Erschienen in: Journal of Medical Ultrasonics 1/2021

Open Access 03.01.2021 | Original Article–Breast & Thyroid

B-mode ultrasound diagnostic flowchart for solid breast masses: JABTS BC-01 study

verfasst von: Takanori Watanabe, Takuhiro Yamaguchi, Eriko Tohno, Hiroko Tsunoda, Setsuko Kaoku, Kanako Ban, Ryoji Watanabe, Takeshi Umemoto, Koichi Hirokaga, Toshikazu Ito

Erschienen in: Journal of Medical Ultrasonics | Ausgabe 1/2021

Abstract

Purpose

Breast ultrasound has been widely used as an essential examination for diagnosing breast cancer. However, standardized diagnostic criteria are as yet lacking. This study aimed to develop a simple diagnostic flowchart for beginners learning breast ultrasonography. The diagnostic flowchart was developed based on the recall criteria widely used in Japan.

Methods

We conducted a multicenter study to examine recall criteria usefulness in the diagnostic phase of breast disease. Women with ultrasound-visible breast masses who underwent B-mode breast ultrasound examination were recruited from 22 hospitals in Japan between September 2009 and January 2010. B-mode images were evaluated by members of the centralized image interpretation committee. We developed the new diagnostic flowchart based on the results. The usefulness of the diagnostic flowchart was assessed by employing datasets from the current study and another study which we conducted (BC-04 study).

Results

We evaluated 1045 solid masses (malignant: 495, benign: 550). Multivariate analysis showed that shape, margin, echogenic halo, interruption of the mammary gland interface, and depth width ratio were significant findings for distinguishing between benign and malignant masses. We modified the recall criteria and developed our novel diagnostic flowchart using these findings. The sensitivity and specificity of the new flowchart (current study: 0.97, 0.45; BC-04 study dataset: 0.95, 0.45) were similar to those of experts (current study: 0.96, 0.54; BC-04 study dataset: 0.98, 0.38).

Conclusion

We developed a simple diagnostic flowchart for breast ultrasound. This flowchart is anticipated to be applicable to educating beginners learning breast ultrasound.
Hinweise

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Introduction

Breast ultrasound has improved remarkably due to advances in imaging technologies, such as tissue harmonic imaging, spatial compounding, Doppler ultrasound, and elastography. These advances have improved the ability to make an accurate differential diagnosis between benign and malignant lesions. However, diagnostic criteria for breast masses have yet to be standardized.
Breast ultrasound and mammography have been widely used as essential examinations for diagnosing breast cancer in Japan. The Japan Association of Breast and Thyroid Sonology (JABTS) was established in 1998. In 2004, JABTS published the Guidelines for Breast Ultrasound Diagnosis (1st edition) [1]. In the Guidelines, we proposed a diagnostic flowchart for breast masses using the recall criteria for ultrasound breast cancer screening. This flowchart and the criteria were developed based on the opinions of breast ultrasound experts. The recall criteria are very simple, and were developed for breast ultrasound screening. The diagnostic flowchart was developed for breast ultrasound diagnosis. However, the diagnostic flowchart was complex and difficult to remember. As a result, the diagnostic flowchart is not now in widespread use in Japan, although the recall criteria are widely used. There is, however, a problem with using the recall criteria for diagnosing breast masses. Since biopsy is performed to confirm the final diagnosis of breast cancer, it is important to decide whether to recommend a biopsy or observation at the time of breast ultrasound diagnosis. The recall criteria are not applicable to making this judgment. Therefore, we cannot use the recall criteria for this purpose. The aim of this study was to develop a simple new diagnostic flowchart for solid breast masses to facilitate the decision as to whether biopsy or observation should be recommended. We conducted a multicenter study and obtained findings useful for distinguishing between benign and malignant masses. Based on the results, we developed the novel diagnostic flowchart presented herein. To evaluate the usefulness of this new diagnostic flowchart, we employed a patient dataset from another of our multicenter studies in addition to the data obtained in this study.

Materials and methods

Data collection

Women with an ultrasound-visible breast mass who underwent B-mode breast ultrasound examination were recruited from 22 hospitals in Japan between September 2009 and January 2010. Ultrasound units with linear transducers exceeding 10 MHz were used in this study. Exclusion criteria were as follows: 1. simple cysts, 2. lesions already being followed by ultrasound, 3. lesions subjected to vacuum-assisted biopsy at another hospital, 4. masses larger than 5 cm in maximum diameter. Biopsy or observation was selected according to the routine clinical practices of each hospital. Lesions with no significant change during the 2 years of observation were regarded as being benign in this study. Static B-mode digital images and histopathological data without personal information were collected at the clinical research data center at Tohoku University Hospital.
The institutional review board or the ethics committee at each hospital approved this prospective observational study. Written informed consent was not required in this trial according to the ethical guidelines for epidemiological research in Japan [2]. There are two reasons for this. First, this trial did not use human biological specimens. Second, B-mode ultrasound is conducted as a routine examination for breast cancer diagnosis. However, public disclosure of information obtained in this study is required by all participating hospitals. When a patient refused to allow use of their clinical data, their data were not used.

