Open Access 01.12.2011 | Research

Application of Benchtop-magnetic resonance imaging in a nude mouse tumor model

verfasst von: Henrike Caysa, Hendrik Metz, Karsten Mäder, Thomas Mueller

Erschienen in: Journal of Experimental & Clinical Cancer Research | Ausgabe 1/2011

Abstract

Background

MRI plays a key role in the preclinical development of new drugs, diagnostics and their delivery systems. However, very high installation and running costs of existing superconducting MRI machines limit the spread of MRI. The new method of Benchtop-MRI (BT-MRI) has the potential to overcome this limitation due to much lower installation and almost no running costs. However, due to the low field strength and decreased magnet homogeneity it is questionable, whether BT-MRI can achieve sufficient image quality to provide useful information for preclinical in vivo studies. It was the aim of the current study to explore the potential of BT-MRI on tumor models in mice.

Methods

We used a prototype of an in vivo BT-MRI apparatus to visualise organs and tumors and to analyse tumor progression in nude mouse xenograft models of human testicular germ cell tumor and colon carcinoma.

Results

Subcutaneous xenografts were easily identified as relative hypointense areas in transaxial slices of NMR images. Monitoring of tumor progression evaluated by pixel extension analyses based on NMR images correlated with increasing tumor volume calculated by calliper measurement. Gd-BOPTA contrast agent injection resulted in a better differentiation between parts of the urinary tissues and organs due to fast elimination of the agent via kidneys. In addition, interior structuring of tumors could be observed. A strong contrast enhancement within a tumor was associated with a central necrotic/fibrotic area.

Conclusions

BT-MRI provides satisfactory image quality to visualize organs and tumors and to monitor tumor progression and structure in mouse models.
Hinweise

Electronic supplementary material

The online version of this article (doi:10.​1186/​1756-9966-30-69) contains supplementary material, which is available to authorized users.

Competing interests

The authors declare that they have no competing interests.

Authors' contributions

HC, HM, KM and TM designed the study. HC, HM and TM performed experiments. HC, HM, KM and TM analysed data. HC and TM wrote the paper. All gave final approval.
Abkürzungen
MRI
magnetic resonance imaging
BT-MRI
benchtop-magnetic resonance imaging
NMR
nuclear magnetic resonance
Gd-BOPTA
gadobenate dimeglumine
s.c.
subcutaneous
HE
hematoxylin/eosin

Background

MRI plays a key role in the preclinical development of new drugs, diagnostics and their delivery systems. However, very high installation and running cost of existing superconducting MRI machines limit the spread of the method. The new method of Benchtop-MRI (BT-MRI) has the potential to overcome this limitation due to much lower installation and almost no running costs. The lower quality of the NMR images is expected due to the low field strength and decreased magnet homogeneity. However, very recently we could show that BT-MRI is able to characterize floating mono- or bilayer tablets, osmotic controlled push-pull tablets [14] or scaffolds for tissue engineering in vitro [5]. A broad, important and increasing range of MRI applications are linked with preclinical studies on small rodents such as mice or rats [68]. Thereby, first developments and testing of more compact MRI systems have been reported [9, 10]. In the present study we have tested a prototype of a new in vivo BT-MRI apparatus.
Clearly, BT-MRI could overcome one of the current main limitations of preclinical MRI, the high costs. However, the question arises, whether BT-MRI can achieve sufficient image quality to provide useful information for preclinical in vivo studies. In a recent paper we have demonstrated that BT-MRI can be used to characterize in situ forming implants in mice [11]. A major application field of preclinical MRI is linked to cancer research. It was therefore the aim of the current study to explore the potential of BT-MRI on tumor models in mice. Nude mouse xenograft models of different human tumors were used to test the suitability of the new BT-MRI system for visualisation of organs and tumors and for quantification of tumor progression.

Methods

NMR system and its characteristics

A 21 MHz NMR benchtop prototype system "MARAN DRX2" (Oxford Instruments) capable of imaging with a horizontal bore of 23 mm diameter was used (Figure 1). The instrument is equipped with a temperature control unit and capable of T1 and T2 relaxation measurements, the determination of diffusion coefficients and imaging.

