The 2012 Mark Coventry Award: A Retrieval Analysis of High Flexion versus Posterior-stabilized Tibial Inserts

verfasst von: Nicholas R. Paterson, BScH, Matthew G. Teeter, BScH, PhD, Steven J. MacDonald, MD, FRCSC, Richard W. McCalden, MD, MPhil(Edin), FRCSC, Douglas D. R. Naudie, MD, FRCSC

Erschienen in: Clinical Orthopaedics and Related Research® | Ausgabe 1/2013

Einloggen, um Zugang zu erhalten

Abstract

Background

High flexion (HF) implants were introduced to increase ROM and patient satisfaction, but design changes to the implant potentially have deleterious effects on polyethylene wear. It is unclear whether the HF implants affect wear.

Questions/purposes

We therefore examined whether the design changes between HF and posterior-stabilized (PS) tibial inserts would affect overall damage or damage on their articular surface, backside, and tibial post and whether flexion angle achieved related to damage.

Methods

We matched 20 retrieved HF inserts to 20 retrieved PS inserts from the same implant system on the basis of duration of implantation, body mass index, and age. Inserts were divided into 16 zones and a microscopic analysis of surface damage was carried out. Five inserts were scanned using micro-CT to further quantify instances of severe post notching. We determined overall damage with a scoring system.

Results

We found greater backside and post damage in the HF group but no difference in the articular surface or overall damage scores. Backside and post damage scores correlated to flexion angle in the HF group. There was no flexion/damage correlation in the PS group. Notch depths around the post in both groups ranged from 0.6 to 1.9 mm.

Conclusions

HF inserts are more susceptible to post damage, possibly as a result of higher contact stresses from greater flexion. The increased backside damage was unexpected because the two groups have the same tibial component, locking mechanism, and sterilization method.

