Three-dimensional surface scanning for accurate patient positioning and monitoring during breast cancer radiotherapy

verfasst von: C. Gaisberger, MSc, P. Steininger, B. Mitterlechner, S. Huber, H. Weichenberger, F. Sedlmayer, H. Deutschmann

Erschienen in: Strahlentherapie und Onkologie | Ausgabe 10/2013

Einloggen, um Zugang zu erhalten

Abstract

Purpose

Clinical evaluation of an optical three-dimensional surface scanning (3D-SS) system for patient positioning and monitoring during radiotherapy (RT) for breast cancer.

Materials and methods

A ceiling-mounted scanner was developed to acquire multiple 3D body surface images and tested in 14 conservatively operated breast cancer patients. A reference skin surface was derived from the planning computed tomography (CT) scan as basis for rigid registration with the surface scans. In addition to electronic portal images (EPIs), optical scans were acquired at three defined time points before and during daily RT. Patient setup was guided by laser alignments and corrected according to EPI findings. The accuracy of the 3D-SS system was validated by comparison of the optical scans to EPIs generated in parallel. Interfraction shifts were investigated by comparison of the first 3D-SS image with the reference body outline. Intrafractional motions were analysed by comparing the three daily surface scans with the first EPI.

Results

Comparison of EPIs and 3D-SS images revealed good accordance (− 0.05 ± 0.94 mm). Analysis of daily patient positions revealed average deviations of 0.4 ± 2.4 mm laterally, 0.3 ± 1.9 mm longitudinally and 0.2 ± 3.3 mm vertically. After 2 weeks, a systematic interfraction shift in patient positioning was noted, particularly in the vertical direction (4.9 ± 0.56 mm), which was attributed to patients progressively relaxing. 3D-SS images showed intrafractional shifts of 1.2 ± 0.7 mm over a time course of 2 min.

