Accuracy and cross-sensitivity of 10 different anesthetic gas monitors

  • Original Articles
  • Published:
Journal of Clinical Monitoring Aims and scope Submit manuscript

Abstract

Objective. The objective of this study was to test the accuracy and cross-sensitivity of commercially available anesthetic gas monitors.Methods. Using gas chromatography (GC) as a reference method, the accuracy, cross-sensitivity, and ability to recognize an erroneously selected agent were determined in the following 10 monitors for volatile anesthetics: Datex Capnomac Ultima-S, Datex Capnomac, Ohmeda 5330 agent monitor, Iris Dräger, Andros Dräger PM 8020 (all monochromatic, infrared analyzers), Nellcor N-2500E, Criticare POET II, Irina Dräger (all polychromatic, infrared analyzers), Siemens Servo Gas Monitor 120 (a piezoelectric analyzer), and Brüel & Kjaer Type 1304 (a photoacoustic analyzer). Accuracy was determined at 0.5, 1, 2, and 4 times the minimal alveolar concentration (MAC) of either halothane or isoflurane in oxygen (O2). The cross-sensitivity tests were performed with 70 vol% nitrous oxide in O2, 5 vol% carbon dioxide in O2, 0.032 vol% alcohol in O2, and 70% water vapor in O2. The photoacoustic analyzer showed a higher accuracy for isoflurane than the polychromatic infrared monitors. The greatest inaccuracy with isoflurane was found in the Iris Dräger monitor, which had a maximal bias percentage by volume (vol%) of 0.09 at 0.5 MAC. (This bias was within the manufacturer's specified tolerance of ±0.1 vol% or 10% relative difference of reading, whichever is greater.) Irina Dräger was the most accurate analyzer with halothane (mean % bias [relative %] ± SD, 0.9 ± 2.0%). The greatest bias with halothane was found in the monochromatic infrared analyzers, with a maximal % bias at 0.5 MAC of 50.3% of the GC reading (12.4% with a new inner Nafion tube) found in the Datex Ultima monitor. The Siemens gas monitor showed a cross-sensitivity for water vapor (−0.248 vol%). The monochromatic infrared analyzers showed a small sensitivity to alcohol (additional deviation of 0.011 to 0.147 vol% at 2 MAC isoflurane) but no sensitivity to nitrous oxide. No cross-sensitivity was found in the polychromatic infrared and photoacoustic analyzers. An incorrect selection of anesthetic agent when using a monochromatic infrared analyzer can be fatal; for example, when using halothane and selecting isoflurane the values measured by the Datex Capnomac monitor were nearly 6 times below the actual value (i.e., 1 vol% “isoflurane” on the display = 6 vol% halothane in reality).Conclusions. The photoacoustic measurement principle is more accurate than the other methods, although the polychromatic infrared analyzers are safer because they detect erroneously selected agents.

Résumé

Objectifs. Le but de cette étude était deverifier l'exactitude et la sensibilité croisée des moniteurs de gaz anesthésiques disponibles sur le marché.Méthodes. En choisissant la chromatographie gazeuse (CG) comme méthode de référence, nous avons déterminé l'exactitude, la sensibilité croisée, et la capacité de reconnaissance d'un agent sélectionné par erreur, chez les 10 moniteurs d'anesthésiques volatils suivants: Datex Capnomac Ultima-S, Datex Capnomac, Ohmeda 5330, Iris Drager, Andros Drager PM 8020 (analyseurs infra-rouges monochromatiques); Nelicor N-2500E, Criticare POET II, Irina Drager (analyseurs infra-rouges polychromatiques), Siemens Servo Gas Monitor 120 (analyseur piezoélectrique), et Bruel & Kjaer Type 1304 (analyseur photoacoustique). L'exactitude a été déterminée à respectivement 0,5; 1; 2; 4 fois la concentration alvéolaire minimale (MAC) soit d'halothane, soit d'isoflurane sous oxygène. Les tests de sensibilité croisée ont été réalisés avec un mélange gazeux contenant de l'oxygène et, respectivement, 70% de protoxyde d'azote, 5% de dioxyde de carbone, 0,032% d'alcool, et 70% de vapeur d'eau. L'analyseur photoacoustique s'est révèlé d'une plus grande exactitude pour l'isoflurane que les moniteurs infra-rouges polychromatiques. La plus grande inexactitude avec l'isoflurane était obtenue avec le moniteur Iris Drager, qui présentait l'erreur maximale en pourcentage de volume (vol %) de 0,09 à 0,5 MAC. Cette erreur était située dans l'intervalle de tolérance spécifié par le constructeur de 0,1 vol% en valeur absolue ou de 10% d'erreur relative. Irina Drager était l'analyseur le plus précis avec l'halothane (erreur moyenne ± écart-type de 0,9 ± 2,0 %). L'erreur la plus importante avec l'halothane était obtenue avec les analyseurs infra-rouges monochromatiques, avec une erreur relative maximale 0,5 MAC de 50.3% des résultats de la CG (12,4% avec le nouveau tube Nafion) obtenue avec le moniteur Ultima de Datex. Le moniteur Siemens a montré une sensibilité croisée pour la vapeur d'eau (−0,248 vol%). Les analyseurs infra-rouges monochromatiques ont montré une petite sensibilité à l'alcool (déviation de 0,011 à 0,147 vol% à 2 MAC d'isoflurane) mais aucune sensibilité au protoxyde d'azote. Aucune sensibilité croisée n'était obtenue chez les analyseurs polychromatiques infrarouges et photo-acoustiques. Une sélection incorrecte d'agent anesthésique d'un agent anesthésique lors de l'utilisation d'un analyseur monochromatique infra-rouge peut être fatal; par exemple, en cas d'utilisation d'halothane et de sélection de l'isoflurane, les valeurs mesurées par le moniteur Capnomac de Datex étaient près de 6 fois inférieures aux valeurs réelles (c'est-à-dire: 1 vol% d' “isoflurane” affiché = 6 vol% d'halothane en réalité).Conclusions. Le principe de mesure photoacoustique est plus exacte que les autres méthodes, bien que les analyseurs infra-rouges polychromatiques sont plus sūrs puisqu'ils détectent les agents sélectionnés par erreur.

