Introduction

Computed tomography (CT) was introduced in the early 1970s and has since then revolutionized diagnostic radiology. It was the first medical imaging machine to use a computer, in this way pioneering the era of digital imaging. CT technology has significantly improved and matured in the past 40 years. Today, CT is the backbone of radiology. In this article we review the different generations of CT system designs, from the first dedicated head scanners to modern multi detector row CTs (MDCT). We introduce novel system concepts, such as CT systems with area detectors and dual source CTs, that aim at solving remaining limitations of MDCT. We explain the basic system options to acquire dual energy CT data or spectrally resolved CT data. Finally, we briefly touch alternative system designs such as electron beam CT (EBCT), CT systems with inverse geometry, interior tomography CT or phase contrast CT.

Basic Principles of CT

In X-ray CT, local X-ray attenuation coefficients are measured in a slice of the patient’s anatomy. They are translated into grey-scale values and displayed in the CT image, usually with a matrix of 512 by 512 pixels. To compute the local X-ray attenuation coefficients, the patient’s X-ray absorption is measured along narrow X-ray beams in many different directions within the desired image plane. A detector records the attenuated intensity of each transmitted X-ray beam after it has passed through the patient’s body—the logarithm of the attenuated intensity is the line-integral of the patient’s attenuation coefficients along the beam path. Multiple line integrals across an object can be acquired by the translation of a narrow X-ray beam and detector system, or a broader, fan-shaped beam and multiple detector elements across the object (see Fig. 1). The combination of line integrals that fully encompass the object along the same angular position of the measurement system comprises one view, or projection. To allow mathematical reconstruction of the attenuation coefficients in a slice of the object, multiple projections at different angles covering half a rotation of the measurement system around the patient have to be acquired—fan beam geometry needs 180° of data plus the total fan angle of the detector. Today, about 1,000 projections are measured in a full 360° rotation, each comprising about 700–900 individual measurement values (line integrals).

Fig. 1
figure 1

Design and acquisition schemes of four generations of X-ray CT scanners

Different Generations of CT Systems

Since its clinical introduction in the early 1970s, the technology used in CT scanners has matured significantly. Figure 1 illustrates the geometries of the first four generations of X-ray CT scanners.

First-generation CT systems, starting with Sir Godfrey Hounsfield’s original head scanner, the EMI Mark I, were based on the “translate–rotate” principle [1]. These employed a narrow pencil X-ray beam and two detectors that simultaneously acquired two views in the through-plane direction (the through-plane direction or z-direction is along the patient’s longitudinal axis). The measurement values of each projection were acquired sequentially by a linear shift of X-ray tube and detector, followed by a rotation of the measurement system by a small angle (1°) to measure the next projection. Figure 2 shows the Siemens head scanner SIRETOM 1, which was quite similar to the EMI Mark I. The narrow-beam geometry required significantly long scan times (5–7 min for two simultaneously acquired anatomic levels). The image was displayed with an 80 by 80 matrix. The scan field of view (SFOV) had a diameter of 25 cm—the system was therefore limited to head scans. Spatial resolution was 1.3 mm (about four line pairs per cm). Despite that these are “poor” image quality parameters from today’s perspective, the first CT images revealed the patient’s anatomy with a then unknown differentiation of low-contrast details, see Fig. 2.

Fig. 2
figure 2

A first-generation CT head scanner, the Siemens SIRETOM 1, and a head image acquired with this scanner

The next, second-generation CT systems were based on the same “translate–rotate” principle, but the single detector was replaced by an array of about 30 detectors that covered a fan angle of approximately 10°. As a consequence, the acquisition time for a slice could be reduced to 18 s. Second-generation CT scanners were single-slice systems. Multiple detectors in the through-plane direction, as already implemented in the first CT scanner generation, were eliminated from most scanner designs until the recent resurgence of “multi-slice” scanning in the mid 1990s.