Conclusion

Optical surface scanning is a simple, fast and reproducible method for breast cancer patient alignment. Particularly for more sophisticated irradiation techniques, it helps to improve accuracy in patient positioning during radiotherapy without the exposure to additional ionizing radiation.
Literatur
1.
Zurück zum Zitat Bartoncini S, Fiandra C, Ruo Redda MG et al (2012) Target registration errors with surface imaging system in conformal radiotherapy for prostate cancer: study on 19 patients. Radiol Med 117(8):1419–1428PubMedCrossRef Bartoncini S, Fiandra C, Ruo Redda MG et al (2012) Target registration errors with surface imaging system in conformal radiotherapy for prostate cancer: study on 19 patients. Radiol Med 117(8):1419–1428PubMedCrossRef
2.
Zurück zum Zitat Bert C, Metheany KG, Doppke K et al (2005) A phantom evaluation of a stereo-vision surface imaging system for radiotherapy setup. Med Phys 2:2753–2762CrossRef Bert C, Metheany KG, Doppke K et al (2005) A phantom evaluation of a stereo-vision surface imaging system for radiotherapy setup. Med Phys 2:2753–2762CrossRef
3.
Zurück zum Zitat Bert C, Metheany KG, Doppke KP et al (2006) Clinical experience with a 3D surface patient setup system for alignment of partial-breast irradiation patients. Int J Radiat Oncol Biol Phys 64:1265–1274PubMedCrossRef Bert C, Metheany KG, Doppke KP et al (2006) Clinical experience with a 3D surface patient setup system for alignment of partial-breast irradiation patients. Int J Radiat Oncol Biol Phys 64:1265–1274PubMedCrossRef
4.
Zurück zum Zitat Besl P, McKay N (1992) A method for registration of 3-D shapes. IEEE trans. Pat Anal Mach Intel 14(2):239–256CrossRef Besl P, McKay N (1992) A method for registration of 3-D shapes. IEEE trans. Pat Anal Mach Intel 14(2):239–256CrossRef
5.
Zurück zum Zitat Brahme A, Nyman P, Skatt B (2008) 4D laser camera for accurate patient positioning, collision avoidance, image fusion and adaptive approaches during diagnostic and therapeutic procedures. Med Phys 35:1670–1681PubMedCrossRef Brahme A, Nyman P, Skatt B (2008) 4D laser camera for accurate patient positioning, collision avoidance, image fusion and adaptive approaches during diagnostic and therapeutic procedures. Med Phys 35:1670–1681PubMedCrossRef
6.
Zurück zum Zitat Calow R, Gademann G, Krell G et al (2002) Photogrammetric measurement of patients in radiotherapy. ISPRS J Photogrammet Remote Sensing 56:347–359CrossRef Calow R, Gademann G, Krell G et al (2002) Photogrammetric measurement of patients in radiotherapy. ISPRS J Photogrammet Remote Sensing 56:347–359CrossRef
7.
Zurück zum Zitat Czigner K, Agoston P, Forgács G et al (2012) Patient positioning variations to reduce dose to normal tissues during 3D conformal radiotherapy for high-risk prostate cancer. Strahlenther Onkol 188(9):816–822PubMedCrossRef Czigner K, Agoston P, Forgács G et al (2012) Patient positioning variations to reduce dose to normal tissues during 3D conformal radiotherapy for high-risk prostate cancer. Strahlenther Onkol 188(9):816–822PubMedCrossRef
8.
Zurück zum Zitat Boer HC de, Sörnsen de Koste JR van, Senan S et al (2001) Analysis and reduction of 3D systematic and random setup errors during the simulation and treatment of lung cancer patients with CT-based external beam radiotherapy dose planning. Int J Radiat Oncol Biol Phys 49(3):857–868PubMedCrossRef Boer HC de, Sörnsen de Koste JR van, Senan S et al (2001) Analysis and reduction of 3D systematic and random setup errors during the simulation and treatment of lung cancer patients with CT-based external beam radiotherapy dose planning. Int J Radiat Oncol Biol Phys 49(3):857–868PubMedCrossRef
9.
Zurück zum Zitat Deutschmann H, Steininger P, Nairz O et al (2008) “Augmented reality” in conventional simulation by projection of 3-D structures into 2-D images: a comparison with virtual methods. Strahlenther Onkol 184(2):93–99PubMedCrossRef Deutschmann H, Steininger P, Nairz O et al (2008) “Augmented reality” in conventional simulation by projection of 3-D structures into 2-D images: a comparison with virtual methods. Strahlenther Onkol 184(2):93–99PubMedCrossRef
10.
Zurück zum Zitat Dickie-Euler C, Parent A, Griffin A et al (2008) Measuring interfraction and intrafraction motion with Cone Beam ComputedTomography (CBCT) and an Optical Localization System (OLS) for lower extremity soft tissue sarcoma patients treated with Preoperative Intensity Modulated Radiation Therapy (IMRT). Int J Radiat Oncol Biol Phys 72:S105–S105CrossRef Dickie-Euler C, Parent A, Griffin A et al (2008) Measuring interfraction and intrafraction motion with Cone Beam ComputedTomography (CBCT) and an Optical Localization System (OLS) for lower extremity soft tissue sarcoma patients treated with Preoperative Intensity Modulated Radiation Therapy (IMRT). Int J Radiat Oncol Biol Phys 72:S105–S105CrossRef
11.
Zurück zum Zitat Gierga DP, Riboldi M, Turcotte JC et al (2008) Comparison of target registration errors for multiple image-guided techniques in accelerated partial breast irradiation. Int J Radiat Oncol Biol Phys 70(4):1239–1246PubMedCrossRef Gierga DP, Riboldi M, Turcotte JC et al (2008) Comparison of target registration errors for multiple image-guided techniques in accelerated partial breast irradiation. Int J Radiat Oncol Biol Phys 70(4):1239–1246PubMedCrossRef
12.
Zurück zum Zitat Giovanna S, Carocci M, Rodella R (1999) Three-dimensional vision based on a combination of gray-code and phase-shift light projection: analysis and compensation of the systematic errors. Appl Optics 38(31):6565–6573CrossRef Giovanna S, Carocci M, Rodella R (1999) Three-dimensional vision based on a combination of gray-code and phase-shift light projection: analysis and compensation of the systematic errors. Appl Optics 38(31):6565–6573CrossRef
13.
Zurück zum Zitat Gong GZ, Yin Y, Xing LG et al (2012) RapidArc combined with the active breathing coordinator provides an effective and accurate approach for the radiotherapy of hepatocellular carcinoma. Strahlenther Onkol 188(3):262–268PubMedCrossRef Gong GZ, Yin Y, Xing LG et al (2012) RapidArc combined with the active breathing coordinator provides an effective and accurate approach for the radiotherapy of hepatocellular carcinoma. Strahlenther Onkol 188(3):262–268PubMedCrossRef
14.
Zurück zum Zitat Gulybán A, Kovács P, Sebestyén Z et al (2008) Multisegmented tangential breast fields: a rational way to treat breast cancer. Strahlenther Onkol 184(5):262–269PubMedCrossRef Gulybán A, Kovács P, Sebestyén Z et al (2008) Multisegmented tangential breast fields: a rational way to treat breast cancer. Strahlenther Onkol 184(5):262–269PubMedCrossRef
15.
