Abstract
Mechanical ventilation (MV) and supplementation of oxygen-enriched gas, often needed in postnatal resuscitation procedures, are known to be main risk factors for impaired pulmonary development in the preterm and term neonates. Unfortunately, current imaging modalities lack in sensitivity for the detection of early stage lung injury. The present study reports a new imaging approach for diagnosis and staging of early lung injury induced by MV and hyperoxia in neonatal mice. The imaging method is based on the Talbot-Lau x-ray grating interferometry that makes it possible to quantify the x-ray small-angle scattering on the air-tissue interfaces. This so-called dark-field signal revealed increasing loss of x-ray small-angle scattering when comparing images of neonatal mice undergoing hyperoxia and MV-O2 with animals kept at room air. The changes in the dark field correlated well with histologic findings and provided superior differentiation than conventional x-ray imaging and lung function testing. The results suggest that x-ray dark-field radiography is a sensitive tool for assessing structural changes in the developing lung. In the future, with further technical developments x-ray dark-field imaging could be an important tool for earlier diagnosis and sensitive monitoring of lung injury in neonates requiring postnatal oxygen or ventilator therapy.
Similar content being viewed by others
Introduction
Early lung injury in the neonatal lung is often provoked by oxygen supplementation or mechanical ventilation (MV) or both (MV-O2), established as critical life-saving treatment strategies in postnatal care. Due to the immature morphology of the lung, these treatment options are also known to induce defective alveolar septation, impaired angiogenesis and pathologic extracellular matrix remodeling resulting in lung growth impairment1,2,3. In the long term, these changes result in neonatal chronic lung disease (nCLD), also known as Bronchopulmonary Dysplasia (BPD)4, frequently complicating the course of preterm or risk term birth. Along with asthma and cystic fibrosis, nCLD is one of the most common chronic lung diseases in children, whose incidence is reported to be as high as 77% in neonates born at less than 32 weeks of gestation5,6. Although outgrowing oxygen dependency by the age of two years, many infants with BPD have episodes of wheezing, require inhalation therapies7 or show indications of poor pulmonary gas transfer and significantly lower peak workload at school age8. In consequence, adolescent nCLD patients show impaired pulmonary function including a reduction in FEV1 that may be regarded as a precursor of COPD at an older age9.
Due to the severity of the disorder and the associated treatment costs, there is an urgent need for a diagnostic tool to early and reliably detect, stage and monitor morphological changes associated with lung injury caused by mechanical ventilation and oxygen toxicity. Current clinical routine is based on the use of conventional x-ray chest radiography10,11,12, which is limited by low sensitivity for the detection of pulmonary morphological changes11. CT has been demonstrated to provide much more meaningful results13,14, yet its use is severely limited due to the associated high radiation exposure of the infants. Access to lung function information is limited in the clinical setting especially after cessation of invasive ventilation. Spirometric tests that largely depend on patients’ compliance have been shown to suffer from a high variability15,16 and do not yield spatial information.
Developing x-ray imaging further, a method has been recently reported that makes it possible to acquire additionally to conventional x-ray absorption information, x-ray phase-contrast and dark-field signals17,18,19. Thereby, information about the small-angle x-ray scattering registered in the dark field19,20 has been shown to significantly increase lung tissue visibility on radiographic images in mice21 and to improve the detection of calcifications in mammographic scans22. The acquisition of this imaging modality is based on the introduction of a three-grating Talbot-Lau interferometer into the x-ray beam. The change in refractive index between tissue and air causes x-rays to be refracted on each air-tissue interface in the lung, resulting in small-angle scattering. Thus, a strong dark-field signal has been observed for healthy lungs in mice21. Moreover, it could be demonstrated that detection of changes to the lung structure can be significantly improved based on dark field compared to absorption x-ray imaging as shown by the analysis of pulmonary emphysema23,24,25,26 and fibrosis27 in mice.
Bringing together the need for a sensitive diagnostic tool in neonatal chronic lung disease and the novel technique of x-ray dark-field imaging in this study, we evaluated for the first time the capability of grating-based x-ray dark-field radiography to detect very early changes in morphology of the neonatal developing lung undergoing injury induced by hyperoxia and MV using a unique, pre-clinical mouse model3,28,29,30,31.
