2 March 2018 In vivo imaging in the oral cavity by endoscopic optical coherence tomography
Julia Walther, Christian Schnabel, Florian Tetschke, Tobias Rosenauer D.D.S., Jonas Golde, Nadja Ebert, Michael Baumann, Christian Hannig D.D.S., Edmund Koch
Author Affiliations +
Abstract
The common way to diagnose hard and soft tissue irregularities in the oral cavity is initially the visual inspection by an experienced dentist followed by further medical examinations, such as radiological imaging and/or histopathological investigation. For the diagnosis of oral hard and soft tissues, the detection of early transformations is mostly hampered by poor visual access, low specificity of the diagnosis techniques, and/or limited feasibility of frequent screenings. Therefore, optical noninvasive diagnosis of oral tissue is promising to improve the accuracy of oral screening. Considering this demand, a rigid handheld endoscopic scanner was developed for optical coherence tomography (OCT). The novelty is the usage of a commercially near-infrared endoscope with fitting optics in combination with an established spectral-domain OCT system of our workgroup. By reaching a high spatial resolution, in vivo images of anterior and especially posterior dental and mucosal tissues were obtained from the oral cavity of two volunteers. The convincing image quality of the endoscopic OCT device is particularly obvious for the imaging of different regions of the human soft palate with highly scattering fibrous layer and capillary network within the lamina propria.

1.

Introduction

Visual examination is usually the first step toward diagnosis of diseases within the oral cavity, such as carious lesions, periodontitis, oral inflammation, dysplasia, and malignancies. Sometimes, the recognition and diagnosis of altered tissue, even by an experienced physician or dentist may be difficult and imprecise, especially in early stages of pathologic phenomena. Due to the multiplicity of diseases of the oral cavity, commonly coupled with changes in tissue structure, high resolution diagnostic imaging modalities in routine clinical dentistry and maxillofacial surgery are beneficial for the detection and treatment of lesions. Current imaging modalities in dentistry include conventional x-ray radiography, computed and digital volume tomography (CT, DVT), magnetic resonance tomography (MRI), and sonography (ultrasound, US) for the in vivo diagnosis of altered tissue in the oral cavity. While all of those imaging techniques are well established, the diagnosis of submillimeter structure is still difficult (for e.g., US) or rather expensive in clinical routine (e.g., CT/DVT, MRI). Present reliable diagnosis of pathological changes in soft tissues, such as the oral mucosa, gingiva, and tongue, normally requires repeated biopsy and histological examination of conspicuous lesions with the drawback of delayed detection of malignant changes for curative treatment. In addition, versatile appearance of mucosal tissue alterations, in combination with absent routine inspection of patients for oral lesions, may lead to late diagnosis. Therefore, effective diagnostic modalities would be of exceptional clinical relevance. Latest established optical imaging methods for in vivo detection of mucosal changes are spectroscopy and confocal imaging, which are still challenging in clinical routine due to small field-of-views (FoV) and low penetration depths. A complementary imaging technique is optical coherence tomography (OCT) providing cross sections (B-scans) and volume views of tissue microstructures of depths up to 1 to 2 mm with the feature of detecting changes in tissue scattering.14 Even though OCT does not deliver subcellular resolution or molecular contrast, it has the potential of label-free optical biopsies of tissue in vivo by a simple noninvasive subsurface view to assess depth structures of oral soft and hard tissue lesions. Most research has presented the feasibility of OCT in various preclinical animal57 and human810 oral cavity studies. Although OCT enables functional imaging, such as Doppler OCT,1113 polarization-sensitive OCT1416 and spectroscopic OCT,17,18 the core strength of oral examination by OCT convincingly lies in the easily applied morphological imaging valued by a physician similar to US examinations.

In dental research, as a main field of application with regard to the oral cavity, OCT has become a promising approach for the investigation of the dental microstructure. So far, the majority of studies using OCT in dentistry are based on ex vivo investigations, either with artificial alteration models or at inactive lesions on extracted teeth. Nevertheless, a small number of promising dental OCT studies dealing with the in vivo detection of caries lesions can be found.1923 Compared to the current gold standard of radiographic examination, OCT provides a higher spatial resolution and contrast for hard tissue and allows the discrimination of enamel and dentin, as well as the recognition of the dentin–enamel junction (EDJ).24,25 Furthermore, de- and remineralization processes can be measured26,27 and thus makes OCT suitable for the spatially resolved detection of early pathologic changes.2831 Moreover, OCT overcomes the drawbacks of superimposed imaging with ionizing radiation and enables the observation of short-term progression in carious dental hard tissue.32,33 Since the decision for or against invasive treatment is dependent on the progress of a lesion, this information could be of great value for the clinician.34 On the other hand, the application of OCT is limited by the relatively low penetration depth, compared to the radiographic imaging.

In oral cancer diagnosis, as another application field regarding the oral cavity, the structural imaging by OCT for early and/or differentiating detection of malign tissue has to be assessed critically. OCT provides assessment of the entire epithelium and partly the underlying lamina propria (LP) and has the potential to distinguish normal from pathologic mucosal tissue of different entities.5,6,3539 The distinction between normal and pathologic tissue is mainly based on the measurement of the epithelial thickness, the identification of the adjacent basement membrane, and the vasculature within the LP, as well as the changed backscattering behavior of the altered tissue lesions. The most promising approach in this context is the potential for biopsy guidance by OCT in conjunction with an improvement of diagnostic accuracy of biopsies.4042

Since OCT is a promising imaging modality for the noninvasive in vivo examination of oral soft and hard tissue, the demand for intraoral miniaturized or rather endoscopic OCT (EOCT) optics is substantial especially for the accessibility of the posterior oral cavity.43 Although a variety of impressive EOCT solutions are proposed in flexible catheter-based configuration,44,45 especially for intravascular,46 bronchial47 as well as gastrointestinal diagnosis,48,49 in rigid implementation, in particular, for otorhinolaryngology5052 and in needle-like design for intratissue examinations in cancer diagnosis,5356 developments of intraoral OCT probes adapted to the oral cavity can only be found occasionally.5759 So far, rigid handheld endoscopic probes,20,6063 miniaturized probes,64,65 and rotary pullback catheters66,67 are developed in forward and side viewing configuration, as well as in contact and noncontact mode for intraoral measurements, which is a crucial step for the application of OCT methodology in clinical practice. In our judgment, most of the reported rigid OCT systems allow the imaging of the well accessible buccal mucosa, the hard and anterior soft palate as well as different parts of the tongue, but none of them has demonstrated contactless OCT images of different areas of the posterior oral mucosa (e.g., palatoglossal arch and fold) with highly resolved structures of the connective tissue and molars in vivo.

The aim of the present study is to expand intraoral application of OCT by a simple adaption of a commercially available endoscope. The development of a handheld endoscopic forward viewing OCT probe enables high quality two- and three-dimensional imaging of teeth and mucosa of the anterior and posterior oral cavity to pave the way for in vivo noninvasive tissue biopsy. The potential of the presented imaging system is demonstrated by in vivo measurements of healthy oral mucosa and conspicuous parts of teeth by means of two volunteers.

2.

Endoscopic System Setup

2.1.

Specification of the Rigid Endoscopic Optics

In this research, the rigid endoscopic optics used for OCT imaging of the oral cavity is a commercially available near-infrared (NIR) endoscope from Karl Storz GmbH & Co. KG. The NIR endoscope has an insertion diameter of 10 mm, a length of 200 mm, and an angle of view of 0 deg (instrument axis consistent with optical axis) and is assumed to be based on afocal optics with a 0.25-fold magnification. For a center wavelength of λcenter=580  nm, a lateral resolution of Δx=62.5  μm at a working distance of WD 49  mm is determined by means of an externally illuminated USAF resolution test target (USAF-1951-target) in combination with a proximally positioned VIS camera focused to infinity. In order to get a focus on the distal side of the endoscope and to increase the lateral resolution, the optics in front of the proximal end of the endoscope was simply adapted. As a result, a lateral resolution of Δx=12.4  μm at WD 7.5  mm and λcenter=580  nm is achieved [cf. Fig. 1(b)], which corresponds to Δx=17.9  μm for the NIR wavelength range with λcenter=835  nm under the assumption of Δxλcenter. The FoV in dependency on WD is determined by a millimeter paper test chart with the result of a linear relationship with a slope of m=1.1, as shown in Fig. 2(c).