NMR imaging parameter

The temperature was set to 37°C. Always 4 slices were simultaneously measured with: slice distance: 3.5 mm, slice width: 3 mm, spin echo time TE: 9.8 ms, repetition time TR: 172 ms, averages: 32 or 16 (for time critical kinetics), total time: 715 s or 357 s, respectively, FOV: 40*40 mm. The pulse sequence was T2SE.
The MRI acquisition parameters were optimized under some hardware restrictions. TE is limited by the bandwidth of 10 KHz to 9.8 ms. An increase of the bandwidth allows shorter TE, however it leads also to stronger image distortions. A TR value of 150 ms gives an optimal contrast for marbled meat and also for mice. For 4 slices TR is limited to 171.4 ms. Therefore 172 ms was used for TR as a good compromise between best contrast and simultaneous acquisition of 4 slices. The resulting images are therefore T1-weighted and range from hyperintense signals for fatty tissues to hypointense signals for water. The higher number of averages was chosen to improve the signal-to-noise ratio. For kinetics of contrast agent distribution a rapid image acquisition may be essential. Therefore measurements with lesser averages were also performed, even though the image quality is reduced.

Cell culture, xenograft tumor model, measurements and analyses

Human colon carcinoma cell lines DLD-1, HCT8 and HT29 and human testicular germ cell tumor cell line 1411HP were maintained as monolayer cultures in RPMI-1640 with 10% FCS and streptomycin/penicillin. Cultures were grown at 37°C in a humidified atmosphere of 5% CO2/95% air.
Eight week old male athymic-nude Foxn1 nu/nu mice (Harlan Winkelmann, Germany) were injected s.c. with 3 × 106 tumor cells in both flanks. NMR Imaging of mice was performed once a week. For comparison, the size of the xenograft tumors was also measured by means of a calliper. For imaging with a positive MRI contrast agent mice received 150 μl of gadobenate dimeglumine (Gd-BOPTA; 0.03 mmol/kg in 0.9% NaCl) via tail vein injection. For investigation of contrast agent associated effects with special focus on xenograft tumors the dose of Gd-BOPTA was increased according to dosage applied in men (0.1 mmol/kg). Animals were anaesthetised via i.p. application of ketamine/xylazine mixture prior to imaging. Body weight was assessed twice weekly. For histological examination tumors were explanted, fixed in 4% formalin and embedded in paraffin. Hematoxylin/Eosin staining of slices was performed according to standard protocols. All animal protocols were approved by the laboratory animal care and use committee of Sachsen-Anhalt, Germany.
Quantification of xenograft tumor growth was performed by
1.) volume calculation based on calliper measurements using the formula a 2 × b × π/6 with a being the short and b the long dimension and
2.) measurement of pixel extensions of tumor sections based on NMR images (128 × 128 JPG) using the measure tool of GNU Image Manipulation Program (GIMP 2.6.8) and calculating the area using formula A = a/2 × b/2 × π.

Results

Imaging of organs and tumors; gadobenate dimeglumine (Gd-BOPTA) induced MRI contrast

A nude mouse xenograft model of different human tumors was used to determine the image sensitivity and quality of the BT-MRI system. Gd-BOPTA as one of the clinically used low molecular weight gadolinium chelates was selected for contrast agent enhanced MRI. A good differentiation between cortex of kidney and renal pelvis could be observed depending on circulation time of the contrast agent (Figure 2A). Furthermore, the fast renal elimination of Gd-BOPTA was visualised. The urinary bladder was visible as a bright, hypertense sphere unlike the NMR image without contrast agent (Figure 2B). Subcutaneous xenograft tumors were easily identified as relative hypointense area at each body site (Figure 2C).
To study the contrast agent associated effects with special focus on xenograft tumors we used a higher dose of Gd-BOPTA according to dosage applied in men. As shown in Figure 3A an interior structuring of tumors could be observed. This was characterized by time dependent alterations of contrast enhancement with initial enhancement of the tumor rim followed by a centripetal progression of the signal. In one case of a strong central contrast enhancement (Figure 3B) the tumor was explanted, fixed and slices were analysed histologically after HE staining. A large central necrotic/fibrotic area could be observed surrounded by peripherally arranged vital tumor cells (Figure 3C).

