Abstract
Objective
Commercially available semi-continuous cardiac output (SCCO) monitoring systems are based on the pulsed warm thermodilution technique. There is evidence that SCCO fails to correlate with standard intermittent bolus cardiac output (ICO) in clinical situations with thermal instability in the pulmonary artery. Furthermore, ventilation may potentially influence thermodilution measurements by enhanced respiratory variations in pulmonary artery blood temperature and by cyclic changes in venous return. Therefore, we evaluated the correlation, accuracy and precision of SCCO versus ICO measurements before and after extubation.
Design
Prospective cohort study.
Setting
Intensive care unit (ICU) of a university hospital.
Patients and participants
22 cardiac surgical ICU patients.
Interventions
None.
Measurements and results
SCCO and ICO data were obtained at nine postoperative time points while the patients were on controlled mechanical ventilation. Further sets of measurements were taken during the weaning phase 20 min before extubation, and 5 min, 20 min and 1 h after extubation. SCCO and ICO measurements yielded 286 data pairs with a range of 1.8–9.9 l/min for SCCO and 1.9–9.8 l/min for ICO. The correlation between SCCO and ICO was highly significant (r=0.92;p<0.01), accompanied by a bias of −0.052 l/min and a precision of 0.56 l/min. Correlation, accuracy and precision were not influenced by the mode of respiration.
Conclusions
Our results demonstrate excellent correlation, accuracy and precision between SCCO and ICO measurements in postoperative cardiac surgical ICU patients. We conclude that SCCO monitoring offers a reliable clinical method of cardiac ouput monitoring in ICU patients following cardiac surgery.
Article PDF
Similar content being viewed by others
References
Mermel LA, Maki DG (1994) Infectious complications of Swan-Ganz pulmonary artery catheters. Am Rev Respir Crit Care Med 149:1020–1036
Lavett JM, Reploge RL (1979) Thermodilution cardiac output: a critical analysis and review of the literature. J Surg Res 27:392–404
Renner RE, Morton MJ, Sakuma GY (1993) Indicator amount, temperature, and intrinsic cardiac output affect thermodilution cardiac output accuracy and reproducebility. Crit Care Med 21:586–597
Keinänen O, Takala J, Kari A (1992) Continuous measurement of cardiac output by the Fick principle: clinical validation in intensive care. Crit Care Med 20:360–365
Hausen B, Schäfers HJ, Rohde R, Haverich A (1992) Clinical evaluation of transtracheal Doppler for continuous cardiac output estimation. Anesth Analg 74:800–804
Perrino AC, O'Connor T, Luther M (1994) Transtracheal Doppler cardiac output monitoring: comparison to thermodilution during noncardiac surgery. Anesth Analg 78:1060–1066
Tannenbaum GA, Mathews D, Weissman C (1993) Pulse contour cardiac output in surgical intensive care unit patients. J Clin Anesth 5:471–478
Balestra B, Malacrida R, Leonardi L, Suter P, Marone C (1992) Esophageal electrodes allow precise measurement of cardiac output by bioimpedance. Crit Care Med 20:62–67
Shoemaker WC, Wo CCJ, Bishop MH, Appel PL, Van de Water JM, Harrington GR, Wang X, Patil RS (1994) Multicenter trial of a new thoracic electrical bioimpedance device for cardiac output estimation. Crit Care Med 22:1907–1912
Segal J, Gaudiani V, Nishimura T (1991) Continuous determination of cardiac output using a flow-directed Doppler pulmonary artery catheter. J Cardiothorac Vasc Anesth 5:309–315
Baum VC, Chait HI, Williams JP (1994) Continual measurement of cardiac output during cardiac surgery by heat exchange (abstract). Anesthesiology 81 [3 A]:A516
Lichtenthal PR, Marchand B, Gordon D, Leissing N, Konno M (1992) A safety comparison between a new continuous cardiac output (CCO) monitoring system and a standard pulmonary artery catheter in sheep (abstract). Anesthesiology 77[3A]:A473
Yelderman M, Quinn MD, McKown RC (1992) Thermal safety of a filamented pulmonary artery catheter. J Clin Monit 8:147–149
Yelderman M (1990) Continuous measurement of cardiac output with the use of stochastic system identification techniques. J Clin Monit 6:322–332
Yelderman M (1993) Continuous cardiac output by thermodilution. Int Anesthesiol Clin 31:127–140
Davis RF, Sakuma G (1992) Comparison of semi-continuous thermodilution to intermittent bolus thermodilution cardiac output determinations (abstract). Anesthesiology 77[3A]:A477
