Open Access 01.12.2023 | Research

Ex vivo detection of mandibular incisors’ root canal morphology using cone-beam computed tomography with 4 different voxel sizes and micro-computed tomography

verfasst von: Bingbing Bai, Ying Tang, Yihan Wu, Fan Pei, Qi Zhu, Peng Zhu, Yongchun Gu

Erschienen in: BMC Oral Health | Ausgabe 1/2023

Abstract

Background

In recent years, cone-beam computed tomography (CBCT) has been widely used to evaluate patients’ root canal anatomy due to its high resolution and noninvasive nature. As voxel size is one of the most important parameters affecting CBCT image quality, the current study evaluated the diagnostic potential of CBCT with 4 different voxel sizes in the detection of double root canal systems and accessory canals (ACs) in permanent mandibular incisors.

Methods

A total of 106 extracted mandibular permanent incisors were collected from the dental clinics, and then were scanned by using micro-CT with a voxel size of 9 μm. The teeth were then fixed in the tooth sockets of human dry mandibles and scanned by using a CBCT device with 4 different voxel sizes (300, 200, 250, and 125 μm). Four observers detected in blind the root canal morphology of the teeth according to the CBCT images, and the presence or absence of a double root canal system, and the presence or absence of ACs, were scored according to a 5-point scale, respectively. Receiver operating characteristic (ROC) analysis was performed, and DeLong test was used to compare the area under the curve (AUC) values and the micro-CT data was taken as a gold standard.

Results

Among 106 sample teeth, 25 specimens with a double root canal system were identified by the micro-CT. ROC curve analysis of the data obtained by the four observers showed that in the detection of double root canal systems, the AUC values ranged from 0.765 to 0.889 for 300 μm voxel size, from 0.877 to 0.926 for 250 μm voxel size, from 0.893 to 0.967 for 200 μm voxel size, and from 0.914 to 0.967 for 125 μm voxel size (all p < 0.01). In general, we observed a trend that the AUC values, sensitivity, and specialty increased with the decrease in the voxel size, and significantly higher AUC values were detected in 125 μm voxel size images. In the detection of ACs, ROC curve analysis showed that among the four observers, the AUC values ranged from 0.554 to 0.639 for 300 μm voxel size, from 0.532 to 0.654 for 250 μm voxel size, from 0.567 to 0.626 for 200 μm voxel size, and from 0.638 to 0.678 for 125 μm voxel size. CBCT images at a voxel size of 125 μm had a weak diagnostic potential (AUC: 0.5–0.7, all p < 0.05) in the detection of AC, with a lower sensitivity ranging from 36.8 to 57.9% and a higher specialty ranging from 73.6 to 98.8%.

Conclusions

CBCT with 300 μm voxel size could only provide moderate diagnostic accuracy in the detection of a double canal system in mandibular incisors. CBCT with a voxel size of 125 μm exhibited high diagnostic value in the detection of double canal systems, while showing low but statistically significant value in the detection of ACs.
Hinweise
Bingbing Bai, Ying Tang, and Yihan Wu contributed equally to this work as first authors.

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Background

An exact knowledge of the individual root canal anatomy and variations is a prerequisite for successful endodontic treatment [1]. The mandibular central and lateral incisors are very similar in morphology and they usually have a single root and a single canal [2]. However, the presence of two canals in the bucco-lingual plane has been reported in mandibular incisors with detection rates ranging widely from 11 to 45%, and the discrepancy can be due to diverse laboratory and clinical methodologies used and different ethnicities of the population groups [315]. The second canal is often tiny in size and prone to be overlooked on conventional periapical radiographs during endodontic treatment, as they are 2-dimensional and cannot provide anatomic information of the root at the sagittal plane. Moreover, the image quality may be degraded due to distortion or superimposition of bone and other dental structures [10]. Failure to detect or treat an existing second canal may lead to unfavorable outcomes of endodontic treatment. The presence of accessory canals (ACs) is another endodontic challenge. They connect the main root canal or pulp chamber with the root surface and the incidence varies widely among different tooth types, and different populations [16]. ACs can result in lateral lesions or post-treatment periapical periodontitis as the conventional techniques are inadequate to access, clean, disinfect, and fill them efficiently [16]. Ji et al. [17] examined CBCT images of 70 extracted mandibular incisors taken at a voxel size of 125, 200 and 250 μm, and the detection accuracy of ACs were 80.0%, 13.3%, and 33.3% respectively (the clearing technique served as the gold standard). This finding indicated that assessment of ACs could only be performed with a voxel size of 125 μm (high-resolution scan mode). Due to the small size of ACs, previous studies were mostly based on laboratory methodologies such as micro-CT or clearing technique, and till now, no efficient diagnostic tool has been reported to clinically detect and identify ACs.
In recent years, dental cone-beam computed tomography (CBCT) has been widely used to evaluate the patients’ tooth root and canal anatomy due to its high resolution and noninvasive nature. Although the scan time and radiation dose of CBCT are significantly reduced as compared with conventional medical CT scans, CBCT scanners can still produce a higher radiation dose than a single traditional dental radiograph [18]. Therefore, the principles of ALARA (as low as reasonably achievable) and ALADA (as low as diagnostically acceptable) have to be followed, and no routine use of CBCT is allowed during dental treatments. The clinicians should be aware that CBCT imaging should be performed with the least amount of radiation required to provide adequate image quality, namely “dose optimization” [19]. The diagnostic accuracy of CBCT examination can be affected by several objective and subjective factors. The former includes the scanning unit itself, the field of view (FOV), scanning time, tube voltage and current, the voxel size, etc. [20, 21], while the latter includes but is not limited to the professional experiences and educational background of the observers. Voxel size is one of the most important parameters affecting the CBCT image quality. Generally speaking, a small voxel size in a CBCT protocol is often associated with better image quality and higher diagnostic accuracy [22, 23], but requires more exposure time and consequently may increase the radiation exposure to the patient [24]. Previous scholars have evaluated the impact of voxel size on different diagnostic tasks, such as detection of root fractures, carious lesions, root resorption, periodontal defects, erosions in the temporomandibular joint, root canal variations, etc., and the results varied depending on the task and the CBCT systems used [21, 2528]. In regard to the diagnosis of root canal anatomy, Bauman et al. [22] and Mouzinho-Machado et al. [25] reported that a smaller voxel size increased the diagnostic accuracy of MB2 canals. While Vizzotto et al. [23] reported that filling of the first mesio-buccal canal (MB1) reduced the second mesio-buccal canal (MB2) detection in the CBCT images of 300 and 250 μm voxel size, but not of 200 μm voxel size; for teeth with MB1 filling, the 200 μm voxel size was the most appropriate, whereas for those with no MB1 filling or after MB1 root canal filling removal, the 300 μm voxel size was adequate to achieve good diagnostic performance and lower radiation exposure.
Micro-computed tomography (micro-CT) is another type of CT, which is often used for the ex vivo study of human teeth. Its resolution can be approximately 10-fold higher than that of CBCT, and thus it can be regarded as the gold standard for the diagnosis of root canal variations [16].
The aim of this study was to evaluate the ability of CBCT with 4 different voxel sizes in the identification of double root canal systems and ACs in permanent mandibular incisors, and the micro-CT images were used as the gold standard.

