Relocatable fixation systems in intracranial stereotactic radiotherapy

Accuracy of serial CT scans and patient acceptance in a randomized design

verfasst von: A. Theelen, J. Martens, BSc, G. Bosmans, PhD, R. Houben, MSc, J.J. Jager, MD PhD, I. Rutten, MD, P. Lambin, MD PhD, A.W. Minken, PhD, Dr. B.G. Baumert, MD PhD

Erschienen in: Strahlentherapie und Onkologie | Ausgabe 1/2012

Einloggen, um Zugang zu erhalten

Abstract

Purpose

The goal was to provide a quantitative evaluation of the accuracy of three different fixation systems for stereotactic radiotherapy and to evaluate patients’ acceptance for all fixations.

Methods

A total of 16 consecutive patients with brain tumours undergoing fractionated stereotactic radiotherapy (SCRT) were enrolled after informed consent (Clinical trials.gov: NCT00181350). Fixation systems evaluated were the BrainLAB® mask, with and without custom made bite-block (fixations S and A) and a homemade neck support with bite-block (fixation B) based on the BrainLAB® frame. The sequence of measurements was evaluated in a randomized manner with a cross-over design and patients’ acceptance by a questionnaire.

Results

The mean three-dimensional (3D) displacement and standard deviations were 1.16 ± 0.68 mm for fixation S, 1.92 ± 1.28 and 1.70 ± 0.83 mm for fixations A and B, respectively. There was a significant improvement of the overall alignment (3D vector) when using the standard fixation instead of fixation A or B in the craniocaudal direction (p = 0.037). Rotational deviations were significantly less for the standard fixation S in relation to fixations A (p = 0.005) and B (p = 0.03). EPI imaging with off-line correction further improved reproducibility. Five out of 8 patients preferred the neck support with the bite-block.

