Open Access 14.09.2023 | Original Article

Effect of simultaneous integrated boost concepts on photoneutron and distant out-of-field doses in VMAT for prostate cancer

verfasst von: Benjamin Gauter-Fleckenstein, M.D., Sebastian Schönig, Lena Mertens, Hans Oppitz, Kerstin Siebenlist, Michael Ehmann, Jens Fleckenstein

Erschienen in: Strahlentherapie und Onkologie | Ausgabe 3/2024

Abstract

Background

A simultaneous integrated boost (SIB) may result in increased out-of-field (DOOF) and photoneutron (HPN) doses in volumetric modulated arc therapy (VMAT) for prostate cancer (PCA). This work therefore aimed to compare DOOF and HPN in flattened (FLAT) and flattening filter-free (FFF) 6‑MV and 10-MV VMAT treatment plans with and without SIB.

Methods

Eight groups of 30 VMAT plans for PCA with 6 MV or 10 MV, with or without FF and with uniform (2 Gy) or SIB target dose (2.5/3.0 Gy) prescriptions (CONV, SIB), were generated. All 240 plans were delivered on a slab-phantom and compared with respect to measured DOOF and HPN in 61.8 cm distance from the isocenter. The 6‑ and 10-MV flattened VMAT plans with conventional fractionation (6- and 10-MV FLAT CONV) served as standard reference groups. Doses were analyzed as a function of delivered monitor units (MU) and weighted equivalent square field size Aeq. Pearson’s correlation coefficients between the presented quantities were determined.

Results

The SIB plans resulted in decreased HPN over an entire prostate RT treatment course (10-MV SIB vs. CONV −38.2%). Omission of the flattening filter yielded less HPN (10-MV CONV −17.2%; 10-MV SIB −22.5%). The SIB decreased DOOF likewise by 39% for all given scenarios, while the FFF mode reduced DOOF on average by 60%. A strong Pearson correlation was found between MU and HPN (r > 0.9) as well as DOOF (0.7 < r < 0.9).

Conclusion

For a complete treatment, SIB reduces both photoneutron and OOF doses to almost the same extent as FFF deliveries. It is recommended to apply moderately hypofractionated 6‑MV SIB FFF-VMAT when considering photoneutron or OOF doses.

Introduction

Recently, moderately hypofractionated intensity modulated radiotherapy (IMRT) regimens were assigned as standard of care in interdisciplinary national and international guidelines [1, 2] for prostate cancer treatments. We are currently faced with an increasing number of simultaneous integrated boost (SIB) concepts being used both in clinical trials and in daily routine (clinicaltrials.gov yielded 85 active or recruiting trials for “simultaneous integrated boost”). Evidence for these approaches is drawn from moderate hypofractionation trials for prostate [38]. For these approaches, an SIB concept is often the method of choice. The SIB increases fractional radiation dose in selected regions of the treated volume. This leads to shortening of overall treatment time and results in steeper dose gradients toward more radiosensitive organs and consequently a higher dose conformity [9]. Furthermore, moderately increased fractional doses potentially result in beneficial fractionation effects in tumors with low alpha/beta values.
High-energy photons may, depending on their energy, generate photoneutrons (PN) when interacting with beam-shaping components in the medical linear accelerator (linac) treatment head. The primary PN-generating structures were identified [10] to be the radiation target, primary collimator, flattening filter, multileaf collimator (MLC), and jaws. The generated PN typically possess kinetic energies between 0.2 and 2.5 MeV. To account for the higher radiobiological effectiveness (RBE) of neutron radiation, the physical dose D in Gray (Gy) is converted into the equivalent dose H in Sievert (Sv) by multiplication of D with a weighting factor, which is between 2.5 and 20.7 (ICRP 103, [11]). Due to the increased RBE, PN have undesired biological and physical effects on surrounding healthy tissue but also on implanted electronic devices such as cardiac pacemakers and cardioverter-defibrillators, insulin pumps, or neural stimulators (AAPM158, [12]).
With respect to radiation protection, it is furthermore relevant to note that matter within the treatment beam leads to formation of scattered radiation that, in combination with radiation transmitted through the beam-shaping devices, leads to distant out-of-field dose (DOOF) in the patient. The OOF radiation contributes to potentially unmonitored and unaccounted radiation exposure at a distance from the treated volume, which means there is an increased risk for secondary cancers in organs distant to the primary cancer region [13]. This dose remains unmonitored because not all treatment planning systems (TPS) explicitly account for distant scattered radiation [14] and if they do, CT scans typically do not exceed the adjacent body regions closest to the target volume.
In conventional three-dimensional conformal radiation therapy, the flattening filter (FF) generates a uniform photon fluence within the therapeutic beam. Here, up to 40 × 40 cm2 uniform photon beams are shaped through position-dependent beam attenuation into a conformal radiation field. In intensity modulated radiation therapy (IMRT), several smaller segments of irregular shaped fields and with differing photon fluences are directed from several angles to superimpose to a uniform dose with high conformity to the treatment volume. Modern IMRT can be applied with continuously moving multileaf collimator (MLC) leaves and jaws during continuously changing gantry positions and varying dose rates (e.g., VMAT [volumetric modulated arc therapy]). While a wedge becomes unnecessary in modulated segmented radiation fields, the FF frequently remains in the beam. Removal of the FF allows for high dose rates (15–25 Gy/min) that are beneficial for irradiation of small and potentially moving volumes with tight margins as present in stereotactic body radiotherapy at the cost of a uniform lateral target fluence profile. Additionally, removal of the FF in IMRT reduces matter in the beam, hence reducing the contamination with PN and also scattered radiation at a distance from the treated volume [13, 1517].
All the aforementioned modifications (SIB, FFF) result in varying modulation depths (number of required monitor units [MU] divided by prescribed dose to target) and differing average equivalent square field sizes (Aeq) describing MLC activity. This might influence PN generation and DOOF. For locations distant to the isocenter, no studies exist that describe the effect of a prostate SIB in combination with FFF-VMAT in view of neutron contamination or scattered radiation. This might be significant especially for implanted electronic devices in this distance from radiation volumes such as cardiac implanted electronic devices (CIEDs; [18, 19]). In parallel to the increasing use for SIB concepts, there is currently an ongoing trend toward hypofractionation in general, and for prostate radiation therapy (RT) in particular. We therefore addressed the question of whether this trend is accompanied by an increase in out-of-field doses originating from either scattered photons or photoneutrons. In this context, we investigated the effect of an SIB concept in flattened and flattening-filter free VMAT with 6 and 10 MV for prostate cancer RT on distant neutron- and out-of-field dose in VMAT plans of comparable dose distributions and comparable therapeutic efficacy. The results of this study might influence the decision process in whether RT for prostate carcinoma should be moderately hypofractionated, especially in CIED-bearing individuals, or not.