Centralized image interpretation committee

Static B-mode digital images were evaluated by members of the centralized image interpretation committee comprised of 26 specialists with no knowledge of the clinical information (except for age) or the histopathological data. These 26 breast ultrasound specialists working in Japan included three radiologists, 19 breast surgeons, and four ultrasonographers. All were members of the Terminology and Diagnostic Criteria Committee of the JABTS. The 26 ultrasound specialists were divided into 13 pairs. Pairs of specialists evaluated each of the ultrasound images. If interpretation was difficult, the images were discussed by all members of the committee. The quality of liquid crystal image displays used for the centralized image interpretation was confirmed by TG18-QC pattern (American Association of Physicists in Medicine) [3]. The ultrasound findings and categories of each mass were reported by the centralized image interpretation committee. After evaluation of findings, such as shape, margin (Fig. 1a), internal echoes, posterior echoes, depth/width ratio (DW ratio, Fig. 1b), echogenic halo (echogenic rim, Fig. 1c), and interruption of the mammary gland interface (Fig. 1d), the B-mode category was determined by consensus.

Japanese category

In Japan, we use Japanese categories; C1: normal, C2: benign, C3a: probably benign (observation is recommended), C3b: probably benign (biopsy is recommended), C4: suspicion of malignancy, C5: malignant [1]. Japanese categories differ from those of the Breast Imaging Reporting and Data System (BI-RADS) [4]. Japanese C3b corresponds to BI-RADS category 4A (biopsy is recommended) (Table 1).
Table 1
Correspondence between Japanese and BI-RADS categories
Japanese Category
BI-RADS Category
1
1
2
2
3a
3
3b
4A
4
4B, 4C
5
5
BI-RADS breast imaging reporting and data system

The recall criteria for ultrasound breast cancer screening

The recall criteria for ultrasound breast cancer screening include criteria pertaining to breast masses (simple cysts, complex cystic and solid masses) and breast non-mass lesions [1, 5]. In this study, we focused only on solid breast masses, and we developed a new diagnostic flowchart for solid breast masses based on the recall criteria. Figure 2 shows the recall criteria pertaining to solid masses (2004 version) [6]. Herein, we show a slightly simplified version of the original recall criteria to enhance understanding. The criteria are divided into three sections. The first section categorizes obviously benign masses (fibroadenomas) as C2. Typical fibroadenomas have an oval shape, circumscribed margin, diameter less than 2 cm, and a very low DW ratio. In this study, we defined “very low DW ratio” as less than 0.5. Typical calcified fibroadenomas are characterized by coarse calcifications. The second section categorizes obviously malignant or highly suspicious masses as C4 or C5. Typical malignant masses show an echogenic halo and/or interruption of the mammary gland interface [7]. Masses with a high possibility of malignancy show echogenic foci within the mass. In the third section, the remaining masses are categorized as C2, 3, or 4 according to the size and DW ratio. Since the third section of the recall criteria does not distinguish between C3a (observation) and C3b (biopsy), it cannot serve to determine whether biopsy or observation should be recommended. In 2014, an item pertaining to complicated cysts was added to the first section of the recall criteria [1, 5]. However, since the B-mode images were evaluated before 2014, this item was not included in the present study.

Statistical analysis

Data collection and statistical analyses were conducted by the Clinical Research Data Center of Tohoku University Hospital. Statistical analyses were conducted using SAS Version 9.4 (SAS Institute, Inc., Cary, NC, USA). Univariate analysis was conducted using chi-square tests. Multivariate analysis was conducted using logistic regression.

New diagnostic flowchart for solid masses

Based on the results of the statistical analysis, the Terminology and Diagnostic Criteria Committee of the JABTS endeavored to develop a new diagnostic flowchart. To facilitate user recollection of the essential points, we developed this new diagnostic flowchart based on the recall criteria already widely used in Japan.

Verification of the usefulness of the new diagnostic flowchart for solid masses

The usefulness of our novel diagnostic flowchart for solid masses was evaluated by comparing the sensitivity and specificity determined by experts with those determined based on the new diagnostic flowchart.
We emphasize the importance of using a dataset from a different patient population, in addition to the one obtained in this study, to allow comparative evaluation of the usefulness of the new diagnostic flowchart. Therefore, in addition to data from the current study (JABTS BC-01), we employed the patient dataset from our JABTS BC-04 study [8]. The JABTS BC-04 study aimed to develop diagnostic criteria for color Doppler examination of solid masses in the breast, and was conducted from 2013 to 2017. The dataset from the JABTS BC-04 study included 839 malignant masses and 569 benign masses.
The datasets from the current (JABTS BC-01) study and the JABTS BC-04 study include findings and categories of solid masses determined by the centralized image interpretation committee. Categories of the new diagnostic flowchart were mechanically converted using the findings contained in the datasets. As a result, there were two categories for each mass; one determined by a specialist and the other based on the new diagnostic flowchart. We calculated and then compared sensitivity and specificity using these two categories. For statistical analyses of sensitivity and specificity, Japanese categories 2 and 3a (only observation is recommended) were considered to be negative, while categories 3b, 4, and 5 (biopsy is recommended) were taken to be positive.

