Abstract
The goal of this paper is to describe the metrics used for the evaluation of accuracy of blood glucose (BG) meters for self-monitoring of blood glucose (SMBG) and continuous-glucose monitoring (CGM) system and their limitations and to discuss the current status of SMBG and CGM accuracy. SMBG measurement is used by patients for therapy control and for calculation of appropriate insulin doses for approximately 30 years. The minimum accuracy criteria for SMBG meters are currently defined by ISO 15197:2003 (at least 95 % of results within \(\pm \)20 % or \(\pm \)15 mg/dL of the comparison method measurement results for BG concentrations above or below 75 mg/dL, respectively). In 2013, these accuracy limits were revised in the standard ISO 15197:2013: at least 95 % of results within \(\pm \)15 % or \(\pm \)15 mg/dL for BG above or below 100 mg/dL, respectively. SMBG systems are also used by patients for calibration of CGM systems. Therefore, precision and trueness of the SMBG system are influencing the accuracy of the CGM results. The timing of the BG measurement used for calibration has to be taken into account because, during rapid glucose changes, a time lag exists between BG and the tissue glucose that is measured by CGM systems. The accuracy of CGM devices is often reported by the mean absolute relative deviation (MARD) between CGM results and BG comparison results. This parameter is influenced by different factors like study procedures, glucose fluctuations during the study, and distribution of comparison BG measurements. It is important to define standard study procedures and evaluations to be able to compare MARD results from different studies. For the correct prediction of glucose concentrations, the specific prediction method as well as the accuracy of the CGM system, which may be affected by the accuracy of the SMBG system used for calibration, and the timing of the calibration are important aspects.
Access this chapter
Tax calculation will be finalised at checkout
Purchases are for personal use only
References
American Diabetes Association: self-monitoring of blood glucose. Diabetes Care 17(1), 81–86 (1994). doi:10.2337/diacare.17.1.81. http://care.diabetesjournals.org/content/17/1/81.short
Bailey, T., Zisser, H., Chang, A.: New features and performance of a next-generation SEVEN-day continuous glucose monitoring system with short lag time. Diabetes Technol. Ther. 11(12), 749–755 (2009)
Baumstark, A., Pleus, S., Schmid, C., Link, M., Haug, C., Freckmann, G.: Lot-to-lot variability of test strips and accuracy assessment of systems for self-monitoring of blood glucose according to ISO 15197. J. Diabetes Sci. Technol. 6(5), 1076–1086 (2012)
Brazg, R.L., Klaff, L.J., Parkin, C.G.: Performance variability of seven commonly used self-monitoring of blood glucose systems: clinical considerations for patients and providers. J. Diabetes Sci. Technol. 7(1), 144–152 (2013)
Clinical and Laboratory Standards Institute: performance metrics for continuous interstitial glucose monitoring; approved guideline (2008). http://shopping.netsuite.com/c.1253739/site/Sample_pdf/POCT05A_sample.pdf. Accessed 24 June 14 A.D
Consensus statement on self-monitoring of blood glucose. Diabetes Care 10(1), pp. 95–99 (1987)
Damiano, E.R., El-Khatib, F.H., Zheng, H., Nathan, D.M., Russell, S.J.: A comparative effectiveness analysis of three continuous glucose monitors. Diabetes Care 36(2), 251–259 (2013)
Del Favero, S., Bruttomesso, D., Di Palma, F., Lanzola, G., Visentin, R., Filippi, A., Scotton, R., Toffanin, C., Messori, M., Scarpellini, S., Keith-Hynes, P., Kovatchev, B.P., Devries, J.H., Renard, E., Magni, L., Avogaro, A., Cobelli, C.: First use of model predictive control in outpatient wearable artificial pancreas. Diabetes Care 37(5), 1212–1215 (2014)