Materials and methods

Collection of sample teeth

Ethnic approval of the use of seven dry human mandibles and extracted human teeth was obtained from the Ethics Committee and the institutional review board (Issuing Number: KY2022-089-01) of the Ninth People’s Hospital of Suzhou. A total of 106 extracted permanent mandibular incisors were randomly collected from a native Chinese population as described previously [29], and the mean age of the subjects was 56.3 ± 16.7 years, ranging from 14 to 84 years.

Micro-CT scanning and analysis

Each mandibular incisor was scanned by using a micro-CT scanner (SkyScan1174; Bruker-microCT, Kontich, Belgium) with a 1-mm aluminum filter at a voxel size of 9 μm, 800 mA, 50 kVp, rotation step of 0.7˚ and arch rotation of 180˚. The data sets were then transferred to Mimics 21.0 (Materialise, Leuven, Belgium) software to perform 3D reconstruction of the teeth and root canal systems. The presence or absence of a double root canal system and the presence or absence of AC(s) (lateral canals and/or apical accessory canals) (Fig. 1) were determined by the first author (Bingbing Bai), a junior dentist with working experience of 4 years.
Before micro-CT diagnosis, the calibration was performed by an endodontic specialist (Yongchun Gu) and the first author. The micro-CT images of 10 single- and 10 double-canaled incisors were randomly selected and examined at different views and magnifications to identify the presence or absence of double canals and ACs. Any disagreements were discussed until a consensus was reached. Cohen’s kappa test was used to evaluate the inter- and intra-observer errors. The micro-CT images were examined twice with an interval of 14 days. The intra-and inter-observer kappa values were both 1.0 (all p = 0.000).
The diameters of the ACs were measured by using Mimics software. Briefly, a cross-section was made at the midpoint of the accessory canal, and after full amplification of the image by zooming, the maximum (D) and minimum (d) diameters were measured.

CBCT scanning and analysis

The extracted sample teeth were placed in the tooth sockets of the mandibular incisors in the dry mandible specimens and fixed with wax. Imaging was performed using a 3D eXam i (KaVo Co., Germany) CBCT machine with 4 different voxel sizes (Fig. 2). The imaging protocols were used in our clinical routine and listed as follows: (1) FOV = 14 × 8.5 cm; tube peak potential = 120 kVp; tube current = 15.44 mA; exposure time = 3 s; voxel size = 0.30 mm. (2) FOV = 14 × 8.5 cm; tube peak potential = 120 kVp; tube current = 30.89 mA; time = 6 s; voxel size = 0.25 mm. (3) FOV = 14 × 8.5 cm; tube peak potential = 120 kVp; tube current = 30.89 mA; time = 6 s; voxel size = 0.20 mm. (4) FOV = 8.5 × 8.5 cm; tube peak potential = 120 kVp; tube current = 30.89 mA; time = 6 s; voxel size = 0.125 mm.
Kavo eXam Vision software (KAVO, Germany) was used to analyze the images. The root canal anatomy was examined in detail by switching the viewing angle (horizontal plane, coronal plane, sagittal plane, and any other arbitrary planes) and adjusting the magnification. A blind visual analysis was conducted by 4 trained observers (a lab researcher with 9 years of experience working in the field of radiology [Ying Tang], an endodontic specialist with 25 years of working experience [Yongchun Gu], and 2 postgraduate students in endodontics [Yihan Wu and Fan Pei) who evaluated the CBCT images independently and scored the detection of a double root canal system on a 5-point rank scale: 1, definitely present of a double canal system; 2, probably present of a double canal system; 3, unsure; 4, probably absent of a double canal system; and 5, definitely absent of a double canal system (a single canal was definitely detected). The detection of ACs was also based on a 5-point rank scale: 1, AC definitely present; 2, AC probably present; 3, unsure; 4, AC probably absent; and 5, AC definitely absent. The observations of CBCT image sets with different voxel sizes were carried out according to the following sequence: 300 μm ◊ 250 μm ◊ 200 μm ◊ 125 μm. The interval between the two observations was 14 days to reduce the recall bias.

Statistical analysis

Statistical analysis was performed using MedCalc statistical software version 15.8 (MedCalc Software bvba, Ostend, Belgium). Receiver operating characteristic (ROC) curve analysis was plotted to evaluate CBCT diagnostic accuracy with 4 different voxel sizes, and DeLong test was used to compare area under curve (AUC) values, and the observation results of micro-CT were taken as the gold standard. According to arbitrary guidelines, the accuracy of prediction is defined as low (AUC:0.5–0.7), moderate (0.7–0.9), and high (0.9-1). The sensitivity, specificity, positive predictive value, negative predictive value, and Youden index were calculated. The significance level was set at p < 0.05.

