Skip to main content
Erschienen in: Strahlentherapie und Onkologie 1/2024

Open Access 02.01.2024 | Review Article

Quality requirements for MRI simulation in cranial stereotactic radiotherapy: a guideline from the German Taskforce “Imaging in Stereotactic Radiotherapy”

verfasst von: PD Dr. Florian Putz, Michael Bock, Daniela Schmitt, Christoph Bert, Oliver Blanck, Maximilian I. Ruge, Elke Hattingen, Christian P. Karger, Rainer Fietkau, Johanna Grigo, Manuel A. Schmidt, Tobias Bäuerle, Andrea Wittig

Erschienen in: Strahlentherapie und Onkologie | Ausgabe 1/2024

Abstract

Accurate Magnetic Resonance Imaging (MRI) simulation is fundamental for high-precision stereotactic radiosurgery and fractionated stereotactic radiotherapy, collectively referred to as stereotactic radiotherapy (SRT), to deliver doses of high biological effectiveness to well-defined cranial targets. Multiple MRI hardware related factors as well as scanner configuration and sequence protocol parameters can affect the imaging accuracy and need to be optimized for the special purpose of radiotherapy treatment planning. MRI simulation for SRT is possible for different organizational environments including patient referral for imaging as well as dedicated MRI simulation in the radiotherapy department but require radiotherapy-optimized MRI protocols and defined quality standards to ensure geometrically accurate images that form an impeccable foundation for treatment planning. For this guideline, an interdisciplinary panel including experts from the working group for radiosurgery and stereotactic radiotherapy of the German Society for Radiation Oncology (DEGRO), the working group for physics and technology in stereotactic radiotherapy of the German Society for Medical Physics (DGMP), the German Society of Neurosurgery (DGNC), the German Society of Neuroradiology (DGNR) and the German Chapter of the International Society for Magnetic Resonance in Medicine (DS-ISMRM) have defined minimum MRI quality requirements as well as advanced MRI simulation options for cranial SRT.
Hinweise

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Introduction

Magnetic Resonance Imaging (MRI) has a long-standing role in radiotherapy treatment planning for brain tumors and is essential today in nearly all cranial treatment indications [1]. Stereotactic Radiosurgery (SRS) and Fractionated Stereotactic Radiotherapy (FSRT), collectively referred to as Stereotactic Radiotherapy (SRT), demand exceptional precision to achieve doses of high biological effectiveness in tumors while simultaneously preserving the adjacent normal tissues and organs-at-risk (OARs) [2, 3]. This paradigm of cranial SRT has been shown to achieve high local cure rates with limited to minimal toxicities [46]. The stringent precision requirements of cranial SRT necessitates not only highly precise treatment planning and delivery, but also critically depends on optimal MRI for target volume definition. The previous German guideline on technological quality requirements for stereotactic radiotherapy explicitly pointed out that imaging for SRT treatment planning needs exceptional attention where “the target volume and all organs-at-risk are defined using organ-specific imaging modalities” and “secondary imaging requires accurate registration with the thin-slice planning computed tomography (CT)” [3]. This is especially important for MRI as an integral part of cranial SRT which most often forms the fundamental basis for all further treatment steps. Errors and uncertainties in MRI can often not be compensated at later stages in the treatment planning process and are propagated throughout the treatment planning chain, possibly leading to suboptimal treatment with inferior outcome. Multiple groups clearly demonstrated that inadequate MRI would affect the accuracy of gross target volume (GTV) delineation, which can easily diminish the clinical outcome [79]. Simply expanding the margins to compensate for uncertainties and imaging errors will increase the planning target volume (PTV), consecutively decrease the therapeutic selectivity, and will ultimately diminish the therapeutic principles of precision radiotherapy. Cranial stereotactic radiotherapy was outside the scope of recently published guidelines on MRI simulation in radiotherapy, like the AAPM task group 284 report [10, 11] and a comprehensive guideline on MRI simulation for cranial SRT was missing. Optimal MRI simulation and target volume definition frequently require interdisciplinary input and close collaboration, integrating radiooncologic, diagnostic, neurosurgical and physics expertise. In this guideline, therefore, an interdisciplinary panel including experts from the working group for radiosurgery and stereotactic radiotherapy of the German Society for Radiation Oncology (DEGRO), the working group for physics and technology in stereotactic radiotherapy of the German Society for Medical Physics (DGMP), the German Society of Neurosurgery (DGNC), the German Society of Neuroradiology (DGNR) and the German Chapter of the International Society for Magnetic Resonance in Medicine (DS-ISMRM) have defined minimum quality requirements as well as advanced simulation options for MRI in cranial SRT to increase its quality in clinical practice and ultimately improve treatment outcomes.

Methods

To formulate this guideline paper, a cross-disciplinary expert task force was constituted. The task force contained experts from radiation oncology, medical physics, neurosurgery, neuroradiology, radiology and MRI physics. The task force conducted biweekly virtual meetings from December 2022 to July 2023. First a systematic literature review on the areas of geometric accuracy, sequence selection and optimization of sequence protocol parameters, contrast agent-related parameters, time interval between MRI simulation and treatment delivery as well as image registration and imaging in SRT position was performed. This literature review was revisited, evaluated, and refined through iterative discussions in the regular meetings. Subsequently, requirements and recommendations were derived from the literature review with consensus being optimized in a two-stage process. Following an initial round of voting, the proposed requirements and recommendations were re-evaluated and fine-tuned in interdisciplinary discussions before proceeding to a final voting round. Each requirement and recommendation was subjected to a vote on agreement (with possible responses: “yes,” “no,” or “abstention”) and, in the case of agreement, the category (“minimum requirement”, “additional recommendation”, or “optional”). Guideline statements that were classified as “minimum requirement” were considered mandatory for MRI simulation in SRT. “Additional recommendations” were defined as recommendations that should be applied for optimal MRI simulation but are not considered mandatory. Finally, “optional” statements provide advanced options that can be implemented by experienced centers. Consensus was quantified as the percentage of agreement, excluding abstentions. For each requirement and recommendation, the rates of consensus, abstention, and the votes on the statement category are provided.

Requirements and recommendations

1. Geometric accuracy

1.1. General geometric accuracy

Distortions in MRI may result from multiple mechanisms that may compromise the precise delivery of treatment [1215] and will be discussed in detail below. Generally, MRI distortions are nonlinear and unevenly distributed across image datasets [13, 16, 17]. Most hardware-related distortions occur in the periphery and least distortions near the isocenter of the magnet [1821]. Therefore, with cranial SRT, hardware-related distortions are expected to occur near the cortical surface of the brain, when the head is placed at the isocenter of the magnet [13, 16]. Another source of distortions is patient-specific: for example, prominent distortions can appear near air-bone interfaces at the frontopolar and orbitofrontal cortex, in addition to the cranial aspects of the prefrontal cortex and the lateral and inferior portions of the temporal lobes [17, 22].
General geometric accuracy—Minimum requirements
  • The position of the patient inside the MR scanner must be optimized, so that the center of the imaged volume is as close as possible to the magnet and gradient isocenter. (Consensus: 100%, abstention: 0%; Minimum requirement: 100%)
  • An end-to-end test including MRI simulation must be performed yearly after commissioning and after changes to the SRT treatment planning chain in accordance with DIN 6864‑1. (Consensus: 100%, abstention: 0%; Minimum requirement: 92%, additional recommendation: 8%)
  • The radiologic report must state that MRI sequences have been acquired for the purpose of SRT treatment planning and optimized for geometric accuracy. (Consensus: 100%, abstention: 8%; Minimum requirement: 92%, additional recommendation: 8%)

1.2. MRI distortion correction using prior knowledge

Two types of MRI distortions are most relevant to cranial stereotactic radiotherapy: distortions that are caused by gradient coil nonlinearities and distortions that originate from inhomogeneities of the main magnetic field (B0) [23]. The B0 inhomogeneities in turn are caused by residual imperfections in the main magnet and by the magnetic susceptibility differences in the tissues of the patient which lead to static field inhomogeneities [15, 23, 24]. These distortions can be substantially reduced using prior knowledge about the image acquisition process. Optimized parameter settings are selected for the MR image acquisition to produce the most geometrically accurate images possible. Remaining distortions in the images are corrected in a distortion correction post-processing step. MRI distortion correction using prior knowledge is the preferred approach to achieve geometric accuracy and will be discussed in detail in subsequent Sects. 1.2.1. and 1.2.2.

1.2.1. Gradient nonlinearity-related distortion (Fig. 1a, b)

Gradient nonlinearity-related distortions are usually the most significant type of distortion in cranial MRI [12]. Spatial encoding in MRI is based on spatially linearly varying magnetic fields (so-called gradients) that are created by 3 independent gradient coils in x, y, and z direction [12, 13, 15, 23, 25, 26]. However, due to inevitable physical constraints (e.g., Maxwell’s equation) in gradient coil design, gradient fields deviate from their linear behavior with increasing distance to the center of the gradient coils (which is typically coincident with the magnet isocenter) [26, 27]. During image reconstruction, these gradient nonlinearities result in spatial distortions that increase with the distance from the isocenter [15, 18, 21, 27, 28]. Gradient nonlinearities are specific to every gradient coil and are therefore constant for a given MRI system [12, 13, 15, 18, 24, 27]. As gradient nonlinearities are linked to the absolute position of the gradient coil, image distortions will vary when the patient is positioned differently relative to the gradient fields [18, 24]. The amount of distortion from gradient nonlinearities therefore depends on the MRI system model (more specifically: the type of installed gradient coil) and the patient position relative to the isocenter. Image distortions can reach up to several millimeters at the brain periphery [18, 27].
Gradient nonlinearities are a fixed property of a given gradient coil type which are known by the manufacturer [26, 27]; thus, they can be corrected using a vendor-specific distortion correction. Vendor-specific gradient non-linearity distortion correction is usually implemented as a post-processing step using deformable registration (image warping), resampling and intensity correction [18, 27, 29]. Interestingly, gradient nonlinearity correction can be performed without having access to the MRI system: some MRI manufacturers offer distortion correction software that contains a library of distortion fields of all their gradient coils. By identifying the specific gradient coil in the DICOM metadata, distortions can be retrospectively corrected in any study acquired on one of their MRI systems. While vendors also offer a 2D distortion correction option, only 3D correction rectifies gradient nonlinearity-related distortion in all dimensions and therefore is the minimum requirement for SRT treatment planning.
The clinical impact of gradient nonlinearity-related distortions was investigated by Seibert et al. in cranial radiosurgery by comparing 3D corrected and uncorrected images. When uncorrected images were used, they found an average GTV displacement of 1.2 mm and a maximum GTV displacement of 3.9 mm. This would have resulted in 8 of 28 lesions experiencing geographic miss [9].
Residual distortion may remain after vendor-specific 3D correction. The amount of residual distortion should be characterized regularly with phantom measurements and corrected if necessary. Evaluating MRI geometric accuracy with phantom measurements is recommended during MRI system installation to obtain baseline data, and after major repair or and inspection work, which is typically scheduled at 6‑month intervals ([10, 3032]; Fig. 1d). Multiple groups have reported how residual gradient nonlinearity distortions can be corrected by obtaining 3D deformation vector fields from phantom measurements [12, 26, 33]. Typically, a grid structure with known geometry is imaged and the deviation of the grid locations in the images from the known positions are quantified and extrapolated to the full image [34].
Gradient nonlinearity-related distortion—Minimum requirements
  • Vendor 3D gradient nonlinearity distortion correction must be applied, when acquiring image datasets for SRT treatment planning. (Consensus: 100%, abstention: 0%; Minimum requirement: 100%)
  • The residual gradient non-linearity-related distortions after vendor 3D correction should be characterized using phantom measurements for quality assurance at the time of commissioning, after scanner upgrades, repairs or maintenance, and at least in yearly intervals. The maximum amount of distortion obtained via phantom measurements for the field-of-view of a typical head MRI scan (sphere of 25 cm diameter) must not exceed 1 mm [30]. If larger distortion is present, this has to be addressed by arranging the repair of hardware or software components or by performing additional correction for the remaining image distortion. (Consensus: 100%, abstention: 8%; Minimum requirement: 92%, additional recommendation: 8%)
Gradient nonlinearity-related distortion—Additional recommendations
  • Residual gradient non-linearity-related distortions after vendor correction should be characterized using phantom measurements for quality assurance in at least monthly intervals. This recommendation is derived from the ESTRO-ACROP guideline for online MRI guided radiotherapy systems and the ACR 2015 MRI quality control manual, which recommend monthly and weekly assessment of gradient non-linearity-related distortion, respectively [11, 32]. (Consensus: 91%, abstention: 15%; Additional recommendation: 100%)
Gradient nonlinearity-related distortion—Optional
  • Consider, correcting residual distortions < 1 mm after vendor distortion correction based on phantom measurements to further minimize remaining distortion. (Consensus: 92%, abstention: 0%; Optional: 100%)

