Optimizing imaging quality in endourology with the Uro Dyna-CT: contrast agent dilution matters

verfasst von: M. Ritter, C. Weiss, M.-C. Rassweiler, M. S. Michel, A. Häcker

Erschienen in: World Journal of Urology | Ausgabe 5/2013

Einloggen, um Zugang zu erhalten

Abstract

Background and purpose

The implementation of the Uro Dyna-CT (Siemens Healthcare Solutions, Erlangen, Germany) with interventional 3D- and cross-sectional imaging necessitates the development of new standards for endourologic procedures such as the customized use of diluted contrast agent. Our aim was to find the ideal contrast agent dilution (CAD) for the interventional use of the Uro Dyna-CT in a standardized setting and prove the experimental findings in a clinical case.

Materials and methods

Retrograde pyelographies were performed with 10 different CADs of Imeron 300 (Bracco, Konstanz, Germany) in 10 different pig’s urinary tracts. Fluoroscopy, X-ray, 3D- and slice-image reconstruction was performed with the Uro Dyna-CT. Image quality was evaluated, blinded and randomized by 5 observers. Small plastic jars were filled with the CADs and 2 artificial stones (Plaster of paris). Images were evaluated by two observers. The ex vivo findings were transferred to a clinical setting in a complex percutaneous lithotomy procedure. Unweighted and weighted kappa coefficients were calculated to indicate the degree of observers’ agreement.

Results

Twenty percent diluted contrast agent provides the best image quality and stone detection when interventional cross-sectional imaging is considered without limitations in fluoroscopy or X-ray quality. This was proved in a percutaneous lithotomy of an obese patient.