Results

The presence/absence of double canal systems or ACs in mandibular incisors detected by micro-CT

The root canal systems in mandibular incisors detected by micro-CT (the gold standard) were described in Table 1; Fig. 1. Among a total of 106 specimens, double root canal systems were detected in 25 specimens, and 26 ACs were identified in 19 specimens. The 25 cases of double canal systems included 21 cases of Vertucci’s type III (1-2-1) canals, 1 case of type II (2 − 1), and 3 cases of type V (1–2)] canals, and in the current study, we have not detected other complex root canal forms.
Table 1
Root canal configuration of mandibular incisors detected by micro-CT (n = 106)
 
Double canals
 
Accessory canals
Presence
Absence
 
Presence
Absence
n
25
81
 
19
87
%
23.58%
76.42%
 
17.92%
82.08%

The diameter of the ACs measured by micro-CT

In 19 specimens, 26 accessory canals were detected by micro-CT scans. Measurement results of the maximum and minimum diameters of the ACs were shown in Fig. 3. The mean maximum and minimum diameter was 192 ± 107 μm and 131 ± 51 μm, respectively. The majority of the ACs (20/26) have a canal diameter (D) below 250 μm.

The presence/absence of a double canal system in mandibular incisors detected by CBCT with 4 different voxel sizes

The AUC values, sensitivity, specificity, PPV, NPV, and Youden index calculated for 4 observers and CBCT image sets with 4 different voxel sizes in the detection of double root canal systems were summarized in Table 2. The AUC values ranged from 0.765 to 0.889 for voxel size of 300 μm, from 0.877 to 0.926 for 250 μm voxel size, from 0.893 to 0.967 for 200 μm voxel size, and from 0.914 to 0.967 for 125 μm voxel size. In general, the AUC values, sensitivity, and specialty increased with the decrease in the voxel size. Statistically significantly higher AUC values were detected in 125 μm voxel size images as compared to the image sets with 300 μm voxel size (p = 0.038) and 250 μm voxel size (p = 0.044) for Observer 1. For Observer 2, significantly higher AUC values were detected in 200 μm voxel size images as compared to those with 300 μm voxel size (p = 0.024). For Observer 3, significantly higher AUC values were detected in 125 μm voxel size images as compared to CBCT images at 300 μm voxel size (p = 0.009), and for Observer 4, AUC values for images at 250, 200, and 125 μm voxel sizes were significantly higher than that for 300 μm voxel size, and the p value was 0.047, 0.027 and 0.015, respectively (Table 2; Fig. 4). Statistical differences were also detected in AUC values between Observer 1 and 4 (p = 0.043), 2 and 4 (p = 0.017) at 300 μm voxel size, and between Observer 2 and 4 (p = 0.043) at 200 μm voxel size.
Table 2
AUC values, Se, Sp, PPV, NPV, and Youden index for 4 observers in the diagnosis of double-canaled mandibular incisors based on CBCT imaging with 4 different voxel sizes
 
Voxel size
AUC
SE
P value
Asymptotic 95% confidence interval
Se
Sp
PPV
NPV
Youden
index
(µm)
Lower bound
Upper bound
Observer 1
300
0.882
0.039
< 0.0001
0.805
0.936
0.760
0.901
0.904
0.924
0.661
(Tang Y)
250
0.926
0.036
< 0.0001
0.859
0.968
0.840
0.938
0.808
0.950
0.778
 
200
0.943
0.035
< 0.0001
0.881
0.979
0.920
0.963
0.885
0.975
0.883
 
125
0.959
0.028
< 0.0001
0.901
0.988
0.920
0.988
0.958
0.976
0.908
Observer 2
300
0.889
0.035
< 0.0001
0.814
0.942
0.760
0.877
0.655
0.922
0.637
(Gu Y)
250
0.917
0.034
< 0.0001
0.847
0.961
0.880
0.889
0.710
0.960
0.769
 
200
0.967
0.023
< 0.0001
0.912
0.992
0.960
0.889
0.727
0.986
0.849
 
125
0.948
0.029
< 0.0001
0.887
0.982
0.920
0.963
0.885
0.975
0.883
Observer 3
300
0.827
0.058
< 0.0001
0.741
0.893
0.800
0.827
0.588
0.931
0.627
(Wu Y)
250
0.877
0.047
< 0.0001
0.799
0.933
0.760
0.975
0.905
0.929
0.735
 
200
0.893
0.043
< 0.0001
0.818
0.945
0.800
0.975
0.909
0.940
0.775
 
125
0.947
0.029
< 0.0001
0.886
0.981
0.920
0.975
0.920
0.975
0.895
Observer 4
300
0.765
0.056
< 0.0001
0.672
0.842
0.640
0.815
0.516
0.880
0.455
(Pei F)
250
0.889
0.042
< 0.0001
0.814
0.942
0.840
0.951
0.840
0.951
0.791
 
200
0.899
0.040
< 0.0001
0.825
0.949
0.840
0.963
0.875
0.951
0.803
 
125
0.914
0.038
< 0.0001
0.843
0.996
0.800
0.975
0.909
0.940
0.775
AUC, area under curve; CBCT, cone beam CT; NPV, negative predictive value; PPV, positive predictive value; Se, sensitivity; SE, standard error; Sp, specificity

CBCT images at a voxel size of 125 μm have a weak diagnostic potential in the detection of ACs in mandibular incisors

As shown in Table 3; Fig. 4, among 4 observers, the AUC values ranged from 0.509 to 0.625 for 300 μm voxel size, from 0.562 to 0.667 for 250 μm voxel size, from 0.565 to 0.638 for 200 μm voxel size, and from 0.644 to 0.678 for 125 μm voxel size. These data also indicated that CBCT images had no (p > 0.05) or only low (AUC < 0.7) diagnostic potential in the detection of ACs in mandibular incisors. No statistically significant differences were found among different voxel sizes in the detection of ACs for each observer (all p > 0.05). In general, CBCT images at a voxel size of 125 μm had a weak diagnostic potential in the detection of ACs in mandibular incisors (for each of all 4 observers, p values were in a range between 0.001 and 0.034), and the sensitivity was at a low level ranged from 36.8 to 42.1%, while the specialty was at a high level ranged between 86.2% and 98.9%.
Table 3
AUC values, Se, Sp, PPV, NPV, and Youden index for 4 observers in the diagnosis of accessory canals in the mandibular incisors based on CBCT imaging with 4 different voxel sizes
 