Conclusion

The mask fixation system in conjunction with a bite-block is the most accurate fixation for SCRT reducing craniocaudal and rotational movements. Patients favoured the more comfortable but less accurate neck support. To optimize the accuracy of SCRT, additional regular portal imaging is warranted.
Literatur
1.
Zurück zum Zitat Alheit H, Dornfeld S, Dawel M et al (2001) Patient position reproducibility in fractionated stereotactically guided conformal radiotherapy using the BrainLAB® Mask system. Strahlenther Onkol 177:264–268PubMedCrossRef Alheit H, Dornfeld S, Dawel M et al (2001) Patient position reproducibility in fractionated stereotactically guided conformal radiotherapy using the BrainLAB® Mask system. Strahlenther Onkol 177:264–268PubMedCrossRef
2.
Zurück zum Zitat Ali I, Tubbs J, Hibbitts K et al (2010) Evaluation of the setup accuracy of a stereotactic radiotherapy head immobilization mask system using kV on-board imaging. J Appl Clin Med Phys 11:3192PubMed Ali I, Tubbs J, Hibbitts K et al (2010) Evaluation of the setup accuracy of a stereotactic radiotherapy head immobilization mask system using kV on-board imaging. J Appl Clin Med Phys 11:3192PubMed
3.
Zurück zum Zitat Astner ST, Theodorou M, Dobrei-Ciuchendea M et al (2010) Tumor shrinkage assessed by volumetric MRI in the long-term follow-up after stereotactic radiotherapy of meningiomas. Strahlenther Onkol 186:423–429PubMedCrossRef Astner ST, Theodorou M, Dobrei-Ciuchendea M et al (2010) Tumor shrinkage assessed by volumetric MRI in the long-term follow-up after stereotactic radiotherapy of meningiomas. Strahlenther Onkol 186:423–429PubMedCrossRef
4.
Zurück zum Zitat Balducci M, Apicella G, Manfrida S et al (2010) Single-arm phase II study of conformal radiation therapy and temozolomide plus fractionated stereotactic conformal boost in high-grade gliomas. Strahlenther Onkol 186:558–564PubMedCrossRef Balducci M, Apicella G, Manfrida S et al (2010) Single-arm phase II study of conformal radiation therapy and temozolomide plus fractionated stereotactic conformal boost in high-grade gliomas. Strahlenther Onkol 186:558–564PubMedCrossRef
5.
Zurück zum Zitat Baumert BG, Egli P, Studer S et al (2005) Repositioning accuracy of fractionated stereotactic irradiation: assessment of isocentre alignment for different dental fixations by using sequential CT scanning. Radiother Oncol 74:61–66PubMedCrossRef Baumert BG, Egli P, Studer S et al (2005) Repositioning accuracy of fractionated stereotactic irradiation: assessment of isocentre alignment for different dental fixations by using sequential CT scanning. Radiother Oncol 74:61–66PubMedCrossRef
6.
Zurück zum Zitat Bednarz G, Machtay M, Werner-Wasik M et al (2009) Report on a randomized trial comparing two forms of immobilization of the head for fractionated stereotactic radiotherapy. Med Phys 36:12–17PubMedCrossRef Bednarz G, Machtay M, Werner-Wasik M et al (2009) Report on a randomized trial comparing two forms of immobilization of the head for fractionated stereotactic radiotherapy. Med Phys 36:12–17PubMedCrossRef
7.
Zurück zum Zitat Bel A, Herk M van, Bartelink H, Lebesque JV (1993) A verification procedure to improve patient set-up accuracy using portal images. Radiother Oncol 29:253–260PubMedCrossRef Bel A, Herk M van, Bartelink H, Lebesque JV (1993) A verification procedure to improve patient set-up accuracy using portal images. Radiother Oncol 29:253–260PubMedCrossRef
8.
Zurück zum Zitat Brada M, Bidmead M (2003) Geometric uncertainties in radiotherapy of the brain. In: Geometric uncertainties in radiotherapy. Defining the planning target volume. The British Institute of Radiology, London, pp 109–126 Brada M, Bidmead M (2003) Geometric uncertainties in radiotherapy of the brain. In: Geometric uncertainties in radiotherapy. Defining the planning target volume. The British Institute of Radiology, London, pp 109–126
9.
Zurück zum Zitat D’Agostino GR, Autorino R, Pompucci A et al (2011) Whole-brain radiotherapy combined with surgery or stereotactic radiotherapy in patients with brain oligometastases: long-term analysis. Strahlenther Onkol 187:421–425CrossRef D’Agostino GR, Autorino R, Pompucci A et al (2011) Whole-brain radiotherapy combined with surgery or stereotactic radiotherapy in patients with brain oligometastases: long-term analysis. Strahlenther Onkol 187:421–425CrossRef
10.
Zurück zum Zitat Ezzell LC, Hansen EK, Quivey JM, Xia P (2007) Detection of treatment setup errors between tow CT scans for patients with head and neck cancer. Med Phys 34:3233–3242PubMedCrossRef Ezzell LC, Hansen EK, Quivey JM, Xia P (2007) Detection of treatment setup errors between tow CT scans for patients with head and neck cancer. Med Phys 34:3233–3242PubMedCrossRef
11.