Monitoring of xenograft tumor growth

Apart from tumor detection the quantification of tumor burden is one important aspect of non-invasive in vivo imaging techniques. To test whether the BT-MRI system is suitable for following s.c. xenograft growth the tumor burden was examined in 2 groups of 3 mice each bearing 2 different tumors: one group with 1411HP germ cell tumor and DLD-1 colon carcinoma, one group with HT29 colon carcinoma and DLD-1 colon carcinoma. Growth of tumors was followed using (a) calliper measurement and volume calculation and (b) BT-MRI and measurement of pixel extensions of tumor sections based on NMR images. For both methods comparable progression profiles could be observed, which was independent of Gd-BOPTA injection. A representative example of one individual is presented in Figure 4A and 4B. In addition, all values calculated by pixel extension analyses were plotted dependent on respective values calculated by calliper measurement. This demonstrates the correlation of both applications (Figure 4C).

Discussion

MRI as a non-invasive imaging technology plays a key role in preclinical in vivo evaluation of tumor therapies. The development of a BT-MRI system for small animal imaging could lead to easy detection of tumor mass and progression with little effort and low costs. Additionally, MRI provides an insight into organs and tissues of laboratory animals.
The experimental results clearly proof that BT-MRI can be used to visualise organs and tumors in nude mouse xenograft models. Subcutaneous xenografts were easily identified as relative hypointense areas in transaxial slices of NMR images. In addition BT-MRI system is suitable for following xenograft tumor growth. Monitoring of tumor progression evaluated by pixel extension analyses based on NMR images correlated with increasing tumor volume calculated by calliper measurement. This is an important requirement for application of BT-MRI system in orthotopic/metastatic tumor models to evaluate the whole tumor burden. For this purpose it is necessary to take serial slices of NMR images to get the largest dimension of the tumor as basis for calculation. In addition the whole tumor shape can be reconstituted.
One critical aspect using orthotopic/metastatic tumor models could be the visualization of metastasis in tissues and organs depending on the model. This may require application of contrast agent for differentiation between tumor and normal tissue. In this study we used Gd-BOPTA as one of the clinically used low molecular weight gadolinium chelates. Gd chelates are commonly used as MRI contrast agents for the detection of solid tumors in patients where an initial tumor rim enhancement is usually observed [1218]. Thereby the characteristic enhancement of the tumor rim can be used for the differentiation between malignant and benign masses [15]. Initially most tumors in our study showed no peripheral contrast enhancement on NMR images. Applying a higher but well tolerated dose of Gd-BOPTA such an effect could be observed, albeit not in each case. This may be due to the artificial location of the tumor as subcutaneous xenograft. Moreover, it was observed that low molar mass Gd chelates show an initial rim enhancement, followed by a washout effect, which requires that the images are obtained within the first 2 min after injection [19]. This probably explains the lack of initial rim enhancement in our models after application of low dose Gd-BOPTA. In this regard the application of macromolecular MRI contrast agents could be useful [20]. They have a longer circulation time and are more confined to the blood pool, therefore giving a longer time window for imaging in mice models.
A main advantage of MRI is the capability to characterize important tumor characteristics (e.g. internal structure, oedema in the tumor environment, necrotic areas). We observed a pronounced interior structuring of an s.c. HT29 tumor after i.v. injection of the contrast agent Gd-BOPTA. Histological analyses revealed that a large central necrotic/fibrotic area was associated with contrast enhancement. Such an effect can also be observed in patient tumors. After the characteristic initial tumor rim enhancement a centripetal progression of the signal can occur depending on the tumor structure, e.g. determined by different vascular architecture [12, 15, 21]. Early peripheral enhancement with centripetal progression was seen in invasive carcinomas with a high peripheral and a low central microvessel density, which was associated with fibrosis and/or necrosis [12, 21]. This demonstrates that depending on the tumor and used contrast agent the BT-MRI system is suitable for observation of intratumoral structures and that characteristic features of patient tumors can be reproduced in the model system. It offers the opportunity to follow intratumoral processes under therapy.
Further work will be done particularly with regard to imaging of different orthotopic installed tumors and their progression as well as the development of metastatic disease. Other contrast agents will also be examined in order to find better enhancement of (small) tumor sites and metastases. Moreover, other contrast enhancer could lead to better results for imaging of interior tumor structures.

