Abstract
The procedure described is based on the acquisition and processing of x-rays of the distal radius obtained under standard conditions. An x-ray was obtained of the forearm together with an aluminum step wedge to automatically normalize the photometric values of the bone with respect to the photometric values of the reference aluminum wedge. Densitometric values for thickness (T) and a coarseness parameter (C) that depends on the trabecular bone pattern are measured on interactively selected rows and regions of interest (ROIs) of the digital image. Twenty-five women were examined and two different measurements were performed. The first measurement considers C in three sites of the radial epiphisis. The trabecular bone coarseness appears to increase from the distal to the very-very distal site and the value of C in the very distal site, which is located 1 cm distally to the distal one tenth of the radius, seems to be related to the pathological variations more than the value of C in the other sites. The second measurement is the C/T ratio of eight ROIs of 15 patients: five healthy and 10 osteoporotic women. This ratio is significantly different for the two groups in all the eight ROIs and the variations are particularly significant at 6 to 12 mm from the subchondral line.
Article PDF
Similar content being viewed by others
Avoid common mistakes on your manuscript.
References
Cann EC: Skeletal structure-function revisited. Radiology 179:607–608, 1991
Griffiths HJ, Virtama P: Cortical thickness and trabecular pattern of the femoral neck as a measure of osteopenia. Invest Radiol 25:1116–1119, 1990
Phillips HB, Owen JS, Chandler B: Quantitative histology of bone: A computerized method for measuring the total mineral content of bone. Calcif Tiss Res 26:85–89, 1978
Colbert C, Mazess RB, Schmidt PB: Bone mineral determination in vitro by radiographic photodensitometry and direct photon absorptiometry. Invest Radiol 5:336–340, 1970
Rockoff SD, Scandrett J, Zacher R: Quantization of relevant image information: Automated radiographic bone trabecular characterization. Radiology 101:435–439, 1971
Durand EP, Ruegsegger P: Cancellous bone structure: Analysis of high-resolution CT images with the run-length method. J Comp Assist Tomogr 15:133–139, 1991
Trouerbach WT, Grashuis JL, Zwamborn AW, et al: Microdensitometric analysis of bone structures in x-ray images. Skeletal Radiol 16:190–195, 1987
Hesp R, Deacon AC, Hulme P, et al: Trends in trabecular and cortical bone in the radius compared with whole body calcium balance in osteoporosis. Clin Sci 66:109–112, 1984
Nilas L, Borg J, Gotfredsen A, et al: Comparison of single and dual photon absorptiometry in postmenopausal bone mineral loss. J Nucl Med 26:1257–1262, 1985
Kalebo P, Strid KG: Bone mass determination from microradiographs by computer-assisted videodensitometry: II-Aluminium as a reference substance. Acta Radiol 29:611–617, 1988
Bevington PR: Data Reduction and Error Analysis for the Physical Sciences. New York, NY, McGraw-Hill, 1969
Lang P, Steiger P, Faulkner K, et al: Osteoporosis. Current techniques and recent developments in quantitative bone densitometry. Radiol Clin North Am 29:49–76, 1991
Johns HE, Cunningham JR: The Physics of Radiology. Springfield, IL, Thomas, 1983
Chimenti M, Trippi D, Bozzi R, et al: Computer characterization of trabecular bone patterns, in Lemke HU, et al (eds): Proc CAR '89. New York, NY, Springer, 1989
Nilas L, Nørgaard H, Pødenphant J, et al: Bone composition in the distal forearm. Scan J Clin Lab Invest 47:41–46, 1987
Schlenker RA: Percentages of cortical and trabecular bone mineral mass in the radius and ulna. in Mazess RB (ed): Third Int Conf Bone Mineral Measurement. AJR Am J Roentgenol 126:1309–1312, 1976
Author information
Authors and Affiliations
Rights and permissions
About this article
Cite this article
Trippi, D., Chimenti, M. & Bozzi, R. A computer-assisted method for the study of the trabecular bone of the distal radius on conventional radiographs. J Digit Imaging 6, 140–147 (1993). https://doi.org/10.1007/BF03168440