Fig. 1

By adapting the proximal optics for light coupling, a lateral resolution of Δx=12.4  μm of the commercial NIR endoscope from Karl Storz GmbH & Co. KG is measured by means of a proximal VIS camera in combination with the USAF resolution test target, which is positioned at a working distance of WD 7.5  mm and illuminated by a center wavelength of λcenter=580  nm.

JBO_23_7_071207_f001.png

Fig. 2

(a,b) Endoscopic VIS images of a millimeter paper test chart, which is illuminated by the integrated illumination system of the endoscope, for measuring the relation between the FoV and the WD. (c) At the favorable WD of 5 to 10 mm, the FoV with fan-shaped beam increased linearly from 8.3 to 14.5 mm.

JBO_23_7_071207_f002.png

2.2.

Endoscopic Optical Coherence Tomography System Setup

The OCT system used in this study is spectrometer-based and consists of a fiber-coupled superluminescent diode (SLD-371-HP1, Superlumdiodes Ltd., Ireland) with a central wavelength of 835 nm and a FWHM of 50 nm, an optical circulator (CIR-0850-40-APC, Thorlabs GmbH, Germany), a fiber-coupled Michelson interferometer and a self-developed spectrometer with a wavelength range from 802 to 867 nm. As presented in Fig. 3, the scanning EOCT probe is equipped with a collimator C1 (Schäfter + Kirchhoff GmbH, Germany), two galvanometer scanners GSx and GSy (Cambridge Technology Inc., USA) for 2-D beam deflection, a lens combination consisting of two achromats L1+L2 for light coupling (Edmund Optics GmbH, Germany), the NIR endoscope from Karl Storz GmbH & Co. KG, and an additional objective lens L3 (AC050-008-B, Thorlabs GmbH, Germany; at 4-mm distance to the distal end of the endoscope) for telecentric OCT imaging. As seen, the difference in glass content between sample and reference arm is not compensated optically but numerically in the processing.68,69

Fig. 3

(a) Scheme of the principle OCT setup. SLD, 5-mW superluminescent diode with λcenter=835  nm and FWHM=50  nm; OC, optical circulator; FC, wideband fiber coupler with a coupling ratio of 7525; C1 and C2, collimators; GS, galvanometer scanners; L1–L4, lenses; endoscope, endoscope from Karl Storz GmbH and Co. KG; CLS, cold-light source; PC, polarization controller; RM, reference mirror; spectrometer, line rate of 11.88 kHz. (b) Telecentric imaging is realized by an additional lens L3 mounted on the distal end of the commercial endoscope at 4 mm distance by a custom adapter made of stainless steel.

JBO_23_7_071207_f003.png

The optical components C1, GSx, GSy, L1 and L2 are integrated in a base body made of polylactide (PLA), at which the chosen endoscope, a CCD VIS camera, and an ergonomic handle are mounted, as shown in Fig. 4(a). Lens L3 is pasted into an additional adapter made of stainless steel by two set screws (M2) and secured by optical glue (Vitralit®1655, Panacol-Elosol GmbH, Germany). Since the distal lens allows imaging only in a small range, a CCD camera (acA1600-20uc, Basler AG) is installed at the base body of the handheld scanner head for online off-axis video guidance by means of an overview image of the oral cavity [cf. Fig. 4(a), yellow beam path], which is displayed on the integrated LCD screen (LCDInfo Oy, Finland). The custom-developed spectrometer, operating at a read-out rate of 11.88 kHz for one A-scan, is mounted in an OCT table device [cf. Fig. 4(b)] along with the SLD and the optical circulator, as well as the drive electronics of the galvanometer scanners and electronic control of the line scan detector (Dalsa IL C6, Teledyne DALSA Inc., Canada).70 The control of the system, the data acquisition, and online display are controlled by custom software developed with LabVIEW® (National Instruments Inc., USA) on a personal computer [cf. Fig. 4(c)]. With typical 480 A-scans per frame (B-scan), the system achieves 23  frames/s, which can be processed and displayed in real time.

Fig. 4

Scheme of the measurement setup for the endoscopic investigation of the oral cavity equipped with (a) the EOCT probe with integrated scanning sample arm, (b) the base OCT device with built-in SLD, spectrometer, partial fiber optics and system electronics, and (c) a personal computer for the control and triggering of the electronic components as well as the data display and acquisition.

JBO_23_7_071207_f004.png

The overall dimensions of the developed scanner head including the proximal handle are 76  mm×442  mm×251  mm (width×depth×height) with an endoscopic part relevant for the mouth of a length of 200 mm and a diameter of 15 mm tapered to 10 mm at the distal end [cf. Fig. 15 in Appendix]. The height of the scanner body is mainly defined by the height of the handle (100 mm) and the height of the LCD screen (130 mm). With this design, most of the oral cavity is reached proven by two experienced dentists and one physician. Limitation of the developed endoscope might occur for the imaging of the buccal side and the interdental space of second and third molars of patients with reduced mouth opening. The endoscopic probe enables imaging in noncontact and contact mode by using an additional spacer (distance piece), which is simply mounted on the lens adapter of lens L3 before medical examination and acts as a disposable item for future clinical use. Advantageously, the endoscope, considered as an attachment to the scanner head, can be cleaned and disinfected (by immersing into 3% Sekusept® aktiv solution, Ecolab Deutschland GmbH, Germany) in the daily hospital routine. If necessary, steam sterilization can be performed by disconnecting the commercial endoscope.

2.3.

EOCT Scanner Performance

In order to evaluate the performance of the developed EOCT probe with telecentric scanning, several measurements were performed for the quantification of the axial and lateral resolution (Δz,Δx), the FoV, the sensitivity, and the spectrometer-induced depth-dependent sensitivity loss. The axial resolution of the OCT system without the endoscopic extension is experimentally measured to be Δz=11  μm in air and in good agreement with the theoretical value of Δz=10.8  μm. The measurement range is 2.75 mm. Since the glass content in the sample arm is higher than in the reference arm, the occurring dispersion mismatch is numerically compensated for optimal resolution.68,71 This is achieved by multiplying the spectral data (which is formerly linearized in k-space, zero-padded to 212  pixels and spectrally shaped) with a compensating phase term eiφ(k).69,72 The empirically determined phase correction amounts to

Eq. (1)

φ(k)=a(kk0Δk)2+b(kk0Δk)3witha=150  radandb=3  rad,
where the coefficient a is the dominant second order term, compensating the group velocity dispersion, and with this, the broadening of the axial resolution in OCT, and coefficient b is the very low third-order term eliminating the asymmetric distortion. k0 is the mean wavenumber and Δk is the half width of the spectrum captured with the detector, so that the terms within the brackets of Eq. (1) run from 1 to +1. The comparison of the experimentally obtained quadratic term of the compensating phase shift with data calculated from the Schott glass database results in glass length from 75 mm of SF66 up to 620 mm of NFK51A or any appropriate mixture of glasses. As the total length of the endoscope is 265 mm in addition to lenses in the base body of the scanner, this result seems to be reasonable. However, this calculation does not allow a more accurate insight into the glass composition of the endoscope. The cubic term of the phase shift gives no additional information on the glass type as this is 2% of the quadratic term for most glasses considering the spectral range of this OCT system. The experimentally obtained axial resolution of Δz=11.6  μm in air (Δz=8.7  μm in tissue) is measured by means of the reflection of a glass plate and corresponds approximately to the one achieved with the scanner head without endoscope.73 The lateral resolution is measured by telecentric OCT imaging of the USAF resolution test target and amounts to Δx=17.5  μm [cf. Fig. 5(a)], which confirms the measurement result in Sec. 2.1. The FoV, achieved with the telecentric OCT imaging (cf. Fig. 3), is determined by means of the mm-paper target and results in xmax=ymax=4.8  mm, as depicted in Fig. 5(b).

Fig. 5

En face projection of the USAF-1951-target (a) as well as a millimeter-chart (b) detected by telecentric EOCT for determining the lateral resolution and the FoV. (c) Corresponding OCT cross-sectional image (B-scan) of the millimeter-chart endoscopically detected. Scale bar in the cross-sectional image (c) corresponds to 500  μm in air.

