Biomechanical analysis of bicortical versus unicortical locked plating of mid-clavicular fractures

verfasst von: Daniel Hamman, Derek Lindsey, Jason Dragoo

Erschienen in: Archives of Orthopaedic and Trauma Surgery | Ausgabe 6/2011

Einloggen, um Zugang zu erhalten

Abstract

Objectives

Operative fixation of displaced mid-shaft clavicle fractures has been shown to improve the functional outcomes and decrease the likelihood of non-union; however, little is known about the need for locking screws versus traditional screws. We, therefore, evaluated the strength of unicortical locked plating versus traditional bicortical non-locking fixation methods.

Methods

Ten matched pairs of fresh, frozen cadaver clavicle specimens were obliquely osteotomized through the mid-shaft to represent the most common fracture pattern. After randomization, the clavicles were repaired using pre-contoured plates with either standard bicortical non-locking screws or unicortical locking screws. The constructs were then potted in cement and tested on a MTS machine using a custom gimble and evaluated for load to failure and axial and rotational stiffness.

Results

There was no significant difference between the constructs in terms of axial stiffness (locking 688.3 ± 306.2 N/mm, non-locking 674.5 ± 613.0 N/mm; p = 0.77) or load to failure (locking 720.1 ± 232.0 N, non-locking 664.8 ± 167.5 N; p = 0.46). However, rotational stiffness varied significantly (locking 1.70 ± 0.91 N-m/mm, non-locking 2.49 ± 0.78 N-m/mm, p = 0.049) with bicortical non-locking constructs exhibiting higher torque values.

