Skip to content
Licensed Unlicensed Requires Authentication Published by De Gruyter May 10, 2013

Recommendations for accurate numerical blood flow simulations of stented intracranial aneurysms

  • Gábor Janiga EMAIL logo , Philipp Berg , Oliver Beuing , Mathias Neugebauer , Rocco Gasteiger , Bernhard Preim , Georg Rose , Martin Skalej and Dominique Thévenin

Abstract

The number of scientific publications dealing with stented intracranial aneurysms is rapidly increasing. Powerful computational facilities are now available; an accurate computational modeling of hemodynamics in patient-specific configurations is, however, still being sought. Furthermore, there is still no general agreement on the quantities that should be computed and on the most adequate analysis for intervention support. In this article, the accurate representation of patient geometry is first discussed, involving successive improvements. Concerning the second step, the mesh required for the numerical simulation is especially challenging when deploying a stent with very fine wire structures. Third, the description of the fluid properties is a major challenge. Finally, a founded quantitative analysis of the simulation results is obviously needed to support interventional decisions. In the present work, an attempt has been made to review the most important steps for a high-quality computational fluid dynamics computation of virtually stented intracranial aneurysms. In consequence, this leads to concrete recommendations, whereby the obtained results are not discussed for their medical relevance but for the evaluation of their quality. This investigation might hopefully be helpful for further studies considering stent deployment in patient-specific geometries, in particular regarding the generation of the most appropriate computational model.


Corresponding author: Gábor Janiga, Department of Fluid Dynamics and Technical Flows, University of Magdeburg “Otto von Guericke”, 39106 Magdeburg, Germany, Phone: +49 391 67 18196, Fax: +49 391 67 12840

This work was financially supported in part by the International Max Planck Research School for Analysis, Design and Optimization in Chemical and Biomedical Process Engineering Magdeburg and the Land Saxony-Anhalt in Germany. The help of S. Seshadhri to produce the CAD geometry of the stent and to support the mesh generation is highly acknowledged. The permission to publish a case from the VISC challenge has been provided by the main organizer, Professor Makoto Ohta, and is warmly acknowledged.

Conflict of interest statement:

We declare that we have no conflict of interest.

References

[1] Antiga L, Steinman DA. Robust and objective decomposition and mapping of bifurcating vessels. IEEE Trans Med Imag 2004; 23: 704–713.10.1109/TMI.2004.826946Search in Google Scholar PubMed

[2] Appanaboyina S, Mut F, Lohner R, Putman C, Cebral J. Simulation of intracranial aneurysm stenting: techniques and challenges. Comput Methods Appl Mech Eng 2009; 198: 3567–3582.10.1016/j.cma.2009.01.017Search in Google Scholar

[3] Augsburger L, Reymond P, Rufenacht DA, Stergiopulos N. Intracranial stents being modeled as a porous medium: flow simulation in stented cerebral aneurysms. Ann Biomed Eng 2011; 39: 850–863.10.1007/s10439-010-0200-6Search in Google Scholar PubMed

[4] Baptista T, Fragata I, Ribeiro C, Reis J. Flow diverters devices for treatment of intra-cranial aneurysms – six months follow-up results. Acta Med Port 2012; 25 Suppl 1: 13–16.10.20344/amp.254Search in Google Scholar

[5] Berg P, Janiga G, Thévenin D. Instationäre Strömungssimulationen und Auswertung von Blutströmungen in zerebralen Aneurysmen mit Stent. In: Burgert O, Schipper J, Zachow S, editors. Computer- und Roboterassistierte Chirurgie (CURAC). Magdeburg, Germany: ISBN 978-3-940961-59-4 2011: 91–94.Search in Google Scholar

[6] Berg P, Janiga G, Thévenin D. Detailed comparison of numerical flow predictions in cerebral aneurysms using different CFD software. In: Vad J, editor. 15th Conference on Modelling Fluid Flow Budapest, Hungary: ISBN 978-963-08-4588-5 2012: 128–135.Search in Google Scholar

