Faster pediatric 3-T abdominal magnetic resonance imaging: comparison between conventional and variable refocusing flip-angle single-shot fast spin-echo sequences

verfasst von: Nichanan Ruangwattanapaisarn, Andreas M. Loening, Manojkumar Saranathan, Daniel V. Litwiller, Shreyas S. Vasanawala

Erschienen in: Pediatric Radiology | Ausgabe 6/2015

Einloggen, um Zugang zu erhalten

Abstract

Background

Single-shot fast spin echo (SSFSE) is particularly appealing in pediatric patients because of its motion robustness. However radiofrequency energy deposition at 3 tesla forces long pauses between slices, leading to longer scans, longer breath-holds and more between-slice motion.

Objective

We sought to learn whether modulation of the SSFSE refocusing flip-angle train could reduce radiofrequency energy deposition without degrading image quality, thereby reducing inter-slice pauses and overall scan times.

Materials and methods

We modulated the refocusing flip-angle train for SSFSE to minimize energy deposition while minimizing blurring and motion-related signal loss. In a cohort of 50 consecutive patients (25 boys, mean age 5.5 years, range 1 month to 17 years) referred for abdominal MRI we obtained standard SSFSE and variable refocusing flip-angle (vrfSSFSE) images and recorded sequence scan times. Two readers independently scored the images in blinded, randomized order for noise, tissue contrast, sharpness, artifacts and left lobe hepatic signal uniformity on a four-point scale. The null hypothesis of no difference between SSFSE and vrfSSFSE image-quality was assessed with a Mann–Whitney U test, and the null hypothesis of no scan time difference was assessed with the paired t-test.

Results

SSFSE and vrfSSFSE mean acquisition times were 54.3 and 26.2 s, respectively (P-value <0.0001). For each reader, SSFSE and vrfSSFSE noise, tissue contrast, sharpness and artifacts were not significantly different (P-values 0.18–0.86). However, SSFSE had better left lobe hepatic signal uniformity (P < 0.01, both readers).