Zurück zum Zitat Hasan Y, Kim L, Martinez A et al (2008) Image guidance in external beam accelerated partial breast irradiation: comparison of surrogates for the lumpectomy cavity. Int J Radiat Oncol Biol Phys 70(2):619–625PubMedCrossRef Hasan Y, Kim L, Martinez A et al (2008) Image guidance in external beam accelerated partial breast irradiation: comparison of surrogates for the lumpectomy cavity. Int J Radiat Oncol Biol Phys 70(2):619–625PubMedCrossRef
16.
Zurück zum Zitat Jin J, Ajlouni M, Ryu S et al (2007) A technique of quantitatively monitoring both respiratory and nonrespiratory motion in patients using external body markers. Med Phys 34:2875–2881PubMedCrossRef Jin J, Ajlouni M, Ryu S et al (2007) A technique of quantitatively monitoring both respiratory and nonrespiratory motion in patients using external body markers. Med Phys 34:2875–2881PubMedCrossRef
17.
Zurück zum Zitat Kang HJ, Grelewicz Z, Wiersma RD (2012) Development of an automated region of interest selection method for 3D surface monitoring of head motion. Med Phys 39(6):3270–3282PubMedCrossRef Kang HJ, Grelewicz Z, Wiersma RD (2012) Development of an automated region of interest selection method for 3D surface monitoring of head motion. Med Phys 39(6):3270–3282PubMedCrossRef
18.
Zurück zum Zitat Krengli M, Gaiano S, Mones E et al (2009) Reproducibility of patient setup by surface image registration system in conformal radiotherapy of prostate cancer. Radiat Oncol 4:9PubMedCrossRef Krengli M, Gaiano S, Mones E et al (2009) Reproducibility of patient setup by surface image registration system in conformal radiotherapy of prostate cancer. Radiat Oncol 4:9PubMedCrossRef
19.
Zurück zum Zitat Liebich J, Licher J, Scherf C et al (2011) Simple proposal for dosimetry with an Elekta iViewGT™ electronic portal imaging device (EPID) using commercial software modules. Strahlenther Onkol 187(5):316–321PubMedCrossRef Liebich J, Licher J, Scherf C et al (2011) Simple proposal for dosimetry with an Elekta iViewGT™ electronic portal imaging device (EPID) using commercial software modules. Strahlenther Onkol 187(5):316–321PubMedCrossRef
20.
Zurück zum Zitat Moore C, Lilley F, Sauret V et al (2003) Opto-electronic sensing of body surface topology changes during radiotherapy for rectal cancer. Int J Radiat Oncol Biol Phys 56(1):248–258PubMedCrossRef Moore C, Lilley F, Sauret V et al (2003) Opto-electronic sensing of body surface topology changes during radiotherapy for rectal cancer. Int J Radiat Oncol Biol Phys 56(1):248–258PubMedCrossRef
21.
Zurück zum Zitat Mulchrone K (2003) Application of Delaunay triangulation to the nearest neighbour method of strain analysis. J Struct Geol 25:689–702CrossRef Mulchrone K (2003) Application of Delaunay triangulation to the nearest neighbour method of strain analysis. J Struct Geol 25:689–702CrossRef
22.
Zurück zum Zitat Nairz O, Merz F, Deutschmann H et al (2008) A strategy for the use of image-guided radiotherapy (IGRT) on linear accelerators and its impact on treatment margins for prostate cancer patients. Strahlenther Onkol 184:663–667PubMedCrossRef Nairz O, Merz F, Deutschmann H et al (2008) A strategy for the use of image-guided radiotherapy (IGRT) on linear accelerators and its impact on treatment margins for prostate cancer patients. Strahlenther Onkol 184:663–667PubMedCrossRef
23.
Zurück zum Zitat Pallotta S, Marrazzo L, Ceroti M et al (2012) A phantom evaluation of Sentinel(™), a commercial laser/camera surface imaging system for patient setup verification in radiotherapy. Med Phys 39(2):706–712PubMedCrossRef Pallotta S, Marrazzo L, Ceroti M et al (2012) A phantom evaluation of Sentinel(™), a commercial laser/camera surface imaging system for patient setup verification in radiotherapy. Med Phys 39(2):706–712PubMedCrossRef
24.
Zurück zum Zitat Pili G, Grimaldi L, Fidanza C et al (2011) Geometric and dosimetric approach to determine probability of late cardiac mortality in left tangential breast irradiation: comparison between wedged beams and field-in-field technique. Int J Radiat Oncol Biol Phys 81(3):894–900PubMedCrossRef Pili G, Grimaldi L, Fidanza C et al (2011) Geometric and dosimetric approach to determine probability of late cardiac mortality in left tangential breast irradiation: comparison between wedged beams and field-in-field technique. Int J Radiat Oncol Biol Phys 81(3):894–900PubMedCrossRef
25.
Zurück zum Zitat Remouchamps VM, Letts N, Yan D et al (2003) Three-dimensional evaluation of intra- and interfraction immobilization of lung and chest wall using active breathing control: a reproducibility study with breast cancer patients. Int J Radiat Oncol Biol Phys 57(4):968–978PubMedCrossRef Remouchamps VM, Letts N, Yan D et al (2003) Three-dimensional evaluation of intra- and interfraction immobilization of lung and chest wall using active breathing control: a reproducibility study with breast cancer patients. Int J Radiat Oncol Biol Phys 57(4):968–978PubMedCrossRef
26.
Zurück zum Zitat Stieler F, Wenz F, Scherrer D et al (2012) Clinical evaluation of a commercial surface-imaging system for patient positioning in radiotherapy. Strahlenther Onkol 188(12):1080–1084PubMedCrossRef Stieler F, Wenz F, Scherrer D et al (2012) Clinical evaluation of a commercial surface-imaging system for patient positioning in radiotherapy. Strahlenther Onkol 188(12):1080–1084PubMedCrossRef
27.
Zurück zum Zitat Subramaniam S, Thirumalaiswamy S, Srinivas C et al (2012) Chest wall radiotherapy with volumetric modulated arcs and the potential role of flattening filter free photon beams. Strahlenther Onkol 188(6):484–490PubMedCrossRef Subramaniam S, Thirumalaiswamy S, Srinivas C et al (2012) Chest wall radiotherapy with volumetric modulated arcs and the potential role of flattening filter free photon beams. Strahlenther Onkol 188(6):484–490PubMedCrossRef
28.
Zurück zum Zitat Tsai R (1987) A versatile camera calibration techniaue for high-accuracy 3D machine vision metrology using off-the-shelf TV cameras and lenses. IEEE J Robotics Automation 3(4):323–344CrossRef Tsai R (1987) A versatile camera calibration techniaue for high-accuracy 3D machine vision metrology using off-the-shelf TV cameras and lenses. IEEE J Robotics Automation 3(4):323–344CrossRef
29.
Zurück zum Zitat Vaerenbergh K et al (2005) Automatic generation of a plan optimization volume for tangential field breast cancer radiation therapy. Strahlenther Onkol 181(2):82–88PubMedCrossRef Vaerenbergh K et al (2005) Automatic generation of a plan optimization volume for tangential field breast cancer radiation therapy. Strahlenther Onkol 181(2):82–88PubMedCrossRef
Metadaten
Titel
Three-dimensional surface scanning for accurate patient positioning and monitoring during breast cancer radiotherapy
verfasst von
C. Gaisberger, MSc
P. Steininger
B. Mitterlechner
S. Huber
H. Weichenberger
F. Sedlmayer
H. Deutschmann
Publikationsdatum
01.10.2013
Verlag
Springer Berlin Heidelberg
Erschienen in
Strahlentherapie und Onkologie / Ausgabe 10/2013
Print ISSN: 0179-7158
Elektronische ISSN: 1439-099X
DOI
https://doi.org/10.1007/s00066-013-0358-6