Zurück zum Zitat Fokas E, Wacker U, Gross MW et al (2009) Hypofractionated stereotactic reirradiation of recurrent glioblastomas. A beneficial treatment option after high-dose radiotherapy? Strahlenther Onkol 185:235–240PubMedCrossRef Fokas E, Wacker U, Gross MW et al (2009) Hypofractionated stereotactic reirradiation of recurrent glioblastomas. A beneficial treatment option after high-dose radiotherapy? Strahlenther Onkol 185:235–240PubMedCrossRef
12.
Zurück zum Zitat Hamilton RJ, Kuchnir FT, Pelizzari CA et al (1996) Repositioning accuracy of a noninvasive head fixation system for stereotactic radiotherapy. Med Phys 23:1909–1917PubMedCrossRef Hamilton RJ, Kuchnir FT, Pelizzari CA et al (1996) Repositioning accuracy of a noninvasive head fixation system for stereotactic radiotherapy. Med Phys 23:1909–1917PubMedCrossRef
13.
Zurück zum Zitat Henzel M, Hamm K, Sitter H et al (2009) Comparison of stereotactic radiosurgery and fractionated stereotactic radiotherapy of acoustic neurinomas according to 3-D tumor volume shrinkage and quality of life. Strahlenther Onkol 185:567–573PubMedCrossRef Henzel M, Hamm K, Sitter H et al (2009) Comparison of stereotactic radiosurgery and fractionated stereotactic radiotherapy of acoustic neurinomas according to 3-D tumor volume shrinkage and quality of life. Strahlenther Onkol 185:567–573PubMedCrossRef
14.
Zurück zum Zitat Kalapurakal JA, Ilahi Z, Kepka AG et al (2001) Repositioning accuracy with the Laitinen frame for fractionated stereotactic radiation therapy in adult and pediatric brain tumours: preliminary report. Radiology 2218:157–161 Kalapurakal JA, Ilahi Z, Kepka AG et al (2001) Repositioning accuracy with the Laitinen frame for fractionated stereotactic radiation therapy in adult and pediatric brain tumours: preliminary report. Radiology 2218:157–161
15.
Zurück zum Zitat Karger CP, Jäkel O, Debus J et al (2001) Three-dimensional accuracy and interfractional reproducibility of patient fixation and positioning using a stereotactic head mask system. Int J Radiat Oncol Biol Phys 49:1493–1504PubMedCrossRef Karger CP, Jäkel O, Debus J et al (2001) Three-dimensional accuracy and interfractional reproducibility of patient fixation and positioning using a stereotactic head mask system. Int J Radiat Oncol Biol Phys 49:1493–1504PubMedCrossRef
16.
Zurück zum Zitat Kooy HM, Dunbar SF, Tarbell NJ et al (1994) Adaptation and verification of the relocatable Gill-Thomas-Cosman frame in stereotactic radiotherapy. Int J Radiat Oncol Biol Phys 30:685–691PubMed Kooy HM, Dunbar SF, Tarbell NJ et al (1994) Adaptation and verification of the relocatable Gill-Thomas-Cosman frame in stereotactic radiotherapy. Int J Radiat Oncol Biol Phys 30:685–691PubMed
17.
Zurück zum Zitat Kortmann RD, Becker G, Perelmouter J et al (1999) Geometric accuracy of field alignment in fractionated stereotactic conformal radiotherapy of brain tumors. Int J Radiat Oncol Biol Phys 43:921–926PubMedCrossRef Kortmann RD, Becker G, Perelmouter J et al (1999) Geometric accuracy of field alignment in fractionated stereotactic conformal radiotherapy of brain tumors. Int J Radiat Oncol Biol Phys 43:921–926PubMedCrossRef
18.
Zurück zum Zitat Kumar S, Burke K, Nalder C et al (2005) Treatment accuracy of fractionated stereotactic radiotherapy. Radiother Oncol 74:53–59PubMedCrossRef Kumar S, Burke K, Nalder C et al (2005) Treatment accuracy of fractionated stereotactic radiotherapy. Radiother Oncol 74:53–59PubMedCrossRef
19.
Zurück zum Zitat Lopatta E, Liesenfeld SM, Bank P et al (2003) Verbesserte Lagerungsreproduzierbarkeit eines nichtinvasiven Hochpräzisions- maskensystems in der stereotaktischen Radiotherapie durch eine integrierte Kieferfixierung. Strahlenther Onkol 179:571–575PubMedCrossRef Lopatta E, Liesenfeld SM, Bank P et al (2003) Verbesserte Lagerungsreproduzierbarkeit eines nichtinvasiven Hochpräzisions- maskensystems in der stereotaktischen Radiotherapie durch eine integrierte Kieferfixierung. Strahlenther Onkol 179:571–575PubMedCrossRef
20.
Zurück zum Zitat Masi L, Casamassima F, Polli C et al (2008) Cone beam CT image guidance for intracranial stereotactic treatments: comparison with a frame guided set-up. Int J Radiat Oncol Biol Phys 71:926–933PubMedCrossRef Masi L, Casamassima F, Polli C et al (2008) Cone beam CT image guidance for intracranial stereotactic treatments: comparison with a frame guided set-up. Int J Radiat Oncol Biol Phys 71:926–933PubMedCrossRef
21.
Zurück zum Zitat Minniti G, Valeriani M, Clarke E et al (2010) Fractionated stereotactic radiotherapy for skull base tumors: analysis of treatment accuracy using a stereotactic mask fixation system. Radiat Oncol 13(5):1CrossRef Minniti G, Valeriani M, Clarke E et al (2010) Fractionated stereotactic radiotherapy for skull base tumors: analysis of treatment accuracy using a stereotactic mask fixation system. Radiat Oncol 13(5):1CrossRef
22.