Voxel size
AUC
SE
P value
Asymptotic 95% confidence interval
Se
Sp
PPV
NPV
Youden
index
(µm)
Lower bound
Upper bound
Observer 1
300
0.604
0.064
0.105
0.505
0.698
0.684
0.517
0.236
0.882
0.202
(Tang Y)
250
0.562
0.077
0.421
0.463
0.658
0.263
0.920
0.417
0.851
0.183
 
200
0.565
0.091
0.473
0.466
0.661
0.316
0.977
0.750
0.867
0.293
 
125
0.671
0.064
0.008
0.573
0.759
0.421
0.862
0.400
0.872
0.283
Observer 2
300
0.509
0.069
0.896
0.410
0.607
0.842
0.184
0.184
0.842
0.026
(Gu Y)
250
0.667
0.072
0.020
0.568
0.755
0.421
0.908
0.500
0.878
0.329
 
200
0.638
0.069
0.047
0.526
0.729
0.526
0.701
0.278
0.871
0.228
 
125
0.644
0.068
0.034
0.545
0.735
0.368
0.885
0.412
0.865
0.254
Observer 3
300
0.539
0.062
0.529
0.440
0.637
0.895
0.184
0.193
0.889
0.079
(Wu Y)
250
0.626
0.064
0.049
0.527
0.718
0.421
0.816
0.333
0.866
0.237
 
200
0.633
0.068
0.051
0.534
0.725
0.368
0.874
0.389
0.864
0.242
 
125
0.661
0.065
0.013
0.563
0.750
0.412
0.899
0.437
0.889
0.311
Observer 4
300
0.625
0.065
0.053
0.526
0.717
0.684
0.609
0.277
0.898
0.293
(Pei F)
250
0.577
0.067
0.251
0.477
0.672
0.263
0.977
0.714
0.859
0.240
 
200
0.588
0.049
0.073
0.488
0.683
0.211
0.966
0.571
0.848
0.176
 
125
0.678
0.057
0.001
0.581
0.766
0.368
0.989
0.875
0.878
0.357
AUC, the area under the curve; CBCT, cone beam CT; NPV, negative predictive value; PPV, positive predictive value; Se, sensitivity; SE, standard error; Sp, specificity

Discussion

CBCT is a valuable diagnostic tool in endodontics. The selection of an appropriate scanning protocol is critically important to balance between maintenance of essential image quality and the reduction of radiation dose [30]. The voxel size is one of the most important scanning parameters, and in most of the cases, an optimal, rather than the minimum voxel size is desirable for achieving the best diagnostic efficiency, as a smaller voxel size is associated with more radiation exposure, increased noise level and decreased contrast-to-noise ratio, while might provide a same diagnostic outcome as lower resolution images [16, 21, 31, 32].
In the current ex vivo study, to identify the presence of a double root canal system and ACs in mandibular incisors, extracted teeth were fixed in the tooth sockets of dry mandibles with wax to simulate the in vivo condition. All observers assessed CBCT image sets taken at 4 different voxel sizes as our clinical routine, while the micro-CT images at 9 μm voxel size were used as the gold standard. During ROC curve analysis, a high AUC value closing 1.0 indicated an ideal diagnostic test, while a low AUC value closing 0.5 indicated no diagnostic effectiveness. The data in Table 3 reveals that all 4 voxel sizes are reliable in identifying the double root canal system (the AUC ranged from 0.765 to 0.967, all p < 0.0001), and for all 4 observers, a voxel size of 125 μm could achieve high diagnostic accuracy (all AUC > 0.9, and all p < 0.0001), while a voxel size of 300 μm could only achieve moderate diagnostic accuracy (the AUC ranged from 0.765 to 0.889). Professional experience is also an influencing factor, as for Observer 1 (a researcher with 9 years of working experience in radiology) and Observer 2 (an experienced endodontic specialist), the AUC values were greater than those of Observer 3 and 4 (both were graduate students of dental school) at each voxel size. Statistical differences were detected between Observer 1 and 4, and Observer 2 and 4 at 300 μm voxel size, and between Observer 2 and 4 at 200 μm voxel size (all p < 0.05). These findings also indicate that sharp CBCT images with high resolution (125 μm voxel size) are more essential for an inexperienced clinician; while for an experienced specialist, CBCT with a lower resolution (200 μm voxel size) is only slightly less accurate than the high-resolution modality (125 μm voxel size), which has an advantage of lower exposure dosage. This is in contrast to the report of Bauman et al. [22], who noted that the accuracy of MB2 canal detection couldn’t be correlated with the observers’ level of clinical experience. Previous scholars demonstrated that the presence of artifacts can also affect the diagnostic ability of CBCT [33]. Vizzotto et al. [23] found that the root-filling materials in MB1 may affect the detection of MB2, and the degree of influence was dependent on the CBCT voxel size during the examination. Recently, Lee et al. [34] retrospectively examined CBCT images of a Korean population and reported that the mandibular first molar showed a high incidence (27.0%) of separate disto-lingual (DL) roots, and this root variation was highly associated with the presence of two‑canaled mandibular incisors (left central incisors: p = 0.001, odds ratio = 4.25; left lateral incisors: p < 0.001, odds ratio = 3.8; right central incisors: p = 0.003, odds ratio = 3.86; right lateral incisors: p = 0.001, odds ratio = 3.44). Similar finding has been reported in the Taiwanese population [10, 11], which indicates that when the voxel size and quality of CBCT images are inadequate for accurate detection of two-canaled mandibular incisors, the root number of the mandibular first molars may provide clinicians some clues. The DL root can more easily be diagnosed, and its presence indicates a high possibility of two canals in mandibular incisors on the same side.
Literature review shows only very few scholars have previously assessed the diagnostic potential of CBCT in identifying ACs [16, 17]. To evaluate the effectiveness of CBCT imaging (using a voxel size of 80 μm) for the detection of lateral canals in endodontically treated premolars, a similar ex vivo study was carried out by Sousa et al. [16]. They reported that the AUC value was only 0.58 and 0.49 (all p < 0.001) before and after root canal treatment, respectively, and CBCT imaging was not an effective diagnostic tool for the detection of ACs. The current study found that CBCT images taken at 125 μm voxel size have a low (AUC: 0.5–0.7) diagnostic potential in the detection of ACs in mandibular incisors for all 4 observers. The AUC values ranged from 0.638 to 0.678 (all p < 0.05), and the sensitivity was at a low level ranging between 36.8% and 57.9%, while the specialty was at a relatively high level ranging between 73.6% and 98.8%. The higher specificity value can be due to the observers’ difficulty in detecting the accessory canals and they tended to score “4 = probably absent” or “5 = absent”, which resulted in a large number of “negative” answers, consequently increasing the number of “true-negative” cases [16]. This finding is similar to that of Ji et al. [17], who reported that partial lateral accessory canals could be detected using CBCT at high-resolution scan mode (voxel size = 125 μm). Theoretically, CBCT cannot identify an anatomic structure with a size below the voxel size. Moreover, the relationship between CBCT spatial resolution and voxel size indicated that the former is at least twice greater than the latter [35]. The odontometric measurement by micro-CT showed that the majority of the ACs had a canal size below 250 μm (Fig. 3), which means even using the smallest voxel size of 125 μm, the spatial resolution (which is often 2-fold greater than the voxel size) is inadequate to identify the majority of ACs. The low sensitivity also indicated that the majority of ACs were missed diagnoses. Theoretically, more advanced modulates with 75–80 μm voxel size can improve diagnostic accuracy.
This ex vivo study has several limitations. First, to simulate the in vivo dental anatomy, the extracted sample teeth were fixed in the tooth sockets of dry mandibles, which were stationary during CBCT scanning. While in the clinical scenario, patient movement is also an influential factor, with the magnitudes even exceeding voxel size and scatter effects, and this may lead to overestimated diagnostic accuracy of the CBCT. Second, in the current study, the majority of the sample teeth were collected from elderly people, and the small size of the main canals or ACs was associated with a lower CBCT detection rate. Finally, this study only used one CBCT machine (3D eXam I, KaVo) with the smallest voxel size of 125 μm, which was not the most advanced machine currently in the CBCT market. Therefore, further studies are essential based on other brands of CBCT units, especially those with a minimum voxel size of 75–80 μm.