The most common CT systems today are third-generation scanners. They employ the so-called “rotate–rotate” geometry, in which both the X-ray tube and a detector array rotate about the patient (see Fig. 3). The fan angle of the detector is wide enough (approximately 45°–55°) to cover a whole-body SFOV with a diameter of 50 cm—in this way, routine body scanning became feasible. A projection was no longer built-up by sequential measurements requiring a translation of detector and X-ray tube, but was instead simultaneously acquired by 700–900 individual detectors in the fan. This way, scan times per slice dropped rapidly. As a next step, slip-ring designs which passed the electrical signals across sliding contacts allowed for continuous rotation of the measurement system and eliminated the need to rewind the gantry after each rotation in order to avoid the tangling of cables. Using this technology, rotation times of less than a second were achieved in the early 1990s. In 2012, the fastest CT scanner needed 0.27 s for a 360° rotation of the tube and detector system. Slip-ring gantries were also a pre-requisite for spiral (helical) CT scanning with continuous data acquisition and continuous table movement, which has revolutionized medical CT since 1990 [2, 3••]. With the use of spiral CT, entire organs, such as lung or liver, could be examined within one breath-hold and without the danger of mis- or double-registration of anatomical details.

Fig. 3
figure 3

Components of a modern third generation CT system. The fan beam covering a whole-body scan field of view with a diameter of 50 cm is indicated

A fourth-generation of CT systems utilized a wide fan beam and a rotating X-ray tube, but a stationary array of detectors that encircled the patient. These systems had fewer moving parts and needed not pass signal data across the moving gantry. They had the potential to enable very fast rotation of the X-ray tube around the patient, and therefore require very short acquisition times per slice. The design principle, however, has currently been abandoned because of high system costs as a consequence of the large number of detector elements and other problems. As an example, focused post-patient collimators (near the detector surface) to suppress scattered radiation could not be used, and the images suffered from severe scatter artifacts.

Multi-Detector Row CT

Since the late 1990s, MDCT systems based on the third generation rotate–rotate geometry have dominated the CT market. The most important advance in these CT scanners is the implementation of X-ray detectors that are physically separated in the z-axis direction and enable simultaneous acquisition of multiple slices of the patient’s anatomy. Key benefits of MDCT systems are faster scan speed, improved through-plane resolution and better utilization of the available X-ray tube power. MDCT also expanded into new clinical areas, such as CT angiography of the coronary arteries with the addition of ECG gating capability [4, 5].

To provide slices at different collimated slice widths, detectors with a larger number of detector rows than finally read-out slices are used. The total beam width in the z-direction is adjusted by pre-patient collimation, and the signals of every two (or more) detectors along the z-axis are electronically combined to thicker slices, see Fig. 4.

Fig. 4
figure 4

Multidetector row CT system utilizing a detector with 24 detector rows to provide 16 slices at different slice widths. If only the central 16 rows are illuminated, they are read-out individually, and the detector provides 16 slices with, e.g., 0.625- or 0.75-mm slice width (4, top). To obtain 16 collimated 1.25- or 1.5-mm slices, the pre-patient collimator is opened. The full z-width of the detector is illuminated, and the signals of every 2 central rows are electronically combined (4, bottom)

The first MDCT systems introduced in 1998 offered simultaneous acquisition of four slices with a thinnest collimated slice width of 1 or 1.25 mm at a rotation time of down to 0.5 s [68]. An 8-slice CT system, introduced in 2000, enabled shorter examination times but no improved spatial resolution (thinnest collimation 8 × 1.25 mm). In 2001, 16-slice CT systems [9] became commercially available, with collimations of 16 × 0.5, 16 × 0.625 or 16 × 0.75 mm and faster gantry rotation (down to 0.42 s and later 0.375 s).

In 2004, all major CT manufacturers introduced MDCT systems with simultaneous acquisition of 64-slices at 0.5-, 0.6- or 0.625-mm collimated slice width, and further reduced rotation times (down to 0.33 s). GE, Philips and Toshiba aimed at increasing volume coverage speed by using detectors with 64 rows, in this way providing 64 collimated 0.5-mm or 0.625-mm slices with a total z-coverage of 32 or 40 mm. Siemens used 32 physical detector rows in combination with a z-flying focal spot to simultaneously acquire 64 overlapping 0.6-mm slices with a total z-coverage of 19.2 mm, with the goal of pitch-independent increase of through-plane resolution and reduction of spiral artifacts [10]. With 64-slice CT systems CT scans with isotropic sub-mm resolution became feasible even for extended anatomical ranges (see Fig. 5).