Zurück zum Zitat Miranpuri AS, Tome WA, Paliwal BR et al (2001) Assessment of patient-independent intrinsic error for a noninvasive frame for fractionated stereotactic radiotherapy. Int J Cancer 20:320–325CrossRef Miranpuri AS, Tome WA, Paliwal BR et al (2001) Assessment of patient-independent intrinsic error for a noninvasive frame for fractionated stereotactic radiotherapy. Int J Cancer 20:320–325CrossRef
23.
Zurück zum Zitat Peng LC, Kahler D, Samant S et al (2010) Quality assessment of frameless fractionated stereotactic radiotherapy using cone beam computed tomography. Int J Radiat Oncol Biol Phys 78:1586–1593PubMedCrossRef Peng LC, Kahler D, Samant S et al (2010) Quality assessment of frameless fractionated stereotactic radiotherapy using cone beam computed tomography. Int J Radiat Oncol Biol Phys 78:1586–1593PubMedCrossRef
24.
Zurück zum Zitat Rosenthal ST, Gall KP, Jackson M, Thornton AF (1995) A precision cranial immobilization system for conformal stereotactic fractionated radiation therapy. Int J Radiat Oncol Biol Phys 35:1239–1245 Rosenthal ST, Gall KP, Jackson M, Thornton AF (1995) A precision cranial immobilization system for conformal stereotactic fractionated radiation therapy. Int J Radiat Oncol Biol Phys 35:1239–1245
25.
Zurück zum Zitat Salter BJ, Fuss M, Vollmer DG et al (2001) The talon removable head frame system for stereotactic radiosurgery/radiotherapy: measurement of the repositioning accuracy. Int J Radiat Oncol Biol Phys 51:555–562PubMedCrossRef Salter BJ, Fuss M, Vollmer DG et al (2001) The talon removable head frame system for stereotactic radiosurgery/radiotherapy: measurement of the repositioning accuracy. Int J Radiat Oncol Biol Phys 51:555–562PubMedCrossRef
26.
Zurück zum Zitat Scorsetti M, Mancosu P, Navarria P et al (2011) Stereotactic body radiation therapy (SBRT) for adrenal metastases: a feasibility study of advanced techniques with modulated photons and protons. Strahlenther Onkol 187:238–244PubMedCrossRef Scorsetti M, Mancosu P, Navarria P et al (2011) Stereotactic body radiation therapy (SBRT) for adrenal metastases: a feasibility study of advanced techniques with modulated photons and protons. Strahlenther Onkol 187:238–244PubMedCrossRef
27.
Zurück zum Zitat Herk M van, Remeijer P, Rasch C, Lebesque JV (2000) The probability of correct target dosage: dose-population histograms for deriving treatment margins in radiotherapy. Int J Radiat Oncol Biol Phys 47:1121–1135PubMedCrossRef Herk M van, Remeijer P, Rasch C, Lebesque JV (2000) The probability of correct target dosage: dose-population histograms for deriving treatment margins in radiotherapy. Int J Radiat Oncol Biol Phys 47:1121–1135PubMedCrossRef
28.
Zurück zum Zitat Santvoort J van, Wiggenraad R, Bos P (2008) Positioning accuracy in stereotactic radiotherapy using a mask system with added vacuum mouth piece and stereoscopic X-ray positioning. Int J Radiat Oncol Biol Phys 72:261–267PubMedCrossRef Santvoort J van, Wiggenraad R, Bos P (2008) Positioning accuracy in stereotactic radiotherapy using a mask system with added vacuum mouth piece and stereoscopic X-ray positioning. Int J Radiat Oncol Biol Phys 72:261–267PubMedCrossRef
29.
Zurück zum Zitat Wiehle R, Koth HJ, Nanko N et al (2009) On the accuracy of isocenter verification with kV imaging in stereotactic radiosurgery. Strahlenther Onkol 185:325–330PubMedCrossRef Wiehle R, Koth HJ, Nanko N et al (2009) On the accuracy of isocenter verification with kV imaging in stereotactic radiosurgery. Strahlenther Onkol 185:325–330PubMedCrossRef
30.
Zurück zum Zitat Willner J, Flentje M, Bratengeier K (1997) CT simulation in stereotactic brain radiotherapy-analysis of isocenter reproducibility with mask fixation. Radiother Oncol 45:83–88PubMedCrossRef Willner J, Flentje M, Bratengeier K (1997) CT simulation in stereotactic brain radiotherapy-analysis of isocenter reproducibility with mask fixation. Radiother Oncol 45:83–88PubMedCrossRef
31.
Zurück zum Zitat Wong VY, Tung SY, Leung TW, Ho KH (2003) verification of isocentre relocatability using stereotactic mask fixation system. Clin Oncol (R Coll Radiol) 15:280–287 Wong VY, Tung SY, Leung TW, Ho KH (2003) verification of isocentre relocatability using stereotactic mask fixation system. Clin Oncol (R Coll Radiol) 15:280–287
Metadaten
Titel
Relocatable fixation systems in intracranial stereotactic radiotherapy
Accuracy of serial CT scans and patient acceptance in a randomized design
verfasst von
A. Theelen
J. Martens, BSc
G. Bosmans, PhD
R. Houben, MSc
J.J. Jager, MD PhD
I. Rutten, MD
P. Lambin, MD PhD
A.W. Minken, PhD
Dr. B.G. Baumert, MD PhD
Publikationsdatum
01.01.2012
Verlag
Urban and Vogel
Erschienen in
Strahlentherapie und Onkologie / Ausgabe 1/2012
Print ISSN: 0179-7158
Elektronische ISSN: 1439-099X
DOI
https://doi.org/10.1007/s00066-011-0018-7