Freckmann, G., Baumstark, A., Jendrike, N., Zschornack, E., Kocher, S., Tshiananga, J., Heister, F., Haug, C.: System accuracy evaluation of 27 blood glucose monitoring systems according to DIN EN ISO 15197. Diabetes Technol. Ther. 12(3), 221–231 (2010)
Freckmann, G., Pleus, S., Link, M., Zschornack, E., Klotzer, H.M., Haug, C.: Performance evaluation of three continuous glucose monitoring systems: comparison of six sensors per subject in parallel. J. Diabetes Sci. Technol. 7(4), 842–853 (2013)
Freckmann, G., Baumstark, A., Schmid, C., Pleus, S., Link, M., Haug, C.: Evaluation of 12 blood glucose monitoring systems for self-testing: system accuracy and measurement reproducibility. Diabetes Technol. Ther. 16(2), 113–122 (2014)
Hasslacher, C., Kulozik, F., Platten, I.: Accuracy of self monitoring blood glucose systems in a clinical setting: application of new planned ISO- standards. Clin. Lab. 59(7–8), 727–733 (2013)
Hirose, T., Mita, T., Fujitani, Y., Kawamori, R., Watada, H.: Glucose monitoring after fruit peeling: pseudohyperglycemia when neglecting hand washing before fingertip blood sampling: wash your hands with tap water before you check blood glucose level. Diabetes Care 34(3), 596–597 (2011)
Hortensius, J., Slingerland, R.J., Kleefstra, N., Logtenberg, S.J., Groenier, K.H., Houweling, S.T., Bilo, H.J.: Self-monitoring of blood glucose: the use of the first or the second drop of blood. Diabetes Care 34(3), 556–560 (2011)
International Organization for Standardization: in vitro diagnostic test systems—requirements for blood-glucose monitoring systems for self-testing in managing diabetes mellitus. EN ISO 15197 (2003)
International Organization for Standardization: in vitro diagnostic test systems—requirements for blood-glucose monitoring systems for self-testing in managing diabetes mellitus. ISO 15197 (2013)
Kamath, A., Mahalingam, A., Brauker, J.: Analysis of time lags and other sources of error of the DexCom SEVEN continuous glucose monitor. Diabetes Technol. Ther. 11(11), 689–695 (2009)
Koschinsky, T., Heinemann, L.: Sensors for glucose monitoring: technical and clinical aspects. Diabetes Metab. Res. Rev. 17(2), 113–123 (2001)
Kovatchev, B.P., Renard, E., Cobelli, C., Zisser, H.C., Keith-Hynes, P., Anderson, S.M., Brown, S.A., Chernavvsky, D.R., Breton, M.D., Mize, L.B., Farret, A., Place, J., Bruttomesso, D., Del Favero, S., Boscari, F., Galasso, S., Avogaro, A., Magni, L., Di Palma, F., Toffanin, C., Messori, M., Dassau, E., Doyle, F.J.: Safety of outpatient closed-loop control: first randomized crossover trials of a wearable artificial pancreas. Diabetes Care 37(7), 1789–1796 (2014)
Kuo, C.Y., Hsu, C.T., Ho, C.S., Su, T.E., Wu, M.H., Wang, C.J.: Accuracy and precision evaluation of seven self-monitoring blood glucose systems. Diabetes Technol. Ther. 13(5), 596–600 (2011)
Leelarathna, L., Thabit, H., Allen, J.M., Nodale, M., Wilinska, M.E., Powell, K., Lane, S., Evans, M.L., Hovorka, R.: Evaluating the performance of a novel embedded closed-loop system. J. Diabetes Sci. Technol. 8(2), 267–272 (2014)
Luijf, Y.M., DeVries, J.H., Zwinderman, K., Leelarathna, L., Nodale, M., Caldwell, K., Kumareswaran, K., Elleri, D., Allen, J.M., Wilinska, M.E., Evans, M.L., Hovorka, R., Doll, W., Ellmerer, M., Mader, J.K., Renard, E., Place, J., Farret, A., Cobelli, C., Del Favero, S., Dalla Man, C., Avogaro, A., Bruttomesso, D., Filippi, A., Scotton, R., Magni, L., Lanzola, G., Di Palma, F., Soru, P., Toffanin, C., De Nicolao, G., Arnolds, S., Benesch, C., Heinemann, L.: Day and night closed-loop control in adults with type 1 diabetes: a comparison of two closed-loop algorithms driving continuous subcutaneous insulin infusion versus patient self-management. Diabetes Care 36(12), 3882–3887 (2013)
Nielsen, J.K., Freckmann, G., Kapitza, C., Ocvirk, G., Koelker, K.H., Kamecke, U., Gillen, R., Amann-Zalan, I., Jendrike, N., Christiansen, J.S., Koschinsky, T., Heinemann, L.: Glucose monitoring by microdialysis: performance in a multicentre study. Diabet. Med. 26(7), 714–721 (2009)