JBO_23_7_071207_f005.png

The sensitivity is obtained from a glass substrate surface (14  dB reflector) and amounts to 88  dB. Due to the decreasing resolvability of interference signals with increasing frequency on the line detector of the spectrometer, a signal loss of 8 dB (sensitivity-roll-off) between zero and maximum depth is determined.

3.

Endoscopic Optical Coherence Tomography Imaging in the Oral Cavity

3.1.

Dental Imaging

The most promising application fields for OCT in dentistry are (1) the detection of carious lesions and the characterization of (de)mineralization processes, (2) the detection of defects at dental restorations, and (3) the assessment of erosion and cracks. Particularly for the imaging of well-defined boundaries, such as the interfacial layer between a dental restoration and the natural tooth surface, OCT might be a beneficial tool for the dentist to assess internal defects.74,75

For testing the developed EOCT system at dental hard tissue, two volunteers were examined (A: male 30 years, B: female 33 years). Prior to the OCT measurements, oral examination was made by an experienced dentist in order to select specific regions for imaging. Additionally, photographic overviews of the oral cavity were made for the allocation of the imaged regions. First, OCT measurements of the incisors of volunteer A are performed with the handheld scanner head in noncontact mode. In order to reduce movement artifacts for the imaging of the premolars and molars, the distal end of the endoscope is equipped with an attached distance piece, which is positioned at the region of interest of the tooth.

Figure 6 shows the photograph of the maxillary central incisors [Figs. 6(a) and 6(b)] as well as the OCT B-scans [Figs. 6(c)6(e)] at the corresponding lines in the zoom view in Fig. 6(b). In Fig. 6(c), the proximal region of tooth 11 and tooth 21 was imaged [at the red line in Fig. 6(b)]. The EDJ can be recognized by a well-defined, low-scattering band between the enamel (E) and the dentin (D).19,24,25 Conspicuously, the enamel of both tooth 11 and 21 appears as light-dark progression containing homogeneous bands of varying intensity, which is assumed to be caused by the birefringent properties. In former research, it was shown that enamel appears as layered structure in polarization-sensitive (PS)OCT.76,77 Although the light source is not polarized, the system presented in this work is sensitive to birefringence.78 Figure 6(d) shows a shadow artifact caused by a strong scattering microcrack in the upper region of the enamel [asterisk, cf. Refs. 31, 79, and 80, at the green line in Fig. 6(b)]. Due to increased scattering, higher intensities and a reduced penetration depth of the OCT signal can be observed at the gingiva in Fig. 6(e) [cf. Refs. 57, 75, and 81, at the blue line in Fig. 6(b)].

Fig. 6

(a) Photographic overview and (b) magnification of the maxillary central incisors with corresponding exemplary OCT B-scans of the proximal region between tooth 11 and 21 (c, red), the enamel (E), the EDJ and the dentin (D) of tooth 11 (d, green) with breakage (asterisk), and the intersection between enamel and gingiva (e, blue). The OCT measurement was performed in noncontact mode. Abbreviations: E, enamel; D, dentin; EDJ, enamel–dentin junction; G, gingiva; CR, composite restoration. Coordinate system in (d) is equivalent in image part (c) and (e). Scale bar in the OCT images (c–e) corresponds to 300  μm. Indicated dynamic range is valid for all depicted OCT images.

JBO_23_7_071207_f006.png

The composite restoration (CR) at tooth 11 [cf. magenta frame in Fig. 6(b)] is characterized by a homogeneous scattering layer in the OCT B-scans, as presented in Figs. 7(a) and 7(b). Interfacial gaps (green arrow in a) or even small cracks (green arrow in b) are visible at different positions in the restoration, which was found by previous nonendoscopic ex vivo research results.31,74,82,83 The 3-D representation reveals a roughened intersectional profile of the restoration (c,d). There, two projections were chosen: (c) with manually manipulated sectional view (artificial edges marked in bright green) to image the interfacial gap of the composite restoration in the three-dimensional context and (d) with intact and untreated surface of the imaged composite-enamel-passage.

Fig. 7

Cross-sectional images (a, b) and 3-D representation (c, d) of the composite restoration (CR) within the magenta frame in Fig. 6(b). Green arrows show gaps (a) or even small cracks (b) in the restoration. The dental surface of the 3-D display in panel (c) is manually manipulated in the postprocessing (artificial edges marked in bright green) to provide a more instructive view of the interfacial gap of the composite restoration. Again, the OCT measurement was performed in noncontact mode. Abbreviations: E, enamel; EDJ, enamel–dentin junction; CR, composite restoration. Scale bar in (a) and (b) corresponds to 300  μm. Indicated dynamic range is valid for all depicted OCT images.

JBO_23_7_071207_f007.png

Beyond the OCT imaging at the maxillary central incisors, the first premolars of both volunteers were examined by the developed EOCT scanner head. The upper row in Fig. 8 [parts 8(a)8(c)] represents the OCT B-scan and 3-D projection of the cervical area of tooth 24 in subject B, showing the exposed dentin at the transition to the gingiva. Furthermore, a small white spot can be identified in the reduced enamel layer without clear diagnosis. The OCT examination of the mineralization defect at tooth 24 in subject A can be recognized by a high scattering band in the OCT scan in Figs. 8(d) and 8(e), appearing in a depth up to 430  μm. Again, the normal enamel (E) appears as light-dark changes in the right image part in Fig. 8(d).

Fig. 8

Upper row shows a single exemplary OCT B-scan (a) and generated 3-D scan (b) of the cervical area of tooth 24 of subject B (c). The OCT examination was performed contactless. Lower row presents the OCT results of the mineralization defect at tooth 24 of proband A (f) in 2-D (d,e). The second in vivo OCT measurement (d,e) was realized in contact mode by an additional distance piece (disposable item) at the distal endoscopic end. Abbreviations: E, enamel; D, dentin; G, gingiva. Scale bar in part (a), (d), and (e) corresponds to 300  μm. Indicated dynamic range is valid for all depicted OCT images.

JBO_23_7_071207_f008.png

EOCT measurements were performed up to the region of the posterior teeth. A small fracture at the buccal side of tooth 16 in subject A is shown in Fig. 9. The B-scan and 3-D representation indicate that also the subjacent enamel layer is affected by the fracture. As already shown at the microcrack in Fig. 6(d), a shadowing effect due to higher backscattering of the incident light at the breakage can be found below the inset fractures.

Fig. 9

Exemplary B-scan (a) and 3-D representation (b) of a fracture at the buccal area of tooth 16 in subject A. The in vivo OCT imaging was executed in contact mode by the aid of a distal distance piece (disposable item). Scale bar in (a) corresponds to 300  μm. Dynamic range of the depicted OCT cross-sectional scans amounts to 50 dB.

JBO_23_7_071207_f009.png

In addition, occlusal surfaces, such as the mesial-occlusal-distal composite restoration at tooth 45, in subject A, were imaged as exemplary presented in Fig. 10. The intersection between the restoration and the surrounding enamel shows a smooth transition without major cracks or gaps. The appearing bright section inside the restoration [cf. Fig. 10(b)] could be caused by an interlayer, possibly resulting from air entrapments or from varying scattering properties of adhesive and composite.

Fig. 10

(a,b) Representative B-scans of a mesial–distal–occlusal composite restoration at (c,d) tooth 45 in subject A. The in vivo OCT measurement was realized in contact mode by an additional distal distance piece (disposable item). Scale bar in (a) and (b) corresponds to 300  μm. Dynamic range of the depicted OCT cross-sectional scans amounts to 50 dB.

JBO_23_7_071207_f010.png

3.2.

Imaging of the Oral Mucosa

For characterizing the performance of the handheld EOCT probe at the oral soft tissue, the mucosa at different parts of the oral cavity is imaged by means of a healthy volunteer (subject B in Sec. 3.1: female, 33 years, nonsmoker, occasional use of alcohol). Of particular interest for OCT imaging is the lining mucosa, which enables a larger depth of light penetration due to the absence of a corneous layer in comparison to the masticatory mucosa, such as the gingiva [cf. Fig. 6(e)], which is covered by keratinized stratified squamous epithelium. Figure 11 displays three exemplary B-scans of the buccal mucosa of subject B. With regard to the structural validation of oral mucosal tissue in OCT images by Hamdoon et al.,38 the epithelial layer (EP) has a lower backscattering signal and appears more homogenous in comparison to the subsequent LP, which is confirmed by the imaging results in Fig. 11. The lower signal intensity of normal epithelium is related to the lower optical density and scattering properties.9 The LP corresponds to a dense fibrous layer with embedded blood vessels and nerves, which can be clearly differentiated from EP due to the higher backscattering signal and inhomogeneity in the OCT cross-sections. The EP and LP are divided by a basement membrane, which appears as a demarcation line between the two different backscattering signals of EP and underlying LP. OCT scans were acquired at the anterior [cf. Figs. 11(a) and 11(b)] and the posterior [cf. Fig. 11(c)] buccal mucosa. The epithelial thickness of the buccal mucosa is measured to be maximum 500  μm in the anterior part and of about 300  μm in the posterior region of the inside of the cheek.