Study registration

The JABTS BC-01 study is registered with the University Hospital Medical Information Network, Japan (No. UMIN000007603).

Results

Between September 2009 and January 2010, 1412 ultrasound-visible breast masses were registered from 22 hospitals. Final enquiries regarding the histopathology and clinical observations were conducted in March 2014. Of the 1412 masses, six were excluded due to patient withdrawal, four due to missing data, two due to being simple cysts, and 18 due to being unevaluable by the centralized image interpretation committee because of inadequate image quality. Three hundred and five (55.3%) of 551 observational masses lacked 2-year observation results. Of the remaining 1077 masses, 1045 were solid. Since the number of mixed masses was only 32, we evaluated the 1045 solid masses (malignant: 495 (468 patients), benign: 550 (459 patients)) in this study (Fig. 3). Mean sizes of malignant and benign masses were 1.6 ± 0.78 cm (0.3–4.4) and 1.2 ± 0.71 cm (0.3–5.5), respectively. The ages of the 468 patients with malignant mases and the 459 patients with benign masses were 56.8 ± 12.5 years (mean ± standard deviation, range: 30–95) and 45.1 ± 11.9 years (13–76), respectively. The histopathological results of the 1045 masses are shown in Table 2. Biopsy was performed for 799 masses, of which 495 were malignant and 304 were benign. Invasive carcinoma of no special type accounted for 80% of malignant masses and ductal carcinoma in situ accounted for 10%.
Table 2
Histopathological results (n = 1045)
Malignant (n = 495)
 Ductal carcinoma in situ
47
 Invasive carcinoma of no special type
402
 Invasive lobular carcinoma
20
 Mucinous carcinoma
18
 Carcinoma with apocrine differentiation
2
 Tubular carcinoma
2
 Other (malignant)
4
Benign (n = 550)
 Fibroadenoma
115
 Fibrocystic disease
71
 Intraductal papilloma
28
 Complicated cyst
25
 Phyllodes tumor
8
 Fibrosis
7
 Ductal hyperplasia
5
 Tubular adenoma
2
 Other (benign)
43
 Unknown (benign)a
246
a2-year observation completed

Utility of the recall criteria as a diagnostic flowchart

We evaluated the usefulness of the recall criteria as a diagnostic flowchart. In the first and second sections of the recall criteria, 12.5% (69/550) of benign masses and 87.1% (431/495) of malignant masses were detected. Table 3 shows the malignancy rate in the first and second sections of the recall criteria. Regarding circumscribed masses with a very low DW ratio (diameter less than 2 cm), 98.4% (61/62) were benign, and 89% (8/9) of masses with coarse calcifications were benign. Furthermore, 97.5% (153/157) of masses with an echogenic halo, 90.8% (246/271) showing interruption of the mammary gland interface, and 59.3% (32/54) with multiple echogenic foci were malignant. Since the third section of the recall criteria for solid masses cannot be used in a diagnostic flowchart, we did not include it in the present evaluation.
Table 3
Malignancy rates in lesions found to be benign or malignant based on JABTS Recall criteria
Findings
Malignancy rate
Obvious benign findings (C2)
 Circumscribed mass with a very low DW ratio (diameter less than 2 cm)
1.6% (1/62)
 Coarse calcification
11.1% (1/9)
Obvious malignant findings (C4, 5)
 Echogenic halo
97.5% (153/157)
 Interruption of mammary gland interface
90.8% (246/271)
 Multiple echogenic foci
59.3% (32/54)

Frequencies of malignant and benign masses according to the ultrasound findings (Table 4)