1.2.2. Distortions due to B0 inhomogeneity and chemical shift

As with non-linearities of gradient coils, inhomogeneities of the main magnetic field (B0 inhomogeneities) also result in distortions. MRI B0 inhomogeneities are caused by imperfections in magnet design, but, more importantly, by magnetic perturbations induced by the patient [13]. Locally, B0 inhomogeneities cause static field gradients that are superimposed onto the dynamic image encoding gradients—thus, they affect both the slice selection and the readout process [13]. In standard 3D sequences, gradient nonlinearity-related displacements occur in all three dimensions, while B0 inhomogeneity-related distortions occur only in frequency-encoding (readout) direction. B0 inhomogeneities also affect the slice selection in 2D sequences resulting in non-ideal slice geometries that diverge from a perfectly planar configuration. Thus, 2D sequences are more susceptible to B0 inhomogeneity-induced distortions than 3D sequences [12, 13, 17].
Changes in magnetic susceptibility at tissue-tissue or air-tissue interfaces lead to patient-induced perturbations of the main magnetic field ([13, 23, 35]; Fig. 1c). Susceptibility-induced magnetic field distortions are directly proportional to the field strength—thus, larger distortions are expected at higher fields if the same imaging protocols are applied. The greatest susceptibility differences and, therefore, the largest distortions occur at air-bone interfaces and near metal implants [23, 35]. In cranial SRT, the most severe susceptibility-related distortions are expected near the paranasal sinuses and mastoid cells [17]. An analysis of a brain imaging protocol using a 3D gradient echo sequence with T1-weighted preparation (T1-MPRAGE) conducted in 2013 by Wang et al. revealed that susceptibility-induced distortions were < 0.5 mm in 86.9% of the imaged volume (3 T; using a rather low readout bandwidth of 180 Hz/pixel; patient-specific automated shimming activated). Although displacement averages were low for the entire imaged volume, sinus air-bone boundaries showed 1.6 mm of average distortion. Despite distortions degraded with distance from the sinuses, they extended into the adjacent brain and optic system, and still measured 0.8 mm at a distance of 12 mm, which is clinically relevant for stereotactic targets located close to the sinuses and mastoid cells [17]. Furthermore, large differences in susceptibility and, consequently, significant distortions can also occur near metallic implants, such as surgical clips [17]. It is possible to reduce distortions due to B0 inhomogeneities by increasing the readout bandwidth, by using 3D rather than 2D sequences [36], and by activating patient-specific active shimming.
Shim coils create additional magnetic fields that can partially compensate both system-related and patient-induced B0 inhomogeneities. Shim coils are typically installed in the MR magnet bore, or, additionally, in specialized RF coils to provide a more local field optimization. During shimming, at first the B0 inhomogeneities are measured using a fast field mapping protocol, followed by the calculation of the shim coil currents needed to minimize the field distortion in a given target region [17, 23, 37, 38]. Shimming is typically an iterative procedure which converges rapidly. In active shimming on 2D multi-slice MRI the shim currents are adapted to each slice separately—however, not all MRI systems support this feature. Even though shimming can improve the geometric accuracy during imaging, it should be used with care, as the local distortion fields will change after each re-adjustment of the shim currents. This can be especially challenging for radiotherapy applications, in which imaging protocols are used that automatically perform a shim (e.g., in diffusion-weighted echo planar (EPI) imaging).
Patient-specific active shimming and RT-optimized bandwidth settings (see below) may not be entirely effective in minimizing B0 inhomogeneity-related distortions. If additional correction is required, advanced techniques have been described that involve acquiring an improved higher resolution B0 map with the patient in the scanner that subsequently can be used for further rectification of patient-induced distortions via image post-processing [15, 39]. Reverse gradient methods have also been proposed to correct B0 inhomogeneity-related distortions but require obtaining every sequence twice with the opposite frequency encoding direction [30].
Furthermore, because wear and tear of individual components, incorrect software settings and small metallic objects such as ear pins left inside the magnet can cause unnoticed distortions [30, 36], regular quality assurance is recommended to ensure optimal images for stereotactic radiotherapy ([30]; Table 3). It is recommended to evaluate the main magnetic field homogeneity after installation (baseline), after each repair and maintenance session at the scanner, and at regular intervals, with a minimum frequency of at least once per year. To prevent the introduction of small metallic objects and dust into the scanner bore that could degrade magnetic field homogeneity and cause subtle distortion [36] it is essential to establish appropriate standard operating procedures. In addition, some MR scanner manufacturers offer a daily quick check of main magnetic field homogeneity for daily quality assurance. Moreover, gradient echo-based localizer images also offer an efficient means of screening for metal-induced distortions in each patient before acquiring sequences for radiotherapy treatment simulation. This MRI simulation quality assurance process is well complemented by an end-to-end test that encompasses MR imaging ([40], Table 3). Chemical shifts, such as the fat-water shift, also belong to the group of patient-related distortions. In the case of the fat-water shift, the different resonance frequencies of fat and water cause a shift of the fat-containing tissue along the frequency encoding direction [41].
Role of the main magnetic field strength (B0)
In general, there is an ongoing debate which magnetic field strength is best suited for RT applications [24]. With increasing B0 also the SNR in the images increases so that smaller lesions can be better detected. On the other hand, the tissue parameters T1 and T2 change with field strength and the imaging protocols need to be adapted to achieve the same contrast [42, 43]. Often similar gradient systems are used at 1.5 T and 3 T so that gradient-related distortions are comparable. Chemical shifts and susceptibility-related distortions on the other hand are proportional to B0 demanding higher readout bandwidths (i.e., stronger readout gradients) to minimize their effect. Thus, the SNR benefit of higher fields is often partially compensated by the need to acquire geometrically accurate images: with increasing readout bandwidth, all distortions resulting from B0 inhomogeneities decrease reciprocally [17, 23, 37, 38], but the SNR is inversely correlated with the square root of the read-out bandwidth [38]. In order to reduce distortions caused by B0 inhomogeneities, radiotherapy planning usually requires a field strength dependent higher read-out bandwidth than routine diagnostic imaging [24]. These higher readout bandwidths will also minimize chemical shifts (see above) [41].
The loss of SNR caused by a higher read-out bandwidth can be compensated by increasing the measurement time, optimizing the coil selection, and reducing motion artifacts with immobilization. Despite that 3 T scanners would be expected to suffer more from distortions related to B0 inhomogeneity, the profound SNR increase associated with higher field strengths also enables larger compensatory readout bandwidths. MR scanners operating at 3 T, therefore, generally can also be suitable for acquiring simulation MRI studies for stereotactic radiotherapy treatment planning [12, 37, 44].
More recently, MRI systems with 0.35 T and less have been built which are combined with a RT system for online tumor imaging. The suitability of these low-field systems for real-time tumor tracking has been demonstrated, but their use in treatment of small brain targets still needs further evaluation [45, 46].
Distortions due to B0 inhomogeneity and chemical shift—Minimum requirements
  • The pixel bandwidth must be set to at least 440 Hz (i.e., twice the fat-water shift at 1.5 T) [10]. (Consensus: 92%, abstention: 0%; Minimum requirement: 100%)
  • Active shimming must be used to actively mitigate magnetic field inhomogeneities from system imperfections and susceptibility-related inhomogeneities caused by the patient anatomy. (Consensus: 100%, abstention: 0%; Minimum requirement: 100%)
  • The main magnetic field homogeneity must be characterized after installation (baseline), after scanner upgrades, repairs or maintenance, and at least in yearly intervals, as detailed in sources such as [32] and [33]. If necessary, arrange for repairs to maintain field homogeneity and ensure necessary geometric accuracy. (Consensus: 100%, abstention: 0%; Minimum requirement: 100%)
  • Standard operating procedures must be established to minimize the introduction of small metallic objects (e.g., hairpins) and metallic dust (e.g., from shoes) into the scanner bore that could degrade magnetic field homogeneity and geometric accuracy. (Consensus: 100%, abstention: 0%; Minimum requirement: 100%)
  • Screening for metallic objects inside the scanner bore that could degrade magnetic field homogeneity and geometric accuracy must be performed daily. In every patient, check for metal artifacts in gradient echo-based localizer images before acquiring images for treatment planning. (Consensus: 100%, abstention: 0%; Minimum requirement: 100%)
Distortions due to B0 inhomogeneity and chemical shift—Optional
  • Consider, increasing the pixel bandwidth to at least 660 Hz to further reduce distortions due B0 inhomogeneities and chemical shift (i.e., three-times the fat-water shift at 1.5 T). Increasing the pixel bandwidth might entail more averages need to be acquired to preserve SNR and lesion conspicuity. (Consensus: 92%, abstention: 8%; Optional: 100%)
  • Consider, individually characterizing main magnetic field inhomogeneities from system imperfections and susceptibility-related inhomogeneities for patients undergoing MRI simulation by using B0 mapping. (Consensus: 92%, abstention: 8%; Optional: 100%)
  • Consider, individually correcting residual distortions because of magnetic field inhomogeneities based on individual B0 mapping or reverse gradient methods. (Consensus: 100%, abstention: 8%; Optional: 100%)

1.3. Distortion correction using image registration

Registration-based distortion correction is a generic method to reduce MR image distortion, but inferior to the prior knowledge-based methods discussed above. With registration-based distortion correction, MRI distortions are reduced via a specialized non-rigid registration to a planning CT. This correction method does not require any prior knowledge about the MRI acquisition process, but it estimates a 3D deformation field via registration to the planning CT which is assumed to be geometrically correct. Registration-based distortion correction was able to improve distortions in past analyses, albeit some studies only analyzed phantom data but no clinical cases [47, 48].
Registration-based distortion correction has some principal limitations: 1) the accuracy of non-rigid registration is known to decrease with increasing deformation between image datasets [47, 49, 50] 2) in contrast to other distortion correction methods that use prior knowledge, registration-based solutions do not guarantee improvement of geometric accuracy of corrected image datasets in every clinical case, as their iterative algorithms can converge locally or may simply be underconstrained in case of poor visibility of targets (homogeneous radiodensity of targets and surrounding tissues) in the CT images [50, 51]. Registration-based distortion correction should therefore only have a role as a supplementary tool after optimal prior knowledge-based distortion correction.
Distortion correction using image registration—Optional
  • The use of registration-based distortion correction in addition to the minimum requirements may have a supplementary role in certain settings. (Consensus 92%, abstention: 0%; Optional: 100%)

2. Optimal sequence selection and optimization of sequence protocol parameters

Patients referred to stereotactic radiotherapy generally have their diagnosis established through diagnostic imaging beforehand. The primary objective of MRI for radiotherapy planning therefore is to accurately depict the tumor location and shape in three-dimensional space so that the gross tumor volume (GTV) can be precisely delineated. Generally, high-resolution isotropic 3D sequences are optimal for this task since they permit accurate multiplanar reconstruction and minimize over- or underestimation of the GTV because of partial-volume effects ([52, 53]; Fig. 2). Additionally, 3D sequences provide continuous imaging of the brain and targets without gaps and demonstrate lower susceptibility to distortions arising from B0 inhomogeneities compared to 2D sequences [12, 13, 17] (see above for a detailed discussion on optimizing sequence parameters for geometric accuracy). As a rule of thumb, the volumetric error will exceed 10%, if the target is visualized on less than 5 slices, which is particularly relevant for small targets such as brain metastases [53]. Partial volume effects because of large slice thickness will mostly result in an overestimation of the GTV volume. Moreover, thick slices and image gaps can also lead to an underestimation of tumor growth perpendicular to the imaging plane or miss small tumor parts [16, 53].
The most used 3D-T1w MR sequences for brain tumors have been T1w inversion-recovery gradient echo (IR-GE) sequences such as the T1-MPRAGE [54, 55]. For cranial radiotherapy target volume delineation, however, multiple studies have recently suggested that 3D-T1w fast or turbo spin echo (FSE/TSE) sequences may in fact be superior to 3D-T1w IR-GE sequences (Fig. 3; [5559]). Since 3D-T1w FSE/TSE sequences show less contrast between gray and white matter [55], they frequently improve the contrast ratio of contrast-enhancing lesions and low-contrasting brain parenchyma as background for targets. An equally beneficial effect for radiotherapy planning purposes is the suppression of vessels in 3D-T1w FSE/TSE sequences [58], which can facilitate discrimination of small tumors from transverse vessel cross-sections. For T1w IR-GE sequences, on the other hand, an additional well-described caveat is that they suffer from reduced visibility of enhancement when low contrast agent uptake is present, which could lead to an underestimation of the lesion boundaries [55, 60].
A further advantage of 3D-T1w FSE/TSE sequences over gradient echo sequences is that metal artefacts are significantly reduced, which is valuable when imaging patients with brain tumors who have shunts or surgical clips in place [23, 61]. However, 3D-T1w FSE/TSE sequences are more susceptible to motion compared to gradient echo sequences and may introduce artifacts due to the reliance on high turbo factors [62].
Despite, diagnostic imaging generally has been performed before MRI simulation, the SRT planning MRI may also need to address diagnostic uncertainty, as targets and patient anatomy may undergo important changes and new complications, like intratumoral bleeding, may arise [8, 63]. In some cases, diagnostic uncertainty may even be the leading factor affecting treatment precision. Neuroradiologic and radiologic sequence protocol recommendations addressing individual patient factors are therefore important to incorporate in addition to the following general guidance.
Optimal sequence selection and optimization of sequence protocol parameters—Minimum requirements
  • MRI protocols must be used that include at least one 3D sequence (e.g., 3D-T1w) with a sufficient signal-to-noise ratio (SNR) for target delineation. (Consensus: 100%, abstention: 0%; Minimum requirement: 100%)
  • Standardized MRI protocols must be set up and used for cranial stereotactic treatment planning indications. These standardized protocols must be characterized by a unique and easily understandable study description (e.g., “RT treatment planning—brain metastases”). (Consensus: 100%, abstention: 0%; Minimum requirement: 100%)
  • The main 3D sequence must be isovolumetric with a voxel size of ≤ 1 mm3. (Consensus: 100%, abstention: 0%; Minimum requirement: 100%)
  • Choose the 3D-T1w sequence that provides the best target conspicuity and the most accurate characterization of 3D tumor boundaries. For a substantial fraction of patients, treatment indications and MR scanners, 3D-T1w FSE/TSE sequence protocols are to be preferred over 3D-T1w IR-GE sequence protocols [55, 59, 64]. If multiple 3D sequence protocols for target delineation are acquired, generally, the gross target volume should encompass the extent of the tumor in all 3D sequences. (Consensus: 100%, abstention: 0%; Minimum requirement: 85%, Additional recommendation: 15%)