Results
Detection of morphological changes in the neonatal lung undergoing MV-O2 using x-ray dark-field radiography
As shown in Fig. 1, x-ray dark-field radiograms, in contrast to conventional x-ray transmission projections, were able to visualize changes in the neonatal mouse lung induced by MV-O2. Unventilated lungs from animals breathing room air appear bright, whereas the diseased lungs display a reduced dark-field signal intensity, due to reduction in air-tissue interfaces. Histology supported the dark-field imaging results with larger and fewer alveoli in the lungs of neonatal mice undergoing MV-O2 (Fig. 1(C)). Dark-field radiograms furthermore showed an increase in signal heterogeneity in neonatal mice undergoing 8 hours of MV-O2 (Fig. 2), which, apart from the overall signal decay, reflects inhomogeneity in lung injury and consecutive remodeling.
Sensitivity of imaging to different degrees of neonatal lung damage
To assess the sensitivity of the imaging method with respect to early changes, a series of transmission and dark-field radiographies were performed in mice that were exposed to moderate hyperoxia for 2 or 8 hours or to short (2 h) or medium length (8 h) MV-O2, respectively. Results were compared to images of mice breathing room air. As presented in Fig. 3, the dark-field signal intensity showed a stepwise decrease when comparing the lungs from neonatal mice exposed to hyperoxia for 8 h and to MV-O2 for 2 and 8 h with respect to the animals from the room air group. In contrast, the transmission radiograms did not allow a reliable discrimination of the groups.
Quantification of the imaging signal intensities (Fig. 4) confirmed these observations. Whereas analysis of the transmission signal revealed only modest differences when comparing the different experimental groups, x-ray dark-field radiography yielded a significantly improved discrimination between the animal groups. Quantification of transmission images showed a maximum change of 14% in logarithmic transmission values, when comparing images obtained from animals after 8 h of MV-O2 with results from mice breathing room air. In contrast, analysis of the dark-field radiograms revealed a stepwise scattering decrease, reaching significance as early as in the group of animals undergoing 2 h of MV-O2. The maximum logarithmic signal difference of 32% was observed between the groups of 8 h MV-O2 and room air controls. A similar clear discrimination result was obtained also for the normalized scatter, with a maximum difference of 40% between the room air control animals and 8 h MV-O2 mice.
Histologic assessment of structural changes to the neonatal lung and pulmonary tests
Histologic analysis showed a step-wise increase in distal airspace size and a decrease in radial alveolar counts (RAC) reaching significance when comparing lungs from mice breathing room air with animals undergoing 8 h of MV-O2 (Fig. 5). Quantification of RACs reached a significant difference when comparing lungs from animals after moderate hyperoxia for 8 h and pups breathing room air. Lung volumes measured by fluid displacement before histology were found to not differ significantly between the groups, e.g. RA: 53.9 ± 3.4 μl/g; 8 h MV-O2: 56.2 ± 3.1 μl/g. Percentage of atelectasis was not significantly different between the groups and did not exceed 30% in relation to total lung tissue.
Lung function analysis, comprising assessment of tidal volume and airway pressure by whole body plethysmography only revealed differences when comparing animals undergoing 8 h of MV-O2 with their respective hyperoxia controls (Fig. 5).
Correlation of imaging with histological parameters
Distal airspace size and alveolar number, estimated by radial alveolar counts show a linear correlation with the normalized scatter (Fig. 6) with a Pearson’s correlation coefficient of −0.75 (distal airspaces size) and 0.82 (RAC), respectively. In contrast, the correlation coefficient with the transmission signal was only −0.67 (distal airspace size) and 0.76 (RAC), respectively.