Conclusions

CBCT with 300 μm voxel size could only provide moderate diagnostic accuracy in the detection of double canal systems in mandibular incisors. CBCT with a voxel size of 125 μm exhibited high diagnostic value in the detection of double canal systems while showing low but statistically significant value in the detection of ACs. For experienced observers, lower resolution CBCT at 200 μm voxel size only slightly reduces the diagnostic accuracy, while having the advantage of radiation reduction.

Acknowledgements

Not applicable.

Declarations

Competing interests

The authors declare no competing interests.

Abbreviations

CBCT Cone-beam computed tomography.
micro-CT micro-computed tomography.
3D three-dimensional.
FOV field of view.
AC accessory canal.
ROC receiver operating characteristic.
AUC area under curve.
MB2 secondary mesio-buccal canal.
MB1 first mesio-buccal canal.
DL disto-lingual.
The human mandibular incisors used in this study were obtained from the Department of Dentistry, Ninth People’s Hospital of Suzhou. All methods have been performed in accordance with the Declaration of Helsinki and have been approved by the Ethics Committee of Ninth People’s Hospital of Suzhou with the approval number # KY2022-089-01. Informed consent was obtained from all subjects and/or their legal guardian(s).
Not applicable.

Authors details

1The Stomatology Hospital Affiliated of Suzhou Vocational Health College, Renmin Road 829#, Gusu Dist., 215002 Suzhou, China.2 Department of Central Laboratory and Dentistry, Ninth People’s Hospital of Suzhou, Soochow University, Ludang Road 2666#, Wujiang Dist., 215200 Suzhou, China.
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article’s Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article’s Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://​creativecommons.​org/​licenses/​by/​4.​0/​. The Creative Commons Public Domain Dedication waiver (http://​creativecommons.​org/​publicdomain/​zero/​1.​0/​) applies to the data made available in this article, unless otherwise stated in a credit line to the data.