[7] Bernardini A, Larrabide I, Petrini L, et al. Deployment of self-expandable stents in aneurysmatic cerebral vessels: comparison of different computational approaches for interventional planning. Comput Methods Biomech Biomed Eng 2012; 15: 303–311.10.1080/10255842.2010.527838Search in Google Scholar PubMed

[8] Bordás R, Seshadhri S, Janiga G, Skalej M, Thévenin D. Experimental validation of numerical simulations on a cerebral aneurysm phantom model. Interv Med Appl Sci 2012; 4: 193–205.10.1556/imas.4.2012.4.4Search in Google Scholar PubMed PubMed Central

[9] Burleson AC, Turitto VT. Identification of quantifiable hemodynamic factors in the assessment of cerebral aneurysm behavior. On behalf of the Subcommittee on Biorheology of the Scientific and Standardization Committee of the ISTH. Thromb Haemostasis 1996; 76: 118–123.10.1055/s-0038-1650533Search in Google Scholar

[10] Casey M, Wintergerste T. Special Interest Group on “Quality and Trust in Industrial CFD” Best Practice Guidelines. ERCOFTAC 2000.Search in Google Scholar

[11] Cebral JR, Mut F, Raschi M, et al. Aneurysm rupture following treatment with flow-diverting stents: computational hemodynamics analysis of treatment. Am J Neuroradiol 2011; 32: 27–33.10.3174/ajnr.A2398Search in Google Scholar PubMed PubMed Central

[12] Cebral JR, Mut F, Weir J, Putman C. Quantitative characterization of the hemodynamic environment in ruptured and unruptured brain aneurysms. Am J Neuroradiol 2011; 32: 145–151.10.3174/ajnr.A2419Search in Google Scholar PubMed PubMed Central

[13] Chen J, Lu XY. Numerical investigation of the non-Newtonian pulsatile blood flow in a bifurcation model with a non-planar branch. J Biomech 2006; 39: 818–832.10.1016/j.jbiomech.2005.02.003Search in Google Scholar PubMed

[14] Dhar S, Tremmel M, Mocco J, et al. Morphology parameters for intracranial aneurysm rupture risk assessment. Neurosurgery 2008; 63: 185–196.10.1227/01.NEU.0000316847.64140.81Search in Google Scholar PubMed PubMed Central

[15] Ding YH, Lewis DA, Kadirvel R, Dai D, Kallmes DF. The Woven EndoBridge: a new aneurysm occlusion device. Am J Neuroradiol 2011; 32: 607–611.10.3174/ajnr.A2399Search in Google Scholar

[16] Gijsen FJ, Allanic E, van de Vosse FN, Janssen JD. The influence of the non-Newtonian properties of blood on the flow in large arteries: unsteady flow in a 90 degrees curved tube. J Biomech 1999; 32: 705–713.10.1016/S0021-9290(99)00014-7Search in Google Scholar

[17] Ha SW, Jang SJ. Clinical analysis of giant intracranial aneurysms with endovascular embolization. J Cerebrovasc Endovasc Neurosurg 2012; 14: 22–28.10.7461/jcen.2012.14.1.22Search in Google Scholar PubMed PubMed Central

[18] Hoi Y, Meng H, Woodward SH, et al. Effects of arterial geometry on aneurysm growth: three-dimensional computational fluid dynamics study. J Neurosurg 2004; 101: 676–681.10.3171/jns.2004.101.4.0676Search in Google Scholar PubMed

[19] Hoi Y, Woodward SH, Kim M, Taulbee DB, Meng H. Validation of CFD simulations of cerebral aneurysms with implication of geometric variations. J Biomed Eng 2006; 128: 844–851.10.1115/1.2354209Search in Google Scholar PubMed PubMed Central

[20] Janiga G, Beuing O, Seshadhri S, et al. Virtual stenting using real patient data. In: Vad J, editor. 14th Conference on Modelling Fluid Flow, Budapest, Hungary: ISBN 978-963-420-987-4 2009: 111–117.Search in Google Scholar

