Skip to content
Licensed Unlicensed Requires Authentication Published by De Gruyter March 14, 2015

Adhesive strength of total knee endoprostheses to bone cement – analysis of metallic and ceramic femoral components under worst-case conditions

  • Philipp Bergschmidt EMAIL logo , Rebecca Dammer , Carmen Zietz , Susanne Finze , Wolfram Mittelmeier and Rainer Bader

Abstract

Evaluation of the adhesive strength of femoral components to the bone cement is a relevant parameter for predicting implant safety. In the present experimental study, three types of cemented femoral components (metallic, ceramic and silica/silane-layered ceramic) of the bicondylar Multigen Plus knee system, implanted on composite femora were analysed. A pull-off test with the femoral components was performed after different load and several cementing conditions (four groups and n=3 components of each metallic, ceramic and silica/silane-layered ceramic in each group). Pull-off forces were comparable for the metallic and the silica/silane-layered ceramic femoral components (mean 4769 N and 4298 N) under standard test condition, whereas uncoated ceramic femoral components showed reduced pull-off forces (mean 2322 N). Loading under worst-case conditions led to decreased adhesive strength by loosening of the interface implant and bone cement using uncoated metallic and ceramic femoral components, respectively. Silica/silane-coated ceramic components were stably fixed even under worst-case conditions. Loading under high flexion angles can induce interfacial tensile stress, which could promote early implant loosening. In conclusion, a silica/silane-coating layer on the femoral component increased their adhesive strength to bone cement. Thicker cement mantles (>2 mm) reduce adhesive strength of the femoral component and can increase the risk of cement break-off.


Corresponding author: PD Dr. med. habil. Philipp Bergschmidt, Department of Orthopaedics, University Medicine Rostock, Doberaner Str. 142, 18057 Rostock, Germany, Phone: +49-381-494-9374, Fax: +49-381-494-9303, E-mail:

Acknowledegment

The authors would like to thank CeramTec GmbH, Plochingen, Germany for supporting this experimental study.

References

[1] Atsu SS, Kilicarslan MA, Kucukesmen HC, Aka PS. Effect of zirconium-oxide ceramic surface treatments on the bond strength to adhesive resin. J Prosthet Dent 2006; 95: 430–436.10.1016/j.prosdent.2006.03.016Search in Google Scholar

[2] Bader R, Bergschmidt P, Fritsche A, Ansorge S, Thomas P, Mittelmeier W. Alternative materials and solutions in total knee arthroplasty for patients with metal allergy. Orthopaede 2008; 37: 136–142.10.1007/s00132-007-1189-xSearch in Google Scholar

[3] Bader R, Mittelmeier W, Steinhauser E. Failure analysis of total knee replacement. Basics and methodological aspects of the damage analysis. Orthopaede 2006; 35: 896–903.10.1007/s00132-006-0976-0Search in Google Scholar

[4] Benazzo F, Macchi F, Ross S, DallaPria P. Ceramic total knee arthroplasty – an update. Eur Muskuloskel Rev 2007; 59–62.Search in Google Scholar

[5] Bergmann G, Heinlein B, Graichen F, Bender A, Rohlmann A, Kutzner I. Loading of the Knee Joint During Frequent Strenuous Activities– Measured in Four Patients. in 55th Annual Meeting of the Orthopaedic Research Society; Las Vegas 2009.Search in Google Scholar

[6] Bergschmidt P, Bader R, Finze S, Ansorge S, Kundt G, Mittelmeier W. Bicondylar knee arthroplasty – influence of preoperative functional restriction on early functional postoperative outcome. Z Orthop Unfall 2008; 146: 344–351.10.1055/s-2008-1038463Search in Google Scholar

[7] Bergschmidt P, Bader R, Ganzer D, et al. Ceramic femoral components in total knee arthroplasty – two year follow-up results of an International Prospective Multi-Centre Study. Open Orthop J 2012; 6: 176–182.10.2174/1874325001206010172Search in Google Scholar

[8] Bergschmidt P, Bader R, Mittelmeier W. Metal hypersensitivity in total knee arthroplasty – revision surgery using a ceramic femoral component – a case report. Knee 2012; 19: 144–147.10.1016/j.knee.2011.01.001Search in Google Scholar

[9] Bergschmidt P, Lohmann C, Ganzer D, et al. Total knee replacement with ceramic femoral components: a national prospective multicenter study of clinical and radiological outcomes. Orthopaede 2011; 40: 224–230.10.1007/s00132-010-1649-6Search in Google Scholar

[10] Cristofolini L, Affatato S, Erani P, Leardini W, Tigani D, Viceconti M. Long-term implant-bone fixation of the femoral component in total knee replacement. Proc Inst Meach Eng J Eng Med 2008; 222: 319–331.10.1243/09544119JEIM328Search in Google Scholar

[11] Cristofolini L, Affatato S, Erani P, Tigani D, Viceconti M. Implant fixation in knee replacement: preliminary in vitro comparison of ceramic and metal cemented femoral components. Knee 2009; 16: 101–108.10.1016/j.knee.2008.08.006Search in Google Scholar

[12] Cristofolini L, Viceconti M, Cappello A, Toni A. Mechanical validation of whole bone composite femur models. J Biomech 1996; 29: 525–535.10.1016/0021-9290(95)00084-4Search in Google Scholar

[13] DallaPria P. Evolution and new application of the alumina ceramics in joint replacement. Eur J Orthop Surg Traumatol 2007; 17: 253–256.10.1007/s00590-006-0181-1Search in Google Scholar

[14] Habenicht G Kleben. Grundlagen, Technologien, Anwendungen. Berlin: Springer 2009.10.1007/978-3-540-85266-7Search in Google Scholar

