The importance of surrounding tissues and window settings for contouring of moving targets

verfasst von: Kai Joachim Borm, Markus Oechsner, PhD, Johannes Berndt, BSc, Stephanie Elisabeth Combs, Michael Molls, MD, Marciana Nona Duma, MD

Erschienen in: Strahlentherapie und Onkologie | Ausgabe 9/2015

Einloggen, um Zugang zu erhalten

Abstract

Aim

The aim of the study was to assess the importance of surrounding tissues for the delineation of moving targets in tissue-specific phantoms and to find optimal settings for lung, soft tissue, and liver tumors.

Materials and methods

Tumor movement was simulated by a water-filled table tennis ball (target volume, TV). Three phantoms were created: corkboards to simulate lung tissue (lung phantom, LunPh), animal fat as fatty soft tissue (fatty tissue phantom, FatPh), and water enhanced with contrast medium as the liver tissue (liver phantom, LivPh). Slow planning three-dimensional compute tomography images (3D-CTs) were acquired with and without phantom movements. One-dimensional tumor movement (1D), three-dimensional tumor movement (3D), as well as a real patient’s tumor trajectories were simulated. The TV was contoured using two lung window settings, two soft-tissue window settings, and one liver window setting. The volumes were compared to mathematical calculated values.

Results

TVs were underestimated in all phantoms due to movement. The use of soft-tissue windows in the LivPh led to a significantunderestimation of the TV (70.8 % of calculated TV). When common window settings [LunPh + 200 HU/-1,000 HU (upper window/lower window threshold); FatPh: + 240 HU/-120 HU; LivPh: + 175 HU/+ 50 HU] were used, the contoured TVs were: LivPh, 84.0 %; LunPh, 93.2 %, and FatPh, 92.8 %. The lower window threshold had a significant impact on the size of the delineated TV, whereas changes of the upper threshold led only to small differences.