[21] Janiga G, Rössl C, Skalej M, Thévenin D. Realistic virtual intracranial stenting and computational fluid dynamics for treatment analysis. J Biomech 2013; 46: 7–12.10.1016/j.jbiomech.2012.08.047Search in Google Scholar PubMed

[22] Kim M, Taulbee DB, Tremmel M, Meng H. Comparison of two stents in modifying cerebral aneurysm hemodynamics. Ann Biomed Eng 2008; 36: 726–741.10.1007/s10439-008-9449-4Search in Google Scholar PubMed PubMed Central

[23] Lanzino G, Wakhloo AK, Fessler RD, Hartney ML, Guterman LR, Hopkins LN. Efficacy and current limitations of intravascular stents for intracranial internal carotid, vertebral, and basilar artery aneurysms. J Neurosurg 1999; 91: 538–546.10.3171/jns.1999.91.4.0538Search in Google Scholar PubMed

[24] Liepsch DW. Flow in tubes and arteries – a comparison. Biorheology 1986; 23: 395–433.10.3233/BIR-1986-23408Search in Google Scholar

[25] Liou TM, Liou SN. Pulsatile flows in a lateral aneurysm anchored on a stented and curved parent vessel. Exp Mech 2004; 44: 253–260.10.1007/BF02427891Search in Google Scholar

[26] Löhner R, Cebral J, Soto O, Yim P, Burgess JE. Applications of patient-specific CFD in medicine and life sciences. Int J Numer Methods Fluids 2003; 43: 637–650.10.1002/fld.544Search in Google Scholar

[27] Marzo A, Singh P, Larrabide I, et al. Computational hemodynamics in cerebral aneurysms: the effects of modeled versus measured boundary conditions. Ann Biomed Eng 2011; 39: 884–896.10.1007/s10439-010-0187-zSearch in Google Scholar PubMed

[28] Morales HG, Kim M, Vivas EE, et al. How do coil configuration and packing density influence intra-aneurysmal hemodynamics? Am J Neuroradiol 2011; 32: 1935–1941.10.3174/ajnr.A2635Search in Google Scholar PubMed PubMed Central

[29] Neugebauer M, Janiga G, Zachow S, Skalej M, Preim B. Generierung qualitativ hochwertiger Modelle für die Simulation von Blutfluss in zerebralen Aneurysmen. In: Hauser H, Straßburger S, Theisel H, editors. Simulation and visualization. Magdeburg, Germany: SCS Publishing House e.V 2008: 221–236.Search in Google Scholar

[30] Owen SJ, White DR. Mesh-based geometry. Int J Numer Methods Eng 2003; 58: 375–395.10.1002/nme.773Search in Google Scholar

[31] Radaelli AG, Augsburger L, Cebral JR, et al. Reproducibility of haemodynamical simulations in a subject-specific stented aneurysm model – a report on the Virtual Intracranial Stenting Challenge 2007. J Biomech 2008; 41: 2069–2081.10.1016/j.jbiomech.2008.04.035Search in Google Scholar PubMed

[32] Ryu CW, Kwon OK, Koh JS, Kim EJ. Analysis of aneurysm rupture in relation to the geometric indices: aspect ratio, volume, and volume-to-neck ratio. Neuroradiology 2011; 53: 883–889.10.1007/s00234-010-0804-4Search in Google Scholar PubMed

[33] Schneiders JJ, Vanbavel E, Majoie CB, Ferns SP, van den Berg R. A flow-diverting stent is not a pressure-diverting stent. Am J Neuroradiol 2013; 34: E1–E4.10.3174/ajnr.A2613Search in Google Scholar PubMed PubMed Central

[34] Schöberl J. NETGEN: an advancing front 2D/3D-mesh generator based on abstract rules. Comput Visual Sci 1997; 1: 41–52.10.1007/s007910050004Search in Google Scholar