Discussion
In the present study we have demonstrated that grating-based x-ray dark-field radiography can be used to significantly improve the visualization of early changes in the neonatal developing mouse lung undergoing MV-O2. The transmission signal is not sensitive enough to visualize the changes in airspace size and quantity. In contrast, x-ray dark-field radiography is picking up the architectural changes in the lung, thereby providing valuable visualization of early pulmonary changes and allowing for the assessment of different degrees of lung injury. As diffracted x-ray photons are detected in the same detector pixel without influencing the transmission signal in the absence of the grating interferometer, the result highlights the complementarity of the x-ray transmission and the dark-field imaging technique. Only the x-ray dark-field signal is able to pick up the small diffraction angles, thereby generating sensitive information about the pulmonary structure, i.e. airspace enlargement. In order to account for the inter-animal variability, a combination of x-ray dark-field and conventional transmission imaging, referred to as normalized scatter, was calculated. The normalized scatter reflects the pulmonary architectural changes induced by increase in airspace size and loss of (septal) tissue and reveals the best differentiation between the different experimental groups. Due to the developmental stage of the murine lung, the airspace can refer to alveoli themselves or the alveolar duct as secondary septation is still ongoing.
Imaging results obtained by dark-field radiography were related to quantitative histological analysis. With respect to the clinical relevance of the model, histological analysis displays changes to the lung structure that mimics the characteristics of pulmonary injury observed in infants undergoing MV-O2. This has been confirmed in previous studies32.
The early pulmonary changes in the lung undergoing postnatal injury can be attributed to an increase in i) apoptosis or ii) extracellular matrix remodeling with the resulting airspace enlargement and interstitial remodeling reflected by the signal alteration in dark-field imaging. The detection of these early pulmonary changes, long-term potentially leading to BPD development, is of uttermost importance to allow for adequate and individualized treatment strategies in the mechanically ventilated neonate. In contrast, pulmonary function tests show a lower sensitivity for the detection of early morphological changes and only differentiate for the changes observed in neonatal mice that underwent 8 h of MV-O2. This result furthermore emphasizes the necessity for a more sensitive, advanced diagnostic approach in the clinical setting, where conventional x-ray techniques and lung function analysis serve as the main diagnostic tools.
The results obtained by the current study will open a new perspective for future management of neonates that require mechanical ventilation with oxygen enriched gas and are at a high risk to develop long-term changes to the pulmonary architecture. X-ray dark-field imaging allows for the detection of early detrimental effects by MV-O2 in neonates and can thereby help to monitor treatment approaches including the development of differentiated ventilation protocols protecting the developing lung from injury. To overcome the limitation of 2D imaging of superimposition of ventral and dorsal lung regions, the additional acquisition of lateral radiograms have to be considered.
Although X-ray dark-field radiography is still a preclinical technique, a recent study reported the development of a system that has a field of view that is suitable for imaging the thorax of an infant33 and showed the feasibility of acquiring x-ray dark field radiograms at a clinically compatible dose33. In mice, the dose used to acquire the small-animal radiographies is estimated with 1.4 mGy, so that longitudinal in vivo small-animal measurements can be performed without restrictions24,25. In order to obtain x-ray transmission and dark-field images, a precise movement of one of the gratings by a fraction of its period is required between the acquisitions of a minimum of two images. As the grating periods are typically in the range of a few μm, a high precision for the movement of the grating is mandatory. As this prerequisite can be challenging for clinical applications, several algorithms have been recently developed to adjust for the impact of imprecise grating movement34. Nonetheless, the technical implementation of x-ray grating interferometer into clinical imaging systems will require additional optimization.
The reasoning for an ex vivo imaging approach was based on two important technical challenges that needed to be addressed. On the one hand, the differing degree of pulmonary inflation induced by the process of MV led to the implementation of standardized inflation volumes immediately after the disconnection from the ventilator in order to allow for the assessment of structural changes by avoiding the artificial impact of differences in inflation levels. On the other hand, the optimization of the experimental conditions in neonatal mice ventilation required the reduction of dead space and optimization of temperature conditions, which could not be achieved inside the scanner.
Nonetheless, future research should focus on applying x-ray dark-field imaging in parallel to the ventilation procedure in order to investigate changes in the lung while undergoing MV-O2. Furthermore, these studies should take the spatial information for the signal analysis into account and seek to identify the most affected areas of the lung.
