Introduction
Methods
Requirements and recommendations
1. Geometric accuracy
1.1. General geometric accuracy
-
The position of the patient inside the MR scanner must be optimized, so that the center of the imaged volume is as close as possible to the magnet and gradient isocenter. (Consensus: 100%, abstention: 0%; Minimum requirement: 100%)
-
An end-to-end test including MRI simulation must be performed yearly after commissioning and after changes to the SRT treatment planning chain in accordance with DIN 6864‑1. (Consensus: 100%, abstention: 0%; Minimum requirement: 92%, additional recommendation: 8%)
-
The radiologic report must state that MRI sequences have been acquired for the purpose of SRT treatment planning and optimized for geometric accuracy. (Consensus: 100%, abstention: 8%; Minimum requirement: 92%, additional recommendation: 8%)
1.2. MRI distortion correction using prior knowledge
1.2.1. Gradient nonlinearity-related distortion (Fig. 1a, b)
-
Vendor 3D gradient nonlinearity distortion correction must be applied, when acquiring image datasets for SRT treatment planning. (Consensus: 100%, abstention: 0%; Minimum requirement: 100%)
-
The residual gradient non-linearity-related distortions after vendor 3D correction should be characterized using phantom measurements for quality assurance at the time of commissioning, after scanner upgrades, repairs or maintenance, and at least in yearly intervals. The maximum amount of distortion obtained via phantom measurements for the field-of-view of a typical head MRI scan (sphere of 25 cm diameter) must not exceed 1 mm [30]. If larger distortion is present, this has to be addressed by arranging the repair of hardware or software components or by performing additional correction for the remaining image distortion. (Consensus: 100%, abstention: 8%; Minimum requirement: 92%, additional recommendation: 8%)
-
Residual gradient non-linearity-related distortions after vendor correction should be characterized using phantom measurements for quality assurance in at least monthly intervals. This recommendation is derived from the ESTRO-ACROP guideline for online MRI guided radiotherapy systems and the ACR 2015 MRI quality control manual, which recommend monthly and weekly assessment of gradient non-linearity-related distortion, respectively [11, 32]. (Consensus: 91%, abstention: 15%; Additional recommendation: 100%)
-
Consider, correcting residual distortions < 1 mm after vendor distortion correction based on phantom measurements to further minimize remaining distortion. (Consensus: 92%, abstention: 0%; Optional: 100%)
1.2.2. Distortions due to B0 inhomogeneity and chemical shift
-
The pixel bandwidth must be set to at least 440 Hz (i.e., twice the fat-water shift at 1.5 T) [10]. (Consensus: 92%, abstention: 0%; Minimum requirement: 100%)
-
Active shimming must be used to actively mitigate magnetic field inhomogeneities from system imperfections and susceptibility-related inhomogeneities caused by the patient anatomy. (Consensus: 100%, abstention: 0%; Minimum requirement: 100%)
-
The main magnetic field homogeneity must be characterized after installation (baseline), after scanner upgrades, repairs or maintenance, and at least in yearly intervals, as detailed in sources such as [32] and [33]. If necessary, arrange for repairs to maintain field homogeneity and ensure necessary geometric accuracy. (Consensus: 100%, abstention: 0%; Minimum requirement: 100%)
-
Standard operating procedures must be established to minimize the introduction of small metallic objects (e.g., hairpins) and metallic dust (e.g., from shoes) into the scanner bore that could degrade magnetic field homogeneity and geometric accuracy. (Consensus: 100%, abstention: 0%; Minimum requirement: 100%)
-
Screening for metallic objects inside the scanner bore that could degrade magnetic field homogeneity and geometric accuracy must be performed daily. In every patient, check for metal artifacts in gradient echo-based localizer images before acquiring images for treatment planning. (Consensus: 100%, abstention: 0%; Minimum requirement: 100%)