Methods

Treatment sequence generation

Based on the anonymized treatment planning CT data of 30 patients previously treated for prostate cancer, volumetric modulated arc therapy treatment plans (all 360° dual arcs) were generated in the TPS (Monaco 5.51, Elekta AB, Stockholm, Sweden). Treatment plans for two different dose prescription approaches were created:
  • A uniform dose-to-target concept (CONV): planning target volume (PTV) was prostate and seminal vesicles plus 10 mm isotropic margin (PTV); 37 fractions (fx) with fractional prescription dose D50%(PTV) = 2.0 Gy uniformly to entire PTV with a total dose prescription of 74 Gy.
  • A simultaneously integrated prostate-boost concept (SIB) in analogy to CHHiP protocol [3]: same PTV as in CONV but with a SIB to the prostate only; 20 fx with fractional prescription dose D50%(PTV) = 2.5 Gy and D50%(prostate) = 3.0 Gy and a total dose of 50 Gy (PTV) and 60 Gy (SIB).
In total, eight different clinically acceptable and dosimetrically similar treatment plans for all permutations of the following parameters were generated (Table 1):
  • CONV or SIB dose prescription to the target volume(s)
  • Nominal acceleration potential (6 MV or 10 MV)
  • With or without flattening filter (FLAT, FFF)
Table 1
Volumetric modulated arc therapy treatment plansa (all 360° dual arcs) for two different dose prescription approaches generated in a 2 × 2 × 2 factorial design: SIB 60 Gy vs. CONV 74 Gy, FLAT vs. FFF, 6 MV vs. 10 MV
Treatment sequence generation
Energy
Filter position: FLAT
Filter position: FFF
Dose concept: CONV
Dose concept: SIB
Dose concept: CONV
Dose concept: SIB
6 MV
6‑MV FLAT CONV (n = 30)
6‑MV FLAT SIB (n = 30)
6‑MV FFF CONV (n = 30)
6‑MV FFF SIB (n = 30)
10 MV
10-MV FLAT CONV (n = 30)
10-MV FLAT SIB (n = 30)
10-MV FFF CONV (n = 30)
10-MV FLAT SIB (n = 30)
aBased on anonymized treatment planning CT data of 30 patients previously treated for prostate cancer
The resulting total number of MU per treatment and the weighted equivalent square field size (Aeq) as a surrogate for MLC activity of the generated treatment plans are presented in Table 2.
Table 2
Prescription doses per fraction (fx) and whole treatment (tx), monitor units (MU), and equivalent square field sizes (Aeq) per treatment as well as average neutron-equivalent (HPN) and out-of-field doses (DOOF) per MU and treatment fxa for all scenarios
Nom. acc. potential
Delivery type
PTV coverage
6 MV
10 MV
FLAT
FFF
FLAT
FFF
CONV
SIB
CONV
SIB
CONV
SIB
CONV
SIB
Prescription dose, fx (Gy)
2
2.5/3
2
2.5/3
2
2.5/3
2
2.5/3
Total dose, tx (Gy)
74
50/60
74
50/60
74
50/60
74
50/60
Monitor units (MU)
839 ± 116
1006 ± 205
962 ± 124
1098 ± 265
717 ± 84
832 ± 137
968 ± 188
1063 ± 209
Aeq (cm)
3.0 ± 0.3
2.9 ± 0.5
2.8 ± 0.3
2.9 ± 0.5
3.0 ± 0.3
2.9 ± 0.5
2.7 ± 0.4
2.7 ± 0.5
HPN/MU (nSv)
14.9 ± 0.8
15.0 ± 0.6
26.5 ± 0.6
27.0 ± 0.8
494.4 ± 11.1
503.3 ± 13.4
303.1 ± 13.1
305.0 ± 6.1
HPN/fx (µSv)
12.5 ± 1.8
15.1 ± 3.1
25.5 ± 3.5
29.7 ± 7.6
354.7 ± 44.2
419.3 ± 72.5
293.8 ± 60.7
324.8 ± 67.3
HPN/tx (mSv)
0.5 ± 0.1
0.3 ± 0.1
0.9 ± 0.1
0.6 ± 0.2
13.1 ± 1.5
8.4 ± 1.5
10.9 ± 2.1
6.5 ± 1.4
DOOF/MU (µGy)
1.7 ± 0.1
1.6 ± 0.1
0.7 ± 0.1
0.7 ± 0.1
2.3 ± 0.3
2.3 ± 0.2
0.7 ± 0.1
0.7 ± 0.1
DOOF/fx (mGy)
1.6 ± 0.3
1.9 ± 0.5
0.7 ± 0.1
0.7 ± 0.2
1.8 ± 0.4
2.0 ± 0.4
0.7 ± 0.2
0.8 ± 0.2
DOOF/tx (mGy)
59.2 ± 11.1
38.0 ± 10.0
25.9 ± 3.7
14.0 ± 4.0
66.6 ± 14.8
40.0 ± 8.0
25.9 ± 7.4
16.0 ± 4.0
r (MU, HPN)
0.93
0.98
0.99
0.99
0.99
0.99
0.98
1
r (MU, DOOF)
0.84
0.92
0.68
0.73
0.75
0.90
0.78
0.76
r (Aeq, HPN)
−0.75
−0.71
−0.74
−0.71
−0.55
−0.73
−0.83
−0.72
r (Aeq, DOOF)
−0.53
−0.66
−0.31
−0.53
−0.17
−0.65
−0.51
−0.49
aMean values and standard deviations, r Pearson’s correlation coefficient
Aeq is defined as
$$A_{eq}={\sum }_{i=1}^{n}\omega _{i}A_{i{,}eq}$$
where ωi is the fraction of MUs of the i‑th control point and Ai,eq is the corresponding control point equivalent square field size. Ai,eq is obtained by the ratio of control point area over control point perimeter.

Measurement set-up and data analysis

Treatment sequences were delivered on a linac (Versa HD, Elekta AB, Stockholm, Sweden) to an anthropomorphous slab phantom arrangement, consisting of three 30 × 30 × 10 cm3 stacks of water equivalent phantom (RW3, PTW, Freiburg, Germany) and a neutron detector. The beam isocenter was located in the “pelvic area” at the center of the lowest RW3 stack. The center positions of each detector and the isocenter were 61.8 cm away from each other. See measurement set-up in Fig. 1.
For each VMAT sequence, PN-dose HPN and out-of-field-dose DOOF measurements were performed simultaneously. Out-of-field-dose, comprising scattered and beam-limiting device-transmitted radiation, was measured using an ionization chamber (0.125 cm3 semiflex, 31010, PTW, Freiburg, Germany) with photon energy response uncertainty above 50 kV of less than ±5% (useful range 140 kV and above radiation quality), connected to a Unidos Webline electrometer (PTW, Freiburg, Germany). The ionization chamber was placed at the center of the uppermost slab of the slab phantom on the opposite side of the neutron detector after calibration to ambient temperature and pressure. Since one slab of the RW3 phantom has a thickness of 1 cm and no bolus material was used, the depth of the ionization chamber was located at a depth of 0.5 cm below the surface. This represents a typical implantation depth for CIEDs in patients.
The wide energy neutron detection instrument (WENDI‑2 FHT 762 with FH 40 G survey meter, Thermo Fisher Scientific, Erlangen, Germany) is a He3-filled proportional counter with a thermal to 5 GeV energy range according to H*(10), ICRP 74 [20] and a measuring range from 0.01 μSv/h to 100 mSv/h. Its reported sensitivity is 0.84 cps/μSv/h Cf-252 and the device has proven to produce reliable dose readings [21]. Measurement mode was set to “neutron dose only”. Internal software-based calibration procedures according to the manufacturer’s specifications including detection of contaminating X‑ray scattered radiation were respected.
The resulting DOOF and HPN were plotted against two characteristic delivery sequence-defining parameters MU and Aeq. Subsequently, absolute values for DOOF in Gray and HPN in Sievert were put in relation to the delivered MUs (DOOF/MU, HPN/MU) to compare values between different dose levels with and without SIB. Measurements were performed once for all 30 plans. To determine the measurement reproducibility, measurements for three representative treatment plans were repeated three times for all eight treatment scenarios. Standard deviation is reported as percent of the corresponding mean value.
The values reported in Table 2 were generated by first determining the relevant quantity (e.g., HPN/MU) for each treatment sequence and subsequently generating the mean and standard deviations of the 30 treatment sequences that belong to one treatment approach.
For a comparison of total PN or out-of-field dose resulting from an entire RT course for prostate cancer with the investigated VMAT plans, summed absolute values for all treatment fractions for CONV (total dose 74 Gy; 37 fx) and SIB (total dose 50/60 Gy; 20 fx) were compared. The 6‑ and 10-MV flattened VMAT plans with conventional fractionation (6- and 10-MV FLAT CONV) served as standard reference groups. Using SAS software (release 9.4, SAS Institute Inc., Cary, NC, USA), a linear mixed model was generated to control interaction effects in our 2 × 2 × 2 model. After investigating single effects of photon energy level (6 and 10 MV), position of the flattening filter (FLAT, FFF), and dose prescription (SIB, CONV), a three-way ANOVA was used to investigate the difference in dose reductions between the most promising plan setting in comparison to the standard reference groups. Fixed factors were FF, photon energy, and homogeneity. Random factor was the plan ID. For comparisons between reference plan settings and plan settings with least HPN and DOOF, paired t tests were used. In general, p < 0.05 was considered significant. The F values describe effect sizes.
Finally, Pearson’s correlation coefficients between DOOF and HPN and the aforementioned variables MU and Aeq were generated. Correlation between parameters was considered strong for r values between 1 and 0.7, moderate for r values between 0.7 and 0.3, and weak for r values between 0.3 and 0.