2.1. Indication-specific considerations for sequence selection

Many treatment indications for cranial SRT benefit from the acquisition of additional 3D sequences for high-resolution target delineation or are not properly visualized on contrast-enhanced 3D-T1w sequences alone. These for example include 3D-True FISP/Dual Excitation (Siemens: 3D-CISS/GE: 3D-FIESTA-C) or 3D-T2w FSE/TSE sequences for cranial nerve targets and vestibular schwannomas [6569]. Recommended sequence parameters and protocols for SRT treatment planning are shown in Tables 1 and 2. As some treatment planning systems still exclusively accept strictly transversal image datasets, reformatting or even acquisition of all sequences in transversal orientation may be an additional requirement for some centers. As described above, individual neuroradiologic and radiologic sequence protocol recommendations should be incorporated and especially complex clinical cases require diagnostic input for optimal target volume definition.
Table 1
MRI Sequence parameters recommended for cranial stereotactic treatment planning at 1.5 T
 
3D-T1w FSE/TSE
3D-T1w IR-GE
3D-T2-FLAIR FSE/TSE
3D-T2w FSE/TSE
3D-True FISP/Dual Excitation
Type
3D
3D
3D
3D
3D
Orientation
Transversal or Sagittal
Transversal or Sagittal
Transversal or Sagittal
Transversal
Transversal
TE
Minimuma
Minimuma
e.g., 374 ms
e.g., 182 ms (heavily T2w)
e.g., 92 ms (moderately T2w)
e.g., 2.45 ms
TR
e.g., 550–750 ms
e.g., 2100–2200 ms
e.g., 7000 ms
e.g., 1200 ms
e.g., 5.47 ms
TI
e.g., 900–1100 ms
e.g., 2050 ms
Acq. matrix
≥ 256 × 256
≥ 256 × 256
≥ 256 × 256
≥ 256 × 256
≥ 256 × 256
Acq. in-plane resolution
≤ 1 × 1 mm
≤ 1 × 1 mm
≤ 1 × 1 mm
≤ 0.7 × 0.7 mm
≤ 0.7 × 0.7 mm
Slice thickness
≤ 1 mm
≤ 1 mm
≤ 1 mm
≤ 0.7 mm
≤ 0.7 mm
Fat saturation
Optional
Post-contrast interval
≥ 5 min
≥ 5 min
GNL Distortion correction
Vendor 3D ± in-houseb
Vendor 3D ± in-houseb
Vendor 3D ± in-houseb
Vendor 3D ± in-houseb
Vendor 3D ± in-houseb
Shimming of B0 inhomogeneities
Patient-specific active shimming
Patient-specific active shimming
Patient-specific active shimming
Patient-specific active shimming
Patient-specific active shimming
Readout bandwidth
≥ 440 Hz
(≥ 660 Hz recommended)
≥ 440 Hz
(≥ 660 Hz recommended)
≥ 440 Hz
(≥ 660 Hz recommended)
≥ 440 Hz
(≥ 660 Hz recommended)
≥ 440 Hz
(≥ 660 Hz recommended)
Interval to treatment
≤ 14 days
(≤ 5 days recommended)
≤ 14 days
(≤ 5 days recommended)
≤ 14 days
(≤ 5 days recommended)
≤ 14 days
(≤ 5 days recommended)
≤ 14 days
(≤ 5 days recommended)
TE echo time, TR repetition time, TI inversion time, Acq. Acquisition, GNL gradient non-linearity, B0 main magnetic field
aUse the minimum possible TE interval, in general the TE interval should be ≤ 10 ms
bInhouse GNL distortion correction refers to obtaining a deformation field via phantom measurements to correct residual distortion after vendor 3D GNL distortion correction
cIf patient-specific active shimming is not available, B0 inhomogeneities during patient image acquisition should be characterized and corrected with advanced methods if relevant (see text)
Table 2
Recommended planning MRI sequence protocols for important stereotactic radiotherapy treatment indications
Brain metastases
Meningioma
Vestibular schwannoma
Pituitary adenoma
Trigeminal neuralgia
AVM
Glomus tumors
2D-T1w pre
(optional: 3D-T1w pre)
3D-T1w pre
2D-T1w pre
(optional: 3D-T1w pre)
2D-T2w FSE/TSE cor
3D-T1w pre
2D-T1w pre
(optional: 3D-T1w GE pre)
3D-T1w pre
Contrast administration
Contrast administration
Contrast administration
2D-T2-FLAIR tra
3D-TOF
3D-TOF
Contrast administration
2D-T2-FLAIR tra
2D-T2-FLAIR tra
2D-T2-FLAIR
3D-T1w pre
Contrast administration
3D-T2w FSE/TSE
3D-T2w FSE/TSE / 3D-True FISP-Dual Excitation
3D-T1w post early
3D-T1w post
3D-T1w post
Contrast administration
3D-T2w FSE/TSE / 3D-True FISP-Dual Excitation
Contrast administration
3D-T1w post
(Optional: 3D-T1w post late)
Subtraction
3D-T1w post − T1w pre
3D-T2w FSE/TSE / 3D-True FISP-Dual Excitation
2D-T1w dynamic cor
3D-T1w post
3D-T1w GE post
Subtraction
3D-T1w post − T1w pre
3D-T1w post
(Optional: Subtraction
3D-T1w GE postpre)
Subtraction
3D-T1w post − T1w pre
Table 3
Test frequencies and intervention thresholds for regular quality assurance
Parameter
Minimum requirement
Additional recommendation
Optional
End-to-end test including MRI simulation
Yearly, after commissioning and after changes to the SRT treatment planning chain in accordance with DIN 6864‑1
Residual gradient non-linearity-related distortions after vendor 3D correction
Characterize using phantom measurements at the time of commissioning, after scanner upgrades, repairs or maintenance, and at least in yearly intervals.
Maximum amount of distortion obtained via phantom measurements for the field-of-view of a typical head MRI scan (sphere of 25 cm diameter) must not exceed 1 mm
In addition to minimum requirements:
Characterization of residual gradient non-linearity related distortions in at least monthly intervals using phantom measurements
In addition to additional recommendations:
Correction of residual distortions ≤ 1 mm based on phantom measurements
Main magnetic field (B0) homogeneity
Characterize after installation (baseline), after scanner upgrades, repairs or maintenance, and at least in yearly intervals, as detailed in sources such as [32] and [33]. If necessary, arrange for repairs to maintain field homogeneity and ensure necessary geometric accuracy
In addition to minimum requirements:
Individually characterize main magnetic field inhomogeneities from system imperfections and susceptibility-related inhomogeneities for patients undergoing MRI simulation by using B0 mapping. Perform individual corrections based on individual B0 mapping or reverse gradient methods
Screening for metallic objects
Daily check the scanner bore for small metallic objects;
Check gradient echo-based localizer images for metal artifacts in every patient
Registration algorithm
Registration error within ≤ 1 mm for registration of phantoms at commissioning
Image quality for flexible coil systems
Monthly
Brain metastases
A high-resolution contrast-enhanced 3D-T1w sequence should be the main sequence for target delineation in brain metastases. Care must be taken to ensure that sufficient time is allowed between contrast administration and sequence acquisition for contrast uptake to occur, enabling all lesions to be clearly visible and the tumor boundaries to be accurately identified (see below). Therefore, a T2-FLAIR sequence should be acquired as a “spacer” between contrast administration and acquisition of the main 3D-T1w sequence. The T2-FLAIR sequence does not add much information on the configuration of the target volume for individual lesions, however it provides complementary diagnostic information, e.g., on the presence of perifocal edema, which indicates an elevated risk for lesion shift [70] and informs the need for corticosteroid dose adaption [43]. A second late contrast-enhanced 3D-T1w sequence can be acquired 15–20 min after the first contrast administration with or without application of a second dose of contrast media to further improve the visibility of metastases and lesion conspicuity [7173]. Prior to contrast administration a T1w sequence should be acquired to discriminate contrast enhancement from other causes for T1w hyperintensity like bleeding or melanin-containing metastases [43].
Meningiomas
Meningiomas are typically strongly contrast-enhancing and are therefore best delineated using 3D-T1w sequences [74]. Subtraction sequences between pre- and postcontrast 3D-T1w sequences can help with tumor delineation near blood vessels and sinuses [75]. Due to their frequently peripheral location at the skull convexity, geometric accuracy is of particular importance in meningiomas.
Vestibular schwannomas
Vestibular schwannomas exhibit intense contrast-enhancement and are well demarcated on 3D-T1w [76]. A 3D-True FISP/Dual Excitation (Siemens: 3D-CISS/GE: 3D-FIESTA-C) or 3D-T2w FSE/TSE sequence provides additional information for target volume definition. In these heavily T2-weighted sequences vestibular schwannomas and accompanying cranial nerves appear hypointense in front of the bright CSF background [77]. 3D-True FISP/Dual Excitation (Siemens: 3D-CISS/GE: 3D-FIESTA-C) and 3D-T2w FSE/TSE sequences enable a particularly high in-plane and slice resolution, which additionally improves the accuracy of target volume delineation for these generally small tumors. Moreover, these sequences are very well suited for delineation of the cochlea, the semi-circular canals and cranial nerve OARs [77].
Pituitary adenomas
Pituitary adenomas typically appear hypointense in relation to residual normal pituitary gland tissue. 3D-T1w sequences alone are frequently not adequate for optimal target volume definition of pituitary adenomas. For accurate delineation of pituitary adenomas, it’s typically beneficial to additionally use 3D-T1w subtraction sequences and dynamic T1w sequences, which offer sequential image datasets at various stages of contrast enhancement [78].
Trigeminal neuralgia
High-resolution strongly T2-weighted 3D-True FISP/Dual Excitation (Siemens: 3D-CISS/GE: 3D-FIESTA-C) or 3D-T2w FSE/TSE should be the main sequences for target volume delineation in trigeminal neuralgia. These sequences depict the hypointense trigeminal nerve fibers against the background of the bright CSF [77]. Due to their high resolution, they are also well suited for target delineation in case of nerval atrophy. 3D-T1w sequences can additionally help with target volume delineation for trigeminal neuralgia and 3D-TOF sequences can depict associated vessels for sparing [79].
Glomus tumors
Glomus tumors are well vascularized benign neoplasms that typically show intense homogeneous contrast-enhancement and are well visualized on 3D-T1w FSE/TSE sequences [80, 81]. The main challenges in MRI for radiotherapy target volume definition is the discrimination from accompanying hyperintense vessels and the occasionally substantial caudal extension through the jugular foramen along the jugular vein [81, 82]. Substantial inferior extension along the carotid sheath requires MR imaging with a sufficiently large craniocaudal field-of-view. As the neck anatomy can be affected by substantial non-rigid tissue deformation with different positioning and 3D-T1w FSE/TSE sequences are susceptible to motion in the neck, imaging in treatment position with mask immobilization can be particularly beneficial in glomus tumors. Distinguishing glomus tumor from surrounding vascular structures can be improved using subtraction sequences [75].
Arteriovenous malformation
Arteriovenous malformations are usually best visualized on 3D-TOF angiography [83]. In addition, 3D-T2w FSE/TSE and contrast-enhanced 3D-T1w are complementary useful for optimal target delineation. With vascular targets, 3D-T1w GE should be used instead of 3D-T1w FSE/TSE sequences because of the improved depiction of vessels in 3D-T1w GE sequences [58]. On 3D-T2w sequences, arteriovenous malformations appear hypointense due to the associated flow void and can thus be discriminated from the surrounding brain and CSF [84].
Indication-specific considerations for sequence selection—Minimum requirements:
MRI protocols must be used that include 3D sequence protocols for all required image contrasts (e.g., 3D-T1w, 3D-T2w, 3D-T2-FLAIR) needed for target and organs-at-risk delineation (Fig. 4). (Consensus: 92%, abstention: 8%; Minimum requirement: 100%).