-
Consider, increasing the pixel bandwidth to at least 660 Hz to further reduce distortions due B0 inhomogeneities and chemical shift (i.e., three-times the fat-water shift at 1.5 T). Increasing the pixel bandwidth might entail more averages need to be acquired to preserve SNR and lesion conspicuity. (Consensus: 92%, abstention: 8%; Optional: 100%)
-
Consider, individually characterizing main magnetic field inhomogeneities from system imperfections and susceptibility-related inhomogeneities for patients undergoing MRI simulation by using B0 mapping. (Consensus: 92%, abstention: 8%; Optional: 100%)
-
Consider, individually correcting residual distortions because of magnetic field inhomogeneities based on individual B0 mapping or reverse gradient methods. (Consensus: 100%, abstention: 8%; Optional: 100%)
1.3. Distortion correction using image registration
-
The use of registration-based distortion correction in addition to the minimum requirements may have a supplementary role in certain settings. (Consensus 92%, abstention: 0%; Optional: 100%)
2. Optimal sequence selection and optimization of sequence protocol parameters
-
MRI protocols must be used that include at least one 3D sequence (e.g., 3D-T1w) with a sufficient signal-to-noise ratio (SNR) for target delineation. (Consensus: 100%, abstention: 0%; Minimum requirement: 100%)
-
Standardized MRI protocols must be set up and used for cranial stereotactic treatment planning indications. These standardized protocols must be characterized by a unique and easily understandable study description (e.g., “RT treatment planning—brain metastases”). (Consensus: 100%, abstention: 0%; Minimum requirement: 100%)
-
The main 3D sequence must be isovolumetric with a voxel size of ≤ 1 mm3. (Consensus: 100%, abstention: 0%; Minimum requirement: 100%)
-
Choose the 3D-T1w sequence that provides the best target conspicuity and the most accurate characterization of 3D tumor boundaries. For a substantial fraction of patients, treatment indications and MR scanners, 3D-T1w FSE/TSE sequence protocols are to be preferred over 3D-T1w IR-GE sequence protocols [55, 59, 64]. If multiple 3D sequence protocols for target delineation are acquired, generally, the gross target volume should encompass the extent of the tumor in all 3D sequences. (Consensus: 100%, abstention: 0%; Minimum requirement: 85%, Additional recommendation: 15%)
2.1. Indication-specific considerations for sequence selection
3D-T1w FSE/TSE | 3D-T1w IR-GE | 3D-T2-FLAIR FSE/TSE | 3D-T2w FSE/TSE | 3D-True FISP/Dual Excitation | |
---|---|---|---|---|---|
Type | 3D | 3D | 3D | 3D | 3D |
Orientation | Transversal or Sagittal | Transversal or Sagittal | Transversal or Sagittal | Transversal | Transversal |
TE | Minimuma | Minimuma | e.g., 374 ms | e.g., 182 ms (heavily T2w) e.g., 92 ms (moderately T2w) | e.g., 2.45 ms |
TR | e.g., 550–750 ms | e.g., 2100–2200 ms | e.g., 7000 ms | e.g., 1200 ms | e.g., 5.47 ms |
TI | – | e.g., 900–1100 ms | e.g., 2050 ms | – | – |
Acq. matrix | ≥ 256 × 256 | ≥ 256 × 256 | ≥ 256 × 256 | ≥ 256 × 256 | ≥ 256 × 256 |
Acq. in-plane resolution | ≤ 1 × 1 mm | ≤ 1 × 1 mm | ≤ 1 × 1 mm | ≤ 0.7 × 0.7 mm | ≤ 0.7 × 0.7 mm |
Slice thickness | ≤ 1 mm | ≤ 1 mm | ≤ 1 mm | ≤ 0.7 mm | ≤ 0.7 mm |
Fat saturation | Optional | – | – | – | – |
Post-contrast interval | ≥ 5 min | ≥ 5 min | – | – | – |
GNL Distortion correction | Vendor 3D ± in-houseb | Vendor 3D ± in-houseb | Vendor 3D ± in-houseb | Vendor 3D ± in-houseb | Vendor 3D ± in-houseb |
Shimming of B0 inhomogeneities | Patient-specific active shimming | Patient-specific active shimming | Patient-specific active shimming | Patient-specific active shimming | Patient-specific active shimming |
Readout bandwidth | ≥ 440 Hz (≥ 660 Hz recommended) | ≥ 440 Hz (≥ 660 Hz recommended) | ≥ 440 Hz (≥ 660 Hz recommended) | ≥ 440 Hz (≥ 660 Hz recommended) | ≥ 440 Hz (≥ 660 Hz recommended) |
Interval to treatment | ≤ 14 days (≤ 5 days recommended) | ≤ 14 days (≤ 5 days recommended) | ≤ 14 days (≤ 5 days recommended) | ≤ 14 days (≤ 5 days recommended) | ≤ 14 days (≤ 5 days recommended) |