Obermaier, K., Schmelzeisen-Redeker, G., Schoemaker, M., Klotzer, H.M., Kirchsteiger, H., Eikmeier, H., del Re, L.: Performance evaluations of continuous glucose monitoring systems: precision absolute relative deviation is part of the assessment. J. Diabetes Sci. Technol. 7(4), 824–832 (2013)
Parkes, J.L., Slatin, S.L., Pardo, S., Ginsberg, B.H.: A new consensus error grid to evaluate the clinical significance of inaccuracies in the measurement of blood glucose. Diabetes Care 23(8), 1143–1148 (2000). doi:10.2337/diacare.23.8.1143. http://care.diabetesjournals.org/content/23/8/1143.abstract
Pfutzner, A., Mitri, M., Musholt, P.B., Sachsenheimer, D., Borchert, M., Yap, A., Forst, T.: Clinical assessment of the accuracy of blood glucose measurement devices. Curr. Med. Res. Opin. 28(4), 525–531 (2012)
Phillip, M., Battelino, T., Atlas, E., Kordonouri, O., Bratina, N., Miller, S., Biester, T., Stefanija, M.A., Muller, I., Nimri, R., Danne, T.: Nocturnal glucose control with an artificial pancreas at a diabetes camp. N. Engl. J. Med. 368(9), 824–833 (2013)
Pleus, S., Schmid, C., Link, M., Zschornack, E., Klotzer, H.M., Haug, C., Freckmann, G.: Performance evaluation of a continuous glucose monitoring system under conditions similar to daily life. J. Diabetes Sci. Technol. 7(4), 833–841 (2013)
Pleus, S., Schmid, C., Link, M., Baumstark, A., Haug, C., Stolberg, E., Freckmann, G.: Accuracy assessment of two novel systems for self-monitoring of blood glucose following ISO 15197:2013. J. Diabetes Sci. Technol. 8(4), 906–908 (2014). doi:10.1177/1932296814536030. http://dst.sagepub.com/content/8/4/906.short
Schmid, C., Haug, C., Heinemann, L., Freckmann, G.: System accuracy of blood glucose monitoring systems: impact of use by patients and ambient conditions. Diabetes Technol. Ther. 15(10), 889–896 (2013)
Thorpe, G.H.: Assessing the quality of publications evaluating the accuracy of blood glucose monitoring systems. Diabetes Technol. Ther. 15(3), 253–259 (2013)
Twomey, P.J.: Plasma glucose measurement with the yellow springs glucose 2300 STAT and the Olympus AU640. J. Clin. Pathol. 57(7), 752–754 (2004)
Zisser, H.C., Bailey, T.S., Schwartz, S., Ratner, R.E., Wise, J.: Accuracy of the SEVEN continuous glucose monitoring system: comparison with frequently sampled venous glucose measurements. J. Diabetes Sci. Technol. 3(5), 1146–1154 (2009)
Zschornack, E., Schmid, C., Pleus, S., Link, M., Klotzer, H.M., Obermaier, K., Schoemaker, M., Strasser, M., Frisch, G., Schmelzeisen-Redeker, G., Haug, C., Freckmann, G.: Evaluation of the performance of a novel system for continuous glucose monitoring. J. Diabetes Sci. Technol. 7(4), 815–823 (2013)
Zueger, T., Diem, P., Mougiakakou, S., Stettler, C.: Influence of time point of calibration on accuracy of continuous glucose monitoring in individuals with type 1 diabetes. Diabetes Technol. Ther. 14(7), 583–588 (2012)
Author information
Authors and Affiliations
Corresponding author
Editor information
Editors and Affiliations
Rights and permissions
Copyright information
© 2016 Springer International Publishing Switzerland
About this chapter
Cite this chapter
Freckmann, G., Pleus, S., Link, M., Haug, C. (2016). Accuracy of BG Meters and CGM Systems: Possible Influence Factors for the Glucose Prediction Based on Tissue Glucose Concentrations. In: Kirchsteiger, H., Jørgensen, J., Renard, E., del Re, L. (eds) Prediction Methods for Blood Glucose Concentration. Lecture Notes in Bioengineering. Springer, Cham. https://doi.org/10.1007/978-3-319-25913-0_2
Download citation
DOI: https://doi.org/10.1007/978-3-319-25913-0_2
Published:
Publisher Name: Springer, Cham
Print ISBN: 978-3-319-25911-6
Online ISBN: 978-3-319-25913-0
eBook Packages: EngineeringEngineering (R0)