Brought to you by:
Paper

Radiation exposure of patients during endourological procedures: IAEA-SEGUR study

, , , , , , , , , , , , and

Published 20 November 2020 © 2020 Society for Radiological Protection. Published on behalf of SRP by IOP Publishing Limited. All rights reserved
, , Citation Jenia Vassileva et al 2020 J. Radiol. Prot. 40 1390 DOI 10.1088/1361-6498/abc351

0952-4746/40/4/1390

Abstract

Fluoroscopy is increasingly used to guide minimally invasive endourological procedures and optimised protocols are needed to minimise radiation exposure while achieving best treatment results. This multi-center study of radiation exposure of patients was conducted by the South-Eastern European Group for Urolithiasis Research (SEGUR), in cooperation with the International Atomic Energy Agency. Seven clinical centers from the SEGUR group collected data for 325 procedures performed within a three-months period, including standard percutaneous nephrolithotomy (PCNL), mini PCNL, retrograde intrarenal surgery (RIRS), semirigid ureterorenoscopy (URS) and flexible URS. Data included: air kerma area product (PKA), air kerma at the patient entrance reference point (Ka,r), fluoroscopy time (FT), number of radiographic images (N) and fluoroscopy pulse rate, as well as total procedure duration, size and location of stones. Data were centrally analysed and statistically compared. Median PKA values per center varied 2-fold for RIRS (0.80–1.79 Gy cm2), 7.1 fold for mini-PCNL (1.39–9.90 Gy cm2), 7.3 fold for PCNL (2.40–17.50 Gy cm2), 19 fold (0.13–2.51 Gy cm2) for semi-rigid URS and 29-fold for flexible URS (0.10–2.90 Gy cm2). Lower PKA and Ka,r were associated with use of lower FT, N and lower fluoroscopy pulse rate. FT varied from 0.1 to 14 min, a small fraction of the total procedure time, ranging from 10 to 225 min. Higher N was associated with higher PKA and Ka,r. Higher median PKA in PCNL was associated with the use of supine compared to prone position. No correlation was found between the concrement size and procedure duration, FT, PKA or Ka,r. Dose values for RIRS were significantly lower compared to PCNL. The maximum Ka,r value of 377 mGy was under the threshold for radiation induced skin erythema. The study demonstrated a potential for patient dose reduction by lowering FT and N, using pulsed fluoroscopy and beam collimation.

Export citation and abstract BibTeX RIS

1. Introduction

Ionising radiation is an essential part of urological practice, used preoperatively for diagnosis, intraoperatively to guide treatment and postoperatively for follow-up. Endourology is a subspecialty area in urology using minimaly invasive techniques to treat renal stones and other abnormalities in the urinary tract with the help of intracoreal scopes and instrumentation. Percutaneous nephrolithotripsy (PCNL) and ureterorenoscopy (URS) are the two main procedures performed worldwide to treat urolithiasis (stones in the urinary tract). During PCNL, the operating physician gains access to the renal tract through a percutaneous needle puncture, under the fluoroscopic or ultrasound guidance. After successful entrance, the percutaneous tract is dilated and the stone is comminuted with instruments using laser or mechanical energy. During URS, physician enters the urinary tract through the urethra and after identifying ureteral orifice, several instruments and guidewires are used under fluoroscopic guidance to reach the level of stone/pathology and manage it accordingly, usually with laser energy [1]. Fluoroscopy, along with ultrasound, is used to guide safe access to kidneys and ureters, while it accounts for about 18% of all fluoroscopically guided interventional procedures [2]. The advantages of endourology procedures over open surgery include a minimally invasive management option, the elimination of prolonged pain after surgery, and faster recovery [1]. Despite its benefits, the use of ionising radiation poses health risks for patients and operative room staff, thus knowledge should be enriched in order to establish optimised protocols.