Conclusion

The decisive factor for window settings is the lower window threshold (for adequate TV delineation in the lung and fatty-soft tissue it should be lower than density values of surrounding tissue). The use of a liver window should be considered.
Literatur
1.
Zurück zum Zitat Langen KM, Jones DT (2001) Organ motion and its management. Int J Radiat Oncol Biol Phys 50:265–278CrossRefPubMed Langen KM, Jones DT (2001) Organ motion and its management. Int J Radiat Oncol Biol Phys 50:265–278CrossRefPubMed
3.
Zurück zum Zitat Wurstbauer K et al (2005) Radiotherapy planning for lung cancer: slow CTs allow the drawing of tighter margins. Radiother Oncol 75:165–170CrossRefPubMed Wurstbauer K et al (2005) Radiotherapy planning for lung cancer: slow CTs allow the drawing of tighter margins. Radiother Oncol 75:165–170CrossRefPubMed
4.
Zurück zum Zitat Chen GT, Kung JH, Beaudette KP (2004) Artifacts in computed tomography scanning of moving objects. Semin Radiat Oncol 14:19–26CrossRefPubMed Chen GT, Kung JH, Beaudette KP (2004) Artifacts in computed tomography scanning of moving objects. Semin Radiat Oncol 14:19–26CrossRefPubMed
5.
Zurück zum Zitat Bundschuh RA et al (2012) Respiratory gated [18F]FDG PET/CT for target volume delineation in stereotactic radiation treatment of liver metastases. Strahlenther Onkol 188:592–598CrossRefPubMed Bundschuh RA et al (2012) Respiratory gated [18F]FDG PET/CT for target volume delineation in stereotactic radiation treatment of liver metastases. Strahlenther Onkol 188:592–598CrossRefPubMed
6.
Zurück zum Zitat Keall P (2004) 4-dimensional computed tomography imaging and treatment planning. Semin Radiat Oncol 14:81–90CrossRefPubMed Keall P (2004) 4-dimensional computed tomography imaging and treatment planning. Semin Radiat Oncol 14:81–90CrossRefPubMed
7.
Zurück zum Zitat Pech M et al (2008) Radiotherapy of liver metastases. Comparison of target volumes and dose-volume histograms employing CT- or MRI-based treatment planning. Strahlenther Onkol 184:256–261CrossRefPubMed Pech M et al (2008) Radiotherapy of liver metastases. Comparison of target volumes and dose-volume histograms employing CT- or MRI-based treatment planning. Strahlenther Onkol 184:256–261CrossRefPubMed
8.
Zurück zum Zitat Dalah EZ, Nisbet A, Bradley D (2010) Effect of window level on target volume delineation in treatment planning. Appl Radiat Isot 68:602–604CrossRefPubMed Dalah EZ, Nisbet A, Bradley D (2010) Effect of window level on target volume delineation in treatment planning. Appl Radiat Isot 68:602–604CrossRefPubMed
9.
Zurück zum Zitat Araki T et al (1980) Dynamic CT densitometry of hepatic tumors. AJR Am J Roentgenol 135:1037–1043CrossRefPubMed Araki T et al (1980) Dynamic CT densitometry of hepatic tumors. AJR Am J Roentgenol 135:1037–1043CrossRefPubMed
10.
Zurück zum Zitat Engelsman M et al (2001) Impact of simple tissue inhomogeneity correction algorithms on conformal radiotherapy of lung tumours. Radiother Oncol 60:299–309CrossRefPubMed Engelsman M et al (2001) Impact of simple tissue inhomogeneity correction algorithms on conformal radiotherapy of lung tumours. Radiother Oncol 60:299–309CrossRefPubMed
11.
Zurück zum Zitat Knoos T et al (1995) Limitations of a pencil beam approach to photon dose calculations in lung tissue. Phys Med Biol 40:1411–1420CrossRefPubMed Knoos T et al (1995) Limitations of a pencil beam approach to photon dose calculations in lung tissue. Phys Med Biol 40:1411–1420CrossRefPubMed
12.
Zurück zum Zitat Shirai K et al (2014) Phantom and clinical study of differences in cone beam computed tomographic registration when aligned to maximum and average intensity projection. Int J Radiat Oncol Biol Phys 88:189–194CrossRefPubMed Shirai K et al (2014) Phantom and clinical study of differences in cone beam computed tomographic registration when aligned to maximum and average intensity projection. Int J Radiat Oncol Biol Phys 88:189–194CrossRefPubMed
13.
Zurück zum Zitat Fayad LM et al (2002) Chest CT window settings with multiscale adaptive histogram equalization: pilot study. Radiology 223:845–852CrossRefPubMed Fayad LM et al (2002) Chest CT window settings with multiscale adaptive histogram equalization: pilot study. Radiology 223:845–852CrossRefPubMed
14.
Zurück zum Zitat Harris KM et al (1993) The effect on apparent size of simulated pulmonary nodules of using three standard CT window settings. Clin Radiol 47:241–244CrossRefPubMed Harris KM et al (1993) The effect on apparent size of simulated pulmonary nodules of using three standard CT window settings. Clin Radiol 47:241–244CrossRefPubMed
15.
Zurück zum Zitat Nino-Murcia M et al (2000) Focal liver lesions: pattern-based classification scheme for enhancement at arterial phase CT. Radiology 215:746–751CrossRefPubMed Nino-Murcia M et al (2000) Focal liver lesions: pattern-based classification scheme for enhancement at arterial phase CT. Radiology 215:746–751CrossRefPubMed
16.
Zurück zum Zitat Chan R et al (2001) Computed tomographic-pathologic correlation of gross tumor volume and clinical target volume in non-small cell lung cancer: a pilot experience. Arch Pathol Lab Med 125:1469–1472PubMed Chan R et al (2001) Computed tomographic-pathologic correlation of gross tumor volume and clinical target volume in non-small cell lung cancer: a pilot experience. Arch Pathol Lab Med 125:1469–1472PubMed
17.
Zurück zum Zitat Macpherson RE et al (2009) Non-small-cell lung cancer dimensions: CT-pathological correlation and interobserver variation. Br J Radiol 82:421–425CrossRefPubMed Macpherson RE et al (2009) Non-small-cell lung cancer dimensions: CT-pathological correlation and interobserver variation. Br J Radiol 82:421–425CrossRefPubMed
18.
Zurück zum Zitat Grills IS et al (2007) Clinicopathologic analysis of microscopic extension in lung adenocarcinoma: defining clinical target volume for radiotherapy. Int J Radiat Oncol Biol Phys 69:334–341CrossRefPubMed Grills IS et al (2007) Clinicopathologic analysis of microscopic extension in lung adenocarcinoma: defining clinical target volume for radiotherapy. Int J Radiat Oncol Biol Phys 69:334–341CrossRefPubMed
19.
Zurück zum Zitat Borm KJ et al (2014) The impact of CT window settings on the contouring of a moving target: a phantom study. Clin Radiol 69:331–336CrossRef Borm KJ et al (2014) The impact of CT window settings on the contouring of a moving target: a phantom study. Clin Radiol 69:331–336CrossRef
20.
Zurück zum Zitat Mayo-Smith WW et al (1999) Detecting hepatic lesions: the added utility of CT liver window settings. Radiology 210:601–604CrossRefPubMed Mayo-Smith WW et al (1999) Detecting hepatic lesions: the added utility of CT liver window settings. Radiology 210:601–604CrossRefPubMed
21.
Metadaten
Titel
The importance of surrounding tissues and window settings for contouring of moving targets
verfasst von
Kai Joachim Borm
Markus Oechsner, PhD
Johannes Berndt, BSc
Stephanie Elisabeth Combs
Michael Molls, MD
Marciana Nona Duma, MD
Publikationsdatum
01.09.2015
Verlag
Springer Berlin Heidelberg
Erschienen in
Strahlentherapie und Onkologie / Ausgabe 9/2015
Print ISSN: 0179-7158
Elektronische ISSN: 1439-099X
DOI
https://doi.org/10.1007/s00066-015-0862-y