Brain metastases | Meningioma | Vestibular schwannoma | Pituitary adenoma | Trigeminal neuralgia | AVM | Glomus tumors |
---|---|---|---|---|---|---|
2D-T1w pre (optional: 3D-T1w pre) | 3D-T1w pre | 2D-T1w pre (optional: 3D-T1w pre) | 2D-T2w FSE/TSE cor | 3D-T1w pre | 2D-T1w pre (optional: 3D-T1w GE pre) | 3D-T1w pre |
Contrast administration | Contrast administration | Contrast administration | 2D-T2-FLAIR tra | 3D-TOF | 3D-TOF | Contrast administration |
2D-T2-FLAIR tra | 2D-T2-FLAIR tra | 2D-T2-FLAIR | 3D-T1w pre | Contrast administration | 3D-T2w FSE/TSE | 3D-T2w FSE/TSE / 3D-True FISP-Dual Excitation |
3D-T1w post early | 3D-T1w post | 3D-T1w post | Contrast administration | 3D-T2w FSE/TSE / 3D-True FISP-Dual Excitation | Contrast administration | 3D-T1w post |
(Optional: 3D-T1w post late) | Subtraction 3D-T1w post − T1w pre | 3D-T2w FSE/TSE / 3D-True FISP-Dual Excitation | 2D-T1w dynamic cor | 3D-T1w post | 3D-T1w GE post | Subtraction 3D-T1w post − T1w pre |
3D-T1w post | (Optional: Subtraction 3D-T1w GE post − pre) | |||||
Subtraction 3D-T1w post − T1w pre |
Parameter | Minimum requirement | Additional recommendation | Optional |
---|---|---|---|
End-to-end test including MRI simulation | Yearly, after commissioning and after changes to the SRT treatment planning chain in accordance with DIN 6864‑1 | – | – |
Residual gradient non-linearity-related distortions after vendor 3D correction | Characterize using phantom measurements at the time of commissioning, after scanner upgrades, repairs or maintenance, and at least in yearly intervals. Maximum amount of distortion obtained via phantom measurements for the field-of-view of a typical head MRI scan (sphere of 25 cm diameter) must not exceed 1 mm | In addition to minimum requirements: Characterization of residual gradient non-linearity related distortions in at least monthly intervals using phantom measurements | In addition to additional recommendations: Correction of residual distortions ≤ 1 mm based on phantom measurements |
Main magnetic field (B0) homogeneity | – | In addition to minimum requirements: Individually characterize main magnetic field inhomogeneities from system imperfections and susceptibility-related inhomogeneities for patients undergoing MRI simulation by using B0 mapping. Perform individual corrections based on individual B0 mapping or reverse gradient methods | |
Screening for metallic objects | Daily check the scanner bore for small metallic objects; Check gradient echo-based localizer images for metal artifacts in every patient | – | – |
Registration algorithm | Registration error within ≤ 1 mm for registration of phantoms at commissioning | – | – |
Image quality for flexible coil systems | – | – | Monthly |
3. Contrast enhancement
-
For intraaxial tumors, the time interval between contrast administration and the start of the acquisition of the main T1w sequence should be at least 5 min (see discussion in main text). (Consensus: 100%, abstention: 0%; Minimum requirement: 100%)
-
Given the improved lesion conspicuity with increased contrast dose, administration of double-dose contrast may be considered in specific circumstances, if the individual benefit of improved tumor delineation for treatment planning clearly outweighs individual GBCA-associated risks. (Consensus: 100%, abstention: 15%; Additional recommendation: 9%, Optional: 91%)
-
An additional delayed T1w sequence protocol may be acquired 15–20 min after the first contrast administration, with or without repeated contrast administration to improve target conspicuity and depiction of target boundaries. (Consensus: 92%, abstention: 8%; Optional: 100%)
4. Time interval between MRI simulation and treatment delivery
-
The time interval between the MRI simulation and the administration of treatment must not be larger than 14 days. (Consensus: 100%, abstention: 8%; Minimum requirement: 100%)
-
In brain metastases and primary brain tumors CNS-WHO-grade 2–4, the time interval between MR simulation and treatment delivery should not be greater than 5 days. (Consensus: 100%, abstention: 8%; Minimum requirement: 17%, additional recommendation: 83%)