Abstrakt

Zielsetzung. Ziel der Untersuchung war es, die Genauigkeit und Kreuzempfindlichkeit kommerziell erhältlicher Narkosegasmonitore zu überprüfen. Ein Gaschromatograph (GC) wurde als Referenz verwendet, um die Genauigkeit, Kreuzempfindlichkeit und die Fähigkeit fälschlich gewāhlter Narkosegase zu erkennen, an 10 verschiedenen Monitoren überprüft: Datex Capnomac® Ultima -S, Datex Capnomac®, Ohmreda 5330® Nakosegasmonitor, Iris® Dräger, Andros® Dräger PM8020 (alle monochromatische Infrarotanalysesysteme), Nellcor® N-2500E, Criticare® POET II, Irina® Dräger (alle polychromatische Infrarotanalysesysteme), Siemens Servo® Gasmonitor 1320 (ein piezoelektrisches System) und das Brüel & Kjaer® Typ 1304 ein photakustische Analysesystem. Die Genauigkeit wurde bei der 0.5–1–2 und 4fachen minimalen alveolären Konzenztration (MAC) in entweder Halothan oder Isofluran in Sauerstoff bestimmt. Eine Kreuzempfindlichkeit wurde mit 70 Vol% Lachgas in Sauerstoff, 5 Vol% Kohlendioxyd in Sauerstoff, 0.032 Vol% Alkohol in Sauerstoff und 70 Vol% Wasserdampf in Sauerstoff untersucht.Ergebnisse. Das photoakustische System wies grōssere Genauigkeit für Isofluran auf als alle polychromatischen Infrarotgasmonitore Die grösste ungenauigheit wurde beim Dräger Iris® Monitor für Isofluran gefunden, der eine prozentuale Abweichung von 0.09 Vol% bei 0.5 MAC aufwies. Diese Abweichung lag innerhalb der vom Hersteller angegebenen Toleranzgrenze von ± 0.1 Vol% bzw. relative10%. Der Irina® Drägermonitor war der genauste für Halothan (mittlere % Abweichung [relative %] ± Stabdardabweichung, 0.9 Vol% ± 2.0%). Die größte Abweichung für Halothan wurde bei den monochromatischen Systemen gefunden; bei 0.5 MAC wurde eine maximale % Abweichung für den Datex Ultima® Monitor von 50.3% (12.4% mit dem neuen Nafionschlauch) von dem am Gaschromatographen gemessenen Wert gefunden. Der Siemens Gasmonitor wies eine Kreuzempfindlichkeit für Wasserdampf auf (−0.248 Vol%). Die monochromatischen Infrarotsysteme wiesen alle eine geringe Kreuzempfindlichkeit für Alkohol (zusätzliche Abweichung von 0.011 bis 0.147 Vol% bei 2 MAC Isofluran) jedoch keine Kreuzempfindlichkeit für Lachgas auf. Keine Kreuzempfindlichkeit wurde bei den polychromatischen Infrarot- und photoakustischen Systemen nachgewiesen. Eine fehlerhafte Auswahl zum verwendeten Narkosegas kann bei einem monochromatischen Infrarot-Narkosegasmonitor fatale Folgen haben, wenn z.B. beim Gebrauch von Halothan und der Wahleinstellung Isofluran, die mit dem Datex Capnomat® gemessenen Werte bis zum 6fachen unterhalb der eigentlichen Konzentration liegen (i.e. 1 Vol% Isofluran in der Anzeige bei eigentlichen 6 Vol% Halothan).Schlußfolgerung. Das photoakustische Messprinzip ist genauer als alle anderen Methoden, obgleich das polychromatischen System sicherer ist, da hiermit fälschlicherweise Narkosegase gewählte entdeckt werden können.