Clinical Relevance

The introduction of a highly crosslinked HF insert will require close scrutiny as a result of the potential for post damage demonstrated in this series.
Literatur
1.
Zurück zum Zitat Angelo FD, Marcolli D, Bulgheroni P, Murena L, Congiu T, Cherubino P. Two stage fracture of a polyethylene post in a 9-year-old posterior-stabilized knee prosthesis: a case report. J Med Case Reports. 2010;4:2–9.CrossRef Angelo FD, Marcolli D, Bulgheroni P, Murena L, Congiu T, Cherubino P. Two stage fracture of a polyethylene post in a 9-year-old posterior-stabilized knee prosthesis: a case report. J Med Case Reports. 2010;4:2–9.CrossRef
2.
Zurück zum Zitat Bal BS, Greenberg D, Li S, Mauerhan DR, Schultz L, Cherry K. Tibial post failures in a condylar posterior cruciate substituting total knee arthroplasty. J Arthroplasty. 2008;23:650–655.PubMedCrossRef Bal BS, Greenberg D, Li S, Mauerhan DR, Schultz L, Cherry K. Tibial post failures in a condylar posterior cruciate substituting total knee arthroplasty. J Arthroplasty. 2008;23:650–655.PubMedCrossRef
3.
Zurück zum Zitat Bourne RB, Chesworth BM, Davis AM, Mahomed NN, Charron KDJ. Patient satisfaction after total knee arthroplasty: who is satisfied and who is not? Clin Orthop Relat Res. 2010;468:57–63.PubMedCrossRef Bourne RB, Chesworth BM, Davis AM, Mahomed NN, Charron KDJ. Patient satisfaction after total knee arthroplasty: who is satisfied and who is not? Clin Orthop Relat Res. 2010;468:57–63.PubMedCrossRef
4.
Zurück zum Zitat Chiu Y, Chen W, Huang C. Case report fracture of the polyethylene tibial post in a NexGen posterior-stabilized knee prosthesis. J Bone Joint Surg Am. 2004;19:1045–1049. Chiu Y, Chen W, Huang C. Case report fracture of the polyethylene tibial post in a NexGen posterior-stabilized knee prosthesis. J Bone Joint Surg Am. 2004;19:1045–1049.
5.
Zurück zum Zitat Crowninshield RD, Wimmer MA, Jacobs JJ, Rosenberg AG. Clinical performance of contemporary tibial polyethylene components. J Arthroplasty. 2006;21:754–761.PubMedCrossRef Crowninshield RD, Wimmer MA, Jacobs JJ, Rosenberg AG. Clinical performance of contemporary tibial polyethylene components. J Arthroplasty. 2006;21:754–761.PubMedCrossRef
6.
Zurück zum Zitat Dolan MM, Kelly NH, Nguyen JT, Wright TM, Haas SB. Implant design influences tibial post wear damage in posterior-stabilized knees. Clin Orthop Relat Res. 2011;469:160–167.PubMedCrossRef Dolan MM, Kelly NH, Nguyen JT, Wright TM, Haas SB. Implant design influences tibial post wear damage in posterior-stabilized knees. Clin Orthop Relat Res. 2011;469:160–167.PubMedCrossRef
7.
Zurück zum Zitat Engh GA, Zimmerman RL, Parks NL, Engh CA. Analysis of wear in retrieved mobile and fixed bearing knee inserts. J Arthroplasty. 2009;24:28–32.PubMedCrossRef Engh GA, Zimmerman RL, Parks NL, Engh CA. Analysis of wear in retrieved mobile and fixed bearing knee inserts. J Arthroplasty. 2009;24:28–32.PubMedCrossRef
8.
Zurück zum Zitat Furman BD, Lipman J, Kligman M, Wright TM, Haas SB. Tibial post wear in posterior-stabilized knee replacements is design-dependent. Clin Orthop Relat Res. 2008;466:2650–2655.PubMedCrossRef Furman BD, Lipman J, Kligman M, Wright TM, Haas SB. Tibial post wear in posterior-stabilized knee replacements is design-dependent. Clin Orthop Relat Res. 2008;466:2650–2655.PubMedCrossRef
9.
Zurück zum Zitat Hamai S, Miura H, Matsuda S, Shimoto T, Higaki H, Iwamoto Y. Contact stress at the anterior aspect of the tibial post in posterior-stabilized total knee replacement. J Bone Joint Surg Am. 2011;92:1765–1773.CrossRef Hamai S, Miura H, Matsuda S, Shimoto T, Higaki H, Iwamoto Y. Contact stress at the anterior aspect of the tibial post in posterior-stabilized total knee replacement. J Bone Joint Surg Am. 2011;92:1765–1773.CrossRef
10.
Zurück zum Zitat Hanson GR, Suggs JF, Kwon Y, Freiberg AA, Li G. In vivo anterior tibial post contact after posterior stabilizing total knee arthroplasty. J Orthop Res. 2007;25:1447–1453.PubMedCrossRef Hanson GR, Suggs JF, Kwon Y, Freiberg AA, Li G. In vivo anterior tibial post contact after posterior stabilizing total knee arthroplasty. J Orthop Res. 2007;25:1447–1453.PubMedCrossRef
11.
Zurück zum Zitat Heyse TJ, Chen DX, Kelly N, Boettner F, Wright TM, Haas SB. The knee matched-pair total knee arthroplasty retrieval analysis: oxidized zirconium vs CoCrMo. Knee. 2011;18:448–452.PubMedCrossRef Heyse TJ, Chen DX, Kelly N, Boettner F, Wright TM, Haas SB. The knee matched-pair total knee arthroplasty retrieval analysis: oxidized zirconium vs CoCrMo. Knee. 2011;18:448–452.PubMedCrossRef