Fig. 11

Single exemplary B-scans of the buccal oral mucosa in vivo, which are detected contactless at selected regions of the inner surface of the human cheek. (a, b) Buccal oral mucosa toward the human mouth in the anterior region of the oral cavity. (c) Buccal oral mucosa toward the oropharynx in the posterior region of the oral cavity. The OCT measurement was performed in noncontact mode. Abbreviations: EP, epithelium; LP, lamina propria. Scale bar: 300  μm. Indicated dynamic range is valid for all depicted OCT images.

JBO_23_7_071207_f011.png

Additionally, the oral mucosa of the tongue and lateral floor of the mouth is imaged contactless by EOCT. Figure 12 demonstrates OCT cross-sectional images of the tongue body especially the inferior surface (a), the lateral border (b), and the curved upper surface (dorsum) (c). As a result, the EP thickness of the tongue strongly depends on its location and lowest for the inferior surface of the tongue with its smooth mucous membrane and thin stratified squamous epithelium [cf. Fig. 12(a)]. Additionally, the sublingual vessel network and the dense fibrous layer within the highly scattering LP are clearly visible in (a). The oral mucosa of the lateral border of the tongue in (b) shows a thicker EP, which is clearly differentiated from the subjacent LP with imaged minor and major blood vessels. In Fig. 12(c), the specialized oral mucosa of the surface of the tongue shows prominent epithelium ridges and tongue papillae. Due to the thick keratinized stratified squamous epithelium of the dorsal surface, the imaging depth is limited for which reason the LP is not visible in the OCT image [Fig. 12(c)]. An additional example for the lining mucosa is shown in Fig. 12(d), where the EP and the LP of the floor of the mouth are demonstrated.

Fig. 12

Single exemplary B-scans showing (a) the oral mucosa in vivo of the inferior surface of the human tongue (facies inferior linguae), (b) the margin of the tongue (margo linguae), (c) the dorsum of the tongue (dorsum linguae), and (d) the lateral floor of the mouth. All measurements are carried out in noncontact mode. Abbreviations: EP, epithelium; LP, lamina propria; B, blood vessel. Scale bar: 300  μm. Indicated dynamic range is valid for all depicted OCT images.

JBO_23_7_071207_f012.png

Besides the representative OCT measurements of the buccal and lingual mucosa, the oral mucosa of the soft palate was imaged exemplarily with the developed OCT endoscope. For test purposes, the soft palate and the palatoglossal arch are examined in contact mode. In Fig. 13, the palatal mucosa is imaged by EOCT for three different regions of the soft palate. Figure 13(a) presents the oral mucosa of the anterior soft palate (toward the posterior border of hard palate), which obviously contains an LP with highly scattering dense connective tissue. This may be interpreted as part of the thin firm fibrous lamella named palatine aponeurosis. Below the dense fibrous layer, submucosal structures with integrated blood vessels are visible. In comparison, the oral mucosa detected at the palatoglossal arch is presented in Fig. 13(b) and shows a comparable EP, a thinner highly scattering fibrous layer as well as submucosal structures beneath. The oral mucosa of the transition part from the palatoglossal fold to the posterior buccal mucosa shows a varying EP thickness with a clearly determinable LP [cf. Figs. 13(c), 13(c1), and 13(c2)].

Fig. 13

(a–c) Cross-sectional images of oral mucosa at three selected regions of the soft palate, which are detected in contact mode. Oral mucosa of the anterior soft palate close to the posterior border of the hard palate (a) and corresponding overview image of the off-axis camera (a1). Oral mucosa of the central soft palate (b) and corresponding video guidance image (b1). Oral mucosa of the palatoglossal arch close to the posterior buccal mucosa in 2-D (c) and 3-D (c1,c2). Abbreviations: EP, epithelium; LP, lamina propria; B, blood vessel. Scale bar: 300  μm. Indicated dynamic range is valid for all depicted OCT images.

JBO_23_7_071207_f013.png

In addition, the oral mucosa of the palatoglossal arch was imaged via contactless EOCT. The results are presented in Fig. 14, where the EP and subsequent LP are obviously differentiable.

Fig. 14

Contactless OCT detection of oral mucosa of the palatoglossal arch of the soft palate with two capillary branches. (a) 2-D OCT with both capillary branches (CN1, CN2) embedded in the papillary layer of the LP. (b,c) OCT image series of the two capillary branches showing cross-section in front, through, and after the small capillary network. Abbreviations: EP, epithelium; LP, lamina propria; CN, capillary network. Scale bar: 300  μm. Indicated dynamic range is valid for all depicted OCT images.

JBO_23_7_071207_f014.png

Moreover, two capillary branches (CN1 and CN2) lying relatively close to each other are detected within the papillary layer of the LP. Both capillary branches running to the papilla are detected separately in 3-D. Representative B-scans of the OCT volume scans are presented as image series in Figs. 14(b) and 14(c), showing the respective cross-section in front, through, and after the capillary networks CN1 and CN2 within the superficial layer of the LP.

4.

Discussion and Summary

In this research, customized EOCT imaging of the anterior and posterior oral cavity was allowed by adapting a commercially available NIR endoscope by Karl Storz GmbH & Co. KG. By developing a scanning optics for 2-D imaging and using the established SD-OCT system of our workgroup, EOCT cross-sections of hard and soft oral tissue are detected with a frame rate of 23 fps, a spatial resolution of 11.6  μm axially and 17.5  μm laterally, and a FoV of 4.8 mm. This allows an image quality highly appropriate for the oral optical biopsy with μm-resolution. All presented imaging results are primary generated in noncontact mode (no contact to the imaged oral tissue but possible contact of the endoscope shaft to, e.g., lips), which is preferential for the investigation of highly sensitive and painful mucous oral regions. The movements of the test person are reduced by a variable forehead and chin support similar to the eye refractometry. Despite the handheld measurement, motion artifacts have rarely been observed. Moreover, investigations are also possible in contact mode by an additional spacer (disposable item) positioned at the tissue to be investigated, which is advantageous for recording dental volumetric data with minimal motion artifacts. In general, the technical development of a handheld endoscopic probe forms the basis for in vivo OCT studies of physicians and dentists in the dental clinic. By this adaption, the potential support of OCT to current diagnostic tools of visual inspection and radiology, which unfortunately show insufficient sensitivity and specificity in clinical routine, can be validated.

Promising application of EOCT in the oral cavity is generally the early detection of initial and small pathologic alterations of hard and soft tissue. Considering dentistry, OCT could particularly improve the detection of early incipient demineralization and primary/secondary caries, the diagnosis of periodontal disease, the monitoring and assessment of dental restorations as well as the visualization of different dental and orthodontic treatment steps in general, which is confirmed by the presented preliminary in vivo endoscopic results. For caries diagnostics, the primary aim is the detection of emerging demineralization. Since these lesions are characterized by porous areas in comparison to the surrounding enamel, OCT has the potential to image these lesions as white spots positioned close to the tooth subsurface up to 2 mm in depth within the enamel. By applying the developed EOCT probe, the detection of noncavitated occlusal carious lesions of premolars and molars in vivo is enabled to a limited extent with regard to the interdental and labial surfaces. Considering dental restorations, the assessment of the composite homogeneity and bonding, especially at the tooth-restoration interface and margin, as well as the early detection of secondary carious lesions, is demanding. Here again, the proposed OCT endoscope enables the subsurface imaging with the exception of interdental treatments. Besides these noninvasive applications, the feasibility of supporting minimally invasive restorative treatment with a customized EOCT scanner will be evaluated in a clinical study in future research.