Materials and Methods
Neonatal mouse ventilation
Animal experiments were performed with permission of the Institutional Animal Care and Use Committee of the Helmholtz Zentrum Munich and carried out in accordance with national (Gesellschaft für Versuchstierkunde - Society for Laboratory Animal Science) and international (Federation for Laboratory Animal Science Associations) animal welfare guidelines. All experimental protocols were approved by the Institutional Animal Care and Use Committee of the Helmholtz Zentrum Muenchen, Munich, Germany. The study was carried out in 5–7-day-old C57B6 wild type mice born at term gestation (3.2 ± 0.4 g). Neonatal mice were assigned to the following groups: unventilated animals spontaneously breathing 40% O2 for 2 or 8 hours or room air (21% O2) (n = 3–6). Randomly chosen littermates received mechanical ventilation with oxygen-enriched gas (MV-O2; 40% O2) for 2 or 8 hours (n = 4–6). For MV-O2, pups received a tracheotomy under sedation with subcutaneous ketamine (~60 μg/g body weight, bw) and xylazine (~12 μg/g bw), before transfer to MV at 180 breaths/min with a customized, small animal respirator (MicroVent 848; Harvard Apparatus, Holliston, MA). The ventilation settings mimic clinical protocols with a volume-guaranteed strategy applying mean tidal volumes of 8.68 μl/g bw, stepwise decreased after the onset of MV, and airway pressures of 12–13 cmH2O (peak) and 11–12 cmH2O (mean)35. All unventilated animals received sham surgery under mild sedation with ketamine and xylazine.
Pulmonary Function Tests
In a subset of animals, tidal volume and airway pressure (maximum tracheal pressure, Ptramax) measurements were performed using whole body plethysmography (Pulmodyn, Hugo Sachs Elektronik, Harvard Apparatus GmbH, Hugstetten, Germany). As the impact of the chest wall can be neglected in the newborn, tidal volume of the ventilator was increased in a stepwise manner from 10 to 25 μl/g bw to assess the quasi-static compliance of the respiratory system. The animals were allowed to relax for 60 seconds between different tidal volumes which were applied for 30 seconds.
Imaging Protocol
In order to obtain x-ray transmission and dark-field images it is necessary to acquire a minimum of two images with a different relative position of the gratings in the grating interferometer. Therefore, one of the gratings is typically moved by a fraction of its period between the images, thereby achieving the possibility to discriminate between x-ray absorption and the scattering processes. From the acquired original images the intrinsically perfectly registered transmission and dark-field images were calculated using the Fourier decomposition18. Imaging was performed with a prototype small-animal x-ray phase-contrast and dark-field scanner36,37. On the ventilator, pups were euthanized with sodium pentobarbital, lungs were filled with a standardized volume of 150 μl room air via the intra-tracheal tubing and animals were transferred immediately to the dark-field scanner. Ex vivo imaging was chosen in order to i) standardize inflation volume to optimize quantification of structural changes of the lung and to ii) meet critical requirements for MV in neonatal mice including the reduction of dead space and temperature optimization. The scanner was operated at 35 kVp (source 17 W), acquiring images for 5 different source grating positions, with an exposure time of 3.3 seconds per image. The sample resolution was approximately 60 μm (10% MTF). The grating interferometer consisted of a gold source grating (period p = 10 μm, height h = 35 μm), a nickel phase grating (p = 3.24 μm, h = 4 μm) and a gold analyzer grating (p = 4.8, h = 45 μm). The distances between the source to the phase and the analyzer grating are 30 and 45 cm, respectively.
Quantitative Morphometry
After imaging, all lungs were harvested for histological analysis through intra-tracheal fixation with buffered 4% paraformaldehyde at a standard pressure of 20 cm H2O as previously reported32. After fixation, lung volume was measured using fluid displacement38. Lungs were subsequently embedded in paraffin. Histologic analysis was performed blinded to the animal’s group assignment and included airspace size, radial alveolar counts (RAC) and septal counts. As published previously, these measures allow estimating airspace enlargement together with a reduction in alveolar/airspace number in the case of unchanged lung volumes as measured by fluid displacement. The percentage of atelectasis in relation to total lung tissue was quantified using the ImageJ software. Measurements of alveolar area, i.e. airspace size was performed in 4 high power fields per tissue section in 2–3 independent random tissue sections (4 μm, H&E) per animal (CAST-Grid 2.1.5; Olympus, Ballerup, Denmark). A median of ≥30 fields of view in 2–3 random tissue sections per animal was used to estimate alveolar number by the use of radial alveolar counts (RAC) as described previously39.