During a fluoroscopy guided procedures, both operating team and patients are exposed to radiation that may cause two types of health effects: deterministic (tissue reactions) and stochastic (cancer or hereditary effects) [3, 4]. To limit the radiation exposure to an acceptable level for occupational exposure of staff working with ionising radiation, the International Commission for Radiological Protection (ICRP) and corresponding international safety standards prescribe annual dose limits, and actions to keep the exposure as low as reasanably achievable [3, 5, 6]. At the same time, because radiation is used intentionally for the patient direct benefit, applying dose limits for medical exposure of patients is considered inappropriate. Instead, the efforts are to limit unnecessary exposure by applying radiation protection principles of justification and optimisation [3, 4, 6]. Patient dose assessment and comparison of typical dose values between departments performing the same diagnostic or interventional procedure are considered to be an important benchmarking tool and a basis for optimisation of radiation protection of patients [3, 5 7]. Typical dose value is defined as the median of the distribution of the data for a commonly and easily measured or determined modality-specific radiation dose metric that is obtained from a local survey or a review of local data in a radiological facility [5, 7]. For fluoroscopy guided procedures, such a metric includes the kerma area product and the air kerma at the patient entrance reference point [7].

The amount of radiation reaching patient during a fluoroscopically-guided procedure depends on the total fluoroscopy time (FT), exposure factors (tube voltage, tube current and beam filtration), fluoroscopy mode (combination of settings of the automatic dose rate control (ADRC) system), x-ray beam geometry, including beam colimation, and the number (N) of the acquired radiographic images and their exposure settings [2]. Practicing urologist is called to increase benefit to risk ratio by minimising radiation dose to patient when appropriate, while achieving the best result regarding treatment of underlying health condition. Another important goal is to avoid tissue reactions, namely skin injuries for patients by keeping dose to the patient skin below the dose threshold for such as a tissue reaction [2, 4]. Although observed in other complex fluoroscopy guided interventional procedures such as in cardiology or neuroradiology, no skin injuries have been reported in the scientific literature from endourology [2]. Minimising radiation exposure to patient contributes to reducing the exposure to surgical team members, which can be further lowered by using personal and in-room protective shielding and increasing distance from patient to the extend possible [2, 8].

It is well known that patient doses from fluoroscopy guided procedures vary significantly depending on the type of procedure, its complexity, the fluoroscopy equipment specification and its operation. Data regarding patient radiation exposure in endourology are scant and come from a single center [2, 9 13]. Limited data of patient exposure levels are reported in few publications, which main focus is staff exposure [14 20]. To fill the gap and study the potential for optimisation of patient exposure during endourology procedures, the South-Eastern European Group for Urolithiasis Research (SEGUR), in cooperation with the International Atomic Energy Agency (IAEA), conducted a multi-center study in seven South-Eastern European hospitals. The aim was to investigate radiation exposure of patients during fluoroscopy guided endourological procedures in correlation with affecting factors. In parallel, doses to the endourology staff were measured and analysed, the results of this part of the study were presented in a separate paper [21].

2. Materials and methods

2.1. Study design

Seven clinical centers from the SEGUR group participated in the study: two from Turkey and by one from Bulgaria, Greece, Italy, North Macedonia, and Serbia. Each center appointed a coordinator, while central organisation, data collection and analysis were undertaken by the IAEA experts. The study protocol was approved by the SEGUR Board, and detailed instructions and data collection forms were provided to the local teams. All teams had an access to a local qualified medical physicist or/and an x-ray engineer to advise on data recording and dose display verification.

The study was conducted during 2018–2019 in three phases. In the planning phase, all centers completed a standard questionnaire regarding fluoroscopy system (type and model), type of image receptor, availability of ADRC, pulsed fluoroscopy, digital acquisition (image recording), last image hold (LIH), patient dose display. The total number of endourology procedures performed during 2017 was recorded.

Based on the feedbacks, data collection in the second phase was focused on five most common procedures: Standard PCNL, Mini PCNL, Retrograde Intrarenal Surgery (RIRS), Semirigid URS and Flexible URS. The main difference between standard and mini PCNL is the size of access sheath and instruments used during the procedure. Specifically during standard PCNL, instruments' diameter ranges between 24 and 30 F, while in mini PCNL 14–20 F (F stands for French, used to measure the size of catheters; a cathether of 1 Fr has an outer diameter of 1/3 mm). RIRS is a type of transurethral surgery, during which the endoscope is guided under fluoroscopic guidance to the renal pelvis and calyces and after proper identification of pathology it is managed accordingly. URS can be performed either with semirigid or flexible ureteroscopes, with the latter being increasingly used [1].