Fig. 5
figure 5

Whole body CT angiographic scan acquired with a 64-slice CT system at sub-millimeter isotropic spatial resolution. The total scan time with 0.5-s gantry rotation was less than 30 s (courtesy of Eberhard Karls University, Tübingen, Germany)

The improved temporal resolution due to faster gantry rotation increased clinical robustness of ECG-gated scanning, thereby facilitating the successful integration of CT coronary angiography into routine clinical algorithms [11, 12], although higher and irregular heart rates were still problematic.

In 2007, one vendor introduced a MDCT system with 128 simultaneously acquired slices, based on a 64-row detector with 0.6-mm collimated slice width and z-flying focal spot. Recently, simultaneous acquisition of 256 slices became available, with a CT system equipped with a 128-row detector (0.625-mm collimated slice width) and z-flying focal spot.

Clinical experience with 64-, 128- or 256-slice CT indicates that the performance level of MDCT has reached a level of saturation, and mere adding of even more detector rows will not by itself translate into increased clinical benefit. Instead, developments are ongoing to solve remaining limitations of MDCT.

CT Systems with Area Detector

One remaining challenge for MDCT is the visualization of dynamic processes in extended anatomical ranges, e.g., to characterize the inflow and outflow of contrast agent in the arterial and venous system in dynamic CT angiographic studies (CTAs), or to determine the enhancement characteristics of the contrast agent in volume perfusion scans. One way to solve this problem is the introduction of area detectors large enough to cover entire organs, such as the heart, the kidneys or the brain, in one axial scan.

In 2007 a CT-scanner with 320 × 0.5–mm collimation, covering 16 cm at iso-center, and 0.35-s gantry rotation time was commercially introduced by one vendor, after a long evaluation phase using prototype systems (see Fig. 6) with 256 × 0.5–mm collimation and 0.5-s gantry rotation time [1316].

Fig. 6
figure 6

Prototype of a CT area detector providing 256 collimated 0.5-mm slices. From [52], with permission

Computed tomography scanners with area detector have advantages in cardiac scanning and in the acquisition of dynamic CT data. Figure 7 shows a clinical example for a pediatric cardiovascular CT examination with a 320-row CT system.

Fig. 7
figure 7

Congenital anomaly in a 4-day old boy. The pulmonary arteries arise from the aortic arch. The pulmonary veins drain normally into the left atrium. The scan was performed in a single 0.35-s gantry rotation using a CT scanner with 320 × 0.5–mm collimation (courtesy of R. Mather, Toshiba Medical Systems, USA)

With the typical MDCT detector z-coverages of 40 mm (and recently up to 80 mm), ECG-controlled CT volume imaging of the heart is comprised of several sub-volumes acquired during two to four consecutive heart beats. CT systems with large area detectors can image the entire heart in one axial scan without table movement. Meanwhile, successful use of the commercially available CT system with 320 × 0.5–mm detector collimation for coronary CTA has been demonstrated [1719].

As a second benefit, CT systems with area detectors can acquire dynamic volume data by repeatedly scanning the same anatomical range without table movement, which is useful in dynamic CTAs or in volume perfusion studies, e.g., in the management of stroke patients or in oncology [20], or for the evaluation of myocardial perfusion defects.

A challenge of larger detector z-coverage in particular for perfusion scanning is increased X-ray scatter. Scattered radiation may cause hypodense artifacts, may affect CT-number stability, and the scatter induced noise may reduce the contrast-to-noise-ratio (CNR) in the images [21]. The magnitude of scatter artifacts scales linearly with the illuminated z-width of the detector [22].

An alternative approach to dynamic CT scanning of a larger volume is the use of a spiral shuttle mode. By periodically moving the patient forward and backward during data acquisition, a scan range larger than the detector z-width is covered. The scan range can be flexibly adapted to the organ of interest. As a downside, the maximum temporal sampling rate is lower than with a wide detector, and data sampling is temporally equidistant only in the center of the scanned area. The non-equidistant data sampling at different z-positions within the volume has to be considered in the calculation of the perfusion parameters; their accuracy, however, is not compromised [23].

