header advert
You currently have no access to view or download this content. Please log in with your institutional or personal account if you should have access to through either of these
The Bone & Joint Journal Logo

Receive monthly Table of Contents alerts from The Bone & Joint Journal

Comprehensive article alerts can be set up and managed through your account settings

View my account settings

Get Access locked padlock

High complication rate in reconstruction of Paprosky type IIIa acetabular defects using an oblong implant with modular side plates and a hook



Download PDF

Abstract

We report the results of 62 hips in 62 patients (17 males, 45 females) with mean age of 62.4 years (37 to 81), who underwent revision of the acetabular component of a total hip replacement due to aseptic loosening between May 2003 and November 2007. All hips had a Paprosky type IIIa acetabular defect. Acetabular revision was undertaken using a Procotyl E cementless oblong implant with modular side plates and a hook combined with impaction allografting.

At a mean follow-up of 60.5 months (36 to 94) with no patients lost to follow-up and one died due to unrelated illness, the complication rate was 38.7%. Complications included aseptic loosening (19 hips), deep infection (3 hips), broken hook and side plate (one hip) and a femoral nerve palsy (one hip). Further revision of the acetabular component was required in 18 hips (29.0%) and a further four hips (6.4%) are currently loose and awaiting revision.

We observed unacceptably high rates of complication and failure in our group of patients and cannot recommend this implant or technique.


Correspondence should be sent to Dr G. C. Babis; e-mail:

For access options please click here