Resumen

Objetivo. El objetivo de este estudio fue determinar la precisión y sensibilidad cruzada de monitores de gases anestésicos disponibles en el comercio.Metodos. Usando cromatografía de gases (GC) como método de referencia, se determinó la precisión, sensibilidad cruzada, y capacidad de reconocer un agente seleccionado en forma errónea, de los siguientes diez monitores de agentes anestésicos volátiles: Datex Capnomac Ultima S, Datex Capnomac, Ohmeda 5330, Iris Drager, Andros Drager PM 8020 (todos analizadores infrarrojos monocromáticos), Nellcor N 2500E, Criticare POET II, Irina Drager (todos analizadores infrarrojos policromáticos), Siemens Servo Gas Monitor 120 (analizador piezoeléctrico), y el Bruel & Kjaer Tipo 1304 (analizador fotoacústico). La precisión fue determinada a 0.5, 1, 2, y 4 veces la concentración alveolar mínima (MAC) de halothane o isoflurane en oxígeno (O2). Las pruebas de sensibilidad cruzada fueron realizadas con 70 vol% de óxido nitroso en O2, 5 vol% de CO2 en O2, 0.032 vol% alcohol in O2 y 70% vapor de agua en O2. El analizador fotoacústico mostró mayor precisión para el isofluorano que los monitores infrarrojos policromáticos. La mayor falta de precisión para el isoflurane se observó en el monitor Iris Drager, cuyo sesgo máximo en porcentaje por volumen (vol%) fue 0.09 at 0.5 MAC. (Este sesgo estaba dentro de la tolerancia especificada por el fabricante, de ±0.1 vol% o 10% diferencia relativa de lectura, lo que resultare mayor). El analizador Irina Drager fue el más preciso con halothane (sesgo % medio [% relativo] ± SD, 0.9 ± 2.0%). El mayor sesgo con halothane se observó en los analizadores infrarrojos monocromáticos, con un sesgo máximo a 0.5 MAC de 50.3% de la lectura del GC (12.4% con un tubo interior de Nafion nuevo) en el monitor Datex Ultima. El monitor de gases Siemens demostró sensibilidad cruzada para el vapor de agua (0.248 vol%). Los analizadores infrarrojos monocromáticos mostraron escasa sensibilidad al alcohol (desviación adicional de 0.011 a 0.147 vol% a 2 MAC isoflurane), pero no sensibilidad al óxido nitroso. No se detectó sensibilidad cruzada a los analizadores infrarrojos policromáticos ni fotoacústico. Una selección incorrecta de agente anestésico, usando un analizador infrarrojo monocromático, puede ser fatal: Por ejemplo, al usar halothano y seleccionar isofluorano los valores medidos por el monitor Datex Capnomac eran casi 6 veces por debajo del valor real (i.e., 1 vol% “isoflurane” en el display = 6 vol% halothane en la realidad).Conclusiones. El principio fotoacústico de medición es más preciso que los otros métodos, si bien los analizadores infrarrojos policromáticos son más seguros porque detectan selección errónea de agente anestésico.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Similar content being viewed by others

References

  1. Eger EI II, Saidman LJ, Brandstater B. Minimum alveolar anesthetic concentration: a standard of anesthetic potency. Anesthesiology 1965;26:756–763

    PubMed  Google Scholar 

  2. Carpenter RL, Eger EI II. Alveolar-to-arterial-to-venous anesthetic partial pressure differences in humans. Anesthesiology 1989;70:630–635

    Article  PubMed  CAS  Google Scholar 

  3. Lin CY. Assessment of vaporiser performance in low flow and closed circuit anesthesia. Anesth Analg 1980;50:359–366

    Google Scholar 

  4. Buckinham JD, Holme AE. Mass spectrometry for respiratory gas analysis. Br J Clin Equip 1977;2:142–148

    Google Scholar 

  5. Luff NP, White DC. Evaluation of the Datex Normac anesthetic agent monitor. Anaesthesia 1985;40:555–559

    Article  PubMed  CAS  Google Scholar 

  6. McPeak H, Palayiwa E, Madgwick R, et al. Evaluation of a multigas anaesthetic monitor: the Datex Capnomac. Anaesthesia 1988;43:1035–1041

