Hand-held compared with isokinetic dynamometry for measurement of static knee extension torque (parallel reliability of dynamometers)

Published under licence by IOP Publishing Ltd
, , Citation R W Bohannon 1990 Clin. Phys. Physiol. Meas. 11 217 DOI 10.1088/0143-0815/11/3/004

0143-0815/11/3/217

Abstract

The author compares measurements of static knee extension torque obtained using a hand-held dynamometer (HHD) and an isokinetic dynamometer (IKD). Twenty young healthy women performed two maximum knee extensions efforts against each device. The two HHD measurements did not differ significantly and were highly reliable (intra-class correlation coefficient (ICC)=0.945). The two IKD measurements did not differ significantly and were highly reliable (ICC=0.932). The mean of the two HHD and two IKD measurements did not differ significantly. The inter-instrument reliability was fair (ICC=0.797). Thus, under limited circumstances the instruments may be interchangeable. The less expensive and more portable HHD may be a practical alternative for the clinical measurement of muscle strength.