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Literatur
1.
Zurück zum Zitat Kabak YS, Abbott PV. Endodontic treatment of mandibular incisors with two root canals: report of two cases. Aust Endod J. 2010;33(1):27–31.CrossRef Kabak YS, Abbott PV. Endodontic treatment of mandibular incisors with two root canals: report of two cases. Aust Endod J. 2010;33(1):27–31.CrossRef
2.
Zurück zum Zitat Scheid RC, Weiss G. Woelfel’s dental anatomy. 8th ed. Philadelphia: Lippincott, Williams & Wilkins; 2012. Scheid RC, Weiss G. Woelfel’s dental anatomy. 8th ed. Philadelphia: Lippincott, Williams & Wilkins; 2012.
3.
Zurück zum Zitat Wolf TG, Stiebritz M, Boemke N, Elsayed I, Paqué F, Wierichs RJ, et al. 3-dimensional analysis and literature review of the root canal morphology and physiological foramen geometry of 125 mandibular incisors by means of micro–computed tomography in a german population. J Endod. 2020;46(2):184–91.CrossRefPubMed Wolf TG, Stiebritz M, Boemke N, Elsayed I, Paqué F, Wierichs RJ, et al. 3-dimensional analysis and literature review of the root canal morphology and physiological foramen geometry of 125 mandibular incisors by means of micro–computed tomography in a german population. J Endod. 2020;46(2):184–91.CrossRefPubMed
4.
Zurück zum Zitat Leoni GB, Versiani MA, Pécora JD, Damio dSM. Micro–computed tomographic analysis of the root canal morphology of mandibular incisors. J Endod. 2014;40(5):710–6.CrossRefPubMed Leoni GB, Versiani MA, Pécora JD, Damio dSM. Micro–computed tomographic analysis of the root canal morphology of mandibular incisors. J Endod. 2014;40(5):710–6.CrossRefPubMed
5.
Zurück zum Zitat Vertucci FJ. Root canal anatomy of the mandibular anterior teeth. J Am Dent Assoc 1974,89(2): 369–71. Vertucci FJ. Root canal anatomy of the mandibular anterior teeth. J Am Dent Assoc 1974,89(2): 369–71.
6.
Zurück zum Zitat Baxter S, Jablonski M, Hulsmann M. Cone-beam-computed-tomography of the symmetry of root canal anatomy in mandibular incisors. J Oral Sci. 2020;62(2):180–3.CrossRefPubMed Baxter S, Jablonski M, Hulsmann M. Cone-beam-computed-tomography of the symmetry of root canal anatomy in mandibular incisors. J Oral Sci. 2020;62(2):180–3.CrossRefPubMed
7.
Zurück zum Zitat Han T, Ma Y, Yang L, Chen X, Zhang X, Wang Y. A study of the root canal morphology of mandibular anterior teeth using cone-beam computed tomography in a chinese subpopulation. J Endod. 2014;40(9):1309–14.CrossRefPubMed Han T, Ma Y, Yang L, Chen X, Zhang X, Wang Y. A study of the root canal morphology of mandibular anterior teeth using cone-beam computed tomography in a chinese subpopulation. J Endod. 2014;40(9):1309–14.CrossRefPubMed
8.
Zurück zum Zitat Miyashita M, Kasahara E, Yasuda E, Yamamoto A, Sekizawa T. Root canal system of the mandibular incisor. J Endod. 1997;23(8):479–84.CrossRefPubMed Miyashita M, Kasahara E, Yasuda E, Yamamoto A, Sekizawa T. Root canal system of the mandibular incisor. J Endod. 1997;23(8):479–84.CrossRefPubMed
9.
10.
Zurück zum Zitat Wu YC, Cheng WC, Chung MP, Su CC, Weng PW, Cathy Tsai YW, et al. Complicated root canal morphology of mandibular lateral incisors is associated with the presence of distolingual root in mandibular first molars: a cone-beam computed tomographic study in a taiwanese population. J Endod. 2018;44(1):73–9.CrossRefPubMed Wu YC, Cheng WC, Chung MP, Su CC, Weng PW, Cathy Tsai YW, et al. Complicated root canal morphology of mandibular lateral incisors is associated with the presence of distolingual root in mandibular first molars: a cone-beam computed tomographic study in a taiwanese population. J Endod. 2018;44(1):73–9.CrossRefPubMed
11.
Zurück zum Zitat Wu YC, Cheng WC, Weng PW, Chung MP, Su CC, Chiang HS, et al. The presence of distolingual root in mandibular first molars is correlated with complicated root canal morphology of mandibular central incisors: a cone-beam computed tomographic study in a taiwanese population. J Endod. 2018;44(5):711–6.CrossRefPubMed Wu YC, Cheng WC, Weng PW, Chung MP, Su CC, Chiang HS, et al. The presence of distolingual root in mandibular first molars is correlated with complicated root canal morphology of mandibular central incisors: a cone-beam computed tomographic study in a taiwanese population. J Endod. 2018;44(5):711–6.CrossRefPubMed
12.
Zurück zum Zitat Geduk G, Deniz Y, Eroglu E, Zengin AZ. Cone-beam computed tomography study of root canal morphology of permanent mandibular incisors in a turkish sub-population. J Oral Maxillofac Radiol. 2015;3(1):7–10.CrossRef Geduk G, Deniz Y, Eroglu E, Zengin AZ. Cone-beam computed tomography study of root canal morphology of permanent mandibular incisors in a turkish sub-population. J Oral Maxillofac Radiol. 2015;3(1):7–10.CrossRef
13.
Zurück zum Zitat Liu J, Luo J, Dou L, Yang D. CBCT study of root and canal morphology of permanent mandibular incisors in a chinese population. Acta Odontol Scand. 2014;72(1):26–30.CrossRefPubMed Liu J, Luo J, Dou L, Yang D. CBCT study of root and canal morphology of permanent mandibular incisors in a chinese population. Acta Odontol Scand. 2014;72(1):26–30.CrossRefPubMed
14.
Zurück zum Zitat Candeiro GTM, Monteiro Dodt Teixeira IM, Olimpio Barbosa DA, Alves FRF. Vertucci’s root canal configuration of 14,413 mandibular anterior teeth in a brazilian population: a prevalence study using cone-beam computed tomography. J Endod. 2021;47(3):404–8.CrossRefPubMed Candeiro GTM, Monteiro Dodt Teixeira IM, Olimpio Barbosa DA, Alves FRF. Vertucci’s root canal configuration of 14,413 mandibular anterior teeth in a brazilian population: a prevalence study using cone-beam computed tomography. J Endod. 2021;47(3):404–8.CrossRefPubMed
15.
Zurück zum Zitat Kartal N, Yanikoglu FC. Root canal morphology of mandibular incisors. J Endod. 1992;18(11):562–4.CrossRefPubMed Kartal N, Yanikoglu FC. Root canal morphology of mandibular incisors. J Endod. 1992;18(11):562–4.CrossRefPubMed
16.
Zurück zum Zitat Sousa TO, Hassan B, Mirmohammadi H, Shemesh H, Haiter-Neto F. Feasibility of cone-beam computed tomography in detecting lateral canals before and after root canal treatment: an ex vivo study. J Endod. 2017;43(6):1014–7.CrossRefPubMed Sousa TO, Hassan B, Mirmohammadi H, Shemesh H, Haiter-Neto F. Feasibility of cone-beam computed tomography in detecting lateral canals before and after root canal treatment: an ex vivo study. J Endod. 2017;43(6):1014–7.CrossRefPubMed
17.
Zurück zum Zitat Ji Y, Wen S, Liu S, Zhu M, Yao M, Wang T, et al. Could cone-beam computed tomography demonstrate the lateral accessory canals? BMC Oral Health. 2017;17(1):142.CrossRefPubMedPubMedCentral Ji Y, Wen S, Liu S, Zhu M, Yao M, Wang T, et al. Could cone-beam computed tomography demonstrate the lateral accessory canals? BMC Oral Health. 2017;17(1):142.CrossRefPubMedPubMedCentral
18.
Zurück zum Zitat Ma RH, Ge ZP, Li G. Detection accuracy of root fractures in cone-beam computed tomography images: a systematic review and meta-analysis. Int Endod J. 2016;49(7):646–54.CrossRefPubMed Ma RH, Ge ZP, Li G. Detection accuracy of root fractures in cone-beam computed tomography images: a systematic review and meta-analysis. Int Endod J. 2016;49(7):646–54.CrossRefPubMed
19.
Zurück zum Zitat The American Dental Association Council on Scientific Affairs. The use of cone-beam computed tomography in dentistry. J Am Dent Assoc. 2012;143(8):899–902.CrossRef The American Dental Association Council on Scientific Affairs. The use of cone-beam computed tomography in dentistry. J Am Dent Assoc. 2012;143(8):899–902.CrossRef