3. Contrast enhancement

Additional critical parameters that can affect the delineation of lesions in T1-based MR sequences include the dose of gadolinium-based contrast agent (GBCA), as well as the time interval between contrast application and measurement [52, 7173]. In a study by Yuh et al., early imaging (10 min) and late imaging (20 min) following standard dose gadoteridol were compared [71]. After 10 min, 40.6% of metastases < 5 mm were visible, while 75.0% were visible after 20 min. Only after an additional bolus of double-dose gadoteridol, the remaining lesions were evident, demonstrating the importance of increasing the doses of GBCA [71]. Likewise, Baleriaux et al. found an increase in the number of metastases and a better visual conspicuity of the lesions with increasing cumulative doses of gadobenate dimeglumine over several sequential injections [72]. Similarly, Kushnisky et al. found more brain metastases at 15 min after GBCA administration compared to 5 min. They also notably found an increase in metastasis volume after 10 min compared to 5 min, as well as 15 min compared to 10 min postcontrast [73].
Contrast enhancement—Minimum requirements
  • For intraaxial tumors, the time interval between contrast administration and the start of the acquisition of the main T1w sequence should be at least 5 min (see discussion in main text). (Consensus: 100%, abstention: 0%; Minimum requirement: 100%)
Contrast enhancement—Optional
  • Given the improved lesion conspicuity with increased contrast dose, administration of double-dose contrast may be considered in specific circumstances, if the individual benefit of improved tumor delineation for treatment planning clearly outweighs individual GBCA-associated risks. (Consensus: 100%, abstention: 15%; Additional recommendation: 9%, Optional: 91%)
  • An additional delayed T1w sequence protocol may be acquired 15–20 min after the first contrast administration, with or without repeated contrast administration to improve target conspicuity and depiction of target boundaries. (Consensus: 92%, abstention: 8%; Optional: 100%)

4. Time interval between MRI simulation and treatment delivery

The interval between MRI and treatment delivery is one of the most crucial parameters for treatment precision ([7, 8]; Fig. 5). The importance of simulation MRI datasets being up-to-date is especially high for rapidly growing tumors such as brain metastases [8, 63] that are also often associated with fluctuating amounts of surrounding edema [85, 86].
This was confirmed by Seymour et al., who found a large detriment in local control following radiosurgery in brain metastases, if the interval between MRI and radiosurgery was greater than or equal to 14 days (local control 56% vs. 95% at 6 months post-SRS) [7]. Subsequently, Salkeld et al. even reported relevant differences with imaging intervals ≤ 7 days before SRS. In this study, 41% of patients with an interval ≤ 7 days required a change in radiooncologic management, whereas even 78% of patients required changes if the delay exceeded 7 days. Replanning was performed most frequently due to an increase in the tumor or resection cavity volume [8, 87]. In a retrospective review of 101 patients and 531 brain metastases, Kutuk et al. confirmed that changes in tumor size and spatial position occurred as a function of time, with the risk of requiring a margin beyond 1 mm increasing every day by 5% [88]. Therefore, the interval between imaging and treatment delivery should be as short as possible.
Repeated MR imaging during fractionated stereotactic radiotherapy
During fractionated stereotactic radiation therapy, brain metastases and malignant primary brain tumors can undergo profound changes due to transient swelling, changes in perifocal edema, and incipient treatment effect. An analysis by Hessen et al. evaluated the significance of repeated MRI scans during fractionated SRT for 18 brain metastases and 20 resection cavities. In cases with in situ brain metastases, there was a reduction in PTV coverage of up to 34.8%, while postoperative cases were less affected (up to −4.5% in PTV coverage) [89]. Importantly, as only 3–5 fractions were employed in the study by Hessen et al., even more pronounced changes would be expected with more prolonged fractionation schemes. Uto et al. investigated interfractional target changes in 27 brain metastases (23 brain metastases) during 13-fraction FSRT based on a mid-treatment MRI acquired after a median of 6 days. Compared to the baseline MRI, the mid-treatment GTV had increased by more than 20% in 5 lesions and decreased by more than 20% in another 5 tumors. Interestingly, in 15 out of 27 brain metastases the initial PTV did not encompass the entire mid-treatment GTV [90]. More recently, similar findings were reported in a series of 114 brain metastases (66 patients) treated with 10 to 20 fractions of gamma knife FSRT. After a median of 7 days between the initial simulation MRI and the interfractional MR imaging, interfractional changes between −48 to 72% in tumor volume were observed and 29% of lesions showed significant volume changes (defined as ≥ +20% or ≤ −20% change in volume). Beyond volumetric changes, 85% of treatment plans needed to be modified, because of the information provided by the mid-treatment MRI [91].
Time interval between MRI simulation and treatment delivery—Minimum requirements
  • The time interval between the MRI simulation and the administration of treatment must not be larger than 14 days. (Consensus: 100%, abstention: 8%; Minimum requirement: 100%)
Time interval between MRI simulation and treatment delivery—Additional recommendations
  • In brain metastases and primary brain tumors CNS-WHO-grade 2–4, the time interval between MR simulation and treatment delivery should not be greater than 5 days. (Consensus: 100%, abstention: 8%; Minimum requirement: 17%, additional recommendation: 83%)
Time interval between MRI simulation and treatment delivery—Optional
  • Consider, performing an additional simulation MRI for adaption of target volumes every 5 fractions in fractionated stereotactic radiotherapy (≤ 12 fractions). (Consensus: 100%, abstention: 0%, Optional: 100%)

6. Image registration and imaging in SRT position (Fig. 6)

The simulation MRI for SRT is most frequently acquired in a diagnostic head coil and subsequently rigidly coregistered to the planning CT obtained in the treatment position with mask immobilization. Therefore, it is an essential prerequisite that the registration of the simulation MRI to the geometry of the planning CT in treatment position is as accurate as possible. In general, rigid registration that involves translations and rotations only is the optimal type of registration for cranial SRT. In case of multiple target volumes, multiple individual registrations in a reduced volume of interest covering each lesion separately should be considered [50]. The software used for image registration must be properly commissioned and validated followed by a quality assurance program for image registration e.g., according to AAPM Task Group report 132 [50]. This involves quantitative validation of registration accuracy with phantoms [50]. For cranial SRT, the employed rigid registration algorithm should have a registration error within 1 mm in registration phantoms [50, 92, 93]. Moreover, in daily clinical practice every MRI-CT registration for treatment planning must be verified qualitatively by a clinical expert with board certification. Software manufacturers developing treatment planning systems for cranial stereotactic radiotherapy are encouraged to offer a quantitative assessment of the registration method’s accuracy using an independent approach from the registration algorithm (e.g., employing an automatic landmark placement technique when the registration is carried out by optimizing the normalized mutual information). For the brain, in individual publications the accuracy of normalized-mutual-information based MRI-CT coregistration has been reported to be ≤ 0.5 mm in plane and ≤ 1 mm along the Z‑axis for CT slices with a thickness between 2 and 3 mm and a 2 mm 3D-T1w IR-GE sequence. However, the registration uncertainty increased by a factor of 2–3 when a 2D-SE sequence with a slice thickness of 5 mm was registered to a 5-mm planning CT [92]. More generally, registration accuracies of ~ 2 mm are reported [94, 95]. This finding highlights the significance of using high-resolution pairs of MRI and CT datasets to ensure adequate registration accuracy. MRI datasets used for registration should be 3D sequences, with equidistant slice planes and without gaps. MRI datasets must also be free from distortions, motion, folding, and metal artifacts and errors in planning CT datasets have to be minimized for optimal registration accuracy [50]. Due to the long scanning times with MRI, it is also important to exclude motion between individual sequences that are not addressed by registration. Many treatment planning systems automatically propagate the transformation from the registration of one MRI sequence to all other sequences in the study, which may lead to poor registration accuracy in case of intermediate motion. It is therefore important to individually verify the registration accuracy of every MRI series used for treatment planning.
Acquiring planning MRI datasets in the treatment position may improve the accuracy of rigid registration, as planning MRI and CT images are acquired in a similar anatomical configuration [30, 93, 96]. Multiple groups have reported imaging setups to acquire simulation MRIs in the treatment position with mask immobilization [24, 96, 97]. Flexible coil setups have been developed most frequently for this purpose, as most stereotactic mask systems will not fit into routine radiological head coils [30, 9799]. Masitho et al. evaluated the registration accuracy of three treatment planning systems for coregistering planning MRIs acquired in a diagnostic head coil and in radiotherapy treatment position with mask immobilization to a common planning CT. Generally, the software used for registration had the most significant impact on the accuracy of the registration. However, for less optimal registration methods, registration accuracy was significantly improved if simulation MRIs had been performed in the treatment position [93]. For targets near the foramen magnum, performing the planning MRI in treatment position could be more beneficial due to slight deformation occurring with different extension of the occipito-atlanto-axial joint complex [61]. In addition to slightly improving the accuracy of the registration, imaging in the treatment position with mask immobilization significantly decreases motion artifacts, which may help with the precise definition of the target volume [97, 99]. However, usually some decrease in image quality is expected from imaging with a flexible coil setup in treatment position as compared to using dedicated diagnostic head coils, which may need to be compensated with a longer measurement time [99]. Since flexible coils are more prone to mechanical damage, monthly quality assurance is recommended as proposed in the AAPM TG 284 report [10]. The procedure can be combined with validating geometric accuracy, gain, and spatial resolution which are part of quality assurance recommendations of the ACR [32].
Acquiring planning MRI studies in the treatment position with mask immobilization becomes a necessity when using an MR-only workflow [100]. Synthetic CT, involving the computation of synthetic CT images from one or multiple MR sequences, eliminates the requirement for an additional planning CT, thereby theoretically removing any registration uncertainties [101]. Current methods provide reasonable dosimetric results in most standard situations and dosimetric differences < 1% have been reported for deep learning methods recently [100, 102]. It has also been reported that sCTs provide adequate means for positioning the patient at the linear accelerator, i.e., as reference for registering the CBCT [103, 104] as well as 2D/2D based positioning systems [93]. However, when synthetic CT is employed in an MR-only workflow, care must be taken that the treatment position is exactly reproduced on the MR scanner as the synthetic CT will serve both as a foundation for dose calculation as well as a reference for image guidance at the radiotherapy treatment unit [100]. It is equally important to ensure that no changes in patient position occur between acquisition of the dedicated synthetic CT sequences and the MR sequences used for target delineation that are not addressed by registration. It is therefore recommended that prior to clinical implementation of an MR-only workflow, experience with MR simulation has been established, with optimized sequence selection and protocol parameters as well as by optimal quality assurance using the advanced recommendations as described above.
Image registration and imaging in SRT position—Minimum requirements
  • When registering simulation MRI datasets to a planning CT, both the planning CT and the MRI dataset, i.e., both registration pairs, must have a slice thickness of ≤ 1 mm. (Consensus: 100%, abstention: 0%; Minimum requirement: 92%, additional recommendation: 8%)
  • A proper registration algorithm, commissioned, and validated for stereotactic radiotherapy/radiosurgery followed by expert correction and verification must be used. (Consensus: 100%, abstention: 0%; Minimum requirement: 100%)
  • With quantitative validation at commissioning, the registration algorithm must achieve a registration error within ≤ 1 mm for registration of phantoms. (Consensus: 100%, abstention: 0%; Minimum requirement: 100%)
  • Artifacts that could impair registration must be minimized: Set a sufficient phase oversampling factor to avoid folding artifacts, minimize distortions in MRI, minimize metal artifacts in MRI (e.g., by using 3D-T1w FSE/TSE instead of 3D-T1w GE sequence protocols) and minimize artifacts in planning CTs (e.g., minimize metal artifacts from dental fillings). (Consensus: 100%, abstention: 0%, Minimum requirement: 92%, additional recommendation: 8%)
  • Motion artifacts inside the diagnostic head coil must be minimized by proper use of cushions and foam elements. (Consensus: 100%, abstention: 0%; Minimum requirement: 92%, additional recommendation: 8%)
  • In daily clinical practice, registration quality for each treatment planning registration must be verified qualitatively by a board-certified physician or medical physics expert with experience in cranial stereotactic radiotherapy. This verification should be performed using an overlay method (e.g., alpha blending with or without varying color schemes for both datasets, a checkerboard pattern, and/or “spy glass” tools), with dynamic assessment of registration accuracy (e.g., by blending between datasets or by moving the checkerboard pattern and “spy glass” tool). Site-specific recommendations can be found in [21], for example. (Consensus: 100%, abstention: 0%; Minimum requirement: 100%)
  • When using multiple sequences for treatment planning, registration accuracy must be individually verified for each sequence used for treatment planning, because of the risk of motion between sequences. (Consensus: 100%, abstention: 0%; Minimum requirement: 100%)
Image registration and imaging in SRT position—Optional
  • Consider, acquiring simulation MRIs in the treatment position with mask immobilization to improve registration accuracy and to reduce/eliminate motion artifacts. Sufficient image quality must be ensured when acquiring simulation MRIs in the treatment position with a flexible coil setup. (Consensus: 85%, abstention: 0%; Additional recommendation: 9%, Optional: 91%)
  • Especially consider acquisition in the treatment position with mask immobilization if nonrigid tissue deformations are expected between the treatment position and the diagnostic imaging position (e.g., for targets near the foramen magnum and if rigidity of the skull is significantly impaired after surgery). (Consensus: 92%, abstention: 0%; Additional recommendation: 8%, Optional: 92%)
  • Consider, performing monthly checks on image quality for flexible coil systems employed for stereotactic radiotherapy simulation [10]. (Consensus: 85%, abstention: 0%; Additional recommendation: 9%, optional: 91%)
  • Use of synthetic CT and an MR-only workflow may be considered to exclude MRI-CT registration uncertainties. If an MR-only workflow is used, it must be ensured that synthetic CT datasets can be used for both treatment planning and image guidance. In addition, motion between MRI sequences used for synthetic CT calculation and sequences used for target delineation must be excluded or addressed by registration. (Consensus: 92%, abstention: 0%; Minimum requirement: 8%, additional recommendation: 8%, optional: 83%)