Table 4
Frequency of malignant and benign masses according to US findings (n = 1045)
Findings
Number of cases
Malignant
Benign
% Malignant
% Benign
Shape
 Oval
313
33
280
10.5%
89.5%
 Round
56
25
31
44.6%
55.4%
 Lobulated
220
104
116
47.3%
52.7%
 (2 or 3 undulations)
141
50
-91
35.5%
64.5%
 (More than 4 undulations)
73
48
-25
65.8%
34.2%
 Polygonal
47
22
25
46.8%
53.2%
 Irregular
409
311
98
76.0%
24.0%
DW ratio
  < 0.7
616
206
410
33.4%
66.6%
 (< 0.5)
(216)
(50)
(166)
(23.1%)
(76.9%)
 (0.5–0.7)
(400)
(156)
(244)
(39.0%)
(61.0%)
  ≥ 0.7
429
289
140
67.4%
32.6%
Margin
 Circumscribed
337
30
307
8.9%
91.1%
 Well-defined and rough
486
298
188
61.3%
38.7%
 Indistinct
220
166
54
75.5%
24.5%
 Obscure
2
1
1
50.0%
50.0%
Echogenic halo
 Present
157
153
4
97.5%
2.5%
 Absent
888
342
546
38.5%
61.5%
Internal echoes
 Homogeneity
  Homogeneous
408
104
304
25.5%
74.5%
  Heterogeneous
637
391
246
61.4%
38.6%
 Echo level
  Anechoic
0
0
0
0.0%
0.0%
  Hypoechoic
920
461
459
50.1%
49.9%
  Isoechoic
119
31
88
26.1%
73.9%
  Hyperechoic
6
3
3
50.0%
50.0%
 Echogenic foci
  Present
97
73
24
75.3%
24.7%
  Absent
948
422
526
44.5%
55.5%
 Coarse calcifications
  Present
9
1
8
11.1%
88.9%
  Absent
1036
494
542
47.7%
52.3%
 Posterior echoes
  Accentuating
323
148
175
45.8%
54.2%
  Not changing
551
214
337
38.8%
61.2%
  Attenuating
168
132
36
78.6%
21.4%
  Shadowing
3
1
2
33.3%
66.7%
 Interruption of mammary gland interface
  Present
271
246
25
90.8%
9.2%
  Absent
774
249
525
32.2%
67.8%
As to shape, 89.5% of oval masses were benign, and 76.0% of irregular masses were malignant. Regarding the DW ratio, 66.6% of the masses with a DW ratio less than 0.7 were benign, and 67.4% of those with a DW ratio of at least 0.7 were malignant. As to the margin, 91.1% of circumscribed masses were benign, and 75.5% with indistinct margins were malignant. Furthermore, 97.5% of the masses with an echogenic halo and 90.8% of those showing interruption of the mammary gland interface were malignant. Regarding echogenic foci, 75.3% of the masses with echogenic foci were malignant.

Multivariate analysis

Multivariate analysis showed that shape, DW ratio, margin, echogenic halo, and interruption of the mammary gland interface were significant findings for distinguishing between benign and malignant masses (Table 5).
Table 5
Multivariate analysis of B-mode features and benign/malignant differential diagnosis (logistic regression)
Findings
Adjusted odds ratio
95% Confidence interval
p
Shape
 Round vs. oval
3.34
1.41
7.93
0.0062
 Polygonal vs. oval
2.04
0.79
5.29
0.141
 Irregular vs. oval
4.07
2.16
7.65
 < .0001
 Lobulated vs. oval
2.82
1.52
5.25
0.001
 DW ratio
  ≥ 0.7 vs < 0.7
2.64
1.7
4.1
 < .0001
Margin
 Indistinct vs. circumscribed
3.38
1.62
7.04
0.0012
 Well-defined and rough vs. circumscribed
5.3
3
9.39
 < .0001
Echogenic halo
 Present vs. absent
9.82
2.79
34.57
0.0004
Internal echoes
 Heterogeneous vs. homogeneous
1.18
0.74
1.86
0.49
Echo level
 Severely hypoechoic vs. hypoechoic
0.6
0.15
2.39
0.47
 Isoechoic vs. hypoechoic
0.58
0.3
1.12
0.1
 Hyperechoic vs. hypoechoic
1.18
0.07
18.89
0.91
Echogenic foci or coarse calcification
 Echogenic foci vs. absent
1.74
0.89
3.42
0.11
 Coarse calcification vs absent
0.18
0.03
1.11
0.06
Posterior echoes
 Accentuating vs. not changing
1.37
0.88
2.16
0.17
 Attenuating vs. not changing
0.98
0.52
1.87
0.96
 Shadowing vs. not changing
2.5
0.06
114.43
0.64
Interruption of mammary gland interface
 Present vs. absent
3.1
1.72
5.57
0.0002

The new diagnostic flowchart for solid masses (Fig. 4)

Using these results, the Terminology and Diagnostic Criteria Committee of the JABTS discussed and proposed a new diagnostic flowchart based on the recall criteria. Since the first and second sections of the recall criteria were demonstrated to be very useful, we applied them as the first and second sections of the new diagnostic flowchart. Then, we developed the third section of the new diagnostic flowchart. Of the five significant findings for distinguishing between benign and malignant masses by multivariate analysis, echogenic halo and interruption of the mammary gland interface were included in the second section. Therefore, we used shape, margin, and DW ratio in the third section. Among these parameters, those raising suspicion of malignancy were irregular shape, well-defined and rough/indistinct margin, and DW ratio ≥ 0.7 (Fig. 4). If none of these three suspicious findings is present, the category is determined to be 3a, and if at least one is present, the category is 3b. The committee proposed a new diagnostic flowchart for solid masses in May of 2015.