Conclusions

Unicortical fixation using pre-contoured plates and locking screws has a similar biomechanical profile compared to gold standard non-locked bicortical screws in cyclic axial compression and axial load to failure. Non-locking constructs were stiffer under rotational testing. This technique may provide a suitable biomechanical environment for bony healing. This may also improve the safety of clavicle plating by protecting infraclavicular structures from injury during drilling or screw penetration as it obviates the need for bicortical fixation.
Literatur
1.
Zurück zum Zitat Zlowodzki M, Zelle BA, Cole PA et al (2005) Treatment of acute midshaft clavicle fractures: systematic review of 2144 fractures: on behalf of the Evidence-Based Orthopaedic Trauma Working Group. J Orthop Trauma 19(7):504–507PubMedCrossRef Zlowodzki M, Zelle BA, Cole PA et al (2005) Treatment of acute midshaft clavicle fractures: systematic review of 2144 fractures: on behalf of the Evidence-Based Orthopaedic Trauma Working Group. J Orthop Trauma 19(7):504–507PubMedCrossRef
2.
Zurück zum Zitat Celestre P, Roberston C, Mahar A et al (2008) Biomechanical evaluation of clavicle fracture plating techniques: does a locking plate provide improved stability? J Orthop Trauma 22(4):241–247PubMedCrossRef Celestre P, Roberston C, Mahar A et al (2008) Biomechanical evaluation of clavicle fracture plating techniques: does a locking plate provide improved stability? J Orthop Trauma 22(4):241–247PubMedCrossRef
3.
Zurück zum Zitat Nordqvist A, Petersson C (1994) The incidence of fractures of the clavicle. Clin Orthop Relat Res 300:127–132PubMed Nordqvist A, Petersson C (1994) The incidence of fractures of the clavicle. Clin Orthop Relat Res 300:127–132PubMed
4.
Zurück zum Zitat Bernstein J (2007) Nonoperative treatment compared with plate fixation of displaced midshaft clavicular fractures. J Bone Joint Surg Am 89(8):1866–1867PubMed Bernstein J (2007) Nonoperative treatment compared with plate fixation of displaced midshaft clavicular fractures. J Bone Joint Surg Am 89(8):1866–1867PubMed
5.
Zurück zum Zitat Hill JM, McGuire MH, Crosby LA (1997) Closed treatment of displaced middle-third fractures of the clavicle gives poor results. J Bone Joint Surg Br 79(4):537–539PubMedCrossRef Hill JM, McGuire MH, Crosby LA (1997) Closed treatment of displaced middle-third fractures of the clavicle gives poor results. J Bone Joint Surg Br 79(4):537–539PubMedCrossRef
6.
Zurück zum Zitat Nordqvist A, Petersson CJ, Redlund-Johnell I (1998) Mid-clavicle fractures in adults: end result study after conservative treatment. J Orthop Trauma 12(8):572–576PubMedCrossRef Nordqvist A, Petersson CJ, Redlund-Johnell I (1998) Mid-clavicle fractures in adults: end result study after conservative treatment. J Orthop Trauma 12(8):572–576PubMedCrossRef
7.
Zurück zum Zitat Nowak J, Holgersson M, Larsson S (2005) Sequelae from clavicular fractures are common: a prospective study of 222 patients. Acta Orthop 76(4):496–502PubMedCrossRef Nowak J, Holgersson M, Larsson S (2005) Sequelae from clavicular fractures are common: a prospective study of 222 patients. Acta Orthop 76(4):496–502PubMedCrossRef
8.
Zurück zum Zitat Johnson B, Thursby P (1996) Subclavian artery injury caused by a screw in a clavicular compression plate. Cardiovasc Surg 4(3):414–415PubMedCrossRef Johnson B, Thursby P (1996) Subclavian artery injury caused by a screw in a clavicular compression plate. Cardiovasc Surg 4(3):414–415PubMedCrossRef
9.
Zurück zum Zitat Shackford SR, Connolly JF (2003) Taming of the screw: a case report and literature review of limb-threatening complications after plate osteosynthesis of a clavicular nonunion. J Trauma 55(5):840–843PubMedCrossRef Shackford SR, Connolly JF (2003) Taming of the screw: a case report and literature review of limb-threatening complications after plate osteosynthesis of a clavicular nonunion. J Trauma 55(5):840–843PubMedCrossRef
10.
Zurück zum Zitat Regel JP, Pospiech J, Aalders TA et al (2002) Intraspinal migration of a Kirschner wire 3 months after clavicular fracture fixation. Neurosurg Rev 25(1):110–112PubMedCrossRef Regel JP, Pospiech J, Aalders TA et al (2002) Intraspinal migration of a Kirschner wire 3 months after clavicular fracture fixation. Neurosurg Rev 25(1):110–112PubMedCrossRef
11.
Zurück zum Zitat Azimi M, Adipepe F, Haykal S et al (1990) Recurrent pneumothorax after osteosynthesis of the clavicle. Apropos of a case. J Chir (Paris) 127(6):365 Azimi M, Adipepe F, Haykal S et al (1990) Recurrent pneumothorax after osteosynthesis of the clavicle. Apropos of a case. J Chir (Paris) 127(6):365
12.
Zurück zum Zitat Marchi E, Reis MP, Carvalho MV (2008) Transmediastinal migration of Kirschner wire. Interact Cardiovasc Thorac Surg 7(5):869–870PubMedCrossRef Marchi E, Reis MP, Carvalho MV (2008) Transmediastinal migration of Kirschner wire. Interact Cardiovasc Thorac Surg 7(5):869–870PubMedCrossRef
13.
Zurück zum Zitat Iannotti MR, Crosby LA, Stafford P et al (2002) Effects of plate location and selection on the stability of midshaft clavicle osteotomies: a biomechanical study. J Shoulder Elbow Surg 11(5):457–462PubMedCrossRef Iannotti MR, Crosby LA, Stafford P et al (2002) Effects of plate location and selection on the stability of midshaft clavicle osteotomies: a biomechanical study. J Shoulder Elbow Surg 11(5):457–462PubMedCrossRef