Image Analysis
The imaging signals were quantified blinded to the animal’s group assignment using a manually created mask identical for x-ray transmission and dark-field radiograms, as both imaging modalities are intrinsically perfectly registered. The mask was chosen to fit as much lung tissue as possible, omitting the heart shadow. As the ribs are not fully developed in the pups and are barely visible on the transmission images, their influence was neglected. The signal intensities were subsequently analyzed using Matlab R2013b (MathWorks, Natick, MA, USA). Since the dark field scales exponentially with the sample thickness20,40 similar to the Beer-Lambert law for transmission imaging, the natural logarithm of both transmission T and dark field V were quantified.
In order to get rid of the sample thickness dependence, associated with projection-based imaging, a ratio of the transmission and dark field was calculated, referred to as normalized scatter N41. The normalized scatter is obtained as:
where V0 and T0 are the dark field and transmission amplitudes in front of the sample and ε and μ are the diffusion and attenuation coefficient, respectively.
Statistical analysis
Data are given as mean and standard deviation (SD). Statistical analysis was performed using Matlab R2013b (MathWorks, Natick, MA, USA). The datasets were compared using Student’s unpaired t-test or the non-parametric Mann-Whitney test for results with skewed distribution when two groups were compared. For the comparison of three or more groups the N-Way ANOVA was applied with respective post-hoc tests as appropriate. Differences were considered statistically significant when the p value was <0.05.
Additional Information
How to cite this article: Yaroshenko, A. et al. Visualization of neonatal lung injury associated with mechanical ventilation using x-ray dark-field radiography. Sci. Rep. 6, 24269; doi: 10.1038/srep24269 (2016).
References
Jobe, A. H. & Bancalari, E. Bronchopulmonary dysplasia. Am J Respir Crit Care Med 163, 1723–1729 (2001).
Merritt, T. A., Deming, D. D. & Boynton, B. R. The ‘new’ bronchopulmonary dysplasia: challenges and commentary. Semin Fetal Neonatal Med 14, 345–357 (2009).
Hargitai, B. et al. Apoptosis in various organs of preterm infants: histopathologic study of lung, kidney, liver, and brain of ventilated infants. Pediatr Res 50, 110–114 (2001).
Northway, W. H., Jr., Rosan, R. C. & Porter, D. Y. Pulmonary disease following respirator therapy of hyaline-membrane disease. Bronchopulmonary dysplasia. N Engl J Med. 276, 357–368 (1967).
Stoll, B. J. et al. Neonatal outcomes of extremely preterm infants from the NICHD Neonatal Research Network. Pediatrics, 10.1542/peds.2009-2959 (2010).
Johnson, A. H. et al. High-frequency oscillatory ventilation for the prevention of chronic lung disease of prematurity. N Engl J Med 347, 633–642 (2002).
May, C. et al. Lung function abnormalities in infants developing bronchopulmonary dysplasia. Arch Dis Child, 10.1136/adc.2011.212332 (2011).
Welsh, L. et al. The EPICure study: maximal exercise and physical activity in school children born extremely preterm. Thorax, 10.1136/thx.2008.107474 (2009).
Doyle, L. W. et al. Bronchopulmonary dysplasia in very low birth weight subjects and lung function in late adolescence. Pediatrics 118, 108–113 (2006).
May, C., Prendergast, M., Salman, S., Rafferty, G. F. & Greenough, A. Chest radiograph thoracic areas and lung volumes in infants developing bronchopulmonary dysplasia. Pediatr Pulmonol 44, 80–85 (2009).
Oppenheim, C. et al. Bronchopulmonary dysplasia: value of CT in identifying pulmonary sequelae. Am J Roentgenol 163, 169–172 (1994).
Northway Jr, W. H. et al. Late pulmonary sequelae of bronchopulmonary dysplasia. N Engl J Med 323, 1793–1799 (1990).
