittrich R, Kobe J, Milberg P, Korsukewitz C, Dechering DG, Pott C, Wasmer K, Leitz P, Guner F, Eckardt L, Mönnig G (2014) Modified phased radiofrequency ablation of atrial fibrillation reduces the number of cerebral microembolic signals. Europace 16:341–346 Zellerhoff S, Ritter MA, Kochhauser S, Dittrich R, Kobe J, Milberg P, Korsukewitz C, Dechering DG, Pott C, Wasmer K, Leitz P, Guner F, Eckardt L, Mönnig G (2014) Modified phased radiofrequency ablation of atrial fibrillation reduces the number of cerebral microembolic signals. Europace 16:341–346
8.
Zurück zum Zitat Wieczorek M, Hoeltgen R, Brueck M (2013) Does the number of simultaneously activated electrodes during phased RF multielectrode ablation of atrial fibrillation influence the incidence of silent cerebral microembolism? Heart Rhythm 10:953–959PubMedCrossRef Wieczorek M, Hoeltgen R, Brueck M (2013) Does the number of simultaneously activated electrodes during phased RF multielectrode ablation of atrial fibrillation influence the incidence of silent cerebral microembolism? Heart Rhythm 10:953–959PubMedCrossRef
9.
Zurück zum Zitat Compier MG, Leong DP, Marsan NA, Delgado V, Zeppenfeld K, Schalij MJ, Trines SA (2013) Duty-cycled bipolar/unipolar radiofrequency ablation for symptomatic atrial fibrillation induces significant pulmonary vein narrowing at long-term follow-up. Europace 15:690–696PubMedCrossRef Compier MG, Leong DP, Marsan NA, Delgado V, Zeppenfeld K, Schalij MJ, Trines SA (2013) Duty-cycled bipolar/unipolar radiofrequency ablation for symptomatic atrial fibrillation induces significant pulmonary vein narrowing at long-term follow-up. Europace 15:690–696PubMedCrossRef
10.
Zurück zum Zitat Dong J, Vasamreddy CR, Jayam V, Dalal D, Dickfeld T, Eldadah Z, Meininger G, Halperin HR, Berger R, Bluemke DA, Calkins H (2005) Incidence and predictors of pulmonary vein stenosis following catheter ablation of atrial fibrillation using the anatomic pulmonary vein ablation approach: results from paired magnetic resonance imaging. J Cardiovasc Electrophysiol 16:845–852PubMedCrossRef Dong J, Vasamreddy CR, Jayam V, Dalal D, Dickfeld T, Eldadah Z, Meininger G, Halperin HR, Berger R, Bluemke DA, Calkins H (2005) Incidence and predictors of pulmonary vein stenosis following catheter ablation of atrial fibrillation using the anatomic pulmonary vein ablation approach: results from paired magnetic resonance imaging. J Cardiovasc Electrophysiol 16:845–852PubMedCrossRef
11.
Zurück zum Zitat von Bary C, Dornia C, Kirchner G, Weber S, Fellner C, Nisenbaum D, Georgieva M, Stroszczynski C, Hamer OW (2013) Esophageal tissue injury following pulmonary vein isolation using the PVAC: assessment by endoscopy and magnetic resonance imaging. Pacing Clin Electrophysiol 36:477–485PubMedCrossRef von Bary C, Dornia C, Kirchner G, Weber S, Fellner C, Nisenbaum D, Georgieva M, Stroszczynski C, Hamer OW (2013) Esophageal tissue injury following pulmonary vein isolation using the PVAC: assessment by endoscopy and magnetic resonance imaging. Pacing Clin Electrophysiol 36:477–485PubMedCrossRef
12.
Zurück zum Zitat Bittner A, Mönnig G, Zellerhoff S, Pott C, Kobe J, Dechering D, Milberg P, Wasmer K, Eckardt L (2011) Randomized study comparing duty-cycled bipolar and unipolar radiofrequency with point-by-point ablation in pulmonary vein isolation. Heart Rhythm 8:1383–1390PubMedCrossRef Bittner A, Mönnig G, Zellerhoff S, Pott C, Kobe J, Dechering D, Milberg P, Wasmer K, Eckardt L (2011) Randomized study comparing duty-cycled bipolar and unipolar radiofrequency with point-by-point ablation in pulmonary vein isolation. Heart Rhythm 8:1383–1390PubMedCrossRef
13.
Zurück zum Zitat Bulava A, Hanis J, Sitek D, Osmera O, Karpianus D, Snorek M, Rehouskova K, Tousek F, Pesl L (2010) Catheter ablation for paroxysmal atrial fibrillation: a randomized comparison between multielectrode catheter and point-by-point ablation. Pacing Clin Electrophysiol 33:1039–1046PubMedCrossRef Bulava A, Hanis J, Sitek D, Osmera O, Karpianus D, Snorek M, Rehouskova K, Tousek F, Pesl L (2010) Catheter ablation for paroxysmal atrial fibrillation: a randomized comparison between multielectrode catheter and point-by-point ablation. Pacing Clin Electrophysiol 33:1039–1046PubMedCrossRef