Conclusions

The results of the current study show that BT-MRI is, despite its limitations with respect to the magnetic field strength and magnet homogeneity, clearly capable of providing satisfactory image slice quality to visualize organs and tumors and to monitor tumor progression in mouse models.

Acknowledgements

We would like to thank Dr. Ian Nicholson and his colleagues from Oxford Instruments for the development, manufacture and installation of the BT-MRI prototype apparatus.
The study was supported in part by grants from the Federal State of Saxonia-Anhalt (FKZ 3646A/0907).
Open Access This article is published under license to BioMed Central Ltd. This is an Open Access article is distributed under the terms of the Creative Commons Attribution License ( https://​creativecommons.​org/​licenses/​by/​2.​0 ), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.

Competing interests

The authors declare that they have no competing interests.

Authors' contributions

HC, HM, KM and TM designed the study. HC, HM and TM performed experiments. HC, HM, KM and TM analysed data. HC and TM wrote the paper. All gave final approval.
Anhänge

Authors’ original submitted files for images

Literatur
1.
Zurück zum Zitat Malaterre V, Metz H, Ogorka J, Gurny R, Loggia N, Mader K: Benchtop-magnetic resonance imaging (BT-MRI) characterization of push-pull osmotic controlled release systems. J Control Release. 2009, 133: 31-36. 10.1016/j.jconrel.2008.09.007.CrossRefPubMed Malaterre V, Metz H, Ogorka J, Gurny R, Loggia N, Mader K: Benchtop-magnetic resonance imaging (BT-MRI) characterization of push-pull osmotic controlled release systems. J Control Release. 2009, 133: 31-36. 10.1016/j.jconrel.2008.09.007.CrossRefPubMed
2.
Zurück zum Zitat Metz H, Mader K: Benchtop-NMR and MRI - a new analytical tool in drug delivery research. Int J Pharm. 2008, 364: 170-175. 10.1016/j.ijpharm.2008.09.033.CrossRefPubMed Metz H, Mader K: Benchtop-NMR and MRI - a new analytical tool in drug delivery research. Int J Pharm. 2008, 364: 170-175. 10.1016/j.ijpharm.2008.09.033.CrossRefPubMed
3.
Zurück zum Zitat Strubing S, Abboud T, Contri RV, Metz H, Mader K: New insights on poly(vinyl acetate)-based coated floating tablets: characterisation of hydration and CO2 generation by benchtop MRI and its relation to drug release and floating strength. Eur J Pharm Biopharm. 2008, 69: 708-717. 10.1016/j.ejpb.2007.12.009.CrossRefPubMed Strubing S, Abboud T, Contri RV, Metz H, Mader K: New insights on poly(vinyl acetate)-based coated floating tablets: characterisation of hydration and CO2 generation by benchtop MRI and its relation to drug release and floating strength. Eur J Pharm Biopharm. 2008, 69: 708-717. 10.1016/j.ejpb.2007.12.009.CrossRefPubMed
4.
Zurück zum Zitat Strubing S, Metz H, Mader K: Characterization of poly(vinyl acetate) based floating matrix tablets. J Control Release. 2008, 126: 149-155. 10.1016/j.jconrel.2007.11.013.CrossRefPubMed Strubing S, Metz H, Mader K: Characterization of poly(vinyl acetate) based floating matrix tablets. J Control Release. 2008, 126: 149-155. 10.1016/j.jconrel.2007.11.013.CrossRefPubMed
5.
Zurück zum Zitat Nitzsche H, Metz H, Lochmann A, Bernstein A, Hause G, Groth T, Mader K: Characterization of scaffolds for tissue engineering by benchtop-magnetic resonance imaging. Tissue Eng Part C Methods. 2009, 15: 513-521. 10.1089/ten.tec.2008.0488.CrossRefPubMed Nitzsche H, Metz H, Lochmann A, Bernstein A, Hause G, Groth T, Mader K: Characterization of scaffolds for tissue engineering by benchtop-magnetic resonance imaging. Tissue Eng Part C Methods. 2009, 15: 513-521. 10.1089/ten.tec.2008.0488.CrossRefPubMed
6.
Zurück zum Zitat Benoit MR, Mayer D, Barak Y, Chen IY, Hu W, Cheng Z, Wang SX, Spielman DM, Gambhir SS, Matin A: Visualizing implanted tumors in mice with magnetic resonance imaging using magnetotactic bacteria. Clin Cancer Res. 2009, 15: 5170-5177. 10.1158/1078-0432.CCR-08-3206.PubMedCentralCrossRefPubMed Benoit MR, Mayer D, Barak Y, Chen IY, Hu W, Cheng Z, Wang SX, Spielman DM, Gambhir SS, Matin A: Visualizing implanted tumors in mice with magnetic resonance imaging using magnetotactic bacteria. Clin Cancer Res. 2009, 15: 5170-5177. 10.1158/1078-0432.CCR-08-3206.PubMedCentralCrossRefPubMed