Participating centers agreed to record data of all patients who were treated with one of these five procedures during the study period of three months. The standard recording form included four groups of data: de-identified patient data (gender, age and weight), data for patient exposure in terms of available dose metrics; exposure factors FT, N and fluoroscopy pulse rate, and complexity of procedures. No information was requested for tube voltage, tube current, filtration, which are controlled by ADRC system based on the object attenuation.

Patient exposure was recorded in two dose metrics provided by the fluoroscopy systems. The first, kerma area product, PKA, is an indicator of the overall energy imparted to a patient body and thus of risk of stochastic effects, and is defined as the product of air kerma (equal to the absorbed dose in air) at the center of a certain plane of the x-ray beam, multiplied by the area of the beam in that plane, with the unit Gy cm2 [7, 22, 23]. The second metric, air kerma at the patient entrance reference point, Ka,r, expressed in Gy (or mGy), is the air kerma at a point in space located at a fixed distance from the focal spot (for C-arms it is located on the central axis of the x-ray beam, 15 cm from the isocenter toward the x-ray source) and is an indicator of dose to the patient skin [7, 22, 23]. PKA and Ka,r and their rates are measured with devises integrated into the fluoroscopy equipment and the cumulated values are displayed and automatically saved after the procedure [7, 22, 23].

Based on the experience, the SEGUR team agreed to describe the complexity of procedure that might influence patient dose, in terms of total procedure duration, size (length and width) and location of the stone (in ureter or in kidney). Stone surface area (SA), estimated from the length (l) and width (w) of the stone using the formula SA = 0.25lwπ, was calculated to describe the complexity of a procedure [1]. Analyses was performed separately for each type of procedure, and corelation was studied between the dose values and the factors accounting for the complexity.

2.2. Data collection and analyses

Data acquisition was performed during a three months period (May–July 2018), then all forms were collected and centrally analysed in Microsoft® Excel. Distribution of each data set was tested for normality by applying Kolmogorov–Smirnov test, and descriptive statistical analysis was performed by calculating minimum, maximum, median, mean and standard deviation (SD) values. For comparing two or more independent data sets, Student's t-test and one-way ANOVA test (parametric or non-parametric, as appropriate) were used. Associations between dose values and complexity factors within a patient sample was evaluated by using Spearman ρ correlation test. For all statistical tests, significance level was set to α = 0.05.

3. Results

The responses received from the endourology centers are presented with their unique ID number, U1–U7. Centers U1, U2, U4, U6, U7 used mobile C-arm fluoroscopy systems, two of Siemens Healthcare, Germany, one of General Electric, USA, two of General Medical Merate, Italy, one of Gilardoni, Italy. U3 used a fixed C-arm (Lithoskop, Siemens Healthcare), and U5 a fixed fluoroscopy system with over-couch x-ray tube (Uroskop, Siemens Healthcare). All with C-arms declared to place the x-ray tube under the table. System U1 had flat panel detector while others used image intensifier. All systems were equipped with ADRC, digital acquisition mode and LIH function. All except U4 and U6 used pulsed fluoroscopy with either 12.5, 25/30 p s-1. Except U7, all had dose readings in PKA displayed after the procedure, but display of Ka,r was available in only systems U1–U3 and U5. Dose displays were verified by independent measurements in U4, U5 and U6 and the reported results were found to be within the tolerable limits, while in U1, U2 and U3 no verification was performed due to the lack of access to qualified medical physicist and unadequate quality control procedures No dose data were provided from the center U7 that contributed with a limited data for one procedure only.

The annual workload differed significantly between centers, the highest in U6 with 865 procedures in 2017, followed by U1 with 750 and U5 with 707 procedures, U3 with 450 procedures and much lower in U2 and U4, with correspondingly 185 and 173 procedures.

Procedures of 315 patients in total were recorded; their distribution per type and center is presented in table 1 for PCNL (standard and mini) and RIRS and in table 2 for URS (semi-rigid and flexible).

Table 1. Number of patients, patient weight in kg, treated stone burden, expressed in stone surface area (SA) in mm2, stone location, total procedure time (PT) in minutes, fluoroscopy time (FT) in minutes, number of acquired images (N), kerma-area product, PKA, and cumulative air kerma, Ka,r, and fluoroscopy pulse rate (p/s) for the standard PCNL, mini PCNL and RIRS, performed in the participating clinical centers U1–U7. For SA, PT, FT, N, PKA, and Ka,r, data are presented with their mean (min–max) median values. Empty cells correspond to missing data or when statistical analysis could not be performed due to a small sample size. For the fluoroscopy pulse rate, the percentage in brackets indicates the frequency of using the corresponding p/s; CF = continuous fluoroscopy.