Aukland, S. M. et al. Neonatal bronchopulmonary dysplasia predicts abnormal pulmonary HRCT scans in long-term survivors of extreme preterm birth. Thorax 64, 405–410 (2009).
Tonson la Tour, A. et al. Chest CT in bronchopulmonary dysplasia: clinical and radiological correlations. Pediatr Pulmonol 48, 693–698 (2013).
Hruby, J. & Butler, J. Variability of routine pulmonary function tests. Thorax 30, 548–553 (1975).
Pellegrino, R. et al. Interpretative strategies for lung function tests. Eur Respir J 26, 948–968, 10.1183/09031936.05.00035205 (2005).
Momose, A. et al. Demonstration of x-ray Talbot interferometry. Jpn J Appl Phys 42, 866–868 (2003).
Pfeiffer, F., Weitkamp, T., Bunk, O. & David, C. Phase retrieval and differential phase-contrast imaging with low brilliance x-ray sources. Nat Phys 2, 258–261 (2006).
Pfeiffer, F. et al. Hard x-ray dark-field imaging using a grating interferometer. Nat Mater 7, 134–137 (2008).
Bech, M. et al. Quantitative x-ray dark-field computed tomography. Phys Med Biol 55, 5529–5539 (2010).
Bech, M. et al. In-vivo dark-field and phase-contrast x-ray imaging. Sci. Rep. 3, 10.1038/srep03209 (2013).
Grandl, S. et al. Improved visualization of breast cancer features in multifocal carcinoma using phase-contrast and dark-field mammography: an ex vivo study. Eur Radiol 25, 3659–3668 (2015).
Yaroshenko, A. et al. Pulmonary Emphysema Diagnosis with a Preclinical Small-Animal X-ray Dark-Field Scatter-Contrast Scanner. Radiology 269, 427–433, 10.1148/radiol.13122413 (2013).
Meinel, F. G. et al. Improved diagnosis of pulmonary emphysema using in vivo dark-field radiography. Invest Radiol 49, 653–658 (2014).
Hellbach, K. et al. in Vivo Dark-Field Radiography for Early Diagnosis and Staging of Pulmonary Emphysema. Invest Radiol 50, 430–435 (2015).
Velroyen, A. et al. Grating-based X-ray Dark-field Computed Tomography of Living Mice. EBioMedicine 2, 1500–1506, 10.1016/j.ebiom.2015.08.014 (2015).
Yaroshenko, A. et al. Improved In vivo Assessment of Pulmonary Fibrosis in Mice using X-Ray Dark-Field Radiography. Sci Rep 5, 17492, 10.1038/srep17492 (2015).
Ekekezie, I. I. et al. Low levels of tissue inhibitors of metalloproteinases with a high matrix metalloproteinase-9/tissue inhibitor of metalloproteinase-1 ratio are present in tracheal aspirate fluids of infants who develop chronic lung disease. Pediatrics. 113, 1709–1714 (2004).
Hislop, A. A., Wigglesworth, J. S., Desai, R. & Aber, V. The effects of preterm delivery and mechanical ventilation on human lung growth. Early Hum Dev. 15, 147–164 (1987).
Margraf, L. R., Tomashefski, J. F., Jr., Bruce, M. C. & Dahms, B. B. Morphometric analysis of the lung in bronchopulmonary dysplasia. Am Rev Respir Dis. 143, 391–400 (1991).
Thibeault, D. W., Mabry, S. M., Ekekezie, I. I. & Truog, W. E. Lung elastic tissue maturation and perturbations during the evolution of chronic lung disease. Pediatrics. 106, 1452–1459 (2000).
Bland, R. D. et al. Mechanical ventilation uncouples synthesis and assembly of elastin and increases apoptosis in lungs of newborn mice. Prelude to defective alveolar septation during lung development? Am J Physiol Lung Cell Mol Physiol. 294, 3–14 (2008).
Koehler, T. et al. Slit-scanning differential x-ray phase-contrast mammography: Proof-of-concept experimental studies. Med Phys 42, 1959–1965, 10.1118/1.4914420 (2015).