Sensitivity and specificity of the new diagnostic flowchart

We calculated sensitivity and specificity using the datasets from the current study (JABTS BC-01) and the JABTS BC-04 study. Details of the BC-04 study dataset are shown in Table 6. With the current study dataset, sensitivity and specificity of the diagnostic flowchart were 0.97 and 0.45, respectively. The sensitivity and specificity based on the evaluations performed by the specialists were 0.96 and 0.54, respectively (Table 7). When we used the BC-04 study dataset, the respective sensitivity and specificity of the diagnostic flowchart were 0.95 and 0.45, while the corresponding sensitivity and specificity for the specialists were 0.98 and 0.38.
Table 6
BC-04 study dataset (n = 1408)
Number of masses
 
 Malignant: 839 (818 patients)
 Benign: 569 (547 patients)
Age (mean ± standard deviation, range)
 Malignant (818 patients): 57.7 ± 13.7 (25–96)
 Benign (547 patients): 44.8 ± 12.8 (12–87)
Mean sizes
 Malignant: 1.7 ± 0.84 cm (0.4–4.8)
 Benign: 1.5 ± 0.92 cm (0.3–4.8)
Histopathological results
 Malignant (839)
  Ductal carcinoma in situ
53
  Invasive carcinoma of no special type
716
  Invasive lobular carcinoma
29
  Mucinous carcinoma
22
  Carcinoma with apocrine differentiation
5
  Tubular carcinoma
1
  Other (malignant)
18
 Benign (569)
  Fibroadenoma
179
  Fibrocystic disease
104
  Intraductal papilloma
43
  Complicated cyst
3
  Phyllodes tumor
29
  Ductal hyperplasia
6
  Tubular adenoma
1
  Other (benign)
29
  Unknown (benign)a
175
a2-year observation completed
Table 7
Sensitivity and specificity using new diagnostic flowchart vs experienced specialists (C2, 3a vs. C3b, 4, 5)
 
Datasets
Current study (JABTS BC-01) (n = 1045a)
JABTS BC-04 study (n = 1408b)
Sensitivity
Specificity
Sensitivity
Specificity
New diagnostic flowchart
0.97
0.45
0.95
0.45
Specialists
0.96
0.54
0.98
0.38
aMalignant: 495, benign: 550
bMalignant: 839, benign: 569

Discussion

Ultrasonography is very useful for diagnosing breast cancer. There has been significant progress in distinguishing between malignant and benign masses using ultrasound since the early 1990s [912]. Several sonographic features based on shape, margin, and echo texture have been proposed for the diagnosis of breast masses [10, 13, 14]. At present, the BI-RADS classification is widely used globally [4]. However, BI-RADS does not include ultrasound diagnostic criteria. Several studies have attempted to develop ultrasound diagnostic criteria [15, 16], but diagnostic criteria have yet to be standardized.
In Japan, JABTS developed a diagnostic flowchart for breast masses in 2004 [6]. The diagnostic flowchart was developed after 4 years of discussions among experts. However, the flowchart was complex and therefore did not come into widespread use in Japan. On the other hand, the recall criteria for ultrasound breast cancer screening are very simple and widely used in Japan for ultrasound screening of breast cancer [5]. It would be optimal if the recall criteria could be used as a diagnostic flowchart. However, the recall criteria cannot be used to decide whether to recommend a biopsy or observation, and thus cannot be applied in diagnostic flowchart form. Therefore, we needed to develop a novel diagnostic flowchart. To encourage widespread use in Japan, we aimed to simplify the new diagnostic flowchart and to apply the recall criteria already widely used in Japan.
The recall criteria were proposed by JABTS in 2004 [6], and a revised version was published in 2016 [5]. Clearly or typically benign (fibroadenoma) and malignant masses are identified by applying the first and second portions of the criteria. Several reports have described fibroadenoma findings [1720]. These findings are oval shape, smooth and well-circumscribed margin, and low DW ratio. The DW ratio was reportedly less than 0.7 in 86% of fibroadenomas and, furthermore, fibroadenomas usually stopped growing at a size of 2–3 cm [18]. According to the recall criteria, typical fibroadenomas are defined as oval circumscribed masses less than 2 cm in diameter, with a very low DW ratio. In this study, 1.6% (1/62) of masses judged to be typical fibroadenomas were ultimately found to be malignant.
The two typically used malignant findings are echogenic halo and interruption of the mammary gland interface. Echogenic halo has been reported to have high predictive value for malignancy by multiple research groups [13, 14, 16]. Interruption of the mammary gland interface was originally proposed by Konishi [7] in 1988. This finding has been used in Japan since the late twentieth century [21]. Our present study showed that this finding had high predictive value for malignancy. However, the predictive value of interruption of the mammary gland interface was slightly lower (90.8%) than that of echogenic halo (97.5%). Applying a combination of these two findings resulted in approximately half of breast cancers being interpreted as malignant. Features typical of malignancy (C4, 5) and of fibroadenoma (C2) in the recall criteria had high diagnostic utility (Table 2). Therefore, we decided to incorporate the first and second sections of the recall criteria into the new diagnostic flowchart.
We also examined the applicability of the third section of the new diagnostic flowchart. We used three findings raising suspicion of malignancy, which multivariate analysis had shown to be useful for distinguishing between benign and malignant breast masses. We advocate performing a biopsy if any of these three suspicious findings is identified in breast masses with neither clearly typical benign nor malignant findings. The sensitivity and specificity of the new diagnostic flowchart using the dataset from the current study were 0.97 and 0.45, respectively. The corresponding sensitivity and specificity of the current study for the specialists (centralized image interpretation committee) were 0.96 and 0.54. Furthermore, we examined the usefulness of the new diagnostic flowchart using our dataset from the BC-04 study [8]. The sensitivity and specificity of the new diagnostic flowchart using the BC-04 study dataset were 0.95 and 0.45, respectively. The sensitivity and specificity for the specialists (centralized image interpretation committee) examining the BC-04 study were 0.98 and 0.38. The specificity of the new flowchart was thus slightly inferior to that of the experts, but the sensitivity was higher. These results indicate that the new diagnostic flowchart is applicable, at a minimum, to diagnostic flowchart use for beginners. This flowchart is just a first step for beginners learning breast ultrasound. As they gain experience, beginners can progress to the intermediate and more advanced skill levels. This flowchart may even serve as a gateway allowing beginners to become experts in performing diagnostic ultrasound examinations of the breast.
We anticipate that this flowchart will become more sophisticated with ongoing revisions. As an example, parameters, such as age, elastographic findings, and color Doppler imaging, could potentially be incorporated into the flowchart. We hope that this flowchart will be useful not only to medical specialists but also to patients.