Regarding the oral examinations, current research is strongly focused on depth-resolved imaging of early malign tissue. OCT in general and as endoscopic adaption in particular enables the imaging of the epithelial thickness end ridges of the oral mucosa, as well as irregular epithelial stratification. Moreover, the measurement of the continuity of the basement membrane is allowed. To the best of our knowledge, OCT imaging of different regions of the human soft palate (e.g., the palatoglossal arch and fold) with highly resolved structures of the LP in vivo is realized for the first time by means of the presented endoscopic development. By this, the findings of previous ex vivo and in vivo studies can be expanded to the posterior oral cavity under in vivo conditions. By knowing the regular structure of the oral mucosa of different parts of the oral cavity accessible by the developed OCT endoscope, the differentiation of normal and morphologically altered oral tissue should be ensured prospectively.35 Since the image quality of the presented EOCT probe offers appropriate axial and lateral resolution, even small structural alteration could be apparent, which has to be validated in a future clinical study. Despite the conspicuousness, premalignant disorders will be challenging to differentiate from benign lesions and to stage by OCT. In our view, the developed EOCT probe will have its most potential primarily in the monitoring of dysplastic lesions and further in biopsy guidance during surgical removal of lesions.

In summary, the developed EOCT probe, based on the commercially available NIR endoscope, ranks among the few in vivo scanners designed for the oral cavity (cf. Table 1 in Appendix) with the advantage of an expanded application field: the posterior oral cavity including different regions of the palatoglossal arch and (pre-)molars. As a result of this presented preliminary research, the contactless imaging is preferable for the examination of the soft tissue due to the high sensitivity, whereas the contact and noncontact mode is equally feasible for the hard tissue diagnosis. Besides the good image quality with high lateral and axial resolution, and sufficient wide FoV, the development convinces primarily with the accessibility to most of the oral sections and further with flexibility, efficiency, and effectiveness in the prospective clinical routine. In this research, the feasibility of in vivo studies within the anterior and posterior oral cavity was validly presented by hard and soft tissue imaging. The evaluation of the promising medical applications in clinical routine will be part of subsequent studies. For the case of successful studies with high reliability, accuracy, and evidence, OCT could have the potential to promote the clinical decision process in the future research.

Table 1

Technical specifications of published rigid oral probes (Appendix).

Ref.Oral regionsSpatial resolutionFoVTechniqueContact/Noncontact modeScan rate
Walther (this work)Dental hard and soft tissue of the anterior and posterior oral cavity; in detail: buccal and lingual oral mucosa, mouth floor, palatoglossal arch and fold (soft palate), incisors, premolars and molarsLateral 17.5  μm Axial 11  μm4.8 mmRigid (commercial endoscope with Hopkins optics), handheldBoth11.88 kHz 23 fps
Lenton et al.20Tooth occlusalLateral 80  μm Axial 11  μmNo specsRigid, handheldContactless30 kHz
Davoudi et al.59Buccal, labial, and lingual oral mucosa, soft palate (no layered structure and no connective tissue visible)Lateral 20.5  μm Axial 7  μm2 mmRigid (no Hopkins optics), mounted on a stageContactless47 kHz
Tsai et al.60Buccal, labial, and lingual oral mucosa (possibly not long enough for the posterior oral cavity)Lateral 10  μm Axial 8  μm2 mmRigid (no Hopkins optics), handheldContact100 kHz
Higgins and Pierce62No specificationsLateral 12.4  μm Axial 9.3  μm2 mmRigid (no Hopkins optics), handheldContact25.3 fps
Wang et al.63Buccal, labial, and lingual oral mucosa, mouth floor, hard palate, anterior soft palate (no layered structure and no connective tissue visible)Lateral and axial 12  μm4.6 mmRigid (GRIN lens), side viewing probe, sharp edges of the endoscopic tip, handheldContact13 kHz 25 fps
Wang et al.64Buccal mucosa, leukoplakia tissueLateral 40  μm Axial 10.6  μm2 mmRigid (GRIN lens), handheldNo specs50 kHz
Sun et al.65Tooth buccalAxial 10  μm2 to 4 mmRigid (fiber-based), handheldNo specs20 kHz
Lee et al.66Buccal and lingual oral mucosaLateral 20 to 40  μm2.3 mmRigid (fiber-based), handheldContact50.4 kHz

Appendices

Appendix

For the objective evaluation of the presented development, important technical specifications of previously published endoscopic probes designed for in vivo imaging of the human oral cavity are listed in Table 1. Moreover, a technical drawing of the handheld endoscopic scanner head is given in Fig. 15.

Fig. 15

Technical drawing of the scanner head for the presentation of relevant geometric features (Appendix).

JBO_23_7_071207_f015.png

Disclosures

The PS-OCT imaging of the human oral cavity by the homebuilt system was reviewed and approved by the developers employed at the Technische Universität Dresden, and informed consent was obtained from all subjects before imaging. Both human subjects included in this pilot study are developers of the applied OCT system and main author of the paper who decided to be volunteers on their own responsibility. Accordingly, the responsible local ethics committee waived the need for an approval. The study followed the tenets of the Declaration of Helsinki and was conducted in compliance with Health Insurance Portability and Accountability Act. Furthermore, the authors declare that there are no conflicts of interest related to this article.

Acknowledgments

For the presented research, we would like to thank Karl Storz GmbH & Co. KG, Tuttlingen, Germany, for provision of the rigid NIR endoscope. The author Florian Tetschke is supported by the European Union/European Social Fund (ESF) and the Free State of Saxony within the ESF junior research group “Optical Technologies in Medicine” (project number 100270108). The author Jonas Golde is supported by the European Union/European Social Fund (ESF) and the Free State of Saxony within a doctoral scholarship (project number 100284305).

References

1. 

D. Huang et al., “Optical coherence tomography,” Science, 254 1178 –1181 (1991). http://dx.doi.org/10.1126/science.1957169 SCIEAS 0036-8075 Google Scholar

2. 

A. F. Fercher et al., “Optical coherence tomography-principles and applications,” Rep. Prog. Phys., 66 (2), 239 –303 (2003). http://dx.doi.org/10.1088/0034-4885/66/2/204 RPPHAG 0034-4885 Google Scholar

3. 

J. Walther et al., “Optical coherence tomography in biomedical research,” Anal. Bioanal. Chem., 400 (9), 2721 –2743 (2011). http://dx.doi.org/10.1007/s00216-011-5052-x ABCNBP 1618-2642 Google Scholar

4. 

S. Marschall et al., “Optical coherence tomography-current technology and applications in clinical and biomedical research,” Anal. Bioanal. Chem., 400 (9), 2699 –2720 (2011). http://dx.doi.org/10.1007/s00216-011-5008-1 ABCNBP 1618-2642 Google Scholar

5. 

P. Wilder-Smith et al., “In vivo optical coherence tomography for the diagnosis of oral malignancy,” Lasers Surg. Med., 35 (4), 269 –275 (2004). http://dx.doi.org/10.1002/(ISSN)1096-9101 LSMEDI 0196-8092 Google Scholar

6. 

W. Jung et al., “Advances in oral cancer detection using optical coherence tomography,” IEEE J. Sel. Top. Quantum Electron., 11 (4), 811 –817 (2005). http://dx.doi.org/10.1109/JSTQE.2005.857678 Google Scholar

7. 

P. Pande et al., “Automated classification of optical coherence tomography images for the diagnosis of oral malignancy in the hamster cheek pouch,” J. Biomed. Opt., 19 (8), 086022 (2014). http://dx.doi.org/10.1117/1.JBO.19.8.086022 JBOPFO 1083-3668 Google Scholar

8. 

C. K. Lee et al., “Diagnosis of oral submucous fibrosis with optical coherence tomography,” J. Biomed. Opt., 14 (5), 054008 (2009). http://dx.doi.org/10.1117/1.3233653 JBOPFO 1083-3668 Google Scholar

9. 

S. Prestin et al., “Measurement of epithelial thickness within the oral cavity using optical coherence tomography,” Head Neck, 34 (12), 1777 –1781 (2012). http://dx.doi.org/10.1002/hed.v34.12 Google Scholar

10. 

I. Grulkowski et al., “Quantitative assessment of oral mucosa and labial minor salivary glands in patients with Sjögren’s syndrome using swept source OCT,” Biomed. Opt. Express, 5 (1), 259 –274 (2014). http://dx.doi.org/10.1364/BOE.5.000259 Google Scholar

11. 