Results

The resulting neutron-equivalent doses HPN per delivered treatment fraction as a function of either MUs per treatment plan (Fig. 2a) or weighted equivalent square field size (Fig. 2b) are presented in Fig. 2. In analogy, the resulting out-of-field doses DOOF are presented in Fig. 2c and d. In all panels, the eight different treatment scenarios with all 30 treatment sequences are displayed. Mean values are displayed in Fig. 2 while the standard deviations as percent of the corresponding mean value were σ(HPN) = 0.80% (between 0.44 and 1.47% for all groups) and σ(DOOF) = 0.54% (between 0.26 and 0.91% for all groups). Figure 3 shows violin and box plots of the 30 treatment plans for each of the eight treatment approaches. Neutron-equivalent doses per treatment fraction (panel a) and per MU (panel b) as well as out-of-field doses per treatment fraction (panel c) and per MU (panel d) are displayed. Boxplots indicate one interquartile range (IQR).
The prescription dose, the total dose, the total number of MUs, and the equivalent square field sizes as well as average neutron-equivalent (HPN) and out-of-field doses (DOOF) per MU and treatment fraction are presented in Table 2 as mean values and standard deviations. Comparing all respective plans together, it becomes apparent that over all of the different scenarios, SIB resulted only in 57 plans in more MUs but actually in another 63 plans in less MUs than their CONV counterparts. Furthermore, the resulting neutron-equivalent—and out-of-field doses per MU and per complete treatment as well as the correlation coefficients are displayed.

Neutron-equivalent dose between SIB and CONV plans with and without flattening filter, as a function of monitor units, and equivalent square field size

As expected, flattened 10-MV beams resulted in the highest neutron dose per MU with comparable values between CONV and SIB treatment plans. For 10-MV beams, omission of the flattening filter yielded 20% less neutron dose per fraction. The 10-MV SIB plans (flattened and FFF altogether) resulted over their entire treatment course (20 fx) in 38.17% (Table 3) decreased absolute neutron dose when compared to their CONV counterparts (37 fx). A negligible neutron dose (< 0.03 µSv/fx) was measured with 6‑MV treatment plans. When summed up over an entire treatment course of 37 treatment fx, 10-MV FLAT beams resulted in 13.1 mSv neutron dose vs. 10.9 mSv for FFF beam dose delivery (0.5 and 0.9 mSv for 6‑MV equivalents). A total of 20 fx of respective SIB plans resulted in 8.4 mSv neutron dose for 10-MV FLAT beams and 6.5 mSv for 10-MV FFF beams (0.3 and 0.6 mSv for their 6‑MV counterparts, Table 3).
Table 3
Difference in percent for the comparison between conventional (74 Gy) and simultaneous integrated boost (SIB 50/60 Gy) treatment concepts (left side) as well for comparison between FLAT and FFF plans (right side) in terms of resulting neutron doses (HPN) and out-of-field doses (DOOF) at respective nominal accelerator potentials (6 and 10 MV)
Effect SIB
Effect FFF
Groups
Nom. acc. potential
HPN (mSv)
DOOF (mGy)
Groups
Nom. acc. potential
HPN (mSv)
DOOF (mGy)
6‑MV FLAT 74 Gy
6 MV
0.5 ± 0.1
59.2 ± 11.1
6‑MV FLAT 74 Gy
6 MV
0.5 ± 0.1
59.2 ± 11.1
6‑MV FLAT SIB 50/60 Gy
6 MV
0.3 ± 0.1
38.0 ± 10.0
6‑MV FFF 74 Gy
6 MV
0.9 ± 0.1
25.9 ± 3.7
Difference (%)
 
−34.74
−36.13
Difference (%)
 
104.08
−59.13
6‑MV FFF 74 Gy
6 MV
0.9 ± 0.1
25.9 ± 3.7
6‑MV FLAT SIB 50/60 Gy
6 MV
0.3 ± 0.1
38.0 ± 10.0
6‑MV FFF SIB 50/60 Gy
6 MV
0.6 ± 0.2
14.0 ± 4.0
6‑MV FFF SIB 50/60 Gy
6 MV
0.6 ± 0.2
14.0 ± 4.0
Difference (%)
 
−37.13
−39.85
Difference (%)
 
96.59
−61.51
10-MV FLAT 74 Gy
10 MV
13.1 ± 1.5
66.6 ± 14.8
10-MV FLAT 74 Gy
10 MV
13.1 ± 1.5
66.6 ± 14.8
10-MV FLAT SIB 50/60 Gy
10 MV
8.4 ± 1.5
40.0 ± 8.0
10-MV FFF 74 Gy
10 MV
10.9 ± 2.1
25.9 ± 7.4
Difference (%)
 
−36.10
−40.90
Difference (%)
 
−17.18
−60.72
10-MV FFF 74 Gy
10 MV
10.9 ± 2.1
25.9 ± 7.4
10-MV FLAT SIB 50/60 Gy
10 MV
8.6 ± 1.5
40.0 ± 8.0
10-MV FFF SIB 50/60 Gy
10 MV
6.5 ± 1.4
16.0 ± 4.0
10-MV FFF SIB 50/60 Gy
10 MV
6.5 ± 1.4
16.0 ± 4.0
Difference (%)
 
−40.23
−38.79
Difference (%)
 
−22.53
−59.31
A strong positive correlation between MUs and HPN exists for all delivery scenarios with Pearson’s correlation coefficients over r = 0.93 (Table 2). The resulting data do not allow for differentiation between SIB and CONV within one standard deviation for all given scenarios. Weighted equivalent square field size (Aeq) calculations show a moderate-to-strong negative correlation between HPN and Aeq with Pearson’s correlation coefficients between r = −0.55 and −0.83 (Table 2). Therefore, neutron contamination depends obviously predominantly on beam energy and number of delivered MUs. Implementation of an SIB by itself does not result in markedly different neutron doses in comparison to CONV when MUs remain comparable.