Weitere Artikel der Ausgabe 10/2013

Strahlentherapie und Onkologie 10/2013 Zur Ausgabe

Alphablocker schützt vor Miktionsproblemen nach der Biopsie

16.05.2024 alpha-1-Rezeptorantagonisten Nachrichten

Nach einer Prostatabiopsie treten häufig Probleme beim Wasserlassen auf. Ob sich das durch den periinterventionellen Einsatz von Alphablockern verhindern lässt, haben australische Mediziner im Zuge einer Metaanalyse untersucht.

Antikörper-Wirkstoff-Konjugat hält solide Tumoren in Schach

16.05.2024 Zielgerichtete Therapie Nachrichten

Trastuzumab deruxtecan scheint auch jenseits von Lungenkrebs gut gegen solide Tumoren mit HER2-Mutationen zu wirken. Dafür sprechen die Daten einer offenen Pan-Tumor-Studie.

Mammakarzinom: Senken Statine das krebsbedingte Sterberisiko?

15.05.2024 Mammakarzinom Nachrichten

Frauen mit lokalem oder metastasiertem Brustkrebs, die Statine einnehmen, haben eine niedrigere krebsspezifische Mortalität als Patientinnen, die dies nicht tun, legen neue Daten aus den USA nahe.

Labor, CT-Anthropometrie zeigen Risiko für Pankreaskrebs

13.05.2024 Pankreaskarzinom Nachrichten

Gerade bei aggressiven Malignomen wie dem duktalen Adenokarzinom des Pankreas könnte Früherkennung die Therapiechancen verbessern. Noch jedoch klafft hier eine Lücke. Ein Studienteam hat einen Weg gesucht, sie zu schließen.

Update Onkologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.