R. A. Leitgeb et al., “Doppler optical coherence tomography,” Prog. Retinal Eye Res., 41 (100), 26 –43 (2014). http://dx.doi.org/10.1016/j.preteyeres.2014.03.004 PRTRES 1350-9462 Google Scholar

12. 

J. Walther et al., “Analysis of in vitro and in vivo bidirectional flow velocities by phase-resolved Doppler Fourier-domain OCT,” Sens. Actuators A: Phys., 156 (1), 14 –21 (2009). http://dx.doi.org/10.1016/j.sna.2009.02.025 Google Scholar

13. 

J. Walther and E. Koch, “Relation of joint spectral and time domain optical coherence tomography (jSTdOCT) and phase-resolved Doppler OCT,” Opt. Express, 22 (19), 23129 –23146 (2014). http://dx.doi.org/10.1364/OE.22.023129 OPEXFF 1094-4087 Google Scholar

14. 

J. F. de Boer, C. K. Hitzenberger and Y. Yasuno, “Polarization sensitive optical coherence tomography–a review,” Biomed. Opt. Express, 8 (3), 1838 –1873 (2017). http://dx.doi.org/10.1364/BOE.8.001838 BOEICL 2156-7085 Google Scholar

15. 

B. Baumann, “Polarization sensitive optical coherence tomography: a review of technology and applications,” Appl. Sci., 7 (5), 474 (2017). http://dx.doi.org/10.3390/app7050474 Google Scholar

16. 

M. Pircher, C. K. Hitzenberger and U. Schmidt-Erfurth, “Polarization sensitive optical coherence tomography in the human eye,” Prog. Retinal Eye Res., 30 (6), 431 –451 (2011). http://dx.doi.org/10.1016/j.preteyeres.2011.06.003 PRTRES 1350-9462 Google Scholar

17. 

U. Morgner et al., “Spectroscopic optical coherence tomography,” Opt. Lett., 25 (2), 111 –113 (2000). http://dx.doi.org/10.1364/OL.25.000111 OPLEDP 0146-9592 Google Scholar

18. 

D. J. Faber et al., “Toward assessments of blood oxygen saturation by spectroscopic optical coherence tomography,” Opt. Lett., 30 (9), 1015 –1017 (2005). http://dx.doi.org/10.1364/OL.30.001015 OPLEDP 0146-9592 Google Scholar

19. 

Y. Shimada et al., “Noninvasive cross-sectional imaging of proximal caries using swept-source optical coherence tomography (SS-OCT) in vivo,” J. Biophotonics, 7 (7), 506 –513 (2014). http://dx.doi.org/10.1002/jbio.201200210 Google Scholar

20. 

P. Lenton et al., “Imaging in vivo secondary caries and ex vivo dental biofilms using cross-polarization optical coherence tomography,” Dent. Mater., 28 (7), 792 –800 (2012). http://dx.doi.org/10.1016/j.dental.2012.04.004 Google Scholar

21. 

T. Louie et al., “Clinical assessment of early tooth demineralization using polarization sensitive optical coherence tomography,” Lasers Surg. Med., 42 (10), 898 –905 (2010). http://dx.doi.org/10.1002/lsm.21013 LSMEDI 0196-8092 Google Scholar

22. 

L. L. Otis et al., “Optical coherence tomography: a new imaging technology for dentistry,” J. Am. Dent. Assoc., 131 (4), 511 –514 (2000). http://dx.doi.org/10.14219/jada.archive.2000.0210 Google Scholar

23. 

C. Wilder-Smith et al., “Quantification of dental erosions in patients with GERD using optical coherence tomography before and after double-blind, randomized treatment with esomeprazole or placebo,” Am. J. Gastroenterol., 104 (11), 2788 –2795 (2009). http://dx.doi.org/10.1038/ajg.2009.441 Google Scholar

24. 

H. Nakagawa et al., “Validation of swept source optical coherence tomography (SS-OCT) for the diagnosis of smooth surface caries in vitro,” J. Dent., 41 (1), 80 –89 (2013). http://dx.doi.org/10.1016/j.jdent.2012.10.007 Google Scholar

25. 

J. A. Warren et al., “Imaging and characterization of dental structure using optical coherence tomography,” in Technical Digest. Summaries of papers presented at the Conf. on Lasers and Electro-Optics (CLEO ’98), 128 (1998). http://dx.doi.org/10.1109/CLEO.1998.675955 Google Scholar

26. 

S. K. Manesh, C. L. Darling and D. Fried, “Imaging natural and artificial demineralization on dentin surfaces with polarization sensitive optical coherence tomography,” Proc. SPIE, 6843 68430M (2008). http://dx.doi.org/10.1117/12.778788 PSISDG 0277-786X Google Scholar

27. 

S. K. Manesh, C. L. Darling and D. Fried, “Polarization-sensitive optical coherence tomography for the nondestructive assessment of the remineralization of dentin,” J. Biomed. Opt., 14 (4), 044002 (2009). http://dx.doi.org/10.1117/1.3158995 Google Scholar

28. 

P. Wilder-Smith et al., “Optical diagnostics in the oral cavity: an overview,” Oral Dis., 16 (8), 717 –728 (2010). http://dx.doi.org/10.1111/odi.2010.16.issue-8 Google Scholar

29. 

Y. S. Hsieh et al., “Dental optical coherence tomography,” Sensors, 13 (7), 8928 –8949 (2013). http://dx.doi.org/10.3390/s130708928 SNSRES 0746-9462 Google Scholar

30. 

S. Canjau et al., “Optical coherence tomography for non-invasive ex vivo investigations in dental medicine–a joint group experience,” Mod. Technol. Med., 7 (1), 97 –114 (2015). http://dx.doi.org/10.17691/stm2015.7.1.13 Google Scholar

31. 

Y. Shimada et al., “Application of optical coherence tomography (OCT) for diagnosis of caries, cracks, and defects of restorations,” Curr. Oral Health Rep., 2 (2), 73 –80 (2015). http://dx.doi.org/10.1007/s40496-015-0045-z Google Scholar

32. 

D. Fried et al., “Imaging caries lesions and lesion progression with polarization sensitive optical coherence tomography,” J. Biomed. Opt., 7 (4), 618 –627 (2002). http://dx.doi.org/10.1117/1.1509752 Google Scholar

33. 

M. H. Van der Veen, “Detecting short-term changes in the activity of caries lesions with the aid of new technologies,” Curr. Oral Health Rep., 2 (2), 102 –109 (2015). http://dx.doi.org/10.1007/s40496-015-0050-2 Google Scholar

34. 

D. Ericson et al., “Minimally invasive dentistry–concept and techniques in cariology,” Oral Health Prev. Dent., 1 (1), 59 –72 (2003). Google Scholar

35. 

M.-T. Tsai et al., “Effective indicators for diagnosis of oral cancer using optical coherence tomography,” Opt. Express, 16 (20), 15847 –15862 (2008). http://dx.doi.org/10.1364/OE.16.015847 OPEXFF 1094-4087 Google Scholar

36. 

M.-T. Tsai et al., “Differentiating oral lesions in different carcinogenesis stages with optical coherence tomography,” J. Biomed. Opt., 14 (4), 044028 (2009). http://dx.doi.org/10.1117/1.3200936 JBOPFO 1083-3668 Google Scholar

37. 

C.-K. Lee et al., “Diagnosis of oral precancer with optical coherence tomography,” Biomed. Opt. Express, 3 (7), 1632 –1646 (2012). http://dx.doi.org/10.1364/BOE.3.001632 BOEICL 2156-7085 Google Scholar

38. 

Z. Hamdoon et al., “Structural validation of oral mucosal tissue using optical coherence tomography,” Head Neck Oncol., 4 (1), 29 (2012). http://dx.doi.org/10.1186/1758-3284-4-29 Google Scholar

39. 

Z. Hamdoon et al., “Optical coherence tomography in the assessment of suspicious oral lesions: an immediate ex vivo study,” Photodiagn. Photodyn. Ther., 10 (1), 17 –27 (2013). http://dx.doi.org/10.1016/j.pdpdt.2012.07.005 Google Scholar

40. 

Z. Hamdoon et al., “Optical coherence tomography in the assessment of oral squamous cell carcinoma resection margins,” Photodiagn. Photodyn. Ther., 13 211 –217 (2016). http://dx.doi.org/10.1016/j.pdpdt.2015.07.170 Google Scholar

41. 