Conclusions

Image quality of the Uro Dyna-CT can be optimized by the use of 20 % diluted contrast agent. This knowledge helps to provide high-quality 3D imaging in the endourological operation room.
Literatur
1.
Zurück zum Zitat Akpek S, Brunner T, Benndorf G, Strother C (2005) Three-dimensional imaging and cone beam volume CT in C-arm angiography with flat panel detector. Diagn Interv Radiol 11(1):10–13PubMed Akpek S, Brunner T, Benndorf G, Strother C (2005) Three-dimensional imaging and cone beam volume CT in C-arm angiography with flat panel detector. Diagn Interv Radiol 11(1):10–13PubMed
2.
Zurück zum Zitat Siewerdsen JH, Moseley DJ, Burch S et al (2005) Volume CT with a flat-panel detector on a mobile, isocentric C-arm: pre-clinical investigation in guidance of minimally invasive surgery. Med Phys 32(1):241–254PubMedCrossRef Siewerdsen JH, Moseley DJ, Burch S et al (2005) Volume CT with a flat-panel detector on a mobile, isocentric C-arm: pre-clinical investigation in guidance of minimally invasive surgery. Med Phys 32(1):241–254PubMedCrossRef
3.
Zurück zum Zitat Michel MS, Ritter M, Schonberg S, Hacker A, Weisser G (2012) The urological Dyna-CT: urological cross-sectional imaging on a newly developed urological intervention table. Urologe A [Epub ahead of print] Michel MS, Ritter M, Schonberg S, Hacker A, Weisser G (2012) The urological Dyna-CT: urological cross-sectional imaging on a newly developed urological intervention table. Urologe A [Epub ahead of print]
4.
Zurück zum Zitat Ritter M, Rassweiler MC, Hacker A, Michel MS (2012) Laser-guided percutaneous kidney access with the Uro Dyna-CT: first experience of three-dimensional puncture planning with an ex vivo model. World J Urol [Epub ahead of print] Ritter M, Rassweiler MC, Hacker A, Michel MS (2012) Laser-guided percutaneous kidney access with the Uro Dyna-CT: first experience of three-dimensional puncture planning with an ex vivo model. World J Urol [Epub ahead of print]
5.
Zurück zum Zitat Fleiss J (1971) Measuring nominal scale agreement among many readers. Psychol Bull 76:378–382CrossRef Fleiss J (1971) Measuring nominal scale agreement among many readers. Psychol Bull 76:378–382CrossRef
6.
Zurück zum Zitat Schorer J, Weiss C (2007) A weighted kappa-coefficient for three observers as a measure of expert ratings on characteristics in handball throwing patterns. Meas Phys Educ Exerc Sci 11:177–187CrossRef Schorer J, Weiss C (2007) A weighted kappa-coefficient for three observers as a measure of expert ratings on characteristics in handball throwing patterns. Meas Phys Educ Exerc Sci 11:177–187CrossRef
7.
Zurück zum Zitat Landis JR, Koch GG (1977) The measurement of observer agreement for categorical data. Biometrics 33(1):159–174PubMedCrossRef Landis JR, Koch GG (1977) The measurement of observer agreement for categorical data. Biometrics 33(1):159–174PubMedCrossRef
8.
Zurück zum Zitat Patel U, Walkden RM, Ghani KR, Anson K (2009) Three-dimensional CT pyelography for planning of percutaneous nephrostolithotomy: accuracy of stone measurement, stone depiction and pelvicalyceal reconstruction. Eur Radiol 19(5):1280–1288PubMedCrossRef Patel U, Walkden RM, Ghani KR, Anson K (2009) Three-dimensional CT pyelography for planning of percutaneous nephrostolithotomy: accuracy of stone measurement, stone depiction and pelvicalyceal reconstruction. Eur Radiol 19(5):1280–1288PubMedCrossRef
9.
Zurück zum Zitat Campbell W (2006) Urology. Saunders, p 170 Campbell W (2006) Urology. Saunders, p 170
10.
Zurück zum Zitat de Petriconi R (2007) Urologische Röntgendiagnostik. In: Jocham D, Miller K (eds) Praxis der Urologie. Thieme, Stuttgart, pp 132–135 de Petriconi R (2007) Urologische Röntgendiagnostik. In: Jocham D, Miller K (eds) Praxis der Urologie. Thieme, Stuttgart, pp 132–135
11.
Zurück zum Zitat Palmer BW, Ramji FG, Snyder CT, Hemphill M, Kropp BP, Frimberger D (2011) Voiding cystourethrogram—are our protocols the same? J Urol 186(4 Suppl):1668–1671PubMedCrossRef Palmer BW, Ramji FG, Snyder CT, Hemphill M, Kropp BP, Frimberger D (2011) Voiding cystourethrogram—are our protocols the same? J Urol 186(4 Suppl):1668–1671PubMedCrossRef
12.
Zurück zum Zitat Vlahos L, Gouliamos A, Clauss W et al (1992) Gd-DTPA as an intestinal contrast agent for MR imaging of the lower abdomen: phase III clinical trial. Gastrointest Radiol 17(4):300–304PubMedCrossRef Vlahos L, Gouliamos A, Clauss W et al (1992) Gd-DTPA as an intestinal contrast agent for MR imaging of the lower abdomen: phase III clinical trial. Gastrointest Radiol 17(4):300–304PubMedCrossRef
13.
Zurück zum Zitat Namba K, Niimi Y, Song JK, Berenstein A (2009) Use of Dyna-CT angiography in neuroendovascular decision-making. A case report. Interv Neuroradiol J Perither Neuroradiol Surg Proced Relat Neurosci 15(1):67–72 Namba K, Niimi Y, Song JK, Berenstein A (2009) Use of Dyna-CT angiography in neuroendovascular decision-making. A case report. Interv Neuroradiol J Perither Neuroradiol Surg Proced Relat Neurosci 15(1):67–72
14.
Zurück zum Zitat Eide KR, Odegard A, Myhre HO, Haraldseth O (2007) Initial observations of endovascular aneurysm repair using Dyna-CT. J Endovasc Ther 14(1):50–53PubMedCrossRef Eide KR, Odegard A, Myhre HO, Haraldseth O (2007) Initial observations of endovascular aneurysm repair using Dyna-CT. J Endovasc Ther 14(1):50–53PubMedCrossRef
15.
Zurück zum Zitat Jung SC, Kim SH, Cho JY (2011) A comparison of the use of contrast media with different iodine concentrations for multidetector CT of the kidney. Korean J Radiol Off J Korean Radiol Soc 12(6):714–721CrossRef Jung SC, Kim SH, Cho JY (2011) A comparison of the use of contrast media with different iodine concentrations for multidetector CT of the kidney. Korean J Radiol Off J Korean Radiol Soc 12(6):714–721CrossRef
16.
Zurück zum Zitat McKeown C, Dempsey MF, Gillen G, Paterson C (2012) Quantitative analysis shows that contrast medium in positron emission tomography/computed tomography may cause significant artefacts. Nuclear medicine communications [Epub ahead of print] McKeown C, Dempsey MF, Gillen G, Paterson C (2012) Quantitative analysis shows that contrast medium in positron emission tomography/computed tomography may cause significant artefacts. Nuclear medicine communications [Epub ahead of print]
17.
Zurück zum Zitat Heiland S, Erb G, Ziegler S, Krix M (2010) Where contrast agent concentration really matters—a comparison of CT and MRI. Invest Radiol 45(9):529–537PubMedCrossRef Heiland S, Erb G, Ziegler S, Krix M (2010) Where contrast agent concentration really matters—a comparison of CT and MRI. Invest Radiol 45(9):529–537PubMedCrossRef
Metadaten
Titel
Optimizing imaging quality in endourology with the Uro Dyna-CT: contrast agent dilution matters
verfasst von
M. Ritter
C. Weiss
M.-C. Rassweiler
M. S. Michel
A. Häcker
Publikationsdatum
01.10.2013
Verlag
Springer Berlin Heidelberg
Erschienen in
World Journal of Urology / Ausgabe 5/2013
Print ISSN: 0724-4983
Elektronische ISSN: 1433-8726
DOI
https://doi.org/10.1007/s00345-012-0903-4