Conclusion

vrfSSFSE is twice as fast as SSFSE, with equivalent image quality with the exception of left hepatic lobe signal heterogeneity.
Literatur
1.
Zurück zum Zitat Brenner DJ (2002) Estimating cancer risks from pediatric CT: going from the qualitative to the quantitative. Pediatr Radiol 32:228–231, discussion 228–231 Brenner DJ (2002) Estimating cancer risks from pediatric CT: going from the qualitative to the quantitative. Pediatr Radiol 32:228–231, discussion 228–231
2.
3.
Zurück zum Zitat Semelka RC, Kelekis NL, Thomasson D et al (1996) Abdominal MR imaging using the HASTE sequence. Proceedings of the Society of Magnetic Resonance, New York, p 219 Semelka RC, Kelekis NL, Thomasson D et al (1996) Abdominal MR imaging using the HASTE sequence. Proceedings of the Society of Magnetic Resonance, New York, p 219
4.
Zurück zum Zitat Hennig J, Weigel M, Scheffler K (2003) Multiecho sequences with variable refocusing flip angles: optimization of signal behavior using smooth transitions between pseudo steady states (TRAPS). Magn Reson Med 49:527–535CrossRefPubMed Hennig J, Weigel M, Scheffler K (2003) Multiecho sequences with variable refocusing flip angles: optimization of signal behavior using smooth transitions between pseudo steady states (TRAPS). Magn Reson Med 49:527–535CrossRefPubMed
5.
Zurück zum Zitat Busse RF (2004) Reduced RF power without blurring: correcting for modulation of refocusing flip angle in FSE sequences. Magn Reson Med 51:1031–1037CrossRefPubMed Busse RF (2004) Reduced RF power without blurring: correcting for modulation of refocusing flip angle in FSE sequences. Magn Reson Med 51:1031–1037CrossRefPubMed
6.
Zurück zum Zitat Busse RF, Brau AC, Vu A et al (2008) Effects of refocusing flip angle modulation and view ordering in 3D fast spin echo. Magn Reson Med 60:640–649PubMedCentralCrossRefPubMed Busse RF, Brau AC, Vu A et al (2008) Effects of refocusing flip angle modulation and view ordering in 3D fast spin echo. Magn Reson Med 60:640–649PubMedCentralCrossRefPubMed
7.
Zurück zum Zitat Loening A, Saranathan M, Ruangwattanapaisarn N et al (2014) Flip angle modulation in single shot fast spin echo imaging greatly increases speed with little change in diagnostic image quality. Proceedings of the Joint Annual Meeting ISMRM-ESMRMB, Milan, p 2174 Loening A, Saranathan M, Ruangwattanapaisarn N et al (2014) Flip angle modulation in single shot fast spin echo imaging greatly increases speed with little change in diagnostic image quality. Proceedings of the Joint Annual Meeting ISMRM-ESMRMB, Milan, p 2174
8.
Zurück zum Zitat Saranathan M, Loening AM, Litwiller DV et al (2014) Optimized refocusing flip angles for efficient single-shot fast spin echo imaging. Proceedings of the Joint Annual Meeting ISMRM-ESMRMB, Milan, p 2175 Saranathan M, Loening AM, Litwiller DV et al (2014) Optimized refocusing flip angles for efficient single-shot fast spin echo imaging. Proceedings of the Joint Annual Meeting ISMRM-ESMRMB, Milan, p 2175
11.
Zurück zum Zitat Madhuranthakam AJ, Busse RF, Brittain JH et al (2007) Sensitivity of low flip angle SSFSE of the abdomen to cardiac motion. Proceedings of the ISMRM, Berlin, p 15 Madhuranthakam AJ, Busse RF, Brittain JH et al (2007) Sensitivity of low flip angle SSFSE of the abdomen to cardiac motion. Proceedings of the ISMRM, Berlin, p 15
12.
Zurück zum Zitat Madhuranthakam AJ, Yu H, Shimakawa A et al (2010) T(2)-weighted 3D fast spin echo imaging with water-fat separation in a single acquisition. J Magn Reson Imaging 32:745–751PubMedCentralCrossRefPubMed Madhuranthakam AJ, Yu H, Shimakawa A et al (2010) T(2)-weighted 3D fast spin echo imaging with water-fat separation in a single acquisition. J Magn Reson Imaging 32:745–751PubMedCentralCrossRefPubMed
13.
Zurück zum Zitat Reeder SB, Wintersperger BJ, Dietrich O et al (2005) Practical approaches to the evaluation of signal-to-noise ratio performance with parallel imaging: application with cardiac imaging and a 32-channel cardiac coil. Magn Reson Med 54:748–754CrossRefPubMed Reeder SB, Wintersperger BJ, Dietrich O et al (2005) Practical approaches to the evaluation of signal-to-noise ratio performance with parallel imaging: application with cardiac imaging and a 32-channel cardiac coil. Magn Reson Med 54:748–754CrossRefPubMed
14.
Zurück zum Zitat Seifert F, Wubbeler G, Junge S et al (2007) Patient safety concept for multichannel transmit coils. J Magn Reson Imaging 26:1315–1321CrossRefPubMed Seifert F, Wubbeler G, Junge S et al (2007) Patient safety concept for multichannel transmit coils. J Magn Reson Imaging 26:1315–1321CrossRefPubMed
Metadaten
Titel
Faster pediatric 3-T abdominal magnetic resonance imaging: comparison between conventional and variable refocusing flip-angle single-shot fast spin-echo sequences
verfasst von
Nichanan Ruangwattanapaisarn
Andreas M. Loening
Manojkumar Saranathan
Daniel V. Litwiller
Shreyas S. Vasanawala
Publikationsdatum
01.06.2015
Verlag
Springer Berlin Heidelberg
Erschienen in
Pediatric Radiology / Ausgabe 6/2015
Print ISSN: 0301-0449
Elektronische ISSN: 1432-1998
DOI
https://doi.org/10.1007/s00247-014-3227-2

Weitere Artikel der Ausgabe 6/2015

Pediatric Radiology 6/2015 Zur Ausgabe

Hermes

Hermes

Endlich: Zi zeigt, mit welchen PVS Praxen zufrieden sind

IT für Ärzte Nachrichten

Darauf haben viele Praxen gewartet: Das Zi hat eine Liste von Praxisverwaltungssystemen veröffentlicht, die von Nutzern positiv bewertet werden. Eine gute Grundlage für wechselwillige Ärzte und Psychotherapeuten.

Akuter Schwindel: Wann lohnt sich eine MRT?

28.04.2024 Schwindel Nachrichten

Akuter Schwindel stellt oft eine diagnostische Herausforderung dar. Wie nützlich dabei eine MRT ist, hat eine Studie aus Finnland untersucht. Immerhin einer von sechs Patienten wurde mit akutem ischämischem Schlaganfall diagnostiziert.

Screening-Mammografie offenbart erhöhtes Herz-Kreislauf-Risiko

26.04.2024 Mammografie Nachrichten

Routinemäßige Mammografien helfen, Brustkrebs frühzeitig zu erkennen. Anhand der Röntgenuntersuchung lassen sich aber auch kardiovaskuläre Risikopatientinnen identifizieren. Als zuverlässiger Anhaltspunkt gilt die Verkalkung der Brustarterien.

S3-Leitlinie zu Pankreaskrebs aktualisiert

23.04.2024 Pankreaskarzinom Nachrichten

Die Empfehlungen zur Therapie des Pankreaskarzinoms wurden um zwei Off-Label-Anwendungen erweitert. Und auch im Bereich der Früherkennung gibt es Aktualisierungen.

Update Radiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.