Distant out-of-field-dose in SIB and CONV plans with and without flattening filter, as a function of monitor units, and equivalent square field size

Out-of-field doses were lowest for all FFF scenarios (0.7 mGy/fx). The 6‑MV FLAT beams yielded higher doses (1.8 mGy/fx) and 10-MV FLAT resulted in the highest doses (1.9 mGy/fx). Within one standard deviation, CONV and the corresponding SIB plans were identical for all scenarios when a single treatment fraction was compared. Pearson’s correlation coefficients between MUs and DOOF exhibited a strong positive correlation in all cases over r = 0.73, but was the highest for FLAT SIB 6‑MV and 10-MV treatment plans with r > 0.90. No clear correlation was found for the relationship between DOOF and Aeq.
In view of a complete treatment course, SIB plans yielded markedly lower out-of-field doses when compared with their respective CONV plans. A comparison between effects exerted from FFF-VMAT and SIB-VMAT resulted in a comparable reduction of DOOF. For example, for 10-MV FLAT VMAT, 60 Gy SIB RT resulted in 40.9% decreased DOOF (compared with CONV RT) while FFF-mode led to 60.7% (CONV) and 59.3% (SIB) less DOOF readings when compared to their referring FLAT counterparts (10-MV FLAT CONV, 10-MV FLAT SIB). The DOOF-sparing effect of SIB plans was similar for flattened and FFF plans (Table 3).

Comparison of integrated effects of photon energy, SIB and flattening filter-free VMAT on photoneutron dose and distant out-of-field dose

As expected, for HPN readings, photon energy exerted the greatest effect (F value 1647, p < 0.0001), while the influence of SIB was significantly smaller (F value 114, p < 0.0001) as was the effect of FFF-RT (F value 13.9, p = 0.0002).
Interactions were noted between FFF-RT and photon energy (F value 70.6, p < 0.0001) but also for SIB and photon energy (F value 215, p < 0.0001). Therefore, the influence of photon energy on HPN is not consistent but depends on filter position and homogeneity. For the reference group 10-MV FLAT CONV, significantly higher HPN readings were noted in comparison to the most promising plan setting 6‑MV FFF SIB (13.1 ± 1.5 mSv vs. 0.6 ± 0.2 mSv, estimate 12.5 ± 0.3 mSv, p < 0.0001) as opposed to slightly decreased HPN readings for 6‑MV FLAT CONV when compared to 6‑MV FFF SIB (0.5 ± 0.1 mSv vs. 0.6 ± 0.2 mSv, estimate 0.13 ± 0.03 mSv, p < 0.0001, Table 4).
Table 4
Estimated marginal means (emmean) for neutron doses (HPN) and out-of-field doses (DOOF), their standard errors (SE), and confidence levels (CL) for the comparison of standard reference groups with the most promising plan setting (6-MV FFF SIB)
 
10-MV Flat CONV vs. 6‑MV FFF SIB
Emmean
SE
Lower 95% CL
Upper 95% CL
F value
p
HPN (mSv)
12.5
0.3
11.9
13.1
1688.6
< 0.0001
DOOF (mGy)
50.8
2.8
45.7
55.9
328.1
< 0.0001
 
6‑MV Flat CONV vs. 6‑MV FFF SIB
Emmean
SE
Lower 95% CL
Upper 95% CL
F value
p
HPN (mSv)
−0.13
0.03
−0.18
−0.8
17.9
< 0.0001
DOOF (mGy)
45.2
2.2
41.6
55.9
437.2
< 0.0001
F value indicates effect size, significance level p< 0.05
DOOF readings were primarily influenced by the position of the FF (F value 1082, p < 0.0001) and to a smaller extent by SIB (F value 326, p < 0.0001) and even less by photon energy (F value 11, p = 0.001). An interaction was noted only between FFF-RT and SIB (F value 65.5, p < 0.0001). A comparison between 10-MV FLAT CONV and 6‑MV FFF SIB yielded significantly decreased DOOF readings for the FFF SIB plan (66.6 ± 14.8 mGy vs. 14.0 ± 4.0 mGy, estimate 50.8 ± 2.8 mGy, p < 0.0001) as did the comparison between 6‑MV FLAT CONV and 6‑MV FFF SIB (59.2 ± 11.1 mGy vs. 14.0 ± 4.0 mGy, estimate 45.2 ± 2.2 mGy, p < 0.0001, Table 4).