 Clinical centerU1U2U3U4U5U6U7
Standard PCNL (111 patients)No. of patients10414636410
 Patient weight73.5 (65–86) 72.577.9 (46–110) 7684.8 (75–92) 8673.7 (40–94) 78.587.6 (60–145) 82.096.0 (79–110) 97.569 (50–92) 66
 SA, mm2 427 (236–589) 412500 (157–1570) 471282 (199–471) 230190 (102–318) 1651014 (120–6005) 641617 (352–942) 589NA
 Stone location100% kidney100% kidney100% kidney100% kidney100% kidney100% kidney100% kidney
 PT, min127 (75–220) 12068 (12–130) 6587.5 (45–110) 97.5103 (53–225) 8456 (30–120) 5091 (65–105) 9630 (24–36) 30
 FT, min1.5 (0.6–3.1) 1.34.0 (1.0–14.0) 3.21.7 (0.9–3.7) 1.14.4 (2.5–8.5) 3.32.7 (0.3–10.3) 2.25.2 (3–7.3) 5.43.0 (2.0–4.0) 2.9
  N 22 (9–40) 2368 (15–156) 69100 (67–142) 9501 (0–2) 102.3 (2–3) 2.0
  PKA, Gy cm2 5.1 (1.9–11.0) 3.714.3 (0.9–78.0) 8.54.5 (2.3–8.1) 3.725.9 (10.7–68.3) 17.53.1 (0.3–22.5) 2.43.0 (1.2–5.2) 2.7
  Ka,r, mGy22.9 (9.8–99.6) 14.958.4 (5.1–377.1) 41.127.9 (15.1–50.7) 23.039.8 (1–249) 30.5
 Pulse rate, p/s (%)30 (100%)CF (100%)12.5 (47%), 25 (53%)CF (100%)
Mini PCNL (30 Patients)No. of patients56001900
 Patient weight78.6 (72–85) 7882.0 (75–87) 8382.3 (48–118) 81.0
 SA, mm2 366 (200–484) 393118 (118–157) 118247 (122–534) 190
 Stone location100% kidney100% kidney95% kidney; 5% ureter
 PT, min109 (80–130) 12055 (50–70) 5560 (38–120) 60
 FT, min2.2 (1.2–4.1) 2.13.0 (0.3–4.9) 3.33.7 (1.2–12.1) 2.3
  N 27 (19–32) 3063 (13–89) 721.3 (0–5) 1
  PKA, Gy cm2 7.6 (3.0–13.3) 8.09.5 (0.5–16.5) 9.94.3 (0.6–37.7) 1.4
  Ka,r, mGy20.4 (12.4–30.1) 19.249.7 (2.8–86.2) 51.746.3 (7–269) 22.0
 Pulse rate, p/s (%)30 (100%)12.5 (79%), 25 (21%)
RIRS (43 patients)No. of patients50012980
 Patient weight76.6 (65–90) 757284.4 (46–150) 82.081.4 (60–110) 80.0
 SA, mm2 144 (57–294) 12682.470 (16–471) 93317(50–942) 294
 Stone location100% kidneykidney90% kidney; 10% ureter100% kidney
 PT, min64 (50–90) 6010246 (25–90) 3861 (25–106) 58
 FT, min1.7 (0.6–5.2) 0.90.51.3 (0.2–6.5) 1.01.0 (0.4–2.3) 0.7
  N 42 (7–180) 800.9 (0–3) 10
  PKA, Gy cm2 5.2 (1.7–16.8) 1.81.71.9 (0.2–8.0) 1.11.27 (0.51–3.38) 0.8
  Ka,r, mGy15.1 (5.9–39.8) 8.99.3 (<1–30) 5.0
 Pulse rate, p/s (%)CF (100%)12.5 (79%), 25 (21%)CF (100%)

Table 2. Number of patients, patient weight in kg, treated stone burden, expressed in stone surface area (SA) in mm2, stone location, total procedure time (PT) in minutes, fluoroscopy time (FT) in minutes, number of acquired images (N), kerma area product, PKA, and cumulative air kerma, Ka,r, and fluoroscopy pulse rate (p/s) for the semi-rigid URS and flexible URS, performed in the participating clinical centers U1–U7. Empty cells correspond to missing data or when statistical analysis could not be performed due to a small sample size; CF = continuous fluoroscopy). For the fluoroscopy pulse rate, the percentage in brackets indicates the frequency of using the corresponding p/s.