B. A. Miles, “How close are we to real time optical margin control in head and neck oncologic surgery?,” Cancer Cell Microenviron., 3 (2), 1305 (2016). http://dx.doi.org/10.14800/ccm.1305 Google Scholar

42. 

G. Tirelli et al., “Will the mininvasive approach challenge the old paradigms in oral cancer surgery?,” Eur. Arch. Oto-Rhino-Laryngol., 274 (3), 1279 –1289 (2017). http://dx.doi.org/10.1007/s00405-016-4221-0 Google Scholar

43. 

M. J. Gora et al., “Endoscopic optical coherence tomography: technologies and clinical applications,” Biomed. Opt. Express, 8 (5), 2405 –2444 (2017). http://dx.doi.org/10.1364/BOE.8.002405 BOEICL 2156-7085 Google Scholar

44. 

G. G. J. Tearney et al., “Scanning single-mode fiber optic catheter-endoscope for optical coherence tomography,” Opt. Lett., 21 (7), 543 –545 (1996). http://dx.doi.org/10.1364/OL.21.000543 OPLEDP 0146-9592 Google Scholar

45. 

G. J. Tearney et al., “In vivo endoscopic optical biopsy with optical coherence tomography,” Science, 276 (5321), 2037 –2039 (1997). http://dx.doi.org/10.1126/science.276.5321.2037 SCIEAS 0036-8075 Google Scholar

46. 

G. J. Tearney et al., “Consensus standards for acquisition, measurement, and reporting of intravascular optical coherence tomography studies: a report from the International Working Group for intravascular optical coherence tomography standardization and validation,” J. Am. Coll. Cardiol., 59 (12), 1058 –1072 (2012). http://dx.doi.org/10.1016/j.jacc.2011.09.079 Google Scholar

47. 

N. Hanna et al., “Two-dimensional and 3-dimensional optical coherence tomographic imaging of the airway, lung, and pleura,” J. Thorac. Cardiovasc. Surg., 129 (3), 615 –622 (2005). http://dx.doi.org/10.1016/j.jtcvs.2004.10.022 Google Scholar

48. 

T. H. Tsai et al., “Ultrahigh speed endoscopic optical coherence tomography for gastroenterology,” Biomed. Opt. Express, 5 (12), 4387 –4404 (2014). http://dx.doi.org/10.1364/BOE.5.004387 Google Scholar

49. 

T. S. Kirtane and M. S. Wagh, “Endoscopic optical coherence tomography (OCT): advances in gastrointestinal imaging,” Gastroenterol. Res. Pract., 2014 376367 (2014). http://dx.doi.org/10.1155/2014/376367 Google Scholar

50. 

A. Burkhardt et al., “Endoscopic optical coherence tomography device for forward imaging with broad field of view,” J. Biomed. Opt., 17 (7), 071302 (2012). http://dx.doi.org/10.1117/1.JBO.17.7.071302 JBOPFO 1083-3668 Google Scholar

51. 

Z. Hubler et al., “Real-time automated thickness measurement of the in vivo human tympanic membrane using optical coherence tomography,” Quant. Imaging Med. Surg., 5 (1), 69 –77 (2015). http://dx.doi.org/10.3978/j.issn.2223-4292.2014.11.32 Google Scholar

52. 

N. H. Cho et al., “Optical coherence tomography for the diagnosis and evaluation of human otitis media,” J. Korean Med. Sci., 30 (3), 328 –335 (2015). http://dx.doi.org/10.3346/jkms.2015.30.3.328 JKMSEH Google Scholar

53. 

L. Scolaro et al., “High-sensitivity anastigmatic imaging needle for optical coherence tomography,” Opt. Lett., 37 (24), 5247 –5249 (2012). http://dx.doi.org/10.1364/OL.37.005247 Google Scholar

54. 

K. M. Kennedy et al., “Needle optical coherence elastography for the measurement of microscale mechanical contrast deep within human breast tissues,” J. Biomed. Opt., 18 (12), 121510 (2013). http://dx.doi.org/10.1117/1.JBO.18.12.121510 JBOPFO 1083-3668 Google Scholar

55. 

B. G. Muller et al., “Prostate cancer diagnosis: the feasibility of needle-based optical coherence tomography,” J. Med. Imaging, 2 (3), 037501 (2015). http://dx.doi.org/10.1117/1.JMI.2.3.037501 JMEIET 0920-5497 Google Scholar

56. 

M. Villiger et al., “Deep tissue volume imaging of birefringence through fibre-optic needle probes for the delineation of breast tumour,” Sci. Rep., 6 28771 (2016). http://dx.doi.org/10.1038/srep28771 SRCEC3 2045-2322 Google Scholar

57. 

F. I. Feldchtein et al., “In vivo OCT imaging of hard and soft tissue of the oral cavity,” Opt. Express, 3 (3), 239 –251 (1998). http://dx.doi.org/10.1364/OE.3.000239 OPEXFF 1094-4087 Google Scholar

58. 

R. Brandenburg, B. Haller and C. Hauger, “Real-time in vivo imaging of dental tissue by means of optical coherence tomography (OCT),” Opt. Commun., 227 (4–6), 203 –211 (2003). http://dx.doi.org/10.1016/j.optcom.2003.09.043 OPCOB8 0030-4018 Google Scholar

59. 

B. Davoudi et al., “Noninvasive in vivo structural and vascular imaging of human oral tissues with spectral domain optical coherence tomography,” Biomed. Opt. Express, 3 (5), 826 –839 (2012). http://dx.doi.org/10.1364/BOE.3.000826 BOEICL 2156-7085 Google Scholar

60. 

M.-T. Tsai et al., “Noninvasive structural and microvascular anatomy of oral mucosae using handheld optical coherence tomography,” Biomed. Opt. Express, 8 (11), 5001 –5012 (2017). http://dx.doi.org/10.1364/BOE.8.005001 BOEICL 2156-7085 Google Scholar

61. 

J. M. Ridgeway et al., “In vivo optical coherence tomography of the human oral cavity and oropharynx,” Arch. Otolaryngol. Head Neck Surg., 132 (10), 1074 –1081 (2006). http://dx.doi.org/10.1001/archotol.132.10.1074 Google Scholar

62. 

L. M. Higgins and M. C. Pierce, “Design and characterization of a handheld multimodal imaging device for the assessment of oral epithelial lesions,” J. Biomed. Opt., 19 (8), 086004 (2014). http://dx.doi.org/10.1117/1.JBO.19.8.086004 JBOPFO 1083-3668 Google Scholar

63. 

J. Wang et al., “Development of a hybrid Raman spectroscopy and optical coherence tomography technique for real-time in vivo tissue measurements,” Opt. Lett., 41 (13), 3045 –3048 (2016). http://dx.doi.org/10.1364/OL.41.003045 OPLEDP 0146-9592 Google Scholar

64. 

D. Wang et al., “Endoscopic swept-source optical coherence tomography based on a two-axis microelectromechanical system mirror,” J. Biomed. Opt., 18 (8), 086005 (2013). http://dx.doi.org/10.1117/1.JBO.18.8.086005 JBOPFO 1083-3668 Google Scholar

65. 

C.-W. Sun, Y. Ho and S. Lee, “Sensing of tooth microleakage based on dental optical coherence tomography,” J. Sens., 2014 984627 (2014). http://dx.doi.org/10.1155/2014/824904 Google Scholar

66. 

A. M. D. Lee et al., “Wide-field in vivo oral OCT imaging,” Biomed. Opt. Express, 6 (7), 2664 –2674 (2015). http://dx.doi.org/10.1364/BOE.6.002664 BOEICL 2156-7085 Google Scholar

67. 

H. Shemesh et al., “The ability of optical coherence tomography to characterize the root canal walls,” J. Endod., 33 (11), 1369 –1373 (2007). http://dx.doi.org/10.1016/j.joen.2007.06.022 Google Scholar

68. 

M. Wojtkowski et al., “Ultrahigh-resolution, high-speed, Fourier domain optical coherence tomography and methods for dispersion compensation,” Opt. Express, 12 (11), 2404 –2422 (2004). http://dx.doi.org/10.1364/OPEX.12.002404 OPEXFF 1094-4087 Google Scholar

69. 

