Feasibility of machine integrated point of care lung ultrasound automatic B-lines tool in the Corona-virus 2019 critical care unit

verfasst von: Gal Tsaban, Ori Galante, Yaniv Almog, Yuval Ullman, Lior Fuchs

Erschienen in: Critical Care | Ausgabe 1/2021

download
DOWNLOAD
print
DRUCKEN
insite
SUCHEN
Hinweise

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
In the last decade, the use of point-of-care lung ultrasound (POC-LUS) has substantially increased and in some intensive care units has largely replaced the routine use of chest X-rays tests [1]. While POC-LUS is highly efficient, it requires expertise, and when wrongfully interpreted may badly influence the treating physician's decision-making. Due to the logistic and medical equipment limitations in quarantined zones in which COVID-19 patients are treated, the importance of POC-LUS increased [2]. We aimed to assess the reliability of a newly designed POC-LUS machine-integrated tool for automatic-real-time bedside quantification of B-lines among critical COVID-19 patients.
We assessed all patients admitted to the COVID-19 intensive-care unit in Soroka University Medical Center (SUMC) in Beersheva between February 2, 2020, and June 23, 2020. POC-LUS was performed daily during the routine medical rounds using VenueTM ultrasound machine, General Electric.
Intensive care specialists acquired data in the COVID-19 ICU by placing the probe longitudinally in the intercostal space and prospectively recording for four seconds. The 4-second time-frame allowed the shortest acquisition time with high sensitivity to detect B-lines [3]. All clips were reviewed and blindly assessed for B-lines quantification by two experienced physician operators (GT, 5 years and LF, 9 years). Separately, post hoc automatic analyses were performed using the VenueTM integrated auto-B-lines tool.
VenueTM-auto-B-lines tool provides a grade between zero to ≥ five B-lines, detected only as lines reaching the bottom of the screen. Since the estimation of ≥ 3 B-lines in LUS screening of the lungs' independent zones is considered pathological [4], we also divided the B-lines score to "dry" (≤ 3 B-lines) or "wet" (≥ 4 B-lines).
Each clip was considered an independent observation point. We divided the B-lines assessment into three severity groups: (1) ≤ 2 (non-pathologic), (2) 3–4, and (3) ≥ 5 (severe-ARDS). To assess the agreement between the automatic and physician-assessed quantifications, we performed Cohen's-Kappa tests. A two-sided p-value ≤ 0.05 was considered statistically significant. All analyses were performed using SPSS 26.0 (Armonk, NY, USA).
During the study, 153 clips were acquired from ten patients. All clips were interpretable and included in the analysis. Patients were primarily males (90%), with a mean age of 61.0 ± 8.0 and a mean body mass index of 24.5 ± 10.5. The mean-p/f-ratio was 203.7 ± 83.2, and 70% of patients were mechanically ventilated.
When assessed by physician, 104 observations were classified as group 3, 33 as group 2, and 16 as group 1. In automatic analysis, 18 observations were classified as group 1, 33 as group 2, and 102 as group 3. An example of automatic-POC-LUS assessment across the spectrum of SARS-CoV2-ARDS severity is illustrated in Fig. 1. The weighted Cohen's-Kappa for agreement between the automatic and physician quantifications was 0.734 (95% CI 0.641, 0.826; p < 0.001). When divided into "dry" (≤ 3 B-lines) vs. "wet" (≥ 4 B-lines), the weighted Cohen's-Kappa was 0.822 (95% CI 0.716, 0.928; p < 0.001).
In conclusion, we found that the machine-integrated VenueTM-auto-B-lines tool is highly reliable among severe COVID-19 ICU patients. To our knowledge, this is the first study to validate a machine-integrated automatic-B-lines quantification tool with high reliability among COVID-19 patients. The small number of patients included in this feasibility trial should be acknowledged as a limitation; thus, further research to validate the results of this study is warranted. Previously, a non-integrated automatic B-lines detection application has shown good overall reliability while also stressing the difference in assessment reliability dependent on the operator experience [5]. The current study's results may help better interpret POC-LUS assessments performed by less-experienced operators and reduce inter-operator variability. This tool may provide technological infrastructure for future telemedicine, even in non-experienced hands or for self-assessment by patients. The VenueTM-auto-B-lines tool may reduce the medical staff's exposure time and promote more accurate and standardized LUS assessment methods.

Acknowledgements

We thank the medical staff of the internal medicine division in SUMC and the nursing staff of the medical intensive care unit in SUMC for their valuable contribution to this study.