12.
Zurück zum Zitat Heyse TJ, Davis J, Haas SB, Chen DX, Wright TM, Laskin RS. Retrieval analysis of femoral zirconium components in total knee arthroplasty preliminary results. J Arthroplasty. 2011;26:445–450.PubMedCrossRef Heyse TJ, Davis J, Haas SB, Chen DX, Wright TM, Laskin RS. Retrieval analysis of femoral zirconium components in total knee arthroplasty preliminary results. J Arthroplasty. 2011;26:445–450.PubMedCrossRef
13.
Zurück zum Zitat Hood R, Wright T, Burstein AH. Retrieval analysis of total knee prostheses: a method and its application to 48 condylar prostheses. J Biomed Mater Res. 1983;17:829–842.PubMedCrossRef Hood R, Wright T, Burstein AH. Retrieval analysis of total knee prostheses: a method and its application to 48 condylar prostheses. J Biomed Mater Res. 1983;17:829–842.PubMedCrossRef
14.
Zurück zum Zitat Implant Research Center Writing Committee, Medel FJ, Kurtz SM, Sharkey PF, Austin MS, Klein GR, Cohen AR, Patel H, Goldberg VM, Kraay MJ, Rimnac CM. Post damage in contemporary posterior-stabilized tibial inserts influence of implant design and clinical relevance. J Arthroplasty. 2011;26:606–614.PubMedCrossRef Implant Research Center Writing Committee, Medel FJ, Kurtz SM, Sharkey PF, Austin MS, Klein GR, Cohen AR, Patel H, Goldberg VM, Kraay MJ, Rimnac CM. Post damage in contemporary posterior-stabilized tibial inserts influence of implant design and clinical relevance. J Arthroplasty. 2011;26:606–614.PubMedCrossRef
15.
Zurück zum Zitat Jung KA, Lee SC, Hawng SH, Kim SM. Fracture of a second-generation highly cross-linked UHMWPE tibial post in a posterior-stabilized Scorpio knee system. Clin Orthop Relat Res. 2008;31:1137–1140. Jung KA, Lee SC, Hawng SH, Kim SM. Fracture of a second-generation highly cross-linked UHMWPE tibial post in a posterior-stabilized Scorpio knee system. Clin Orthop Relat Res. 2008;31:1137–1140.
16.
Zurück zum Zitat Kurtz SM, Walker PS. How have new designs and new types of joint replacement influenced wear behaviour? J Am Acad Orthop Surg. 2008;16:107–110. Kurtz SM, Walker PS. How have new designs and new types of joint replacement influenced wear behaviour? J Am Acad Orthop Surg. 2008;16:107–110.
17.
Zurück zum Zitat Lachiewicz PF. How to treat a tibial post fracture in total knee arthroplasty: a systematic review. Clin Orthop Relat Res. 2010;496:1709–1715. Lachiewicz PF. How to treat a tibial post fracture in total knee arthroplasty: a systematic review. Clin Orthop Relat Res. 2010;496:1709–1715.
18.
Zurück zum Zitat Lee C, Chen W, Kou H, Lo W, Chen C. Case report early nontraumatic fracture of the polyethylene tibial post in a NexGen LPS-flex posterior stabilized knee prosthesis. J Arthroplasty. 2009;24:1295–1299. Lee C, Chen W, Kou H, Lo W, Chen C. Case report early nontraumatic fracture of the polyethylene tibial post in a NexGen LPS-flex posterior stabilized knee prosthesis. J Arthroplasty. 2009;24:1295–1299.
19.
Zurück zum Zitat Lim HC, Bae JH, Hwang JH, Kim SJ, Yoon JY. Fracture of a polyethylene tibial post in a Scorpio posterior-stabilized knee prosthesis. Clin Orthop Relat Res. 2009;1:118–121. Lim HC, Bae JH, Hwang JH, Kim SJ, Yoon JY. Fracture of a polyethylene tibial post in a Scorpio posterior-stabilized knee prosthesis. Clin Orthop Relat Res. 2009;1:118–121.
20.
Zurück zum Zitat Lombardi AV, Ellison BS, Berend KR. Polyethylene wear is influenced by manufacturing technique in modular TKA. Clin Orthop Relat Res. 2008;466:2798–2805.PubMedCrossRef Lombardi AV, Ellison BS, Berend KR. Polyethylene wear is influenced by manufacturing technique in modular TKA. Clin Orthop Relat Res. 2008;466:2798–2805.PubMedCrossRef
21.
Zurück zum Zitat Lu YC, Huang HC, Chang TK, Ho FY, Cheng CK, Huang CH. Wear-pattern analysis in retrieved tibial inserts of mobile-bearing and fixed-bearing total knee prostheses. J Bone Joint Surg Am. 2010;92:500–507.CrossRef Lu YC, Huang HC, Chang TK, Ho FY, Cheng CK, Huang CH. Wear-pattern analysis in retrieved tibial inserts of mobile-bearing and fixed-bearing total knee prostheses. J Bone Joint Surg Am. 2010;92:500–507.CrossRef
22.
Zurück zum Zitat Luo SX, Su W, Zhao JM, Sha K, Wei QJ, Li XF. High-flexion vs conventional prostheses total knee arthroplasty: a meta-analysis. J Arthroplasty. 2011;26:847–854.PubMedCrossRef Luo SX, Su W, Zhao JM, Sha K, Wei QJ, Li XF. High-flexion vs conventional prostheses total knee arthroplasty: a meta-analysis. J Arthroplasty. 2011;26:847–854.PubMedCrossRef
23.