-
Consider, performing an additional simulation MRI for adaption of target volumes every 5 fractions in fractionated stereotactic radiotherapy (≤ 12 fractions). (Consensus: 100%, abstention: 0%, Optional: 100%)
6. Image registration and imaging in SRT position (Fig. 6)
-
When registering simulation MRI datasets to a planning CT, both the planning CT and the MRI dataset, i.e., both registration pairs, must have a slice thickness of ≤ 1 mm. (Consensus: 100%, abstention: 0%; Minimum requirement: 92%, additional recommendation: 8%)
-
A proper registration algorithm, commissioned, and validated for stereotactic radiotherapy/radiosurgery followed by expert correction and verification must be used. (Consensus: 100%, abstention: 0%; Minimum requirement: 100%)
-
With quantitative validation at commissioning, the registration algorithm must achieve a registration error within ≤ 1 mm for registration of phantoms. (Consensus: 100%, abstention: 0%; Minimum requirement: 100%)
-
Artifacts that could impair registration must be minimized: Set a sufficient phase oversampling factor to avoid folding artifacts, minimize distortions in MRI, minimize metal artifacts in MRI (e.g., by using 3D-T1w FSE/TSE instead of 3D-T1w GE sequence protocols) and minimize artifacts in planning CTs (e.g., minimize metal artifacts from dental fillings). (Consensus: 100%, abstention: 0%, Minimum requirement: 92%, additional recommendation: 8%)
-
Motion artifacts inside the diagnostic head coil must be minimized by proper use of cushions and foam elements. (Consensus: 100%, abstention: 0%; Minimum requirement: 92%, additional recommendation: 8%)
-
In daily clinical practice, registration quality for each treatment planning registration must be verified qualitatively by a board-certified physician or medical physics expert with experience in cranial stereotactic radiotherapy. This verification should be performed using an overlay method (e.g., alpha blending with or without varying color schemes for both datasets, a checkerboard pattern, and/or “spy glass” tools), with dynamic assessment of registration accuracy (e.g., by blending between datasets or by moving the checkerboard pattern and “spy glass” tool). Site-specific recommendations can be found in [21], for example. (Consensus: 100%, abstention: 0%; Minimum requirement: 100%)
-
When using multiple sequences for treatment planning, registration accuracy must be individually verified for each sequence used for treatment planning, because of the risk of motion between sequences. (Consensus: 100%, abstention: 0%; Minimum requirement: 100%)
-
Consider, acquiring simulation MRIs in the treatment position with mask immobilization to improve registration accuracy and to reduce/eliminate motion artifacts. Sufficient image quality must be ensured when acquiring simulation MRIs in the treatment position with a flexible coil setup. (Consensus: 85%, abstention: 0%; Additional recommendation: 9%, Optional: 91%)
-
Especially consider acquisition in the treatment position with mask immobilization if nonrigid tissue deformations are expected between the treatment position and the diagnostic imaging position (e.g., for targets near the foramen magnum and if rigidity of the skull is significantly impaired after surgery). (Consensus: 92%, abstention: 0%; Additional recommendation: 8%, Optional: 92%)
-
Consider, performing monthly checks on image quality for flexible coil systems employed for stereotactic radiotherapy simulation [10]. (Consensus: 85%, abstention: 0%; Additional recommendation: 9%, optional: 91%)
-
Use of synthetic CT and an MR-only workflow may be considered to exclude MRI-CT registration uncertainties. If an MR-only workflow is used, it must be ensured that synthetic CT datasets can be used for both treatment planning and image guidance. In addition, motion between MRI sequences used for synthetic CT calculation and sequences used for target delineation must be excluded or addressed by registration. (Consensus: 92%, abstention: 0%; Minimum requirement: 8%, additional recommendation: 8%, optional: 83%)