Discussion

This study is aimed to investigate whether implementation of a simultaneous integrated boost (SIB) results in an effect on neutron or scattered photon doses, possibly due to increased MLC activity, which has been identified as one primary source for neutron generation [10]. Furthermore, we explored whether the FFF mode may influence such an effect, if present.
In our 240 treatment sequences, the following case-dependent results were found: In 115 (for DOOF) and 118 (for HPN) out of 120 treatment courses, implementation of an SIB resulted in less DOOF and HPN when compared to CONV. The FFF mode was preferable over FLAT in all cases with respect to DOOF. For 6 MV, neutron doses were higher for FFF beams in all cases (60/60). For 10 MV, neutron doses were lower in 54 of 60 cases with the FFF mode. As expected, neutron doses were always higher for 10 MV than for 6 MV. On the other hand, DOOF values were lower in 104 of the 120 cases for 6 MV.
The SIB concept leads to an increased fractional RT dose within a selected volume. When calculated for comparable efficacy, the EQD2Gy model calculates for a given alpha/beta value a reduced total RT dose, which in turn implies a reduction in the fraction number, therefore shortening the overall treatment time. We found that our SIB scenario (2.5 Gy/3.0 Gy to 50 Gy/60 Gy) results in decreased total neutron dose as well as DOOF when compared to conventional (CONV) fractionated 2 Gy to 74 Gy. Therefore, due to the higher fractional doses and thus reduced number of treatment fractions, an SIB is clearly preferable in this context. Flattening filter-free VMAT, effective in sparing both neutron dose (in 10-MV beams) and DOOF, respectively, increased this effect.
Omission of the flattening filter (FFF) significantly decreases both the photoneutron as well as the out-of-field dose in VMAT for prostate cancer at distant locations, even though FFF plans typically require more MUs than their flattened counterparts to compensate for the non-flatness of its profiles. Monitor units as surrogate for the modulation degree have a strong linear correlation with the formation of photoneutrons. By contrast, with respect to out-of-field dose, this strong linear correlation with MUs was only noted in flattened SIB plans.
We show that no increased neutron (HPN) or scattered photon dose (DOOF) results from increasing the fractional dose from 2.0 Gy to 2.5/3.0 Gy. Therefore, implementation of an SIB in a hypofractionated dose concept leads to less neutron and out-of-field dose in a complete treatment course and this effect was as distinct as the neutron dose-sparing and scattered radiation-sparing effect exerted by FFF VMAT.
Since the measurement position was kept constant at all times at 61.8 cm away from the beam isocenter, we were unable to detect a relationship between the distance to the radiation field and scattered dose. Hauri et al. [22] demonstrated a dependency of scattered dose on the distance from the isocenter. Between 40 and 80 cm away from the isocenter, total doses decrease slowly and the resulting scatter dose results mainly from collimator scatter and head leakage (and less from patient scatter). Haelg et al. [23] showed that neutron contamination depends only weakly on distance from the radiation field. Thus, a comparison between different treatment plan parameters (6 vs. 10 MV, FFF vs. FLAT, and SIB vs. CONV) appears to be appropriate at a fixed distance and the results presented may be used to approximate a whole-body exposure.
In our scenarios the boost dose is 3.0 Gy while the surrounding PTV receives 2.5 Gy, hence a ratio of 1.25 to 1.5 when compared to CONV treatment plans (2 Gy). For the sake of comparison, all absolute neutron and out-of-field dose readings were taken in relation with MUs to account for these differing dose levels. When analyzing all 30 patients, we found both increased and decreased numbers of MUs for the respective SIB plans compared to their corresponding CONV treatment plans. When the 120 SIB treatment plans were corrected by 3 Gy/2.5 Gy (to yield MU per Gray PTV dose), 57 SIB treatment plans had more MUs and 63 SIB plans exhibited less MUs than their CONV counterparts. Non-uniform target coverage therefore requires in every second case less MUs than their uniform equivalents. This also implies less neutron dose.
Due to the higher number of MUs required in FFF dose deliveries, the reduced neutron generation has to be set in relation to the difference in MUs to judge whether FFF or FLAT deliveries generate less neutron dose to the patient. In flattened SIB plans (6 MV and 10 MV), we noticed a comparable correlation between MUs and DOOF, an effect which was circumvented by the omission of the FF.
In AAPM report 158 [12], deposited neutron dose equivalents per MU for 6‑MV and 10-MV Elekta linacs are not provided. However, our finding of 0.5 µSv/MU for 10-MV FLAT are in line with the reported Varian 10-MV FLAT beam results of 0.9 µSv/MU and 0.3 µSv/MU for Siemens linacs. It is, moreover, noted in AAPM report 158 that Varian linacs possess a roughly twofold higher neutron contamination than comparable Elekta linacs.
Even though no relevant neutron dose was measured in 6‑MV beams, a stable signal was obtained at a very low level, which also showed a correlation between MUs and HPN with r values between 0.93 and 1.0 (Table 2), and a slight increase in average neutron dose was noted for 6‑MV FFF when compared with 6‑MV FLAT. Of note, even for this beam energy, a reduction in HPN is induced for an overall RT course with implementation of SIB (34.7 and 37.1% for 6‑MV FLAT and 6‑MV FFF, respectively; Table 3).
At the nominal energy of 6‑MV FLAT, the flattening filter does not contribute to PN contamination, because the bremsstrahlung is filtered with a stainless steel disc [24] and energy thresholds for PN production (γ, n) for iron and copper are 11.2 MeV (91.7%, Fe-56), 10.9 MeV (69.2%, Cu-63), and 9.9 MeV (30.3%, Cu-65; [25]). The same IAEA report lists all energy thresholds for PN production (γ, n) for the respective isotope and its natural abundance for tungsten. These were reported to be 8.1 MeV (26.3%, W‑182), 7.4 MeV (30.7% W‑184), 6.2 MeV (14.3%, W‑183), and 5.8 MeV (28.6%, W‑186). The 6‑MV Elekta linac’s maximum photon energies were reported to be 6.6 or 6.7 MeV, respectively [26, 27]. Furthermore, Elekta linacs—in contrast to Varian linacs—increase the initial electron energy for FFF beams up to a point where the D10cm of FLAT and FFF is identical (e.g., 67.5/73.5% of the maximum dose at 10 cm depth for 6/10 MV; [28]). This, in summary explains why detected Hpn values for 6‑MV FFF are higher than 6‑MV FLAT: 0.027 µSv/MU > 0.015 µSv/MU. For 10 MV this should apply as well; however, the flattening filter as an additional neutron source dominates and reverses the behavior 0.5 µSv/MU (10-MV FLAT) > 0.3 µSv/MU (10-MV FFF).
In our study, the SIB scenario reduced for 10-MV plans the total neutron dose from 13.1 to 8.4 mSv (36.1%) in flattened VMAT and from 10.9 to 6.5 mSv for FFF-VMAT (40.2%). FFF-VMAT reduced neutron dose from 13.1 mSV to 10.9 mSV (17.2%) for the conventional fractionated scenario and from 8.4 to 6.5 mSv (22.5%) for SIB-RT. Scattered radiation (DOOF) was likewise reduced for both 6‑MV and 10-MV beams with SIB by approximately 36.1–40.9% and even more effectively with FFF by approximately 59.1–61.5%. The lowest neutron and scattered radiation dose was derived with 6‑MV SIB (HPN 0.3–0.6 mSv; DOOF 14.0 mGy for 6‑MV FFF).
Treutwein et al. [29] compared the excess absolute risk (EAR) of secondary cancer in localized prostate IMRT with 6‑MV beams with and without flattening filter. For the secondary malignancy risk (SMR) in the periphery—corresponding to our distant measurement localization—they noted a reduced peripheral dose without flattening filter, which led to significantly reduced SMR. Murray et al. [13] described lower second cancer risks for FFF with increasing impact in organs at greater distance. This beneficial effect may be increased with SIB-RT. These reported SMR induced by scattered photons and electrons are very low for regions in the body at a distance from the primary treated region. Nevertheless, photoneutrons pose a significant risk in view of secondary cancers (ICRP 60; [30]) and even more so to implanted electronic devices because of their interaction with complementary metal oxide semiconductors and random access memory [3133]. Hauri et al. presented data that showed that neutron dose equivalent remained unaffected by the distance from the isocenter and neutron dose peaked in the neck region along an axis from the pelvis (isocenter) to the head in an Alderson phantom, probably due to different thickness of penetrated tissue [22]. Corresponding to our data, Haelg et al. showed that neutron dose resulting from 15-MV IMRT produced by linacs from three different vendors (Elekta, Varian, and Siemens) was at least one order lower than the photon stray dose, therefore contributing with a very low amount to the total integral dose of a patient [23]. Still, regarding neutron dose, significant effects were noted in patients and phantom studies with respect to cardiac implanted electronic devices (CIEDs, [32, 33]). These devices are implanted near the neck region under the clavicular bone in a subcutaneous pouch, therefore in the peak region for neutron dose [22]. In a phantom study, 10-MV FFF-VMAT of the prostate resulted in pacing inhibition, data loss and reset in 4 of 15 ICDs that were located at a distance from the primary beam [19]. In an exemplary clinical survey, 10-MV FFF-VMAT resulted in an ICD error [32]. Other clinical studies demonstrate that photon beams >6 MV result in CIED errors [3133] even though beam energies were widespread between 10 and 23 MV. International guidelines advocate the use of non-neutron-producing beam energies in CIED-bearing patients, but discrepancy exists between a lower beam energy threshold (6 vs. 10 MV, [18, 31, 3335]). No data exist for the disaggregation between 10-MV FLAT and 10-MV FFF in regard to CIED effects even though we have shown a 20% reduction in neutron dose in 10-MV FFF VMAT when compared to their 10-MV FLAT counterpart plans (for the same fx dose level). Therefore, although we show that SIB can further reduce HPN in FFF-VMAT (HPN 10-MV FLAT > 10-MV FFF > 10-MV FLAT SIB > 10-MV FFF SIB), we cannot advise for the use of 10-MV FFF beams even for RT with SIB in CIED-bearing patients and have to wait for further research on this topic.
Our study has some limitations that need consideration. We only used one measurement point approximately 60 cm from the field edge 0.5 cm below the surface of the phantom. We chose this location with reference to Hauri et al. [22]. At 60 cm distance, there is almost exclusively head leakage contribution and almost no patient scatter. Thus, past this distance the measured doses remain approximately constant. Closer to the radiation field, both Hauri et al. and AAPM report 158 [12] report an approximately exponential decay of patient scatter and a constant head leakage contribution. Furthermore, this distance between detectors and radiation field reflects clinically the distance between prostate RT and CIEDs [19] and considers HPN and DOOF at their typical location in the body.

Conclusion

In conclusion, implementation of a simultaneous integrated boost (SIB) and omission of the flattening filter (FFF) reduces neutron contamination and out-of-field doses per monitor unit for volumetric modulated arc therapy (VMAT) for prostate cancer. SIB treatments lead to lower neutron and out-of-field doses for an overall treatment course. Neutron contamination in VMAT for prostate cancer depends, as expected, predominantly on the beam energy and the number of delivered monitor units. Due to the higher number of monitor units typically required in flattening filter-free (FFF) dose deliveries with this particular set-up, the reduced neutron generation per monitor unit has to be set in relation to the difference in monitor units to judge whether FFF or FLAT deliveries generate a lower neutron and out-of-field dose to the patient. Respecting radiation protection principles, an implementation of an SIB in moderately hypofractionated prostate RT may be advisable not only for shortening of overall treatment time.