Weitere Artikel der Ausgabe 1/2012

Strahlentherapie und Onkologie 1/2012 Zur Ausgabe

Alphablocker schützt vor Miktionsproblemen nach der Biopsie

16.05.2024 alpha-1-Rezeptorantagonisten Nachrichten

Nach einer Prostatabiopsie treten häufig Probleme beim Wasserlassen auf. Ob sich das durch den periinterventionellen Einsatz von Alphablockern verhindern lässt, haben australische Mediziner im Zuge einer Metaanalyse untersucht.

Mammakarzinom: Senken Statine das krebsbedingte Sterberisiko?

15.05.2024 Mammakarzinom Nachrichten

Frauen mit lokalem oder metastasiertem Brustkrebs, die Statine einnehmen, haben eine niedrigere krebsspezifische Mortalität als Patientinnen, die dies nicht tun, legen neue Daten aus den USA nahe.

Labor, CT-Anthropometrie zeigen Risiko für Pankreaskrebs

13.05.2024 Pankreaskarzinom Nachrichten

Gerade bei aggressiven Malignomen wie dem duktalen Adenokarzinom des Pankreas könnte Früherkennung die Therapiechancen verbessern. Noch jedoch klafft hier eine Lücke. Ein Studienteam hat einen Weg gesucht, sie zu schließen.

Viel pflanzliche Nahrung, seltener Prostata-Ca.-Progression

12.05.2024 Prostatakarzinom Nachrichten

Ein hoher Anteil pflanzlicher Nahrung trägt möglicherweise dazu bei, das Progressionsrisiko von Männern mit Prostatakarzinomen zu senken. In einer US-Studie war das Risiko bei ausgeprägter pflanzlicher Ernährung in etwa halbiert.

Update Onkologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.