Zurück zum Zitat Mauerhan DR. Case report fracture of the polyethylene tibial post in a posterior cruciate-substituting total knee arthroplasty mimicking patellar clunk syndrome: a report of 5 cases. J Arthroplasty. 2003;18:942–945.PubMedCrossRef Mauerhan DR. Case report fracture of the polyethylene tibial post in a posterior cruciate-substituting total knee arthroplasty mimicking patellar clunk syndrome: a report of 5 cases. J Arthroplasty. 2003;18:942–945.PubMedCrossRef
24.
Zurück zum Zitat McCalden RW, MacDonald SJ, Bourne RB, Marr JT. A randomized controlled trial comparing ‘high-flex’ vs ‘standard’ posterior cruciate substituting polyethylene tibial inserts in total knee arthroplasty. J Arthroplasty. 2009;24:33–38.PubMedCrossRef McCalden RW, MacDonald SJ, Bourne RB, Marr JT. A randomized controlled trial comparing ‘high-flex’ vs ‘standard’ posterior cruciate substituting polyethylene tibial inserts in total knee arthroplasty. J Arthroplasty. 2009;24:33–38.PubMedCrossRef
25.
Zurück zum Zitat Pinskerova V, Samuelson KM, Stammers J, Maruthainar K, Sosna A, Freeman M. The knee in full flexion: an anatomical study. J Bone Joint Surg Br. 2009;91:830–834.PubMed Pinskerova V, Samuelson KM, Stammers J, Maruthainar K, Sosna A, Freeman M. The knee in full flexion: an anatomical study. J Bone Joint Surg Br. 2009;91:830–834.PubMed
26.
Zurück zum Zitat Puloski SK, McCalden RW, MacDonald SJ, Rorabeck CH, Bourne RB. Tibial post wear in posterior stabilized total knee arthroplasty. An unrecognized source of polyethylene debris. J Bone Joint Surg Am. 2001;83:390–397.PubMed Puloski SK, McCalden RW, MacDonald SJ, Rorabeck CH, Bourne RB. Tibial post wear in posterior stabilized total knee arthroplasty. An unrecognized source of polyethylene debris. J Bone Joint Surg Am. 2001;83:390–397.PubMed
27.
Zurück zum Zitat Silva M, Schmalzried TP. Polyethylene in total knee arthroplasty. In: Callaghan JJ, Rosenberg AG, Rubash HE, Simonian PT, Wickieqicz TL, eds. The Adult Knee. 1st ed. Philadelphia, PA, USA: Lippincott Williams & Wilkins; 2003:279–286. Silva M, Schmalzried TP. Polyethylene in total knee arthroplasty. In: Callaghan JJ, Rosenberg AG, Rubash HE, Simonian PT, Wickieqicz TL, eds. The Adult Knee. 1st ed. Philadelphia, PA, USA: Lippincott Williams & Wilkins; 2003:279–286.
28.
Zurück zum Zitat Stoller AP, Johnson TS, Popoola OO, Humphrey SM, Blanchard CR. Highly crosslinked polyethylene in posterior-stabilized total knee arthroplasty: in vitro performance evaluation of wear, delamination, and tibial post durability. J Arthroplasty. 2011;26:483–491.PubMedCrossRef Stoller AP, Johnson TS, Popoola OO, Humphrey SM, Blanchard CR. Highly crosslinked polyethylene in posterior-stabilized total knee arthroplasty: in vitro performance evaluation of wear, delamination, and tibial post durability. J Arthroplasty. 2011;26:483–491.PubMedCrossRef
29.
Zurück zum Zitat Teeter MG, Naudie DDR, McErlain DD, Brandt J-M, Yuan X, MacDonald SJ, Holdsworth DW. In vitro quantification of wear in tibial inserts using microcomputed tomography. Clin Orthop Relat Res. 2011;469:107–112.PubMedCrossRef Teeter MG, Naudie DDR, McErlain DD, Brandt J-M, Yuan X, MacDonald SJ, Holdsworth DW. In vitro quantification of wear in tibial inserts using microcomputed tomography. Clin Orthop Relat Res. 2011;469:107–112.PubMedCrossRef
30.
Zurück zum Zitat Teeter MG, Naudie DDR, Milner JS, Holdsworth DW. Determination of reference geometry for polyethylene tibial insert wear analysis. J Arthroplasty. 2011;26:497–503.PubMedCrossRef Teeter MG, Naudie DDR, Milner JS, Holdsworth DW. Determination of reference geometry for polyethylene tibial insert wear analysis. J Arthroplasty. 2011;26:497–503.PubMedCrossRef
31.
Zurück zum Zitat Tsao AK, Jones LC, Lewallen DG. What patient and surgical factors contribute to implant wear and osteolysis in total joint arthroplasty. J Am Acad Orthop Surg. 2008;16:7–13. Tsao AK, Jones LC, Lewallen DG. What patient and surgical factors contribute to implant wear and osteolysis in total joint arthroplasty. J Am Acad Orthop Surg. 2008;16:7–13.
Metadaten
Titel
The 2012 Mark Coventry Award: A Retrieval Analysis of High Flexion versus Posterior-stabilized Tibial Inserts
verfasst von
Nicholas R. Paterson, BScH
Matthew G. Teeter, BScH, PhD
Steven J. MacDonald, MD, FRCSC
Richard W. McCalden, MD, MPhil(Edin), FRCSC
Douglas D. R. Naudie, MD, FRCSC
Publikationsdatum
01.01.2013
Verlag
Springer-Verlag
Erschienen in
Clinical Orthopaedics and Related Research® / Ausgabe 1/2013
Print ISSN: 0009-921X
Elektronische ISSN: 1528-1132
DOI
https://doi.org/10.1007/s11999-012-2387-3

