Clinical relative biological effectiveness of low-energy x-rays emitted by miniature x-ray devices

, , and

Published under licence by IOP Publishing Ltd
, , Citation David J Brenner et al 1999 Phys. Med. Biol. 44 323 DOI 10.1088/0031-9155/44/2/002

0031-9155/44/2/323

Abstract

Several groups are developing ultra-miniature x-ray machines for clinical use in radiation therapy. Current systems are for interstitial radiosurgery and for intravascular insertion for irradiation to prevent re-stenosis. Typical generating voltages are low, in the 20 to 40 kV range. It is well established that the biological effectiveness of such low-energy photons is large compared with higher-energy gamma rays, because of the dominance of photoelectric absorption at low energies. We have used microdosimetric analyses to estimate RBEs for such devices, both at low doses and clinically relevant doses, relative to radiations from , , and Y. The RBEs at clinically relevant doses and dose rates for these low-energy x-ray sources are considerably above unity, both relative to and to photons, and also relative to and Y brachytherapy sources. As a function of depth, the overall effect of the change in dose and the change in beam spectrum results in beams whose biologically weighted dose (doseRBE) decreases with depth somewhat more slowly than does the physical dose. The estimated clinically relevant RBEs are sufficiently large that they should be taken into account during the treatment design stage.