Design and implementation of a “cheese” phantom-based Tomotherapy TLD dose intercomparison

verfasst von: Dr. Hans Schiefer, PhD, Konrad Buchauer, PhD, Simon Heinze, MS, Dr. Guido Henke, Prof. Dr. Ludwig Plasswilm

Erschienen in: Strahlentherapie und Onkologie | Ausgabe 11/2015

Einloggen, um Zugang zu erhalten

Abstract

Background

The unique beam-delivery technique of Tomotherapy machines (Accuray Inc., Sunnyvale, Calif.) necessitates tailored quality assurance. This requirement also applies to external dose intercomparisons. Therefore, the aim of the 2014 SSRMP (Swiss Society of Radiobiology and Medical Physics) dosimetry intercomparison was to compare two set-ups with different phantoms.

Materials and methods

A small cylindrical Perspex phantom, which is similar to the IROC phantom (Imaging and Radiation Oncology Core, Houston, Tex.), and the “cheese” phantom, which is provided by the Tomotherapy manufacturer to all institutions, were used. The standard calibration plans for the TomoHelical and TomoDirect irradiation techniques were applied. These plans are routinely used for dose output calibration in Tomotherapy institutions. We tested 20 Tomotherapy machines in Germany and Switzerland. The ratio of the measured (Dm) to the calculated (Dc) dose was assessed for both phantoms and irradiation techniques. The Dm/Dc distributions were determined to compare the suitability of the measurement set-ups investigated.

Results

The standard deviations of the TLD-measured (thermoluminescent dosimetry) Dm/Dc ratios for the “cheese” phantom were 1.9 % for the TomoHelical (19 measurements) and 1.2 % (11 measurements) for the TomoDirect irradiation techniques. The corresponding ratios for the Perspex phantom were 2.8 % (18 measurements) and 1.8 % (11 measurements).