Limitation

Our proposed diagnostic flowchart was developed based on expert-judged imaging findings data. This flowchart may not work well if the inter-observer agreement between the beginners and the experts in evaluating each finding is low. Therefore, there is a need for education for beginners using this flowchart.

Conclusion

In this study, we developed a simple diagnostic flowchart for B-mode breast ultrasound. This flowchart is anticipated to be applicable to educating beginners learning breast ultrasound.

Acknowledgements

We thank all of the patients who consented to the use of their ultrasound images for this study. We also thank all participating investigators and their support staff for their contributions to this study.

Compliance with ethical standards

Conflict of interest

The authors have no conflicts of interest to declare.

Ethical statements

The institutional review board or the ethics committee at each hospital approved this study.
Open AccessThis article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://​creativecommons.​org/​licenses/​by/​4.​0/​.

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Unsere Produktempfehlungen

e.Med Interdisziplinär

Kombi-Abonnement

Für Ihren Erfolg in Klinik und Praxis - Die beste Hilfe in Ihrem Arbeitsalltag

Mit e.Med Interdisziplinär erhalten Sie Zugang zu allen CME-Fortbildungen und Fachzeitschriften auf SpringerMedizin.de.

e.Med Radiologie

Kombi-Abonnement

Mit e.Med Radiologie erhalten Sie Zugang zu CME-Fortbildungen des Fachgebietes Radiologie, den Premium-Inhalten der radiologischen Fachzeitschriften, inklusive einer gedruckten Radiologie-Zeitschrift Ihrer Wahl.