14.
Zurück zum Zitat De Greef Y, Buysschaert I, Schwagten B, Stockman D, Tavernier R, Duytschaever M (2014) Duty-cycled multi-electrode radiofrequency vs. conventional irrigated point-by-point radiofrequency ablation for recurrent atrial fibrillation: comparative 3-year data. Europace 16:820–825 De Greef Y, Buysschaert I, Schwagten B, Stockman D, Tavernier R, Duytschaever M (2014) Duty-cycled multi-electrode radiofrequency vs. conventional irrigated point-by-point radiofrequency ablation for recurrent atrial fibrillation: comparative 3-year data. Europace 16:820–825
15.
Zurück zum Zitat Malmborg H, Lonnerholm S, Blomstrom P, Blomstrom-Lundqvist C (2013) Ablation of atrial fibrillation with cryoballoon or duty-cycled radiofrequency pulmonary vein ablation catheter: a randomized controlled study comparing the clinical outcome and safety; the AF-COR study. Europace 15:1567–1573PubMedCrossRef Malmborg H, Lonnerholm S, Blomstrom P, Blomstrom-Lundqvist C (2013) Ablation of atrial fibrillation with cryoballoon or duty-cycled radiofrequency pulmonary vein ablation catheter: a randomized controlled study comparing the clinical outcome and safety; the AF-COR study. Europace 15:1567–1573PubMedCrossRef
16.
Zurück zum Zitat Malmborg H, Christersson C, Lonnerholm S, Blomstrom-Lundqvist C (2013) Comparison of effects on coagulation and inflammatory markers using a duty-cycled bipolar and unipolar radiofrequency pulmonary vein ablation catheter vs. a cryoballoon catheter for pulmonary vein isolation. Europace 15:798–804PubMedCrossRef Malmborg H, Christersson C, Lonnerholm S, Blomstrom-Lundqvist C (2013) Comparison of effects on coagulation and inflammatory markers using a duty-cycled bipolar and unipolar radiofrequency pulmonary vein ablation catheter vs. a cryoballoon catheter for pulmonary vein isolation. Europace 15:798–804PubMedCrossRef
17.
Zurück zum Zitat Ahsan SY, Flett AS, Lambiase PD, Segal OR (2010) First report of phrenic nerve injury during pulmonary vein isolation using the ablation frontiers pulmonary vein ablation catheter. J Interv Card Electrophysiol 29:187–190PubMedCrossRef Ahsan SY, Flett AS, Lambiase PD, Segal OR (2010) First report of phrenic nerve injury during pulmonary vein isolation using the ablation frontiers pulmonary vein ablation catheter. J Interv Card Electrophysiol 29:187–190PubMedCrossRef
18.
Zurück zum Zitat Mulder AA, Wijffels MC, Wever EF, Boersma LV (2011) Pulmonary vein isolation and left atrial complex-fractionated atrial electrograms ablation for persistent atrial fibrillation with phased radio frequency energy and multi-electrode catheters: efficacy and safety during 12 months follow-up. Europace 13:1695–1702PubMedCrossRef Mulder AA, Wijffels MC, Wever EF, Boersma LV (2011) Pulmonary vein isolation and left atrial complex-fractionated atrial electrograms ablation for persistent atrial fibrillation with phased radio frequency energy and multi-electrode catheters: efficacy and safety during 12 months follow-up. Europace 13:1695–1702PubMedCrossRef
19.
Zurück zum Zitat Hummel J, Michaud G, Hoyt R, DeLurgio D, Rasekh A, Kusumoto F, Giudici M, Dan D, Tschopp D, Calkins H, Boersma L (2014) Phased RF ablation in persistent atrial fibrillation. Heart Rhythm 11:202–209PubMedCrossRef Hummel J, Michaud G, Hoyt R, DeLurgio D, Rasekh A, Kusumoto F, Giudici M, Dan D, Tschopp D, Calkins H, Boersma L (2014) Phased RF ablation in persistent atrial fibrillation. Heart Rhythm 11:202–209PubMedCrossRef
Metadaten
Titel
Multielectrode Pulmonary Vein Ablation Catheter (PVAC®)
Current data on results and risks
verfasst von
Dr. Gerold Mönnig
Lars Eckardt
Publikationsdatum
01.12.2014
Verlag
Springer Berlin Heidelberg
Erschienen in
Herzschrittmachertherapie + Elektrophysiologie / Ausgabe 4/2014
Print ISSN: 0938-7412
Elektronische ISSN: 1435-1544
DOI
https://doi.org/10.1007/s00399-014-0330-x

Weitere Artikel der Ausgabe 4/2014

Herzschrittmachertherapie + Elektrophysiologie 4/2014 Zur Ausgabe

Deutsches Herzschrittmacher- und Defibrillator-Register

Jahresbericht 2012 des Deutschen Herzschrittmacher- und Defibrillatorregisters

Update Kardiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.