Wen, H., Miao, H., Bennett, E. E., Adamo, N. M. & Chen, L. Flexible Retrospective Phase Stepping in X-Ray Scatter Correction and Phase Contrast Imaging Using Structured Illumination. Plos One 8, e78276, 10.1371/journal.pone.0078276 (2013).
Hilgendorff, A. et al. Inhibiting lung elastase activity enables lung growth in mechanically ventilated newborn mice. Am J Respir Crit Care Med 184, 537–546, 10.1164/rccm.201012-2010OC (2011).
Tapfer, A. et al. Development of a prototype gantry system for preclinical x-ray phase-contrast computed tomography. Med Phys 38, 5910–5915, 10.1118/1.3644844 (2011).
Tapfer, A. et al. Experimental results from a preclinical X-ray phase-contrast CT scanner. Proc Natl Acad Sci 109, 15691–15696, 10.1073/pnas.1207503109 (2012).
Scherle, W. A simple method for volumetry of organs in quantitative stereology. Mikroskopie. 26, 57–60 (1970).
Emery, J. L. & Mithal, A. The number of alveoli in the terminal respiratory unit of man during late intrauterine life and childhood. Arch Dis Child. 35, 544–547 (1960).
Yashiro, W., Terui, Y., Kawabata, K. & Momose, A. On the origin of visibility contrast inx-ray Talbot interferometry. Opt Express 18, 16890–16901, 10.1364/OE.18.016890 (2010).
Schleede, S. et al. Emphysema diagnosis using X-ray dark-field imaging at a laser-driven compact synchrotron light source. Proc Natl Acad Sci 109, 17880–17885, 10.1073/pnas.1206684109 (2012).
Acknowledgements
We acknowledge financial support through the DFG Cluster of Excellence Munich-Centre for Advanced Photonics (MAP), the DFG Gottfried Wilhelm Leibniz program, the Helmholtz Association, the German Center of Lung Research (DZL DPLD), the European Research Council (ERC, FP7, StG 240142), and the Langmatz Foundation. This work was carried out with the support of the Karlsruhe Nano Micro Facility (KNMF, www.kit.edu/knmf), a Helmholtz Research Infrastructure at Karlsruhe Institute of Technology (KIT). This work was supported by the German Research Foundation (DFG) and the Technische Universität München within the funding programme Open Access Publishing.
Author information
Authors and Affiliations
Contributions
A.Y., T.P., K.H., S.A., M.R., O.E., F.P. and A.H. conceived the experiments. A.Y., T.P., M.K. and A.H. conducted the experiments. A.Y., T.P., M.K., N.K., K.W., L.G., K.H., F.P. and A.H. analyzed the results. All authors reviewed the manuscript.
Corresponding author
Ethics declarations
Competing interests
The authors declare no competing financial interests.
Rights and permissions
This work is licensed under a Creative Commons Attribution 4.0 International License. The images or other third party material in this article are included in the article’s Creative Commons license, unless indicated otherwise in the credit line; if the material is not included under the Creative Commons license, users will need to obtain permission from the license holder to reproduce the material. To view a copy of this license, visit http://creativecommons.org/licenses/by/4.0/
About this article
Cite this article
Yaroshenko, A., Pritzke, T., Koschlig, M. et al. Visualization of neonatal lung injury associated with mechanical ventilation using x-ray dark-field radiography. Sci Rep 6, 24269 (2016). https://doi.org/10.1038/srep24269
Received:
Accepted:
Published:
DOI: https://doi.org/10.1038/srep24269
This article is cited by
-
X-ray dark-field chest radiography: a reader study to evaluate the diagnostic quality of attenuation chest X-rays from a dual-contrast scanning prototype
European Radiology (2023)
-
Dark-field chest x-ray imaging: first experience in patients with alpha1-antitrypsin deficiency
European Radiology Experimental (2022)
-
Whole-body x-ray dark-field radiography of a human cadaver
European Radiology Experimental (2021)
-
A 3-D Projection Model for X-ray Dark-field Imaging
Scientific Reports (2019)
-
Optimization of tube voltage in X-ray dark-field chest radiography
Scientific Reports (2019)
Comments
By submitting a comment you agree to abide by our Terms and Community Guidelines. If you find something abusive or that does not comply with our terms or guidelines please flag it as inappropriate.