Dual Source CT

Another remaining challenge for MDCT is insufficient temporal resolution for cardiac CT examinations, which require very short exposure time of the individual axial slices to freeze the motion of the heart. To improve temporal resolution in a clinically reliable way, the gantry has to rotate faster. Despite gantry rotation times of 0.3 s and less, motion artifacts at higher and irregular heart rates remain a challenge for coronary CTA. An alternative scanner concept that provides significantly improved temporal resolution but does not require faster gantry rotation is a CT with multiple tubes and corresponding detectors [24, 25]. In 2006, a dual source CT (DSCT), i.e., a CT scanner with two X-ray tubes and two corresponding detectors offset by 90° [26] was commercially introduced by one vendor (see Fig. 8).

Fig. 8
figure 8

Dual source CT-scanner with two independent measurement systems. First generation (center). The measurement systems are at an angle of 90°. Second generation (right). To enlarge the SFOV of detector B, the system angle was increased to 95°

In 2009, the second generation of DSCT-systems was introduced, offering a larger SFOV of the B-detector (33 cm instead of 26 cm), wider z-coverage and faster gantry rotation.

The key benefit of DSCT for cardio-thoracic scanning is improved temporal resolution. In parallel geometry, 180° of scan data (a half-scan sinogram) are necessary for image reconstruction. Due to the 90° angle between both X-ray tubes, the half-scan sinogram can be split up into two 90° data segments which are simultaneously acquired by the two acquisition systems in the same phase of the patient’s cardiac cycle and at the same anatomical level. Therefore, the total data acquisition time per image is a quarter of the gantry rotation time t rot/4 in a sufficiently centered region of the SFOV [26]. For the first-generation DSCT with t rot = 0.33 s, the temporal resolution is t rot/4 = 83 ms. For the second-generation DSCT with t rot = 0.28 s, it is 75 ms as a consequence of the increased system angle of 95°. Meanwhile, several clinical studies have demonstrated the potential of DSCT to provide diagnostic image quality in cardiac CT with little or no dependence on the patient’s heart rate [2731].

Dual source CT systems offer an alternative way to scan the heart within one heartbeat. With a single-source CT, the spiral pitch p is limited to p ≤ 1.5 to ensure gapless volume coverage along the z-axis; p is defined as the table feed per rotation divided by the total detector z-coverage. If the pitch is increased beyond p = 1.5, sampling gaps along the z-axis will lead to severe image artifacts. With DSCT systems, data acquired with the second measurement system a quarter rotation later can be used to fill these gaps. In this way, the pitch can be increased up to p = 3.4 in a limited SFOV that is covered by both detectors [32••, 33]. Using this technique, the second generation DSCT can be operated at a speed of 458 mm/s, which is sufficient to cover the heart (12 cm) in about 0.27 s. Each of the individual axial images has a temporal resolution of 75 ms. The patient’s ECG is used to trigger both table motion and data acquisition. Meanwhile, several clinical studies have demonstrated the successful use of the high-pitch scan technique for coronary CT angiography in patients with sufficiently low and stable heart rates at very low radiation doses [34, 35]. Other potential applications of the high-pitch spiral are ECG-triggered scanning of the entire thorax at low radiation dose, see Fig. 9, or fast scanning of un-cooperative patients, such as in pediatric CT.

Fig. 9
figure 9

ECG-triggered scan of the entire thorax using the dual source high-pitch spiral scan technique. Patient with bilateral pulmonary embolism. The total scan time to cover 269 mm was 0.6 s, the radiation dose at 100 kV and 370 mAs/rot was 2.3 mSv (courtesy of M. Lell, University of Erlangen, Germany)

The main challenge for dual source CT is cross-scattered radiation, i.e., scattered radiation from X-ray tube B detected by detector A, and vice versa. Cross-scattered radiation can produce artefacts and degrade the CNR of the images, if not adequately corrected for.

Dual source CT systems show interesting properties for general radiology applications, too. Both X-ray tubes can be operated at different kV- and mA-settings, allowing for the acquisition of dual energy data. The use of dual energy CT can in principle add functional information to the morphological information based on X-ray attenuation that is usually obtained in a CT examination.