Acknowledgements

We thank Professor Christel Weiss, head of the Department of Medical Statistics, Biomathematics and Information Processing, Medical Faculty Mannheim, University of Heidelberg, for advice, statistical calculations and revisional contributions to this document.

Conflict of interest

B. Gauter-Fleckenstein, S. Schönig, L. Mertens, H. Oppitz, K. Siebenlist, M. Ehmann and J. Fleckenstein declare that they have no competing interests.
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article’s Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article’s Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://​creativecommons.​org/​licenses/​by/​4.​0/​.

Unsere Produktempfehlungen

e.Med Interdisziplinär

Kombi-Abonnement

Für Ihren Erfolg in Klinik und Praxis - Die beste Hilfe in Ihrem Arbeitsalltag

Mit e.Med Interdisziplinär erhalten Sie Zugang zu allen CME-Fortbildungen und Fachzeitschriften auf SpringerMedizin.de.

e.Med Radiologie

Kombi-Abonnement

Mit e.Med Radiologie erhalten Sie Zugang zu CME-Fortbildungen des Fachgebietes Radiologie, den Premium-Inhalten der radiologischen Fachzeitschriften, inklusive einer gedruckten Radiologie-Zeitschrift Ihrer Wahl.

Strahlentherapie und Onkologie

Print-Titel

•Übersichten, Originalien, Kasuistiken

•Kommentierte Literatur aus der Radioonkologie, Strahlenbiologie und -physik

Literatur
1.
Zurück zum Zitat Schaeffer EM, Srinivas S, Adra N, An Y, Barocas D, Bitting R et al (2022) NCCN Guidelines(R) Insights: Prostate Cancer, Version 1.2023. J Natl Compr Canc Netw 20:1288–1298PubMed Schaeffer EM, Srinivas S, Adra N, An Y, Barocas D, Bitting R et al (2022) NCCN Guidelines(R) Insights: Prostate Cancer, Version 1.2023. J Natl Compr Canc Netw 20:1288–1298PubMed
2.
Zurück zum Zitat German S3 Evidence-Based Guideline For the therapy of Prostate Cancer Leitlinienprogramm onkologie (Deutsche Krebsgesellschaft DK, AWMF): S3-leitlinie prostatakarzinom, leitlinienreport, version 6.0, Mai 2021, AWMF registernummer: 043/022OL. German S3 Evidence-Based Guideline For the therapy of Prostate Cancer 2021. German S3 Evidence-Based Guideline For the therapy of Prostate Cancer Leitlinienprogramm onkologie (Deutsche Krebsgesellschaft DK, AWMF): S3-leitlinie prostatakarzinom, leitlinienreport, version 6.0, Mai 2021, AWMF registernummer: 043/022OL. German S3 Evidence-Based Guideline For the therapy of Prostate Cancer 2021.
3.
Zurück zum Zitat Dearnaley D, Syndikus I, Mossop H, Khoo V, Birtle A, Bloomfield D et al (2016) Conventional versus hypofractionated high-dose intensity-modulated radiotherapy for prostate cancer: 5‑year outcomes of the randomised, non-inferiority, phase 3 CHHiP trial. Lancet Oncol 17:1047–1060CrossRefPubMedPubMedCentral Dearnaley D, Syndikus I, Mossop H, Khoo V, Birtle A, Bloomfield D et al (2016) Conventional versus hypofractionated high-dose intensity-modulated radiotherapy for prostate cancer: 5‑year outcomes of the randomised, non-inferiority, phase 3 CHHiP trial. Lancet Oncol 17:1047–1060CrossRefPubMedPubMedCentral
4.
Zurück zum Zitat Brand DH, Tree AC, Ostler P, van der Voet H, Loblaw A, Chu W et al (2019) Intensity-modulated fractionated radiotherapy versus stereotactic body radiotherapy for prostate cancer (PACE-B): acute toxicity findings from an international, randomised, open-label, phase 3, non-inferiority trial. Lancet Oncol 20:1531–1543CrossRefPubMedPubMedCentral Brand DH, Tree AC, Ostler P, van der Voet H, Loblaw A, Chu W et al (2019) Intensity-modulated fractionated radiotherapy versus stereotactic body radiotherapy for prostate cancer (PACE-B): acute toxicity findings from an international, randomised, open-label, phase 3, non-inferiority trial. Lancet Oncol 20:1531–1543CrossRefPubMedPubMedCentral
5.
Zurück zum Zitat Catton CN, Lukka H, Gu CS, Martin JM, Supiot S, Chung PWM et al (2017) Randomized trial of a hypofractionated radiation regimen for the treatment of localized prostate cancer. J Clin Oncol 35:1884–1890CrossRefPubMed Catton CN, Lukka H, Gu CS, Martin JM, Supiot S, Chung PWM et al (2017) Randomized trial of a hypofractionated radiation regimen for the treatment of localized prostate cancer. J Clin Oncol 35:1884–1890CrossRefPubMed
6.
Zurück zum Zitat Incrocci L, Wortel RC, Alemayehu WG, Aluwini S, Schimmel E, Krol S et al (2016) Hypofractionated versus conventionally fractionated radiotherapy for patients with localised prostate cancer (HYPRO): final efficacy results from a randomised, multicentre, open-label, phase 3 trial. Lancet Oncol 17:1061–1069CrossRefPubMed Incrocci L, Wortel RC, Alemayehu WG, Aluwini S, Schimmel E, Krol S et al (2016) Hypofractionated versus conventionally fractionated radiotherapy for patients with localised prostate cancer (HYPRO): final efficacy results from a randomised, multicentre, open-label, phase 3 trial. Lancet Oncol 17:1061–1069CrossRefPubMed
7.
Zurück zum Zitat Kerkmeijer LGW, Groen VH, Pos FJ, Haustermans K, Monninkhof EM, Smeenk RJ et al (2021) Focal boost to the Intraprostatic tumor in external beam radiotherapy for patients with localized prostate cancer: results from the FLAME randomized phase III trial. J Clin Oncol 39:787–796CrossRefPubMed Kerkmeijer LGW, Groen VH, Pos FJ, Haustermans K, Monninkhof EM, Smeenk RJ et al (2021) Focal boost to the Intraprostatic tumor in external beam radiotherapy for patients with localized prostate cancer: results from the FLAME randomized phase III trial. J Clin Oncol 39:787–796CrossRefPubMed
8.
Zurück zum Zitat Murray JR, Tree AC, Alexander EJ, Sohaib A, Hazell S, Thomas K et al (2020) Standard and Hypofractionated dose escalation to Intraprostatic tumor nodules in localized prostate cancer: efficacy and toxicity in the DELINEATE trial. Int J Radiat Oncol Biol Phys 106:715–724CrossRefPubMed Murray JR, Tree AC, Alexander EJ, Sohaib A, Hazell S, Thomas K et al (2020) Standard and Hypofractionated dose escalation to Intraprostatic tumor nodules in localized prostate cancer: efficacy and toxicity in the DELINEATE trial. Int J Radiat Oncol Biol Phys 106:715–724CrossRefPubMed
9.
Zurück zum Zitat Craft D, Khan F, Young M, Bortfeld T (2016) The price of target dose uniformity. Int J Radiat Oncol Biol Phys 96:913–914CrossRefPubMed Craft D, Khan F, Young M, Bortfeld T (2016) The price of target dose uniformity. Int J Radiat Oncol Biol Phys 96:913–914CrossRefPubMed
10.
Zurück zum Zitat Kry SF, Titt U, Ponisch F, Vassiliev ON, Salehpour M, Gillin M et al (2007) Reduced neutron production through use of a flattening-filter-free accelerator. Int J Radiat Oncol Biol Phys 68:1260–1264CrossRefPubMed Kry SF, Titt U, Ponisch F, Vassiliev ON, Salehpour M, Gillin M et al (2007) Reduced neutron production through use of a flattening-filter-free accelerator. Int J Radiat Oncol Biol Phys 68:1260–1264CrossRefPubMed
11.
Zurück zum Zitat The 2007 Recommendations of the International Commission on Radiological Protection. ICRP publication 103. Ann ICRP. 2007;37:1–332. The 2007 Recommendations of the International Commission on Radiological Protection. ICRP publication 103. Ann ICRP. 2007;37:1–332.
12.
Zurück zum Zitat Kry SF, Bednarz B, Howell RM, Dauer L, Followill D, Klein E et al (2017) AAPM TG 158: Measurement and calculation of doses outside the treated volume from external-beam radiation therapy. Med Phys 44:e391–e429CrossRefPubMed Kry SF, Bednarz B, Howell RM, Dauer L, Followill D, Klein E et al (2017) AAPM TG 158: Measurement and calculation of doses outside the treated volume from external-beam radiation therapy. Med Phys 44:e391–e429CrossRefPubMed
13.
Zurück zum Zitat Murray LJ, Thompson CM, Lilley J, Cosgrove V, Franks K, Sebag-Montefiore D et al (2015) Radiation-induced second primary cancer risks from modern external beam radiotherapy for early prostate cancer: impact of stereotactic ablative radiotherapy (SABR), volumetric modulated arc therapy (VMAT) and flattening filter free (FFF) radiotherapy. Phys Med Biol 60:1237–1257CrossRefPubMed Murray LJ, Thompson CM, Lilley J, Cosgrove V, Franks K, Sebag-Montefiore D et al (2015) Radiation-induced second primary cancer risks from modern external beam radiotherapy for early prostate cancer: impact of stereotactic ablative radiotherapy (SABR), volumetric modulated arc therapy (VMAT) and flattening filter free (FFF) radiotherapy. Phys Med Biol 60:1237–1257CrossRefPubMed
14.
Zurück zum Zitat Sanchez-Nieto B, Medina-Ascanio KN, Rodriguez-Mongua JL, Doerner E, Espinoza I (2020) Study of out-of-field dose in photon radiotherapy: a commercial treatment planning system versus measurements and monte Carlo simulations. Med Phys 47:4616–4625CrossRefPubMed Sanchez-Nieto B, Medina-Ascanio KN, Rodriguez-Mongua JL, Doerner E, Espinoza I (2020) Study of out-of-field dose in photon radiotherapy: a commercial treatment planning system versus measurements and monte Carlo simulations. Med Phys 47:4616–4625CrossRefPubMed
15.
Zurück zum Zitat Kragl G, Baier F, Lutz S, Albrich D, Dalaryd M, Kroupa B et al (2011) Flattening filter free beams in SBRT and IMRT: dosimetric assessment of peripheral doses. Z Med Phys 21:91–101CrossRefPubMed Kragl G, Baier F, Lutz S, Albrich D, Dalaryd M, Kroupa B et al (2011) Flattening filter free beams in SBRT and IMRT: dosimetric assessment of peripheral doses. Z Med Phys 21:91–101CrossRefPubMed