14.
Zurück zum Zitat Kemper A, Stitzel J, Gabler C et al (2006) Biomechanical response of the human clavicle subjected to dynamic bending. Biomed Sci Instrum 42:231–236PubMed Kemper A, Stitzel J, Gabler C et al (2006) Biomechanical response of the human clavicle subjected to dynamic bending. Biomed Sci Instrum 42:231–236PubMed
15.
Zurück zum Zitat Bostman O, Manninen M, Pihlajamaki H (1997) Complications of plate fixation in fresh displaced midclavicular fractures. J Trauma 43(5):778–783PubMedCrossRef Bostman O, Manninen M, Pihlajamaki H (1997) Complications of plate fixation in fresh displaced midclavicular fractures. J Trauma 43(5):778–783PubMedCrossRef
16.
Zurück zum Zitat Der Tavitian J, Davison JN, Dias JJ (2002) Clavicular fracture non-union surgical outcome and complications. Injury 33(2):135–143PubMedCrossRef Der Tavitian J, Davison JN, Dias JJ (2002) Clavicular fracture non-union surgical outcome and complications. Injury 33(2):135–143PubMedCrossRef
17.
Zurück zum Zitat Roberts JW, Grindel SI, Rebholz B et al (2007) Biomechanical evaluation of locking plate radial shaft fixation: unicortical locking fixation versus mixed bicortical and unicortical fixation in a sawbone model. J Hand Surg (Am) 32(7):971–975CrossRef Roberts JW, Grindel SI, Rebholz B et al (2007) Biomechanical evaluation of locking plate radial shaft fixation: unicortical locking fixation versus mixed bicortical and unicortical fixation in a sawbone model. J Hand Surg (Am) 32(7):971–975CrossRef
18.
Zurück zum Zitat Gardner MJ, Brophy RH, Campbell D et al (2005) The mechanical behavior of locking compression plates compared with dynamic compression plates in a cadaver radius model. J Orthop Trauma 19(9):597–603PubMedCrossRef Gardner MJ, Brophy RH, Campbell D et al (2005) The mechanical behavior of locking compression plates compared with dynamic compression plates in a cadaver radius model. J Orthop Trauma 19(9):597–603PubMedCrossRef
19.
Zurück zum Zitat Wu CC, Shih CH, Chen WJ et al (1998) Treatment of clavicular aseptic nonunion: comparison of plating and intramedullary nailing techniques. J Trauma 45(3):512–516PubMedCrossRef Wu CC, Shih CH, Chen WJ et al (1998) Treatment of clavicular aseptic nonunion: comparison of plating and intramedullary nailing techniques. J Trauma 45(3):512–516PubMedCrossRef
20.
Zurück zum Zitat Lee YS, Lin CC, Huang CR et al (2007) Operative treatment of midclavicular fractures in 62 elderly patients: knowles pin versus plate. Orthopedics 30(11):959–964PubMed Lee YS, Lin CC, Huang CR et al (2007) Operative treatment of midclavicular fractures in 62 elderly patients: knowles pin versus plate. Orthopedics 30(11):959–964PubMed
21.
Zurück zum Zitat Khalil A (2009) Intramedullary screw fixation for midshaft fractures of the clavicle. Int Orthop 33(5):1421–1424PubMedCrossRef Khalil A (2009) Intramedullary screw fixation for midshaft fractures of the clavicle. Int Orthop 33(5):1421–1424PubMedCrossRef
22.
Zurück zum Zitat Mueller M, Rangger C, Striepens N et al (2008) Minimally invasive intramedullary nailing of midshaft clavicular fractures using titanium elastic nails. J Trauma 64(6):1528–1534PubMedCrossRef Mueller M, Rangger C, Striepens N et al (2008) Minimally invasive intramedullary nailing of midshaft clavicular fractures using titanium elastic nails. J Trauma 64(6):1528–1534PubMedCrossRef
23.
Zurück zum Zitat Ruedi TP, Buckley RE, Moran CG (2007) AO principles of fracture management, vol 1, 2nd edn. Thieme, AO publishing, AO Foundation, Switzerland, pp 266–268 Ruedi TP, Buckley RE, Moran CG (2007) AO principles of fracture management, vol 1, 2nd edn. Thieme, AO publishing, AO Foundation, Switzerland, pp 266–268
24.
Zurück zum Zitat Harnroongroj T, Vanadurongwan V (1996) Biomechanical aspects of plating osteosynthesis of transverse clavicular fracture with and without inferior cortical defect. Clin Biomech 11(5):290–294CrossRef Harnroongroj T, Vanadurongwan V (1996) Biomechanical aspects of plating osteosynthesis of transverse clavicular fracture with and without inferior cortical defect. Clin Biomech 11(5):290–294CrossRef
25.
Zurück zum Zitat Robertson C, Celestre P, Mahar A et al (2009) Reconstruction plates for stabilization of mid-shaft clavicle fractures: differences between nonlocked and locked plates in two different positions. J Shoulder Elbow Surg 18(2):204–209PubMedCrossRef Robertson C, Celestre P, Mahar A et al (2009) Reconstruction plates for stabilization of mid-shaft clavicle fractures: differences between nonlocked and locked plates in two different positions. J Shoulder Elbow Surg 18(2):204–209PubMedCrossRef
Metadaten
Titel
Biomechanical analysis of bicortical versus unicortical locked plating of mid-clavicular fractures
verfasst von
Daniel Hamman
Derek Lindsey
Jason Dragoo
Publikationsdatum
01.06.2011
Verlag
Springer-Verlag
Erschienen in
Archives of Orthopaedic and Trauma Surgery / Ausgabe 6/2011
Print ISSN: 0936-8051
Elektronische ISSN: 1434-3916
DOI
https://doi.org/10.1007/s00402-010-1212-2