Dual Energy CT

In CT, the two relevant interaction mechanisms of X-ray photons with matter are the photo effect and the Compton effect. Both show different dependence on the photon energy E and on the atomic number Z of the investigated materials. A material can therefore be identified by its characteristic change of attenuation when scanned with two different mean X-ray energies.

One option to provide CT scan data at different mean energies is to run one X-ray tube in a dual source CT at, e.g., 80 kV and the other at 140 kV (see Fig. 8). The scan parameters can be individually adjusted for both measurement systems, resulting in a flexible choice of scan modes. Optimization of the spectral separation is possible by introducing additional pre-filtration into the 140 kV beam, e.g., by means of a filter that can be moved into the beam when needed and moved out for standard applications. An additional tin filter with a thickness of 0.4 mm shifts the mean energy of the 140-kV spectrum behind the bow-tie filter from 69 to 89 keV (see Fig. 10). The mean energy of the 80-kV spectrum is 52 keV.

Fig. 10
figure 10

Standard 80- and 140-kV spectra (left). 80-kV spectrum and 140-kV spectrum with additional 0.4 mm tin pre-filtration (right). Note the shift of the 140-kV spectrum to higher mean energy which improves spectral separation (courtesy S. Kappler, Siemens Healthcare, Forchheim, Germany)

As a downside, dual energy evaluation with DSCT is restricted to the smaller central SFOV of detector B (26 resp. 33 cm diameter). Raw data based dual energy algorithms are difficult to realize, because high-energy and low-energy projections are not simultaneously acquired at the same z-position. Dual energy algorithms are therefore image-based. Furthermore, images of moving objects show slightly different motion artifacts due to the 90° offset between both measurement systems.

Another option to acquire dual energy data is fast switching of the X-ray tube voltage from one projection to the next in a single source CT (“fast kV-switching”), see Fig. 11.

Fig. 11
figure 11

Principle of fast kV-switching to acquire dual energy CT data

Dual energy scans can be performed in the full SFOV of the respective CT scanner, and raw-data based evaluation approaches are possible. The nearly simultaneous acquisition of low-energy and high-energy projections prevents registration problems due to organ motion or contrast agent dynamics. However, as a down-side, spectral optimization by inserting additional pre-filters is not possible. Using current tube technology, only the tube voltage can be switched between consecutive readings, but not the tube current. Equal dose at 80 and 140 kV can only be achieved if each 80 kV reading is a factor 2–4 longer than the corresponding 140 kV reading. This reduces the number of available projections per rotation. Another drawback is the fixed and relatively high tube current that is required to obtain stable switching between both tube potentials (in terms of undershoots and overshoots of both voltage and current, stable shape and size of the focal spot). In one realization, a fixed tube current of 600 mA has to be applied [36, 37]. Radiation dose can only be adapted to the clinical application by variation of the spiral pitch and the rotation time.

Yet another option is the use of energy resolving detectors that allow the acquisition of spectral CT data with a single polychromatic X-ray spectrum. Pertinent examples are dual layer detectors consisting of two conventional scintillation detectors on top of each other, and direct converting photon counting detectors. So far, only prototype CT systems relying on both detector technologies have been realized. In particular photon-counting detectors, however, are a promising technology for future CT-systems.

Clinical applications of dual energy CT include tissue characterization, calcium quantification, calculation of pseudo-monochromatic images and quantification of the local blood volume in contrast-enhanced scans [3841]. Figure 12 shows examples of dual energy applications.

Fig. 12
figure 12

Dual source, dual energy scan of a patient with metallic implants (top). The standard image shows severe metal artifacts (top left). Metal artifacts are significantly reduced by dual energy–based computation of pseudo-mono-energetic images at 120 keV (top right). Dual source, dual energy scan of a patient with gout (bottom). Uric acid crystals, which are a characteristic sign of gout, are identified and highlighted in green (courtesy of Sir Run Run Shaw Hospital, Hangzhou, China, and Clinical Innovation Center, Mayo Clinic Rochester, MN, USA)

Alternative CT Designs

A fifth-generation of CT scanners, the EBCT, uses a stationary X-ray tube and a stationary detector ring, thereby avoiding any mechanically moving parts. In these systems, an electron beam is emitted from a powerful electron gun and magnetically deflected to hit a semi-circular anode surrounding the patient. The magnetic deflection sweeps the electron beam over the target, thus generating an X-ray source that virtually rotates around the patient. Given the absence of mechanically moving parts, a sweep can be accomplished in as little as 50 ms (see Fig. 13).

