Mechanical load exerted by PET-G aligners during mesial and distal derotation of a mandibular canine

An in vitro study

verfasst von: F. Elkholy, B. Mikhaiel, F. Schmidt, B. G. Lapatki

Erschienen in: Journal of Orofacial Orthopedics / Fortschritte der Kieferorthopädie | Ausgabe 5/2017

Einloggen, um Zugang zu erhalten

Abstract

Introduction

The six force–moment (F/M) components exerted by aligners of different thickness during simulated mesiorotation and distorotation of a mandibular canine were studied.

Materials and methods

An acrylic mandibular model with a separated right canine mounted on a hexapod via a 3D F/M sensor was used. Duran+® aligners (Scheu Dental, Germany) of thickness 0.5, 0.625, and 0.75 mm were fabricated on plaster models with the measurement tooth in its neutral position. The F/M values were recorded during progressive mesiorotation or distorotation of tooth 43 in 1° steps up to ±15°, corresponding to 0.5 mm displacements of the tooth’s interdental contacts. Each rotation step included renewed seating of the aligner on the acrylic model. Three aligners were tested three times each for each thickness and direction of rotation.

Results

The median rotational moments for the 0.5 mm aligner and 15° distorotation of tooth 43 was 27.49 Nmm (interquartile range, IQR 1.45 Nmm). The corresponding values for the 0.625 and 0.75 mm aligners were 41.04 Nmm (IQR 5.62 Nmm) and 42.48 Nmm (IQR 2.17 Nmm), respectively. The average rotational moments for distorotation were 15% higher than for mesiorotation (p = 0.01). Relatively high collateral F/M components, specifically an intrusive force and labiolingual and mesiodistal tipping moments, were observed.

Conclusion

To avoid overloading of periodontal structures, derotation of lower canines should be limited to 10° per setup step, leading to rotational moments of about 15 Nmm. The mechanical behavior of the 0.625 and 0.75 mm aligners were similar; thus, it may be omitted from the aligner sequence. Further studies are required to investigate specific aligner modifications or attachments for minimizing collateral F/M components or unwanted movements, respectively, during canine derotation.
Literatur
2.
Zurück zum Zitat Elkholy F, Panchaphongsaphak T, Kilic F et al (2015) Forces and moments delivered by PET-G aligners to an upper central incisor for labial and palatal translation. J Orofac Orthop 76:460–475CrossRefPubMed Elkholy F, Panchaphongsaphak T, Kilic F et al (2015) Forces and moments delivered by PET-G aligners to an upper central incisor for labial and palatal translation. J Orofac Orthop 76:460–475CrossRefPubMed
3.
Zurück zum Zitat Elkholy F, Schmidt F, Jäger R et al (2016) Forces and moments delivered by novel, thinner PET-G aligners during labiopalatal bodily movement of a maxillary central incisor: an in vitro study. Angle Orthod 86:883–890CrossRefPubMed Elkholy F, Schmidt F, Jäger R et al (2016) Forces and moments delivered by novel, thinner PET-G aligners during labiopalatal bodily movement of a maxillary central incisor: an in vitro study. Angle Orthod 86:883–890CrossRefPubMed
4.
Zurück zum Zitat Elkholy F, Schmidt F, Jäger R et al (2017) Forces and moments applied during derotation of a maxillary central incisor with thinner aligners: an in vitro study. Am J Orthod Dentofac Orthop 151:407–415CrossRef Elkholy F, Schmidt F, Jäger R et al (2017) Forces and moments applied during derotation of a maxillary central incisor with thinner aligners: an in vitro study. Am J Orthod Dentofac Orthop 151:407–415CrossRef
6.
Zurück zum Zitat Göz G, Diedrich P, Bourauel C et al (2004) Stellungnahme zur Invisalign-Methode. Zahnärztl Mitt 94:2632–2633 Göz G, Diedrich P, Bourauel C et al (2004) Stellungnahme zur Invisalign-Methode. Zahnärztl Mitt 94:2632–2633
7.
Zurück zum Zitat Hahn W, Engelke B, Jung K et al (2010) Initial forces and moments delivered by removable thermoplastic appliances during rotation of an upper central incisor. Angle Orthod 80:239–246CrossRefPubMed Hahn W, Engelke B, Jung K et al (2010) Initial forces and moments delivered by removable thermoplastic appliances during rotation of an upper central incisor. Angle Orthod 80:239–246CrossRefPubMed
8.
Zurück zum Zitat Hahn W, Engelke B, Jung K et al (2011) The influence of occlusal forces on force delivery properties of aligners during rotation of an upper central incisor. Angle Orthod 81:1057–1063CrossRefPubMed Hahn W, Engelke B, Jung K et al (2011) The influence of occlusal forces on force delivery properties of aligners during rotation of an upper central incisor. Angle Orthod 81:1057–1063CrossRefPubMed
9.
Zurück zum Zitat Hahn W, Zapf A, Dathe H et al (2010) Torquing an upper central incisor with aligners–acting forces and biomechanical principles. Eur J Orthod 32:607–613CrossRefPubMed Hahn W, Zapf A, Dathe H et al (2010) Torquing an upper central incisor with aligners–acting forces and biomechanical principles. Eur J Orthod 32:607–613CrossRefPubMed
10.
Zurück zum Zitat Kravitz ND, Kusnoto B, Agran B et al (2008) Influence of attachments and interproximal reduction on the accuracy of canine rotation with Invisalign. A prospective clinical study. Angle Orthod 78:682–687CrossRefPubMed Kravitz ND, Kusnoto B, Agran B et al (2008) Influence of attachments and interproximal reduction on the accuracy of canine rotation with Invisalign. A prospective clinical study. Angle Orthod 78:682–687CrossRefPubMed
11.
Zurück zum Zitat Li X, Ren C, Wang Z et al (2016) Changes in force associated with the amount of aligner activation and lingual bodily movement of the maxillary central incisor. Korean J Orthod 46:65–72CrossRefPubMedPubMedCentral Li X, Ren C, Wang Z et al (2016) Changes in force associated with the amount of aligner activation and lingual bodily movement of the maxillary central incisor. Korean J Orthod 46:65–72CrossRefPubMedPubMedCentral
12.
Zurück zum Zitat Lombardo L, Martines E, Mazzanti V et al (2017) Stress relaxation properties of four orthodontic aligner materials: a 24-hour in vitro study. Angle Orthod 87:11–18CrossRefPubMed Lombardo L, Martines E, Mazzanti V et al (2017) Stress relaxation properties of four orthodontic aligner materials: a 24-hour in vitro study. Angle Orthod 87:11–18CrossRefPubMed
13.
Zurück zum Zitat Proffit WR (2007) The biologic basis of orthodontic therapy. In: Proffit WR, Fields HW, Sarver DM (eds) Contemporary orthodontics. Mosby Elsevier, St. Louis Mo, pp 331–358 Proffit WR (2007) The biologic basis of orthodontic therapy. In: Proffit WR, Fields HW, Sarver DM (eds) Contemporary orthodontics. Mosby Elsevier, St. Louis Mo, pp 331–358
14.
Zurück zum Zitat Rossini G, Parrini S, Castroflorio T et al (2015) Efficacy of clear aligners in controlling orthodontic tooth movement: a systematic review. Angle Orthod 85:881–889CrossRefPubMed Rossini G, Parrini S, Castroflorio T et al (2015) Efficacy of clear aligners in controlling orthodontic tooth movement: a systematic review. Angle Orthod 85:881–889CrossRefPubMed
15.
Zurück zum Zitat Simon M, Keilig L, Schwarze J et al (2014) Forces and moments generated by removable thermoplastic aligners: incisor torque, premolar derotation, and molar distalization. Am J Orthod Dentofac Orthop 145:728–736CrossRef Simon M, Keilig L, Schwarze J et al (2014) Forces and moments generated by removable thermoplastic aligners: incisor torque, premolar derotation, and molar distalization. Am J Orthod Dentofac Orthop 145:728–736CrossRef
Metadaten
Titel
Mechanical load exerted by PET-G aligners during mesial and distal derotation of a mandibular canine
An in vitro study
verfasst von
F. Elkholy
B. Mikhaiel
F. Schmidt
B. G. Lapatki
Publikationsdatum
29.03.2017
Verlag
Springer Medizin
Erschienen in
Journal of Orofacial Orthopedics / Fortschritte der Kieferorthopädie / Ausgabe 5/2017
Print ISSN: 1434-5293
Elektronische ISSN: 1615-6714
DOI
https://doi.org/10.1007/s00056-017-0090-4