 Clinical centerU1U2U3U4U5U6U7
Semi-rigid URS(125 patients)No. of patients26003548160
 Patient weight74.3 (45–95) 7573.6 (45–100) 74.581.3 (58–120) 80.580.2 (62–103) 79.5
 SA, mm2 178 (14–1570) 6748 (13–207) 3870 (16–471) 5577 (28–118) 77
 Stone location50% kidney, 50% ureter100% ureter96% ureter, 4% kidney100% ureter 
 Proced. time, min47 (20–80) 4343 (18–151) 3936 (10–80) 3030 (10–50) 30
 FT, min0.4 (0.08–1.12) 0.30.8 (0.22–3.10) 0.60.4 (0–3.3) 0.10.5 (0.1–1.35) 0.3
  N 3.8 (1–10) 300.3 (0–5) 00
  PKA, Gy cm2 1.0 (0.5–1.9) 0.93.2 (0.8–12.2) 2.50.6 (0–6.7) 0.11.1 (0.02–5.3) 0.5
  Ka,r, mGy4.9 (1.9–10.8) 4.13.7 (<1–63) 1.0
 Pulse rate, p/s (%)CF (100%)12.5 (79%), 25 (21%)CF (100%)
Flexible URS (16 Patients)No. of patients1420180
 Patient weight8276.3 (50–90) 82.578 (70–86) 7812377.6 (47–97) 81.0 
 SA, mm2 11885 (27–141) 86129.6(117.8–141.3)129.670.6157 (64–491) 79 
 Stone locationkidney100% kidneykidneykidney75% ureter, 25% kidney 
 Proced. time, min8553 (30–90) 4590 (80–100) 904057 (30–120) 50
 FT, min0.10.2 (01.-0.4) 0.10.8 (0.5–1.11) 0.81.31.6 (0.1–4.2) 1.4
  N 1111 (8–17) 9119 (101–136) 11920
  PKA, Gy cm2 0.20.3 (0.1–0.9) 0.12.8 (2.7–2.9) 2.82.91.1 (0.2–2.2) 1.0
  Ka,r, mGy10.31.6 (0.3–4.8) 0.615.9 (15.1–16.8) 15.911.0
 Pulse rate, p/s (%)30 (100%)12.5CF (100%)

Majority of patients were treated with semi-rigid URS (40.0%), followed by standard PCNL (32.1%), RIRS (13.7%), mini-PCNL (9.5%) and flexible URS (5.1%). Different case-mix was observed: centers U2, U3 and U7 performed mostly PCNL, while U4 performed more URS and U5 and U6 had the whole spectrum. For URS, U1, U4 and U5 used semi-rigid scope, while U2 and U3 performed only flexible URS, and U6 had 44.4% semi-rigid and 22.2% flexible. All centers performed PCNL in prone position, except U4 routinely using supine.

Tables 1 and 2 present the patient weight, stone location, stone SA, as well as the total procedure time (PT), FT, N, PKA, Ka,r, and fluoroscopy pulse rate (p/s), per procedure and per clinical center. Data are presented with their mean and medical values, and the range (minimum–maximum).

The stone SA varied from 13 to 1570 mm2 in ureter and from 16 to 6005 mm2 in kidney (tables 1 and 2). The largest kidney stones were treated with standard PCNL (101 patients), followed by mini-PCNL (30 patients) and RIRS (43 patients). For ureteral stones, semi-rigid URS was used in majority of cases (125) versus flexible URS (16).

All patients were adult, 214 (68%) males and 101 (32%) females. All samples had a mean patient weight within 80 ± 5 kg, except two, in which the mean weight was 73.5 kg and 96 kg correspondingly. This indicates that the patient weight might not be a factor when comparing dose values between centers.