Conclusion

Accurate MRI simulation is a critical basis for precise treatment in cranial stereotactic radiotherapy and an integral part of stereotactic radiotherapy treatment planning. This report recommends measures and procedures to optimize MRI sequence protocols, to verify spatial precision and to optimize the clinical workflow to secure the fidelity and spatial precision of MR images in SRT. Broadly implementable minimum requirements provide major improvements in accuracy, while additional recommendations and options allow advanced centers to further optimize MR-based treatment planning. This guideline and especially the topics of repeated imaging during fractionated treatment, as well as the role of MR-only treatment planning, will require an update when more evidence becomes available.

Acknowledgements

We sincerely thank Prof. Dr. Daniela Thorwarth, Prof. Dr. Matthias Guckenberger, Dipl.-Phys. Eric Beckers, Prof. Dr. Thomas Brunner, Prof. Dr. Lothar R. Schad, Prof. Dr. Marc Schlamann and Prof. Dr. Mirko Pham for their thoughtful comments and help during the review of the initial manuscript draft.ss

Conflict of interest

F. Putz, C. Bert and R. Fietkau report grants and personal fees from Siemens Healthineers, outside the submitted work. M. Bock, D. Schmitt, O. Blanck, M.I. Ruge, E. Hattingen, C.P. Karger, J. Grigo, M.A. Schmidt, T. Bäuerle and A. Wittig declare that they have no competing interests.
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article’s Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article’s Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://​creativecommons.​org/​licenses/​by/​4.​0/​.

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Unsere Produktempfehlungen

e.Med Interdisziplinär

Kombi-Abonnement

Für Ihren Erfolg in Klinik und Praxis - Die beste Hilfe in Ihrem Arbeitsalltag

Mit e.Med Interdisziplinär erhalten Sie Zugang zu allen CME-Fortbildungen und Fachzeitschriften auf SpringerMedizin.de.

e.Med Radiologie

Kombi-Abonnement

Mit e.Med Radiologie erhalten Sie Zugang zu CME-Fortbildungen des Fachgebietes Radiologie, den Premium-Inhalten der radiologischen Fachzeitschriften, inklusive einer gedruckten Radiologie-Zeitschrift Ihrer Wahl.