Conclusion

Compared with the Perspex phantom-based set-up, the “cheese” phantom-based set-up without individual planning was demonstrated to be more suitable for Tomotherapy dose checks. Future SSRMP dosimetry intercomparisons for Tomotherapy machines will therefore be based on the “cheese” phantom set-up.
Literatur
1.
Zurück zum Zitat AAPM (2004) Quality assurance for clinical trials: a primer for physicists. Report No. 86. Medical Physics Publishing, Madison AAPM (2004) Quality assurance for clinical trials: a primer for physicists. Report No. 86. Medical Physics Publishing, Madison
2.
Zurück zum Zitat Ahnen L, Schiefer H (2013) Results of the TLD intercomparison 2013. SSRPM Bull 78:18–19 Ahnen L, Schiefer H (2013) Results of the TLD intercomparison 2013. SSRPM Bull 78:18–19
3.
Zurück zum Zitat Alfonso R, Andreo P, Capote R et al (2008) A new formalism for reference dosimetry of small and nonstandard fields. Med Phys 35:5179–5186CrossRefPubMed Alfonso R, Andreo P, Capote R et al (2008) A new formalism for reference dosimetry of small and nonstandard fields. Med Phys 35:5179–5186CrossRefPubMed
4.
Zurück zum Zitat Almond PR, Biggs PJ, Coursey BM et al (1999) AAPM’s TG-51 protocol for clinical reference dosimetry of high-energy photon and electron beams. Med Phys 26:1847–1870CrossRefPubMed Almond PR, Biggs PJ, Coursey BM et al (1999) AAPM’s TG-51 protocol for clinical reference dosimetry of high-energy photon and electron beams. Med Phys 26:1847–1870CrossRefPubMed
5.
Zurück zum Zitat Andreo P, Burns DT, Hohlfeld K et al (2000) Absorbed dose determination in external beam radiotherapy. IAEA Technical Report Series No. 398. International Atomic Energy Agency, Vienna Andreo P, Burns DT, Hohlfeld K et al (2000) Absorbed dose determination in external beam radiotherapy. IAEA Technical Report Series No. 398. International Atomic Energy Agency, Vienna
6.
Zurück zum Zitat Duane S, Nicholas D, Palmans H et al (2006) SU-FF-T-195: dosimetry audit for tomotherapy using alanine/EPR. Med Phys 33:2093CrossRef Duane S, Nicholas D, Palmans H et al (2006) SU-FF-T-195: dosimetry audit for tomotherapy using alanine/EPR. Med Phys 33:2093CrossRef
7.
Zurück zum Zitat Fenwick JD, Tomé WA, Jaradat HA et al (2004) Quality assurance of a helical tomotherapy machine. Phys Med Biol 49:2933CrossRefPubMed Fenwick JD, Tomé WA, Jaradat HA et al (2004) Quality assurance of a helical tomotherapy machine. Phys Med Biol 49:2933CrossRefPubMed
8.
Zurück zum Zitat Ferreira IH, Dutreix A, Bridier A et al (1999) The ESTRO-QUALity assurance network (EQUAL). Radiother Oncol 55:273–284CrossRef Ferreira IH, Dutreix A, Bridier A et al (1999) The ESTRO-QUALity assurance network (EQUAL). Radiother Oncol 55:273–284CrossRef
9.
Zurück zum Zitat Followill D, Molineu A, Lowenstein J et al (2008) SU-GG-T-213: quality audits of the calibration for TG-51 non-compliant beams by the Radiological Physics Center. Med Phys 35:2774CrossRef Followill D, Molineu A, Lowenstein J et al (2008) SU-GG-T-213: quality audits of the calibration for TG-51 non-compliant beams by the Radiological Physics Center. Med Phys 35:2774CrossRef
10.
Zurück zum Zitat Gago-Arias A, Rodríguez-Romero R, Sánchez-Rubio P et al (2012) Correction factors for A1SL ionization chamber dosimetry in TomoTherapy: machine-specific, plan-class, and clinical fields. Med Phys 39:1964–1970CrossRefPubMed Gago-Arias A, Rodríguez-Romero R, Sánchez-Rubio P et al (2012) Correction factors for A1SL ionization chamber dosimetry in TomoTherapy: machine-specific, plan-class, and clinical fields. Med Phys 39:1964–1970CrossRefPubMed
11.
Zurück zum Zitat Geier M, Astner ST, Duma MN et al (2012) Dose-escalated simultaneous integrated-boost treatment of prostate cancer patients via helical tomotherapy. Strahlenther Onkol 188:410–416CrossRefPubMed Geier M, Astner ST, Duma MN et al (2012) Dose-escalated simultaneous integrated-boost treatment of prostate cancer patients via helical tomotherapy. Strahlenther Onkol 188:410–416CrossRefPubMed
12.
Zurück zum Zitat IAEA (2000) IAEA TRS 398, Absorbed dose determination in external beam radiotherapy: An international code of practice for dosimetry based on standards of absorbed dose to water. International Atomic Energy Agency, Vienna IAEA (2000) IAEA TRS 398, Absorbed dose determination in external beam radiotherapy: An international code of practice for dosimetry based on standards of absorbed dose to water. International Atomic Energy Agency, Vienna
13.
Zurück zum Zitat Ibbott GS (2010) QA in radiation therapy: the RPC perspective. J Phys Conf Ser 250:012001CrossRef Ibbott GS (2010) QA in radiation therapy: the RPC perspective. J Phys Conf Ser 250:012001CrossRef
14.
Zurück zum Zitat Ibbott GS, Followill DS, Molineu HA et al (2008) Challenges in credentialing institutions and participants in advanced technology multi-institutional clinical trials. Int J Radiat Oncol Biol Phys 71:71–75CrossRef Ibbott GS, Followill DS, Molineu HA et al (2008) Challenges in credentialing institutions and participants in advanced technology multi-institutional clinical trials. Int J Radiat Oncol Biol Phys 71:71–75CrossRef
15.
Zurück zum Zitat Langen KM, Papanikolaou N, Balog J et al (2010) QA for helical tomotherapy: report of the AAPM Task Group 148. Med Phys 37:4817–4853CrossRefPubMed Langen KM, Papanikolaou N, Balog J et al (2010) QA for helical tomotherapy: report of the AAPM Task Group 148. Med Phys 37:4817–4853CrossRefPubMed
17.