Lichtenthal PR, Wade LD (1993) Clinical evaluation of a continuous cardiac output system in post-OP cardiac surgical patients (abstract). Crit Care Med 21 [Suppl]:S214
Schmid ER, Tornic M (1994) Accuracy of continuous cardiac output monitoring by thermodilution (abstract). Anesthesiology 81 [3A]:A519
Ryan MK, Fan YP, Lee TS, Bongard FS (1993) Comparison of continuous vs manual bolus cardiac output following hemodynamic alterations in a porcine model (abstract). Anesthesiology 79[3A]:A469
Yelderman M, Quinn MD, McKown RC (1992) Continuous thermodilution cardiac output measurement in sheep. J Thorac Cardiovasc Surg 104:315–320
Böttiger BW, Rauch H, Böhrer H, Motsch J, Soder M, Fleischer F, Martin E (1995) Continuous versus intermittent cardiac output measurement in cardiac surgical patients undergoing hypothermic cardiopulmonary bypass. J Cardiothorac Vasc Anesth 9:505–411
Latson TW, Whitten CW, O'Flaherty D (1993) Ventilation, thermal noise, and errors in cardiac output measurements after cardiopulmonary bypass. Anesthesiology 79:1233–1243
Spackman TN, Abenstein JP (1993) Continuous cardiac output may be more accurate than bolus thermodilution output during the use of an upperbody warming blanket (abstract). Anesthesiology 79[3A]:A473
Wessel HU, James GW, Paul MH (1966) Effects of respiration and circulation on central blood temperature of the dog. Am J Physiol 211:1403–1412
Nishikawa T, Dohi S (1993) Errors in the measurement of cardiac output by thermodilution. Can J Anaesth 40:142–153
Stetz CW, Miller RG, Kelly GE (1982) Reliability of the thermodilution method in the determination of cardiac output in clinical practice. Am Rev Respir Dis 126:1001–1004
Boyd O, Mackay J, Newman P, Bennett ED, Grounds RM (1994) Effects of insertion depth and use of the sidearm of the introducer sheath of pulmonary artery catheters in cardiac output measurement. Crit Care Med 22:1132–1135
Wetzel RC, Latson TW (1985) Major errors in thermodilution cardiac output measurement during rapid volume infusion. Anesthesiology 62:684–687
Bland JM, Altman DG (1986) Statistical methods for assessing agreement between two methods of clinical measurement. Lancet I:307–310
Yelderman ML, Ramsay MA, Quinn MD, Paulsen AW, McKown RC, Gillman PH (1992) Continuous thermodilution cardiac output measurement in intensive care unit patients. J Cardiothorac Vasc Anesth 6:270–274
Weisel RD, Berger RL, hechtman HB (1975) Measurement of cardiac output by thermodilution. N Engl J Med 292:682–684
Hogue CW, Cerza RF, Rosenbloom M, Lappas DG (1993) Comparison of continual thermodilution cardiac output with electromagnetometry (abstract). Anesthesiology 79[3A]:A467
Armengol J, Man GCW, Balsys AJ, Wells AL (1981) Effects of the respiratory cycle on cardiac output measurements: reproducibility of data enhanced by timing the thermodilution injections in dogs. Crit Care Med 9:852–854
Snyder JV, Powner DJ (1982) Effects of mechanical ventilation on the measurement of cardiac output by thermodilution. Crit Care Med 10:677–682
Stevens JH, Raffin TA, Mihm FG et al. (1985) Thermodilution cardiac output measurement. Effects of the respiratory cycle on its reproducibility. JAMA 253:2240–2242
Thrush DN, Varlotta D (1992) Thermodilution cardiac output. Comparison between automated and manual injection of indicator. J Cardiothorac Vasc Anesth 6:17–19
Woods M, Scott RN, Harken AH (1976) Practical considerations for the use of a pulmonary artery thermistor catheter. Surgery 79:469–475
Jansen JRC, Versprille A (1986) Improvement of cardiac output estimation by the thermodilution method during mechanical ventilation. Intensive Care Med 12:71–79
Jansen JRC, Schreuder JJ, Settels JJ, Kloek JJ, Versprille A (1990) An adequate strategy for the thermodilution technique in patients during mechanical ventilation. Intensive Care Med 16:422–425
Ham TH, Shen SC, Fleming EM, Castle WB (1948) Studies in destruction of red blood cells. IV. Thermal injury. Blood 3:373–403
Author information
Authors and Affiliations
Rights and permissions
About this article
Cite this article
Böttiger, B.W., Soder, M., Rauch, H. et al. Semi-continuous versus injectate cardiac output measurement in intensive care patients after cardiac surgery. Intensive Care Med 22, 312–318 (1996). https://doi.org/10.1007/BF01700452
Received:
Accepted:
Issue Date:
DOI: https://doi.org/10.1007/BF01700452