[35] Schumann C, Neugebauer M, Bade R, Preim B, Peitgen H-O. Implicit vessel surface reconstruction for visualization and CFD simulation. Int J Comput Assist Radiol and Surg 2008; 2: 275–286.10.1007/s11548-007-0137-xSearch in Google Scholar

[36] Seshadhri S, Janiga G, Beuing O, Skalej M, Thevenin D. Impact of stents and flow diverters on hemodynamics in idealized aneurysm models. J Biomech Eng 2011; 133: 071005.10.1115/1.4004410Search in Google Scholar PubMed

[37] Sforza DM, Putman CM, Cebral JR. Hemodynamics of cerebral aneurysms. Annu Rev Fluid Mech 2009; 41: 91–107.10.1146/annurev.fluid.40.111406.102126Search in Google Scholar PubMed PubMed Central

[38] Sforza DM, Putman CM, Scrivano E, Lylyk P, Cebral JR. Blood-flow characteristics in a terminal basilar tip aneurysm prior to its fatal rupture. Am J Neuroradiol 2010; 31: 1127–1131.10.3174/ajnr.A2021Search in Google Scholar PubMed PubMed Central

[39] Shobayashi Y, Tanoue T, Tateshima S, Tanishita K. Mechanical design of an intracranial stent for treating cerebral aneurysms. Med Eng Phys 2010; 32: 1015–1024.10.1016/j.medengphy.2010.07.002Search in Google Scholar PubMed

[40] Shobayashi Y, Tateshima S, Kakizaki R, Sudo R, Tanishita K, Vinuela F. Intra-aneurysmal hemodynamic alterations by a self-expandable intracranial stent and flow diversion stent: high intra-aneurysmal pressure remains regardless of flow velocity reduction. J Neuro Intervent Surg 2012. doi:10.1136/neurintsurg-2012-010488. http://jnis.bmj.com/content/early/2012/10/08/neurintsurg-2012-010488.long.Search in Google Scholar

[41] Steinman DA, Hoi Y, Fahy P, et al. Variability of computational fluid dynamics solutions for pressure and flow in a giant aneurysm: the ASME 2012 Summer Bioengineering Conference CFD challenge. J Biomech Eng 2013; 135: 021015/1–13.Search in Google Scholar

[42] Tanoue T, Tateshima S, Villablanca JP, Vinuela F, Tanishita K. Wall shear stress distribution inside growing cerebral aneurysm. Am J Neuroradiol 2011; 32: 1732–1737.10.3174/ajnr.A2607Search in Google Scholar PubMed PubMed Central

[43] Ujiie H, Liepsch DW, Goetz M, Yamaguchi R, Yonetani H, Takakura K. Hemodynamic study of the anterior communicating artery. Stroke 1996; 27: 2086–2093; discussion 2094.10.1161/01.STR.27.11.2086Search in Google Scholar PubMed

[44] Valencia A, Morales H, Rivera R, Bravo E, Galvez M. Blood flow dynamics in patient-specific cerebral aneurysm models: the relationship between wall shear stress and aneurysm area index. Med Eng Phys 2008; 30: 329–340.10.1016/j.medengphy.2007.04.011Search in Google Scholar PubMed

[45] Wakhloo AK, Lanzino G, Lieber BB, Hopkins LN. Stents for intracranial aneurysms: the beginning of a new endovascular era? Neurosurgery 1998; 43: 377–379.10.1097/00006123-199808000-00126Search in Google Scholar PubMed

[46] Xiang J, Natarajan SK, Tremmel M, et al. Hemodynamic-morphologic discriminants for intracranial aneurysm rupture. Stroke 2011; 42: 144–152.10.1161/STROKEAHA.110.592923Search in Google Scholar PubMed PubMed Central

Received: 2012-12-21
Accepted: 2013-4-3
Published Online: 2013-05-10
Published in Print: 2013-06-01

©2013 by Walter de Gruyter Berlin Boston

Downloaded on 29.4.2024 from https://www.degruyter.com/document/doi/10.1515/bmt-2012-0119/html