Weitere Artikel der Ausgabe 1/2013

Clinical Orthopaedics and Related Research® 1/2013 Zur Ausgabe

Symposium: Papers Presented at the Annual Meetings of the Knee Society

Does Interlimb Knee Symmetry Exist After Unicompartmental Knee Arthroplasty?

Symposium: Papers Presented at the Annual Meetings of the Knee Society

Durability of a Cruciate-retaining TKA With Modular Tibial Trays at 20 Years

Symposium: Papers Presented at the Annual Meetings of the Knee Society

Is Administratively Coded Comorbidity and Complication Data in Total Joint Arthroplasty Valid?

Arthropedia

Grundlagenwissen der Arthroskopie und Gelenkchirurgie. Erweitert durch Fallbeispiele, Videos und Abbildungen. 
» Jetzt entdecken

Knie-TEP: Kein Vorteil durch antibiotikahaltigen Knochenzement

29.05.2024 Periprothetische Infektionen Nachrichten

Zur Zementierung einer Knie-TEP wird in Deutschland zu über 98% Knochenzement verwendet, der mit einem Antibiotikum beladen ist. Ob er wirklich besser ist als Zement ohne Antibiotikum, kann laut Registerdaten bezweifelt werden.

Häusliche Gewalt in der orthopädischen Notaufnahme oft nicht erkannt

28.05.2024 Häusliche Gewalt Nachrichten

In der Notaufnahme wird die Chance, Opfer von häuslicher Gewalt zu identifizieren, von Orthopäden und Orthopädinnen offenbar zu wenig genutzt. Darauf deuten die Ergebnisse einer Fragebogenstudie an der Sahlgrenska-Universität in Schweden hin.

Fehlerkultur in der Medizin – Offenheit zählt!

28.05.2024 Fehlerkultur Podcast

Darüber reden und aus Fehlern lernen, sollte das Motto in der Medizin lauten. Und zwar nicht nur im Sinne der Patientensicherheit. Eine negative Fehlerkultur kann auch die Behandelnden ernsthaft krank machen, warnt Prof. Dr. Reinhard Strametz. Ein Plädoyer und ein Leitfaden für den offenen Umgang mit kritischen Ereignissen in Medizin und Pflege.

Mehr Frauen im OP – weniger postoperative Komplikationen

21.05.2024 Allgemeine Chirurgie Nachrichten

Ein Frauenanteil von mindestens einem Drittel im ärztlichen Op.-Team war in einer großen retrospektiven Studie aus Kanada mit einer signifikanten Reduktion der postoperativen Morbidität assoziiert.

Update Orthopädie und Unfallchirurgie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.