Strahlentherapie und Onkologie

Print-Titel

•Übersichten, Originalien, Kasuistiken

•Kommentierte Literatur aus der Radioonkologie, Strahlenbiologie und -physik

Literatur
1.
Zurück zum Zitat Potter R et al (1992) Sagittal and coronal planes from MRI for treatment planning in tumors of brain, head and neck: MRI assisted simulation. Radiother Oncol 23(2):127–130PubMedCrossRef Potter R et al (1992) Sagittal and coronal planes from MRI for treatment planning in tumors of brain, head and neck: MRI assisted simulation. Radiother Oncol 23(2):127–130PubMedCrossRef
2.
Zurück zum Zitat Guckenberger M et al (2020) Definition and quality requirements for stereotactic radiotherapy: consensus statement from the DEGRO/DGMP Working Group Stereotactic Radiotherapy and Radiosurgery. Strahlenther Onkol 196(5):417–420PubMedPubMedCentralCrossRef Guckenberger M et al (2020) Definition and quality requirements for stereotactic radiotherapy: consensus statement from the DEGRO/DGMP Working Group Stereotactic Radiotherapy and Radiosurgery. Strahlenther Onkol 196(5):417–420PubMedPubMedCentralCrossRef
3.
Zurück zum Zitat Schmitt, D., et al., Technological quality requirements for stereotactic radiotherapy : Expert review group consensus from the DGMP Working Group for Physics and Technology in Stereotactic Radiotherapy. Strahlenther Onkol, 2020. 196(5): p. 421–443. Schmitt, D., et al., Technological quality requirements for stereotactic radiotherapy : Expert review group consensus from the DGMP Working Group for Physics and Technology in Stereotactic Radiotherapy. Strahlenther Onkol, 2020. 196(5): p. 421–443.
4.
Zurück zum Zitat Kirkpatrick JP et al (2015) Defining the optimal planning target volume in image-guided stereotactic radiosurgery of brain metastases: results of a randomized trial. Int J Radiat Oncol Biol Phys 91(1):100–108PubMedCrossRef Kirkpatrick JP et al (2015) Defining the optimal planning target volume in image-guided stereotactic radiosurgery of brain metastases: results of a randomized trial. Int J Radiat Oncol Biol Phys 91(1):100–108PubMedCrossRef
5.
Zurück zum Zitat Lo A et al (2018) Population-Based Study of Stereotactic Radiosurgery or Fractionated Stereotactic Radiation Therapy for Vestibular Schwannoma: Long-Term Outcomes and Toxicities. Int J Radiat Oncol Biol Phys 100(2):443–451PubMedCrossRef Lo A et al (2018) Population-Based Study of Stereotactic Radiosurgery or Fractionated Stereotactic Radiation Therapy for Vestibular Schwannoma: Long-Term Outcomes and Toxicities. Int J Radiat Oncol Biol Phys 100(2):443–451PubMedCrossRef
6.
Zurück zum Zitat Sheehan J et al (2022) An international multicenter matched cohort analysis of incidental meningioma progression during active surveillance or after stereotactic radiosurgery: the IMPASSE study. Neuro Oncol 24(1):116–124PubMedCrossRef Sheehan J et al (2022) An international multicenter matched cohort analysis of incidental meningioma progression during active surveillance or after stereotactic radiosurgery: the IMPASSE study. Neuro Oncol 24(1):116–124PubMedCrossRef
7.
Zurück zum Zitat Seymour ZA et al (2015) Interval From Imaging to Treatment Delivery in the Radiation Surgery Age: How Long Is Too Long? Int J Radiat Oncol Biol Phys 93(1):126–132PubMedCrossRef Seymour ZA et al (2015) Interval From Imaging to Treatment Delivery in the Radiation Surgery Age: How Long Is Too Long? Int J Radiat Oncol Biol Phys 93(1):126–132PubMedCrossRef
8.
Zurück zum Zitat Salkeld AL et al (2018) Changes in Brain Metastasis During Radiosurgical Planning. Int J Radiat Oncol Biol Phys 102(4):727–733PubMedCrossRef Salkeld AL et al (2018) Changes in Brain Metastasis During Radiosurgical Planning. Int J Radiat Oncol Biol Phys 102(4):727–733PubMedCrossRef
9.
Zurück zum Zitat Seibert TM et al (2016) Distortion inherent to magnetic resonance imaging can lead to geometric miss in radiosurgery planning. Pract Radiat Oncol 6(6):e319–e328PubMedPubMedCentralCrossRef Seibert TM et al (2016) Distortion inherent to magnetic resonance imaging can lead to geometric miss in radiosurgery planning. Pract Radiat Oncol 6(6):e319–e328PubMedPubMedCentralCrossRef
10.
Zurück zum Zitat Glide-Hurst CK et al (2021) Task group 284 report: magnetic resonance imaging simulation in radiotherapy: considerations for clinical implementation, optimization, and quality assurance. Med Phys 48(7):e636–e670PubMedCrossRef Glide-Hurst CK et al (2021) Task group 284 report: magnetic resonance imaging simulation in radiotherapy: considerations for clinical implementation, optimization, and quality assurance. Med Phys 48(7):e636–e670PubMedCrossRef
11.
Zurück zum Zitat Tanadini-Lang S et al (2023) An ESTRO-ACROP guideline on quality assurance and medical physics commissioning of online MRI guided radiotherapy systems based on a consensus expert opinion. Radiother Oncol 181:109504PubMedCrossRef Tanadini-Lang S et al (2023) An ESTRO-ACROP guideline on quality assurance and medical physics commissioning of online MRI guided radiotherapy systems based on a consensus expert opinion. Radiother Oncol 181:109504PubMedCrossRef
12.
Zurück zum Zitat Baldwin LN et al (2007) Characterization, prediction, and correction of geometric distortion in 3 T MR images. Med Phys 34(2):388–399PubMedCrossRef Baldwin LN et al (2007) Characterization, prediction, and correction of geometric distortion in 3 T MR images. Med Phys 34(2):388–399PubMedCrossRef
13.
Zurück zum Zitat Sumanaweera TS et al (1994) Characterization of spatial distortion in magnetic resonance imaging and its implications for stereotactic surgery. Neurosurgery 35(4):696–703 (discussion 703–4)PubMedCrossRef Sumanaweera TS et al (1994) Characterization of spatial distortion in magnetic resonance imaging and its implications for stereotactic surgery. Neurosurgery 35(4):696–703 (discussion 703–4)PubMedCrossRef
14.
Zurück zum Zitat Khoo VS et al (1997) Magnetic resonance imaging (MRI): considerations and applications in radiotherapy treatment planning. Radiother Oncol 42(1):1–15PubMedCrossRef Khoo VS et al (1997) Magnetic resonance imaging (MRI): considerations and applications in radiotherapy treatment planning. Radiother Oncol 42(1):1–15PubMedCrossRef
15.
Zurück zum Zitat Baldwin LN, Wachowicz K, Fallone BG (2009) A two-step scheme for distortion rectification of magnetic resonance images. Med Phys 36(9):3917–3926PubMedCrossRef Baldwin LN, Wachowicz K, Fallone BG (2009) A two-step scheme for distortion rectification of magnetic resonance images. Med Phys 36(9):3917–3926PubMedCrossRef
16.
Zurück zum Zitat Putz, F., et al., Magnetic resonance imaging for brain stereotactic radiotherapy : A review of requirements and pitfalls. Strahlenther Onkol, 2020. 196(5): p. 444–456. Putz, F., et al., Magnetic resonance imaging for brain stereotactic radiotherapy : A review of requirements and pitfalls. Strahlenther Onkol, 2020. 196(5): p. 444–456.
17.
Zurück zum Zitat Wang H, Balter J, Cao Y (2013) Patient-induced susceptibility effect on geometric distortion of clinical brain MRI for radiation treatment planning on a 3T scanner. Phys Med Biol 58(3):465–477PubMedCrossRef Wang H, Balter J, Cao Y (2013) Patient-induced susceptibility effect on geometric distortion of clinical brain MRI for radiation treatment planning on a 3T scanner. Phys Med Biol 58(3):465–477PubMedCrossRef
18.
Zurück zum Zitat Jovicich J et al (2006) Reliability in multi-site structural MRI studies: effects of gradient non-linearity correction on phantom and human data. Neuroimage 30(2):436–443PubMedCrossRef Jovicich J et al (2006) Reliability in multi-site structural MRI studies: effects of gradient non-linearity correction on phantom and human data. Neuroimage 30(2):436–443PubMedCrossRef
19.
Zurück zum Zitat Pappas EP et al (2017) MRI-Related Geometric Distortions in Stereotactic Radiotherapy Treatment Planning: Evaluation and Dosimetric Impact. Technol Cancer Res Treat 16(6):1120–1129PubMedPubMedCentralCrossRef Pappas EP et al (2017) MRI-Related Geometric Distortions in Stereotactic Radiotherapy Treatment Planning: Evaluation and Dosimetric Impact. Technol Cancer Res Treat 16(6):1120–1129PubMedPubMedCentralCrossRef
20.
Zurück zum Zitat Kondziolka D et al (1992) A comparison between magnetic resonance imaging and computed tomography for stereotactic coordinate determination. Neurosurgery 30(3):402–406 (discussion 406–7)PubMedCrossRef Kondziolka D et al (1992) A comparison between magnetic resonance imaging and computed tomography for stereotactic coordinate determination. Neurosurgery 30(3):402–406 (discussion 406–7)PubMedCrossRef
21.
Zurück zum Zitat Stanescu T et al (2008) A study on the magnetic resonance imaging (MRI)-based radiation treatment planning of intracranial lesions. Phys Med Biol 53(13):3579–3593PubMedCrossRef Stanescu T et al (2008) A study on the magnetic resonance imaging (MRI)-based radiation treatment planning of intracranial lesions. Phys Med Biol 53(13):3579–3593PubMedCrossRef
22.
Zurück zum Zitat Mengling, V., et al., Evaluation of the influence of susceptibility-induced magnetic field distortions on the precision of contouring intracranial organs at risk for stereotactic radiosurgery. Phys Imaging Radiat Oncol, 2020. 15: p. 91–97. Mengling, V., et al., Evaluation of the influence of susceptibility-induced magnetic field distortions on the precision of contouring intracranial organs at risk for stereotactic radiosurgery. Phys Imaging Radiat Oncol, 2020. 15: p. 91–97.
23.
Zurück zum Zitat Fransson A, Andreo P, Potter R (2001) Aspects of MR image distortions in radiotherapy treatment planning. Strahlenther Onkol 177(2):59–73PubMedCrossRef Fransson A, Andreo P, Potter R (2001) Aspects of MR image distortions in radiotherapy treatment planning. Strahlenther Onkol 177(2):59–73PubMedCrossRef
24.
Zurück zum Zitat Paulson ES et al (2015) Comprehensive MRI simulation methodology using a dedicated MRI scanner in radiation oncology for external beam radiation treatment planning. Med Phys 42(1):28–39PubMedCrossRef Paulson ES et al (2015) Comprehensive MRI simulation methodology using a dedicated MRI scanner in radiation oncology for external beam radiation treatment planning. Med Phys 42(1):28–39PubMedCrossRef
25.
Zurück zum Zitat Reinsberg SA et al (2005) A complete distortion correction for MR images: II. Rectification of static-field inhomogeneities by similarity-based profile mapping. Phys Med Biol 50(11):2651–2661PubMedCrossRef Reinsberg SA et al (2005) A complete distortion correction for MR images: II. Rectification of static-field inhomogeneities by similarity-based profile mapping. Phys Med Biol 50(11):2651–2661PubMedCrossRef
26.
Zurück zum Zitat Doran SJ et al (2005) A complete distortion correction for MR images: I. Gradient warp correction. Phys Med Biol 50(7):1343–1361PubMedCrossRef Doran SJ et al (2005) A complete distortion correction for MR images: I. Gradient warp correction. Phys Med Biol 50(7):1343–1361PubMedCrossRef
27.
Zurück zum Zitat Karger CP et al (2006) Accuracy of device-specific 2D and 3D image distortion correction algorithms for magnetic resonance imaging of the head provided by a manufacturer. Phys Med Biol 51(12):N253–61PubMedCrossRef Karger CP et al (2006) Accuracy of device-specific 2D and 3D image distortion correction algorithms for magnetic resonance imaging of the head provided by a manufacturer. Phys Med Biol 51(12):N253–61PubMedCrossRef
28.
Zurück zum Zitat Torfeh T et al (2016) Characterization of 3D geometric distortion of magnetic resonance imaging scanners commissioned for radiation therapy planning. Magn Reson Imaging 34(5):645–653PubMedCrossRef Torfeh T et al (2016) Characterization of 3D geometric distortion of magnetic resonance imaging scanners commissioned for radiation therapy planning. Magn Reson Imaging 34(5):645–653PubMedCrossRef
29.
Zurück zum Zitat Janke A et al (2004) Use of spherical harmonic deconvolution methods to compensate for nonlinear gradient effects on MRI images. Magn Reson Med 52(1):115–122PubMedCrossRef Janke A et al (2004) Use of spherical harmonic deconvolution methods to compensate for nonlinear gradient effects on MRI images. Magn Reson Med 52(1):115–122PubMedCrossRef
30.
Zurück zum Zitat Paulson ES et al (2016) Consensus opinion on MRI simulation for external beam radiation treatment planning. Radiother Oncol 121(2):187–192PubMedCrossRef Paulson ES et al (2016) Consensus opinion on MRI simulation for external beam radiation treatment planning. Radiother Oncol 121(2):187–192PubMedCrossRef
31.
Zurück zum Zitat Matheoud R et al (2023) EFOMP’s protocol quality controls in PET/CT and PET/MR. Phys Med 105:102506PubMedCrossRef Matheoud R et al (2023) EFOMP’s protocol quality controls in PET/CT and PET/MR. Phys Med 105:102506PubMedCrossRef
32.
Zurück zum Zitat Price R et al (2015) MRI Quality Control. Manual, vol 2015. American College of Radiology Price R et al (2015) MRI Quality Control. Manual, vol 2015. American College of Radiology
33.
Zurück zum Zitat Wang D, Doddrell DM, Cowin G (2004) A novel phantom and method for comprehensive 3‑dimensional measurement and correction of geometric distortion in magnetic resonance imaging. Magn Reson Imaging 22(4):529–542PubMedCrossRef Wang D, Doddrell DM, Cowin G (2004) A novel phantom and method for comprehensive 3‑dimensional measurement and correction of geometric distortion in magnetic resonance imaging. Magn Reson Imaging 22(4):529–542PubMedCrossRef
34.
Zurück zum Zitat Schad, L., et al., Correction of spatial distortion in MR imaging: a prerequisite for accurate stereotaxy. J Comput Assist Tomogr, 1987. 11(3): p. 499–505. Schad, L., et al., Correction of spatial distortion in MR imaging: a prerequisite for accurate stereotaxy. J Comput Assist Tomogr, 1987. 11(3): p. 499–505.
35.
Zurück zum Zitat Stanescu T, Wachowicz K, Jaffray DA (2012) Characterization of tissue magnetic susceptibility-induced distortions for MRIgRT. Med Phys 39(12):7185–7193PubMedCrossRef Stanescu T, Wachowicz K, Jaffray DA (2012) Characterization of tissue magnetic susceptibility-induced distortions for MRIgRT. Med Phys 39(12):7185–7193PubMedCrossRef
36.
Zurück zum Zitat Bednarz G et al (1999) Evaluation of the spatial accuracy of magnetic resonance imaging-based stereotactic target localization for gamma knife radiosurgery of functional disorders. Neurosurgery 45(5):1156–1161 (discussion 1161–3)PubMedCrossRef Bednarz G et al (1999) Evaluation of the spatial accuracy of magnetic resonance imaging-based stereotactic target localization for gamma knife radiosurgery of functional disorders. Neurosurgery 45(5):1156–1161 (discussion 1161–3)PubMedCrossRef
37.
Zurück zum Zitat Schmidt MA et al (2017) Stereotactic radiosurgery planning of vestibular schwannomas: Is MRI at 3 Tesla geometrically accurate? Med Phys 44(2):375–381PubMedCrossRef Schmidt MA et al (2017) Stereotactic radiosurgery planning of vestibular schwannomas: Is MRI at 3 Tesla geometrically accurate? Med Phys 44(2):375–381PubMedCrossRef
38.
Zurück zum Zitat Walker A et al (2014) MRI distortion: considerations for MRI based radiotherapy treatment planning. Australas Phys Eng Sci Med 37(1):103–113PubMedCrossRef Walker A et al (2014) MRI distortion: considerations for MRI based radiotherapy treatment planning. Australas Phys Eng Sci Med 37(1):103–113PubMedCrossRef
39.
Zurück zum Zitat Jezzard P, Balaban RS (1995) Correction for geometric distortion in echo planar images from B0 field variations. Magn Reson Med 34(1):65–73PubMedCrossRef Jezzard P, Balaban RS (1995) Correction for geometric distortion in echo planar images from B0 field variations. Magn Reson Med 34(1):65–73PubMedCrossRef
40.
Zurück zum Zitat Shariff M et al (2022) End-to-end testing for stereotactic radiotherapy including the development of a Multi-Modality phantom. Z Med Phys Shariff M et al (2022) End-to-end testing for stereotactic radiotherapy including the development of a Multi-Modality phantom. Z Med Phys
41.
Zurück zum Zitat Smith AS et al (1990) Intracranial chemical-shift artifacts on MR images of the brain: observations and relation to sampling bandwidth. AJR Am J Roentgenol 154(6):1275–1283PubMedCrossRef Smith AS et al (1990) Intracranial chemical-shift artifacts on MR images of the brain: observations and relation to sampling bandwidth. AJR Am J Roentgenol 154(6):1275–1283PubMedCrossRef
42.
Zurück zum Zitat Soher, B.J., B.M. Dale, and E.M. Merkle, A review of MR physics: 3T versus 1.5T. Magn Reson Imaging Clin N Am, 2007. 15(3): p. 277–90, v. Soher, B.J., B.M. Dale, and E.M. Merkle, A review of MR physics: 3T versus 1.5T. Magn Reson Imaging Clin N Am, 2007. 15(3): p. 277–90, v.
43.
Zurück zum Zitat Kaufmann TJ et al (2020) Consensus recommendations for a standardized brain tumor imaging protocol for clinical trials in brain metastases. Neuro Oncol 22(6):757–772PubMedPubMedCentralCrossRef Kaufmann TJ et al (2020) Consensus recommendations for a standardized brain tumor imaging protocol for clinical trials in brain metastases. Neuro Oncol 22(6):757–772PubMedPubMedCentralCrossRef
44.
Zurück zum Zitat Zhang B et al (2010) Development of a geometrically accurate imaging protocol at 3 Tesla MRI for stereotactic radiosurgery treatment planning. Phys Med Biol 55(22):6601–6615PubMedCrossRef Zhang B et al (2010) Development of a geometrically accurate imaging protocol at 3 Tesla MRI for stereotactic radiosurgery treatment planning. Phys Med Biol 55(22):6601–6615PubMedCrossRef
45.
Zurück zum Zitat Wen N et al (2018) Evaluation of a magnetic resonance guided linear accelerator for stereotactic radiosurgery treatment. Radiother Oncol 127(3):460–466PubMedPubMedCentralCrossRef Wen N et al (2018) Evaluation of a magnetic resonance guided linear accelerator for stereotactic radiosurgery treatment. Radiother Oncol 127(3):460–466PubMedPubMedCentralCrossRef
46.
Zurück zum Zitat Slagowski JM et al (2020) Dosimetric feasibility of brain stereotactic radiosurgery with a 0.35 T MRI-guided linac and comparison vs a C-arm-mounted linac. Med Phys 47(11):5455–5466PubMedCrossRef Slagowski JM et al (2020) Dosimetric feasibility of brain stereotactic radiosurgery with a 0.35 T MRI-guided linac and comparison vs a C-arm-mounted linac. Med Phys 47(11):5455–5466PubMedCrossRef
47.
Zurück zum Zitat Retif P et al (2022) Evaluation of the ability of the Brainlab Elements Cranial Distortion Correction algorithm to correct clinically relevant MRI distortions for cranial SRT. Strahlenther Onkol 198(10):907–918PubMedCrossRef Retif P et al (2022) Evaluation of the ability of the Brainlab Elements Cranial Distortion Correction algorithm to correct clinically relevant MRI distortions for cranial SRT. Strahlenther Onkol 198(10):907–918PubMedCrossRef
48.
Zurück zum Zitat Calvo-Ortega JF et al (2019) Evaluation of a novel software application for magnetic resonance distortion correction in cranial stereotactic radiosurgery. Med Dosim 44(2):136–143PubMedCrossRef Calvo-Ortega JF et al (2019) Evaluation of a novel software application for magnetic resonance distortion correction in cranial stereotactic radiosurgery. Med Dosim 44(2):136–143PubMedCrossRef
49.
Zurück zum Zitat Shi L et al (2021) Benchmarking of Deformable Image Registration for Multiple Anatomic Sites Using Digital Data Sets With Ground-Truth Deformation Vector Fields. Pract Radiat Oncol 11(5):404–414PubMedCrossRef Shi L et al (2021) Benchmarking of Deformable Image Registration for Multiple Anatomic Sites Using Digital Data Sets With Ground-Truth Deformation Vector Fields. Pract Radiat Oncol 11(5):404–414PubMedCrossRef
50.
Zurück zum Zitat Brock KK et al (2017) Use of image registration and fusion algorithms and techniques in radiotherapy: Report of the AAPM Radiation Therapy Committee Task Group No. 132. Med Phys 44(7):e43–e76PubMedCrossRef Brock KK et al (2017) Use of image registration and fusion algorithms and techniques in radiotherapy: Report of the AAPM Radiation Therapy Committee Task Group No. 132. Med Phys 44(7):e43–e76PubMedCrossRef
51.
Zurück zum Zitat Zhong H, Kim J, Chetty IJ (2010) Analysis of deformable image registration accuracy using computational modeling. Med Phys 37(3):970–979PubMedPubMedCentralCrossRef Zhong H, Kim J, Chetty IJ (2010) Analysis of deformable image registration accuracy using computational modeling. Med Phys 37(3):970–979PubMedPubMedCentralCrossRef
52.
Zurück zum Zitat Anzalone, N., et al., Optimizing contrast-enhanced magnetic resonance imaging characterization of brain metastases: relevance to stereotactic radiosurgery. Neurosurgery, 2013. 72(5): p. 691–701. Anzalone, N., et al., Optimizing contrast-enhanced magnetic resonance imaging characterization of brain metastases: relevance to stereotactic radiosurgery. Neurosurgery, 2013. 72(5): p. 691–701.