F. Köttig et al., “An advanced algorithm for dispersion encoded full range frequency domain optical coherence tomography,” Opt. Express, 20 (22), 24925 –24948 (2012). http://dx.doi.org/10.1364/OE.20.024925 OPEXFF 1094-4087 Google Scholar

70. 

J. Walther et al., “Signal power decrease due to fringe washout as an extension of the limited Doppler flow measurement range in spectral domain optical coherence tomography,” J. Biomed. Opt., 15 (4), 041511 (2010). http://dx.doi.org/10.1117/1.3466578 JBOPFO 1083-3668 Google Scholar

71. 

B. Cense et al., “Ultrahigh-resolution high-speed retinal imaging using spectral-domain optical coherence tomography,” Opt. Express, 12 (11), 2435 –2447 (2004). http://dx.doi.org/10.1364/OPEX.12.002435 OPEXFF 1094-4087 Google Scholar

72. 

L. Pan et al., “Depth-dependent dispersion compensation for full-depth OCT image,” Opt. Express, 25 (9), 10345 –10354 (2017). http://dx.doi.org/10.1364/OE.25.010345 OPEXFF 1094-4087 Google Scholar

73. 

C. Schnabel, M. Gaertner and E. Koch, “Optical coherence tomography (OCT) for time-resolved imaging of alveolar dynamics in mechanically ventilated rats,” Appl. Sci., 7 (3), 287 (2017). http://dx.doi.org/10.3390/app7030287 Google Scholar

74. 

K. J. Park, H. Schneider and R. Haak, “Assessment of interfacial defects at composite restorations by swept source optical coherence tomography,” J. Biomed. Opt., 18 (7), 076018 (2013). http://dx.doi.org/10.1117/1.JBO.18.7.076018 JBOPFO 1083-3668 Google Scholar

75. 

K. Ishibashi et al., “Swept-source optical coherence tomography as a new tool to evaluate defects of resin-based composite restorations,” J. Dent., 39 (8), 543 –548 (2011). http://dx.doi.org/10.1016/j.jdent.2011.05.005 JDENAB 0300-5712 Google Scholar

76. 

Y. Chen et al., “Characterization of dentin, enamel, and carious lesions by a polarization-sensitive optical coherence tomography system,” Appl. Opt., 44 (11), 2041 –2048 (2005). http://dx.doi.org/10.1364/AO.44.002041 APOPAI 0003-6935 Google Scholar

77. 

Y. Chen, L. Otis and Q. Zhu, “Polarization memory effect in optical coherence tomography and dental imaging application,” J. Biomed. Opt., 16 (8), 086005 (2011). http://dx.doi.org/10.1117/1.3606573 JBOPFO 1083-3668 Google Scholar

78. 

P. Cimalla et al., “Simultaneous dual-band optical coherence tomography in the spectral domain for high resolution in vivo imaging,” Opt. Express, 17 (22), 19486 –19500 (2009). http://dx.doi.org/10.1364/OE.17.019486 OPEXFF 1094-4087 Google Scholar

79. 

J.-M. Kim, S.-R. Kang and W.-J. Yi, “Automatic detection of tooth cracks in optical coherence tomography images,” J. Periodontal Implant Sci., 47 (1), 41 –50 (2017). http://dx.doi.org/10.5051/jpis.2017.47.1.41 Google Scholar

80. 

K. Imai et al., “Noninvasive cross-sectional visualization of enamel cracks by optical coherence tomography in vitro,” J. Endod., 38 (9), 1269 –1274 (2012). http://dx.doi.org/10.1016/j.joen.2012.05.008 Google Scholar

81. 

H. Kakuma, K. Ohbayashi and Y. Arakawa, “Optical imaging of hard and soft dental tissues using discretely swept optical frequency domain reflectometry optical coherence tomography at wavelengths from 1560 to 1600 nm,” J. Biomed. Opt., 13 (1), 014012 (2008). http://dx.doi.org/10.1117/1.2839042 JBOPFO 1083-3668 Google Scholar

82. 

K. J. Park, H. Schneider and R. Haak, “Assessment of defects at tooth/self-adhering flowable composite interface using swept-source optical coherence tomography (SS-OCT),” Dent. Mater., 31 (5), 534 –541 (2015). http://dx.doi.org/10.1016/j.dental.2015.02.005 Google Scholar

83. 

K. Horie et al., “Monitoring of cariogenic demineralization at the enamel-composite interface using swept-source optical coherence tomography,” Dent. Mater., 32 (9), 1103 –1112 (2016). http://dx.doi.org/10.1016/j.dental.2016.06.017 Google Scholar

Biography

Julia Walther is a senior researcher at Technische Universität Dresden. She received her diploma degree in physics technology and biomedical engineering from the University of Applied Sciences Mittweida in 2006 and her PhD degree in medical physics and biomedical optics from the Technische Universität (TU) Dresden in 2010. Her current research interests include the advancement of optical coherence tomography for different applications in dentistry and otolaryngology as well as the analysis and extension of diverse theoretical models for Doppler OCT (DOCT).

Christian Schnabel finished his studies of physical engineering at the University of Applied Sciences Mittweida in 2010 and subsequently joined the Department of Clinical Sensoring and Monitoring at the Faculty of Medicine Carl Gustav Carus in Dresden. He received his PhD degree in 2015 in the field of lung imaging, mechanical ventilation, and total liquid ventilation. His main field of research is currently the imaging of different biological tissues by optical coherence tomography.

Florian Tetschke received his MSc degree in biomedical engineering from the University of Applied Sciences, Jena, in 2013. He contributed to research projects at the Biomagnetic Center of the University Hospital in Jena, at the Polytechnical University of Catalonia and the Institute of Biomedical Engineering at the Technische Universität Dresden. Currently, he is a research associate at the young scientist group, “Optical Technologies in Medicine” at Technische Universität Dresden. His research interest focuses on the development and application of optical technologies, such as hyperspectral imaging, optical coherence tomography, and Raman spectroscopy.

Tobias Rosenauer is a dentist and a research associate of the University Clinic Carl Gustav Carus in Dresden. He graduated in 2016 and has since then worked for the Department of Operative Dentistry, TU Dresden. He is currently writing his doctoral thesis and his present research concerns cariological aspects as well as optical imaging.

Jonas Golde is a PhD student at the Technische Universität Dresden. He received his MSc degree in physics from the Technische Universität Dresden in 2016 with a thesis on polarization sensitive optical coherence tomography (PS-OCT). Since then, he has been focused on the development and application of PS-OCT systems in the fields of dentistry and otolaryngology. His research interests include the combination of PS-OCT with complementary optical technologies, such as multiphoton microscopy and hyperspectral imaging.

Christian Hannig is the head of the department of Operative Dentistry, TU Dresden. Furthermore, he is the dean for dental education. He is a dentist since 2000 and has worked at Göttingen and Freiburg University before coming to Dresden. His main field of research is bioadhesion in the oral cavity. He is the author of more than 100 international and national journal papers. His research has been funded several times by the German Research Foundation.

Edmund Koch received his diploma and PhD degrees in physics from the University of Marburg. He worked as a senior researcher at the Drägerwerke AG, Lübeck, and thereafter, he became professor for laser and optics at the Lübeck University of Applied Sciences. He is currently head of the work group Clinical Sensoring and Monitoring at the Faculty of Medicine at the TU Dresden. His main research interests are optical coherence tomography and spectroscopic imaging.

Biographies for the other authors are not available.

CC BY: © The Authors. Published by SPIE under a Creative Commons Attribution 4.0 Unported License. Distribution or reproduction of this work in whole or in part requires full attribution of the original publication, including its DOI.
Julia Walther, Christian Schnabel, Florian Tetschke, Tobias Rosenauer D.D.S., Jonas Golde, Nadja Ebert, Michael Baumann, Christian Hannig D.D.S., and Edmund Koch "In vivo imaging in the oral cavity by endoscopic optical coherence tomography," Journal of Biomedical Optics 23(7), 071207 (2 March 2018). https://doi.org/10.1117/1.JBO.23.7.071207
Received: 10 August 2017; Accepted: 12 January 2018; Published: 2 March 2018
Lens.org Logo
CITATIONS
Cited by 28 scholarly publications.
Advertisement
Advertisement
KEYWORDS
Optical coherence tomography

Endoscopy

In vivo imaging

Tissues

Endoscopes

Tissue optics

Teeth

Back to Top