16.
Zurück zum Zitat Kry SF, Howell RM, Titt U, Salehpour M, Mohan R, Vassiliev ON (2008) Energy spectra, sources, and shielding considerations for neutrons generated by a flattening filter-free Clinac. Med Phys 35:1906–1911CrossRefPubMed Kry SF, Howell RM, Titt U, Salehpour M, Mohan R, Vassiliev ON (2008) Energy spectra, sources, and shielding considerations for neutrons generated by a flattening filter-free Clinac. Med Phys 35:1906–1911CrossRefPubMed
17.
Zurück zum Zitat Vassiliev ON, Kry SF, Kuban DA, Salehpour M, Mohan R, Titt U (2007) Treatment-planning study of prostate cancer intensity-modulated radiotherapy with a Varian Clinac operated without a flattening filter. Int J Radiat Oncol Biol Phys 68:1567–1571CrossRefPubMed Vassiliev ON, Kry SF, Kuban DA, Salehpour M, Mohan R, Titt U (2007) Treatment-planning study of prostate cancer intensity-modulated radiotherapy with a Varian Clinac operated without a flattening filter. Int J Radiat Oncol Biol Phys 68:1567–1571CrossRefPubMed
20.
Zurück zum Zitat Olsher RH, Hsu HH, Beverding A, Kleck JH, Casson WH, Vasilik DG et al (2000) WENDI: an improved neutron rem meter. Health Phys 79:170–181CrossRefPubMed Olsher RH, Hsu HH, Beverding A, Kleck JH, Casson WH, Vasilik DG et al (2000) WENDI: an improved neutron rem meter. Health Phys 79:170–181CrossRefPubMed
21.
Zurück zum Zitat Gutermuth F, Fehrenbacher G, Radon T, Siekmann R (2004) Test of the rem-counter WENDI-II from Eberline in different energy-dispersed neutron fields. CERN report EXT-2004-085. Gutermuth F, Fehrenbacher G, Radon T, Siekmann R (2004) Test of the rem-counter WENDI-II from Eberline in different energy-dispersed neutron fields. CERN report EXT-2004-085.
22.
Zurück zum Zitat Hauri P, Schneider U (2019) Whole-body dose equivalent including neutrons is similar for 6 MV and 15 MV IMRT, VMAT, and 3D conformal radiotherapy. J Appl Clin Med Phys 20:56–70CrossRefPubMedPubMedCentral Hauri P, Schneider U (2019) Whole-body dose equivalent including neutrons is similar for 6 MV and 15 MV IMRT, VMAT, and 3D conformal radiotherapy. J Appl Clin Med Phys 20:56–70CrossRefPubMedPubMedCentral
23.
Zurück zum Zitat Halg RA, Besserer J, Boschung M, Mayer S, Lomax AJ, Schneider U (2014) Measurements of the neutron dose equivalent for various radiation qualities, treatment machines and delivery techniques in radiation therapy. Phys Med Biol 59:2457–2468CrossRefPubMed Halg RA, Besserer J, Boschung M, Mayer S, Lomax AJ, Schneider U (2014) Measurements of the neutron dose equivalent for various radiation qualities, treatment machines and delivery techniques in radiation therapy. Phys Med Biol 59:2457–2468CrossRefPubMed
24.
Zurück zum Zitat Xiao Y, Kry SF, Popple R, Yorke E, Papanikolaou N, Stathakis S et al (2015) Flattening filter-free accelerators: a report from the AAPM Therapy Emerging Technology Assessment Work Group. J Appl Clin Med Phys 16:5219CrossRefPubMed Xiao Y, Kry SF, Popple R, Yorke E, Papanikolaou N, Stathakis S et al (2015) Flattening filter-free accelerators: a report from the AAPM Therapy Emerging Technology Assessment Work Group. J Appl Clin Med Phys 16:5219CrossRefPubMed
25.
Zurück zum Zitat Agency IIAE (2000) Handbook on photonuclear data for applications: cross-sections and spectra Agency IIAE (2000) Handbook on photonuclear data for applications: cross-sections and spectra
26.
Zurück zum Zitat Fleckenstein J, Jahnke L, Lohr F, Wenz F, Hesser J (2013) Development of a Geant4 based Monte Carlo Algorithm to evaluate the MONACO VMAT treatment accuracy. Z Med Phys 23:33–45CrossRefPubMed Fleckenstein J, Jahnke L, Lohr F, Wenz F, Hesser J (2013) Development of a Geant4 based Monte Carlo Algorithm to evaluate the MONACO VMAT treatment accuracy. Z Med Phys 23:33–45CrossRefPubMed
27.
Zurück zum Zitat Gholampourkashi S, Cygler JE, Belec J, Vujicic M, Heath E (2019) Monte Carlo and analytic modeling of an Elekta infinity linac with agility MLC: investigating the significance of accurate model parameters for small radiation fields. J Appl Clin Med Phys 20:55–67CrossRefPubMed Gholampourkashi S, Cygler JE, Belec J, Vujicic M, Heath E (2019) Monte Carlo and analytic modeling of an Elekta infinity linac with agility MLC: investigating the significance of accurate model parameters for small radiation fields. J Appl Clin Med Phys 20:55–67CrossRefPubMed
28.
Zurück zum Zitat Fogliata A, Fleckenstein J, Schneider F, Pachoud M, Ghandour S, Krauss H et al (2016) Flattening filter free beams from TrueBeam and Versa HD units: Evaluation of the parameters for quality assurance. Med Phys 43:205CrossRefPubMed Fogliata A, Fleckenstein J, Schneider F, Pachoud M, Ghandour S, Krauss H et al (2016) Flattening filter free beams from TrueBeam and Versa HD units: Evaluation of the parameters for quality assurance. Med Phys 43:205CrossRefPubMed
29.
Zurück zum Zitat Treutwein M, Loeschel R, Hipp M, Koelbl O, Dobler B (2020) Secondary malignancy risk for patients with localized prostate cancer after intensity-modulated radiotherapy with and without flattening filter. J Appl Clin Med Phys 21:197–205CrossRefPubMedPubMedCentral Treutwein M, Loeschel R, Hipp M, Koelbl O, Dobler B (2020) Secondary malignancy risk for patients with localized prostate cancer after intensity-modulated radiotherapy with and without flattening filter. J Appl Clin Med Phys 21:197–205CrossRefPubMedPubMedCentral
30.
31.
Zurück zum Zitat Gauter-Fleckenstein B, Israel CW, Dorenkamp M, Dunst J, Roser M, Schimpf R et al (2015) DEGRO/DGK guideline for radiotherapy in patients with cardiac implantable electronic devices. Strahlenther Onkol 191:393–404CrossRefPubMed Gauter-Fleckenstein B, Israel CW, Dorenkamp M, Dunst J, Roser M, Schimpf R et al (2015) DEGRO/DGK guideline for radiotherapy in patients with cardiac implantable electronic devices. Strahlenther Onkol 191:393–404CrossRefPubMed
32.
Zurück zum Zitat Gauter-Fleckenstein B, Barthel C, Buttner S, Wenz F, Borggrefe M, Tulumen E (2020) Effectivity and applicability of the German DEGRO/DGK-guideline for radiotherapy in CIED-bearing patients. Radiother Oncol 152:208–215CrossRefPubMed Gauter-Fleckenstein B, Barthel C, Buttner S, Wenz F, Borggrefe M, Tulumen E (2020) Effectivity and applicability of the German DEGRO/DGK-guideline for radiotherapy in CIED-bearing patients. Radiother Oncol 152:208–215CrossRefPubMed
33.
Zurück zum Zitat Zecchin M, Severgnini M, Fiorentino A, Malavasi VL, Menegotti L, Alongi F et al (2018) Management of patients with cardiac implantable electronic devices (CIED) undergoing radiotherapy: A consensus document from Associazione Italiana Aritmologia e Cardiostimolazione (AIAC), Associazione Italiana Radioterapia Oncologica (AIRO), Associazione Italiana Fisica Medica (AIFM). Int J Cardiol 255:175–183CrossRefPubMed Zecchin M, Severgnini M, Fiorentino A, Malavasi VL, Menegotti L, Alongi F et al (2018) Management of patients with cardiac implantable electronic devices (CIED) undergoing radiotherapy: A consensus document from Associazione Italiana Aritmologia e Cardiostimolazione (AIAC), Associazione Italiana Radioterapia Oncologica (AIRO), Associazione Italiana Fisica Medica (AIFM). Int J Cardiol 255:175–183CrossRefPubMed
34.
Zurück zum Zitat Hurkmans CW, Knegjens JL, Oei BS, Maas AJ, Uiterwaal GJ, van der Borden AJ et al (2012) Management of radiation oncology patients with a pacemaker or ICD: a new comprehensive practical guideline in The Netherlands. Dutch Society of Radiotherapy and Oncology (NVRO). Radiat Oncol 7:198CrossRefPubMedPubMedCentral Hurkmans CW, Knegjens JL, Oei BS, Maas AJ, Uiterwaal GJ, van der Borden AJ et al (2012) Management of radiation oncology patients with a pacemaker or ICD: a new comprehensive practical guideline in The Netherlands. Dutch Society of Radiotherapy and Oncology (NVRO). Radiat Oncol 7:198CrossRefPubMedPubMedCentral
35.
Zurück zum Zitat Indik JH, Gimbel JR, Abe H, Alkmim-Teixeira R, Birgersdotter-Green U, Clarke GD et al (2017) 2017 HRS expert consensus statement on magnetic resonance imaging and radiation exposure in patients with cardiovascular implantable electronic devices. Heart Rhythm 14:e97–e153CrossRefPubMed Indik JH, Gimbel JR, Abe H, Alkmim-Teixeira R, Birgersdotter-Green U, Clarke GD et al (2017) 2017 HRS expert consensus statement on magnetic resonance imaging and radiation exposure in patients with cardiovascular implantable electronic devices. Heart Rhythm 14:e97–e153CrossRefPubMed
Metadaten
Titel
Effect of simultaneous integrated boost concepts on photoneutron and distant out-of-field doses in VMAT for prostate cancer
verfasst von
Benjamin Gauter-Fleckenstein, M.D.
Sebastian Schönig
Lena Mertens
Hans Oppitz
Kerstin Siebenlist
Michael Ehmann
Jens Fleckenstein
Publikationsdatum
14.09.2023
Verlag
Springer Berlin Heidelberg
Erschienen in
Strahlentherapie und Onkologie / Ausgabe 3/2024
Print ISSN: 0179-7158
Elektronische ISSN: 1439-099X
DOI
https://doi.org/10.1007/s00066-023-02138-x