7.
Zurück zum Zitat McConville P, Hambardzumyan D, Moody JB, Leopold WR, Kreger AR, Woolliscroft MJ, Rehemtulla A, Ross BD, Holland EC: Magnetic resonance imaging determination of tumor grade and early response to temozolomide in a genetically engineered mouse model of glioma. Clin Cancer Res. 2007, 13: 2897-2904. 10.1158/1078-0432.CCR-06-3058.CrossRefPubMed McConville P, Hambardzumyan D, Moody JB, Leopold WR, Kreger AR, Woolliscroft MJ, Rehemtulla A, Ross BD, Holland EC: Magnetic resonance imaging determination of tumor grade and early response to temozolomide in a genetically engineered mouse model of glioma. Clin Cancer Res. 2007, 13: 2897-2904. 10.1158/1078-0432.CCR-06-3058.CrossRefPubMed
8.
Zurück zum Zitat Brockmann MA, Kemmling A, Groden C: Current issues and perspectives in small rodent magnetic resonance imaging using clinical MRI scanners. Methods. 2007, 43: 79-87. 10.1016/j.ymeth.2007.07.001.CrossRefPubMed Brockmann MA, Kemmling A, Groden C: Current issues and perspectives in small rodent magnetic resonance imaging using clinical MRI scanners. Methods. 2007, 43: 79-87. 10.1016/j.ymeth.2007.07.001.CrossRefPubMed
9.
Zurück zum Zitat Inoue Y, Nomura Y, Haishi T, Yoshikawa K, Seki T, Tsukiyama-Kohara K, Kai C, Okubo T, Ohtomo K: Imaging living mice using a 1-T compact MRI system. J Magn Reson Imaging. 2006, 24: 901-907. 10.1002/jmri.20713.CrossRefPubMed Inoue Y, Nomura Y, Haishi T, Yoshikawa K, Seki T, Tsukiyama-Kohara K, Kai C, Okubo T, Ohtomo K: Imaging living mice using a 1-T compact MRI system. J Magn Reson Imaging. 2006, 24: 901-907. 10.1002/jmri.20713.CrossRefPubMed
10.
Zurück zum Zitat Shirai T, Haishi T, Utsuzawa S, Matsuda Y, Kose K: Development of a compact mouse MRI using a yokeless permanent magnet. Magn Reson Med Sci. 2005, 4: 137-143. 10.2463/mrms.4.137.CrossRefPubMed Shirai T, Haishi T, Utsuzawa S, Matsuda Y, Kose K: Development of a compact mouse MRI using a yokeless permanent magnet. Magn Reson Med Sci. 2005, 4: 137-143. 10.2463/mrms.4.137.CrossRefPubMed
11.
Zurück zum Zitat Kempe S, Metz H, Pereira PG, Mader K: Non-invasive in vivo evaluation of in situ forming PLGA implants by benchtop magnetic resonance imaging (BT-MRI) and EPR spectroscopy. Eur J Pharm Biopharm. 2009 Kempe S, Metz H, Pereira PG, Mader K: Non-invasive in vivo evaluation of in situ forming PLGA implants by benchtop magnetic resonance imaging (BT-MRI) and EPR spectroscopy. Eur J Pharm Biopharm. 2009
12.
Zurück zum Zitat Buadu LD, Murakami J, Murayama S, Hashiguchi N, Sakai S, Toyoshima S, Masuda K, Kuroki S, Ohno S: Patterns of peripheral enhancement in breast masses: correlation of findings on contrast medium enhanced MRI with histologic features and tumor angiogenesis. J Comput Assist Tomogr. 1997, 21: 421-430. 10.1097/00004728-199705000-00016.CrossRefPubMed Buadu LD, Murakami J, Murayama S, Hashiguchi N, Sakai S, Toyoshima S, Masuda K, Kuroki S, Ohno S: Patterns of peripheral enhancement in breast masses: correlation of findings on contrast medium enhanced MRI with histologic features and tumor angiogenesis. J Comput Assist Tomogr. 1997, 21: 421-430. 10.1097/00004728-199705000-00016.CrossRefPubMed
13.
Zurück zum Zitat Geirnaerdt MJ, Bloem JL, van der Woude HJ, Taminiau AH, Nooy MA, Hogendoorn PC: Chondroblastic osteosarcoma: characterisation by gadolinium-enhanced MR imaging correlated with histopathology. Skeletal Radiol. 1998, 27: 145-153. 10.1007/s002560050354.CrossRefPubMed Geirnaerdt MJ, Bloem JL, van der Woude HJ, Taminiau AH, Nooy MA, Hogendoorn PC: Chondroblastic osteosarcoma: characterisation by gadolinium-enhanced MR imaging correlated with histopathology. Skeletal Radiol. 1998, 27: 145-153. 10.1007/s002560050354.CrossRefPubMed
14.
15.