The correlation coefficient r2 between the stone SA and the total PT varied between 0.001 for the semi-rigid URS in U1 (26 patients) and 0.409 for the standard PCNL in U5 (36 patients). The correlation coefficient r2 between SA and FT was between 0.001 for RIRS in U5 (29 patients) and 0.17 for the standard PCNL in U1 (10 patients). The correlation coefficient r2 between PKA and FT varied from 0.63 to 0.80 for the standard PCNL performed in centers U1, U2 and U5, from 0.75 to 0.90 for mini PCNL performed in U1, U2 and U5, from 0.20 to 0.99 for RIRS performed in U1 and U5, and from 0.08 to 0.52 for both semi-rigid and flexible URS performed in U1, U4, U5 and U6.

Table 3 presents the results for all type of procedures of the nonparametric ANOVA using the Kruskal–Wallis test of the significance of the difference between values of FT and PKA. A p-value less than 0.05 indicates that there is a significant difference somewhere among the various samples. The difference between FT values for the same procedure performed in different centers was statistically significant for the standard PCNL (p = 0.0125) and semi-rigid URS (p = 0.0226), while no significant difference was observed for the other procedures with less statistical data like mini PCNL (p =0.4894), RIRS (p = 0.7477) and flexible URS (p = 0.1896). The difference between PKA values for the same procedure performed in different centers was statistically significant for all procedures (table 3, last column).

Table 3. Results of the nonparametric ANOVA using the Kruskal–Wallis Test of the difference between values of fluoroscopy time, FT, and kerma-area product, PKA, for all type of procedures. A p-value less than 0.05 indicates that there is a significant difference somewhere among the various groups.

Procedure typeNumber of casesData groups p-value for FT p-value for PKA
Standard PCNL111U1, U2, U3, U4, U5, U6, U7 0.0125 0.0001
Mini PCNL30U1, U2, U50.4894 0.0009
RIRS43U1, U5, U60.7477 0.0415
Semi-rigid URS125U1, U4, U5, U6 0.0226 0.0001
Flexible URS16U2, U3, U60.1896 0.0019

Figures 1 and 2 show the median values of PKA and Ka,r , for all procedures performed in different clinical centers. Medians are selected for this presentation because of the non-normal distribution of some samples.

Figure 1.

Figure 1. Median values of the kerma-area product, PKA, for all procedures performed in centers U1U6.

Standard image High-resolution image
Figure 2.

Figure 2. Median values of the cumulative air kerma, Ka,r, for all procedures performed in centers U1, U2, U3 and U5.

Standard image High-resolution image

Table 4 presents the findings of other published studies in terms of PKA and FT for PCNL and URS.

Table 4. Patient exposure in kerma area product, PKA, and fluoroscopy time (FT) for PCNL and URS in other published studies.

Authors [Ref]Procedure typeNumber of casesFT (min) Mean (Min–Max) Median PKA (cGy cm2) Mean (Min–Max) Median
Hristova-Popova et al [7]PCNL URS52 524.0 (0.4–13.7) 3.7 1.7 (0.3–6.1) 1.14.57 (0.54–24.40) 3.54 5.90 (0.19–23.10) 3.46
Hanna et al [8]PCNL3481.64.52
Kumari et al [9]PCNL506.04 (1.8–2.16)
Safak et al [11]PCNL201227.00 (calculated)
Hristova-Popova et al [15]PCNL URS16 154.5 (0.5–12.5) 3.7 0.9 (0.1–3.4) 0.410.10 (1.96–42.67) 6.64 3.65 (0.22–13.93) 2.28
Medici et al [16]URS210.9 (0.2–2.4) 0.88.50 (0.50–23.10) 5.70
Ritter et al [17]PCNL URS11 397.3 (5.3–15.7) 1.1 (0.2–13.9)
Vano et al [18]PCNL3411.5 (5.1–23.9) 7.335.70 (10.60–85.40) 29.60

4. Discussion

Urolithiasis is a common health issue affecting about 13% of male and 7% of female population and represents a major cause of emergency department visits and a great economic health burden [24]. Minimally invasive procedures rapidly replace open surgery and become the most effective treatment option even for larger stones in anatomically difficult systems. Safe entrance into urinary tract and fragmentation-removal of stones is accomplished with fluoroscopic guidance. To our best knowledge, this is the first multicentric study of radiation exposure associated with endourological procedures and related factors.