[15] Hamadouche M, Boutin P, Daussange J, Bolander ME, Sedel L. Alumina-on-alumina total hip arthroplasty: a minimum 18.5-year follow-up study. J Bone Joint Surg Am 2002; 84-A: 69–77.10.2106/00004623-200201000-00011Search in Google Scholar

[16] Han HS, Kang SB, Yoon KS. High incidence of loosening of the femoral component in legacy posterior stabilised-flex total knee replacement J Bone Joint Surg Br 2007; 89: 1457–1461.10.1302/0301-620X.89B11.19840Search in Google Scholar

[17] Hannouche D, Hamadouche M, Nizard R, Bizot P, Meunier A, Sedel L. Ceramics in total hip replacement. Clin Orthop Relat Res 2005; 430: 62–71.10.1097/01.blo.0000149996.91974.83Search in Google Scholar

[18] Inacio MC, Cafri G, Paxton EW, Kurtz SM, Namba RS. Alternative bearings in total knee arthroplasty: risk of early revision compared to traditional bearings: an analysis of 62,177 primary cases. Acta Orthop 2013; 84: 145–152.10.3109/17453674.2013.784660Search in Google Scholar

[19] Kircher J, Bergschmidt P, Bader R, et al. The importance of wear couples for younger endoprosthesis patients. Orthopaede 2007; 36: 337–346.10.1007/s00132-007-1069-4Search in Google Scholar

[20] Lee AJC. The time-dependent properties of polymethylmethacrylate bone cement: the interaction of shape of femoral stems, surface finish and bone cement. In: Learmonth IA, editor. Interfaces in arthroplasty. Berlin, Heidelberg, New York: Springer 2000.10.1007/978-1-4471-0477-3_2Search in Google Scholar

[21] Lung CY, Matinlinna JP. Aspects of silane coupling agents and surface conditioning in dentistry: an overview. Dent Mater 2012; 28: 467–477.10.1016/j.dental.2012.02.009Search in Google Scholar

[22] Markus A, Wimmer MA, Andriacchi TP. Tractive forces during rolling motion of the knee: implications for wear in total knee replacement. J Biomech 1997; 30; 131–137.10.1016/S0021-9290(96)00112-1Search in Google Scholar

[23] Marx R, Faramarzi R, Oberbach T, Begand S, Grätz N, Wirtz DC. PVD-silicoating before cementation of zirconia-based knee prostheses effects better cement adhesion and lower aseptic loosening rates. Z Orthop Unfall 2012; 150: 40–47.10.1055/s-0031-1280249Search in Google Scholar PubMed

[24] Müller K. Mechanische Eigenschaften nicht-modifizierter Knochenzemente auf PMMA-Basis. In: Willert HG, Buchhorn G, editors. Knochenzement-Werkstoff, klinische Erfahrungen, Weiterentwicklungen. Bern: Huber 1987.Search in Google Scholar

[25] Mumme T, Marx R, Müller-Rath R, Gravius S, Andereya S, Wirtz DC. Surface pretreatment of endoprostheses by silica/silane to optimise the hydrolytic stability between bone cement and metal. Total hip and knee arthroplasty. Orthopaede 2008; 37: 240–244, 246–250.10.1007/s00132-008-1198-4Search in Google Scholar PubMed

[26] Nakamura S, Kobayashi M, Ito H, Nakamura K, Ueo T, Nakamura T. The Bi-Surface total knee arthroplasty: minimum 10-year follow-up study. Knee 2010; 17: 274–278.10.1016/j.knee.2010.02.015Search in Google Scholar PubMed

[27] Oonishi H, Aono M, Murata N, Kushitani S. Aluminia versus polyethylen in total knee arthroplasty. Clin Orthop 1992; 282: 95–104.10.1097/00003086-199209000-00012Search in Google Scholar

[28] Plueddemann EP. Silane coupling agents. New York: Plenum 1982.10.1007/978-1-4899-0342-6Search in Google Scholar

[29] Schultze C, Kluess D, Martin H, et al. Finite element analysis of a cemented ceramic femoral component for the real assembly situation in total knee arthroplasty. Biomed Tech 2007; 52: 301–307.10.1515/BMT.2007.051Search in Google Scholar PubMed

[30] The Swedish Knee Arthroplasty Register – Annual report 2010, Lund: Wallin & Dallholm AB 2010.Search in Google Scholar

[31] Willert HG. Die Zerrüttung des Zementköchers. In: Willert HG, Buchhorn G, editors. Knochenzement-Werkstoff, klinische Erfahrungen, Weiterentwicklungen. Bern: Huber 1987.Search in Google Scholar

[32] Zelle J, Janssen D, Van Eijden J, De Waal Malefijt M, Verdonschot N. Does high-flexion total knee arthroplasty promote early loosening of the femoral component? J Orthop Res 2011; 29: 976–983.10.1002/jor.21363Search in Google Scholar PubMed

[33] Zietz C, Bergschmidt P, Lange R, Mittelmeier W, Bader R. Third-body abrasive wear of tibial polyethylene inserts combined with metallic and ceramic femoral components in a knee simulator study. Int J Artif Organs 2013; 36: 47–55.10.5301/ijao.5000189Search in Google Scholar PubMed

[34] Zietz C, Kluess D, Bergschmidt P, Haenle M, Mittelmeier W, Bader R. Tribological aspects of ceramics in total hip and knee arthroplasty. Semin Arthroplasty 2011; 22: 258–263.10.1053/j.sart.2011.10.006Search in Google Scholar

Received: 2014-8-25
Accepted: 2015-2-13
Published Online: 2015-3-14
Published in Print: 2016-6-1

©2016 by De Gruyter

Downloaded on 26.4.2024 from https://www.degruyter.com/document/doi/10.1515/bmt-2014-0090/html