Weitere Artikel der Ausgabe 9/2015

Strahlentherapie und Onkologie 9/2015 Zur Ausgabe

Umsetzung der POMGAT-Leitlinie läuft

03.05.2024 DCK 2024 Kongressbericht

Seit November 2023 gibt es evidenzbasierte Empfehlungen zum perioperativen Management bei gastrointestinalen Tumoren (POMGAT) auf S3-Niveau. Vieles wird schon entsprechend der Empfehlungen durchgeführt. Wo es im Alltag noch hapert, zeigt eine Umfrage in einem Klinikverbund.

CUP-Syndrom: Künstliche Intelligenz kann Primärtumor finden

30.04.2024 Künstliche Intelligenz Nachrichten

Krebserkrankungen unbekannten Ursprungs (CUP) sind eine diagnostische Herausforderung. KI-Systeme können Pathologen dabei unterstützen, zytologische Bilder zu interpretieren, um den Primärtumor zu lokalisieren.

Sind Frauen die fähigeren Ärzte?

30.04.2024 Gendermedizin Nachrichten

Patienten, die von Ärztinnen behandelt werden, dürfen offenbar auf bessere Therapieergebnisse hoffen als Patienten von Ärzten. Besonders gilt das offenbar für weibliche Kranke, wie eine Studie zeigt.

Adjuvante Immuntherapie verlängert Leben bei RCC

25.04.2024 Nierenkarzinom Nachrichten

Nun gibt es auch Resultate zum Gesamtüberleben: Eine adjuvante Pembrolizumab-Therapie konnte in einer Phase-3-Studie das Leben von Menschen mit Nierenzellkarzinom deutlich verlängern. Die Sterberate war im Vergleich zu Placebo um 38% geringer.

Update Onkologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.