Weitere Artikel der Ausgabe 6/2011

Archives of Orthopaedic and Trauma Surgery 6/2011 Zur Ausgabe

Arthropedia

Grundlagenwissen der Arthroskopie und Gelenkchirurgie. Erweitert durch Fallbeispiele, Videos und Abbildungen. 
» Jetzt entdecken

Knie-TEP: Kein Vorteil durch antibiotikahaltigen Knochenzement

29.05.2024 Periprothetische Infektionen Nachrichten

Zur Zementierung einer Knie-TEP wird in Deutschland zu über 98% Knochenzement verwendet, der mit einem Antibiotikum beladen ist. Ob er wirklich besser ist als Zement ohne Antibiotikum, kann laut Registerdaten bezweifelt werden.

Häusliche Gewalt in der orthopädischen Notaufnahme oft nicht erkannt

28.05.2024 Häusliche Gewalt Nachrichten

In der Notaufnahme wird die Chance, Opfer von häuslicher Gewalt zu identifizieren, von Orthopäden und Orthopädinnen offenbar zu wenig genutzt. Darauf deuten die Ergebnisse einer Fragebogenstudie an der Sahlgrenska-Universität in Schweden hin.

Fehlerkultur in der Medizin – Offenheit zählt!

28.05.2024 Fehlerkultur Podcast

Darüber reden und aus Fehlern lernen, sollte das Motto in der Medizin lauten. Und zwar nicht nur im Sinne der Patientensicherheit. Eine negative Fehlerkultur kann auch die Behandelnden ernsthaft krank machen, warnt Prof. Dr. Reinhard Strametz. Ein Plädoyer und ein Leitfaden für den offenen Umgang mit kritischen Ereignissen in Medizin und Pflege.

Mehr Frauen im OP – weniger postoperative Komplikationen

21.05.2024 Allgemeine Chirurgie Nachrichten

Ein Frauenanteil von mindestens einem Drittel im ärztlichen Op.-Team war in einer großen retrospektiven Studie aus Kanada mit einer signifikanten Reduktion der postoperativen Morbidität assoziiert.

Update Orthopädie und Unfallchirurgie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.