    PubMed  CAS  Google Scholar 

  7. Cooper JB, Edmondson JH, Joseph DM, et al. Piezoelectric sorption anesthetic sensor. IEEE Trans Biomed Eng 1981;6:BME 28

    Google Scholar 

  8. Gedeon A, Hamilton K, Kindlund A, et al. A new sensitive method for breath by breath monitoring of anesthetic gases. Anesthesiology 1980;9:14–21

    Google Scholar 

  9. Kay B, Cohen AT, Wheeler MF. A laboratory investigation of a multigas monitor for anaesthesia (EMMA). Anaesthesia 1982;37:446–450

    Article  PubMed  CAS  Google Scholar 

  10. Westenskow DR, Smith KW, Coleman DL, et al. Clinical evaluation of a RAMAN scattering multiple gas analyzer for the operating room. Anesthesiology 1989;70:350–355

    Article  PubMed  CAS  Google Scholar 

  11. Westenskow DR, Coleman DL. Can the RAMAN scattering analyzer compete with mass spectrometers? An affirmative reply. J Clin Monit 1989;5:34–36

    Article  PubMed  CAS  Google Scholar 

  12. Nielsen J, Kann T, Moller JT. Evaluation of two newly developed anesthetic agent monitors: Brüel & Kjaer anesthetic agent monitor 1304 and Datex capnomac ultima. Anesthesiology 1990;73:A538

    Google Scholar 

  13. Schütz H. Alkohol im Blut. Weinheim: Verlag Chemie, 1983:6–8

    Google Scholar 

  14. Zbinden AM, Frei FJ, Funk B, et al. Determination of halothane (or isoflurane) in blood. Br J Anaesth 1985;57:796–802

    Article  PubMed  CAS  Google Scholar 

  15. Merkel G, Eger EI II. A comparative study of halothane and halopropane anesthesia. Including method for determining equipotency. Anesthesiology 1963;24:346–357

    Article  PubMed  CAS  Google Scholar 

  16. Saidman LJ, Eger EI II. Effect of nitrous oxide and of narcotic premedication on the alveolar concentration of halothane required for anesthesia. Anesthesiology 1964;25:302–306

    Article  PubMed  CAS  Google Scholar 

  17. Stevens WC, Dolan WM, Gibbons RT, et al. Minimum alveolar concentration (MAC) of isoflurance with and without nitrous oxide in patients of various ages. Anesthesiology 1975;42:197–200

    PubMed  CAS  Google Scholar 

  18. Comité Européen de Normalisation (1991) CEN/TC 215/WG1 N71. Anesthetic Gas Monitors. Particular Requirements. Version 0.3, April 17, 1991

  19. Mollgaard K. Acoustic gas measurement. Biomed Instrument Technol 1989;23:495–497

    CAS  Google Scholar 

  20. Severinghaus JW. Water vapor errors in some capnometers. Respiratory conventions misunderstood by manufacturers. Anesthesiology 1989;70:996–998

    Article  PubMed  CAS  Google Scholar 

  21. Lindstromberg JW, Muir JJ. Cross-sensitivity in water vapor in the Engström EMMA. Anesth Analg 1984;63:75–78

    Google Scholar 

  22. Foley MA, Wood PR, Peel WJ, et al. The effect of exhaled alcohol on the performance of the Datex Capnomac. Anaesthesia 1990;45:232–234

    Article  PubMed  CAS  Google Scholar 

  23. Guyton DC, Gravenstein N. Infrared analysis of volatile anesthetics: impact of monitor agent setting, volatile mixtures, and alcohol. J Clin Monit 1990;6:203–206

    Article  PubMed  CAS  Google Scholar 

  24. Zbinden AM, Westenskow D, Thomson DA, et al. A laboratory investigation of two new portable gas analysers. Int J Clin Monit Comput 1986;2:151–161

    Article  PubMed  CAS  Google Scholar 

  25. Baras E, Benmosbah L, Deriaz H, et al. Erroneous selection of the agent on a monochromatic infrared analyzer: which displayed concentrations? Anesthesiology 1989;71:3A-A449

    Article  Google Scholar 

Download references

Author information

Authors and Affiliations

Authors

Rights and permissions

Reprints and permissions

About this article

Cite this article

Walder, B., Lauber, R. & Zbinden, A.M. Accuracy and cross-sensitivity of 10 different anesthetic gas monitors. J Clin Monitor Comput 9, 364–373 (1993). https://doi.org/10.1007/BF01618679

Download citation

  • Received:

  • Revised:

  • Accepted:

  • Issue Date:

  • DOI: https://doi.org/10.1007/BF01618679

Key words

Navigation