Literatur
1.
Zurück zum Zitat Japan association of breast and thyroid sonology (JABTS). Guidelines for breast ultrasound: Management and diagnosis. 3rd edition Tokyo: Nankodo; 2014 [In Japanese]. Japan association of breast and thyroid sonology (JABTS). Guidelines for breast ultrasound: Management and diagnosis. 3rd edition Tokyo: Nankodo; 2014 [In Japanese].
3.
Zurück zum Zitat Samei E. Assessment of display performance for medical imaging systems: executive summary of AAPM TG 18 report. Med Phys. 2005;32:1205–25.CrossRef Samei E. Assessment of display performance for medical imaging systems: executive summary of AAPM TG 18 report. Med Phys. 2005;32:1205–25.CrossRef
4.
Zurück zum Zitat American College of Radiology. ACR BI-RADS atlas: Breast imaging reporting and data system. 5th ed. Reston, VA: Author; 2013. American College of Radiology. ACR BI-RADS atlas: Breast imaging reporting and data system. 5th ed. Reston, VA: Author; 2013.
5.
Zurück zum Zitat Terminology and diagnostic criteria committee of Japan society of ultrasonic in medicine. Recall criteria for ultrasound breast cancer screening. J Med Ultrasonics. 2016;43:301–13.CrossRef Terminology and diagnostic criteria committee of Japan society of ultrasonic in medicine. Recall criteria for ultrasound breast cancer screening. J Med Ultrasonics. 2016;43:301–13.CrossRef
6.
Zurück zum Zitat Japan Association of Breast and Thyroid Sonology (JABTS). Guidelines for breast ultrasound: Management and diagnosis. Tokyo:Nankodo; 2004 [In Japanese]. Japan Association of Breast and Thyroid Sonology (JABTS). Guidelines for breast ultrasound: Management and diagnosis. Tokyo:Nankodo; 2004 [In Japanese].
7.
Zurück zum Zitat Konishi Y, Ogata M, Kuroki T, Tocjiko H, Fujimoto T, Morimoro Y, et al. Assessment on Differential Diagnosis between Tumor-forming Mastopathy on Ultrasonography and T1 Breast Cancer. Jpn J Med Ultrasonics. 1988;15:44–51. Konishi Y, Ogata M, Kuroki T, Tocjiko H, Fujimoto T, Morimoro Y, et al. Assessment on Differential Diagnosis between Tumor-forming Mastopathy on Ultrasonography and T1 Breast Cancer. Jpn J Med Ultrasonics. 1988;15:44–51.
8.
Zurück zum Zitat Watanabe T, Kaoku S, Yamaguchi T, Izumori A, Konno S, Okuno T, et al. Multicenter prospective study of color Doppler ultrasound for breast masses: Utility of our color Doppler method. Ultrasound in Med Biol. 2019;45:1367–79.CrossRef Watanabe T, Kaoku S, Yamaguchi T, Izumori A, Konno S, Okuno T, et al. Multicenter prospective study of color Doppler ultrasound for breast masses: Utility of our color Doppler method. Ultrasound in Med Biol. 2019;45:1367–79.CrossRef
9.
Zurück zum Zitat Jackson VP, Reynolds HE, Hawes DR. Sonography of the breast. Semin Ultrasound CT MR. 1996;17:460–75.CrossRef Jackson VP, Reynolds HE, Hawes DR. Sonography of the breast. Semin Ultrasound CT MR. 1996;17:460–75.CrossRef
10.
Zurück zum Zitat Stavros AT, Thickman D, Rapp CL, Dennis MA, Parker SH, Sisney GA. Solid breast nodules: use of sonography to distinguish between benign and malignant lesions. Radiology. 1995;196:123–34.CrossRef Stavros AT, Thickman D, Rapp CL, Dennis MA, Parker SH, Sisney GA. Solid breast nodules: use of sonography to distinguish between benign and malignant lesions. Radiology. 1995;196:123–34.CrossRef
11.
Zurück zum Zitat Jackson VP. Management of solid breast nodules: what is the role of sonography? Radiology. 1995;196:14–5.CrossRef Jackson VP. Management of solid breast nodules: what is the role of sonography? Radiology. 1995;196:14–5.CrossRef
12.
Zurück zum Zitat Paulinelli RR, Freitas-Junior R, Moreira MA, Moraes VA, Bernardes-Júnior JR, Vidal CS, et al. Risk of malignancy in solid breast nodules according to their sonographic features. J ultrasound Med. 2005;24:635–41.CrossRef Paulinelli RR, Freitas-Junior R, Moreira MA, Moraes VA, Bernardes-Júnior JR, Vidal CS, et al. Risk of malignancy in solid breast nodules according to their sonographic features. J ultrasound Med. 2005;24:635–41.CrossRef
13.
Zurück zum Zitat Stavros AT. Breast ultrasound. In: Cynthia L. Rapp, Steve H. Parker, Editors. Philadelphia: Lippincott Williams & Wilkins; 2004. Stavros AT. Breast ultrasound. In: Cynthia L. Rapp, Steve H. Parker, Editors. Philadelphia: Lippincott Williams & Wilkins; 2004.
14.
Zurück zum Zitat Helmut M. The practice of breast ultrasound: techniques, findings, differential diagnosis. New York: Thieme; 2000. Helmut M. The practice of breast ultrasound: techniques, findings, differential diagnosis. New York: Thieme; 2000.
15.
Zurück zum Zitat Costantini M, Belli P, Lombardi R, Franceschini G, Mulè A, Bonomo L. Characterization of solid breast masses -use of the sonographic breast imaging reporting and data system lexicon-. J Ultrasound Med. 2006;25:649–59.CrossRef Costantini M, Belli P, Lombardi R, Franceschini G, Mulè A, Bonomo L. Characterization of solid breast masses -use of the sonographic breast imaging reporting and data system lexicon-. J Ultrasound Med. 2006;25:649–59.CrossRef
16.
Zurück zum Zitat Kim SJ, Ko EY, Shin JH, Kang SS, Mun SH, Han BK, et al. Application of sonographic BI-RADS to synchronous breast nodules detected in patients with breast cancer. Am J Roentgenol. 2008;191:653–8.CrossRef Kim SJ, Ko EY, Shin JH, Kang SS, Mun SH, Han BK, et al. Application of sonographic BI-RADS to synchronous breast nodules detected in patients with breast cancer. Am J Roentgenol. 2008;191:653–8.CrossRef
17.
Zurück zum Zitat Fornage BD, Lorigan JG, Andry E. Fibroadenoma of the breast: sonographic appearance. Radiology. 1989;172:671–5.CrossRef Fornage BD, Lorigan JG, Andry E. Fibroadenoma of the breast: sonographic appearance. Radiology. 1989;172:671–5.CrossRef
18.
Zurück zum Zitat Cole-Beuglet C, Soriano RZ, Kurts AB, Goldberg BB. Fibroadenoma of the breast: sonomammography correlated with pathology in 122 patients. AJR. 1983;140:369–75.CrossRef Cole-Beuglet C, Soriano RZ, Kurts AB, Goldberg BB. Fibroadenoma of the breast: sonomammography correlated with pathology in 122 patients. AJR. 1983;140:369–75.CrossRef
19.
Zurück zum Zitat Jackson VP, Rothschild PA, Kreipke DL, Mail JT, Holden RW. The spectrum of sonpgraphic findings of fibroadenoma of the breast. Invest Radiol. 1986;21:34–40.CrossRef Jackson VP, Rothschild PA, Kreipke DL, Mail JT, Holden RW. The spectrum of sonpgraphic findings of fibroadenoma of the breast. Invest Radiol. 1986;21:34–40.CrossRef
20.
Zurück zum Zitat Venta LA, Dudiak CM, Salomon CG, Flisak ME. Sonographic evaluation of the breast. RadioGraphics. 1994;14:29–50.CrossRef Venta LA, Dudiak CM, Salomon CG, Flisak ME. Sonographic evaluation of the breast. RadioGraphics. 1994;14:29–50.CrossRef
21.
Zurück zum Zitat Ueno E, Tohno E, Tsunoda H, Hirano M. Current status of ultrasound examination of the superficial organs (breast and thyroid). Rinsho Byori. 1993;41:15–25.PubMed Ueno E, Tohno E, Tsunoda H, Hirano M. Current status of ultrasound examination of the superficial organs (breast and thyroid). Rinsho Byori. 1993;41:15–25.PubMed
Metadaten
Titel
B-mode ultrasound diagnostic flowchart for solid breast masses: JABTS BC-01 study
verfasst von
Takanori Watanabe
Takuhiro Yamaguchi
Eriko Tohno
Hiroko Tsunoda
Setsuko Kaoku
Kanako Ban
Ryoji Watanabe
Takeshi Umemoto
Koichi Hirokaga
Toshikazu Ito
Publikationsdatum
03.01.2021
Verlag
Springer Singapore
Erschienen in
Journal of Medical Ultrasonics / Ausgabe 1/2021
Print ISSN: 1346-4523
Elektronische ISSN: 1613-2254
DOI
https://doi.org/10.1007/s10396-020-01072-0