Zurück zum Zitat Mackie TR, Holmes T, Swerdloff S et al (1993) Tomotherapy: a new concept for the delivery of dynamic conformal radiotherapy. Med Phys 20:1709–1719CrossRefPubMed Mackie TR, Holmes T, Swerdloff S et al (1993) Tomotherapy: a new concept for the delivery of dynamic conformal radiotherapy. Med Phys 20:1709–1719CrossRefPubMed
18.
Zurück zum Zitat NAR (2008) Dosismessverfahren nach der Sondenmethode für Photonen- und Elektronenstrahlung—Teil 2: Dosimetrie hochenergetischer Photonen- und Elektronenstrahlung mit Ionisationskammern. DIN Norm NAR (2008) Dosismessverfahren nach der Sondenmethode für Photonen- und Elektronenstrahlung—Teil 2: Dosimetrie hochenergetischer Photonen- und Elektronenstrahlung mit Ionisationskammern. DIN Norm
19.
Zurück zum Zitat Nguyen NP, Krafft SP, Vos P et al (2011) Feasibility of tomotherapy for Graves’ ophthalmopathy: dosimetry comparison with conventional radiotherapy. Strahlenther Onkol 187:568–574CrossRefPubMed Nguyen NP, Krafft SP, Vos P et al (2011) Feasibility of tomotherapy for Graves’ ophthalmopathy: dosimetry comparison with conventional radiotherapy. Strahlenther Onkol 187:568–574CrossRefPubMed
20.
Zurück zum Zitat Schiefer H, Fogliata A, Nicolini G et al (2010) The Swiss IMRT dosimetry intercomparison using a thorax phantom. Med Phys 37:4424–4431CrossRefPubMed Schiefer H, Fogliata A, Nicolini G et al (2010) The Swiss IMRT dosimetry intercomparison using a thorax phantom. Med Phys 37:4424–4431CrossRefPubMed
21.
Zurück zum Zitat Sedlmayer F, Sautter-Bihl ML, Budach W et al (2013) Is the simultaneously integrated boost (SIB) technique for early breast cancer ready to be adopted for routine adjuvant radiotherapy? Statement of the German and the Austrian Societies of Radiooncology (DEGRO/OGRO). Strahlenther Onkol 189:193–196CrossRefPubMed Sedlmayer F, Sautter-Bihl ML, Budach W et al (2013) Is the simultaneously integrated boost (SIB) technique for early breast cancer ready to be adopted for routine adjuvant radiotherapy? Statement of the German and the Austrian Societies of Radiooncology (DEGRO/OGRO). Strahlenther Onkol 189:193–196CrossRefPubMed
22.
Zurück zum Zitat SSRMP (2000) High-Energy Photon Beam Therapy Dosimetry with Ionisation Chambers. SSRMP Recommendations 8 SSRMP (2000) High-Energy Photon Beam Therapy Dosimetry with Ionisation Chambers. SSRMP Recommendations 8
23.
Zurück zum Zitat Sterzing F, Welzel T, Sroka-Perez G et al (2009) Reirradiation of multiple brain metastases with helical tomotherapy. A multifocal simultaneous integrated boost for eight or more lesions. Strahlenther Onkol 185:89–93CrossRefPubMed Sterzing F, Welzel T, Sroka-Perez G et al (2009) Reirradiation of multiple brain metastases with helical tomotherapy. A multifocal simultaneous integrated boost for eight or more lesions. Strahlenther Onkol 185:89–93CrossRefPubMed
24.
Zurück zum Zitat Swinnen A (2005) Quality assurance in radiotherapy: Development and validation of a mailed dosimetry procedure for external audits using a multipurpose phantom and in vivo dosimetry. PhD Thesis, Katholieke Universiteit Leuven, Belgium Swinnen A (2005) Quality assurance in radiotherapy: Development and validation of a mailed dosimetry procedure for external audits using a multipurpose phantom and in vivo dosimetry. PhD Thesis, Katholieke Universiteit Leuven, Belgium
25.
Zurück zum Zitat Swinnen A, Verstraete J, Huyskens DP (2002) The use of a multipurpose phantom for mailed dosimetry checks of therapeutic photon beams: ‘OPERA’ (operational phantom for external radiotherapy audit). Radiother Oncol 64:317–326CrossRefPubMed Swinnen A, Verstraete J, Huyskens DP (2002) The use of a multipurpose phantom for mailed dosimetry checks of therapeutic photon beams: ‘OPERA’ (operational phantom for external radiotherapy audit). Radiother Oncol 64:317–326CrossRefPubMed
26.
Zurück zum Zitat Thomas SD, Mackenzie M, Rogers DWO et al (2005) A Monte Carlo derived TG-51 equivalent calibration for helical tomotherapy. Medical Phys 32:1346–1353CrossRef Thomas SD, Mackenzie M, Rogers DWO et al (2005) A Monte Carlo derived TG-51 equivalent calibration for helical tomotherapy. Medical Phys 32:1346–1353CrossRef
27.
Zurück zum Zitat Thomas SJ, Aspradakis MM, Byrne JP et al (2014) Reference dosimetry on TomoTherapy: an addendum to the 1990 UK MV dosimetry code of practice. Phys Med Biol 59:1339–1352CrossRefPubMed Thomas SJ, Aspradakis MM, Byrne JP et al (2014) Reference dosimetry on TomoTherapy: an addendum to the 1990 UK MV dosimetry code of practice. Phys Med Biol 59:1339–1352CrossRefPubMed
28.
Zurück zum Zitat Vallet V, Pisaturo O, Pachoud M et al (2010) 18 Months experience of Filmless patient’s Dqa on Tomotherapy units. Strahlenther Onkol 186:744 Vallet V, Pisaturo O, Pachoud M et al (2010) 18 Months experience of Filmless patient’s Dqa on Tomotherapy units. Strahlenther Onkol 186:744
Metadaten
Titel
Design and implementation of a “cheese” phantom-based Tomotherapy TLD dose intercomparison
verfasst von
Dr. Hans Schiefer, PhD
Konrad Buchauer, PhD
Simon Heinze, MS
Dr. Guido Henke
Prof. Dr. Ludwig Plasswilm
Publikationsdatum
01.11.2015
Verlag
Springer Berlin Heidelberg
Erschienen in
Strahlentherapie und Onkologie / Ausgabe 11/2015
Print ISSN: 0179-7158
Elektronische ISSN: 1439-099X
DOI
https://doi.org/10.1007/s00066-015-0850-2

Weitere Artikel der Ausgabe 11/2015

Strahlentherapie und Onkologie 11/2015 Zur Ausgabe

Update Onkologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.