Weitere Artikel der Ausgabe 5/2017

Journal of Orofacial Orthopedics / Fortschritte der Kieferorthopädie 5/2017 Zur Ausgabe

Mitteilungen DGKFO

Mitteilungen der DGKFO

Parodontalbehandlung verbessert Prognose bei Katheterablation

19.04.2024 Vorhofflimmern Nachrichten

Werden Personen mit Vorhofflimmern in der Blanking-Periode nach einer Katheterablation gegen eine bestehende Parodontitis behandelt, verbessert dies die Erfolgsaussichten. Dafür sprechen die Resultate einer prospektiven Untersuchung.

Invasive Zahnbehandlung: Wann eine Antibiotikaprophylaxe vor infektiöser Endokarditis schützt

11.04.2024 Endokarditis Nachrichten

Bei welchen Personen eine Antibiotikaprophylaxe zur Prävention einer infektiösen Endokarditis nach invasiven zahnärztlichen Eingriffen sinnvoll ist, wird diskutiert. Neue Daten stehen im Einklang mit den europäischen Leitlinienempfehlungen.

Zell-Organisatoren unter Druck: Mechanismen des embryonalen Zahnwachstums aufgedeckt

08.04.2024 Zahnmedizin Nachrichten

Der Aufbau von Geweben und Organen während der Embryonalentwicklung wird von den Zellen bemerkenswert choreografiert. Für diesen Prozess braucht es spezielle sogenannte „Organisatoren“. In einer aktuellen Veröffentlichung im Fachjournal Nature Cell Biology berichten Forschende durch welchen Vorgang diese Organisatoren im Gewebe entstehen und wie sie dann die Bildung von Zähnen orchestrieren.

Die Oralprophylaxe & Kinderzahnheilkunde umbenannt

11.03.2024 Kinderzahnmedizin Nachrichten

Infolge der Umbenennung der Deutschen Gesellschaft für Kinderzahnheilkunde in Deutsche Gesellschaft für Kinderzahnmedizin (DGKiZ) wird deren Mitgliederzeitschrift Oralprophylaxe & Kinderzahnheilkunde in Oralprophylaxe & Kinderzahnmedizin umbenannt. Aus diesem Grunde trägt die erste Ausgabe in 2024 erstmalig den neuen Titel.

Newsletter

Bestellen Sie unseren kostenlosen Newsletter Update Zahnmedizin und bleiben Sie gut informiert – ganz bequem per eMail.