53.
Zurück zum Zitat Snell, J.W., et al., Assessment of imaging studies used with radiosurgery: a volumetric algorithm and an estimation of its error. Technical note. J Neurosurg, 2006. 104(1): p. 157–62. Snell, J.W., et al., Assessment of imaging studies used with radiosurgery: a volumetric algorithm and an estimation of its error. Technical note. J Neurosurg, 2006. 104(1): p. 157–62.
54.
Zurück zum Zitat Mugler JP 3rd, Brookeman JR (1990) Three-dimensional magnetization-prepared rapid gradient-echo imaging (3D MP RAGE). Magn Reson Med 15(1):152–157PubMedCrossRef Mugler JP 3rd, Brookeman JR (1990) Three-dimensional magnetization-prepared rapid gradient-echo imaging (3D MP RAGE). Magn Reson Med 15(1):152–157PubMedCrossRef
55.
Zurück zum Zitat Danieli L et al (2019) Brain Tumor-Enhancement Visualization and Morphometric Assessment: A Comparison of MPRAGE, SPACE, and VIBE MRI Techniques. AJNR Am J Neuroradiol 40(7):1140–1148PubMedPubMedCentralCrossRef Danieli L et al (2019) Brain Tumor-Enhancement Visualization and Morphometric Assessment: A Comparison of MPRAGE, SPACE, and VIBE MRI Techniques. AJNR Am J Neuroradiol 40(7):1140–1148PubMedPubMedCentralCrossRef
56.
Zurück zum Zitat Chappell PM et al (1994) Comparison of lesion enhancement on spin-echo and gradient-echo images. AJNR Am J Neuroradiol 15(1):37–44PubMedPubMedCentral Chappell PM et al (1994) Comparison of lesion enhancement on spin-echo and gradient-echo images. AJNR Am J Neuroradiol 15(1):37–44PubMedPubMedCentral
57.
Zurück zum Zitat Reichert, M., et al., Contrast-enhanced 3‑dimensional SPACE versus MP-RAGE for the detection of brain metastases: considerations with a 32-channel head coil. Invest Radiol, 2013. 48(1): p. 55–60. Reichert, M., et al., Contrast-enhanced 3‑dimensional SPACE versus MP-RAGE for the detection of brain metastases: considerations with a 32-channel head coil. Invest Radiol, 2013. 48(1): p. 55–60.
58.
Zurück zum Zitat Komada T et al (2008) Contrast-enhanced MR imaging of metastatic brain tumor at 3 tesla: utility of T(1)-weighted SPACE compared with 2D spin echo and 3D gradient echo sequence. Magn Reson Med Sci 7(1):13–21PubMedCrossRef Komada T et al (2008) Contrast-enhanced MR imaging of metastatic brain tumor at 3 tesla: utility of T(1)-weighted SPACE compared with 2D spin echo and 3D gradient echo sequence. Magn Reson Med Sci 7(1):13–21PubMedCrossRef
59.
Zurück zum Zitat Welzel T et al (2022) Stereotactic radiotherapy of brain metastases: clinical impact of three-dimensional SPACE imaging for 3T-MRI-based treatment planning. Strahlenther Onkol 198(10):926–933PubMedPubMedCentralCrossRef Welzel T et al (2022) Stereotactic radiotherapy of brain metastases: clinical impact of three-dimensional SPACE imaging for 3T-MRI-based treatment planning. Strahlenther Onkol 198(10):926–933PubMedPubMedCentralCrossRef
60.
Zurück zum Zitat Mugler JP 3rd, Brookeman JR (1993) Theoretical analysis of gadopentetate dimeglumine enhancement in T1-weighted imaging of the brain: comparison of two-dimensional spin-echo and three-dimensional gradient-echo sequences. J Magn Reson Imaging 3(5):761–769PubMedCrossRef Mugler JP 3rd, Brookeman JR (1993) Theoretical analysis of gadopentetate dimeglumine enhancement in T1-weighted imaging of the brain: comparison of two-dimensional spin-echo and three-dimensional gradient-echo sequences. J Magn Reson Imaging 3(5):761–769PubMedCrossRef
62.
Zurück zum Zitat Wu X et al (2014) Contrast-enhanced radial 3D fat-suppressed T1-weighted gradient-recalled echo sequence versus conventional fat-suppressed contrast-enhanced T1-weighted studies of the head and neck. AJR Am J Roentgenol 203(4):883–889PubMedCrossRef Wu X et al (2014) Contrast-enhanced radial 3D fat-suppressed T1-weighted gradient-recalled echo sequence versus conventional fat-suppressed contrast-enhanced T1-weighted studies of the head and neck. AJR Am J Roentgenol 203(4):883–889PubMedCrossRef
63.
Zurück zum Zitat Garcia MA et al (2018) Brain metastasis growth on preradiosurgical magnetic resonance imaging. Pract Radiat Oncol 8(6):e369–e376PubMedCrossRef Garcia MA et al (2018) Brain metastasis growth on preradiosurgical magnetic resonance imaging. Pract Radiat Oncol 8(6):e369–e376PubMedCrossRef
64.
Zurück zum Zitat Kutuk T et al (2022) Dedicated isotropic 3‑D T1 SPACE sequence imaging for radiosurgery planning improves brain metastases detection and reduces the risk of intracranial relapse. Radiother Oncol 173:84–92PubMedCrossRef Kutuk T et al (2022) Dedicated isotropic 3‑D T1 SPACE sequence imaging for radiosurgery planning improves brain metastases detection and reduces the risk of intracranial relapse. Radiother Oncol 173:84–92PubMedCrossRef
65.
Zurück zum Zitat Naganawa S (2015) The Technical and Clinical Features of 3D-FLAIR in Neuroimaging. Magn Reson Med Sci 14(2):93–106PubMedCrossRef Naganawa S (2015) The Technical and Clinical Features of 3D-FLAIR in Neuroimaging. Magn Reson Med Sci 14(2):93–106PubMedCrossRef
66.
Zurück zum Zitat Jaju A et al (2022) Imaging of pediatric brain tumors: A COG Diagnostic Imaging Committee/SPR Oncology Committee/ASPNR White Paper. Pediatr Blood Cancer p:e30147 Jaju A et al (2022) Imaging of pediatric brain tumors: A COG Diagnostic Imaging Committee/SPR Oncology Committee/ASPNR White Paper. Pediatr Blood Cancer p:e30147
67.
Zurück zum Zitat Gürün E et al (2021) Evaluation of schwannoma using the 3D-SPACE sequence: comparison with the 3D-CISS sequence in 3T-MRI. Turk J Med Sci 51(3):1123–1135PubMedPubMedCentralCrossRef Gürün E et al (2021) Evaluation of schwannoma using the 3D-SPACE sequence: comparison with the 3D-CISS sequence in 3T-MRI. Turk J Med Sci 51(3):1123–1135PubMedPubMedCentralCrossRef
68.
Zurück zum Zitat Yang D et al (2000) Increased conspicuity of intraventricular lesions revealed by three-dimensional constructive interference in steady state sequences. AJNR Am J Neuroradiol 21(6):1070–1072PubMedPubMedCentral Yang D et al (2000) Increased conspicuity of intraventricular lesions revealed by three-dimensional constructive interference in steady state sequences. AJNR Am J Neuroradiol 21(6):1070–1072PubMedPubMedCentral
69.
Zurück zum Zitat Yousry I et al (2000) Visualization of cranial nerves I–XII: value of 3D CISS and T2-weighted FSE sequences. Eur Radiol 10(7):1061–1067PubMedCrossRef Yousry I et al (2000) Visualization of cranial nerves I–XII: value of 3D CISS and T2-weighted FSE sequences. Eur Radiol 10(7):1061–1067PubMedCrossRef
70.
Zurück zum Zitat Hessen, E.D., et al., Significant tumor shift in patients treated with stereotactic radiosurgery for brain metastasis. Clin Transl Radiat Oncol, 2017. 2: p. 23–28. Hessen, E.D., et al., Significant tumor shift in patients treated with stereotactic radiosurgery for brain metastasis. Clin Transl Radiat Oncol, 2017. 2: p. 23–28.
71.
Zurück zum Zitat Yuh WT et al (1995) The effect of contrast dose, imaging time, and lesion size in the MR detection of intracerebral metastasis. AJNR Am J Neuroradiol 16(2):373–380PubMedPubMedCentral Yuh WT et al (1995) The effect of contrast dose, imaging time, and lesion size in the MR detection of intracerebral metastasis. AJNR Am J Neuroradiol 16(2):373–380PubMedPubMedCentral
72.
Zurück zum Zitat Baleriaux D et al (2002) Magnetic resonance imaging of metastatic disease to the brain with gadobenate dimeglumine. Neuroradiology 44(3):191–203PubMedCrossRef Baleriaux D et al (2002) Magnetic resonance imaging of metastatic disease to the brain with gadobenate dimeglumine. Neuroradiology 44(3):191–203PubMedCrossRef
73.
Zurück zum Zitat Kushnirsky M et al (2016) Time-delayed contrast-enhanced MRI improves detection of brain metastases and apparent treatment volumes. J Neurosurg 124(2):489–495PubMedCrossRef Kushnirsky M et al (2016) Time-delayed contrast-enhanced MRI improves detection of brain metastases and apparent treatment volumes. J Neurosurg 124(2):489–495PubMedCrossRef
75.
Zurück zum Zitat Voigt R et al (2023) Kranielle MRT-Simulation für die Bestrahlungsplanung: Erfahrungen und Bildgebungsprotokolle nach 1000 dedizierten Planungs-MRT-Untersuchungen. Strahlenther Onkol 199(Suppl 1):8 Voigt R et al (2023) Kranielle MRT-Simulation für die Bestrahlungsplanung: Erfahrungen und Bildgebungsprotokolle nach 1000 dedizierten Planungs-MRT-Untersuchungen. Strahlenther Onkol 199(Suppl 1):8
76.
Zurück zum Zitat Bonneville F, Savatovsky J, Chiras J (2007) Imaging of cerebellopontine angle lesions: an update. Part 1: enhancing extra-axial lesions. Eur Radiol 17(10):2472–2482PubMedCrossRef Bonneville F, Savatovsky J, Chiras J (2007) Imaging of cerebellopontine angle lesions: an update. Part 1: enhancing extra-axial lesions. Eur Radiol 17(10):2472–2482PubMedCrossRef
77.
Zurück zum Zitat Cavallaro M et al (2022) Three-Dimensional Constructive Interference in Steady State (3D CISS) Imaging and Clinical Applications in Brain Pathology. Biomedicines 10(11) Cavallaro M et al (2022) Three-Dimensional Constructive Interference in Steady State (3D CISS) Imaging and Clinical Applications in Brain Pathology. Biomedicines 10(11)
78.
Zurück zum Zitat Tsukamoto T, Miki Y (2023) Imaging of pituitary tumors: an update with the 5th WHO Classifications—part 2. Neoplasms other than PitNET and tumor-mimicking lesions. Jpn J Radiol Tsukamoto T, Miki Y (2023) Imaging of pituitary tumors: an update with the 5th WHO Classifications—part 2. Neoplasms other than PitNET and tumor-mimicking lesions. Jpn J Radiol
79.
Zurück zum Zitat Cai J et al (2015) Diagnostic value of 3D time-of-flight MRA in trigeminal neuralgia. J Clin Neurosci 22(8):1343–1348PubMedCrossRef Cai J et al (2015) Diagnostic value of 3D time-of-flight MRA in trigeminal neuralgia. J Clin Neurosci 22(8):1343–1348PubMedCrossRef
80.
Zurück zum Zitat Bonne, N.-X., et al., Imaging of Jugular Paragangliomas, in Contemporary Management of Jugular Paraganglioma, G.B. Wanna, M.L. Carlson, and J.L. Netterville, Editors. 2018, Springer International Publishing: Cham. p. 49–62. Bonne, N.-X., et al., Imaging of Jugular Paragangliomas, in Contemporary Management of Jugular Paraganglioma, G.B. Wanna, M.L. Carlson, and J.L. Netterville, Editors. 2018, Springer International Publishing: Cham. p. 49–62.
81.
Zurück zum Zitat Weissmann, T., et al., Paragangliomas of the Head and Neck: Local Control and Functional Outcome Following Fractionated Stereotactic Radiotherapy. Am J Clin Oncol, 2019. 42(11): p. 818–823. Weissmann, T., et al., Paragangliomas of the Head and Neck: Local Control and Functional Outcome Following Fractionated Stereotactic Radiotherapy. Am J Clin Oncol, 2019. 42(11): p. 818–823.
82.
Zurück zum Zitat Glazebrook KN et al (2011) Imaging features of glomus tumors. Skelet Radiol 40(7):855–862CrossRef Glazebrook KN et al (2011) Imaging features of glomus tumors. Skelet Radiol 40(7):855–862CrossRef
83.
Zurück zum Zitat Loy, D.N., et al., Time-of-flight magnetic resonance angiography imaging of a residual arteriovenous malformation nidus after Onyx embolization for stereotactic radiosurgery planning. Technical note. Neurosurg Focus, 2009. 26(5): p. E13. Loy, D.N., et al., Time-of-flight magnetic resonance angiography imaging of a residual arteriovenous malformation nidus after Onyx embolization for stereotactic radiosurgery planning. Technical note. Neurosurg Focus, 2009. 26(5): p. E13.
84.
Zurück zum Zitat Geibprasert S et al (2010) Radiologic assessment of brain arteriovenous malformations: what clinicians need to know. Radiographics 30(2):483–501PubMedCrossRef Geibprasert S et al (2010) Radiologic assessment of brain arteriovenous malformations: what clinicians need to know. Radiographics 30(2):483–501PubMedCrossRef
85.
Zurück zum Zitat Nardone V et al (2019) Role of perilesional edema and tumor volume in the prognosis of non-small cell lung cancer (NSCLC) undergoing radiosurgery (SRS) for brain metastases. Strahlenther Onkol 195(8):734–744PubMedCrossRef Nardone V et al (2019) Role of perilesional edema and tumor volume in the prognosis of non-small cell lung cancer (NSCLC) undergoing radiosurgery (SRS) for brain metastases. Strahlenther Onkol 195(8):734–744PubMedCrossRef
86.
Zurück zum Zitat Andersen C, Astrup J, Gyldensted C (1994) Quantitative MR analysis of glucocorticoid effects on peritumoral edema associated with intracranial meningiomas and metastases. J Comput Assist Tomogr 18(4):509–518PubMedCrossRef Andersen C, Astrup J, Gyldensted C (1994) Quantitative MR analysis of glucocorticoid effects on peritumoral edema associated with intracranial meningiomas and metastases. J Comput Assist Tomogr 18(4):509–518PubMedCrossRef
87.
Zurück zum Zitat Scharl S et al (2019) Cavity volume changes after surgery of a brain metastasis-consequences for stereotactic radiation therapy. Strahlenther Onkol 195(3):207–217PubMedCrossRef Scharl S et al (2019) Cavity volume changes after surgery of a brain metastasis-consequences for stereotactic radiation therapy. Strahlenther Onkol 195(3):207–217PubMedCrossRef
88.
Zurück zum Zitat Kutuk T et al (2021) Impact of MRI timing on tumor volume and anatomic displacement for brain metastases undergoing stereotactic radiosurgery. Neurooncol Pract 8(6):674–683PubMedPubMedCentral Kutuk T et al (2021) Impact of MRI timing on tumor volume and anatomic displacement for brain metastases undergoing stereotactic radiosurgery. Neurooncol Pract 8(6):674–683PubMedPubMedCentral
89.
Zurück zum Zitat Hessen E et al (2019) Predicting and implications of target volume changes of brain metastases during fractionated stereotactic radiosurgery. Radiother Oncol Hessen E et al (2019) Predicting and implications of target volume changes of brain metastases during fractionated stereotactic radiosurgery. Radiother Oncol
90.
91.
Zurück zum Zitat Kawashima M et al (2022) Interfractional change of tumor volume during fractionated stereotactic radiotherapy using gamma knife for brain metastases. J Neurooncol 159(2):409–416PubMedCrossRef Kawashima M et al (2022) Interfractional change of tumor volume during fractionated stereotactic radiotherapy using gamma knife for brain metastases. J Neurooncol 159(2):409–416PubMedCrossRef
92.
Zurück zum Zitat Veninga T et al (2004) Clinical validation of the normalized mutual information method for registration of CT and MR images in radiotherapy of brain tumors. J Appl Clin Med Phys 5(3):66–79PubMedPubMedCentralCrossRef Veninga T et al (2004) Clinical validation of the normalized mutual information method for registration of CT and MR images in radiotherapy of brain tumors. J Appl Clin Med Phys 5(3):66–79PubMedPubMedCentralCrossRef
93.
Zurück zum Zitat Masitho S et al (2022) Accuracy of MRI-CT registration in brain stereotactic radiotherapy: Impact of MRI acquisition setup and registration method. Z Med Phys 32(4):477–487PubMedPubMedCentralCrossRef Masitho S et al (2022) Accuracy of MRI-CT registration in brain stereotactic radiotherapy: Impact of MRI acquisition setup and registration method. Z Med Phys 32(4):477–487PubMedPubMedCentralCrossRef
94.
Zurück zum Zitat Owrangi AM, Greer PB, Glide-Hurst CK (2018) MRI-only treatment planning: benefits and challenges. Phys Med Biol 63(5):5–tr1CrossRef Owrangi AM, Greer PB, Glide-Hurst CK (2018) MRI-only treatment planning: benefits and challenges. Phys Med Biol 63(5):5–tr1CrossRef
95.
Zurück zum Zitat Ulin K, Urie MM, Cherlow JM (2010) Results of a multi-institutional benchmark test for cranial CT/MR image registration. Int J Radiat Oncol Biol Phys 77(5):1584–1589PubMedPubMedCentralCrossRef Ulin K, Urie MM, Cherlow JM (2010) Results of a multi-institutional benchmark test for cranial CT/MR image registration. Int J Radiat Oncol Biol Phys 77(5):1584–1589PubMedPubMedCentralCrossRef
96.
Zurück zum Zitat Hanvey S, Glegg M, Foster J (2009) Magnetic resonance imaging for radiotherapy planning of brain cancer patients using immobilization and surface coils. Phys Med Biol 54(18):5381–5394PubMedCrossRef Hanvey S, Glegg M, Foster J (2009) Magnetic resonance imaging for radiotherapy planning of brain cancer patients using immobilization and surface coils. Phys Med Biol 54(18):5381–5394PubMedCrossRef
97.
Zurück zum Zitat Mandija S et al (2019) Brain and Head-and-Neck MRI in Immobilization Mask: A Practical Solution for MR-Only Radiotherapy. Front Oncol 9:647PubMedPubMedCentralCrossRef Mandija S et al (2019) Brain and Head-and-Neck MRI in Immobilization Mask: A Practical Solution for MR-Only Radiotherapy. Front Oncol 9:647PubMedPubMedCentralCrossRef
98.
Zurück zum Zitat Wong OL et al (2017) Image quality assessment of a 1.5T dedicated magnetic resonance-simulator for radiotherapy with a flexible radio frequency coil setting using the standard American College of Radiology magnetic resonance imaging phantom test. Quant Imaging Med Surg 7(2):205–214PubMedPubMedCentralCrossRef Wong OL et al (2017) Image quality assessment of a 1.5T dedicated magnetic resonance-simulator for radiotherapy with a flexible radio frequency coil setting using the standard American College of Radiology magnetic resonance imaging phantom test. Quant Imaging Med Surg 7(2):205–214PubMedPubMedCentralCrossRef
99.
Zurück zum Zitat Mengling V et al (2021) Implementation of a dedicated 1.5T MR scanner for radiotherapy treatment planning featuring a novel high-channel coil setup for brain imaging in treatment position. Strahlenther Onkol 197(3):246–256PubMedCrossRef Mengling V et al (2021) Implementation of a dedicated 1.5T MR scanner for radiotherapy treatment planning featuring a novel high-channel coil setup for brain imaging in treatment position. Strahlenther Onkol 197(3):246–256PubMedCrossRef
100.
Zurück zum Zitat Masitho, S., et al., Feasibility of artificial-intelligence-based synthetic computed tomography in a magnetic resonance-only radiotherapy workflow for brain radiotherapy: Two-way dose validation and 2D/2D kV-image-based positioning. Phys Imaging Radiat Oncol, 2022. 24: p. 111–117. Masitho, S., et al., Feasibility of artificial-intelligence-based synthetic computed tomography in a magnetic resonance-only radiotherapy workflow for brain radiotherapy: Two-way dose validation and 2D/2D kV-image-based positioning. Phys Imaging Radiat Oncol, 2022. 24: p. 111–117.
101.
Zurück zum Zitat Johnstone E et al (2018) Systematic Review of Synthetic Computed Tomography Generation Methodologies for Use in Magnetic Resonance Imaging-Only Radiation Therapy. Int J Radiat Oncol Biol Phys 100(1):199–217PubMedCrossRef Johnstone E et al (2018) Systematic Review of Synthetic Computed Tomography Generation Methodologies for Use in Magnetic Resonance Imaging-Only Radiation Therapy. Int J Radiat Oncol Biol Phys 100(1):199–217PubMedCrossRef
102.
Zurück zum Zitat Kazemifar S et al (2019) MRI-only brain radiotherapy: Assessing the dosimetric accuracy of synthetic CT images generated using a deep learning approach. Radiother Oncol 136:56–63PubMedCrossRef Kazemifar S et al (2019) MRI-only brain radiotherapy: Assessing the dosimetric accuracy of synthetic CT images generated using a deep learning approach. Radiother Oncol 136:56–63PubMedCrossRef
103.
Zurück zum Zitat Edmund JM et al (2015) Cone beam computed tomography guided treatment delivery and planning verification for magnetic resonance imaging only radiotherapy of the brain. Acta Oncol 54(9):1496–1500PubMedCrossRef Edmund JM et al (2015) Cone beam computed tomography guided treatment delivery and planning verification for magnetic resonance imaging only radiotherapy of the brain. Acta Oncol 54(9):1496–1500PubMedCrossRef
104.
Zurück zum Zitat Buhl SK et al (2010) Clinical evaluation of 3D/3D MRI-CBCT automatching on brain tumors for online patient setup verification—A step towards MRI-based treatment planning. Acta Oncol 49(7):1085–1091PubMedCrossRef Buhl SK et al (2010) Clinical evaluation of 3D/3D MRI-CBCT automatching on brain tumors for online patient setup verification—A step towards MRI-based treatment planning. Acta Oncol 49(7):1085–1091PubMedCrossRef
Metadaten
Titel
Quality requirements for MRI simulation in cranial stereotactic radiotherapy: a guideline from the German Taskforce “Imaging in Stereotactic Radiotherapy”
verfasst von
PD Dr. Florian Putz
Michael Bock
Daniela Schmitt
Christoph Bert
Oliver Blanck
Maximilian I. Ruge
Elke Hattingen
Christian P. Karger
Rainer Fietkau
Johanna Grigo
Manuel A. Schmidt
Tobias Bäuerle
Andrea Wittig
Publikationsdatum
02.01.2024
Verlag
Springer Berlin Heidelberg
Erschienen in
Strahlentherapie und Onkologie / Ausgabe 1/2024
Print ISSN: 0179-7158
Elektronische ISSN: 1439-099X
DOI
https://doi.org/10.1007/s00066-023-02183-6