Zurück zum Zitat Ma LD, Frassica FJ, McCarthy EF, Bluemke DA, Zerhouni EA: Benign and malignant musculoskeletal masses: MR imaging differentiation with rim-to-center differential enhancement ratios. Radiology. 1997, 202: 739-744.CrossRefPubMed Ma LD, Frassica FJ, McCarthy EF, Bluemke DA, Zerhouni EA: Benign and malignant musculoskeletal masses: MR imaging differentiation with rim-to-center differential enhancement ratios. Radiology. 1997, 202: 739-744.CrossRefPubMed
16.
Zurück zum Zitat Mitchell DG, Saini S, Weinreb J, De Lange EE, Runge VM, Kuhlman JE, Parisky Y, Johnson CD, Brown JJ, Schnall M, et al: Hepatic metastases and cavernous hemangiomas: distinction with standard- and triple-dose gadoteridol-enhanced MR imaging. Radiology. 1994, 193: 49-57.CrossRefPubMed Mitchell DG, Saini S, Weinreb J, De Lange EE, Runge VM, Kuhlman JE, Parisky Y, Johnson CD, Brown JJ, Schnall M, et al: Hepatic metastases and cavernous hemangiomas: distinction with standard- and triple-dose gadoteridol-enhanced MR imaging. Radiology. 1994, 193: 49-57.CrossRefPubMed
17.
Zurück zum Zitat Mussurakis S, Gibbs P, Horsman A: Peripheral enhancement and spatial contrast uptake heterogeneity of primary breast tumors: quantitative assessment with dynamic MRI. J Comput Assist Tomogr. 1998, 22: 35-46. 10.1097/00004728-199801000-00007.CrossRefPubMed Mussurakis S, Gibbs P, Horsman A: Peripheral enhancement and spatial contrast uptake heterogeneity of primary breast tumors: quantitative assessment with dynamic MRI. J Comput Assist Tomogr. 1998, 22: 35-46. 10.1097/00004728-199801000-00007.CrossRefPubMed
18.
Zurück zum Zitat Tsien C, Gomez-Hassan D, Ten Haken RK, Tatro D, Junck L, Chenevert TL, Lawrence T: Evaluating changes in tumor volume using magnetic resonance imaging during the course of radiotherapy treatment of high-grade gliomas: Implications for conformal dose-escalation studies. Int J Radiat Oncol Biol Phys. 2005, 62: 328-332. 10.1016/j.ijrobp.2004.10.010.CrossRefPubMed Tsien C, Gomez-Hassan D, Ten Haken RK, Tatro D, Junck L, Chenevert TL, Lawrence T: Evaluating changes in tumor volume using magnetic resonance imaging during the course of radiotherapy treatment of high-grade gliomas: Implications for conformal dose-escalation studies. Int J Radiat Oncol Biol Phys. 2005, 62: 328-332. 10.1016/j.ijrobp.2004.10.010.CrossRefPubMed
19.
Zurück zum Zitat Morris EA: Breast cancer imaging with MRI. Radiol Clin North Am. 2002, 40: 443-466. 10.1016/S0033-8389(01)00005-7.CrossRefPubMed Morris EA: Breast cancer imaging with MRI. Radiol Clin North Am. 2002, 40: 443-466. 10.1016/S0033-8389(01)00005-7.CrossRefPubMed
20.
Zurück zum Zitat Daldrup H, Shames DM, Wendland M, Okuhata Y, Link TM, Rosenau W, Lu Y, Brasch RC: Correlation of dynamic contrast-enhanced MR imaging with histologic tumor grade: comparison of macromolecular and small-molecular contrast media. AJR Am J Roentgenol. 1998, 171: 941-949.CrossRefPubMed Daldrup H, Shames DM, Wendland M, Okuhata Y, Link TM, Rosenau W, Lu Y, Brasch RC: Correlation of dynamic contrast-enhanced MR imaging with histologic tumor grade: comparison of macromolecular and small-molecular contrast media. AJR Am J Roentgenol. 1998, 171: 941-949.CrossRefPubMed
21.
Zurück zum Zitat Buadu LD, Murakami J, Murayama S, Hashiguchi N, Sakai S, Masuda K, Toyoshima S, Kuroki S, Ohno S: Breast lesions: correlation of contrast medium enhancement patterns on MR images with histopathologic findings and tumor angiogenesis. Radiology. 1996, 200: 639-649.CrossRefPubMed Buadu LD, Murakami J, Murayama S, Hashiguchi N, Sakai S, Masuda K, Toyoshima S, Kuroki S, Ohno S: Breast lesions: correlation of contrast medium enhancement patterns on MR images with histopathologic findings and tumor angiogenesis. Radiology. 1996, 200: 639-649.CrossRefPubMed
Metadaten
Titel
Application of Benchtop-magnetic resonance imaging in a nude mouse tumor model
verfasst von
Henrike Caysa
Hendrik Metz
Karsten Mäder
Thomas Mueller
Publikationsdatum
01.12.2011
Verlag
BioMed Central
Erschienen in
Journal of Experimental & Clinical Cancer Research / Ausgabe 1/2011
Elektronische ISSN: 1756-9966
DOI
https://doi.org/10.1186/1756-9966-30-69