Declarations

The study was approved by the SUMC institutional review board (#2020-0050-17). As this was a retrospective analysis of existing data, participants were not required to supply informed consent.
All authors have read and reviewed the manuscript and consent to the publication of the manuscript.

Competing interests

LF serves as a medical consulter for General Electric.
Open AccessThis article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://​creativecommons.​org/​licenses/​by/​4.​0/​. The Creative Commons Public Domain Dedication waiver (http://​creativecommons.​org/​publicdomain/​zero/​1.​0/​) applies to the data made available in this article, unless otherwise stated in a credit line to the data.

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Literatur
1.
Zurück zum Zitat Bouhemad B, Zhang M, Lu Q, Rouby JJ. Clinical review: bedside lung ultrasound in critical care practice. Crit Care. 2007;11:205.CrossRef Bouhemad B, Zhang M, Lu Q, Rouby JJ. Clinical review: bedside lung ultrasound in critical care practice. Crit Care. 2007;11:205.CrossRef
2.
Zurück zum Zitat Buonsenso D, Pata D, Chiaretti A. COVID-19 outbreak: less stethoscope, more ultrasound. Lancet Respir. 2020;8:e27.CrossRef Buonsenso D, Pata D, Chiaretti A. COVID-19 outbreak: less stethoscope, more ultrasound. Lancet Respir. 2020;8:e27.CrossRef
3.
Zurück zum Zitat Platz E, Pivetta E, Merz AA, Peck J, Rivero J, Cheng S. Impact of device selection and clip duration on lung ultrasound assessment in patients with heart failure. Am J Emerg Med. 2015;33:1552.CrossRef Platz E, Pivetta E, Merz AA, Peck J, Rivero J, Cheng S. Impact of device selection and clip duration on lung ultrasound assessment in patients with heart failure. Am J Emerg Med. 2015;33:1552.CrossRef
4.
Zurück zum Zitat Dietrich CF, Mathis G, Blaivas M, Volpicelli G, Seibel A, Wastl D, et al. Lung B-line artefacts and their use. J Thorac Dis. 2016;8:1356–65.CrossRef Dietrich CF, Mathis G, Blaivas M, Volpicelli G, Seibel A, Wastl D, et al. Lung B-line artefacts and their use. J Thorac Dis. 2016;8:1356–65.CrossRef
Metadaten
Titel
Feasibility of machine integrated point of care lung ultrasound automatic B-lines tool in the Corona-virus 2019 critical care unit
verfasst von
Gal Tsaban
Ori Galante
Yaniv Almog
Yuval Ullman
Lior Fuchs
Publikationsdatum
01.12.2021
Verlag
BioMed Central
Erschienen in
Critical Care / Ausgabe 1/2021
Elektronische ISSN: 1364-8535
DOI
https://doi.org/10.1186/s13054-021-03770-8

Weitere Artikel der Ausgabe 1/2021

Critical Care 1/2021 Zur Ausgabe

Häusliche Gewalt in der orthopädischen Notaufnahme oft nicht erkannt

28.05.2024 Traumatologische Notfälle Nachrichten

In der Notaufnahme wird die Chance, Opfer von häuslicher Gewalt zu identifizieren, von Orthopäden und Orthopädinnen offenbar zu wenig genutzt. Darauf deuten die Ergebnisse einer Fragebogenstudie an der Sahlgrenska-Universität in Schweden hin.

Fehlerkultur in der Medizin – Offenheit zählt!

28.05.2024 Fehlerkultur Podcast

Darüber reden und aus Fehlern lernen, sollte das Motto in der Medizin lauten. Und zwar nicht nur im Sinne der Patientensicherheit. Eine negative Fehlerkultur kann auch die Behandelnden ernsthaft krank machen, warnt Prof. Dr. Reinhard Strametz. Ein Plädoyer und ein Leitfaden für den offenen Umgang mit kritischen Ereignissen in Medizin und Pflege.

Mehr Frauen im OP – weniger postoperative Komplikationen

21.05.2024 Allgemeine Chirurgie Nachrichten

Ein Frauenanteil von mindestens einem Drittel im ärztlichen Op.-Team war in einer großen retrospektiven Studie aus Kanada mit einer signifikanten Reduktion der postoperativen Morbidität assoziiert.

„Übersichtlicher Wegweiser“: Lauterbachs umstrittener Klinik-Atlas ist online

17.05.2024 Klinik aktuell Nachrichten

Sie sei „ethisch geboten“, meint Gesundheitsminister Karl Lauterbach: mehr Transparenz über die Qualität von Klinikbehandlungen. Um sie abzubilden, lässt er gegen den Widerstand vieler Länder einen virtuellen Klinik-Atlas freischalten.

Update AINS

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.