Safety of intrauterine contraceptive devices during MR imaging

  • Magnetic Resonance Imaging
  • Published:
European Radiology Aims and scope Submit manuscript

Abstract

Three intrauterine contraceptive devices (IUD) were examined in an experimental in vitro set-up to test their MR compatibility: Multiload CU375, Nova T (containing copper and silver), and Gyne T. The devices were fixed in a polycrylamide gel and exposed to the magnetic field and the radiofrequence (RF) of standard MR sequences (spin-echo, turbo-spin-echo, turboFLASH, and magnetization transfer prepared FLASH). The RF power of a turbo-spin-echo sequence was increased by a factor of 5. Temperatures of the IUDs and the gel were measured during the MR examination. No deflection of the IUDs exposed to the magnetic field of 1.5 Tesla was detected. Under normal imaging conditions no temperature increase could be observed. Using an increased RF power, a maximal temperature rise of 0.4°C per examination was seen. No differences between the gel and the IUDs were observed, indicating that there is no specific heating of the tested IUDs.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Similar content being viewed by others

References

  1. Brown MA, Carden JA, Coleman RE et al. (1987) Magnetic field effects on surgical ligation clips. Magn Reson Imaging 5: 443

    Google Scholar 

  2. Davis PL, Crooks L, Arakawa M et al. (1981) Potential hazards in NMR imaging: heating effects of changing magnetic fields and RF fields on small metallic implants. AJR 137: 857

    Google Scholar 

  3. Gold JP, Pulsinelli W, Winchester P et al. (1989) Safety of metallic surgical clips in patients undergoing high-field-strength magnetic resonance imaging. Ann Thorac Surg 48: 643

    Google Scholar 

  4. Kagetsu NJ, Litt AW (1991) Important considerations in measurement of attractive force on metallic implants in MR imagers. Radiology 179: 505

    Google Scholar 

  5. Kanal E, Shellock FG, Talagala L (1990) Safety considerations in MR. Radiology 176: 593

    Google Scholar 

  6. Mark AS, Hricak H (1987) Intrauterine contraceptive devices: MR imaging. Radiology 162(2): 311

    Google Scholar 

  7. Mesgarzadeh M, Revesz G, Bonakdarpour A et al. (1985) The effect on medical metal implants by magnetic fields of magnetic resonance imaging. Skeletal Radiol 14: 205

    Google Scholar 

  8. New PFJ, Rosen BR, Brady TJ et al. (1983) Potential hazards and artifacts of ferromagnetic and non ferromagnetic surgical and dental materials and devices in nuclear magnetic resonance imaging. Radiology 147:139

    Google Scholar 

  9. Shellock FG, Crues JV (1988) High-field-strength MR imaging and metallic biomedical implants: an ex vivo evaluation of deflection forces. AJR 151: 389

    Google Scholar 

  10. Shellock FG (1988) MR imaging of metallic implants and materials: a compilation of the literature. AJR 151: 811

    Google Scholar 

  11. Shellock FG, Litwer C, Kanal E (1992) MRI bioeffects, safety, and patient management: a review. Rev Magn Reson Imaging 4: 21

    Google Scholar 

  12. Shellock FG, Curtis JS (1991) MR imaging and biomedical implants, materials, and devices: an updated review. Radiology 180: 541

    Google Scholar 

  13. Shellock FG, Morisoli S, Kanal E (1993) MR procedures and biomedical implants, materials, and devices: 1993 update. Radiology 189: 587

    Google Scholar 

Download references

Author information

Authors and Affiliations

Authors

Rights and permissions

Reprints and permissions

About this article

Cite this article

Hess, T., Stepanow, B. & Knopp, M.V. Safety of intrauterine contraceptive devices during MR imaging. Eur. Radiol. 6, 66–68 (1996). https://doi.org/10.1007/BF00619957

Download citation

  • Received:

  • Revised:

  • Accepted:

  • Issue Date:

  • DOI: https://doi.org/10.1007/BF00619957

Key words

Navigation