Weitere Artikel der Ausgabe 1/2021

Journal of Medical Ultrasonics 1/2021 Zur Ausgabe

Akuter Schwindel: Wann lohnt sich eine MRT?

28.04.2024 Schwindel Nachrichten

Akuter Schwindel stellt oft eine diagnostische Herausforderung dar. Wie nützlich dabei eine MRT ist, hat eine Studie aus Finnland untersucht. Immerhin einer von sechs Patienten wurde mit akutem ischämischem Schlaganfall diagnostiziert.

Screening-Mammografie offenbart erhöhtes Herz-Kreislauf-Risiko

26.04.2024 Mammografie Nachrichten

Routinemäßige Mammografien helfen, Brustkrebs frühzeitig zu erkennen. Anhand der Röntgenuntersuchung lassen sich aber auch kardiovaskuläre Risikopatientinnen identifizieren. Als zuverlässiger Anhaltspunkt gilt die Verkalkung der Brustarterien.

S3-Leitlinie zu Pankreaskrebs aktualisiert

23.04.2024 Pankreaskarzinom Nachrichten

Die Empfehlungen zur Therapie des Pankreaskarzinoms wurden um zwei Off-Label-Anwendungen erweitert. Und auch im Bereich der Früherkennung gibt es Aktualisierungen.

Fünf Dinge, die im Kindernotfall besser zu unterlassen sind

18.04.2024 Pädiatrische Notfallmedizin Nachrichten

Im Choosing-Wisely-Programm, das für die deutsche Initiative „Klug entscheiden“ Pate gestanden hat, sind erstmals Empfehlungen zum Umgang mit Notfällen von Kindern erschienen. Fünf Dinge gilt es demnach zu vermeiden.

Update Radiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.