Abstract
The facilities for instrumental neutron activation analysis in Delft are described. Technical details of the fast rabbit systems, the normal pneumatic rabbit system, the large sample facility and the various coaxial and well-type Ge-detector spectrometers and sample changers are given. The capacity of the facilities is in the order of 15,000 samples per annum for full multi-element analysis. The capacity of the fast rabbit systems for INAA for use with short half-life radionuclides is even larger. The facilities are accessible for use by scientists from other establishments and countries as well.
Similar content being viewed by others
References
P. J. M. Korthoven, M. De Bruin, J. Radioanal. Chem., 35 (1977) 127.
P. J. M. Korthoven, M. De Bruin, Computer aspects of large-scale routine instrumental activation analysis, Proc. Conf. on Computers in Activation Analysis and Gamma-Ray Spectrometry, Mayaguez, Puerto Rico, 1978, p. 639.
M. De Bruin, P. Bode, P. J. M. Korthoven, Kerntechnik, suppl. to vol. 44 (1984) 683.
P. Bode, M. De Bruin, P. J. M. Korthoven, Research Reactor Activities in Support of National Nuclear Programs, IAEA-TECDOC 409, IAEA, Vienna, 1987, p. 369.
M. De Bruin, P. J. M. Korthoven, P. Bode, J. Radioanal. Chem., 70 (1982) 497.
P. Bode, J. Trace Microprobe Techn., 8 (1990) 139.
P. Bode, Trans. Am. Nucl. Soc., 68 (1993) 14.
P. Bode, J. P. J. Van Dalen, J. Radioanal. Nucl. Chem., 179 (1994) 141.
P. Bode, Analyst, 120 (1995) 1527.
P. Bode, J. Radioanal. Nucl. Chem., 215 (1997) 51.
P. Bode, Proc. IAEA Intern. Symp. on Harmonization of Health-related Environmental Measurements using Nuclear and Isotopic Techniques, IAEA-SM-344/13, IAEA, Vienna, 1997, p. 49.
P. Bode, R. M. W. Overwater, J. Radioanal. Nucl. Chem., 167 (1993) 169.
P. Bode, M. De Bruin, P. J. M. Korthoven, J. Radioanal. Chem., 26 (1975) 209.
M. De Bruin, S. S. Then, P. Bode, P. J. M. Korthoven, Nucl. Instr. Meth., 121 (1974) 611.
P. Bode, M. De Bruin, J. Radioanal. Chem., 112 (1987) 375.
P. Bode, P. J. M. Korthoven, M. De Bruin, J. Radioanal. Chem., 113 (1987) 371.
P. Bode, M. De Bruin, J. Radioanal. Chem., 132 (1987) 365.
P. Bode, M. Blaauw, I. ObrusnÍk, J. Radioanal. Nucl. Chem., 157 (1992) 301.
P. Bode, R. M. W. Overwater, J. J. M. De Goeij, J. Radioanal. Nucl. Chem., 216 (1997) 5.
S. S. Then, F. D. P. Geurink, P. Bode, R. M. Lindstrom, J. Radioanal. Nucl. Chem., 215 (1997) 249.
M. Blaauw, Nucl. Instr. Meth., A353 (1994) 269.
M. Blaauw, The κ 0-Consistent IRI Gamma-Ray Catalogue for INAA, IRI, TU Delft, Delft, 1996, ISNB 90–73861–32–2, 141 p.
P. Bode, C. P. Van Dijk, J. Radioanal. Nucl. Chem., 215 (1997) 87.
M. J. J. Ammerlaan, P. Bode, J. Radioanal. Nucl. Chem., 215 (1997) 253.
P. Bode, M. Blaauw, C. P. Van Dijk, Transact. Am. Nucl. Soc., 77 (1997) 33.
M. Blaauw, Nucl.Instr. Meth., A332 (1993) 493.
R. M. W. Overwater, P. Bode, J. J. M. De Goeij, Anal. Chem., 68 (1996) 341.
Author information
Authors and Affiliations
Rights and permissions
About this article
Cite this article
Bode, P. Automation and Quality Assurance in the NAA Facilities in Delft. Journal of Radioanalytical and Nuclear Chemistry 245, 127–132 (2000). https://doi.org/10.1023/A:1006777230207
Issue Date:
DOI: https://doi.org/10.1023/A:1006777230207