Fig. 13
figure 13

Schematic illustration of an electron-beam CT (EBCT), a CT concept without mechanically moving parts. An electron beam is magnetically deflected to hit a semi-circular anode surrounding the patient (angular coverage 210°). The electron beam is swept over the target in 50–100 ms

Electron-beam CT systems were already introduced in 1984 as a non-invasive imaging modality for the diagnosis of coronary artery disease [42, 43]. The technical principles were discussed in [44, 45]. Despite its excellent temporal resolution, EBCT suffers from inherent disadvantages of the measurement principle, which have prevented a more wide-spread use in cardiology or general radiology. Due to the fourth-generation system geometry, the use of anti-scatter collimator blades on the detector is not possible. As a consequence, image quality is degraded by scattered radiation. The housing of the electron gun limits the movement of the patient table and thus the scan range. Because of the fixed beam current of 640 mA in the commercially released EBCT-product, the exposure of only 32–64 mAs per slice was insufficient for larger patients.

In summary, the EBCT principle is not considered adequate either for state-of-the-art cardiac imaging or for general radiology applications, and has currently been abandoned.

Currently, newer experimental CT designs are being evaluated, such as inverse geometry CT systems [46, 47]. They use an X-ray tube with multiple distributed focal spots which sequentially illuminate a small detector to cover the examination object with several “sub-projections”, see Fig. 14. These systems provide interesting advantages, such as the potential to shape the beam profile within a projection, thereby flexibly adapting the radiation dose to the size and shape of the patient. As a downside, they require very short acquisition times of the sub-projections to keep the total scan time within acceptable limits, and as a consequence short X-ray pulses with very high tube output. Currently, tube technologies that could potentially be used in inverse geometry CT systems are not mature enough to enable routine clinical use.

Fig. 14
figure 14

Schematic illustration of a third-generation fan beam CT geometry (left) and an inverse CT geometry (right) (Courtesy of Dr. N. Pelc, Stanford University, USA)

Other, even more experimental CT designs, such as interior tomography [48, 53••] have not left the concept stage so far. Multisource interior tomography is intended to enable ultrafast imaging by irradiating a region of interest, such as the heart, in parallel with narrow X-ray beams defined by many source—detector pairs for data acquisition. This region of interest can be then reconstructed using the interior tomography approach.

All CT concepts discussed so far rely on the attenuation of X-rays in the examined object and on measuring the intensity of the attenuated X-ray beam. Phase-contrast CT, on the other hand, is based on measuring the phase shift of the incoming X-rays, which are an electromagnetic wave, by the measurement object. Phase-contrast imaging has recently gained considerable attention in the scientific literature as a new method with potential applications in medical imaging, in particular after its successful implementation using low-brilliance sources [49]. For a typical photon energy of, e.g., 60 keV, the attenuation coefficient of relevant materials, such as water, is three orders of magnitude smaller than the phase shift. This suggests that phase contrast CT should be much more sensitive than absorption CT and could allow for a significantly reduced radiation dose to the patient. Unfortunately, this potential advantage of phase contrast CT can only be utilized at very high spatial resolution [50]. At the resolution level of today’s medical CT, the relative performance of phase contrast CT is inferior to absorption CT. Imaging techniques already operating at higher spatial resolution, such as mammography, may benefit from phase contrast for some applications in terms of reduced patient dose at equal image quality [51].

Conclusion

We have reviewed the different system concepts of CT scanners from the introduction of CT in the early 1970s to modern MDCT, CT with area detectors, and dual source CT. Interestingly, despite all technological advances and improvements, all modern CT scanners are still based on the third-generation rotate–rotate geometry. Other system concepts, such as the fourth-generation CT scanners with rotating X-ray tube and stationary detector, or the fifth-generation EBCT systems, have been abandoned because of severe practical drawbacks and limitations. Although several novel concepts for CT systems are currently being investigated, convincing approaches ready for routine clinical use and not relying on the classic “third-generation design” have not yet been presented—it will be interesting to witness the first real breakthrough in this aspect.