The total PT varied from 10 to 225 min, with the higher values for PCNL (12–225 min) and lower for URS (10–151 min). Weak or no correlation was found between the stone SA and the total PT (r2 from 0.001 to maximum 0.4), indicating that there are other factors affecting the procedure duration, such as the stone composition and location.

Fluoroscopy was used in a small part of the procedure; FT varied from 0.1 min to 14 min for different patients, with the maximum for standard PCNL. FT was similar to data reported in other studies (table 4), except the study of Vano et al where FT for PCNL was higher [9 11, 13, 17 20].

Median FT, which better represents the typical value in samples with non-normal distribution, varied 5-fold (from 1.1 to 5.4 min) for the standard PCNL, and 6-fold (from 0.1 to 0.6 min) for the semi-rigid URS (tables 1 and 2). This might be influenced by the different number of cases in each center, but considering that for both procedures the Kruskal–Wallis test found statistical differences of FT between different centers, variations might be mostly related to a different practice of use of fluoroscopy and different complexity of the procedures due to the stone size, structure and location.

No correlation was found between stone SA and FT (the maximum r2 was 0.17), showing a potential impact of confounding factors, also discussed by Safak et al [13]. Specifically, the lithotripsy phase does not require fluoroscopy guidance and exposure depends on body habitus and morphology of the urinary tract that can make more difficult the percutaneous access, or the passage of the scope requiring repeated fluoroscopic checks. For example, the largest kidney stone with a SA of 6015 mm2 in U5 was treated with the standard PCNL that needed 120 min total procedural time, of which only in 2.1 min fluoroscopy was used. Much smaller kidney stone of SA of 236 mm2 required 7.5 min FT due to the more difficult access to the kidney and only 60 min to finish the procedure because of the easier lithotripsy phase.

High variations were found in the number of acquired images—from zero in U4 and U6, up to 156 for standard PCNL and 89 for mini-PCNL in U2, 180 for RIRS and 10 for semi rigid URS in U1 and 136 for flexible URS in U3. It is well known that the use of radiography increases patient dose, and whenever available, LIH has to be used instead [2].

The differences in FT, N and system settings resulted in significant variations of doses between individual patients and between centers. The lowest individual PKA of 0.01 Gy cm2 was recorded for a semi rigid URS case, and the maximum 72.30 Gy cm2 for a standard PCNL case. The values of Ka,r varied between < 1 mGy in short RIRS and semi-rigid URS, to 377 mGy for PCNL. The highest PKA and Ka,r values were observed for the same patient case in U2, treating with PCNL by multiple access a kidney stone with SA of 942 mm2, that resulted in the highest FT of 14 min and 156 radiography images. In addition, pulsed fluoroscopy with 30 p s-1 was routinely used in this center, indicating a potential for dose reduction if lowering pulse rate and reducing the number of acquired radiography images.

Median PKA (tables 1 and 2, and figure 1) varied 2 fold for RIRS (0.80–1.79 Gy cm2), 7.1 fold for mini-PCNL (1.39–9.90 Gy cm2), 7.3 fold for PCNL (2.40–17.50 Gy cm2), 19 fold (0.13–2.51 Gy cm2) for semi-rigid URS and 29 fold for flexible URS (0.10–2.90 Gy cm2). The difference between median PKA for the same procedure in different centers was statistically significant for all procedure types (table 3). Higher median PKA (from 1.4 to 17.5 Gy cm2) were observed for the PCNL procedures compared to the URS procedures (from 0.1 to 2.9 Gy cm2).

Good correlation was found in most departments and procedures between PKA and FT (r2 between 0.5 and 0.9 in most samples) that is reasonable since FT is the main factor affecting patient dose along with N and pulse rate. For example, in U5, for 71% of cases 12.5 p s-1 was used and 25 p s-1 in the rest, which explains the relatively lower doses compared to U2 where 30 p s-1 was always used for all procedures, and U4 and U6 where continuous fluoroscopy (CF) is used. Much higher values of PKA for PCNL in U4 might be also due to the use of supine position.