Weitere Artikel der Ausgabe 5/2013

World Journal of Urology 5/2013 Zur Ausgabe

ADT zur Radiatio nach Prostatektomie: Wenn, dann wohl länger

24.05.2024 Prostatakarzinom Nachrichten

Welchen Nutzen es trägt, wenn die Strahlentherapie nach radikaler Prostatektomie um eine Androgendeprivation ergänzt wird, hat die RADICALS-HD-Studie untersucht. Nun liegen die Ergebnisse vor. Sie sprechen für länger dauernden Hormonentzug.

„Überwältigende“ Evidenz für Tripeltherapie beim metastasierten Prostata-Ca.

22.05.2024 Prostatakarzinom Nachrichten

Patienten mit metastasiertem hormonsensitivem Prostatakarzinom sollten nicht mehr mit einer alleinigen Androgendeprivationstherapie (ADT) behandelt werden, mahnt ein US-Team nach Sichtung der aktuellen Datenlage. Mit einer Tripeltherapie haben die Betroffenen offenbar die besten Überlebenschancen.

„Übersichtlicher Wegweiser“: Lauterbachs umstrittener Klinik-Atlas ist online

17.05.2024 Klinik aktuell Nachrichten

Sie sei „ethisch geboten“, meint Gesundheitsminister Karl Lauterbach: mehr Transparenz über die Qualität von Klinikbehandlungen. Um sie abzubilden, lässt er gegen den Widerstand vieler Länder einen virtuellen Klinik-Atlas freischalten.

Alphablocker schützt vor Miktionsproblemen nach der Biopsie

16.05.2024 alpha-1-Rezeptorantagonisten Nachrichten

Nach einer Prostatabiopsie treten häufig Probleme beim Wasserlassen auf. Ob sich das durch den periinterventionellen Einsatz von Alphablockern verhindern lässt, haben australische Mediziner im Zuge einer Metaanalyse untersucht.

Update Urologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.