Weitere Artikel der Ausgabe 3/2024

Strahlentherapie und Onkologie 3/2024 Zur Ausgabe

Erhöhtes Risiko fürs Herz unter Checkpointhemmer-Therapie

28.05.2024 Nebenwirkungen der Krebstherapie Nachrichten

Kardiotoxische Nebenwirkungen einer Therapie mit Immuncheckpointhemmern mögen selten sein – wenn sie aber auftreten, wird es für Patienten oft lebensgefährlich. Voruntersuchung und Monitoring sind daher obligat.

Costims – das nächste heiße Ding in der Krebstherapie?

28.05.2024 Onkologische Immuntherapie Nachrichten

„Kalte“ Tumoren werden heiß – CD28-kostimulatorische Antikörper sollen dies ermöglichen. Am besten könnten diese in Kombination mit BiTEs und Checkpointhemmern wirken. Erste klinische Studien laufen bereits.

Positiver FIT: Die Ursache liegt nicht immer im Dickdarm

27.05.2024 Blut im Stuhl Nachrichten

Immunchemischer Stuhltest positiv, Koloskopie negativ – in solchen Fällen kann die Blutungsquelle auch weiter proximal sitzen. Ein Forschungsteam hat nachgesehen, wie häufig und in welchen Lokalisationen das der Fall ist.

Mammakarzinom: Brustdichte beeinflusst rezidivfreies Überleben

26.05.2024 Mammakarzinom Nachrichten

Frauen, die zum Zeitpunkt der Brustkrebsdiagnose eine hohe mammografische Brustdichte aufweisen, haben ein erhöhtes Risiko für ein baldiges Rezidiv, legen neue Daten nahe.

Update Onkologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.