20.
Zurück zum Zitat Kamburoğlu K, Koç C, Sönmez G, Elbahary S, Rosen E, Tsesis I. Effect of cone beam computed tomography voxel size and dental specialty status on the agreement of observers in the detection and measurement of periapical lesions. Oral Surg Oral Med Oral Pathol. 2021;132(3):346–51.CrossRef Kamburoğlu K, Koç C, Sönmez G, Elbahary S, Rosen E, Tsesis I. Effect of cone beam computed tomography voxel size and dental specialty status on the agreement of observers in the detection and measurement of periapical lesions. Oral Surg Oral Med Oral Pathol. 2021;132(3):346–51.CrossRef
21.
Zurück zum Zitat Spin-Neto R, Gotfredsen E, Wenzel A. Impact of voxel size variation on CBCT-based diagnostic outcome in dentistry: a systematic review. J Digit Imaging 2013,26(4):813–20. Spin-Neto R, Gotfredsen E, Wenzel A. Impact of voxel size variation on CBCT-based diagnostic outcome in dentistry: a systematic review. J Digit Imaging 2013,26(4):813–20.
22.
Zurück zum Zitat Bauman R, Scarfe W, Clark SJ, Morelli J, Scheetz J, Farman A. Ex vivo detection of mesiobuccal canals in maxillary molars using CBCT at four different isotropic voxel dimensions. Int Endod J 2011,44(8): 752–8. Bauman R, Scarfe W, Clark SJ, Morelli J, Scheetz J, Farman A. Ex vivo detection of mesiobuccal canals in maxillary molars using CBCT at four different isotropic voxel dimensions. Int Endod J 2011,44(8): 752–8.
23.
Zurück zum Zitat Vizzotto MB, Silveira PF, Arus NA, Montagner F, Gomes BP, da Silveira HE. CBCT for the assessment of second mesiobuccal (MB2) canals in maxillary molar teeth: effect of voxel size and presence of root filling. Int Endod J. 2013;46(9):870–6.CrossRefPubMed Vizzotto MB, Silveira PF, Arus NA, Montagner F, Gomes BP, da Silveira HE. CBCT for the assessment of second mesiobuccal (MB2) canals in maxillary molar teeth: effect of voxel size and presence of root filling. Int Endod J. 2013;46(9):870–6.CrossRefPubMed
24.
Zurück zum Zitat Kim S, Song H, Samei E, Yin FF, Yoshizumi TT. Computed tomography dose index and dose length product for cone-beam CT: Monte Carlo simulations. J Appl Clin Med Phys. 2011;12(2):3395.CrossRefPubMed Kim S, Song H, Samei E, Yin FF, Yoshizumi TT. Computed tomography dose index and dose length product for cone-beam CT: Monte Carlo simulations. J Appl Clin Med Phys. 2011;12(2):3395.CrossRefPubMed
25.
Zurück zum Zitat Mouzinho-Machado S, Rosado LPL, Coelho-Silva F, Neves FS, Haiter-Neto F, de-Azevedo-Vaz SL. Influence of voxel size and filter application in detecting second mesiobuccal canals in cone-beam computed tomographic images. J Endod. 2021;47(9):1391–7.CrossRefPubMed Mouzinho-Machado S, Rosado LPL, Coelho-Silva F, Neves FS, Haiter-Neto F, de-Azevedo-Vaz SL. Influence of voxel size and filter application in detecting second mesiobuccal canals in cone-beam computed tomographic images. J Endod. 2021;47(9):1391–7.CrossRefPubMed
26.
Zurück zum Zitat Guo XL, Li G, Zheng JQ, Ma RH, Liu FC, Yuan FS et al. Accuracy of detecting vertical root fractures in non-root filled teeth using cone beam computed tomography: effect of voxel size and fracture width. Int Endod J 2019,52(6):887–98. Guo XL, Li G, Zheng JQ, Ma RH, Liu FC, Yuan FS et al. Accuracy of detecting vertical root fractures in non-root filled teeth using cone beam computed tomography: effect of voxel size and fracture width. Int Endod J 2019,52(6):887–98.
27.
Zurück zum Zitat Librizzi ZT, Tadinada AS, Valiyaparambil JV, Lurie AG, Mallya SM. Cone-beam computed tomography to detect erosions of the temporomandibular joint: effect of field of view and voxel size on diagnostic efficacy and effective dose. Am J Orthod Dentofacial Orthop. 2011;140(1):e25–30.CrossRefPubMed Librizzi ZT, Tadinada AS, Valiyaparambil JV, Lurie AG, Mallya SM. Cone-beam computed tomography to detect erosions of the temporomandibular joint: effect of field of view and voxel size on diagnostic efficacy and effective dose. Am J Orthod Dentofacial Orthop. 2011;140(1):e25–30.CrossRefPubMed
28.
Zurück zum Zitat Icen M, Orhan K, Seker Ç, Geduk G, Cakmak Özlü F, Cengiz M. Comparison of CBCT with different voxel sizes and intraoral scanner for detection of periodontal defects: an in vitro study. Dentomaxillofac Radiol 2020,49(5): 20190197. Icen M, Orhan K, Seker Ç, Geduk G, Cakmak Özlü F, Cengiz M. Comparison of CBCT with different voxel sizes and intraoral scanner for detection of periodontal defects: an in vitro study. Dentomaxillofac Radiol 2020,49(5): 20190197.
29.
Zurück zum Zitat Tang Y, Wu Y, Pei F, Liu C, Qiu Y, Yang T, et al. A micro-computed tomographic analysis of the root canal systems in the permanent mandibular incisors in a chinese population. BMC Oral Health. 2023;23(1):129.CrossRefPubMedPubMedCentral Tang Y, Wu Y, Pei F, Liu C, Qiu Y, Yang T, et al. A micro-computed tomographic analysis of the root canal systems in the permanent mandibular incisors in a chinese population. BMC Oral Health. 2023;23(1):129.CrossRefPubMedPubMedCentral
30.
Zurück zum Zitat Scarfe WC. All that glitters is not gold: standards for cone-beam computerized tomographic imaging. Oral Surg Oral Med Oral Pathol Oral Radiol Endod. 2011;111(4):402–8.CrossRefPubMed Scarfe WC. All that glitters is not gold: standards for cone-beam computerized tomographic imaging. Oral Surg Oral Med Oral Pathol Oral Radiol Endod. 2011;111(4):402–8.CrossRefPubMed
31.
32.
Zurück zum Zitat Bechara B, McMahan CA, Moore WS, Noujeim M, Geha H, Teixeira FB. Contrast-to-noise ratio difference in small field of view cone beam computed tomography machines. J Oral Sci. 2012;54(3):227–32.CrossRefPubMed Bechara B, McMahan CA, Moore WS, Noujeim M, Geha H, Teixeira FB. Contrast-to-noise ratio difference in small field of view cone beam computed tomography machines. J Oral Sci. 2012;54(3):227–32.CrossRefPubMed
33.
Zurück zum Zitat Patel S, Dawood A, Ford TP, Whaites E. The potential applications of cone beam computed tomography in the management of endodontic problems. Int Endod J 2007,40(10): 818–30. Patel S, Dawood A, Ford TP, Whaites E. The potential applications of cone beam computed tomography in the management of endodontic problems. Int Endod J 2007,40(10): 818–30.
34.
Zurück zum Zitat Lee JB, Seo MS. Mandibular incisors with two canals are associated with the presence of the distolingual root in mandibular first molars: a cone-beam computed tomographic study. BMC Oral Health. 2022;22(1):14530.CrossRef Lee JB, Seo MS. Mandibular incisors with two canals are associated with the presence of the distolingual root in mandibular first molars: a cone-beam computed tomographic study. BMC Oral Health. 2022;22(1):14530.CrossRef
35.
Zurück zum Zitat Brüllmann D, Schulze RK. Spatial resolution in CBCT machines for dental/maxillofacial applications-what do we know today? Dentomaxillofac Radiol. 2015;44(1):20140204.CrossRefPubMed Brüllmann D, Schulze RK. Spatial resolution in CBCT machines for dental/maxillofacial applications-what do we know today? Dentomaxillofac Radiol. 2015;44(1):20140204.CrossRefPubMed
Metadaten
Titel
Ex vivo detection of mandibular incisors’ root canal morphology using cone-beam computed tomography with 4 different voxel sizes and micro-computed tomography
verfasst von
Bingbing Bai
Ying Tang
Yihan Wu
Fan Pei
Qi Zhu
Peng Zhu
Yongchun Gu
Publikationsdatum
01.12.2023
Verlag
BioMed Central
Erschienen in
BMC Oral Health / Ausgabe 1/2023
Elektronische ISSN: 1472-6831
DOI
https://doi.org/10.1186/s12903-023-03376-2

