The measurement of proton stopping power using proton-cone-beam computed tomography

, , and

Published under licence by IOP Publishing Ltd
, , Citation Piotr Zygmanski et al 2000 Phys. Med. Biol. 45 511 DOI 10.1088/0031-9155/45/2/317

0031-9155/45/2/511

Abstract

A cone-beam computed tomography (CT) system utilizing a proton beam has been developed and tested. The cone beam is produced by scattering a 160 MeV proton beam with a modifier that results in a signal in the detector system, which decreases monotonically with depth in the medium. The detector system consists of a Gd2 O2 S:Tb intensifying screen viewed by a cooled CCD camera. The Feldkamp-Davis-Kress cone-beam reconstruction algorithm is applied to the projection data to obtain the CT voxel data representing proton stopping power. The system described is capable of reconstructing data over a 16 × 16 × 16 cm3volume into 512 × 512 × 512 voxels. A spatial and contrast resolution phantom was scanned to determine the performance of the system. Spatial resolution is significantly degraded by multiple Coulomb scattering effects. Comparison of the reconstructed proton CT values with x-ray CT derived proton stopping powers shows that there may be some advantage to obtaining stopping powers directly with proton CT. The system described suggests a possible practical method of obtaining this measurement in vivo .