Weitere Artikel der Ausgabe 1/2011

Journal of Experimental & Clinical Cancer Research 1/2011 Zur Ausgabe

Bei seelischem Stress sind Checkpoint-Hemmer weniger wirksam

03.06.2024 NSCLC Nachrichten

Wie stark Menschen mit fortgeschrittenem NSCLC von einer Therapie mit Immun-Checkpoint-Hemmern profitieren, hängt offenbar auch davon ab, wie sehr die Diagnose ihre psychische Verfassung erschüttert

Antikörper mobilisiert Neutrophile gegen Krebs

03.06.2024 Onkologische Immuntherapie Nachrichten

Ein bispezifischer Antikörper formiert gezielt eine Armee neutrophiler Granulozyten gegen Krebszellen. An den Antikörper gekoppeltes TNF-alpha soll die Zellen zudem tief in solide Tumoren hineinführen.

Erhebliches Risiko für Kehlkopfkrebs bei mäßiger Dysplasie

29.05.2024 Larynxkarzinom Nachrichten

Fast ein Viertel der Personen mit mäßig dysplastischen Stimmlippenläsionen entwickelt einen Kehlkopftumor. Solche Personen benötigen daher eine besonders enge ärztliche Überwachung.

15% bedauern gewählte Blasenkrebs-Therapie

29.05.2024 Urothelkarzinom Nachrichten

Ob Patienten und Patientinnen mit neu diagnostiziertem Blasenkrebs ein Jahr später Bedauern über die Therapieentscheidung empfinden, wird einer Studie aus England zufolge von der Radikalität und dem Erfolg des Eingriffs beeinflusst.

Update Onkologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.