Weitere Artikel der Ausgabe 1/2023

BMC Oral Health 1/2023 Zur Ausgabe

Darf man die Behandlung eines Neonazis ablehnen?

08.05.2024 Gesellschaft Nachrichten

In einer Leseranfrage in der Zeitschrift Journal of the American Academy of Dermatology möchte ein anonymer Dermatologe bzw. eine anonyme Dermatologin wissen, ob er oder sie einen Patienten behandeln muss, der eine rassistische Tätowierung trägt.

Ein Drittel der jungen Ärztinnen und Ärzte erwägt abzuwandern

07.05.2024 Klinik aktuell Nachrichten

Extreme Arbeitsverdichtung und kaum Supervision: Dr. Andrea Martini, Sprecherin des Bündnisses Junge Ärztinnen und Ärzte (BJÄ) über den Frust des ärztlichen Nachwuchses und die Vorteile des Rucksack-Modells.

Endlich: Zi zeigt, mit welchen PVS Praxen zufrieden sind

IT für Ärzte Nachrichten

Darauf haben viele Praxen gewartet: Das Zi hat eine Liste von Praxisverwaltungssystemen veröffentlicht, die von Nutzern positiv bewertet werden. Eine gute Grundlage für wechselwillige Ärztinnen und Psychotherapeuten.

Parodontalbehandlung verbessert Prognose bei Katheterablation

19.04.2024 Vorhofflimmern Nachrichten

Werden Personen mit Vorhofflimmern in der Blanking-Periode nach einer Katheterablation gegen eine bestehende Parodontitis behandelt, verbessert dies die Erfolgsaussichten. Dafür sprechen die Resultate einer prospektiven Untersuchung.

Newsletter

Bestellen Sie unseren kostenlosen Newsletter Update Zahnmedizin und bleiben Sie gut informiert – ganz bequem per eMail.