The effect of pulse rate was studied further for the center U5 by comparing two normally distributed samples of PCNL with the same mean FT = 2.5 min performed with different pulse rate. The use of 12.5 p s-1 resulted in a significantly lower PKA (mean = 1.78 Gy cm2, SD = 1.17 Gy cm2) compared to the use of 25 p s-1 (mean = 3.25 Gy cm2, SD = 1.34 Gy cm2), t(33) = −3.5, p = 0.002. A similar significant decrease was found for Ka,r from PCNL performed with 12.5 p s-1 (mean = 25.0 mGy, SD = 18.3 mGy) compared to the use of 25 p s-1 (mean = 42.2 mGy, SD = 23.4 mGy), t(33) = −2.41, p = 0.022.

As reported in the first paper from this study [21], which presents the measured doses to the whole body and eye lens of the operating room staff, a strong correlation (r2 between 0.91 and 0.99) was found between the measured eye lens dose and the total PKA of procedures performed by the corresponding urologist. This confirms that the optimal use of fluoroscopy has a positive impact on not only patient dose, but staff dose as well. As an example, although the fluoroscopy system in U5 has an x-ray tube over the table, because of the use of lower FT, N, pulsed fluoroscopy of 12.5 p s-1 and good collimation, doses to eye lens of staff and patient doses per procedure were lower in this center compared to U4 and U6 [21].

Although not studied separately, some higher PKA values can be associated also with an inappropriate beam collimation—when keeping other exposure factors unchanged, decreasing the field area linearly decreases PKA. Additional factors not studied are the distance of x-ray tube and image detector relative to patient, as well as the tube voltage and tube filtration defined by ADRC [2].

The maximum Ka,r value of 377 mGy for a patient treated with PCNL in U2 is much lower than the dose threshold of around 2 Gy that is needed to observe radiation induced skin erythema [2, 4], confirming the findings of Safak et al [13]. The typical (median) values of Ka,r varied between 0.6 mGy and 51.7 mGy.

PKA and Ka,r values for RIRS were significantly lower compared to PCNL performed in the same centers. Considering also its higher safety and repeatability, RIRS is a valuable option for renal stone treatment [1, 24].

There are few published studies of patient doses in endourology (table 4). Median PKA for PCNL in our study ranges between 2.40 and 17.50 Gy cm2, which is comparable to the findings of Hristova-Popova et al [9, 17] and Medici et al [18], and lower than the value reported by Vano et al who also observed higher FT [20]. For URS, median PKA is comparable to the values reported by Hristova-Popova et al [9, 17] and lower compared to Medici et al [18].

5. Conclusions

The study demonstrated variations in the endourology practice between urology centers, that results in high variations in patient doses, as well as variation in staff doses [21]. This is mainly related to the use of different fluoroscopy systems (C-arm with x-ray tube under the table and fixed system with tube over the table), use of CF versus pulsed fluoroscopy with different pulse rate, different patient positioning for PCNL (supine or prone), and different pattern of recording radiography images.

No direct correlation was found between the concrement size and total duration of procedure, FT or dose parameters kerma area product and air kerma at the patient entrance reference point that indicates that there are other concurring factors, e.g. patient condition, concernment structure, stone position, etc, which have not been studied here. Further studies might be needed to find a better descriptor of the complexity of the endouroly procedures and link it to dose data.

Skin doses were significantly lower the threshold and no skin injury can be expected for patients in endourology.

The study showed potential for patient dose reduction by paying attention to the following factors: use of low-dose pulsed fluoroscopy (12.5 p s-1 or lower when available), better beam collimation to the area of interest, no or less use of electronic magnification, reduction of FT and number of recorded images and use of LIH instead of radiography whenever possible. Additional reduction can be achieved by keeping the x-ray tube further and image detector closer to the patient body [2, 25]. Endourologists must understand and optimally use these factors to reduce simultaneously patient and staff exposure and optimise the benefit-risk ratio. Radiation protection needs to be an integral part of the training programs of endourologists, with an emphasis on the practical dose management and optimisation [2, 26]. The study demonstrated the need to further strentgen quality control program and involve qualified medical physicist to advise on dose reduction approaches.

Acknowledgments

The study was organised by the South-Eastern European Group for Urolithiasis Research (SEGUR) and was supported under the IAEA Technical cooperation projects RER9135 'Strengthening Radiation Protection of Patients and Medical Exposure Control'. The authors acknowledge the contribution of Ognyan Gatsev, Stefan Hristoforov and Petar Petrov, all from the Department of Urology and Nephrology, Military Medical Academy, Sofia, Bulgaria.

Please wait… references are loading.
10.1088/1361-6498/abc351