Weitere Artikel der Ausgabe 1/2024

Strahlentherapie und Onkologie 1/2024 Zur Ausgabe

Alphablocker schützt vor Miktionsproblemen nach der Biopsie

16.05.2024 alpha-1-Rezeptorantagonisten Nachrichten

Nach einer Prostatabiopsie treten häufig Probleme beim Wasserlassen auf. Ob sich das durch den periinterventionellen Einsatz von Alphablockern verhindern lässt, haben australische Mediziner im Zuge einer Metaanalyse untersucht.

Antikörper-Wirkstoff-Konjugat hält solide Tumoren in Schach

16.05.2024 Zielgerichtete Therapie Nachrichten

Trastuzumab deruxtecan scheint auch jenseits von Lungenkrebs gut gegen solide Tumoren mit HER2-Mutationen zu wirken. Dafür sprechen die Daten einer offenen Pan-Tumor-Studie.

Mammakarzinom: Senken Statine das krebsbedingte Sterberisiko?

15.05.2024 Mammakarzinom Nachrichten

Frauen mit lokalem oder metastasiertem Brustkrebs, die Statine einnehmen, haben eine niedrigere krebsspezifische Mortalität als Patientinnen, die dies nicht tun, legen neue Daten aus den USA nahe.

Labor, CT-Anthropometrie zeigen Risiko für Pankreaskrebs

13.05.2024 Pankreaskarzinom Nachrichten

Gerade bei aggressiven Malignomen wie dem duktalen Adenokarzinom des Pankreas könnte Früherkennung die Therapiechancen verbessern. Noch jedoch klafft hier eine Lücke. Ein Studienteam hat einen Weg gesucht, sie zu schließen.

Update Onkologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.